PASCO CI-6559 User Manual

Instruction Sheet for the PASCO Model CI-6559
012-06437B
11/99 $1.00
HUMIDITY SENSOR
Humidity
Sensor
vented top
plate
aperture for
sensing element
%
R
H
U
%
E
L
A
M
T
I
ID
V
E
IT
Y
CI-6559
HUMIDITY
SENSOR
DIN connector
Introduction
The PASCO CI-6559 Humidity Sensor is designed to be used with the PASCO ScienceWorkshop® computer interface and DataStudio™ or ScienceWorkshop® soft­ware. The Humidity Sensor may be used for making single readings or continuous monitoring of relative humidity (RH).
Equipment Included:
• CI-6559 Humidity Sensor
• 6-foot cable with 8-pin DIN connectors
• instruction sheet
• storage bag
Additional Equipment Required:
• Any PASCO ScienceWorkshop®
computer
interface.
PASCO DataStudio™ software
Or PASCO ScienceWorkshop®
2.2.5 or higher.
software, version
to computer interface
interface cable with
to Humidity Sensor
DIN connectors
Specifications
Range: 0% - 100% RH Accuracy:
-no calibration ± 5% RH at 60% R H ± 8% at 90% RH
-with saturated salt calibration ± 2% 0% - 100% @25 °C
Resolution: 0.1% RH Response Rate: 15 seconds in slow moving air Repeatability: ± 0.5% RH Connector Type: 8-pin DIN, ScienceWorkshop
interface compatible
Voltage Output: 0 - 10V (0% - 100% RH)
DIN Connector
1: analog output (+), 0-10 V 2: analog output (-), signal ground 3: no connection 4: +5 VDC power 5: power ground 6: +12 VDC power 7: -12 VDC power 8: no-connection
© 1997 PASCO scientific Author: Steve Miller
6
1
4
7
8
3
5
2
HUMIDITY
RELATIVE
%
®
10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-9011 USA
Phone (916) 786-3800 • FAX (916) 786-8905 • email: techsupp@PASCO.com
better
ways to
teach science
Humidity Sensor
P
R
E
S
S
T
O
L
O
G
Interface
500
®
1
2
GAIN
=
1,10:ISOLATED
GAIN
=
1,10: REF TO GND
GAIN
=
1: REF TO GND
A
s
C
ON
ANALOG CHANNELS
B
n
DIGITAL CHANNELS
ScienceWorkshop
R E
L A
T I
V E
H U M
I D
I T Y
%
H
U
M
ID
IT
Y
S
E
N
S
O
R
CI-6559
%
012-06437B
Theory
The Humidity Sensor consists of two basic elements: the humidity sensing element and the signal condition amplifier.
The humidity sensing element is powered by +5 volts DC and outputs a DC voltage that is proportional to the relative humidity of the air surrounding the sens­ing element. The element is an integrated circuit, which features high reliability and fast response, and is contamination resistant. The output voltage of the sensing element varies between 0.8 and 3.9 VDC, which corresponds to a relative humidity ranging from 0 to 100%, respectively.
Two stages of amplification are provided to condition the signal from the humidity sensing element for input to the ScienceWorkshop computer interface. The output of these two stages will range from 0 to 10 VDC, which will correspond to 0 to 100% humidity.
The first stage is used to reference the sensor output range from 0 to 3.1 volts. This establishes 0 volts as the 0% relative humidity point. The second stage applies a gain of about 3.2, which causes the maxi­mum sensor output, 3.1VDC, to increase to 10VDC, which corresponds to a 100% relative humidity. Both DataStudio and ScienceWorkshop software take the 0 - 10V output from the sensor, multiply it by 10 and display the result as relative humidity.
The Humidity Sensor can be plugged directly into any ScienceWorkshop computer interface or connected to the interface box using the supplied cable with 8-pin DIN connectors.
Setup Procedure
1. Connect the Humidity Sensor and any analog channel into the PASCO Computer Interface with interface cable (Figure 1A),
or
Insert the DIN plug of the Humidity Sensor directly into any analog channel on the PASCO Computer Interface (Figure 1B).
2. Open the Experiment Setup window in DataStudio or ScienceWorkshop. Click and drag the analog plug icon to the analog channel icon that matches the analog port you are using for the Humidity Sensor.
AB
ANALOG CHANNELS
ScienceWorkshop
®
s
n
A
C
500
Interface
E
S
R
S
1
P
T
G
O
O
L
DIGITAL CHANNELS
B
2
ON
=
=
GAIN
GAIN
=
GAIN
1,10:ISOLATED
1,10: REF TO GND
1: REF TO GND
Plug into any
analog channel.
Plug into any
analog channel.
SENSOR
9
HUMIDITY
5
5
I-6
C
%
E Y V I T I T D
%
A
I L M E U R H
Figure 1. Connecting the Humidity Sensor into the computer interface.
3 . Select "Humidity Sensor" from the Sensors list
(DataStudio™) or drop-down menu (ScienceWorkshop®).
4. Open a display window, such as the graph display. In DataStudio, doubleclick the graph icon in the display list. In ScienceWorkshop, drag and drop the graph icon over the Humidity Sensor icon.
5. Additional display windows can be opened by fol­lowing the procedure in step 4.
Operational Notes
To prevent damage to the Humidity Sensor or other equipment, do not immerse the Humidity Sensor in fluid.
The sensing element resists contamination vapors, such as organic solvents, chlorine, and ammonia. The element is also unaffected by water condensation.
For best results: The Humidity Sensor is capable of responding fairly quickly to changes in the humidity. However, in order to respond, a sample of the environment (air) must reach the sensing element (positioned near the sensor aperture). The sensor responds quickest when it is in a slow moving air stream. This may be accomplished by moving the sensor slowly with your hand during data collection.
To avoid errors: The sensing element can be affected by the bright light of the sun. Do not point the aperture for the sensing element directly toward the sun or a bright light.
2
012-06437B
Humidity Sensor
Calibration of the Humidity Sensor
For most applications, calibration of the Humidity Sensor is not required.
If the sensor is to be calibrated with DataStudio or ScienceWorkshop, one of two methods may be used:
1) "single point" calibration method or 2) "two point" calibration method. (Each method is described, in detail, in the DataStudio online help manual and the ScienceWorkshop software user manual.)
Use of either method requires that the sensor be exposed to a sample of known relative humidity.
Single Point Calibration
The "single point" calibration method is fairly easy to implement. The "single point" calibration is a "two point" calibration (see below) where the output of the Humidity Sensor is assumed to be 0 volts at 0% RH.
To open the Calibration window, double-click on the Humidity Sensor icon in the Experiment Setup win­dow. The other point for calibration is gained by exposing the Humidity Sensor to an atmosphere of known relative humidity. When the humidity reading is relatively constant, enter the known % RH value in the High Value area in the Calibration window and click "Take Reading" (DataStudio) or "Read" (ScienceWorkshop). Click OK to close the Calibration window.
Two Point Calibration
The "two point" calibration is more difficult because the sensor must be exposed to two distinctively different atmospheres of different levels of known relative humidities to read high and low values for calibration. This difficulty may be overcome by using a device known as a saturated salt cell.
Constructing a Saturated Salt Cell
1. A one-liter glass or plastic-covered container may be used as a vessel for a cell.
2. Place distilled or deionized water approximately one centimeter deep into the vessel.
3. Add enough "salt" to create the desired saturated solution.
Hint: Excess "salt" should be present (visible as it settles to the bottom of the vessel) to ensure saturation of the solution.
A closed cell constructed with NaCl as the excess "salt" will have a RH of about 75%. A similar con­structed cell using MgCl will have a RH of about 33%.
The "single point" calibration method may also be accomplished with a saturated salt cell.
When performing the calibration, the Humidity Sensor should be placed in the saturated salt cell and allowed to equilibrate. This could take as long as 30 minutes or more.
Mounting on an Experimental Apparatus
Use the 1/4-20 screw connector located on the bottom of the sensor box to secure the Humidity Sensor to an experimental apparatus (Figure 2). The alignment hole fits over an alignment pin included on some PASCO apparatuses.
alignment hole
1/4-20 screw
connector
A saturated salt cell is an enclosed system containing a saturated salt solution and air. The air in the closed system will come to equilibrium at a particular relative humidity. The specific RH at which the system comes to equilibrium is based on the ability of the saturated salt solution to extract or add (through evaporation) water vapor from or to the air. This is a common tool for calibration of relative humidity instruments.
The high and low % RH readings should be entered in the Calibration window, as explained in the "Single Point Calibration" section.
Figure 2. Mounting connector and alignment hole.
3
Humidity Sensor
Copyright, Warranty, and Equipment Return
012-06437B
Please—Feel free to duplicate this manual subject to the copyright restric­tions below.
Copyright Notice
The PASCO scientific 012-06437A instruction sheet is copyrighted and all rights reserved. However, permis­sion is granted to non-profit educational institutions for reproduction of any part of the Humidity Sensor instruction sheet providing the reproductions are used only for their laboratories and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO scientific, is prohibited.
Limited Warranty
PASCO scientific warrants the product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace, at its option, any part of the product which is deemed to be defective in material or workmanship. The warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return of equipment for warranty repair belongs to the cus­tomer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equip­ment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment after repair will be paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO scientific for any reason, notify PASCO scientific by letter, phone, or fax BEFORE returning the product. Upon notification, the return authorization and shipping instructions will be promptly issued.
äÿ
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units must be packed properly. Carriers will not accept responsi­bility for damage caused by improper packing. To be certain the unit will not be damaged in shipment, observe the following rules:
The packing carton must be strong enough for the
item shipped.
Make certain there are at least two inches of
packing material between any point on the appara­tus and the inside walls of the carton.
Make certain that the packing material can not shift
in the box, or become compressed, allowing the instrument to come in contact with the edge of the packing carton.
Address: PASCO scientific
10101 Foothills Blvd. P.O. Box 619011
Roseville, CA 95678-9011 Phone: (916) 786-3800 FAX: (916) 786-8905 email: techsupp@pasco.com
4
Loading...