PASCO CI-6499 User Manual

Instruction Sheet for the PASCO Model CI-6499
012-08549A
Drop Counter
Label with light-emitting (LED) indicators
1/2” Probe Holders
1/4” Probe Holder
Drop
Set screws
detection opening
Introduction
The PASCO CI-6499 Drop Counter counts drops/ sample and displays the drop count (drops) and fluid volume (mL) of liquids and fine particles.
When the Drop Counter is used with a ScienceWorkshop® interface and DataStudio software, drop count and fluid volume can be viewed in real-time in a graph display.
Typical applications for the Drop Counter include measuring drop count and fluid volume, titration experiments, and particle counts.
Equipment Included:
• CI-6499 Drop Counter
• Syringe Assembly with two stopcocks
Plastic tips for buret (1 small, 1 medium)
Stainless steel rod
5’6” cable
Stereo plug
Plastic syringe
Additional Equipment Required:
ScienceWorkshop® interface (500 or 750)
• DataStudio® software (version 1.8.5 or later)
Specifications:
Sensor range: 0 to 40 drops/second Default sample rate: 5 samples/second Maximum sample rate: 10 samples/second Object detection size: 0.64 mm diameter (liquid drop or fine particle) Plastic enclosure: polypropylene, water resistant, chemically resistant to mild acids and bases Detector: Infrared photodetector with collimating lens
Probe holders: 2 x 1/2”, 1 x 1/4”
stopcocks
plastic tip
© 2003 PASCO scientific
Drop Counter 012-08549A
Drop Counter Detector Operation
The Drop Counter includes a photodetector, which records the change in infrared light emitting from the aperture. As the drop falls between the lenses, it penetrates a thin sheet of light. Light emitted from the infrared light source enters the aperture, travels through a collimating lens, and exits another aperture before reaching the photodetector. In Figure 1, the arrows indicate the direction of light travel.
infrared light source
drop
collimating lens
sheet of light
Figure 1: Detector Operation
When the detector recognizes a drop, the green light flashes. The light remains on when something is blocking its path. The Drop Counter blocks all visible light and works off of infrared light. On power up, the detector automatically adjusts the light level for the best sensitivity.
detector
focusing lens
thumbscrews for mounting up to three sensor probes. Typical sensors used are the pH Sensor (CI-6507A) and the Temperature Sensor (CI-6505B). A Teflon
®
sleeve is
recommended for use with these probes (part no. 669-075).
Drop Rate and Drop Size
For best results, use the included syringe with the Drop Counter. The syringe has two stopcocks. Use the top stopcock to select a flow rate during calibration and adjust the rate during the run. Use the bottom stopcock to start and stop the liquid flow. The tip provided with the Drop Counter helps to ensure a consistent drop size. The drops from a glass tip are commonly twice the size of drops from plastic tips, and the volume per drop for a glass tip is not as consistent.
Shaking the syringe during the titration can affect the drop size and the flow rate by causing small drops to fall. Avoid making unnecessary adjustments during the run. Exceptions include the following: the system needs to be refilled or the drop rate falls below 1 drop/second. If the drop rate is below 1 drop per second, the drop size increases (up to 2 - 3 times larger) and this causes inaccurate volumes. If the flow rate falls below 1 drop per second, and a large volume of liquid remains in the syringe, adjust the stopcock to increase the flow. If the liquid level is low, turn off the stopcock and refill the syringe before turning it on at the higher drop rate.
Sensor Usage Tips
• Avoid operating the Drop Counter in direct sunlight.
• The green LED will flash once per drop. Two flashes per drop indicates that a splash caused a count. If the LED stays lit, then the lenses need to be cleaned.
• If the lenses become dirty or wet, clean or dry with
a cotton swab or tissue. The lenses are located in the
slits in the rectangular opening. WARNING: Do not
use chemicals to clean the lenses. Abrasive
chemicals may damage the lenses!
• To avoid miscount errors from splashing and/or potential lens damage, use a magnetic stirrer.
• The Drop Counter has two different size holes with
It can take several seconds for each drop to mix, but the pH probe detects more concentrated fractions as often as less concentrated fractions. Therefore, the mixing tends to occur without systematic error.
CAUTION: If you use a buret, slowly turn the stopcock to allow drops of liquid to drip and avoid streams of liquid bursting from the valve early on in the count. Bursts of liquid will distort the actual count or volume dispensed.
The common 25 or 50 mL buret is long enough that the pressure variation from the bottom to the top of the buret will change the flow rate. With some burets, the rate can drop below 1 drop per second, which may cause drops to become larger and unpredictable. If using a glass buret, start at a rate of 3 to 4 drops/second. When the rate slows, stop, take a volume reading, refill the buret, and start again at the faster rate.
2
Loading...
+ 4 hidden pages