PASCO CI-6499 User Manual

Instruction Sheet for the PASCO Model CI-6499
012-08549A
Drop Counter
Label with light-emitting (LED) indicators
1/2” Probe Holders
1/4” Probe Holder
Drop
Set screws
detection opening
Introduction
The PASCO CI-6499 Drop Counter counts drops/ sample and displays the drop count (drops) and fluid volume (mL) of liquids and fine particles.
When the Drop Counter is used with a ScienceWorkshop® interface and DataStudio software, drop count and fluid volume can be viewed in real-time in a graph display.
Typical applications for the Drop Counter include measuring drop count and fluid volume, titration experiments, and particle counts.
Equipment Included:
• CI-6499 Drop Counter
• Syringe Assembly with two stopcocks
Plastic tips for buret (1 small, 1 medium)
Stainless steel rod
5’6” cable
Stereo plug
Plastic syringe
Additional Equipment Required:
ScienceWorkshop® interface (500 or 750)
• DataStudio® software (version 1.8.5 or later)
Specifications:
Sensor range: 0 to 40 drops/second Default sample rate: 5 samples/second Maximum sample rate: 10 samples/second Object detection size: 0.64 mm diameter (liquid drop or fine particle) Plastic enclosure: polypropylene, water resistant, chemically resistant to mild acids and bases Detector: Infrared photodetector with collimating lens
Probe holders: 2 x 1/2”, 1 x 1/4”
stopcocks
plastic tip
© 2003 PASCO scientific
Drop Counter 012-08549A
Drop Counter Detector Operation
The Drop Counter includes a photodetector, which records the change in infrared light emitting from the aperture. As the drop falls between the lenses, it penetrates a thin sheet of light. Light emitted from the infrared light source enters the aperture, travels through a collimating lens, and exits another aperture before reaching the photodetector. In Figure 1, the arrows indicate the direction of light travel.
infrared light source
drop
collimating lens
sheet of light
Figure 1: Detector Operation
When the detector recognizes a drop, the green light flashes. The light remains on when something is blocking its path. The Drop Counter blocks all visible light and works off of infrared light. On power up, the detector automatically adjusts the light level for the best sensitivity.
detector
focusing lens
thumbscrews for mounting up to three sensor probes. Typical sensors used are the pH Sensor (CI-6507A) and the Temperature Sensor (CI-6505B). A Teflon
®
sleeve is
recommended for use with these probes (part no. 669-075).
Drop Rate and Drop Size
For best results, use the included syringe with the Drop Counter. The syringe has two stopcocks. Use the top stopcock to select a flow rate during calibration and adjust the rate during the run. Use the bottom stopcock to start and stop the liquid flow. The tip provided with the Drop Counter helps to ensure a consistent drop size. The drops from a glass tip are commonly twice the size of drops from plastic tips, and the volume per drop for a glass tip is not as consistent.
Shaking the syringe during the titration can affect the drop size and the flow rate by causing small drops to fall. Avoid making unnecessary adjustments during the run. Exceptions include the following: the system needs to be refilled or the drop rate falls below 1 drop/second. If the drop rate is below 1 drop per second, the drop size increases (up to 2 - 3 times larger) and this causes inaccurate volumes. If the flow rate falls below 1 drop per second, and a large volume of liquid remains in the syringe, adjust the stopcock to increase the flow. If the liquid level is low, turn off the stopcock and refill the syringe before turning it on at the higher drop rate.
Sensor Usage Tips
• Avoid operating the Drop Counter in direct sunlight.
• The green LED will flash once per drop. Two flashes per drop indicates that a splash caused a count. If the LED stays lit, then the lenses need to be cleaned.
• If the lenses become dirty or wet, clean or dry with
a cotton swab or tissue. The lenses are located in the
slits in the rectangular opening. WARNING: Do not
use chemicals to clean the lenses. Abrasive
chemicals may damage the lenses!
• To avoid miscount errors from splashing and/or potential lens damage, use a magnetic stirrer.
• The Drop Counter has two different size holes with
It can take several seconds for each drop to mix, but the pH probe detects more concentrated fractions as often as less concentrated fractions. Therefore, the mixing tends to occur without systematic error.
CAUTION: If you use a buret, slowly turn the stopcock to allow drops of liquid to drip and avoid streams of liquid bursting from the valve early on in the count. Bursts of liquid will distort the actual count or volume dispensed.
The common 25 or 50 mL buret is long enough that the pressure variation from the bottom to the top of the buret will change the flow rate. With some burets, the rate can drop below 1 drop per second, which may cause drops to become larger and unpredictable. If using a glass buret, start at a rate of 3 to 4 drops/second. When the rate slows, stop, take a volume reading, refill the buret, and start again at the faster rate.
2
012-08549A Drop Counter
Drop Counter Setup
1. Using a right-angle clamp to horizontally mount the Drop Counter to a rod stand (Figure 2).
2. Use a three-fingered clamp to mount an empty buret or the included syringe apparatus above the rectangular hole in the Drop Counter.
3. Connect the Drop Counter to a ScienceWorkshop interface.
4. Place a 150 mL (or larger) beaker on the stirring apparatus, but underneath the Drop Counter. Insert a magnetic stirrer into the solution.
5. Insert the probes of any other sensors (i.e. pH Sensor or Temperature Sensor) into the holes; hold in place with the set screws on the Drop Counter.
Experiment Setup with DataStudio Software
1. Plug the Drop Counter into a digital channel on a ScienceWorkshop interface.
2. Open DataStudio and double click “Create Experiment.”
3. In the Experiment Setup window, scroll to the
Sensors list. Click and drag the Drop Counter icon to a digital channel on the picture of the interface (the same digital channel to which you have your sensor connected.)
4. In the Setup window, double click on the Drop Counter icon.
5. In the Measurement tab, place checks (or verify that checks exist) in the boxes next to the drop count and fluid volume. Click the OK button to accept the changes.
6. Follow the calibration procedure instructions.
Note: When using the Drop Counter with DataStudio,
DataStudio must be in automatic sampling mode (the default mode). The Drop Counter does not operate in manual sampling mode. To set DataStudio to automatic sampling, open the Sampling Options dialog (under Experiment>Set Sampling Options) and deselect all options checked in the Manual Sampling tab.
Figure 2: Drop Counter Setup
3
Drop Counter 012-08549A
Calibration Procedure
Note: DataStudio provides an indirect calibration procedure for calibrating the fluid volume per drop.
1. Follow the apparatus setup instructions under “Drop Counter Setup.”
2. Fill the buret/syringe with deionized water. Place a
waste beaker under the Drop Counter to catch any liquid from the buret.
3. Flush bubbles out of the area below the stopcocks by
opening the stopcock valves completely (handles parallel for the syringe) or tapping below the stopcock. When the bubbles disappear, close the stopcock and refill the syringe/buret.
4. If using the syringe: The top stopcock regulates flow
and the bottom stopcock turns the flow on or off. With the bottom stopcock completely open, adjust the top stopcock so that the flow rate is 2-4 drops/ second. To stop the flow, turn off the bottom stopcock (on/off valve).
Calibration Tips
To calibrate the Drop Counter with DataStudio, you need to measure the dropped fluid volume (from a syringe or buret) and enter the volume per drop in the Constant tab of the Drop Counter window.
-If using a 2-stopcock syringe, measure the fluid volume:
a) Take the mass of an empty, light cup. Record the mass to
0.01 precision. b) Start DataStudio and allow the drops to fall into a cup. c) At some point, stop the flow of drops and take the cup with fluid back and weigh it again to the same precision. d) Use the difference in mass to calculate the volume difference. If you liquid comes close to the density of water, which is 1 g/ mL, you can convert from grams to mL volume.
-If using a glass buret, record the volume change:
a) Take the volume before and after the calibration (using the
buret scale or a graduated cylinder).
b) Subtract the difference (mL) to determine the
actual volume.
If using a buret: Adjust the flow rate to 3-4 drops/ second. Important: Open the valve slowly to avoid rapid fluid bursts, which will affect your reading. (Please read “Drop Rate and Drop Size” on page 2.)
5. If you have not set up the experiment, follow the
“Experiment Setup with DataStudio Software” on page 3 of this manual.
6. From the Data list, drag the Drop Count icon to a Digits display.
7. Follow the directions under “Calibration Tips” to
collect drops and measure fluid volume.
8. The Digits display shows the number of drops
collected during the volume determination procedure. Divide the actual volume measured by the number of drops to find the volume per drop.
9. In the Setup window, double click on the Drop
Counter icon. Click the Constant tab. Enter the volume per drop and click the OK button.
Collecting Data with the Drop Counter
Note: Before collecting data, please read the “Sensor Usage Tips” on page 2 of this manual.
1. Open DataStudio and double click “Create Experiment.”
Note: When using the Drop Counter with DataStudio, DataStudio must be in automatic sampling mode (the default mode). The Drop Counter does not operate in manual sampling mode. To set DataStudio to automatic sampling, open the Sampling Options dialog (under Experiment>Set Sampling Options) and deselect all options checked in the Manual Sampling tab.
2. Calibrate the Drop Counter.
3. In DataStudio, create a display of Drop Count and a display of pH vs. fluid volume.
4. Click the Start button to prepare the Drop Counter
for data collection.
4
012-08549A Drop Counter
5. If using a syringe: With the top control valve at the same set position as in the calibration, turn the bottom On/Off valve to allow fluid to drip into the beaker. Watch the drop count and/or fluid volume display in real-time in DataStudio.
OR
If using a glass buret with one stopcock: Start very slowly to avoid liquid surges. After the drops begin flowing, slowly increase the rate to 3 to 4 drops per second.
Suggested Experiment
Equipment required: CI-6499 Drop Counter, buret
or syringe with 2 stopcocks, ScienceWorkshop interface, DataStudio software (version 1.8.5 or later), 0.1 M NaOH, 0.1 M HCI, beaker (150 mL), pH Sensor, buffered solution of pH 4.0 and 7.0, magnetic stirring rod, rod stand, rod, right-angle clamp, three-fingered clamp, mass balance (0.1 g resolution)
1. Set up the Drop Counter Apparatus (See page 3 for instructions.)
6. Insert the pH electrode through one of the holes on the Drop Counter. Ensure that the tip of the pH electrode is fully immersed in the solution. The drops from the buret/syringe should fall through the Drop Counter and into the beaker.
7. In DataStudio, create a graph of pH vs. Fluid Volume. Click the Start button to prepare the Drop Counter to read drops.
8. If using a syringe: Turn the bottom valve to the full “on” position and release the titrant.
Or
If using a buret: Slowly open the stopcock to avoid liquid surges. After drops begin flowing, slowly increase the rate to 3 or 4 drops per second. Observe the titration curve in DataStudio.
9. Watch the pH change. After the equivalence point, allow about 5 mL of titrant to drop until the curve flattens, then click the Stop button to end the data collection.
2. Open DataStudio and setup and calibrate the pH Sensor. (See the instructions provided with the sensor.)
3. Calibrate the Drop Counter in mL fluid/drop (See page 4) or open the saved DataStudio calibration file.
4. Fill the buret (or syringe) with 0.1M NaOH. Place a waste beaker under the Drop Counter. Turn the stopcock and release a few mL of NaOH to flush other liquid or bubbles from the stopcock.
5. Fill a 150 ml beaker with 100 ml of deionized water. Add 5 ml of 0.1M HCl to the beaker. (Optional: Add a few drops of phenophthalein indicator to the beaker.) Add a spin bar to the beaker and place the beaker on a stir plate (set the stir plate to medium speed).
Sample Data
The following is sample data from a titration experiment performed with the Drop Counter.
equivalence point
Figure 3
5
Drop Counter 012-08549A
Limited Warranty
PASCO scientific warrants the product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace, at its option, any part of the product which is deemed to be defective in material or workmanship. The warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return of equipment for warranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment after repair will be paid by PASCO scientific.
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: (916) 786-3800
FAX: (916) 786-8905
E-mail: techsupp@pasco.com
Web site: www.pasco.com
6
Loading...