Gravitational Torsion BalanceCopyright, Warranty, and Equipment Return
Copyright, Warranty, and Equipment Return
Please—Feel free to duplicate this manual subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific 012-11032C Gravitational Torsion Balance manual is copyrighted and all rights reserved.
However, permission is granted to non-profit educational institutions for reproduction of any part of the manual
providing the reproductions are used only for their laboratories and are not sold for profit. Reproduction under
any other circumstances, without the written consent of PASCO scientific, is prohibited.
Limited Warranty
PASCO scientific warrants the product to be free from defects in materials and workmanship for a period of one
year from the date of shipment to the customer. PASCO will repair or replace at its option any part of the product
which is deemed to be defective in material or workmanship. The warranty does not cover damage to the product
caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing
defect or improper use by the customer shall be made solely by PASCO scientific. Responsibility for the return
of equipment for warranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equipment for return
shipment will not be covered by the warranty.) Shipping costs for returning the equipment after repair will be
paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO scientific for any reason, notify PASCO scientific by letter,
phone, or fax BEFORE returning the product. Upon notification, the return authorization and shipping instructions will be promptly issued.
•NOTE: NO EQUIPMENT WILL BE ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units must be packed properly. Carriers will not accept responsibility
for damage caused by improper packing. To be certain the unit will not be damaged in shipment, observe the following rules:
•The packing carton must be strong enough for the item shipped.
•Make certain there are at least two inches of packing material between any point on the apparatus and the
inside walls of the carton.
•Make certain that the packing material cannot shift in the box or become compressed, allowing the
instrument come in contact with the packing carton.
Address:PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone:(916) 786-3800 (worldwide) or 800-772-8700 (US)
web:www.pasco.com
ii
®
Model No. AP-8215A Introduction
Base with leveling feet
Large
masses
Sight for
leveling
Grounding
wire
Head of
torsion ribbon
Zero adjust
knob
Mirror on
pendulum
bob
Figure 1: Assembled Gravitational
Torsion Balance, ready to begin Henry
Cavendish’s classic experiment to
determine the gravitational constant.
Newton’s Law of Universal Gravitation:
where m
1
and m2 are the masses of the
objects, r is the distance between their
centers, and G is the universal gravitational constant, 6.67 x 10
-11
Nm2/kg2.
FG
m
1m2
r
2
--------------
=
Introduction
The PASCO scientific AP-8215A Gravitational Torsion Balance
reprises one of the great experiments in the history of physics—the
measurement of the gravitational constant, as performed by Henry Cavendish in 1798.
The Gravitational Torsion Balance consists of two 38.3 gram masses
suspended from a highly sensitive torsion ribbon and two 1.5 kilogram
masses that can be positioned as required. The Gravitational Torsion
Balance is oriented so the force of gravity between the small balls and
the earth is negated (the pendulum is nearly perfectly aligned vertically
and horizontally). The large masses are brought near the smaller
masses, and the gravitational force between the large and small masses
is measured by observing the twist of the torsion ribbon.
An optical lever, produced by a laser light source and a mirror affixed to
the torsion pendulum, is used to accurately measure the small twist of
the ribbon. Three methods of measurement are possible: the final
deflection method, the equilibrium method, and the acceleration
method.
A Little Background
The gravitational attraction of all objects toward the Earth is obvious.
The gravitational attraction of every object to every other object, however, is anything but obvious. Despite the
lack of direct evidence for any such attraction between everyday objects, Isaac Newton was able to deduce his
law of universal gravitation.
However, in Newton's time, every measurable example of this gravitational force included the Earth as one of the masses. It was therefore
impossible to measure the constant, G, without first knowing the mass of
the Earth (or vice versa).
The answer to this problem came from Henry Cavendish in 1798, when he
performed experiments with a torsion balance, measuring the gravitational
attraction between relatively small objects in the laboratory. The value he
determined for G allowed the mass and density of the Earth to be determined. Cavendish's experiment was so well constructed that it was a hundred years before more accurate measurements were made.
012-11032C
1
®
Gravitational Torsion BalanceEquipment
Attach to
Ear
th
Ground.
G
R
A
V
IT
A
T
IO
N
A
L
TO
R
S
IO
N
B
A
L
A
N
C
E
AP-8215
1.5 kg Tungsten
masses
Adapter rings
Plastic demonstration
plate
Replacement
torsion ribbon
Zero adjust
knob
Torsion ribbon
head
Aluminum plate
Leveling feet
Pendulum
mirror
Optical
grade glass
window
Large mass
swivel support
Leveling sight
2-56 x 1/8 Phillips
head screws
Figure 2: Equipment included
Equipment
Included:
Gravitational Torsion BalancePlastic Plate
Support Base with Leveling FeetReplacement Torsion Ribbon*
1.5 kg Tungsten Balls (2)2-56 x 1/8 Phillips head screws (4)
Adapter Rings (2)Phillips screwdriver (not shown)
(*Model No. AP-8218)
Additional Required:
Laser light source (such as the PASCO OS-9171 Helium-Neon Laser)
Meter stick
2
012-11032C
®
Model No. AP-8215A Equipment Setup
IMPORTANT NOTES
1. The Gravitational Torsion Balance is a delicate instrument. We
recommend that you set it up in a
relatively secure area where it is
safe from accidents and from
those who don’t fully appreciate
delicate instruments.
2. The first time you set up the torsion balance, do so in a place
where you can leave it for at least
one day before attempting measurements, allowing time for the
slight elongation of the torsion ribbon that will occur initially.
3. Keep the pendulum bob secured
in the locking mechanisms at all
times, except while setting up and
conducting experiences.
A
t
t
a
c
h
t
o
E
a
r
t
h
G
r
o
u
n
d
.
GRA
VIT
ATIONAL
TORSIONB
ALANCE
A
P
-8
2
1
5
Figure 3: Removing a plate from
the chamber box
Aluminum plate
Pendulum
chamber
Pendulum
bob
Equipment Parameters
•Small tungsten balls
Mass: 38.3 g ±0.2 g (m
)
2
Radius: 8.19 mm
Distance from ball center to torsion axis: d = 50.0 mm
•Large tungsten balls
Mass: 1500 ±10 g (m
)
1
Radius: 27.6 mm
•Distance from the center of the large ball to the center of mass of
the small ball when the large ball is against the aluminum plate
and the small ball is in the center position within the case: b =
42.2 mm. (Note: Tolerances may vary depending on the accuracy
of the horizontal alignment of the pendulum.)
•Distance from the surface of the mirror to the outer surface of the
glass window: 11.4 mm
•Torsion Ribbon
Material: Beryllium Copper
Length: approximately 260 mm
Cross-section: 0.017 by 0.150 mm
Equipment Setup
Initial Setup
1.Place the support base on a flat, stable table that is located
such that the Gravitational Torsion Balance will be at least 5
meters away from a wall or screen.
Note: For best results, use a very sturdy table, such as an optics
table.
2.Carefully remove the Gravitational Torsion Balance from the
box, and secure it in the base.
3.Remove the front aluminum plate by removing the thumb-
Note: Save the packing foam, and reinstall it each time the Gravitational Torsion Balance is transported.
4.Fasten the clear plastic plate to the pendulum chamber with
Note: Do not touch the mirror on the pendulum.
screws (Figure 3), and carefully remove the packing foam
from the pendulum chamber.
the thumbscrews.
012-11032C
3
®
Gravitational Torsion BalanceEquipment Setup
Pendulum
bob must be
centered over
the mirror
Mirror
Pendulum
Torsion
ribbon
Torsion ribbon
head
Look through the
sight to view the
reflection of the
pendulum bob in
the mirror.
Figure 5: Using the leveling sight to level
the Gravitational Torsion Balance.
Side,
Cutaway
View
Attach to
Ear
th
Ground.
Figure 4: Lowering the locking mechanism
to release the pendulum bob arms
Figure 6: Adjusting the height of the pendulum bob
The bottom of the pendulum
bob should be flush with the
floor of the chamber.
a
b
Leveling the Gravitational Torsion Balance
1.Release the pendulum from the locking mechanism by unscrewing the locking screws on the case, lowering the locking mechanisms to their lowest positions (Figure 4).
2.Adjust the feet of the base until the pendulum is centered in the
leveling sight (Figure 5). (The base of the pendulum will appear
as a dark circle surrounded by a ring of light).
3.Orient the Gravitational Torsion Balance so the mirror on the pendulum bob faces a screen or wall that is at least 5 meters away.
Vertical Adjustment of the Pendulum
The base of the pendulum should be flush with the floor of the pendulum chamber. If it is not, adjust the height of
the pendulum:
1.Grasp the torsion ribbon head and loosen the
Phillips retaining screw (Figure 6a).
2.Adjust the height of the pendulum by moving
the torsion ribbon head up or down so the base
of the pendulum is flush with the floor of the
pendulum chamber (Figure 6b).
3.Tighten the retaining (Phillips head) screw.
Note: Vertical adjustment is only necessary at initial
setup and when you change the torsion ribbon or if
someone has loosened the retaining screw by mistake; it is not normally done during each experimental setup.
Grasp the torsion ribbon head
and loosen the Phillips screw.
4
012-11032C
®
Model No. AP-8215A Equipment Setup
Case plates
Small mass
The pendulum bob arm must
be centered rotationally
between the plates.
Figure 7: Aligning the pendulum bob
rotationally
Top, Cutaway View
L
Top View
Figure 8: Setting up the optical lever
Side View
Screen with
scale
Laser
Reflected beam
(from mirror)
0
Figure 9: Ideal rotational alignment
(zeroing) of the pendulum
Location of the
projected laser
beam from the
glass window
Location of the
projected laser
beam from the
mirror
Rotational Alignment of the Pendulum Bob Arms (Zeroing)
The pendulum bob arms must be centered rotationally in the case —
that is, equidistant from each side of the case (Figure 7). To adjust
them:
1.Mount a metric scale on the wall or other projection surface that
is at least 5 meters away from the mirror of the pendulum.
2.Replace the plastic cover with the aluminum cover.
3.Set up the laser so it will reflect from the mirror to the projec-
tion surface where you will take your measurements (approximately 5 meters from the mirror). You will need to point the
laser so that it is tilted upward toward the mirror and so the
reflected beam projects onto the projection surface (Figure 8).
There will also be a fainter beam projected off the surface of the
glass window.
L
4.Rotationally align the case by rotating it until the laser beam
projected from the glass window is centered on the metric scale
(Figure 9).
5.Rotationally align the pendulum arm:
a.Raise the locking mechanisms by turning the locking screws
until both of the locking mechanisms barely touch the pendulum arm. Maintain this position for a few moments until the
oscillating energy of the pendulum is dampened.
b.Carefully lower the locking mechanisms slightly so the pen-
dulum can swing freely. If necessary, repeat the dampening
exercise to calm any wild oscillations of the pendulum bob.
c.Observe the laser beam reflected from the mirror. In the optimally aligned system, the equilibrium point of
the oscillations of the beam reflected from the mirror will be vertically aligned below the beam reflected
from the glass surface of the case (Figure 9).
012-11032C
5
®
Gravitational Torsion BalanceEquipment Setup
Zero adjust knob
Zero adjust thumbscrew
Figure 10: Refining the rotational
alignment of the pendulum bob
Attach to
E
ar
th
G
round.
G
R
A
V
ITATIO
N
A
L
TOR
S
IO
N
B
A
L
A
N
C
E
A
P
-82
15
Grounding
screw
Copper wire to
earth ground
Light beam
Large Masses:
Position I
Large Masses:
Position II
Glass window
Mirror
Case
d.If the spots on the projection surface (the laser beam reflec-
tions) are not aligned vertically, loosen the zero adjust thumbscrew, turn the zero adjust knob slightly to refine the
rotational alignment of the pendulum bob arms (Figure 10),
and wait until the movement of the pendulum stops or nearly
stops.
e.Repeat steps 4a – 4c as necessary until the spots are aligned
vertically on the projection surface.
6.When the rotational alignment is complete, carefully tighten
the zero adjust thumbscrew, being careful to avoid jarring the
system.
Hints for speedier rotational alignments:
•Dampen any wild oscillations of the pendulum bob with the locking mechanisms, as described;
•Adjust the rotational alignment of the pendulum bob using small, smooth adjustments of the zero adjust
knob;
•Exercise patience and finesse in your movements.
Setting up for the Experiment
1.Take an accurate measurement of the distance from the mirror to
the zero point on the scale on the projection surface (L) (Figure 8).
(The distance from the mirror surface to the outside of the glass
window is 11.4 mm.)
Note: Avoid jarring the apparatus during this setup procedure.
2.Attach copper wire to the grounding screw (Figure 11), and ground
it to the earth.
3.Place the adapter rings on the support arm and place the large tungsten masses on the adapter rings, and rotate the arm to Position I
(Figure 12), taking care to avoid bumping the case with the masses.
4.Allow the pendulum to come to resting equilibrium.
You are now ready to make a measurement using one of three methods: the final deflection method, the equilibrium method, or the
acceleration method.
Note: The pendulum may require several hours to reach resting
equilibrium. To shorten the time required, dampen the oscillation of
the pendulum by smoothly raising the locking mechanisms up (by
turning the locking screws) until they just touch the crossbar, holding for several seconds until the oscillations are dampened, and then
carefully lowering the locking mechanisms slightly.
Figure 11: Attaching the ground strap
to the grounding screw
Small
mass
6
012-11032C
®
Model No. AP-8215A Measuring the Gravitational Constant
Note: 5% accuracy is possible in
Method I if the experiment is set up
on a sturdy table in an isolated location where it will not be disturbed by
vibration or air movement.
Note: 5% accuracy is possible in
Method II if the resting equilibrium
points are determined using a
graphical analysis program.
Large Masses:
Position I
Large Masses:
Position II
Figure 13: Origin of
variables b and d
FG
m
1m2
b
2
--------------
1.1=
grav
2Fd1.2=
Measuring the Gravitational Constant
Overview of the Experiment
The gravitational attraction between a 38.3 gram mass and a 1.5 kg mass when their centers are separated by a
dis-tance of approximately 42.2 mm (a situation similar to that of the Gravitational Torsion Balance) is about
-10
7 x 10
mass is more than two hundred million times this amount.
The enormous strength of the Earth's attraction for the small masses, in comparison with their attraction for the
large masses, is what originally made the measurement of the gravitational constant such a difficult task. The torsion balance (invented by Charles Coulomb) provides a means of negating the otherwise overwhelming effects
of the Earth's attraction in this experiment. It also provides a force delicate enough to counterbalance the tiny
gravitational force that exists between the large and small masses. This force is provided by twisting a very thin
beryllium copper ribbon.
The large masses are first arranged in Position I, as shown in Figure 12, and
the balance is allowed to come to equilibrium. The swivel support that holds
the large masses is then rotated, so the large masses are moved to Position
II, forcing the system into disequilibrium. The resulting oscillatory rotation
of the system is then observed by watching the movement of the light spot
on the scale, as the light beam is deflected by the mirror.
newtons. If this doesn’t seem like a small quantity to measure, consider that the weight of the small
Any of three methods can be used to determine the gravitational constant,
G, from the motion of the small masses. In Method I, the final deflection
method, the motion is allowed to come to resting equilibrium—a process
that requires several hours—and the result is accurate to within approximately 5%. In Method II, the equilibrium method, the experiment takes 90
minutes or more and produces an accuracy of approximately 5% when graphical analysis is used in the procedure. In Method III, the acceleration method, the motion is observed for only 5 minutes, and the result is accurate
to within approximately 15%.
With the large masses in Position I (Figure 13), the gravitational
attraction, F, between each small mass (m
large mass (m
where b is the distance between the centers of the two
masses.
The gravitational attraction between the two small masses and
their neighboring large masses produces a net torque (
the system:
where d is the length of the lever arm of the pendulum bob cross-
piece.
) is given by the law of universal gravitation:
1
) and its neighboring
2
) on
grav
012-11032C
b
d
7
®
Gravitational Torsion BalanceMETHOD I: Measurement by Final Deflection
T
20
60
S (cm)
Time (min)
s2
s1
Figure 14: Graph of Small Mass Oscillations
band
1.3–=
2dGm
1m2
b
2
-------------------------
=
G
b
2
2dm1m
2
---------------------
1.4=
S1
S2
ΔS
2
θ
L
Figure 15: Diagram of the experiment
showing the optical lever.s
Position I
Position II
22tan»
S
2L
-------
=
4
S
L
-------
=
S
4L
-------
15=
T
2
42I
-----------
16=
I2m2d
2
2
5
---
r
2
+
1.7=
Since the system is in equilibrium, the twisted torsion
band must be supplying an equal and opposite torque.
This torque (
) is equal to the torsion constant for the
band
band () times the angle through which it is twisted (),
or:
Combining equations 1.1, 1.2, and 1.3, and taking into
account that
grav
= –
band
, gives:
Rearranging this equation gives an expression for G:
To determine the values of and — the only unknowns in equation 1.4 — it is necessary to observe the oscillations of the small mass system when the equilibrium is disturbed. To disturb the equilibrium (from S
swivel support is rotated so the large masses are moved to Position II. The system will then oscillate until it
finally slows down and comes to rest at a new equilibrium position (S
) (Figure 14).
2
), the
1
At the new equilibrium position S
, the torsion wire will
2
still be twisted through an angle , but in the opposite
direction of its twist in Position I, so the total change in
angle is equal to 2. Taking into account that the angle is
also doubled upon reflection from the mirror (Figure 15):
S = S2 – S1,
or
The torsion constant can be determined by observing the
period (T) of the oscillations, and then using the equation:
where I is the moment of inertia of the small mass system.
The moment of inertia for the mirror and support system
for the small masses is negligibly small compared to that
of the masses themselves, so the total inertia can be
expressed as:
8
012-11032C
®
Model No. AP-8215A METHOD I: Measurement by Final Deflection
82m
2
d
2
2
5
---
r
2
+
T
2
--------------------
1.8=
G2Sb
2
d
2
2
5
---
r
2
+
T
2
m1Ld
--------------------
1.9=
Position I
Position II
Intermediate
position
Figure 16: Two-step process of moving the
large masses to reduce the time required to
stop oscillating.
Therefore:
Substituting equations 1.5 and 1.8 into equation 1.4 gives:
All the variables on the right side of equation 1.9 are known or measurable:
r = 9.55 mm
d = 50 mm
b = 42.2 mm
m
= 1.5 kg
1
L = (Measure as in step 1 of the setup.)
By measuring the total deflection of the light spot (S) and the period of oscillation (T), the value of G can therefore be determined.
Procedure
1.Once the steps for leveling, aligning, and setup have been
completed (with the large masses in Position I), allow the pendulum to stop oscillating.
2.Turn on the laser and observe the Position I end point of the
balance for several minutes to be sure the system is at equilibrium. Record the Position I end point (S
) as accurately as
1
possible, and indicate any variation over time as part of your
margin of error in the measurement.
3.Carefully rotate the swivel support so that the large masses are
moved to Position II. The spheres should be just touching the
case, but take care to avoid knocking the case and disturbing
the system.
Note: You can reduce the amount of time the pendulum requires to
move to equilibrium by moving the large masses in a two-step process: first move the large masses and support to an intermediate
position that is in the midpoint of the total arc (Figure 16), and
wait until the light beam has moved as far as it will go in the
period; then move the sphere across the second half of the arc until
the large mass support just touches the case. Use a slow, smooth
motion, and avoid hitting the case when moving the mass support.
L
S1
S2
4.Immediately after rotating the swivel support, observe the
light spot and record its position (S
).
1
012-11032C
9
®
Gravitational Torsion BalanceMETHOD I: Measurement by Final Deflection
5.Use a stop watch to determine the time required for one period of oscillation (T). For greater accuracy,
include several periods, and then find the average time required for one period of oscillation.
Note: The accuracy of this period value (T) is very important, since the T is squared in the calculation of G.
6.Wait until the oscillations stop, and record the resting equilibrium point (S
Analysis
1.Use your results and equation 1.9 to determine the value of G.
The value calculated in step 2 is subject to the following systematic error. The small sphere is attracted not only to its
neighboring large sphere, but also to the more distant large
sphere, though with a much smaller force. The geometry for
this second force is shown in Figure 17 (the vector arrows
shown are not proportional to the actual forces).
From Figure 17,
The force, F
lates, in this case, to:
is given by the gravitational law, which trans-
0
).
2
and has a component ƒ that is opposite to the direction of the force F:
This equation defines a dimensionless parameter,
Using the equation F = Gm
/b2, it can be determined that:
1m2
From Figure 17,
where F
is the value of the force acting on each small sphere from both large masses, and F is the force of
net
, that is equal to the ratio of the magnitude of ƒ to that of F.
= F - f = F - F = F(1 - )
F
net
attraction to the nearest large mass only.
Similarly,
G = G
where G is your experimentally determined value for the gravitational constant, and G
for the systematic error.
Finally, G
(1 - )
0
= G/(1 - )
0
is corrected to account
0
Use this equation with equation 1.9 to adjust your measured value.
10
012-11032C
®
Model No. AP-8215A METHOD II: Measurement by Equilibrium Positions
Note: To obtain an accuracy of 5%
with this method, it is important to
use graphical analysis of the position and time data to extrapolate the
resting equilibrium positions,
S
1
and S2.
60
S (cm)
Time (min)
S2
S1
0
20
Figure 18: Typical pendulum oscillation pattern
showing equilibrium positions.
METHOD II: Measurement by Equilibrium Positions
Observation Time: ~90+ minutes
Accuracy: ~5%
Theory
When the large masses are placed on the swivel support and moved to
either Position I or Position II, the torsion balance oscillates for a time
before coming to rest at a new equilibrium position. This oscillation can be
described by a damped sine wave with an offset, where the value of the offset represents the equilibrium point for the balance. By finding the equilibrium point for both Position I and Position II and taking the difference, the
value of S can be obtained. The remainder of the theory is identical to that
described in Method I.
Procedure
1.Set up the experiment following steps 1–3 of Method I.
2.Immediately after rotating the swivel support to Position II, observe the light spot. Record the position of the
light spot (S) and the time (t) every 15 seconds.Continue recording the position and time for about 45 minutes.
3.Rotate the swivel support to Position I. Repeat the procedure described in step 2.
Note: Although it is not imperative that step 3 be performed immediately after step 2, it is a good idea to proceed
with it as soon as possible in order to minimize the risk that the system will be disturbed between the two measurements. Waiting more than a day to perform step 3 is not advised.
Analysis
1.Construct a graph of light spot position versus time for
both Position I and Position II. You will now have a
graph similar to Figure 18.
2.Find the equilibrium point for each configuration by
analyzing the corresponding graphs using graphical
analysis to extrapolate the resting equilibrium points S
and S
(the equilibrium point will be the center line
2
about which the oscillation occurs). Find the difference
between the two equilibrium positions and record the
result as S.
1
3.Determine the period of the oscillations of the small
mass system by analyzing the two graphs. Each graph
will produce a slightly different result. Average these
results and record the answer as T.
4.Use your results and equation 1.9 to determine the value of G.
5.The value calculated in step 4 is subject to the same systematic error as described in Method I. Perform the
correction procedure described in that section (Analysis, step 3) to find the value of G
012-11032C
.
0
11
®
Gravitational Torsion BalanceMETHOD III: Measurement by Acceleration
FG
m
1m2
b
2
--------------
3.1=
F
total
2F2 G
m
1m2
b
2
--------------
3.2==
m2a02G
m
1m2
b
2
--------------
3.3=
Gb
2
a
0
2m
1
----------
3.4=
METHOD III: Measurement by Acceleration
Observation Time: ~ 5 minutes
Accuracy: ~ 15%
Theory
With the large masses in Position I, the gravitational attraction, F, between each small mass (m2) and its neighboring large mass (m
This force is balanced by a torque from the twisted torsion ribbon, so that the system is in equilibrium. The angle
of twist, , is measured by noting the position of the light spot where the reflected beam strikes the scale. This
position is carefully noted, and then the large masses are moved to Position II. The position change of the large
masses disturbs the equilibrium of the system, which will now oscillate until friction slows it down to a new
equilibrium position.
) is given by the law of universal gravitation:
1
Since the period of oscillation of the small masses is long (approximately 10 minutes), they do not move significantly when the large masses are first moved from Position I to Position II. Because of the symmetry of the
setup, the large masses exert the same gravitational force on the small masses as they did in Position I, but now in
the opposite direction. Since the equilibrating force from the torsion band has not changed, the total force (F
total
that is now acting to accelerate the small masses is equal to twice the original gravitational force from the large
masses, or:
Each small mass is therefore accelerated toward its neighboring large mass, with an initial acceleration (a
) that
0
is expressed in the equation:
Of course, as the small masses begin to move, the torsion ribbon becomes more and more relaxed so that the
force decreases and their acceleration is reduced. If the system is observed over a relatively long period of time,
as in Method I, it will be seen to oscillate. If, however, the acceleration of the small masses can be measured
before the torque from the torsion ribbon changes appreciably, equation 3.3 can be used to determine G. Given
the nature of the motion—damped harmonic—the initial acceleration is constant to within about 5% in the first
one tenth of an oscillation. Reasonably good results can therefore be obtained if the acceleration is measured in
the first minute after rearranging the large masses, and the following relationship is used:
)
12
012-11032C
®
Model No. AP-8215A METHOD III: Measurement by Acceleration
Figure 19: Source of data for
calculations in Method III
Ss
2L
d
------
3.5=
a
0
2s
t
2
---------
Sd
t
2
L
----------
3.6==
Best Fit Line
Curve
Through
Data
Figure 20: Sample data and best-fit line
The acceleration is measured by observing the displacement of the
light spot on the screen. If, as is shown in Figure 19:
s = the linear displacement of the small masses,
d = the distance from the center of mass of the small masses to
the axis of rotation of the torsion balance,
S= the displacement of the light spot on the screen, and
L= the distance of the scale from the mirror of the balance,
then, taking into account the doubling of the angle on reflection,
Using the equation of motion for an object with a constant acceleration (x = 1/2 at
2
), the acceleration can be calculated:
S1
S2
ΔS
2
L
2θ
By monitoring the motion of the light spot over time, the acceleration can be determined using equation 3.6, and the gravitational
constant can then be determined using equation 3.4.
Procedure
1.Begin the experiment by completing steps 1–3 of the procedure detailed in Method I.
2.Immediately after rotating the swivel support, observe the light spot. Record the position of the light spot (S)
and the time (t) every 15 seconds for about two minutes.
Analysis
1.Construct a graph of light spot displacement (S = S -
S
) versus time squared (t2), with t2 on the horizontal
1
axis (Figure 20). Draw a best-fit line through the
observed data points over the first minute of observation.
2.Determine the slope of your best-fit line.
3.Use equations 3.4 and 3.6 to determine the gravita-
tional constant.
4.The value calculated in step 3 is subject to a systematic error. The small sphere is attracted not only to its
neighboring large sphere, but also to the more distant
large sphere, although with a much smaller force. Use
the procedure detailed in Method I (Analysis, step 3)
to correct for this force.
9
8
7
6
5
Δ S
4
3
2
1
0
90036005625810011025
225 2025
t2 (sec2)
012-11032C
13
®
Gravitational Torsion BalanceMaintenance
Broken
torsion
ribbon
1. Turn locking screws until the
locking mechanism anchors the
pendulum arms.
Locking
mechanism
3. Loosen the
Phillips screw.
4. Loosen the
Phillips screw.
5. Grasp the torsion ribbon head
and remove the top portion of the
broken ribbon assembly.
b
Figure 21: Securing the pendulum bob
before removing a broken torsion ribbon,
and loosening the torsion ribbon head.
Zero adjust
knob
Torsion
ribbon
The copper
disk must
contact the
torsion ribbon
head.
Ribbon tab
Torsion ribbon
head
Attach the tab
with the Phillips
screw.
Figure 22: Attaching the top tab of the torsion ribbon
assembly to the torsion ribbon head.
Note: Be sure the ribbon is not twisted.
Maintenance
Replacing the Torsion Ribbon Assembly
If the torsion ribbon breaks, replace it as follows:
1.Remove the plates, and raise the locking mechanism using
the locking screws until the pendulum arms are securely
anchored (Figure 21a).
2.Grasp the pendulum bob near the bottom ribbon tab to stabilize it.
3.Loosen the Phillips screw on the bottom tab of the torsion
ribbon assembly (Figure 21a), and remove the bottom half
of the broken ribbon assembly.
4.Loosen the Phillips screw at the top of the balance assembly (Figure 21b).
5.Grasp the torsion ribbon head and remove the top portion
6.Attach the top tab of the new torsion ribbon to the torsion
of the broken torsion ribbon assembly.
ribbon head using the Phillips screw, being sure the copper disc on the tab is in contact with the torsion ribbon
head (Figure 22). Align the tab with the face of the torsion
ribbon head.
2. Grasp the pendulum
bob here to stabilize it.
a
7.Thread the ribbon through the shaft.
8.Using the zero adjust knob, align the bottom tab with the
face of the pendulum bob.
9.Tighten the Phillips screw on the top of the balance to secure the torsion ribbon head.
10. Attach the bottom tab of the ribbon to the pendulum bob using the Phillips screw.
11. Replace the back plate.
12. Level and align the pendulum according to the instructions in the Equipment Setup section of this manual.
14
012-11032C
®
Model No. AP-8215A Transporting and Storing
Transporting and Storing
1.To prepare the Gravitational Torsion Balance for transporting or storing:
a.Remove the front plate.
b.Raise the locking mechanism to securely anchor the pendulum bob.
c.Check to be sure that the torsion ribbon is hanging straight down the center of the tube. If it is not, lower
the locking mechanisms, be sure the torsion wire is centered, and raise the locking mechanisms again.
Repeat as necessary until the ribbon is centered in the tube.
d.Reinstall the packing foam into the chamber to secure the pendulum bob.
e.Replace the plate.
2.The Gravitational Torsion Balance may be stored flat in its shipping container.
3.Store in a cool, dry place, and protect the device from any jarring or rough handling.
Safety Precaution
The large masses are made of tungsten, which is not toxic. Be careful not to drop the 1.5 kg masses.
Technical Support
Feedback
If you have any comments about the product or manual, please let us know. If you have any suggestions on alternate experiments or find a problem in the manual, please tell us. PASCO appreciates any customer feedback.
Your input helps us evaluate and improve our product.
Contacting Technical Support
Before you call PASCO technical support, have the apparatus and this user's guide available. Please note the following:
•Product name and model number (e.g., Gravitational Torsion Balance, AP-8215A)
•Approximate age of the product;
•Detailed description of the problem/sequence of events required to duplicate the problem.
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone:+1 916 462 8384 (worldwide)
877-373-0300 (U.S.)
Web:www.pasco.com
Email:support@pasco.com
For the latest information about this product or the latest revision of the Instruction Manual, visit the PASCO web site
and enter AP-8215A in the Search window.
012-11032C
15
®
Gravitational Torsion BalanceTechnical Support
16
012-11032C
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.