Precautions
READ this manual BEFORE operating or servicing this equipment.
FOLLOW these instructions carefully.
SAVE this manual for future reference.
DO NOT allow untrained personnel to operate, clean, inspect, service or tamper with
this equipment.
ALWAYS DISCONNECT this equipment from the power source before cleaning or
performing maintenance.
Contact your local Parker Division for parts, information and service.
DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING,
SERVICING, CLEANING OR REMOVING THE FUSE. FAILURE TO DO
SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC
SENSITIVE DEVICES.
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS
EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TEST AND
ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO
OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM.
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD, CONNECT
TO PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE
GROUND PRONG.
WARNING
CAUTION
WARNING
WARNING
USE ONLY CLASS 2 LISTED EXTERNAL 24 VDC POWER SUPPLY WITH
THESE DEVICES.
4
CAUTION
Précautions
LISEZ ce manual AVANT de faire fonctionner ou d’entretenir cet équipment.
SUIVEZ attentivement ces instructions.
CONSERVEZ ce manuel pour future référence.
NE LAISSEZ PAS du personnel non qualifié utiliser, nettoyer, inspecter, entretenir, réparer ou
manipuler cet équipement.
DÉBRANCHEZ TOUJOURS cet équipement de la source de courant avant de nettoyer ou
d’exécuter l’entretien.
APPELEZ PARKER pour pièces détachées, renseignements et entretien.
DÉBRANCHEZ TOUT COURANT DE CETTE UNITÉ AVANT DE FAIRE
L’INSTALLATION, D’EFFECTUER L’ENTRETIEN, LE NETTOYAGE OU
AVANT DE RETIRER LE FUSIBLE. NE PAS OBSERVER CES
PRÉCAUTIONS RISQUERAIT DE CAUSER DES BLESSURES
CORPORELLES OU/ET D’ENDOMMAGER L’ÉQUIPEMENT.
SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES APPAREILS
SENSIBLES À L’ÉLECTROSTATIQUE.
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS
EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TEST AND
ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO
OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM.
POUR ASSURER UNE PROTECTION CONTINUE CONTRE UNE
DÉCHARGE ÉLECTRIQUE, BRANCHEZ UNIQUEMENT SUR UNE PRISE
CORRECTEMENT RELIÉE Á LA TERRE. NE RETIREZ PAS LA FICHE DE
TERRE.
ATTENTION
PRUDENCE
ATTENTION
ATTENTION
PRUDENCE
N'UTILISER QUE DE L'ALIMENTATION ÉLECTRIQUE 24 VDC
EXCLUSIVE DE CLASSE 2 AVEC CES DISPOSITIFS.
5
Introduction
You will find the SciLog® DINCon, DINPres and DINTemp Modules and sensors easy to use. The stateof-the-art hardware and software design of the SciLog
to measure and document the conductivity, pressure or temperature of many filtration / separation
processes. With proper maintenance, the SciLog
service and performance.
Please read the following instructions carefully!
Inspections: Remove the products carefully from the shipping container. Check the contents against
the purchase order to verify that all parts are included and undamaged.
Please do the inspection now, even if the products are not used immediately. Many carriers must
receive damage claims within seven days of delivery. Please retain all packing material so unit may be
shipped safely, if necessary.
Customer Service: Parker Bioscience customer service: If assistance is required, please contact us at:
Serial number and model name of the equipment
Installation procedure being used
Concise list of symptoms
List of operating procedures and conditions in use when problem arose
Calibration
The SciLog® DIN Modules are calibrated devices, and are calibrated with test equipment that is
traceable to SI through NIST.
A Calibration Certificate is included with each unit, and has a one year expiration date.
It is highly recommended that the Module be calibrated on an annual basis or more often if your
Metrology Department deems it necessary.
Parker Bioscience provides this Calibration Service, complete with a new certificate, showing the “as
found” and “as left” data.
Contact your local Parker Bioscience representative.
6
Warranty
Country specific information can be found at: www.parker.com/termsandconditions
Standards
EN 61326-1:2006, Class B
EN 61010-1:2010 Ed.3+ C1; C2
Conforms to UL STD 61010-1:2012 Ed.3+ R: 29Apr2016
Certified to CSA STD C22.2 No 61010-1-12:2012 Ed.3+U1;U2
Caution: The maximum recommended pressure for the sensors is 60 psi.
If this is exceeded, problems with leakage and functionality can occur.
Installation & Start Up
Installation of the SciLog® DINCon® DINPres®, and DINTemp® Modules must be carried out only by
trained personnel in accordance with the relevant regulations and this operations manual.
Make sure that the technical specifications and input ratings of the SciLog
the appropriate Sensor Specifications page.
The protection provided by this equipment may be impaired if the SciLog
manner inconsistent with this manual or for purposes not specified by the manufacturer.
®
sensors are observed. See
®
sensors are used in a
Maintenance & Cleaning
The SciLog® DIN Modules are maintenance free. The single-use SciLog® sensors come pre-calibrated
from the factory and require no maintenance.
To remove dust, dirt and stains, the outer surfaces of the DIN Modules may be wiped using a soft, nonfluffing cloth moistened with water. If required, you may also use a mild detergent or 2-propanol.
The single-use sensors may be sanitized with 0.1 Molar NaOH, or 2-propanol. They may be autoclaved
up to 2 times, or gamma-irradiated with exposure up to 45 kGy.
8 Pin Sensor Connection, DINCon has 1, DINPres and
DINTemp have 3.
Power Status LED, Sensor Status LED’s
Environmental
8
Software
Modbus
Temperature
Altitude
IP Rating
Rel. Humidity
Voltage
Pollution
Modbus RTU via RS-232
Modbus RTU via RS-485
Modbus RTU via USB
Modbus TCP via Ethernet
4 – 60° C
Up to 2000 Meters
IP20, Indoor dry environments, unit is wipe-down only
0 – 95%
Fluctuation allowed: +/- 10%
Degree: 2
Specifications – Sensors
SciCon:
Power
Fluid Connections
Wetted Materials
®
5 VDC provided by DINCon
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Module
SciPres:
Sensor Type
Conductivity Range
Accuracy
Temperature Range
Temperature Accuracy
Temperature Probe
Pre-Calibration
Sensor Microchip
Sensor Connector & Cables
Power
4-Electrode Conductivity Cell, gold plated pins.
1 μS/cm to 200 mS/cm; Resolution: 0.1 μS/cm.
High: ± 0.25 mS in the 0.1 - 200 mS range;
Low: ± 3 μS in the 0 - 100 μS range
4-50° C
+/- 0.5° C
Thermistor
0.10 M KCI Solution, 12.88 mS at 25.0° C
EPROM, Stores Sensor ID, and
Calibration Factor.
Lockable and Waterproof.
5 VDC provided by DINPres
®
Module
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
Fluid Connections
Wetted Materials
Sensor Type
Sensor Isolation
Pressure Range
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Medical grade, silicone piezoresistive sensing
element with on-chip temperature compensation.
Insoluble silicone dielectric gel isolates sensing
element from process solution. The gel is a non-
toxic, non-allergenic elastomeric system.
-5 to 60 psi (-0.34 to 4.14 bar).
9
+/- 0.30 psi at -5 to 30 psi for 2 point calibrated
Accuracy
sensors (can be improved to -5 to 60 psi with 3
point calibration, also available)
Sensor Connector & Cables
SciTemp:
Resolution
Temperature Range
Sensor Microchip
Power
Fluid Connections
Wetted Materials
Sensor Type
Temperature Range
0.01 psi
0 to 60 °C (32 to 140 °F).
EPROM, Stores Sensor ID, and
Calibration Factor.
Lockable and Waterproof.
5 VDC provided by DINTemp
®
Module
Choice of five sizes: Luer, 3/8” Barb, 1/2” Barb,
3/4” Tri-Clover (TC) Sanitary and 1” Tri-Clover
(‘Ladish’) Sanitary.
Polysulphone meets USP Class VI and FDA 21
CFR 177.1520, all wetted materials made with
animal-free compounds.
Thermistor, Epoxy coated, 2252 Ω
-40 to +150° Celsius
Accuracy
Resolution
Sensor Microchip
Sensor Connector & Cables
+/- 0.10° Celsius in the 4.0 to 70° Celsius range.
0.01 ° Celsius
EPROM, Stores Sensor ID, and
Calibration Factor.
Lockable and Waterproof.
10
Installation:
Overview:
The DINCon, DINPres, and DINTemp Modules are designed to be installed as part of a
control system inside a control cabinet that has a 24 VDC source available and has
Modbus communication protocol capability.
The DIN Modules are installed on standard DIN rail. They are usually mounted with the
red locking tab located at the bottom of the unit. This places the Power and
Communication connections at the top of the Module and the Sensor connections at the
bottom.
Wiring Power and Communication:
Connect 24 Volts DC, 250 mA power to Pins 1 (+) and 2 (-) of the 8 pin terminal block.
(Note - pin 1 is at the rear.)
All data acquisition is via Modbus registers in either RTU or ASCII. The available
physical layers are RS-232, RS-485, Ethernet and USB. The recommended physical
layer for installation of multiple units is Modbus/RTU over RS-485 with Modbus/TCP via
Ethernet a close second.
For RS-232, connect to the following: Pin 4 = Signal Ground, Pin 5 = RX, Pin 6 = TX.
For RS-485, connect to the following: Pin 4 = Signal Ground, Pin 7 = B, Pin 8 = A.
For Ethernet connect an Ethernet cable to the provided RJ45 jack.
For USB connect to the provided USB Type A jack.
11
A
Pin 1: +24VDC
Pin 2: Common
Pin 3: Not
Pin 4: Signal
Pin 5: RS-232 RX
Pin 6: RS-232 TX
Pin 7: RS-485
Pin 8: RS-485 B
RJ45 Ethernet
USB
Connecting Sensors:
All Sensors, SciCon for use with the DINCon Module, SciPres for use with the DINPres
Module and SciTemp for use with the DINTemp Module are disposable flow through
sensors that are purchased separately from Parker and come Pre-Calibrated.
The Sensors connect to a provided cable that has a watertight connector on the sensor
end and an 8 pin connector on the other end that plugs into a matching port on the
bottom of the Module. The DINCon has only one port, and uses only one SciCon
Sensor. The DINPres and DINTemp have 3 ports, can support up to 3 SciPres or
SciTemp Sensors, and they are referred to as P1 – P3 or T1 – T3 beginning with P1
and T1 at the rear, or the closest connector to the DIN rail.
12
DIN Rail locking clip
Sensor 1
Sensor 2
Sensor 3
When powered up and connected the Sensor Status LEDs on the front of the Module
are colored as follows:
Green = Connected
Red = Disconnected
Yellow = Connected but not calibrated.
If the LED shows Yellow, it’s time to replace the affected Sensor with a new one.
Connecting the wrong type of Sensor to the module also results in a Yellow indication.
13