LIGHT-TRANSMITTING ULTRASONIC INSERTS
NOT RETURNABLE IF OPENED.
If necessary to return insert, please use a shipping box,
not an envelope or padded mailer.
IMPORTANT:
Parkell’s TurboVue™ Light-Transmitting Ultrasonic
Inserts are designed for
operation in Parkell’s TurboVue™ Ultrasonic Scaler. They will provide the
benet of a bright, white light to the operating eld when
used with Parkell TurboVue™
Scalers, but will function as
standard scaling inserts when used with other scaling units.
PRECAUTIONS:
These inserts utilize a
durable, autoclavable glass
sleeve to transmit the light
Rx ONLY
CAUTION: NON-STERILE.
STERILIZE BEFORE USE
This manual, in whole or in
part, should not be
considered a substitute for
formal training in Ultrasonic
Scaling. Appropriate
professional education is
STRONGLY
RECOMMENDED prior to
using this device clinically.
from the handpiece to the
operating eld. While
breakage is highly unlikely
under normal use, abuse
of the insert may create a
safety hazard. Inspect the
insert and the light guide for any visible damage before each use.
Do not bend or reshape
insert. Inserts should be
handled with care at all times. Ultrasonic insert tips
that have been bent,
reshaped or otherwise damaged are susceptible to in-use breakage and should be discarded and replaced immediately.
Make sure the insert’s green O-ring is in place before use, and lubricate
with water prior to inserting into the handpiece. Replace worn or damaged
O-rings with Parkell Stock No. DORING-GR.
During scaling procedures,
all intra-oral soft tissues should be protected to avoid contact with the vibrating tip. To avoid
gouging the tooth, keep the
tip parallel to (not perpendicular to) the
tooth’s long axis. Use a controlled wiping motion,
with copious water irrigation.
While Parkell “Universal” tips and “Burnett Power-Tips™” may be
used at any power setting
from “low” to “high”, thin “Perio-style” tips should
not be used above your
scaler’s “medium” power
setting.
CLEANING AND STERILIZATION:
Inserts must be carefully scrubbed to remove debris,
either by hand with a brush or in an ultrasonic cleaner. Use non-ammoniated detergent or dishwashing
soap, followed by a cold
water rinse. Do not use ammoniated cleansers or disinfectants.
Steam Autoclaving following manufacturer’s instructions is the preferred
sterilization technique, as it
minimizes degradation of the insert’s materials.
“Chemical Vapor”
sterilization may also be
used. “Dry Heat” or “Cold Sterilization” are not recommended, as they may not be as effective, and may cause the O-ring
to deteriorate.
SUGGESTED CYCLES :
Gravity Displacement:
15 minutes @ 132-135º C; 15 minutes cool down.
Prevacuum:
4 minutes @ 132-135º C; 15 minutes cool down.
WARRANTY:
Parkell ultrasonic inserts
are warranted for 90 days
from date of delivery. This warranty does not cover
insert abuse, acts of God, or tips that have been bent, reground, dropped, or
altered in some other way. Parkell reserves the right to replace or return the purchase price for defective inserts at its discretion. Parkell cannot
assume liability for incidental or consequential damage resulting from the use of the insert.
Parkell’s Quality System is
certied to ISO 13485 and ISO 9001.
European Authorized Representative (Not a dealer/distributor):
Directa AB - P.O. Box 723, Finvids väg 8 SE-194 27
Upplands Väsby - Sweden
300 Executive Drive, Edgewood, NY 11717 Toll-Free: 1-800-243-7446 Phone: 631-249-1134 Fax: 631-249-1242
E-mail: info@parkell.com Parkell.com
A00768revB1217