INCLUDES INSTRUCTIONS FOR THE PARI LC® SPRINT NEBULIZER
IMPORTANT: READ INSTRUCTIONS AND SAFETY PRECAUTIONS BEFORE USE
Instructional Video Available Online
Scan this code with your smartphone or tablet to view the
Trek® S Instructional Video, or simply visit
www.PARI.com/instructional-videos online.
READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE. USE THE PRODUCT ONLY FOR ITS INTENDED USE
AS DESCRIBED IN THESE INSTRUCTIONS FOR USE. IMPROPER USE CAN RESULT IN SERIOUS OR FATAL ILLNESS/
INJURY, IMPROPER TREATMENT OR PROPERTY DAMAGE.
This booklet is part
of a greater PARI
initiative to reduce
our carbon footprint.
REMEMBER
REPLACE YOUR REUSABLE NEBULIZER CUP EVERY SIX MONTHS
INDICATIONS FOR USE: LC® SPRINT
The LC® Sprint is a handheld nebulizer, designed to
aerosolize medication approved for nebulization and
prescribed by a physician. The LC
adult and pediatric patients consistent with the indications
for the aerosol medication.
®
Sprint is intended for
A. SAFETY PRECAUTIONS - LC® SPRINT
Congratulations on your new PARI Trek®S
Portable Aerosol System. Your doctor has
prescribed the highest quality of delivery
systems for your nebulizer medication(s) to
treat your respiratory condition. Selected
for use in more than 25 clinical trials for new
respiratory medications, PARI LC
®
reusable
nebulizers deliver consistent particle size and
fast aerosol treatments for targeted delivery
to the lungs.
Your compressor and nebulizer play an
important role in treating your respiratory
condition. Follow your physician’s advice and
prescribed treatment regimen closely to achieve
the most effective therapeutic outcomes.
Read and understand the entire PARI Trek
®
S
Instructions for Use before using. Take
special note of all safety precautions marked
CAUTION, WARNING, and DANGER.
SAVE THESE INSTRUCTIONS
FOR FUTURE USE
CAUTION
US Federal Law restricts this device to sale by or on the
order of a physician. Use only when prescribed by a physician
and with the prescribed medication.
®
The LC
medication approved for nebulization and prescribed by a
physician. The LC
patients consistent with the indications for the aerosol
medication. Improper use can result in serious or fatal illness
or injury. Nebulizer is not sterile.
Sprint is a handheld nebulizer, designed to aerosolize
®
Sprint is intended for adult and pediatric
WARNING
Use only when prescribed by a physician and with the
prescribed medication. In order to achieve the specified
droplet characteristics the compressor or central air
supply must provide 3.5 - 8 lpm at a minimum of 20psi.
WARNING
1. To reduce the risk of increased bacterial growth,
infection, illness, or injury from contamination:
thoroughly clean and dry all parts of the nebulizer
and thoroughly dry any moisture or condensation in
the tubing at the end of every treatment following the
enclosed instructions.
2. Cleaning the nebulizer after each treatment is not
enough---you must regularly disinfect or sterilize the
nebulizer between treatments to prevent serious or
fatal illness caused by contamination of the nebulizer.
3. Never use a nebulizer with a clogged nozzle. If the
nozzle is clogged, the aerosol mist will be reduced,
altered or stopped, which will diminish or eliminate
the effectiveness of the treatment. If clogging occurs,
replace with the same model PARI Reusable Nebulizer.
4. Close supervision is necessary when this product is
used by or near children and physically or mentally
impaired. Never use when sleeping or drowsy.
5. This product contains small parts that may present
a choking hazard to small children. The tubing also
presents a strangulation hazard. ALWAYS USE CLOSE
ADULT SUPERVISION WHEN ADMINISTERING A
TREATMENT TO A CHILD. DO NOT LEAVE CHILD
UNATTENDED DURING TREATMENT.
6. Patients under the age of 4 or any patients who have
special needs should use a facemask instead of the
valved mouthpiece.
7. Before use, check PARI Reusable Nebulizer for proper
assembly. All parts should be seated firmly in place.
Use of an improperly assembled nebulizer could
diminish or prevent adequate delivery of medication
and the effectiveness of the treatment.
EN-1
INDICATIONS FOR USE: PARI TREK® S PORTABLE AEROSOL SYSTEM
The PARI Trek® S nebulizer compressor is a portable, DC powered air compressor intended to provide a source
of compressed air for use with jet nebulizers. The Trek
nonsterile, prescription-use only, intended for use in hospital, clinic, or home environments.
B. SAFETY PRECAUTIONS - TREK® S PORTABLE AEROSOL SYSTEM
®
S is intended for adult and pediatric patients. The device is
WARNING
Use only with nebulizers and tubing authorized by PARI for
this device. Other nebulizers and tubing may result in improper
treatment.
When using electrical and medical products, especially when
children are present, basic safety precautions must
be followed.
DANGER
To reduce the risk of serious or fatal injury from electrocution:
1. Do not place or store the product where it can fall or be
knocked into bathtub, sink, water, or other liquid. Do not
place or drop into water or other liquid. Do not use while
bathing.
2. Do not reach for a product that has fallen into water or
other liquid. Unplug immediately.
WARNING
To reduce the risk of serious or fatal injury from electrocution,
fire or burns and to reduce the risk of damage and malfunction
to the product:
1. Before use, check the serial number label on the bottom of
the compressor, battery and on the Multi-Voltage
AC Adapter to ensure that the voltage and current
indicated on the unit correspond with the voltage and
current available.
2. Never operate this product if it has a damaged cord or
plug, if it is not working properly, if it has been dropped or
damaged, or if it has been exposed to any liquids inside the
case. Return the product to PARI for examination
and repair.
3. Keep all electrical cords away from heated surfaces. Do
not overload wall outlets or use extension cords.
4. Do not operate where oxygen is being administered in a
closed environment such as an oxygen tent.
5. Never drop or insert any object into any openings.
6. Be sure the PARI TREK
Voltage AC Adapter are unplugged prior to cleaning.
7. Do not open or disassemble the PARI TREK
Compressor. Refer all servicing to an authorized PARI
Service Center.
8. Use only adapters that are authorized for the PARI TREK
S Compressor. Use of unapproved accessories can lead
to improper treatment or damage to the compressor,
adapters, or the battery.
9. Never spray liquids onto the compressor housing or
battery. Fluid could cause damage to the electrical and
compressor parts and could lead to a malfunction. In the
event that fluids enter the unit, return the product to PARI
for examination and repair.
10. Do not cover up or use the PARI Trek® S compressor
inside the carry case. Covering the device may cause
excessive heat buildup in the compressor housing.
®
S compressor and the Multi-
®
S
®
WARNING
11. The compressor has auto-resettable thermal protection.
If the compressor shuts down due to temperature, it will
restart automatically when it cools off if the switch is not
set to the off position. Ensure the switch is set to the off
position if the compressor shuts off due to temperature.
12. Always unplug immediately after using.
13. Do not use the PARI TREK
Reusable Nebulizer while operating a vehicle.
14. Portable and mobile RF communication devices can disrupt
electrical medical equipment. The electrical field generated
when the compressor is running can interfere with screens
and monitors (flickering). For this reason, the compressor
must be placed far enough away from screens/monitors
to prevent such interference. Technical electromagnetic
compatibility data is available in table form upon request
from PARI Respiratory Equipment Inc./Patient Connection
1.800.FAST.NEB (327.8632), press 2.
®
S Compressor or the PARI
BATTERY SAFETY PRECAUTIONS
WARNING
1. Use only the PARI TREK® S Rechargeable Battery described
in these Instructions For Use for operating your PARI
®
S Compressor.
TREK
2. Do not immerse the PARI TREK
water.
3. Do not touch the battery with wet or moist hands.
4. Do not expose battery to direct sunlight, humidity, extreme
temperatures or chemicals.
5. Make sure battery poles do not come into contact with keys
or other metal objects, causing a “short circuit”.
6. Do not attempt to open the plastic case or to replace any
components. Do not disassemble Rechargeable Battery. The
Rechargeable Battery is a sealed unit and can not
be repaired.
7. Do not incinerate (burn).
®
S Rechargeable Battery in
CAUTION
1. To charge the PARI TREK® S Rechargeable Battery use only
the PARI TREK
12 V DC Adapter. Using other adapters may damage the
battery.
2. In case of failure in charging or operating, unplug the
Rechargeable Battery from the Multi-Voltage AC Adapter
and turn the unit off. Should the Rechargeable Battery fail to
charge or operate properly, even if the Instructions For Use
are followed, notify the Authorized Service Center by calling
PARI Patient Connection toll free (USA) at 1.800.FAST.NEB
(327.8632), then press 2.
3. Protect the environment: when replacing the Rechargeable
Battery, recycle old battery or dispose of properly in
accordance with local laws.
1. Plug the AC adapter into a polarized wall outlet.
(Fig. A)
2. Plug the other end of the power cord into the power
receptacle on the side of the compressor or optional
rechargeable battery. (Fig. C)
With 12V DC Adapter
NOTE: DO NOT USE WHILE OPERATING A VEHICLE
1. Plug the 12V DC adapter into the vehicle outlet.
(Fig. B)
2. Plug the other end of the adapter into the power
receptacle on the side of the compressor or optional
rechargeable battery. (Fig. C)
With Optional Rechargeable Battery
The PARI Trek® S Li-Ion rechargeable battery is available
as an option to help maximize your freedom with
portable treatments. Speak with your physician and
homecare company about your daily schedule’s needs and
if a rechargeable battery is a good fit for you.
1. Completely charge the battery for the first time before
use. Refer to the “Li-Ion Rechargeable Battery”
section on page 5.
2. The rechargeable battery cannot be connected
directly to the PARI Trek
or 12V DC adapter are plugged into the power
receptacle. (Fig. D)
3. Attach the rechargeable battery to the PARI Trek
by placing the battery onto the rear section of the
compressor and sliding it into place. Ensure battery
is securely locked onto the compressor prior to
operation. (Fig. D)
4. Use the indicator lights to determine status of the
battery. Refer to “Li-Ion Rechargeable Battery” section
on page 5.
When the battery is completely discharged all indicator
lights will extinguish and the compressor will shut off.
NOTE: If battery fails to operate, connect the AC adapter into
the wall outlet. Plug the other end of the AC adapter into the
charge receptacle on the side of the rechargeable battery. The
charge indicator should turn “ON” indicating that the battery
has re-set and is now charging. If the charge indicator does
not appear, or if all four indicators are blinking, the battery
has failed and should not be used. Please contact PARI Patient
Connection at 1.800.FAST.NEB (327.8632)
5. To remove the battery, push and hold the release
button located on the underside of the battery. Slide
the battery away from the compressor. (Fig. E)
®
S if either the AC adapter
®
S
A
C
Battery OnlyCompressor OnlyBattery with Compressor
1. Completely charge the battery before using for
the first time for more than two hours or until all
three battery level indicators are lit, and the charge
indicator turns off.
2. To charge the battery, connect the AC adapter
into a polarized wall outlet (Fig. A). Plug the other
end of the AC adapter into the rechargeable battery
charge receptacle (Fig. C). The charge indicator will
turn “ON”.
3. When the rechargeable battery is fully charged, the
charge indicator will turn “OFF”. Unplug the multivoltage AC adapter from the rechargeable battery
charge receptacle. The amount of time it takes to
recharge the battery will depend on the age and
charge level of the battery.
NOTE: Do not leave the multi-voltage AC Adapter plugged into
the rechargeable battery for more than 7 days. Leaving the
battery connected to the power adapter for long periods of
time may decrease the life of the battery.
When not regularly using the battery, the battery should
be fully charged every six months to optimize and
preserve the life of the rechargeable battery.
NOTE: If the battery goes for a long period of time without
charging, battery life will be decreased.
Storing
The battery should be stored fully charged in a cool dry
place.
Charging:
Charge Indicator “ON” .................................................................. Battery is being charged
Charge Indicator “OFF” ................................................................... Battery is fully charged
Operation with Battery:
Three Green Lights ................................................ More than 75% remaining battery life
Two Green Lights .................................................... More than 50% remaining battery life
One Green Light ..................................................... More than 25% remaining battery life
One Green Light Flashing ........................................ Less than 25% remaining battery life
You are 10 Steps away from starting your treatment with
®
S. Please follow step by step.
Tre k
1. Clean/Disinfect Nebulizer Cup: Refer to Sections E-F to
properly clean and/or disinfect your nebulizer prior
to first use. Re-assemble nebulizer.
2. Set-up Compressor: Place the compressor on a flat,
stable surface free of dust. Do not operate the
compressor while it is in the carry case or covered,
as this can cause excessive heat buildup in the
compressor housing. Avoid cushioned, upholstered or
carpeted surfaces. Make sure power switch is turned
“OFF”
3. Connect Tubing To Compressor: Insert one end of
Wing Tip
.
®
tubing to the air outlet on the compressor.
It should fit snugly (Fig. F).
4. Load Medication: Open the inspiratory valve cap
of nebulizer by lifting the cap (Fig. G). Pour the
prescribed dose of medication into nebulizer cup.
Never fill past the max. fill line of 8 ml.
5. Attach Mouthpiece or Mask: If using the mouthpiece,
attach the mouthpiece to the nebulizer outlet
with the expiratory valve facing up. Then close
the inspiratory valve cap by pressing firmly on the
nebulizer top. If using a mask, refer to the mask
instructions for use.
6. Attach Tubing To Nebulizer Cup: Firmly press the other
end of the Wing Tip
®
tubing to bottom of PARI
reusable nebulizer (Fig. H).
7. Starting Treatment: Sit in a relaxed, upright position and
turn the compressor “ON”
.
8. Nebulizer Position: Place the mouthpiece on top of
tongue and close lips around it. Take slow, deep
breaths. As you inhale, the inhalation valve opens.
9. Breathing: For mouthpiece, inhale and exhale slowly
through your mouth. As you exhale, the valve on
the mouthpiece opens to allow the exhaled mist to
escape. For mask, place mask over nose and mouth
and inhale and exhale slowly.
10. Complete Treatment: Continue until the medication
is gone or you hear a slight “sputtering” sound. Turn
compressor “OFF”
NOTE: The PARI TREK
automatic resettable thermal switch. If the PARI TREK
.
®
S Compressor is protected by an
®
S Compressor has reached its temperature limit, it will
automatically turn off to prevent damage to the compressor.
In the unlikely event this occurs, switch off the PARI Trek
®
S Compressor, ensure it is on a flat, stable surface and is
not covered, and allow the compressor to cool down. If the
PARI Trek
®
S Compressor is not manually switched off, it will
automatically restart when the compressor cools.
®
The PARI TREK
minutes to cool down prior to resuming the treatment. If the
PARI TREK
S Compressor may need approximately 30
®
S Compressor does not start after it is switched on
after cooling for 30 minutes, please switch it off, allow another
few minutes for the compressor to cool, and try again. If you
need to use the PARI Trek
®
S Compressor for longer than 30
minutes, please contact PARI at 1.800.FAST.NEB (327.8632).