
© Panduit Corp. 2017 FS133
Page: 1 of 28
Terminating OptiCam Connectors Using OptiCam 2 Termination Tool
Panduit Corporation 18900 Panduit Drive, Tinley Park, IL 60487
WARNING
Read and understand the instructions and safety information in this manual
before operating this tool.
Failure to observe this warning can result in bodily injury.
Warning: Risk of fire. Battery can explode or leak and cause injury if
installed backwards, disassembled, charged, crushed or exposed to fire or
high temperature.
CAUTION
Caution - Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazardous radiation
exposure. The laser is emitted from the LC connector located at the top of
the tool. See the next page illustration for connector location.
• Never point the laser into the eyes of others.
• Do Not stare directly at the laser beam.
• Do Not set up tool to work at eye level or operate the tool on a
reflective surface as the laser could be projected into your eyes or the
eyes of others.
Viewing the laser output with certain optical instruments (for example, eye
loupes, magnifiers, and microscopes) within a distance of 3.9” (99 mm) may
pose an eye hazard.
Remove the batteries when storing for an extended period of time to avoid
damage to the tool should the batteries deteriorate.
TECHNICAL INFORMATION
Recommended Use: With available launch cords and cradles only
Laser Diode Type: IEC 60825-1 Class 1 Laser product
Connectors: LC, SC, ST, single mode and multimode,
Panduit connectors only
Power Supply: 2 size AA (type L91) 1.5volt Lithium Iron
Disulfide batteries, Max Discharge: 2.5 amp
continuous, included. Size AA alkaline
batteries can be used at a reduced rate battery
life.
Tool can also be powered without batteries
through the power/data (5 VDC, 1 amp max)
port on the side of the tool.
Rated Battery Life: Approximately 8 hours (continuous use)
Firmware Updates: Use power/data port on the side of the tool to
connect to PC, visit www.panduit.com for
latest firmware
Environment: FOR INDOOR USE ONLY
Max altitude 2000m
Pollution degree 2
Operating Temperature: +32° F to +104° F (0° C to +40° C), <93% RH,
non-condensing
Storage Temperature: -40° F to +158°F (-40°C to +70°C)
Overall Dimensions: 2.1" H x 2.8" W x 9.1" L (53mm x 71mm x
231mm)
Weight: 9.0 oz (255 g) without batteries
Tool Storage
After all connector terminations are complete, detach the launch cord from the tool by disconnecting the LC connector at the top of the tool. Detach the
ferrule adapter from the cradle and put dust caps on both ends of the launch cord. Place tool, launch cord, and cradle in a protective case.
Tool Use
This tool is to be used for the termination of Panduit OptiCam connectors only. The tool emits a non-visible laser beam, which, when used in conjunction
with the launch cord and cradle, aids in proper connector termination.
• Laser tools are precision instruments, which should be handled with
care. See Safety Precautions.
• Avoid shock, vibrations, and extreme heat.
• Avoid dust and water that could obstruct laser.
• Keep tool dry and clean.
• Check batteries regularly to avoid deterioration.
• Remove batteries if the tool is to be stored for an extended period.
• This tool contains no user serviceable parts.
• The laser output is not user adjustable.
• Contact Panduit Corp. for service needs.
Do not Throw Away
This symbol indicates the need for separate collection of electrical and electronic equipment waste.
Separating electronic waste can halt the potential adverse effects on the environment and human health as a result of the hazardous
substances in electrical and electronic equipment.
This waste should be returned to the proper collection facility.

© Panduit Corp. 2017 FS133
Page: 2 of 28
Compliance symbols
Complies with IEC 60825-1 Ed. 3 (2014).
Complies with FDA performance standards for laser products except for deviations pursuant to Laser Notice No. 50, dated May 15, 2014.
This device compiles with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including any interference that may cause
undesired operation.
La operación de este equipo está sujeta a las siguientes dos condiciones: (1) es posible que este equipo o dispositivo no cause interferencia perjudicial y (2)
este equipo o dispositivo debe aceptar cualquier interferencia, incluyendo la que pueda causar su operación no deseada.
CAN ICES-3(B)/NMB-3(B)
IMPORTANT
DO NOT remove any labels from this tool
Compliance Label
Table of Contents
Safety Precautions ............................................................................................................................................................ 3
Tool Set Up ........................................................................................................................................................................ 4
Component Identification ................................................................................................................................................. 9
Buffer Stripping Steps ..................................................................................................................................................... 11
Cleave Fiber Steps ........................................................................................................................................................... 14
OptiCam 2 Tool Operation -LC ........................................................................................................................................ 15
Connect-It App Use ......................................................................................................................................................... 18
Assemble Connector and Boot ........................................................................................................................................ 22
Tool Care and Maintenance ............................................................................................................................................ 26

© Panduit Corp. 2017 FS133
Page: 3 of 28
Safety Glasses
WARNING: It is strongly recommended
that safety glasses be worn when handling
bare optical fiber. The bare fiber is very
sharp and can easily damage the eye.
Isopropyl Alcohol (91% or better)
WARNING: FLAMMABLE.
Contact with Alcohol can cause irritation
to the eyes. In case of contact with eyes,
flush with water for at least 15 minutes.
Always us Alcohol with proper levels of
ventilation.
In case of ingestion consult a physician
immediately.
Disposal of Bare Fibers
WARNING: Pick up and discard all pieces
of bare fiber with sticky tabs. Do not let
cut pieces of fiber stick to clothing or drop
in the work area, where they are hard to
see and can cause injury.
Laser Light
DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS. CLASS 1M LASER PRODUCT.
Viewing the laser output with certain optical instruments (for example, magnifiers, and microscopes) within a distance of 100 mm
may pose an eye hazard.
CAUTION: use of controls of adjustments or performance of procedures other than those specified herein may result in hazardous
radiation exposure.
• Never point the laser into the eyes of others
• Do not stare directly at the laser beam.
Do not set up tool to work at eye level or operate the tool on a reflective surface as the laser could be projected into your eyes, or
the eyes of others.
Cable Handling
WARNING: Fiber optic cable can be
damaged by excessive pulling, twisting,
crushing or bending stresses. Consult the
appropriate specification sheets as
provided by your cable vendor. Any
damage may decrease optical
performance.
To Ensure Best Termination
• Make sure the batteries are new and fresh.
• Make sure the launch cord is in good condition, the split sleeve is in good
condition, and the ferrule end face is clean and not scratched.

© Panduit Corp. 2017 FS133
Page: 4 of 28
OptiCam 2 Tool Overview
Various components of the
OptiCam 2 Tool
Select Proper Launch Cord and
Cradle
For each connector type, select the
correct launch cord and cradle for
termination use.
Launch cord, Cradle and connector cross reference table
yellow cable, 1.25mm single mode
(P/N: FOLPC-1.25SM)
LC cradle, black
(P/N: FLCC2)
orange cable, 1.25mm multimode
(P/N: FOLPC-1.25MM)
LC cradle, black
(P/N: FLCC2)
LC multimode; OM1,
OM2, OM3, and OM4
yellow cable, 2.5mm single mode
(P/N: FOLPC-2.5SM)
SC cradle, gray
(P/N: FSCC2)
orange cable, 2.5mm multimode
(P/N: FOLPC-2.5MM)
SC cradle, gray
(P/N: FSCC2)
SC multimode; OM1,
OM2, OM3, and OM4
yellow cable, 2.5mm single mode
(P/N: FOLPC-2.5SM)
ST cradle, gray
(P/N: FSTC2)
orange cable, 2.5mm multimode
(P/N: FOLPC-2.5MM)
ST cradle, gray
(P/N: FSTC2)
ST multimode; OM1,
OM2, OM3, and OM4
* Each cradle contains marking to identify as type LC, SC, or ST.
Left orientation for cradle
shown.
Removeable and reversible
cradle

© Panduit Corp. 2017 FS133
Page: 5 of 28
Load Batteries
To load or replace batteries,
remove the battery cover using a
Phillips screw driver.
Insert two (2) size AA (type L91)
1.5volt Lithium Iron Disulfide
batteries, follow the polarity
markings in the compartment.
Then replace the cover and secure
using the Phillips screw driver.
WARNING: Risk of Fire.
Battery can explode or leak
and cause injury if installed
backwards, disassembled,
charged, crushed or exposed
to fire or high temperature.
Note: OptiCam 2 Tool does not charge batteries
Register and Update Firmware
Please visit Panduit
The OptiCam 2 has a data port on
the side of the tool which can be
used to install firmware updates.
Please update the tool to the most
recent firmware available at
Panduit.com > Support > Download
Center >
Software/Firmware/Printers >
OptiCam 2 Termination Tool.
Remove protective films
The protective film on the Camera
Aperture has a removal tab. The
protective film on the LCD display
can be removed by peeling from
one corner.
Note: Remove the protective film
from the Camera Aperture before
operating the tool.
Must be power
and data cord

© Panduit Corp. 2017 FS133
Page: 6 of 28
Attach Launch Cord
Install the correct OptiCam 2
launch cord and OptiCam 2 cradle.
An LC connector on the launch cord
attaches to a port at the top of the
tool and the unique adapter
attaches it to the cradle.
Please refer to the cross-reference
table on page 4, to make sure the
proper combination of cord, cradle,
connector, and orientation is being
used.
Position Cradle
The cradle attaches to the tool
magnetically. Position the cradle
onto the tool with either a left or
right orientation and the magnets
will properly align the cradle. Right
and left orientations shown (LC
launch cord and LC Cradle are
shown).
Rgiht Orientation for the cradle Left Orientation for the cradle
Turn on Power
Power on by using the center of the
navigation button, pressing directly
down.
Right orientation
for cradle shown
Launch cord
snaps into
the cradle

© Panduit Corp. 2017 FS133
Page: 7 of 28
Select Connector Options
Use the navigation button to
navigate to the menus and to
“Connector Type” and select the
correct connector options
Menu > Connector Type
Use the navigation button left and
right button to select the correct
information. The selected text will
become red, “Save” this option.
Note: To ensure proper function of the
tool, check that settings on the tool
match the cord, cradle, connector, and
orientation being used.
Load Connector
LC and SC connector loading: A) Insert
a connector while the launch cord is
tilted up to receive the connector. B)
Tilt the connector down, ensure the
connector backbone snaps in to the
cradle.
ST connector loading: shown on the
next page.
LC Load Connector SC Load Connector
Connector,
orientation, & color,
change to match
selection
Next icon. The
outline indicates
this is current
selection

© Panduit Corp. 2017 FS133
Page: 8 of 28
Load Connector Continued
ST connector loading: A) Rotate the ST
connector into the cradle. B) Rotate
the retaining clip over the connector C)
Secure the clip with the audible “click”
D) Place the launch cord on the
connector end with the split sleeve
attaching to the connector ferrule.
Note: ST requires that the launch cord
be held in place by the user.

© Panduit Corp. 2017 FS133
Page: 9 of 28
LC launch cord and cradle

© Panduit Corp. 2017 FS133
Page: 10 of 28
SC Launch Cord and Cradle
ST Launch Cord and Cradle
Connector Types with Colors

© Panduit Corp. 2017 FS133
Page: 11 of 28
250μm Coated Fiber Strip
First install the boot. Insert the
fiber end through the small end of
the 250μm boot. Slide the boot
back out of the way.
Place one build-up tube (F250BT)
onto the fiber and slide it back out
of the way.
Place an insertion mark on the
build-up tube at 0.59” (15mm) for
LC and 0.43” (11mm) for SC/ST,
from the end of the tube closest to
the end of the fiber.
Use buffer stripper to strip at least
1.18" (30mm) of coating.
Place an insertion mark on the
coated fiber 2" (51mm) from the
end of the stripped coating.
Slide the leading edge of the buildup tube to the start of the coating
after cleaving the fiber.
900μm Tight-Buffered Fiber Strip
Insert the fiber end through the
small end of the 900μm boot. Slide
the boot back out of the way.
Use buffer stripper to strip at least
1.18" (30mm) of buffer.
Use the marking pen to place an
insertion mark at 0.59” (15mm) for
LC and 0.43” (11mm) for SC/ST,
back from the end of the stripped
buffer.
1.18” (30mm)
Coating Removal
0.59” (15mm)
Insertion Mark
0.43” (11mm)
Insertion Mark
250µm Coated Fiber Stripping Dimensions
IMPORTANT! Due to printing variations, use a rigid
scale for stripping dimensions, or the appropriate
Panduit laminated template.
900µm Coated Fiber Stripping Dimensions
IMPORTANT! Due to printing variations, use a rigid
scale for stripping dimensions, or the appropriate
Panduit laminated template.
1.18” (30mm)
Coating Removal
0.59” (15mm)
Insertion Mark
0.43” (11mm)
Insertion Mark

© Panduit Corp. 2017 FS133
Page: 12 of 28
Jacketed Cable Strip
If using duplex cable, split the two
cables approximately 6" (152mm)
or as needed.
Following the stripping dimensions,
use the marking pen to mark the
cable jacket at 1.89” (48mm) for
LC, and 1.73" (44mm) for SC/ST,
from the end.
Strip off the cable at the jacket
removal mark. To reduce wear on
the jacket stripper blades, do not
slide the blades along the aramid
yarn. Instead, use the tool to cut
through the jacket, then pull off the
jacket by hand.
Insert fiber through the small end
of appropriate boot for your
specific cable type. Use the boot to
fold the aramid yarn back over the
jacket, holding it out of the way.
Use the marking pen to place a
buffer length mark on the buffer at
0.71” (18mm) for LC, and 0.55"
(14mm) for SC/ST, from the edge of
the jacket. Place an insertion mark
at 0.59” (15mm) for LC and 0.43"
(11mm) for SC/ST, back from the
first mark.
Use the marking pen to place a
mark on the buffer at the edge of
the jacketing. This mark should
remain at the edge of the jacket
after buffer stripping.
Use the buffer stripper to strip the
buffer to the buffer length mark.
250μm Coated Fiber Termination
using Opticom Fanout Kit
Install the Fanout Kit (FO6CB,
FO12CB) according to the
installation instructions FS047*.
Feed the 250μm coated fibers
through the 900μm fanout tubing
until at least 30mm of fiber
protrudes from the end of the
tubing. Trim the coated fibers to
length at 30mm.
To reduce the amount of fiber
movement, place a drop of
adhesive (Loctite 401 or equivalent
For 3.0mm jacketed fiber, use the third hole (marked "1.3mm" or #16AWG) from
the tip of the jacket stripper. For 1.6mm - 2.0mm jacketed fiber, use the first hole
(marked "0.8mm" or #20AWG) from the tip of the jacket stripper.
Jacketed Cable Stripping Dimensions
IMPORTANT! Due to printing variations, use a rigid
scale for stripping dimensions, or the appropriate
Panduit laminated template.
1.18” (30mm)
Buffer Removal
Insertion Mark: LC 0.59” (15mm), SC/ST 0.43” (11mm)
Buffer Length: LC 0.71” (18mm), SC/ST 0.55” (14mm)
Jacket Removal: LC 1.89” (48mm), SC/ST 1.73” (44mm)
250µm Coated
Fibers (taut)

© Panduit Corp. 2017 FS133
Page: 13 of 28
Cyanoacrylate adhesive) at the
point that the coated fibers exit the
3mm cable jacket. 250μm coated
fibers need to be taut between
cable jacket and the fanout tube
assembly.
Place the 900μm boots onto the
fanout tubing and slide them back
out of the way.
Place an insertion mark on the
fanout tubing, at 0.59” (15mm) for
LC and 0.43" (11mm) for SC/ST,
back from the end of the tubing.
Place a coating removal mark on
the coated fiber 3mm from the end
of the fanout tubing. Strip the
coating up to the coating removal
mark. Note: The 3mm of coating
extending from the end of the
tubing is intended to compensate
for fiber movement within the
fanout tubing during termination.
Buffer Stripping
Remove no more than 1/4" (6mm)
of buffer at a time to avoid
breaking the fiber.
Hold the buffer stripper such that
the tool is at a 90 degree angle to
the fiber to ensure clean cut of
buffer.
Position the fiber in the tool's Vnotches. Squeeze the handles and
pull toward the end of the fiber.
Note: When closing the tools
handles, the fiber may bend.
Rotate the tool while closing the
handles to keep the fiber straight.

© Panduit Corp. 2017 FS133
Page: 14 of 28
Prep Stripped Fiber
Remove and dispose of any
remaining fiber scraps in the
cleaving tool.
Clean bare fiber using an isopropyl
alcohol (92% concentration) soaked
lint-free wipe. The fiber should be
free of all coating and residue. To
prevent contamination, do not set
fiber down or allow it to touch
anything after cleaning.
Hold the cleaving tool in your hand
and depress the fiber clamp lever
to raise the fiber clamp. Insert the
end of the fiber into the V-groove
and under the fiber clamp. Release
the fiber clamp lever when the end
of the buffer is located at the 7mm
mark (±0.5mm), this length is
critical to termination.
Score & Cleave
This is critical step for successful
termination. Gently hold the
buffered fiber against the groove in
the base plate using thumb and
forefinger of your free hand. With
your other thumb, gently push
down on the housing until the
ceramic blade contacts the fiber
only once. DO NOT cut the fiber
with the ceramic blade. Release the
housing so that the blade is no
longer contacting the fiber.
Hold the tool in one hand while still
holding the fiber against the base
plate. Bend the base plate down
until the fiber breaks. To prevent
contamination, do not re-clean the
fiber, do not set fiber down, or
allow it to touch anything after
cleaving.
IMPORTANT: Do not clean fiber after cleaving. Replace the ceramic
blade in the cleaving tool after every 1000 cleaves, or if installation
yield decreases. Replacement blade (FJQCVRB) is available.

© Panduit Corp. 2017 FS133
Page: 15 of 28
OptiCam 2 Tool Operation - LC
Select Connector Options
Use the navigation button to
navigate to the menus and to
“Connector Type” and select the
correct connector options
Menu > Connector Type
Use the navigation button left and
right button to select the correct
information. The selected text will
become red, “Save” this option.
Note: To ensure proper function of the
tool, check that settings on the tool
match the cord, cradle, connector, and
orientation being used.
Load Connector
A) Insert a connector while the launch
cord is tilted up to receive the
connector. B) Tilt the connector down,
ensure the connector backbone snaps
in to the cradle.
Note: Each connector has slightly
different cradle operation. Refer to
the individual connectors’ specific
instructions from page 7-10.
Connector Test
To start the connector test, select the
“Next” icon, pressing the navigation
button directly down.
The tool will perform a background
test of the connector. After passing the
background test, the “Insert Fiber”
command will display automatically.
Insert Cleaved Fiber
Insert cleaved fiber and maintain
gentle forward pressure while waiting
for the “CAM & TEST” command to
appear on the screen automatically.
Connector,
orientation, & color,
change to match
selection
Next icon. The
outline indicates
this is current
selection

© Panduit Corp. 2017 FS133
Page: 16 of 28
Cam Connector
When the screen displays the
“CAM & Test” command, A)
Maintain gentle forward pressure
on the field fiber, and B) “Cam” the
connector, then select the “Next”
arrow to perform test.
LC Camming requires that the
whole connector rotate.
SC Camming requires that the key
“Fin” rotates, and the rest of the
connector remains in place.

© Panduit Corp. 2017 FS133
Page: 17 of 28
ST Camming is done by rotating the
key lever of the cradle, which
engages the internal camming
parts of the connector.
Review Test Results
Verify IL value meets project
requirements.
If needed, re-test by using the “Re-
test” icon, or re-terminate using
the “Back” icon and reperform
steps 5 and 6. The OptiCam
connector can be reused up to
three times.
Test the Next Connector
Select the “Next” icon on the
display to reset the tool to the next
connector.

© Panduit Corp. 2017 FS133
Page: 18 of 28
Turn on Wireless
Navigate to the menus of the
OptiCam 2 tool and select:
Tool Settings > Wireless > On.
Then select “Exit”.
Enter the wireless password given when registering the tool at Panduit.com
Then turn wireless on.
Note: The tool name is found under Menus > Tool Settings > Tool Name. Rename for
ease of use.
Create a Project
Create a new project in
Connect-It app.
First, enter project credentials
then select “Create Project
File”
Select an
existing project
Enter Project
information

© Panduit Corp. 2017 FS133
Page: 19 of 28
Connect the App to the Tool
Connect to the OptiCam 2 tool,
pull down to refresh the list,
and select the appropriate tool
name from the list. Then select
“Start Project”.
Transfer Un-Synced Records
The app will indicate that the
tool has un-synced records if
terminations were done on the
tool without the app and tool
being synced. Select “OK” to
sync which will transfer the
records of the termination(s) to
the project file in the app.
Note: The records will no
longer be available on the
OptiCam 2 tool.
Share Test Records
To export the connector
termination data, navigate to
the “Project Overview”, select
the current project.
Three available options:
Share – enables the file to be
emailed;
Select – opens the file;
Delete – deletes the file from
the app.
To share select the “Share”
option. Your mobile device’s
default email will open with a
CSV attachment containing the
connector data.
Select the OptiCam 2
tool name

© Panduit Corp. 2017 FS133
Page: 20 of 28
Tool Operation with Mobile
App
While the mobile app
connected to the OptiCam 2
tool, open a previous project or
create a new project. Verify the
tool settings and connector
settings on the “Tool Settings”
tab. Then select the “Test
Window” tab.
Load Connector
Verify the connector type in
the “Test Window”, then install
the connector into the cradle
and select “Load Connector.”
The tool will perform a
Background Test, and
automatically display Insert
Fiber.
Insert Fiber
Insert the fiber into the
connector, when fully inserted
the graphic on the tool will
change to “Cam & Test”.

© Panduit Corp. 2017 FS133
Page: 21 of 28
Cam and Test the Connector
Cam the connector by rotating
it, and maintaining pressure on
the field fiber at the same time.
Select “Perform Test” on the
app.
Test Results
Test result will display.
Verify IL value meets project
requirements.
If needed, re-test by using the
“Re-test” icon, or re-terminate
using the “ReTerminate” icon
and repeat the fiber clean and
cleave steps. The OptiCam
connector can be reused up to
three times.
Share Test Records
To export the connector
termination data, navigate to
the “Project Overview”, select
the current project.
Three available options:
Share – enables the file to be
emailed;
Select – opens the file;
Delete – deletes the file from
the app.
To share select the “Share”
option. Your mobile device’s
default email will open with a
CSV attachment containing the
connector data.

© Panduit Corp. 2017 FS133
Page: 22 of 28
Assemble Connector and Boot
LC Connector and Boot
Grip the boot by the large end
and push the boot forward
onto the grooved area of the
backbone until it is against the
flange.
Insertion mark should be at
end of backbone.
LC Connector with Dust Cap
Clean the ferrule end face with
an alcohol soaked wipe, and
then a dry wipe. Place dust cap
over the ferrule end.
Attach Duplex Clip (Optional)
With a connector held as
shown (latch on top, ferrule
facing away), insert the
connector into one side of the
duplex clip as shown (clip held
with the “A->B” polarity
marking upright and facing
forward). The upper tab of the
clip should slide into the pocket
underneath the latch of the
connector, and the lower tab
should slide under the
connector housing and “snap”,
locking it into place.
Repeat this procedure for the
other connector, completing
the duplexing step.
Note: When making cable
assemblies, be sure to follow
the correct “A->B” polarity
cross-over between
connectors.

© Panduit Corp. 2017 FS133
Page: 23 of 28
SC Connector and Boot
Grip the boot by the large end
and push the boot forward
onto the grooved area of the
backbone until it is against the
flange.
Insertion mark should be at
end of backbone.
SC Outer Housing
Align the external chamfers on
the front of the inner housing
assembly with the internal
chamfers at the front of the
outer housing.
Push the inner housing
assembly into the outer
housing until it snaps into
place. Once in place the outer
housing should be able to
move freely back and forth
over the inner housing and
boot.
SC Outer Housing Removal
The connector can be reterminated if needed. Make
sure that the dust cap is on the
ferrule. Grip the outer housing
and push tip of connector
down onto a hard surface. The
inner housing will release from
the outer housing. Remove
outer housing. Slide the boot
off the backbone. Un-cam the
connector and remove the
fiber.
CAUTION: DO NOT push down
without a dust cap installed on
the ferrule. Doing so could
damage the pre-polished fiber
endface.

© Panduit Corp. 2017 FS133
Page: 24 of 28
SC Connector with Dust Cap
Clean the ferrule end face with
an alcohol soaked wipe, and
then a dry wipe. Place dust cap
over the ferrule end.
Attach SC Duplex Clip
(Optional)
Orient the connector as shown
with the connector key acing
upward in the same direction
as the ‘A’, ‘B’ and ‘KEY’
markings on the duplex clip.
Insert the connector into one
side of the clip on a slight angle
and rotate into place. The
upper and lower legs of the clip
slide over the connector and
“snap” when the connector is
in place. The connector is fully
seated once the top and
bottom connector flanges are
inside of the rectangular slots
in the clip.
Repeat this procedure for the
other connector, completing
the duplexing step.
ST Connector and Boot
Grip the boot by the large end
and push the boot forward
onto the grooved area of the
backbone until it is against the
flange.
Insertion mark should be at
end of backbone.
Note: When making cable assemblies, be sure to follow the
correct ‘A->B’ polarity cross-over between connectors.

© Panduit Corp. 2017 FS133
Page: 25 of 28
ST Connector with Dust Cap
Clean the ferrule end face with
an alcohol soaked wipe, and
then a dry wipe. Place dust cap
over the ferrule end.
Connector Boot Assembly with
Jacketed Cable
Slide the boot back to release
the aramid yarn. Cut the
aramid yarn approximately
0.20" (5mm) past the flange.
Flare the aramid yarn evenly
around the perimeter of the
grooved area of the backbone.
Use tweezers for best results.
There will be a slight gap
between the cable jacket and
the backbone. Push the boot
forward and thread clockwise
onto the grooved area of the
backbone until it seats against
the flange. Best cable retention
is achieved when the aramid
yarn protrudes slightly from
the flange area.

© Panduit Corp. 2017 FS133
Page: 26 of 28
Tool Care and Maintenance
Tool Care and Maintenance
The LC connector port and data
port protective dust caps must
be installed when the tool is
not in use to prevent
contamination.
Launch Cord Care and
Maintenance
The dust caps for both the LC
connector (tool end) and the
Ferrule Adapter (field
connector end) must be
installed when the cord is not
in use. Do not coil the cable to
less than one-inch diameter at
any time to avoid damaging the
fiber.
Note: 1.25 millimeter cord is
shown for reference. The
process is the same for the 2.5
millimeter cord.
SC Launch Cord Ferrule Adapter
The camera aperture should be
cleaned periodically with a soft, lint
free cloth or with isopropyl alcohol
to remove dust and debris which
may interfere with proper camera
operation. Avoid any scratches on
the surface. The LC connector port
can be cleaned with clean dry air, do
not use alcohol.
LC Connector End
with Dust Cap

© Panduit Corp. 2017 FS133
Page: 27 of 28
LC Launch Cord Ferrule Adapter
Remove the Ferrule Adapter
Fiber tip of the launch cords
should be cleaned every fifty
(50) connector terminations.
To clean, retract the boot to
expose the cable crimp sleeve.
Remove the Ferrule Adapter by
holding the crimp sleeve and
unthreading from the Ferrule
Adapter Assembly to expose
the white ceramic split sleeve.
Remove Split Sleeve
Twist the split sleeve and pull
to remove it from the
assembly, exposing the tip of
the fiber.
Prep Lint Free Cloth
Wet an area of the fiber optic
cleaning system wipes with
isopropyl alcohol.
Note: 1.25 millimeter cord is shown for reference. The process is
the same for the 2.5 millimeter cord.
Ferrule Adapter
Ceramic Split Sleeve
Adapter Assembly
Ceramic Split Sleeve

© Panduit Corp. 2017 FS133
Page: 28 of 28
Wet to Dry Clean the Fiber Tip
Clean the end face of the fiber
by wiping the tip of the fiber
along one of the fiber cleaning
slots (shown in blue) wet with
alcohol on the top of the fiber
cleaning system. Then repeat
the wiping procedure with a
dry slot (shown in red).
Reassemble Ferrule Adapter
Inspect the split sleeve for
chipping or cracks. Thread the
Ferrule Adapter onto the
Ferrule Adapter Assembly until
the parts make contact. Hold
the Cable Crimp Sleeve and
turn the Ferrule Adapter about
5° to tighten the adapter.
Clean LC Connector End of
Cord
Always clean the LC connector
end of the launch cord before
inserting the connector into
the tool. Remove the dust cap
and wipe the end face of the
fiber, using the same wet dry
procedure. Insert the
connector into the tool
immediately after cleaning.