· For continued safety, no modification of any circuit must be
attempted.
· Unplug the power cord from the power outlet before
disassembling this projector.
· It is advisable to use an isolation transformer in the AC
power line before the service.
· Observe the original lead dress during the service. If a short
circuit is found, replace all the parts overheated or
damaged by the short circuit.
· After the service, all the protective devices such as
insulation barriers, insulation papers, shields, and isolation
R-C combinations must be properly installed.
· After the service, check the leakage current to prevent the
customer from getting an electric shock.
1.2. Leakage Current Check
1. Prepare the measuring circuit as shown in Fig.1.
Be sure to use a voltmeter having the performance
described in Table 1.
3. Connect M1 to T1 according to Fig. 2 and measure the
voltage.
4. Change the connection of M1 from T1 to T2 and measure
the voltage again.
5. The voltmeter must read 0.375 V or lower in both of steps
3 and 4. This means that the current must be 0.75 mA or
less.
6. If the reading is out of the above standard, the projector
must be repaired and rechecked before returning to the
customer because of a possibility of an electric shock.
1.3. UV Precaution and UHM Lamp
Precautions
· Be sure to unplug the power cord from the power outlet
when replacing the lamp.
· Because the lamp reaches a very high temperature during
its operation, wait until it cools completely when replacing
the Lamp Unit.
· The lamp emits small amounts of UV-radiation, avoid directeye contact with the light.
· Because the high pressure lamp involves a risk of failure,
never touch the lamp wire lead during the service. (See Fig.
3)
Fig. 1
Table 1
Fig. 2
2. Assemble the circuit as shown in Fig. 2. Plug the power
cord in a power outlet.
This mode is used to display seven kinds of test patterns [Horizontal lines, Vertical lines, Dots, Crosshatch, White cross, Black
cross and White (No pattern)] in the four colors (White, Red, Green and Blue)..
Note:
· On the service mode, displays above patterns by each color without test equipment such as PC or SG. Use the service
mode for simplified adjustments by your eyes and so on.
4.1. Procedure to enter the service mode
Select "SERVICE MODE" on "EXT OPTION" menu and press "ENTER" button on the main unit or remote control unit.
Note:
· In the service mode, pressing the up-arrow "
left-arrow "
" or right-arrow "" button the color selection (White / Red / Green / Blue).
" or down-arrow "" button allows the test pattern selection and the
4.2. Canceling the service mode
Press "MENU" button on the main unit or remote control unit.
5 Flicker Adjustment Mode
If replacing the optical parts (Analysis / LCD / Lens block) of this projector and/or A-P.C.Board (assembly), enter the flicker
adjustment mode and minimize the flicker.
5.1. Procedure to enter the adjustment mode
Select "FLICKER ADJ" on "EXT OPTION" menu and press "ENTER" button on the main unit or remote control unit.
Note:
"DESK setting (blue)" is displayed when entering the adjustment mode.
5.2. Adjustment Display and Contents
· Setting value is increased and decreased with the right-arrow "" and left-arrow "" buttons.
"
": Decrease, "": Increase
−
− Adjust the setting value to minimize the flicker on the screen.
− −
−
− Execute the adjustment by 6 patterns below.
− −
·The pattern (adjustment display) is switched with the up-arrow "
"
": Forward direction, "": Reverse direction
−
− There are 6 patterns of "DESK setting (blue)", "DESK setting (red)", "DESK setting (green)", "CEILING setting (blue)",
− −
"CEILING setting (red)" and "CEILING setting (green)".
−
− The setting value is saved into this projector when the pattern is switched.
− −
" and down-arrow "" buttons.
5.3. Canceling the flicker adjustment mode
Press "MENU" button on the main unit or remote control unit.
Note:
When "MENU" button is pressed, the setting value at that time is saved into this projector and the adjustment mode is canceled.
The serial connector which is on the back connector panel of the projector conforms to RS-232C standard. This projector can
be controlled by a PC which is connected as shown in "6.1. Connection".
For controlling this projector by a PC, requires communication software on the market, and inputs control commands according
to communication settings and basic format below.
6.1. Connection
Note:
Use a proper communication cable which is suitable for the PC to connect the optional serial adapter, which is connected with
SERIAL connector of this projector, and the PC.
6.2. Pin Layout and Signal Names for SERIAL Connector
Pin No Signal NameContents
3RXDReceive data
4GNDGround
5TXDTransmit data
1--2---Connected internally
6--7---NC
8---NC
6.3. Communication Settings
Signal LevelContentsDescription
Sync. methodAsynchronousSynchronizes every 1 character (8 bits)
Baud rateConforms to9 600 bpsData transfer speed
ParityRS-232CNoneError detection method
Character lengthstandard8bitsNumber of bit composing 1 character
Stop bit1bitUses stop bit when asynchronous method
X parameterNot used
S parameterNot used
6.7. Signal Selector Connecting Cable Specifications
When connecting to a signal selector (ex. TW-SWS62J), use a cable with specifications below.
Connecting method: Connects a video signal cable from the signal selector to "VIDEO IN", and an RGB signal cable to "RGB1 IN".
Note:
Set VP control terminal switch of the signal selector to VP TYPE "B".
7.4. Removal of WL-P.C.Board (PTLB20NTU/E/EA only)
1. Remove the upper case according to the section 7.2.
"Removal of Upper Case".
2. Unscrew the 4 screws and remove the shield cover and
wireless LAN module.
Notes:
· Because the wireless LAN module is connected with
the WL-P.C.Board, work carefully when removing it.
· When installing/detaching the antenna connector,
insert or pull out it straight along the center of the
connector pin. It causes damage and/or poor
contactingoftheconnectorwhenforcibly
installing/detaching it.
· Connect the connector from the antenna with the
original position (connector on the wireless LAN
module) after reassembling.
3. Unscrew the 1 screw and remove the WL-P.C.Board.
7.6. Removal of S2-P.C.Board
1. Remove the S2-P.C.Board block according to the section
7.3. "Removal of A-P.C.Board".
2. Remove the holder while expanding the resin hooks
outside.
Note:
·Work carefully not to damage the resin hook.
7.5. Removal of S1-P.C.Board
1. Remove the upper case according to the section 7.2.
"Removal of Upper Case".
2. Remove the water-guide plate.
7.7. Removal of K-P.C.Board
1. Remove the analysis block, LCD block and lens according
to the steps 1 through 3 in the section 7.11. "Removal of
Analysis Block and Lens".
2. For PT-LB20NTU/E/EA, unscrew the 4 screws and remove
the shield cover and wireless LAN module.
Notes:
· Because the wireless LAN module is connected with
the WL-P.C.Board, work carefully when removing it.
· When installing/detaching the antenna connector,
insert or pull out it straight along the center of the
connector pin. It causes damage and/or poor
contactingoftheconnectorwhenforcibly
installing/detaching it.
· Connect the connector from the antenna with the
original position (connector on the wireless LAN
module) after reassembling.
3. For PT-LB20NTU/E/EA, unscrew the 2 screws and remove
the WL-P.C.Board (with metal fittings for securing WL).
3. Unscrew the 2 screws and remove the S1-P.C.Board.
according to the section 7.4. "Removal of WL-P.C.Board".
3. For PT-LB20NTU/E/EA, unscrew the 2 screws and remove
the metal fittings for securing WL.
4. Unscrew the screws (PT-LB20NTU/E/EA: 2, Others: 3) and
remove the grounding metal (2).
5. Unscrew the 2 screws and remove the P-Module.
Note:
· When installing the lamp unit (or a new one) in the
main unit, place it in a specified position and press
the connector side and the opposite side of the lamp
unit (arrow positions shown in the figure below), and
confirm the lamp unit is inserted securely.
Then, tighten the 2 screws fixing the lamp unit, and
attach the lamp unit cover.
7.10. Removal of Lamp Unit
1. Turn the projector upside down.
2. Loosen the 2 screws until they idle, remove the lamp unit
cover.
3. Loosen the 2 screws until they idle, remove the lamp unit
with the handle.
7.11. Removal of Analysis Block
and Lens
1. Remove the lamp unit according to the section 7.10.
"Removal of Lamp Unit".
2. Remove the A-P.C.Board block according to the steps 1
through 6 in the section 7.3. "Removal of A-P.C.Board".
3. Unscrew the 3 screws and remove the analysis block, LCD
· Since the ID-tag seal is pasted to the FPC of LCD Panl, (R),
(G) or (B) can be easily identified by the color of the seal.
· Finally, identify the panel color by the part number printed
on the FPC.
6. Reassemble the projector in the reverse order of
disassembling, but leave the upper case and the screws
fixing the A-P.C.Board block as they are removed.
7. Adjust the convergence according to the section 8.4.
"Convergence Adjustment".
8. After the adjustment, while paying attention not to vary the
adjusting result, tighten the 2 screws (upper) fixing the
panel attachment temporarily with a hexagon head wrench.
Notes:
·Prepare a hexagon head wrench processed short.
·Service tool (Part No. TZSH07010), hexagon head
wrench processed short, is available.
9. Remove the LCD block again.
10. Tighten the 3 screws fixing the panel attachment.
7.15. LCD Panel Combination
·Part number is printed on the FPC of LCD Panel.
·When replacing LCD Panel, use a component which has
the same part number as the original.
Model numberLCD
PT-LB20NTU/E/EA
PT-LB20U/E/EA
PT-LB20SU/E/EAGL5BDAXN00074
PT-LB20VU/E/EAGL5BDAXQ00249
panel
RL5BDAXQ00221
GL5BDAXQ00222
BL5BDAXQ00223
RL5BDAXN00073
BL5BDAXN00075
RL5BDAXQ00248
BL5BDAXQ00250
Part No.
(L3P07X-46G10)
(L3P07X-46G10)
(L3P07X-46G10)
(L3P07S-46G00)
(L3P07S-46G00)
(L3P07S-46G00)
(L3P06X-55G10)
(L3P06X-55G10)
(L3P06X-55G10)
7.16. Replacement of Incidence
Polarizer (R and B)
11. Reassemble the projector as it was.
7.14. LCD Panel Discrimination
ID-tag seal colorLCD panel
RedLCD panel (R)
BlueLCD panel (B)
(No seal)LCD panel (G)
1. Remove the LCD block according to the section 7.12.
"Removal of LCD Block".
2. Unscrew the 2 screws.
3. Remove the analysis case (upper) while expanding the