-86°C Ultra-Low Freezers
MDF-U5486SC MDF-U7486SC
PRO Series
-86ºC Freezers
Features:
• Energy-saving storage solutions for uniformity, reliability and environmentally safe operation.
• New Cool Safe refrigeration systems designed to deliver energy-saving, high performance cooling.
Panasonic… the new name for SANYO
www.panasonic.com/biomedical
Page 2
MDF-U5486SC MDF-U7486SC
Panasonic PRO Series -86°C Freezers
Designed for long-term storage of stem cells, cord blood, T-cells, engineered tissue, organ/tissue, bone marrow, hybridomas, lymphocytes, cancer cells, clinical specimens, broblasts and other life science samples.
PRO Series Features & Benets
What It Is What It Does Why It Is Important
Energy-Efcient Refrigeration
Cool Safe Compressors
Environmentally Friendly Refrigerants
Integrated Control Center
Microprocessor control over all cooling functions delivers cooling on demand.
Specic to ultra- low applications. Reduces compressor temperatures internally and externally.
Eliminates potential for ozone depletion while maintaining cooling capacit y.
Combines all control, alarm, monitoring and data management functions into a single system.
Optimizes run time to minimize energy consumption.
Increases compressor longevit y and reliability. Reduces heat output to room and lowers HVAC loads in room.
Complies with the Montreal Protocol, IEC, and OSH A for safety and ef ciency.
High visibility LED display provides a convenient user interface to setpoints, alarm parameters, internal diagnostics, communications and securit y.
Structural Enhancement
Compliant to International Standards
Green Design Attributes
Ergonomic Design
Integrates inventory management, access and site installation.
Assures quality standards, safety and per formance criteria are met or exceeded.
Integrates design, construction, refrigeration and operation functions into an environmentally friendly system.
One- handed outer and inner door latches and quiet-running compressors improve convenience, minimize sound.
PRO Series -86ºC Upright Freezer Selection
Model Number
MDF-U5486SC-PA
MDF-U7486SC-PA
Temperature
Range
-50ºC to -86 ºC
-50ºC to -86 ºC
Panasonic is a pioneer in development of high performance commercial products worldwide. As a result, Panasonic products include many innovative features such as, Cool Safe quiet-running refrigeration compressors, and intuitive microprocessor electronics. These innovations are made possible by our exclusive Vertical Component Integration program. This powerful corporate technology transfer capability, together with modern robotic manufacturing and quality assurance processes, guarantee that only the very best components are used in each Panasonic product.
Interior Volume
17.1 cu.ft.
487 L
23.5 cu.ft
668 L
Area Footprint
(nominal)
8.37 sq.ft.
0.78 m
10.63 sq.ft.
0.99 m
2
2
Cabinet design attributes include high- strength, lockable door latch and door, latchable inner doors, adjustable shelves, locking casters and seismic restraints to simplify operation, installation and to satisfy local codes.
Essential for compliance with LEEDs, RoHS, OSHA and other third -par ty standards and recommended practices.
Minimizes carbon footprint, assists building owners in LEED Cer tication best practices.
Easy access to controls, displays, inventory racks, while low noise operation permits a wider choice of installation locations.
Storage
(2"/ 51 mm boxes)
320 192 32,000
480 336 48,000
Storage
(3"/76 mm boxes)
Storage
(2ml vials in boxes)
Electrical,
60Hz
208/230V
NEMA 6 -15
208/230V
NEMA 6 -15
2 PRO Series -86°C Freezers www.panasonic.com/biomedical
Page 3
PRO Series -86ºC Freezer Performance Advantages
Panasonic conventionally insulated PRO Series ultra-low freezers provide energy savings while minimizing carbon footprint throughout the laboratory without compromising performance. The conventionally insulated series also maintains internal temperatures as low as
-86°C. All models use Panasonic-designed compressors for ultra-low temperature applications. Manufactured with foam-in-place insulation to maximize interior temperature uniformity, they are ideally suited for use in hospitals and laboratories for long-term preservation and storage of blood, specimens and components, as well as material testing.
PRO Series ULT Freezers
Panasonic PRO Series -86ºC ultra-low temperature freezers maintain internal temperatures as low as -86°C (-123°F). All models use Panasonic uniquely designed compressors for ultra-low temperature applications. Manufactured with foamed-in-place insulation, they are ideally suited for use in laboratories and hospitals for long-term preservation and storage of blood, specimens and components, as well as materials test­ing. Whatever your preservation needs are, Panasonic provides the right equip­ment for the right application. Panasonic preservation systems employ advanced technology to insure a high precision temperature environment.
Enhanced Performance, Energy Efciency, and Reliability
Panasonic PRO Series upright freezers represent the industry’s most advanced combination of cabinet design, electron­ics, refrigeration and critical components for enhanced security, better perfor­mance, product safety and cost effective operation at -86°C.
• The newly developed energy efcient refrigeration system and sound abate­ment feature which minimizes intrusive operating noise.
• The newly developed compressor deck combined with strategically placed components in the refrigeration compartment provide superior air ow to drastically reduce the stress to the freezer and contributing to excellent durability.
• Two independent sealing and insulated inner doors ensure maximum interior chamber uniformity at all shelf levels
Panasonic Refrigeration Design Delivers Uniform Temperature
The Panasonic cascade refrigeration sys­tem uses Panasonic designed compres­sors for high-performance, reliability and peace of mind. Refrigeration compo­nents are carefully selected and matched for optimum operation under demanding laboratory environments.
• Increased cooling capacity improves temperature recovery after door openings.
• Quieter operation is achieved through condenser fan blade design, noise abatement insulation and anti-vibration mounts.
• Voltage boost and surge protection devices permit dependable operation over wider power ranges and environ­mental conditions.
MDF-U7486 SC
Because modern laboratories are energy-intensive, Panasonic has developed a corporate-
wide energy savings and environmental impact approach to new product development. The PRO Series freezers offer signicant benets through a balance of refrigeration power, cabinet construction and intelligent control over all functions.
Power Consumption by Capacity
3.0
2.5
2.0
1.5
1.0
0.5
Watts
0.0
Panasonic freezers provide reduced operational cost for highly efcient sample storage*. (See page 9)
(KW/day/box Ef ciency)
2.25
2.21
1.73
Panasonic
MDF -U748 6SC
Brand A
Brand B
3
Page 4
ICD Digital
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
Battery
SAVE
VIP Plus Vacuum
Insulation Panel
VIP+
Filterless
Design
VIP Vacuum
Insulation Panel
VIP
VIP Plus Vacuum
Insulation Panel
VIP+
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
Rechargeable
Battery
VIP Plus Vacuum
Insulation Panel
VIP+
Filterless
Design
Insulated
Inner Doors
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
Service
Automatic
Alarm System
Automatic
Alarm System
Rechargeable
Battery
Rechargeable
Battery
Energy
SAVE
VIP Plus Vacuum
Insulation Panel
VIP+
Filterless
Design
VIP Vacuum
VIP
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
VIP
Insulated
Inner Doors
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
Energy
SAVE
VIP Plus Vacuum
Insulation Panel
VIP+
Filterless
Design
Insulated
Inner Doors
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
Service
Battery
Energy
SAVE
VIP Plus Vacuum
Insulation Panel
VIP+
Filterless
Design
ICD Digital
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
Insulation Panel
VIP
VIP Plus Vacuum Insulation Panel
VIP+
Filterless
Design
Insulated
Inner Doors
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
VIP
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
VIP
Insulated
Inner Doors
ICD Digital
Display
Remote Alarm
CPU and
Touch Pad
VIP Vacuum
VIP
Quiet, Reliable
Compressor
PRO Series -86ºC Freezer Construction
Preservation System Features
Inner Doors Improve Uniformity
The center shelf forms the base of a
Cascade Cooling System
convenient and exible inventory system using conventional racks with boxes,
Insulated Inner Doors
microplates or other storage protocols. Adjustable stainless steel shelves create choices of interior compartments
CPU and Touch Pad
that can be arranged for long-term, low access biologicals or short-term,
LED Digital Display
frequently accessed material.
• Easy-In / Easy Out 'Eagle Beak' inner
Automatic Alarm System
door latches feature ergonomic design to seal rmly against the cabinet with one hand.
Power Failure Alarm
• High strength, sealing insulated inner doors help minimize change in
Air Filter
interior temperatures during routine door openings.
CFC-Free
• Two inner doors are standard, creating two compartments that can be sub-
SAVE
Energy Savings
divided with interior shelves.
Quiet, Reliable Compressors
Service
Rechargeable Battery
4 PRO Series -86°C Freezers www.panasonic.com/biomedical
Patented Panasonic refrigerants are non-ozone depleting, non-ammable and environmentally safe in compliance with the Montreal Protocol.
Product Features
1. The insulation used in Panasonic’s PRO Series ultra-low temperature freezers is high density rigid, foamed-in-place polyurethane. This design prevents the insulation layers from distortion and cracking that might occur due to temperature differences inside and out.
2. Ergonomic one handed Easy-In / Easy Out outer door latch will accept a padlock.
3. Universal keyed door lock provides additional level of security.
4. Microprocessor-based controller.
5. Remote alarm contacts can
be connected to independent monitoring system.
6. Circular chart or touch panel digital (mounted on front door) temperature recorder (optional) installs easily in pre-engineered mounting space.
7. Insulated sealing inner doors help maintain interior temperature during routine openings. Two inner doors are standard.
8. 'Eagle Beak' inner door latches.
9. Easy front access to washable,
electrostatic condenser lter for routine condenser air lter cleaning.
10. High impact recessed casters simplify installation.
11. Seismic restraints secure cabinet to building superstructure to meet local codes.
12. Panasonic designed compressors specically designed for low temperature use offer maximum reliability.
13. Multiple access ports permit independent probes, instrumentation or liquid CO back-up system injection.
14. Internal voltage and power management systems assure component protection and best operation efciencies.
15. Vacuum release port.
2
Page 5
PRO Series -86ºC Freezer Features
Designed by Panasonic specically for rugged ultra-low temperature applications in a laboratory environment; CFC-free refrigerants only. Panasonic refrigeration system delivers uniform temperatures with increased cooling capacity. High performance refrigeration system with foam-in-place cabinet insulation maximizes interior temperature uniformity and protects against uctuating ambient temperatures.
1
11
15
2
The Panasonic ergonomic design features a high-security Easy-In/Easy-Out door latch designed for simplicity and safety. The outer door latch closes smoothly with a single-hand action over a cam action bearing. The outer door can be secured with an independent padlock provided by customer.
8
Inner Door Latches provide single-hand access.
7
13
5
3
9
6
12
14
10
4
Panasonic microprocessor-based controller includes comprehensive setpoint, alarm, monitoring and diagnostic functions.
5
Page 6
PRO Series -86ºC Freezer Energy Efcient Cascade Cooling System
By apportioning the oil cooling function between specially designed Cool Safe compressors, and by cooling the compressor oil to minimize compressor operating temperatures, the Panasonic ultra-low temperature freezer refrigeration system is balanced to decrease component stress, increase system longevity and reliability, and improve temperature uniformity necessary for better cell viability regardless of where the specimen is stored within the chamber.
Low Stage Capillary Tube. Liquid refrigerant under pressure is passed through the capillary tube where it ash evaporates in the low stage evaporator to absorb energy (heat) from the product stored in the freezer.
Freezer Cabinet with Evaporator.
The evaporator coil is strategically wrapped around the interior chamber and concealed within the composite wall of vacuum insulation panels and conventional foamed-in-place urethane insulation.
Low Stage Refrigerant. Commonly available worldwide, R508.
Low Stage Heat Exchanger. Energy is absorbed by the refrigerant gas and transferred to the low-stage heat exchanger to cool discharge gas.
Low Stage Compressor. The compressor pumps refrigerant through the low stage circuit.
Interstage Heat Exchanger. Energy is transferred to the high stage circuit.
High Stage Capillary Tube. Liquid refrigerant under pressure is passed through the capillary tube where it ash evaporates in the interstage heat exchanger to absorb energy (heat) from the low stage refrigerant circuit.
Main Condenser and Motor/Fan Assembly. The motor/fan assembly
blows ambient air across condenser coils to move energy (heat) from the high stage refrigerant to the ambient environment.
High Stage Refrigerant. Commonly available worldwide. Selected for optimum cooling performance in compliance with international environmental protection laws.
Air Cooled Pre-Condenser. Removes energy (heat) from the high stage refrigerant enroute to the high stage oil reservoir.
Low Stage Oil Reservoir. High stage refrigerant passes through the low-stage oil sump to cool lubricating oil resulting in high­stage compressor energy being used to minimize the workload on the low stage compressor.
Air Cooled Pre-Condenser. Removes energy (heat) from the high stage refrigerant enroute to the low stage compressor.
Instrumentation (Not Shown). Temperature and pressure sensors throughout the high and low stage circuits transmit information to the Panasonic status alert central controller for operation, monitoring and interpretation.
6 PRO Series -86°C Freezers www.panasonic.com/biomedical
High Stage Compressor. The com­pressor pumps refrigerant through the high stage circuit.
High Stage Oil Reservoir. High stage refrigerant passes through the high stage sump to cool lubricating oil enroute to the low stage compressor through the air-cooled pre-condenser.
Page 7
PRO Series -86ºC Freezer Control, Alarm and Monitoring
The Panasonic microcompressor control system is secure, easy to use and comprehensive. Control inputs are managed with convenient push-pad buttons on a unitized, sealed control overlay. The panel is conveniently located and is easy to view and set.
PRO Series -86ºC Upright Freezer Alarm Systems
Alarm Safety Features Event Visual Audible
High Temperature (Adjustable to +40°C)
Low Temperature (Adjustable to - 40°C)
Power Failure Loss of Power Intermittent tone
Control Temperature Sensor Failure
Filter Sensor Failure
Batter y Life Time
Fan Motor Life Time
Filter Check Clogged or dusty condenser lter FILTER lamp ashes
Auto Return Touch key is not pressed for 90 sec. Reverts to chamber temp. display
Interior chamber warms
beyond high temp setpoint
Interior chamber cools
beyond low temp setpoint
Sensor probe disconnect,
short-circuit or failure
Preventative Bat tery Replacement
Notic ation (≈3 years)
Preventative Replacement Notication
(≈6 years)
Intermittent tone; time
ALARM lamp ashes;
digital display ashes actual
chamber temp.
ALARM lamp ashes Solid tone No
Battery Lamp Flashes None No
delay of 15 min. after
reaching alarm setpoint
avoids false alarms.
Intermittent tone
None No
Signal to
Alarm Contact
Yes
Digital display defaults to actual chamber temperature.
Display mode changes when setpoint, alarm parameters,
programming and diagnostic functions are performed
Adjustable compressor start-up delay sequence delays re-start after building power failure; allows facility power to reach equilibrium to permit smooth compressor start-up
Buzzer silence temporarily
mutes audible alarm
Alarm test veries readiness
of alarm function and high capacity
Ni-MH battery charge
Panasonic exclusive high capacity Ni-MH battery powers control memory and alarm functions during power failure
Press to set temperature; set button is also
used for other diagnostic functions
Filter indicator lamp glows when electrostatic lter requires removal for cleaning
Alarm indicator lamp glows when freezer is in alarm condition; alarm ringback is factory set for 30 minutes, adjustable in 10 minute increments from 10 to 60 minutes
Audible alarm
Setpoint entry
Setpoint entry advances digital display to next position
Alarm setpoints are factory set 10°C above and below setpoint. Alarm parameters are
adjustable by pressing up or down arrow keys and following the prompt
7
Page 8
PRO Series -86ºC Freezer Green Features
Because modern laboratories are energy-intensive, Panasonic has developed a corporate-wide energy savings and environmental impact approach to new product development. The PRO Series freezers offer signicant benets through a balance of refrigeration power, cabinet construction and intelligent control over all functions.
Eco Friendly Technology
• Panasonic freezers are designed to support LEED certication associated with the U.S. Green Building Council recommendations.
• Components are compliant with RoHS directives on the use of hazardous materials in electrical and electronic equipment.
• Noise reduction and operating cost efciencies are integrated into the refrigeration system.
• Heat output is limited to minimize the impact on facility HVAC demands.
• A microprocessor controller oversees the refrigeration system to regulate cooling cycles, reducing energy consumption.
• Panasonic Cool Safe compressor technology for lower compressor heat internally and externally lower HVAC loads.
Panasonic is conscious of the need to protect our environment and conserve energy. As a corporate pioneer in life science laboratory equipment and appliances, and as a global source of solutions ranging from energy manage­ment to solar power and alternative energies, Panasonic remains committed to providing the best possible labora­tory equipment for research and clinical needs.
CFC Free Refrigerants
Panasonic was the rst ultra-low freezer manufacturer to employ non-HCFC R508 low-stage refrigerant, now recognized as today’s industry standard and widely available. This non-proprietary refrigerant is commonly available to refrigeration service professionals on the open market.
The high stage refrigeration system is commonly available to refrigeration professionals on the open market as well.
RoHS Compliance
In 2006, RoHS (Re­striction of Hazardous
Substances) legislation (EU Directive 2003/95/EU) became ef­fective. RoHS relates to the restriction of hazardous substances and reductions in environmental pollution.
Through RoHS legislation the EU and other participating countries are banning toxic substances in electrical equipment such as lead, cadmium, mercury, chro­mium 6+, PBB and PBDE.
Improved cell viability through Panasonic Ultra-Low Refrigeration Technology
8 PRO Series -86°C Freezers www.panasonic.com/biomedical
Page 9
PRO Series -86ºC Freezer Green Features
Energy Efcient
With a capacity of up to 48,000 two-inch vials in divided boxes, the energy-saving advantages of this efcient Panasonic system extend to lower per-unit storage costs regardless of the preferred inventory conguration.
Electrical Standards
All Panasonic products including ultra­low temperature freezers are tested and certied by an NRTL (National Recognized Testing Laboratory) to assure compliance with US and International standards for electrical safety prescribed in 29 CFR 1910.7( c ).
Noise Reduction
Ultra-low freezers are often located within research and hospital laboratories or production facilities. Users prefer close proximity for easy access to valuable stored products.
If operating noise from refrigeration compressors is excessive, and/or compounded by installation of multiple freezers in adjacent locations, the working environment is severely compromised.
Panasonic has included advanced noise abatement in all contemporary ultra-low freezers and noise reduction levels are well below those of competitive freezers. Panasonic PRO series -86 freezers oper­ate at < 51dB(A).
Uniformity Performance
The placement of evaporator surfaces within the cabinet walls achieve excep­tional documented ultra-low temperature uniformity, thereby permitting investiga­tors more freedom in placing valuable cell lines and biologicals within the interior cabinet, and assuring uniform cell viability when harvesting products from the ultra-low archive.
PRO Series Comparative Energy Performance
Power Consumption by Capacity
(KW/day/box Ef ciency)
3.0
2.5
1.73
2.25
2.21
Panasonic
MDF -U748 6SC
Brand A
Brand B
Watts
2.0
1.5
1.0
0.5
0.0
Panasonic freezers provide reduced operational cost for highly efcient sample storage*.
Total Annual Electrical Cost
per Box Capacity
12
10.27
11.11
10.93
Panasonic
MDF -U748 6SC
Brand A
Brand B
10
8
6
4
2
Dollars / Cubic Foot
0
Panasonic freezers emit less heat into the laboratory, minimizing air conditioning costs*.
Dollars ($)
Panasonic freezers help the environment by reducing carbon footprint*.
*Based on internal performance data. Tested in 25°C ambient environment. Freezer cycling at -80°C. Cabinet volume, 25 cu. ft. Average cabinet temperature based on temperature mapping (15 thermocouples).
Annual CO2 Emissions
by Box Capacit y
1.0
0.9
0.8
0.82
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0.89
0.88
Panasonic
MDF -U748 6SC
Brand A
Brand B
9
Page 10
PRO Series -86ºC Freezer Storage Solutions
PRO Series Storage Capacity and Inventory Systems
Inventory Conguration MDF-U5486SC-PA MDF-U7486SC-PA
Rack Positions, Standard Conguration 4 columns / 4 rows 6 columns / 4 rows
Fiberboard Boxes, 2" high (2ml) in Panasonic racks 384 480
Sample V ials, 2ml (2 " box), 100 -cell dividers 32,000 48,000
Fiberboard Boxes, 3" high (4ml) in Panasonic racks 192 336
Sample V ials, 4ml (3" box), 100- cell dividers 19, 200 33,600
Standard Microplate with foil tape, in racks 2016 3024
Standard Microplate with cover lid, in racks 1632 2448
Note: Maximum storage volume varies according to storage vessel, vial conguration, preservation protocol, inventory technique and user preference. Contact Panasonic for detailed information on inventory racks, boxes and storage options.
Fully Loaded Inventory System
Maximum Vial Capacity (Includes full quantity of story racks, boxes and dividers)
Sliding Drawer Inventory Racks, 2" high boxes (2ml sample) 100 cell -divider (16) SDR- 524 -2S100
Sliding Drawer Inventory Racks, 3" high boxes (4ml sample) 100 cell-divider (16) SDR-3 34 -3S100
Individual Storage Rack Systems
Sliding Drawer Sample Racks for Fiberboard Storage Boxes (boxes not included)
Sliding Drawer Rack for 2" high (2ml sample) boxes, holds 5 boxes high x 4 boxes deep SDR-524- N; Capacity 16
Sliding Drawer Rack for 3" high (4ml sample) boxes, holds 4 boxes high x 4 boxes deep
Sliding Drawer Rack for 3" high (4ml sample) boxes, holds 3 boxes high x 4 boxes deep SDR-33 4- N; Capacity 16
Standard Cellular Inventory Racks for Fiberboard Storage Boxes (boxes not included)
Inventory Rack for 2" high (2ml sample) boxes, holds 5 boxes high x 4 boxes deep SUR- 524 -N; Capacit y 16
Inventory Rack for 3" high (4ml sample) boxes, holds 4 boxes high x 4 boxes deep
Inventory Rack for 3" high (4ml sample) boxes, holds 3 boxes high x 4 boxes deep SUR- 334-N; Capacity 16
Fiberboard Storage Boxes and Dividers
Inventory Box, 2" high (2ml sample) with 100 cell-divider B2C-S100 B2C-S100
Inventory Box, 3" high (4ml sample) with 100 cell -divider B3C -S100 B3C- S100
Inventory Box, 2" high (2ml sample) without divider B2C B2C
Inventory Box, 3" high (4ml sample) without divider B3C B3C
Box Divider, 81-Cell D81 D81
Box Divider, 64- Cell D64 D64
MDF-U5486SC-PA
Catalo g No.
MDF-U7486SC-PA
Catalo g No.
(12) SDR- 424-2S100 (12) SDR- 524-2S100
(12) SDR-334-3S100 (12) SDR- 43 4- 3S100
SDR-424 -N; Capacit y 12 SDR-524 -N; Capacit y 12
SDR-33 4- N; Capacity 12 SDR-4 34 -N; Capacity 12
SDR-424 -N; Capacit y 12 SDR-524 -N; Capacit y 12
SDR-33 4- N; Capacity 12 SDR-4 34 -N; Capacity 12
10 PRO Series -86°C Freezers www.panasonic.com/biomedical
Page 11
PRO Series -86ºC Freezer Specications
1196 (47.1)
867 (34.1)
867 (34.1)
PRO Series Facilities
Description
Model Number MDF-U5486SC-PA MD F- U748 6SC -PA
Refrigeration System, HFC Refrigerants, Cool Safe High Performance Compressors
High Stage 1100 W 1100 W
Low Stage 1100 W 1100 W
Voltage Booster, Built-In Standard Standard
Sound Abatement Standard (< 51dB(A)) Standard (< 51dB(A))
Construction
Shelf
Access Port
Refrigerants HFC refrigerants
Alarm High/ low temp, Power failure, Filter check, Batter y check
Remote Alarm Contact N.O. and N.C., allowable contact capacity: DC30V, 2A
Accessories 1 set keys, 1 scraper
Stainless steel, adjust able 3 shelves
(23. 9" w x 21" d) max. load: 110lbs (50kg) / shelf
.06 6" (17 mm) diameter,
2 locations (back /bottom left)
PRO Series Dimensions
Model Number
MDF-U5486SC-PA
MDF-U7486SC-PA
956 (37.6)
48
(1.9)
890 (35)
630 (24.8)
Volume
(cu.ft.)
17.1
487 L
23.5
668 L
MDF-U5486SC
18 (0.7)
1540 (60.6)450 (17.7)
1990 (78.3)
Footprint
(sq.ft.)
8.37
0.78 m
10.63
0.99 m
130 (5.1)130 (5.1)
1280(50.4)
Exterior Dimensions
35.0 " x 34.3" x 78. 3"
2
890 x 870 x 1990 mm
44.5" x 34.3" x 78. 3"
2
1130 x 870 x 1990 mm
795 (31.3)
605.5 (23.8)
(w x f-b x h)
105
(4.1)
140
(5.5)
8
(0.3)
1625 (64)
Interior Dimensions
(w x f-b x h)
24.8 " x 23.6 " x 50.4"
630 x 6 00 x 1280 mm
34.2" x 23.6" x 50.4"
870 x 60 0 x 1280 mm
48
(1.9)
Temperature
Range
-50ºC to -86 ºC
-50ºC to -86 ºC
1130 (44.5)
870 (34.3)
Voltage, Power
MDF-U7486 SC
18
(0.7)
1280 (50.4)
1540 (60.6)450 (17.7)
1990 (78.3)
Connection
208/230V
NEMA 6 -15
208/230V
NEMA 6 -15
795 (31.3)
605.5 (23.8)
130 (5.1)130 (5.1)
Net Weight
(empty)
672 lbs
305 kg
783 lbs
355 kg
105
(4.1)
140
(5.5)
8
(0.3)
1625 (64)
11
Page 12
Pro Series -86ºC Freezer Options
PRO Series Accessories
Accessory Description Catalog Number
Digital Temperature Recorder Auxiliary Data Logger HAMSTERDT2
Liquid CO
Liquid N
Chart Recorder Circular Chart Temperature Recorder, 7 Day MTR-C954
Chart Paper 6" Diameter, 7 Day Chart C710 03 86 RE V
Replacement Pen, Red Felt tip pens, 6 per pack R252
Replacement Pen, Blue Felt tip pens, 6 per pack R253
4-20 MA Output Independent Analogue Output MTR-420MAC
Touch-Screen Temperature Recorder Full Color Touch-Screen Digital Display FT- 56 0SN
Back-Up System Auxiliary tank back in event of power failure CVK- UB2(I) -PW
2
Back-Up System Auxiliary tank back in event of power failure CVK-UBN2- PW
2
Inventory Rack Systems
Panasonic freezers are available with a variety of inventory racks to meet specic applications. Freezers can be ordered with full-load inventory systems by selecting one catalog number. For individual racks, boxes and dividers see the Storage Capac­ity Chart.
Panasonic's sliding drawer rack with berboard boxes, Catalog Number SDR-624, shown here.
Circular Recorder
(MTR-C954)
Strip Chart Recorder
(MTR-85H)
Recorder Mounting Bracket
(MDF-S3085)
(for MTR-85H)
The Digital Temperature Recorder (optional)
The recorder mounts on the cabinet door and is programmable to capture and store non-corruptible data points for download­ing to an independent data management system to meet criteria outlined in U.S. FDA21CFR Part II directives. An indepen-
dent probe is placed in the interior chamber.
Manufactured by Elpro, Switzerland.
LCO2 Backup System
(CVK-UB2 [I])
LN
2
Backup System
(CVK-UBN2)
Touch-Screen Display (optional)
A touch-screen digital display with full color graphing capabilities of your complete tem­perature history is at your ngertips with scrolling and zoom.
The touch screen monitor ts on the front panel of your Panasonic equipment either in place of the recorder or at eye level and is designed to work without software. Software is optionally available if you wish to download your data to PC via USB or ash memory for archiving.
Touchscreen Catalog Number FT-560SN shown here.
Product conforms to RoHS (European Restriction of Hazardous Substance directives)
Panasonic Healthcare Company of North America
1300 Michael Drive, Suite A, Wood Dale, IL 60191 Toll Free USA (800) 858-8442, Fax (630) 238-0074 www.panasonic.com/biomedical
© Panasonic Printed in USA 2012.02 JB JB201202V1