palintest Turbimeter Plus Instruction Manual

Turbimeter Plus
Turbimeter Plus
See page 1
Instruction Manual
Manual de
instrucciones
ZIINST57ENES
Palintest are committed to making water analysis technologies simple and accessible.
A rich history of innovation
USA UK Middle East China Australia
A global company with a local approach
Years of research
75
Extensive online library, we are proud to share our research.
About us
From multiparameter photometer kits to visual test Palintest has an instrument solution for every application.
Instrumental
Start Up Guide
4
Instrument Overview 5 Battery Life Replacement 7 Guidelines for Use 9
Main Menu
11
Read Menu
12
Normal 14 Average 15 Continuous-Capture (Indexing) 16 Total Suspended Solids (TSS) Measurement 18
Calibration Menu
20
Calibrate/Check Mode 21 Check Standard Mode 22 SDVB 24 Formazin 26 Total Suspended Solids (TSS) 28
Restore Factory Calibration 29
System Settings
30
Log 32 Operator ID and Sample ID 33 USB Connection 36 Units of Measurements, Language 39 Time & Date, Version/Serial Number 40 Backlight, LCD Contrast 42
Appendices
43
Error Messages, Support and Warranty 43 Care and Maintenance, Guidelines for Storage
44
Spares & Consumables, Technical Specification 46
Contents
Turbidity and Total Suspended Solids (TSS) are two of the most important indicator parameters for water quality monitoring.
The new Turbimeter Plus has been developed with leading water utility companies and provides the ideal testing platform for drinking water, wastewater and safe water monitoring.
Fast and reliable results within 5 seconds
Specifically designed for field use with storage for 1000 results and battery power for approximately 10,000 tests
Unique Quadoptix technology with four individual measurement points for increased reliability
Variety of operating and reading modes including TSS, Average and Continuous Capture
USB connectivity for power, downloading data and installing software upgrades
IP67 certified with a waterproof USB socket
1
Introduction
The Palintest Turbimeter Plus utilises QuadoptiX™ technology for turbidity measurement.
QuadoptiX technology uses two independent sources and two independent detectors to provide four entirely autonomous measurement systems, allowing multiple validation of all results for greater accuracy.
Turbidity measurements are carried out by analysis of light scattered at 90° to the incident light (nephelometric measurement) at levels below 40 NTU. Above 40 NTU the recommendation is to measure also at 180° to 'compensate’ the 90° readings.
2
Quadoptix™ Technology
LED
sensor
sensor
LED
LED
sensor
LED
sensor
Please read the entire manual prior to setting up or operating this equipment.
Make sure that the protection provided by this equipment is not impaired, do not use this instrument in any manner other than specified in this manual.
The instrument is IP67 certified, including the USB port, however it is recommended that the Turbimeter Plus is clean and dry before use.
Review the ‘Care and Maintenance’ section for recommendations on best practice.
3
Safety Information
4
Start Up Guide
Turbimeter Plus
1
5
2
4
3
6
5
Front of product
1
1
2
3
4
5
Power
Left
Right
Up
Down
Press the power button and release. An audible beep will confirm the instrument is on. Switch off by holding the POWER button for 1 second.
To Switch Turbimeter Plus On
Start Up Guide
6
Sample Holder
The LCD screen is provided with a backlight option that can be set to on/off in the ‘System’ menu (see Page 42). The initial default setting is ON.
6
Back of product
Made in UK
12345678910
7
8
9
7
8
9
Battery
USB
Serial number
LCD Screen
The Turbimeter Plus is supplied with batteries that will provide power for approximately 10,000 tests. The instrument can also be powered through the USB port.
An indication of remaining battery life is displayed on screen with the battery status icon. To save power, the instrument will switch off after five minutes of inactivity.
When battery life is very low, the instrument will display the following error messages:
7
Replacing batteries will not lead to settings or calibration data being lost. All user data is retained.
Battery Life Replacement
Message Error To solve
Error 110
Battery Low
Instrument has insufficient power to perform reading. Menu functions still available.
Error 111
Battery Critical
Insufficient battery to power instrument. Instrument will save data and shut down.
Made in UK
12345678910
To replace the batteries turn over the instrument and remove the battery compartment cover, insert two AA 1.5V alkaline batteries and replace the compartment cover.
Icon Description
Information Panel Icons
8
DISK / COM
Battery status
USB connected
Hard drive/COM port mode
When replacing batteries, ensure the gasket seal is correctly fitted to maintain the IP67 rating of the instrument. A spare gasket is included with your instrument and should be used if the gasket looks to be damaged, stretched or twisted.
Ensure the sample tube cap is
in place during measurement.
The following guidelines for sample handing should be followed to ensure accurate and consistent results are obtained
.
Use only clean sample
tubes and rinse with sample
before use.
Remove all marks and fingerprints from the sample tube before measuring
.
9
Guidelines for Use
1
0
m
l
1
0
m
l
Guidelines for storage on page 45.
Heavily scratched or contaminated sample tubes should be discarded.
Remove bubbles from sample tube wall. Hold the capped tube at an angle and rotate.
To reduce the effect of scratches
on the glass, place a drop of
silicone oil onto the
tube and wipe using the lint
free polishing cloth.
Place the sample tube in
the instrument in the same
orientation for all readings
.
10
1
0
m
l
1
0
m
l
Select the read mode to measure turbidity or suspended solids.
1
Turbimeter
Mode
OK
Calibrate/ Check
System
Read
01/01/2017 09:30:30
User 1
123456
1. Read
3
2
Select this mode to check or set the instrument calibration.
2. Calibration/Check
General system and instrument settings.
3. System
Main Menu
11
See page 13
See page 20
See page 30
12
Read Menu
Turbimeter
Mode
OK
Calibrate/ Check
System
Read
01/01/2017 09:30:30
User 1
123456
From the main menu select ‘Read’ mode and you will be given the following options:
This mode reads the turbidity of the sample in approximately 5 seconds. See page 14
Normal
This mode takes multiple measurements and averages the results. This mode supports measurements where sample turbidity may be changing over time, for example due to settling of particles. See page 15
Average
This mode performs repeated normal readings of sample turbidity until measurement is stopped by the user. This mode will support sample tube indexing and trending turbidity on settling samples. Trending arrows show the direction of turbidity change. See page 16
Continuous Capture
This mode will measure the site specific relationship between turbidity and suspended solids.
Total Suspended Solids (TSS)
13
Read
See page 18
Select the ‘Read’ menu and highlight ‘Normal’.
Turbimeter
Reading Mode
Read
Average
Continuous-Capture
Normal
Back
01/01/2017 09:30:30
User 1
123456
Insert the sample with the orientation mark ◊ facing forward and press ‘Read’. An audible tone will sound at the start and end of measurement.
Result is displayed on-screen and is automatically stored in the results log with date, time, Reading Mode, Operator ID and Sample ID.
Turbimeter
Normal
Read
01/01/2017 09:30:30
User 1
123456
2.21
NTU
Back
1
3
2
14
Reading Mode - Normal
15
This mode takes multiple measurements and averages the results. Three options are available: Short (3 readings), Medium (6 readings) and Long (12 readings).
Turbimeter
Reading Mode
Read
Continuous-Capture
Normal
Back
01/01/2017 09:30:30
User 1
123456
Average
Select the number of readings required: Short (3), Medium (6), Long (12). Insert the sample with the orientation mark ◊ facing forward and press ‘Read’.
An audible tone will sound at the start and end of measurement. Result is displayed on the screen and stored in the log with time, date, Reading Mode, Operator ID and Sample ID.
Turbimeter
Short (3)
Read
01/01/2017 09:30:30
User 1
123456
2.21
NTU
Back
1
3
Reading Mode - Average
Turbimeter
Average Reading
Read
Medium (6) Long (12)
Short (3)
Back
01/01/2017 09:30:30
User 1
123456
2
Reading Mode - Continuous-Capture
This mode performs repeated Normal readings of the sample until stopped by the user.
Turbimeter
Reading Mode
Read
Normal
Back
01/01/2017 09:30:30
User 1
123456
Average
Continuous-Capture
1
0
m
l
Insert the tube with the orientation mark ◊ facing forward and press read. An audible tone will sound after each reading.
Results are not stored automatically. To store a result in the log, press ‘Capture’ (repeat as required). To stop measurement, press ‘Stop’ at any time.
Turbimeter
Continuous-Capture
Capture
01/01/2017 09:30:30
User 1
123456
2.21
NTU
Stop
1
3
2
16
The lowest turbidity result returned corresponds to the optimal sample tube orientation. Mark the tube in this position to indicate the optimum orientation. See page 44 for more information.
5
17
To index a sample tube:
In continuous capture mode, insert the tube with the orientation mark ◊ facing forward.
Lift and rotate the sample tube by a fixed amount (e.g. 45°) in between measurements for a full 360°.
4
Indexing a Sample Tube
Lift and rotate
Reading Mode - Total Suspended Solids (TSS) Measurement
This mode will only be active if a site calibration has been stored within the Turbimeter. Once a calibration has been created, the Sample ID can be used to recall the site specific relationship between turbidity and suspended solids.
Turbimeter
Reading Mode
Read
Average
Back
01/01/2017 09:30:30
User 1
123456
Continuous-Capture
Total Suspended Solids
Select the Sample ID for the site/sample. The calibration data will be recalled automatically. Insert the sample with the orientation mark ◊ facing forward and press ‘Read’.
An audible tone will sound at the start and end of measurement. Result is displayed on the screen in mg/l and stored in the log with time, date, Operator ID and Sample ID.
Turbimeter
Total Suspended Solids
Read
6
mg/l
01/01/2017 09:30:30
User 1
123456
Back
1
3
2
18
20
Calibration Menu
Calibrate/Check Mode
Turbimeter
Mode
OK
System
Read
01/01/2017 09:30:30
User 1
123456
Calibrate/ Check
Select ‘Calibrate/Check’ from the main menu and choose from the following options:
21
For verifying the instrument calibration.
Check Standard Mode
For field or laboratory calibration using SDVB standards.
SDVB
For laboratory calibration using prepared Formazin standards.
Formazin
For creating the site specific calibration between TSS and turbidity.
Total Suspended Solids (TSS)
See page 22
See page 24
See page 26
See page 28
In check standard mode There are two options for selection:
Check Standard Values Check Standard Read
Turbimeter
Check Standard Mode
OK
Back
01/01/2017 09:30:30
User 1
123456
Check Standard Read
Check Standard Values
Turbimeter
Enter Check Standard Values
OK
Back
1.00
Use the up/down keys to adjust the displayed values to match the assigned target for each check standard.
22
Check Standard Mode
Check Standard Care Points:
Check standards are manufactured to precise values/tolerances and provided in sealed sample tubes. Do not remove the sealed cap from the check standard.
Use the lint free polishing cloth to ensure the check standard tubes are clean before they are inserted into the sample holder.
Check Standard Values
Turbimeter
Check Standard Mode
OK
Back
01/01/2017 09:30:30
User 1
123456
Check Standard Read
Check Standard Values
Turbimeter
1.00 NTU
Read
Insert Standard
Press ‘Read’
01/01/2017 09:30:30
User 1
123456
Cancel
Select the Check Standard Read option to perform a recorded reading of each turbidity check standard.
‘Pass’ will appear on­screen if the result is within the tolerance displayed on the standard. If the result is outside of the tolerance, the instrument may need to be re-calibrated.
Turbimeter
1.00 NTU
Next
Pass
01/01/2017 09:30:30
User 1
123456
Back
NTU
2.13
1 Check Standard Read
3
23
Follow the on-screen prompts and insert the Check Standards as requested. Select the Read button to perform the measurement.
2
24
For field use, a set of SDVB (styrene divinylbenzene) calibration standards are provided to check the factory calibration and recalibrate the meter if necessary. Only SDVB standards from Palintest can be used with the Turbimeter Plus. Note: Recalibration is only required if the instrument does not pass the check standard mode.
Turbimeter
Calibrate/Check
OK
Check Standard Mode
Back
01/01/2017 09:30:30
User 1
123456
SDVB
Formazin
Choose ‘SDVB’ from the calibration menu. Insert the lowest value turbidity calibration standard when prompted.
Use the Up/Down buttons to adjust the value on screen to that on the standard and press ‘Read’.
Turbimeter
SDVB
Read
Insert Standard
Back
01/01/2017 09:30:30
User 1
123456
1.00
and press Read
2
1
SDVB
An "Operator ID" will be logged with a recalibration, this can be changed in 'System Settings' before recalibrating.
If calibration is unsuccessful, the instrument will display ‘Error 101’ for one of the following reasons:
Standards inserted in incorrect sequenceStandards have expiredStandards are scratched or damaged
Continued...
The instrument will request the remaining standards in order of increasing turbidity value. After the final standard has been read, the screen will show a countdown timer.
Turbimeter
SDVB
Read
and press read
16.0
Back
Insert sample
When the calibration process is complete
the screen will show ‘Successful’. The time, date and Operator ID for successful calibrations are stored in the calibration log.
3
25
SDVB Standards should not be inverted or shaken before use. If they have been inverted or the kit containing them has recently been moved they will need to be allowed to stand for at least 10 minutes before being used.
26
Turbimeter
Calibrate/Check
OK
SDVB
Back
01/01/2017 09:30:30
User 1
123456
Formazin
Total Suspended Solids
From the Calibrate/ Check Mode chose Formazin.
( User-prepared Formazin
standards can also be used for calibration where available)
Formazin standards should be prepared within the following ranges:
First Standard
10.0 – 30.0 NTU
Second Standard 90.0 – 110.0 NTU
Third Standard 720.0 – 880.0 NTU
Formazin
1
Insert the lowest turbidity value calibration standard when prompted. Use the Up/Down buttons to adjust the value on screen to that on the standard and press ‘Read.
The instrument will request the remaining standards in order of increasing turbidity value. After the final standard has been read, the screen will show a countdown timer. When the calibration process is complete the screen will show ‘Successful’.
Turbimeter
Formazin
Read
and press Read
10.0
01/01/2017 09:30:30
User 1
123456
Back
and press Read
Insert Standard
3
Continued...
27
The time, date and Operator ID for successful calibrations are stored in the calibration log. To manage the calibration log see page 38.
Turbimeter
Formazin
Read
and press Read
System
0.40
01/01/2017 09:30:30
User 1
123456
Back
and press Read
Insert Standard
2
28
The Turbimeter Plus can store 50 user calibrations for Total Suspended Solids. These can either be entered by providing TSS values for results already in the log or by entering a definition (slope and intercept) for the calibration to be used.
Total Suspended Solids (TSS)
Use Results From Log
Select the appropriate Sample_ID or scroll to the bottom to create a new one.
Select 'Use Results from Log'. The default (0, 0) automatically appears, this is normally used as the first data point but can be edited or deleted.
To recall a result from the log, press 'Down' then ‘Add’ and scroll through the log to select a value.
Repeat step 3 to add further results, or when data entry is complete, press ‘Done’. The data is used to generate slope and intercept values, which are automatically saved and displayed on screen.
1
2
3
4
Only the slope and intercept are saved. Individual data points are not held in memory.
Turbimeter
Calibrate/Check
OK
Formazin
Back
01/01/2017 09:30:30
User 1
123456
Total Suspended Solids
Restore Factory Cal
Select ‘Calibrate/Check’ from the main menu then ‘Restore Factory Calibration’. Select ‘YES’ to confirm the restoration.
Restore Factory Calibration
29
A slope and intercept will first need to be determined from externally existing data.
Select the Sample ID from the list or scroll to the bottom of the list to create a new one.
Press ‘Edit’ to enter a value for Slope, and repeat for Intercept. The saved slope and intercept are displayed on screen.
Use Slope and Intercept
1
2
3
Measurement of TSS by calibration of turbidity results are indicative. Changes in the nature of the sample will affect the accuracy of results.
30
System Settings
31
System Mode
Turbimeter
Mode
OK
Read
01/01/2017 09:30:30
User 1
123456
Calibrate/ Check
System
Select ‘System’ from the main menu and choose from the following options:
Set Time and Date, Version
See page 40
Units of Measurements, Language
See page 39
USB Connection
See page 36
Sample ID
See page 34
Operator ID
See page 33
Log
See page 32
Backlight and LCD Contrast
See page 42
32
Turbimeter
Log
OK
Results
01/01/2017 09:30:30
User 1
123456
Calibrations
Data is stored in chronological order. When the log is full, the oldest stored result will automatically be overwritten
From the main menu select ‘System mode’ then ‘Log’ where the ‘Results’ and ‘Calibrations’ logs are stored. In both sections there is a choice to either view or clear data.
Log
The results log holds up to 1000 data points. Each log entry records date, time, Sample ID, Operator ID, Reading Mode and result for each measurement.
The log can be viewed on the device or downloaded to a PC as a text file using the USB connection.
To access operator ID from the home screen press ‘System’ mode then ‘Operator ID’. Up to 12 Operator IDs can be created. When selected, the ID will appear in the info panel on screen as well as in the log.
To create a new Operator ID
Scroll to the bottom of the list, then select ‘New’ to create an ID.
Note: The screen will be blank on first use.
Turbimeter
Operator
New
Back
Turbimeter
System Mode
OK
Log
Back
01/01/2017 09:30:30
User 1
123456
Operator ID
Sample ID USB
33
Use the Up/Down buttons to select characters. When the correct character is shown, press and hold the [+] key to move to the next character. To delete characters press and hold the [Del] key. Select ‘Done’ to save the ID.
Turbimeter
New Operator
[+]Done
A B
System
2
3
1. Operator ID
34
Up to 24 Sample IDs can be created.
Turbimeter
System Mode
OK
Log
Back
01/01/2017 09:30:30
User 1
123456
Operator ID
Sample ID
USB
Select the ‘System’ mode then ‘Sample ID’.
The Sample ID is required to generate TSS (Total Suspended Solids) calibrations.
To delete characters press and hold the [Del] key. Select ‘Done’ to save the ID.
To create a new Sample ID
Scroll to the bottom of the list, then select ‘New’. Use the Up/Down buttons to select characters. When the correct character is shown, press and hold the [+] key to move to the next character.
Turbimeter
Sample ID
New
Back
Turbimeter
Sample ID
[+]Done
A B
System
2
3
1. Sample ID
Turbimeter
Sample ID
OK
Sample 1
01/01/2017 09:30:30
User 1
123456
Sample 2
Back
Edit or Delete a Sample ID
35
Select ‘OK’ on the Sample ID to be used. When selected, the ID will appear in the information panel on screen as well as in the log when results are stored.
Highlight the ID and select ‘OK’. The screen shows the option to select either ‘Edit’ to change the ID, or ‘Delete’. Deleting the sample ID will not affect results stored in the log.
To record results without a Sample ID. Select a blank entry by scrolling to the bottom of the list and select ‘Back’.
Continued...
Turbimeter
Sample ID
OK
Sample 1
Sample 2
Back
Turbimeter
Sample ID
OK
Sample 1
01/01/2017 09:30:30
User 1
123456
Sample 2
Back
4
5
36
To access USB from the home screen press ‘System’ mode then ‘USB’.
There are 2 options:
Turbimeter
System Mode
OK
Log
Back
01/01/2017 09:30:30
User 1
123456
Operator ID Sample ID
USB
Use ‘Hard Drive’ mode for to downloading results and calibration logs or updating the instrument firmware.
Hard Drive
In COM Port Mode results are streamed across the USB connection in real time using a serial communication format (emulating RS-232). In this mode of operation, Palintest COM Port drivers are required. Please contact Palintest to access the drivers for installation
.
COM Port
See page 38
USB
37
To download results and calibration logs, or to update the instrument firmware the device must be in ‘Hard Drive’ mode.
From the home screen press ‘System’ mode then ‘USB’ then ‘Hard Drive’.
Turbimeter
USB
OK
Com Port
Hard Drive
Back
1. Hard Drive
Once the device is in ‘Hard Drive’ mode connect to a PC, using the supplied USB cable.
2
38
When the instrument is connected it will be accessible as a removable drive, containing two log files:
CAL LOG
The calibration data log file.
LOG
The results data log file.
The log files should be copied to the PC before being opened or edited in a text viewer or other application. The CAL LOG will only appear if a calibration has been completed.
This data can be imported in to most common spreadsheet applications by using the import wizard and setting the columns as ‘tab’ delimited.
Turbimeter_Plus_XXX_YYY
This is the operating firmware of the instrument which can be upgraded if required. XXX & YYY represent the major and minor release version numbers of the firmware file (e.g Turbimeter_Plus_001_001).
Please ensure your device is up to date with the latest firmware releases to ensure the best performance from your device. Any upgrade files will be made available by Palintest.
3
Results can be displayed in:
NTU -Nephelometric Turbidity Units
FTU -Formazin Turbidity Units
FNU -Formazin Nephelometric Units
FAU -Formazin Attenuated Units
39
To change units which results are shown in, from the home screen press ‘System’ mode then ‘Units’.
Turbimeter
System Mode
OK
USB
Sample ID
01/01/2017 09:30:30
User 1
123456
Language
Units
Back
To change the language of your device from the home screen press ‘System’ mode then ‘Language’. Choose from English, French or Spanish.
Turbimeter
System Mode
OK
USB
Sample ID
01/01/2017 09:30:30
User 1
123456
Language
Units
Back
Units
Language
40
Use up/down keys to adjust the hour; press the right key to select and adjust the minutes. The time is displayed in 24Hr format.
Turbimeter
Set Time
OK
01/01/2017 09:30:30
User 1
123456
Set Time
09:30
Back
Language
Set Date
Turbimeter
System Mode
OK
01/01/2017 09:30:30
User 1
123456
Set Time
Units
Back
Language
Set Date
From the main menu select ‘System' mode then ‘Set Time’.
2
Turbimeter
System Mode
OK
Set Time
Back
Language
Set Date
Version
LCD Contrast
From the main menu select ‘System mode’ then ‘Set Date’.
Set Date
1. Set Time
Use the up/down keys to adjust each value in the date display, using the left/right keys to select the field. The date format can be set to either DD/MM/ YY or MM/DD/YY
41
Turbimeter
Set Date
OK
01/01/2017 09:30:30
User 1
123456
Set Time
01/01/17
Back
Language
Set Date
2
From the main menu select ‘System' then ‘Version’.
The serial numbers of the instrument and the software version are displayed.
Turbimeter
System Mode
OK
Version
Back
Date Format
Backlight
Version
LCD Contrast
To change the date format select ‘System Mode’ from the main menu then ‘Date Format’. DD/MM/ YYYY OR MM/DD/YYY.
Version/ Serial Number
42
Turbimeter
System Mode
OK
Version
Back
Date Format
Backlight
Version
LCD Contrast
The instrument display features a high intensity backlight for use in low light conditions. Select either ‘On’ or ‘Off’.
Turbimeter
2
The LCD screen contrast can be manually adjusted. Use the up/down buttons to adjust the contrast until the alternating shapes are clearly visible and press ‘Save’.
Turbimeter
System Mode
OK
Version
Back
Date Format
Backlight
Version
LCD Contrast
From the main menu select ‘System mode’ then ‘LCD Contrast’.
LCD Contrast
1. Backlight
Message Error To solve
Error 100
Measurement The reading is above 1050 NTU.
Error 101
Incorrect Standard
The standard inserted is either the incorrect value or is no longer within an acceptable tolerance for calibration. Use the correct standard or replace the calibration set.
Error 107
Light Detection
The sample tube cap is not in place or the sample tube is not fully inserted. Replace the cap and read again or fully insert the sample tube. If the error persists, set the zero reference level by pressing the ‘Ref Reset’ whilst a capped tube is in the holder
Error 110
Battery Low Instrument has insufficient power to
perform reading. Menu functions still available.
Error 111
Battery Critical
Insufficient battery to power instrument. Instrument will save data and shut down.
The Turbimeter Plus has a two year warranty.
Calibration standards should not be used after their stated expiry date. Replacement sets are available with part code PTC 092.
For technical and product support please visit
www.palintest.com
or contact Palintest at
support@palintest.com
43
Technical Support and Warranty
Error Messages
The Turbimeter Plus is IP67 compliant, so contact with water will not damage the internal components. It is recommended that the instrument is dry prior to use. The USB connector and cable must be dry before use. The instrument does not require regular cleaning for normal operation. If required, clean the instrument with a dust and lint free cloth. Do not clean the Turbimeter Plus with solvents or other cleaning agents.
Sample Tube
The sample tube should be clean and dry prior to insertion into the sample holder. Lint free polishing cloths are provided for this purpose. Small scratches on the sample tube can cause discrepancies in the test results. To reduce the effect of these scratches, place one drop of silicone oil onto the tube and wipe across the glass using the lint free polishing cloth. Sample tubes should be rinsed with distilled water. Use a soft cloth to dry the sample tube. Do not air dry the sample cells as this may leave marks or residue on the surface.
Indexing a Tube
Indexing is used to identify the optimum orientation of the sample tube, particularly when taking measurements of low turbidity (<1NTU). Our continuous capture mode has been designed for this
use, instructions shown on pages 17.
Check and Calibration Standards
SDVB turbidity standards are provided in the kit. Do not use the standards after the stated expiry date. Exposure to freezing temperatures can affect the quality of the standards. Do not store the standards at temperatures close to 0°C. The use of a heat pad is recommended for shipping. Replacement SDVB turbidity standards are available from Palintest and authorised
distributors.
Care and Maintenance
44
45
 When storing the unit in the case, always keep a clean, dry tube with its cap on in the optical cell.
 Ensure all glassware is kept clean and dry before storage.
 If there is water in the optical cell this needs to be removed before replacing it in the case. (Be especially careful that dirt is not wiped into the optical windows – consider using a cotton bud or similar to remove water without touching the sides of the optical cell)
 Do not keep damp cleaning cloths inside the case.
 Be especially careful when changing the batteries to place the sealing gasket over the raised flange on the battery compartment cover before replacing.
 If any moisture gets into the case keep the lid propped open until it dries out thoroughly.
 Never attempt to wipe the recessed optical lenses or the areas around these recesses. Dirt or debris pushed into the recessed window will cause erroneous readings or error messages to be displayed.
Guidelines for storage
Replacement Sample Tubes
PTC555
USB Cable
PT126
Replacement Calibration Kit
PTC092
Silicone Oil
PT120
4000 NTU Formazin Stock Solution
PT121
Lint Free Polishing Cloth
PT399
Sample Tube Brush
PT663
Battery Compartment Gasket and Screws
PT123
Battery Compartment Kit (Cover, Gasket and Screws)
PT125
Replacement Gaskets For Battery Cover, Pack of 5.
PT124
Absorbent cloths, Pack of 5
PT122
46
Spares and Consumables
Optical System
Instrument Type QuadoptiX™ technology for turbidity measurement dual light source.
IP Rating
IP67
Optical Source
Dual infrared LED sources 860nm with optical focusing of beams
Optical Detectors
Photodiodes 750nm to 1100nm
Ranges
0.01 – 1050 NTU
Accuracy
± 2.0% NTU
Display
240 x 160 pixel LCD with contrast adjustment
User Interface
On-screen prompts available in English, French and Spanish
Result Units
NTU, FNU, FTU, mg/L (TSS mode)
Modes
Turbidity (normal, average, continuous capture), Total Suspended Solids
Data Storage
1000 results including date, time, sample ID, operator ID and mode. Last 12 good calibrations including Operator ID
47
Technical Specification
Size (W x L x H) & Weight
82 x 225 x 50mm, 340 g
Power Supply
2x 1.5V ‘AA’ alkaline batteries
Battery Life
10,000 tests (typical use, backlight off, ‘AA’ alkaline cells) and auto-off
Mains
5V DC, 900mA delivered via USB port
Power Management
Auto switch off or continous
Power Saving
User control for backlight
USB
Micro- USB type B connector
Instrument Memory
Non-volatile storage
Memory Capacity
Up to 1000 data sets. Each data set includes date, time, Sample ID, Operator ID, method number, method name, result, units
Sample IDs
Up to 24 at any time
Operator IDs
Up to 12 at any time
Data Output Format
Plain text
Software Upload
Software update by ‘drag and drop’ in USB Hard Disk Mode.
48
Loading...