PAJ WIO 200B User Manual

PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 1 af 53
WIO 200B
Water in oil sensor
User manual Rev. 1.08
111881-944 Rev. 1.08 WIO200B User manual
Date: 04-02-2016
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 2 af 53
Table of contents
Table of contents ......................................................................................................................................................................................... 2
Introduction ...................................................................................................................................................................................................... 3
Safety Advice .................................................................................................................................................................................................. 4
Function ............................................................................................................................................................................................................... 6
Normal operation ......................................................................................................................................................................................... 7
Failure indication .......................................................................................................................................................................................... 8
Self test for alarm functions ............................................................................................................................................................... 9
Before installation .................................................................................................................................................................................... 11
WIO Sensor dimensions in mm..................................................................................................................................................... 14
Terminal box dimensions in mm .................................................................................................................................................. 16
WIO Integrated (WIOI) dimensions in mm ........................................................................................................................ 17
Technical data, Sensors ...................................................................................................................................................................... 19
Technical Data Terminal Box .......................................................................................................................................................... 22
Technical Data Terminal Display Box ....................................................................................................................................... 23
Technical Data Terminal Box Alarm .......................................................................................................................................... 26
Technical Data Terminal Box Alarm Buzzer ........................................................................................................................ 29
Technical data, WIOI (WIO Integrated) ................................................................................................................................ 33
EC-Declaration of Conformity ......................................................................................................................................................... 39
Appendix A1: Install 1 Terminal Box with 1 display ....................................................................................................................... 42
Appendix A2: Install 1 Terminal Box with no display .................................................................................................................... 43
Appendix A3: Install 2 Terminal Boxes with 2 display .................................................................................................................. 44
Appendix A4: Install 2 Terminal Boxes with 1 display .................................................................................................................. 45
Appendix A5: Install 2 Terminal Boxes with 1 display .................................................................................................................. 46
Appendix A6: Install 3 Terminal Boxes with 3 display .................................................................................................................. 47
Appendix A7: Install 3 Terminal Boxes with 2 display .................................................................................................................. 48
Appendix A9: CABLE Specification ............................................................................................................................................. 49
Appendix A10: Error / failure FAQ .............................................................................................................................................. 50
Company Information ........................................................................................................................................................................... 53
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 3 af 53
Introduction
This user manual provides technical data regarding the WIO (water in lubrication oil, gear box oil, diesel oil, hydraulic oil or transformer oil) sensor system. The system consist of a WIO Sensor, terminal box and cables. The WIO Sensor is an insitu sensor which measures water content in lubrication oils on ship engine or similar. The Terminal box connects the sensor to power supply and to the ships surveillance system. The supplied cables are for connection between the WIO Sensor and the Terminal box
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 4 af 53
Safety Advice
For your safety
The WIO200 (WIO) is designed as water in oil monitoring system, with a relay switch-off function.
Before use please read the user manual and carefully store it in a safe place. Install and operate the device only after reading and comprehending the instruction manual, and after you are familiar with the valid rules on work safety and accident prevention.
Please use the device only as specified. For this purpose please also note the values in section “Technical data”.
During transport, storage and operation please adhere to the conditions
listed in “Technical data”.
IMPORTANT
Only qualified personnel may assemble start-up and maintain the unit when it is in a zero voltage state. Only qualified electricians may work on electrical installations.
The connection and instruction manual contains information that enables the proper and efficient use of the WIO monitoring system. PAJ Sensors A/S is not liable for damage caused be improper use of this device. This manual is an integral part of the basic knowledge necessary for proper use of the system.
Rights
PAJ Sensors A/S reserves the right to alter its products without prior notice.
Copying and using the instruction manual for other purposes is only allowed with the acceptance from PAJ Sensors A/S.
PAJ Sensors A/S accepts no responsibility for possible errors and deficiencies in brochures, catalogues and other printed material.
PAJ Sensors A/S guarantees correct function as well as fulfillment of the safety requirements only when connections are made in accordance with instructions.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 5 af 53
Safety observations
The comments about safety in this document will not discuss safety observations of individual machine parts where safety devices (usually safety relays) are applied. Here the respective instruction manuals will be referenced! This document merely describes how to establish a WIO and how to start it up.
WARNING
The monitoring system during machine operation cannot be guaranteed if the system is connected incorrectly or not used as specified. This may lead to fatal injuries. Interventions and changes to the WIO monitoring system are not permitted, unless they are explicitly described in this user manual or be written from PAJ Sensors A/S.
The regional legal regulations and conditions of the liability insurance of the employer must be maintained.
Area of application
The WIO monitoring system is usually used in engines were a WIO will monitor the water content in lubrication oil, gear box oil, diesel oil, hydraulic oil or transformer oil.
The advantage of this monitoring system is the simple 4-20 mA interface and the two relays indicating alarms, with digital readout.
Normally the WIO sensor is calibrated in Taro30 lubrication oil, if nothing else are defined from customer.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 6 af 53
Function
Short description of the function
PAJ Sensors A/S has developed a water in oil (WIO) monitoring system. The fundamental idea of the WIO system is to easily and continuously monitor the water activity (aw) in oil. The output is converted into a standard 4-20 mA output.
The system also includes two relays indicating when too much water is present in the oil. Default alarm values are set to 0,5 aw (high alarm) and 0,9 aw (high high alarm)
By use of the WIO-Software it is possible to change HHA and HA levels. Furthermore you can continuously read out and graph the aw (or PPM) of water in the oil.
WARNING
Do not use the sensor or relays to automatically enable/turn on or disable/shut down engine. The decision to enable/turn on or disable/shut down an engine, based on the sensor data, must be taken by proper trained crew personnel, not by the sensor system.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 7 af 53
Normal operation
Water activity (aw) provides the relative availability of water in oil where pure oil has an activity of zero and oil saturated with distilled water has an activity of exactly one. Under normal operation the WIO200 continuously supplies an output analogue signal of 4 to 20 mA corresponding to 0.01 to
1.00 aw. Alarm indications with 0.03 aw hysteresis via 2 relays are provided for each sensor. These data can also be read out from the sensor on the RS485 Modbus communication line.
Default alarm values are 0,5 aw and 0,9 aw. Assumes these alarm values:
High alarm on at 0,50 aw (12 mA). High alarm off at 0,47 aw (11,5 mA) High high alarm on at 0,90 aw (18,4 mA). High high alarm off at 0,87 aw (17,9 mA). High high alarm will disable high alarm, thus only one alarm can be active.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 8 af 53
Failure indication
Internal function failure will be indicated by activating both alarm relays at the same time, or indicated as defined in the NAMUR NE43.
Internal failures includes:
Sensor reading out of range. Sensor reading CRC error. Unstable Sensor reading.
NAMUR NE43 is a German fault detection standard for 4-20mA analogue signals. It allows the user to know if there is a fault within the instrument, by sending analogue signal below 4mA and/or above 20mA. In accordance with NAMUR NE43, the failure is indicated by:
Fault indicated by analogue output =<2,0 mA
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 9 af 53
Self test for alarm functions
On the WIO Sensor there is a button for testing the alarm functions. By pushing the Test button for 5 sec the high alarm relay turns on, and by pushing the Test button for 10 sec the high alarm relay turns off and the high high alarm turns on. When the Test button has been pushed continuously for more than 15 sec, it will be ignored by the software, and the relays return to normal operation.
Test Button
Mode
Output Relay
Analogue Output
Un-pressed
Normal operation and no alarm
Relay 1 Open Relay 2 Open
Normal
Pressed longer
than 5 sec
Test of high alarm
Relay 1 Closed Relay 2 Open
Normal
Pressed longer
than 10 sec
Test of high high alarm
Relay 1 Open Relay 2 Closed
Normal
Pressed longer
than 15 sec
Normal operation
Normal
Normal
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 10 af 53
Test Button
Mode
Output Relay
Analogue Output
Un-pressed
Normal operation, with high alarm active
Relay 1 Closed Relay 2 Open
Normal
Pressed longer
than 5 sec
Test of high alarm
Relay 1 Closed Relay 2 Open
Normal
Pressed longer
than 10 sec
Test of high high alarm
Relay 1 Open Relay 2 Closed
Normal
Pressed longer
than 15 sec
Normal operation
Normal
Normal
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 11 af 53
Before installation
IMPORTANT
Only qualified personnel may assemble and maintain the unit. Only qualified personnel may install the unit.
Screw in probe pressure-tight directly in the center of the oil pipe where the measurement is to take place. The threads should be sealed with Loctite® 271. The sensor should be mounted with max. torque of 35 Nm. Connect the cables from the WIO sensor to the Terminal box. Use a hexagon wrench size 30.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 12 af 53
To sensor Power and Alarms
IMPORTANT
Only qualified personnel may assemble start-up and maintain the terminal box(s) when it is in a zero voltage state. Only qualified electricians may work on electrical installations.
Install +24 VDC and GND into the terminal box and connect the 4-20 mA to an external device if needed. Connect the relay to external device if needed.
If the terminal box has a (aw) or PPM display, there must be a short circuit between terminals 11 and 12. This short circuit produces the needed closed loop for mA measurement.
If the terminal box has a °C or °F display, there must be a short circuit between terminals 9 and 10. This short circuit produces the needed closed loop for mA measurement. Only for WIO and WIOI 300/400/500
After installing the WIO-Software can PC and terminal box be connected through DB9 connector. Only for WIO and WIOI 300/400/500
Last connect the cables from the Terminal box to the sensor.
It is allowed (but not necessary) to connect screw terminal 10 and 12 ­4-20 mA” to screw terminal 14 “GND”, thus creating common ground for the output and the power.
Terminal boxes with display. Screw terminal pin 11 and 12 or screw terminal pin 9 and 10 can be connected to external equipment, but only in a closed loop for mA measurement.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 13 af 53
Terminal Box Connection
Contact Assignment
There is only one connector CN3 for using of the terminal box contacts.
For con
nection more than one terminal box please refer to:
- Appendix A1
- Appendix A2
- Appendix A3
Contact­Number
Description
9
+4 – 20 mA positive analog output signal for indication of temperature in oil. This option is not available for WIO200
Only for WIO and WIOI 300/400/500
10
-4 – 20 mA return analog signal for indication of temperature in oil. This option is not available for WIO200
Only for WIO and WIOI 300/400/500
11
+4 – 20 mA positive analog output signal for indication of aw or PPM in oil.
12
-4 – 20 mA return analog signal for indication of aw or PPM in oil.
13
Power supply, 24VDC ±10% and with max. residual voltage ripple 10%
14
Power supply ground terminal contact
15
Relay contact for HH Alarm
16
Relay contact for HH Alarm
17
Relay contact for H Alarm
18
Relay contact for H Alarm
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 14 af 53
WIO Sensor dimensions in mm
WIO Standard ½” thread Sensor, Short dimensions in mm:
30mm Wrench max. torque 35 Nm
WIO Standard ¾” thread Sensor, Short dimensions in mm:
30mm Wrench max. torque 35 Nm
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 15 af 53
WIO Ball Valve pipe ½” thread Sensor, Long dimensions in mm:
30mm Wrench max. torque 35 Nm
WIO Ball Valve pipe ¾” thread Sensor, Long dimensions in mm:
30mm Wrench max. torque 35 Nm
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 16 af 53
Terminal box dimensions in mm
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 17 af 53
WIO Integrated (WIOI) dimensions in mm
WIOI Sensor, short pipe, ½” thread, dimensions in mm:
WIOI Sensor, long pipe, ½” thread, dimensions in mm:
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 18 af 53
WIOI Sensor, long pipe, ¾” thread, dimensions in mm:
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 19 af 53
Technical data, Sensors
WIO Standard ¾” (ISO 228-1 G¾”), Short
WIO Standard ½” (ISO 228-1 G½”), Short
WIO Ball Valve pipe, Long
Output
Analogue output (galvanic isolated)
4 – 20 mA for a
w
Max. Load (analogue output)
< 500Ω
Measurement Range (4 – 20 mA)
0,01 – 1,00 aw
Accuracy (0,05-0,95 aw)
± 0,03 a
w
Accuracy (outside 0,05-0,95 aw)
± 0,05 aw
Resolution
< 0,004 aw
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8 VA
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 20 af 53
Relays
Contact arrangement
2 x Normally Closed (NC)
Rated voltage
60V
Rated current (40C)
1A
Max. DC Load breaking capacity
Relay 1
„High Alarm“
Relay 2
„High High Alarm“
Default High Alarm
0,50 aw
Default High High Alarm
0,90 aw
Socket specification
Connector design
1x male socket, 1x female socket
Connector locking system
Screw-locking, M12x1
Wire gauge
0,25 mm2
Contacts
8 Pol
Rated voltage
60V
Rated current (40C)
1A
Cable specification
Cable design
Multipair overall screened cable
Outlet diameter
9,9 mm
Voltage class
0,6/1kV
Wires
4x2 twisted pair
Wire gauge
0,75 mm2
Media for measurement
Lubrication oil
Grade SAE 30/TBN 5-10
Max. Oil temperature
90°C
Max. Oil pressure
Sensor mounting method in application: Ball valve mounted: Max. 10 Bar Thread mounted: Max. 20 Bar
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 21 af 53
Response times
Delay before valid data from start-up
< 30 s
Delay before valid data from installation (first use)
10 minutes Device Failure Indication
Analogue output
< 2 mA
Manual test
Press sensor test button for 5 seconds
High Alarm turns on for 5 seconds
Press sensor test button for 10 seconds
High Alarm turns off High High Alarm turns on for 5 seconds
Press sensor test button for 15 seconds
Normal operating and test button ignored
Miscellaneous
Ambient Temperature, running / storage
0 - 90°C / -30 - +95°C
Relative humidity for running and storage
10% up to 95%, no condensation
Re calibration
Recommended with max 3 years interval
Warranty
2 years
Approvals
Germanischer Loyd
WIO products are EU approval under 75 965 – 09 HH date 2009-11-30
Enclosure
Weight for WIO standard
650 grams
Weight for WIO Ball valve pipe
790 grams
Connection (mechanical)
ISO 228-1 G ½” or ¾” thread 2 x 8-pole connectors, male and female,
M12x1 thread
Enclosure material
Stainless Steel/Aluminum EN44100 (Former
4261)
Protective type
IP66
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 22 af 53
Technical Data Terminal Box
FUNCTIONING
Terminal box shall be provided with 24vdc. Terminal box supply 24vdc on to WIO sensor. WIO sensor delivers mA, digital values and relay status to the terminal box.
Output
Analogue output
See the specifications for sensor’s analogue output
Digital output
See the specifications for sensor’s digital
output
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8 VA
Relays
Contact arrangement
2 x Normally Open (NO)
Rated voltage
250 VAC
Max. switching voltage
400VAC
Rated current
2A
Breaking capacity max.
1250VA
Enclosure
Weight
650 grams
Connection to sensor (mechanical)
2 x 8-pole connectors, male and female, M12x1 thread
Connection (mechanical)
2 x M20 gland, cable diameter 6 to 12 mm
Enclosure material
Aluminum
Protective type
IP66
Warranty
2 years
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 23 af 53
Technical Data Terminal Display Box
Output
Analogue output
See the specifications for sensor’s analogue output
Digital output
See the specifications for sensor’s digital output
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8 VA
Relays
Contact arrangement
2 x Normally Open (NO)
Rated voltage
250 VAC
Max. switching voltage
400VAC
Rated current
2A
Breaking capacity max.
1250VA
Display version - aw (water activity from 0,01 to 1,00)
Accuracy (0,05-0,95 aw)
± 0,03 aw
Resolution
<0,004 aw
Display version - PPM (H20)
Accuracy (0,05-0,95 aw)
±30%
Resolution
1 PPM
Enclosure
Weight
650 grams
Connection to sensor (mechanical)
2 x 8-pole connectors, male and female, M12x1 thread
Connection (mechanical)
2 x M20 gland, cable diameter 6 to 12 mm
Enclosure material
Aluminum
Protective type
IP66
Warranty
2 years
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 24 af 53
FUNCTIONING
Terminal box shall be provided with 24vdc. Terminal box supply 24vdc on to WIO sensor. Terminal box measuring mA from the sensor, and convert it to aw or PPM value. WIO sensor delivers mA, digital value and relay status to the terminal box.
IMPORTANT
The display is an indication of the water level in the oil. Display can be shipped as PPM or aw.
Upon ordering a PPM display the customer must specify the water saturation point in PPM of the used oil at the preferred working temperature of the oil. The saturation point is typically between 3000­10000 PPM. Alternatively the customer can ship 10 liters of oil to PAJ Sensor A/S for inspection including specification of the working oil temperature.
WARNING
All displays will show from positive values when operating properly. There will appear a negative value during start-up, after 5 seconds the value will be in positive measurement range. Otherwise, if a negative value is present the Terminal box might have been installed incorrectly.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 25 af 53
NOTE
WIO200 uses 3 different units for the measurement of water in oil:
1) a
w
: WIO measures humidity in aw. The range is 0.01…1.00 aw.
2) %RH (relative humidity) is identical to the unit a
w
multiplied by
100. The range is 1…100 %RH
3) ppm: WIO can convert a
w
to ppm. In this case the saturation point, at a given temperature, must be known. Output in ppm is only reliable at this temperature.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 26 af 53
Technical Data Terminal Box Alarm
Output
Analogue output
See the specifications for sensor’s analogue output
Digital output
See the specifications for sensor’s digital output
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8 VA
Relays
Contact arrangement
2 x Normally Open (NO)
Rated voltage
250 VAC
Max. switching voltage
400VAC
Rated current
2A
Breaking capacity max.
1250VA
Buzzer
Oscillation frequency
3000±500 Hz
Sound pressure level
85db by open housing
Tone
Pulsed
Button with LED
Blink frequency
2 Hz
Color
RED
Enclosure
Weight
530 grams
Connection to sensor (mechanical)
2 x 8-pole connectors, male and female, M12x1 thread
Connection (mechanical)
2 x M20 gland, cable diameter 6 to 12 mm
Enclosure material
Aluminum
Protective type
IP66
Warranty
2 years
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 27 af 53
FUNCTIONING
Terminal box shall be provided with 24vdc. Terminal box supply 24vdc on to WIO sensor. Terminal box indicate the relay status from the WIO sensor. WIO sensor delivers mA, digtal values and relay status to the terminal box.
H-Alarm LED is for indication of the High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High Alarm relay is open and the H-Alarm LED flash red ON/OFF sequence to show the High Alarm state, and the alarm signal buzzer will be switched ON.
HH-Alarm LED is for indication of the High High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High High Alarm relay is open and the HH-Alarm LED flash red ON/OFF sequence to show the High High Alarm state, and the alarm signal buzzer will be switched ON.
The H-Alarm LED or HH-Alarm LED will stay in blinking mode until the H­Alarm or HH-Alarm button is pressed, then the LED constantly turned on. The alarm signal buzzer will stay on until the H-Alarm or HH-Alarm button is pressed.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 28 af 53
Abstract representation of the H-Alarm and HH-Alarm button and LED function:
t
0,5
0,9
a
w
HA-LED
off
HHA-LED
off
flashing
HA-Button pressed
HHA-Button pressed
ON
OFF
flashing
flashing
ON
flashing
H-Alarm Relay
HH-Alarm Relay
Figure 1: H-Alarm and HH-Alarm button function
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 29 af 53
Technical Data Terminal Box Alarm Buzzer
Output
Analogue output
See the specifications for sensor’s analogue
output
Digital output
See the specifications for sensor’s digital
output
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8 VA
Relays
Contact arrangement
2 x Normally Open (NO)
Rated voltage
250 VAC
Max. switching voltage
400VAC
Rated current
2A
Breaking capacity max.
1250VA
Display version - aw (water activity from 0,01 to 1,00)
Accuracy (0,05-0,95 aw)
± 0,03 aw
Resolution
<0,004 aw
Display version - PPM (H20)
Accuracy (0,05-0,95 aw)
±30%
Resolution
1 PPM
Buzzer
Oscillation frequency
3000±500 Hz
Sound pressure level
85db by open housing
Tone
pulsed
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 30 af 53
FUNCTIONING
Terminal box shall be provided with 24vdc. Terminal box supply 24vdc on to WIO sensor. Terminal box measuring mA from the sensor, and convert it to aw or PPM value. WIO sensor delivers mA, digital values and relay status to the terminal box.
H-Alarm LED is for indication of the High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High Alarm relay is open and the H-Alarm LED flash red ON/OFF sequence to show the High Alarm state, and the alarm signal buzzer will be switched ON.
HH-Alarm LED is for indication of the High High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High High Alarm relay is open and the HH-Alarm LED flash red ON/OFF sequence to show the High High Alarm state, and the alarm signal buzzer will be switched ON.
The H-Alarm LED or HH-Alarm LED will stay in blinking mode until the H­Alarm or HH-Alarm button is pressed, then the LED constantly turned on. The alarm signal buzzer will stay on until the H-Alarm or HH-Alarm button is pressed.
Button with LED
Blink frequency
2 Hz
Color
RED
Enclosure
Weight
650 grams
Connection to sensor (mechanical)
2 x 8-pole connectors, male and female, M12x1 thread
Connection (mechanical)
2 x M20 gland, cable diameter 6 to 12 mm
Enclosure material
Aluminum
Protective type
IP66
Warranty
2 years
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 31 af 53
Abstract representation of the H-Alarm and HH-Alarm button and LED function:
t
0,5
0,9
a
w
HA-LED
off
HHA-LED
off
flashing
HA-Button pressed
HHA-Button pressed
ON
OFF
flashing
flashing
ON
flashing
H-Alarm Relay
HH-Alarm Relay
Figure 2: H-Alarm and HH-Alarm button function
IMPORTANT
The display is an indication of the water level in the oil. Display can be shipped as PPM or aw.
Upon ordering a PPM display the customer must specify the water saturation point in PPM of the used oil at the preferred working temperature of the oil. The saturation point is typically between 3000­10000 PPM. Alternatively the customer can ship 10 liters of oil to PAJ Sensor A/S for inspection including specification of the working oil temperature.
WARNING
All displays will show from positive values when operating properly. There will appear a negative value during start-up, after 5 seconds the value will be in positive measurement range. Otherwise, if a negative value is present the Terminal box might have been installed incorrectly.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 32 af 53
NOTE
WIO200 uses 3 different units for the measurement of water in oil:
1) a
w
: WIO measures humidity in aw. The range is 0.01…1.00 aw.
2) %RH (relative humidity) is identical to the unit a
w
multiplied by
100. The range is 1…100 %RH
3) ppm: WIO can convert a
w
to ppm. In this case the saturation point, at a given temperature, must be known. Output in ppm is only reliable at this temperature.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 33 af 53
Technical data, WIOI (WIO Integrated)
Output
Analogue output (galvanic isolated)
4 – 20 mA for a
w
Max. Load (analogue output)
< 500Ω
Measurement Range (4 – 20 mA)
0,01 – 1,00 aw
Accuracy (0,05-0,95 aw)
± 0,03 aw
Accuracy (outside 0,05-0,95 aw)
± 0,05 a
w
Resolution
< 0,004 aw
Input
Supply nominal voltage
24V DC ± 10%
Max. residual voltage ripple
10%
Maximum Load current
200 mA
Max. Power input
< 4,8VA
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 34 af 53
Relays
Contact arrangement
2 x Normally Open (NO)
Rated voltage
250VAC
Max. switching voltage
400VAC
Rated current
2A
Breaking capacity max.
1250VA
Max. DC Load breaking capacity
Relay 1
„High Alarm“
Relay 2
„High High Alarm“
Default High Alarm
0,50 aw
Default High High Alarm
0,90 aw
Cable specification
Cable design
Multipair overall screened cable
Outlet diameter
9,9 mm
Voltage class
0,6/1kV
Wires
4x2 twisted pair
Wire gauge
0,75 mm2
Media for measurement
Lubrication oil
Grade SAE 30/TBN 5-10
Max. Oil temperature
90°C
Max. Oil pressure
Sensor mounting method in application: Ball valve mounted: Max. 10 Bar Thread mounted: Max. 20 Bar
Response times
Delay before valid data from start-up
< 30 s
Delay before valid data from installation (first use)
10 minutes
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 35 af 53
Device Failure Indication
Analogue output
< 2 mA
Manual test
Press sensor test-button for 5 seconds
High Alarm turns on for 5 seconds
Press sensor test-button for 10 seconds
High Alarm turns off High High Alarm turns on for 5 seconds
Press sensor test-button for 15 seconds
Normal operating and test button ignored
Display version - aw (water activity from 0,01 to 1,00)
Accuracy (0,05-0,95 aw)
± 0,03 aw
Resolution
<0,004 aw
Display version - PPM (H20)
Accuracy (0,05-0,95 aw)
±30%
Resolution
1 PPM
Buzzer
Oscillation frequency
3000±500 Hz
Sound pressure level
85db by open housing
Tone
pulsed
Button with LED
Blink frequency
2 Hz
Color
RED
Enclosure
Weight
4000 grams
Connection (mechanical)
ISO 228-1 G ½” or ¾” thread 2 x gland M20, cable diameter 6 to 12 mm
Enclosure material
Stainless Steel/Aluminum EN44100 (Former
4261)
Protective type
IP66
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 36 af 53
FUNCTIONING
Terminal box shall be provided with 24vdc. Terminal box supply 24vdc on to WIO sensor. Terminal box measuring mA from the sensor, and convert it to aw or PPM value. WIO sensor delivers mA, digital values and relay status to the terminal box.
H-Alarm LED is for indication of the High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High Alarm relay is open and the H-Alarm LED flash red ON/OFF sequence to show the High Alarm state, and the alarm signal buzzer will be switched ON.
HH-Alarm LED is for indication of the High High Alarm relay state from the WIO sensor. When the alarm condition is generated by the WIO sensor, then the High High Alarm relay is open and the HH-Alarm LED flash red ON/OFF sequence to show the High High Alarm state, and the alarm signal buzzer will be switched ON.
The H-Alarm LED or HH-Alarm LED will stay in blinking mode until the H­Alarm or HH-Alarm button is pressed, then the LED constantly turned on. The alarm signal buzzer will stay on until the H-Alarm or HH-Alarm button is pressed.
Miscellaneous
Ambient Temperature, running / storage
0 - 90°C / -30 - +95°C
Relative humidity for running and storage
10% up to 95%, no condensation
Re calibration
Recommended with max 3 years interval
Warranty
2 years
Approvals
Germanischer Loyd
WIO products are EU approval under 75 965 – 09 HH date 2009-11-30
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 37 af 53
Abstract representation of the H-Alarm and HH-Alarm button and LED function:
t
0,5
0,9
a
w
HA-LED
off
HHA-LED
off
flashing
HA-Button pressed
HHA-Button pressed
ON
OFF
flashing
flashing
ON
flashing
H-Alarm Relay
HH-Alarm Relay
Figure 2: H-Alarm and HH-Alarm button function
IMPORTANT
The display is an indication of the water level in the oil. Display can be shipped as PPM or aw.
Upon ordering a PPM display the customer must specify the water saturation point in PPM of the used oil at the preferred working temperature of the oil. The saturation point is typically between 3000­10000 PPM. Alternatively the customer can ship 10 liters of oil to PAJ Sensor A/S for inspection including specification of the working oil temperature.
WARNING
All displays will show from positive values when operating properly. There will appear a negative value during start-up, after 5 seconds the value will be in positive measurement range. Otherwise, if a negative value is present the Terminal box might have been installed incorrectly.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 38 af 53
NOTE
WIOI200 uses 3 different units for the measurement of water in oil:
1) a
w
: WIO measures humidity in aw. The range is 0.01…1.00 aw.
2) %RH (relative humidity) is identical to the unit a
w
multiplied by
100. The range is 1…100 %RH
3) ppm: WIO can convert a
w
to ppm. In this case the saturation point, at a given temperature, must be known. Output in ppm is only reliable at this temperature.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 39 af 53
EC-Declaration of Conformity
PAJ Group, Grundtvigs Allé 163, DK-6400 Sønderborg, Denmark
Manufactory Declare under our sole responsibility that the product:
unit identification: product classification:
WIO200 – WIO300 – WIO400 – WIO500 Water In Oil measuring instrument
is a safety tested component according to EC Guideline:
2004/108/EC Electromagnetic compatibility 2006/95/EC Low voltage directive
to which this declaration relates is in conformity with the following standard(s) or other normative documents(s):
IEC 61326-1:2006-11-03, EN 61010-1:2001, EN 60529 Edition 2.1 2001-02, IEC 60068-1:1988, IEC 60068-2-2:2007, IEC 60068-2-1:2007, IEC 60068-2-30:2005, IEC 60068-2-6:2007, IEC 61000-4-16:1998-01, IEC 61000-4-6:2007, IEC 61000-4-3:2006, IEC 61000-4-5:2005, IEC 61000-4-2:2001, IEC 61000-4-4:2004
The described product corresponds to the following European Directives:
2004/108/EC Electromagnetic compatibility 2006/95/EC Low voltage directive
Consistency of a production sample device with the marked product in accordance with the Directives No:
2004/108/EC Electromagnetic compatibility 2006/95/EC Low voltage directive
Notified agency/Address: Germanischer Lloyd AG Head Office Deputy Head of Department
Automation, Navigation and Communication (MC-EA) Vorsetzen 35
D-20459 Hamburg/Germany
75 956 – 09 HH 2009-11-30
Certification number Date of issue
The marked product is consistent with the examined production sample device.
Technical documentation:
Poul Jessen
______________________________________________________________________ Grundtvigs Allé 163 • DK-6400 Sønderborg • Tel: +45 74 43 71 81 • E-Mail: paj@paj.dk
Sønderborg, 15.04.2011 Poul Jessen, Managing Director
Date of issue Name and signature
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 40 af 53
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 41 af 53
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 42 af 53
Appendix A1: Install 1 Terminal Box with 1 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 43 af 53
Appendix A2: Install 1 Terminal Box with no display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 44 af 53
Appendix A3: Install 2 Terminal Boxes with 2 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 45 af 53
Appendix A4: Install 2 Terminal Boxes with 1 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 46 af 53
Appendix A5: Install 2 Terminal Boxes with 1 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 47 af 53
Appendix A6: Install 3 Terminal Boxes with 3 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 48 af 53
Appendix A7: Install 3 Terminal Boxes with 2 display
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 49 af 53
Appendix A9: CABLE Specification
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 50 af 53
Appendix A10: Error / failure FAQ
Failure type
Description
Possible reasons
PPM display failure
Display shows only negative numbers
1. Sensor is not connected.
2. The error caused by missing jumper in terminal box pin 11 and pin 12. See Appendix A1
3. Check that the jumper is properly installed, it is seen that the jumper has been installed but the jumper had no proper connection
4. The wires are reversed in terminal box pin 7 and pin 8 lower part. After changing and there is no change in the display, then the fault is either 1, 2 or 3.
a
w
display
failure
Dispay shows only negative numbers
1. Sensor is not connected.
2. The error caused by missing jumper in terminal box pin 11 and pin 12. See Appendix A1
3. Check that the jumper is properly installed, it is seen that the jumper has been installed but the jumper had no proper connection
4. The wires are reversed in terminal box pin 7 and pin 8 lower part. After changing and there is no change in the display, then the fault is either 1, 2 or 3.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 51 af 53
Dead display
Display indicate nothing
Maybe is 24V dc supply interrupted to the display. There shall be 24V dc at this connector.
Relay
Relay-state while alarm is on
Sensor
How to check the mA output from sensor?
Remove jumper in pin 11 and pin
12. In order to ensure that there can be measured mA must pin 11 and pin 12 must be screwed. Remember to set the measuring instrument in dc mA and connect test leads correctly in the measuring instrument. Sensor is alright when measuring a value between 4 and 20 mA. Is the result about 2 mA then there is an internal sensor error, so the sensor must be sent for repair. Is the result 0 mA, tjek if the sensor get 24 Vdc power.
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 52 af 53
Dip Switch
a
w
Are the dip switches still positioned correctly?
Correct positions:
Dip Switch PPM
Are the dip switches still positioned correctly?
Correct positions:
Broken cables between Sensor and terminal box
No mA will can be measured on:
Analog 1: Terminal 11 and 12 and Analog 2: Terminal 9 and 10
PAJ Sensor A/S • Grundtvigs Allé 163 • DK-6400 Sønderborg
Tel: +45 74 43 71 81 • Fax: +45 74 43 71 91 • CVR: 3459-1229 • www.paj.dk • e-mail: paj@paj.dk
APPROVALS: ISO 9001, ISO 14001, ISO 13485, IEC 61340-51 & IPC-A-610 CLASS 3
03-01-0501-CRJ-04 Side 53 af 53
Company Information
PAJ Sensor A/S Grundtvigsallé 163 DK-6400 Sønderborg Tel: (0045) 7443 7181 Mail: paj@paj.dk
Loading...