1: Site Preparation
1.1: Locating the Unit
• Place the SCA on a sturdy horizontal surface or cart capable of supporting 200 lbs
(the weight when filled with water).
• Ensure there is no Aluminum or Zinc in the area where the SCA will be used. THE
CLEANING SOLUTION CONTAINS S ODIUM HYDROXIDE AND REACTS
VIOLENTLY WITH ALUMINUM OR ZINC.
• The selected location should be isolated from unintentional contact from those
working in the area.
• It should be near a grounded wall outlet such that the power cord does not pose a
hazard to people or equipment passing by.
• The power cord plug-in receptacle is a disconnecting device that should be easily
accessible at all times. Position the unit so that it can be easily reached.
• It is strongly recommended that the unit be plac ed on a circuit with its ow n breaker
with a GFCI.
• Place the unit in a well-ventilated area.
1.2: Preparations for Drai ning and Disposal
• Placing the unit near a drain or sink will make draining the unit much easier.
Alternatively, the SCA can be kept on a cart for easy transport to a sink or drain.
• Use a 1” inner diameter hose made of PVC or equivalent on the drain tube to
facilitate draining. Ensure the tubing or hose can withstand the temperature and
caustic properties of the cleaning solution.
Used solution must be disposed of under applicable local waste disposal regulations. It is
the responsibility of the user to determine and verify what the local disposal
regulations are and to follow those regulations.
SCAHTUM-03 4 SCA 1200HT User Manual
Page 5
Do not get in eyes, on skin or on clothing. Do not swallow or inhale powder. Keep
2: Important Safeguards
These notes are intended to draw your attention to risks which only you can recognize and
avoid or overcome. They are intended to enhance your own safety consciousness.
THE CORRECT UNIT USAGE AND PROPER HANDLING IS SOLELY THE
USER’S RESPONSIBILITY.
ELECTRICAL PROTECTION WILL BE IMPAIRED IF USED IN A MANNER
NOT SPECIFIED BY THE MANUFACTURER.
Failure to follow these safeguards and/or common sense may result in significant personal
injury and will void the warranty.
The WaterWorks solution that may be added to the unit to aid in support removal contains
Sodium Hydroxide. Please read and remember the following safety information about this
material:
DANGER: WaterWorks SOLUTE CONTAINS SODIUM
HYDROXIDE WHICH CAN CAUSE SEVERE BURNS TO EYES,
SKIN AND RESPIRATORY TRACT.
container closed. Use only with adequate ventilation. Wash thoroughly after
handling.
First Aid:
In case of contact, immediately flush eyes and skin with plenty of water for at least 15
minutes while removing contaminated clothing and shoes. Get medical attention
immediately.
NFPA Rating: Health H=3
Flammability F=0
Reactivity R=1
SCAHTUM-03 5 SCA 1200HT User Manual
Page 6
3: Safety Warnings
READ ALL INSTRUCTIONS BEFORE OPERATING
WHAT TO DO WHAT NOT TO DO
Always wear thermal gloves and safety
glasses when working near the unit or
when touching any part of the unit.
Always keep this manual near the unit.
Always place unit on a flat, stable
surface.
Always wipe away any cleaning solution
spilled near the control panel.
Always locate the unit away from
locations where it might be bumped.
Always unplug the unit from the wall
when not in regular use.
Always use the power cord supplied by
the manufacturer.
Always make sure that the power plug
receptacle is easily accessib le.
Always unplug the unit from the wall
when it is being cleaned, moved or
serviced in any way.
Always connect the device to a power
receptacle with a protective ground.
Always change the cleaning solution as
recommended in this manual.
Always clean the unit with mild soap and
a sponge or rag. Rinse tank completely
before refilling.
Always stand upright with your head
away from the tank when opening the lid
to avoid vapors.
Always remove some liquid from the
tank before adding large parts to avoid
overflow.
Always operate the unit in a well-
ventilated location.
Always operate within environmental
temperature range of 10 to 30 degree C.
Do not use the SCA for any purpose
other than removing Soluble Suppor t
Technology (SST) material from parts
created on Stratasys 3D Printers.
Do not allow aluminum, zinc or acids to
come into contact with the WaterWorks
cleaning solution. WaterWorks cleaning
solution contains Sodium Hydroxide.
Do not use any liquid other than water
and Stratasys cleaning solution
products in the unit.
Do not add more than the recommended
amount of cleaning solution to the unit.
Do not overfill the unit.
Do not allow the fluid temperature to
exceed 85ºC (185ºF).
Do not clean the unit with solvents.
Do not immerse the unit in liquid of any
kind.
Do not operate the device if there are
any doubts regarding safe operation due
to the outer appearance (e.g. damages)
of the system or the operating
environment.
Do not operate the unit until you have
read this manual and understand all
aspects of the unit’s operation.
Do not move the unit until you are
certain that it is cool and all liquid has
been drained from the unit.
Do not operate the unit unless the water
level is above the low liquid level
indicator mark.
Do not operate the unit dry.
SCAHTUM-03 6 SCA 1200HT User Manual
Page 7
4: Warranty Information
WARRANTY CERTIFICATE
Covering PADT, Inc. Support Cleaning Apparatus
Effective January 1, 2009
One Year Limited Warranty Coverage
All new Support Clea ning Apparatus (SCA) s y s tems are warranted ex c l usiv ely
by PADT, Inc.’s (“Manufacturer”) limited warranty as follows:
Each Support Cleaning Apparatus system (“System”) and its components
(“Components”), except those listed below under limits and exclusions, is
warranted against defects in the materials and workmanship for a period of one
(1) year from the date of installation at the end user’s (“Customer”) facility.
Repair or replacement only: manufacturer's liability under this agreement shall
be limited to repairing or replacing, at the discretion of manufacturer, parts, or
components sufficient to return the system to conform to the marketing
specifications of the system.
Components subject to wear during normal use and over time such as paint,
finish, light bulbs, seals, etc., are excluded from this warranty.
This warranty i s void if the syst em is subjected to mishandling , misuse, negl ect,
accident, improper installation, improper maintenance, or improper operation
or application, or if the machine was improperly repaired or serviced by the
customer. This warranty is void if the system is not installed by a certified
distributor and the proper installation documentation provided by the
manufacturer has not been submitted.
Liability, whether based on warranty, negligence or other cause, arising out of
and/or incidental to sale, use or operation of the system, or any part thereof,
shall not in any case exceed the cost of repair or replacement of the defective
equipment, and such repair or replacement shall be the exclusive remedy of
the purchaser, and in no case will manufacturer be responsible for any and/or
all consequential or incidental damages including without limitation, and/or all
consequential damages arising out of commercial losses.
This warranty is transferrable from the original end user to another party if the
machine is sold via private sale before the end of the warranty period.
The foregoing i s a l imited warranty and it is th e o nl y warranty by manu fac tur er .
MANUFACTURER DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED, INCLUDING BUT NOT LIMITED TO ALL WARRANTIES OF
MERCHANTABILITY AND FITNE SS FOR A PARTICULAR PURPOSE.
SCAHTUM-03 7 SCA 1200HT User Manual
Page 8
5: Declaration of Conformity, Regulatory
Manufacturer
Phoenix Analysis & Design Technologies, Inc.
Tempe, AZ 85284, USA
Type of Equipment
Electrical Equipment for Measurement,
(Support Cleaning Apparatus)
Model Number(s)
SCA-1200ES and SCA-1200HT
We declare under our sole responsibility that the devices mentioned above
comply with the following EU Directives:
Low Voltage
2014/35/EU
Electromagnetic Compatibility (EMC)
(RoHS)
2004/108/EC
Common Technical Specifications Used for Demonstration of Compliance:
EN61010-1:2010 / CAN/CSA-C22.2 No. 61010-1:2012 / UL 61010-1:2012;
EN61326-1:2013
Design and Technical Construction
Hui Yang Zing Ear Industry Co., Ltd.
Name of Authorized Signatory
Eric Miller
Position Held in Company
Director
Signature
7755 S Research Dr, Suite 110, Tempe, AZ 85284 • www.PADTINC.com • 480.813.4884 V • 480.813.4807 F
Information
Declaration of Conformity
7755 S Research Dr, Suite 110
Control and Laboratory Use
Waste Electrical and Electronic Equipment
(WEEE)
Restriction of Hazardous Substances
EN61010-2-010:2014 / CAN/CSA-C22.2 No. 61010-2-010:2015 / UL 61010-2-010:2015;
The Support Cleaning Apparatus (SCA) is a device specifically developed to provide
robust and efficient removal of support material from parts created with 3 D pri nters from
Stratasys, Inc. that utilize their Soluble Support Technology (SST). This is an improved
model based on the feedback received from SCA users in the Stratasys 3D Printer
community. The original SCA system was developed with input from Stratasys, users and
distributors of SST based systems.
6.1: How it Works
The system removes support material by immersing parts created with Stratasys’ SST
material in a warm bath of water with a specific amount of cleaning solution added (see
operating instructions). The SCA heats the water and circulates the heate d water around
the parts in the tank. The hot solution dissolves the support material without harming the
underlying model material. Over time, depending on geometry and the amount of support
material, all the support material is dissolved and the part is ready to be rinsed, dried and
used for its intended purpose.
6.2: Key Components
The SCA consists of several key components that all users should be familiar with.
Understanding the function of each component is important for the safe operation of the
system, and will aid in extending the life of your SCA.
Many aspects of the SCA have been designed specifically for safety. All electrical
components are properly grounded and labeled for safety protection.
6.2.1: Control Panel
SCAHTUM-03 10 SCA 1200HT User Manual
Figure 1: Control Panel
Page 11
1 Up/Down arrows (Add/Subtract time)
2 Set timer
3 Start/Pause
4 No heat selection
5 Temperature selection
6 Silence alarm
7 Alert indicator
8 Low water indicator
9 High water indicator
10 Temperature LED display
11 Pump operating LEDs
12 P ro gress LEDs
13 Time remaining LED display
6.2.2: Tank Area
SCAHTUM-03 11 SCA 1200HT User Manual
Figure 2: Tank Area
Page 12
The tank area is used to hold the liquid and it is where the parts are cleaned. The SC A
Figure 4: Small
Figure 3: Large Parts
improves part cleaning by circulating the liquid in the tank.
The Large Part s Bas ket is used to hold parts during cleaning and to
lift parts out of the tank. Users should wear thermal gloves when
lifting the basket from the tank.
The Small Parts Basket is used to hold small parts
during cleaning.
The Strainer over the drain opening in the tank
Basket
prevents debris from entering the drain when the
Parts Basket
tank is emptied. It should be rinsed after emptying
the tank before refilling with fresh liquid.
The Heater is located behind the perforated Heater Shield. It heats the water in the tank.
The Spray Nozzle direc ts the flow of liquid into the tank. The spray nozzle can be removed
for cleaning if the holes are clogged. Simply unscrew the nozzle and rinse with tap water
to un-clog holes.
The Pump Intake Screen prevents large particulates from entering the pump. It may
occasionally need to be cleaned with a soft brush or toothbrush to remove debris trapped
in the holes.
The Temperature and Level sensors ar e used to sense the t emperature and height of the
liquid in the tank. They are mounted behind the rear support bracket.
SCAHTUM-03 12 SCA 1200HT User Manual
Page 13
Drain Tube
Drain Valve
(closed position)
Power Cord
Receptacle
6.2.3: Unit Side Area
Power Switch
Figure 5: Front and Side views
The Main Power Switch is located on the right-hand side of the unit. This turns power to
the unit on and off.
The Power Cord (not shown in figure) is used to connect the system to a wall outlet for
power. Only use a power cord provided by the manufacturer.
The Power Cord Receptacle is where the pow er cord is connected to the unit. The plug is
a disconnecting device and should be easily accessible at all times.
6.2.4: Front Area
The Drain Tube and Drain Valve are on the front of the tank for easy accessibility.
The tank is drained by connecting a hose to the Drain Tube and turning the Drain Valve to
the open position. The liquid in the tank should be at room temperature and the unit should
be turned off and unplugged from the wall before it is drained.
SCAHTUM-03 13 SCA 1200HT User Manual
Page 14
Hot Surface,
Do Not Touch
General Warning
Danger Electricity
GB/T 5465.2-5041
ISO 7000
6.3: Labels
The SCA includes labels that impart important product and safety information. Users
should take the time to familiarize themselves with each label and its meaning.
6.3.1: Product Identification Tag
The Product Identification tag is located on the back of the unit. It contains ke y support
information for the SCA including Model Number, Part Number, Serial Number and
contact information for support. It should never be tampered with or removed.
6.3.2: Warning Labels
Figure 6: Product Identification Tag
SCAHTUM-03 14 SCA 1200HT User Manual
Page 15
6.4: Assembly Drawing
Large basket
Small parts basket
Pump subassembly w/nozzle
Heater subassembly
Sensor subassembly
Display Control Panel
PCB subassembly
Power switch
Spray nozzle
Figure 7 shows the entire SCA assembly. Table 1 lists the replaceable components in the
system. Please consult this information when calling for support or ordering replacement
parts.
SCAHTUM-03 15 SCA 1200HT User Manual
Figure 7: SCA Assembly
Table 1: List of Replaceable Components
Page 16
7: Unpacking, Inspecting and Installing your SCA
Drain Tube
Drain Valve
(closed position)
Power Cord
Receptacle
THE SCA SHOULD BE INSTALLED BY AN AUTHORIZED DISTRIBUTOR ONLY.
INSTALLATION BY THE USER MAY VOID THE WARRANTY.
Before removing your SCA from its packaging, completely read this manual. Follow all
safety recommendations while unpacking, installing and operating the system.
7.1: Package Contents
The SCA is shipped with the following contents:
SCA System (1)
Large Part Basket (1)
Small Part Basket (1)
Quick Reference Card (1)
Power Cords:
110V North America std.
220V Europe std.
User’s Manual & Warranty (1)
If any parts are missing you should contact customer support immediately.
7.2: Unpacking
Place the package on a stable horizontal surface and check inside for the
items listed above. Some items may be stored inside the tank for
shipping. Remove everything from the SCA unit itself before lifting the
SCA from the shipping container.
Unit weight is approximately 64 lb/29 kg. Two or more people should
lift the unit out of the packaging and place it on a table or cart.
7.3: Inspecting Unit
Carefully inspect the unit for damage during shipping. If any damage is found DO NOT
ATTEMPT TO OPERATE THE UNIT. Contact technical support immediately for
assistance.
7.4: Installation
1. Place the Large Part Basket into the tank after ensuring all packing mat erial has
been removed.
2. Close the lid over the tank.
3. Attach the power cord to the system and then to a
grounded wall socket. Only use a power cord
provided by the manufacturer.
4. Press the ON/OFF Power Switch and check to
make sure that the display control panel lights up.
5. Turn the unit off.
6. Make sure the drain valve is in the closed
position. Remove the blue plastic cap from the
drain tube before use.
7. Complete the online registration form at
www.SupportRemoval.com.
SCAHTUM-03 16 SCA 1200HT User Manual
Power Switch
Figure 8: Power Switch and Plug Locations
Page 17
8: Operating Your SCA
Figure 9: Fill Unit to MAX line
Although the SCA is simple to operate, it is critical that all users follow the instructions in
this manual and adhere to all safety guidelines.
8.1: Fill the Unit with Water
Make sure that the strainer is clean. Make sure
that the drain valve is in the closed position.
Fill the tank up almost to the MAX slot on the
rear support bracket. If you plan on running
large parts, it is advisable to fill to a lower level
(but above the Minimum level) to make room
for the parts and basket. DO NOT use hot water.
The addition of WaterWorks to water is a heatreleasing reaction and if added to HOT water
will result in dangerous spattering.
Never operate the tank unless the liquid leve l is ab ove the liquid leve l MIN slot on the heat
shield. This is just below the bottom edge of the rear support bracket.
8.2: Add Cleaning Concentrate to the Tank
Users have two approved options for cleaning concentrate: either WaterWorks or
Ecoworks, both available from your local Stratasys reseller. The WaterWorks soluble
concentrate contains Sod ium Hydroxide and care must be taken when working with the
solution and when disposing of it. Refer to section on Maintaining your SCA for disposal
instructions. The Ecoworks cleaning agent is both user- and eco- friendly. The
WaterWorks concentrate will generally remove support material in less time than the
Ecoworks. However, no personal protection equipment is required when handling the
Ecoworks concentrate and only dilution is required when disposing of the used solution.
If Using WaterWorks Soluble Concentrate:
Wearing protective gloves and proper eye protection, carefully and slowly add one
bottle (950 grams) of WaterWorks Soluble Concentrate P400SC into the filled tank. Never
add concentrate to a tank without water. Optimal pH for the cleaning solution using
WaterWorks is 12.6, and the solution begins to lose effectiveness below 11.5pH. Using
litmus paper on a room temperature sample of the solution is sufficiently accurate to
determine the pH.
If Using Ecoworks Cleaning Agent:
Carefully remove the entire contents of six packages (foil bags) and gently drop them into
the water in the tank. Make sure to use both sides A and B. Ne ver place the cleaning agent
in a tank without water.
Note: As the parts a re cleaned the cleanin g solution becomes saturated with the support
material and loses its effectiveness. Refer to the Troubleshooting section for further
information about determining when the solution requires replacement.
SCAHTUM-03 17 SCA 1200HT User Manual
Page 18
8.3: Preheating and Starti ng the Cleaning Cycle
Recommended Temperature Settings for Dissolving
Build
Temperature Setting
WaterWorks
EcoWorks
ABS
70°C
Yes
Yes
PC
85°C
Yes
No
Nylon
50, 60 or 70°C**
Yes
Yes
**depending on part wall thickness and desired dimensional accuracy.
For Polyjet parts select the ‘HEAT OFF’ option.
Soluble Support from 3D Printed Parts
Material
Connect the power cord to the side of the unit then to a suitable grounded wall socket.
The unit is compatible with both North America 110V~50/60Hz and International
220V~50/60HZ power. The unit ships with both power cords. Onl y use a power cord
provided by the manufacturer.
It is strongly recommended that the unit be placed on a circuit with its own breaker with a
GFCI. The power circuit must be grounded.
Press the ON/OFF rocker switch. The display control panel should be on.
Press one of the four temperature preset buttons or the Heat Off button to select the cleaning
temperature. The temperatu re can be ch anged at any time duri ng the c ycle by select ing a
different temperat ur e button. If heating the water from room temperature to 70°C, it takes
about 2 hours to reach temperature.
Press the SET button and then press the up and down arrows to set the minutes. Press the
SET button again and use the up and down arrows to set the hours. Press the SET button
again to accept the time. Press the button to start the heater and pump.
8.4: Loading Parts and Testing Le v el Sensor s
Wait until the SCA has reached the desired temperature before loading parts into the tank.
This minimizes the time the parts are immersed in the cleaning solution.
Open the lid on the large parts basket and place the parts inside.
For smaller parts, open the spring-latched door of the smaller, square basket, place the
part(s) inside and place small basket inside larger basket.
Before placing the basket into the cleaning solution, press the button to pause the
heater and pump. With the basket cutout facing the nozzle, slowly lower the large parts
basket into the solution-filled tank. The spray nozzle should be centered on the basket
cutout.
If the high level alert lights and the alarm sounds, press the button to silence the
alarm, and remove liquid by draining or scooping liquid from the tank until the alert
indicator is no longer lit.
Note: It is recommended that the level sensors be tested periodically. Adding water to
just above the MAX fill line should cause the high level alert to illuminate and the buzzer
to sound. Draining water until the liquid level is below the MIN fill line should cause the
SCAHTUM-03 18 SCA 1200HT User Manual
Page 19
low level alert to illuminate and the buzzer to sound. If the level sensors are properly
Figure 10: Rest basket on tank
working the system will not operate until the liquid level is between the MIN and MAX
fill lines.
Never operate the tank unless the liquid level is above the liquid level MIN slot on the heat
shield. This is just below the bottom edge of the rear support bracket.
The system will not run until the liquid level is properly adjusted.
When adding or removing liquid from the tank, always remember that the tank liquid
contains cleaning solution. Always use gloves and wear eye protection when adding or
removing water from the tank. Never use aluminum or zinc containers.
Close the lids on the SCA to prevent heat loss and evaporation. Never place your face
near the tank when opening the lid – vapors from the cleaning solution may cause eye
and respiratory irritation.
8.5: Cleaning the Parts
To activate the cleanin g cycle after setting the time and temperature, and adjusting the
liquid level, press the button. Illuminated bars will sweep under the temperature display
when the pump is operating.
Parts can take a few hou rs or longer to clean. Cleaning time is affected by the size of the
part, the amount of support material, part geometry, and the pH level of the cleaning
solution.
Parts can be checked at any time during a cleaning cycle by opening the lid, carefully lifting
the Large Parts Basket from the tank, turning the basket 90° and resting it on the rear
support bracket and front tank ledge. Nev er place your face near the tank when opening
the lid - vapors from the cleaning solution may cause eye and respiratory irritation.
Protective gloves and eyewear should always be worn when lifting the basket from the
tank.
If the parts are clean before the end of the cleaning cycle, you can stop the cycle by pressing
the button on the display control panel.
After several cleaning cycles the cleaning solution becomes saturated with the support
material and loses its effectiveness. Refer to the Troubleshooting section for further
information about determining when the solution requires replacement.
8.6: Removing Parts from the Uni t
Once the cleaning cycle is completed or has been stopped, the
parts can be removed fro m the system. Open the lid, slowly
lift the Large Parts Basket from the t ank, turn the basket 90°
and rest it on the rear support bracket and front tank ledge.
Never place your face ne ar the tank when removing the lid vapors from the cleaning solution may cause eye and
respiratory irritation.
Let the basket rest abo ve t he t ank to let the liquid drain off of
the parts and the basket.
The proper gloves and eyewear should always be worn when
lifting the basket from the tank or handling wet parts before they
are washed.
SCAHTUM-03 19 SCA 1200HT User Manual
ledge and bracket
Page 20
Remove the parts and rin se them with wate r to wa sh awa y an y residual cleaning solution.
Air dry or pat dry with a cloth or paper towels.
Lower the basket into the tank and close the lid on t he tank when part removal is completed.
8.7: Readiness for the Next Cleaning Cycle
Once the cleaning cycle is completed and the parts have been removed, the system should
be turned off by pressing the ON/OFF rocker switch. Alternatively, if the SCA will be used
periodically use the 50°C temperature preset to maintain the water at an elevated
temperature. This will shorten the time required to heat the water to the optimal cleaning
temperature of 70 or 85°C.
If the system is not going to be utilized for an extended period, all liquid should be drained
from the system and the power cord should be removed from the wall socket.
SCAHTUM-03 20 SCA 1200HT User Manual
Page 21
9: Maintaining Your SCA
As with any piece o f equip ment, prop er mainten ance is cri tical for safety and long life of
the SCA.
9.1: Cooling the Tank Liquid
Before draining the t ank, make sure that all parts h ave been removed from the unit. If
possible, open the lid to decrease the cooling time as well. Never place your face near the
tank when opening the lid - vapors from the cleaning solution may cause eye and
respiratory irritation.
Leave the power on to the system to indicate the liquid temperature. Once it has reached
30ºC the tank can be drained safely.
9.2: Draining the Tank
Never drain the tank unless the fluid is at or below 30ºC.
Prior to draining, turn off power to the tank and unplug from wall socket.
Always wear protective gloves and eyewear when draining the tank.
You should drain the liquid from the tank if the unit will not be used in th e near future or
if the liquid needs to be changed out because the cleaning solution is saturated with support
material.
Make sure that the strainer is in place at the bottom of the tank.
To drain the tank, attach a hose to the drain tube on the front of the unit. P lace the other
end of the hose in a sink, or a plastic or stainless steel container that will be used to transport
the liquid to a drain after dilution. Secure the end of the hose in the sink or container so
that it cannot come loose during draining.
Turn the drain valve to the “Open” position.
Once the tank is drained, make sure the drain valve is in the “Closed” Position.
Note: use 1” inner diameter PVC tubing or equivalent on the drain tube. Ensure the tubing
or hose can withstand the temperature and caustic properties of the cleaning solution.
SCAHTUM-03 21 SCA 1200HT User Manual
Figure 11: Drain Valve
Page 22
9.3: Disposing of Solution
Used solution must be disposed of under applicable local waste disposal regulations. It is
the responsibility of the user to determine and verify what the local disposal
regulations are and to follow those regulations.
Proper disposal requires that the alkalinity (pH) of the solution be reduced to allowable
levels before it is disposed of. The pH may be lowered by either diluting the solution or by
neutralizing it by adding an acid. Litmus paper or an alternative method of measuring pH
levels should be used to d etermine if the p H level is within allow able levels bef ore disposal.
If you used the WaterWorks Soluble Concentrate, it is recommended that you dilute the
solution at a ratio of 5 parts clean water to 1 part used solution. More dilution may be
required, please check your local regulations.
If you are using Ecoworks cleaning agent, a dilution ratio of 1 part water to 1 part used
solution is sufficient for most applications. More dilution may be required, please check
your local regulations.
If you are ever unsure as to which solution was used in your SCA, assume that the
WaterWorks Soluble Concentrate was used and dilute accordingly.
9.4: Inspecting and Cleaning Drain S tr ainer
The strainer should be cleaned ev ery time the tank is refilled so it is clear of debris. You
should never remove the strainer when the tank is full of dirty liquid. Doing so will allow
debris to get into the drain tube.
The proper gloves and eyewear should always be worn when cleani ng and removing the
strainer. Never place your face near the tank when removin g the lid - vapors from the
cleaning solution may cause eye and respiratory irritation.
Figure 12: Strai n e r Removal
To remove the drain strainer, simply lift it up and out of the tank drain.
To clean debris from the strainer, scrub it with a nylon brush or a tooth brush and rinse
with warm water. Make sure that all holes are unplugged.
9.5: Inspecting and cleaning the pum p int a k e s cr e en
The pump intake screen may become clogged with debris. This will noticeably reduce the
flow in the tank.
SCAHTUM-03 22 SCA 1200HT User Manual
Page 23
Remove the intake screen by pulling down with sufficient force to unseat the fastening
Figure 14: Lifting the Unit
Figure 13 Removing pump intake screen
clips from the mating surface. Scrub debris from the intake screen using a stiff brush.
Rinse with water. Verify all holes are unplugged before replacing the screen on the pump
intake.
9.6: Refilling the Tank
Refill the tank and add cleaning solution as described in sections 9.1 and 9.2.
9.7: Cleaning
The SCA should only be cleaned with mild soap and water with a sponge or cloth. Never
immerse the system or use a spray nozzle or hose to clean it. Any liquid in the electronics
area of the system may damage the system and void the warranty. Before using any
cleaning or decontamination method except those recommended by the manufacturer,
users should check with the manufacturer that the proposed method will not damage the
equipment.
9.8: Moving
Always use two people to lift the SCA. The SCA should always be drained before moving.
To lift, have each person stand on either side of the unit and place their hands into the
indentations under the unit. Lift the unit to a cart or wheeled table for transport.
SCAHTUM-03 23 SCA 1200HT User Manual
Page 24
10: Troubleshooting
High Risk of Electrical Shock
Probable
Causes
Power Indicator
Power switch is not
“On”
Press rocker switch on the right-hand side of the
unit to the “On” position.
Power cord is not
wall
Check the power cord and make sure it is pushed
is securely connected to a grounded wall socket.
Fuse is blown
Contact technical support.
Power circuit has
Check your building circuit breakers, and any
replace the breaker or fuse as required.
Power indicator has
If the time and temperature displays are lit, and the
technical support.
Pump and
Power is not on
Problem.
Timer has not been
The system will only run when the timer is counting
add time and start the system.
Liquid level is too
Check the indicator lights on the right side of the
LIQUID FROM THE TANK.
The SCA has been designed to allow the end user to resolve most operating problems.
When a problem is encountered, please read through this troubleshooting section. If a
resolution is not found, please contact technical support.
Always disconnect the unit from power
before removing the back panel!
Problem
is not Lit after
pressing the
button
heater will not
start
connected to unit or
tripped
failed
What to Do
all the way in to the receptacle on the system and
power strips that the unit is plugged into for a
tripped circuit breaker or blown fuse. Reset or
rocker switch on the side of the unit is ON, the
power indicator has probably failed. Contact
Check the power indicator on the front control
panel. If it is off press the button. If still unlit,
then proceed to the “Power Indicator is not Lit”
set and started
high or too low
SCAHTUM-03 24 SCA 1200HT User Manual
down. Check the timer display on the control panel.
• If it is not counting down, press the
button.
• If the time being displayed is 00:00, then
display control panel. If either level indicator is on,
add or remove water from the tank until the
indicator turns off. Press the button to restart.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN ADDING OR REMOVING
Page 25
Problem
Probable
Causes
Liquid temperature is
Check the indicator lights on the right side of the
AND RESPIRATORY IRRITATION.
Over-temperature
A resettable over temperature sensor located on
technical support.
Pump or Heater
have failed
Contact technical support.
Buzzer is going
Liquid level is too
Check the indicator lights on the right side of the
LIQUID FROM THE TANK.
Liquid temperature is
Check the indicator lights on the right side of the
AND RESPIRATORY IRRITATION.
Sensor failure,
error
If the Alert Indicator light is also illuminated this is
What to Do
off
Silence the
buzzer by
pressing the
button
too high
Error Code:
EO = over
temperature
sensor has tripped
high or too low
display control panel. If the Alert light is on, open
the lid and lift the part basket from the tank. Let the
liquid cool and try starting the cycle again. If it
overheats a second time, contact technical support.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN MANIPULATING THE PART
BASKET.
NEVER PLACE YOUR FACE NEAR THE TANK
WHEN REMOVING THE LID, VAPORS FROM
THE CLEANING SOLUTION MAY CAUSE EYE
the back wall of the tank turns off the unit if the
temperature of the bath exceeds 95ºC. For
instructions on resetting this sensor, contact
control panel. If either level indicator is on, add or
remove water from the tank until the indicators go
off.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN ADDING OR REMOVING
too high
Error Code:
EO = over
temperature
power supply failure,
or pump failure.
Error Codes:
ES = Vcc out of
range, power supply
control panel. If the Alert Indicator light is on
remove the lid and lift the part basket from the tank.
Let the liquid cool and try starting the cycle again.
If it overheats a second time, contact technical
support.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN MANIPULATING THE PART
BASKET.
NEVER PLACE YOUR FACE NEAR THE TANK
WHEN REMOVING THE LID, VAPORS FROM
THE CLEANING SOLUTION MAY CAUSE EYE
indicative of one of several types of failures.
Contact technical support.
SCAHTUM-03 25 SCA 1200HT User Manual
Page 26
Problem
Probable
Causes
EO = over
sensor failure
Basket is raised or
If the basket is lowered into or raised out of the
until the button is pushed.
Flow in tank
Nozzle is clogged
Remove the nozzle and clean with water. If
COVERED IN CLEANING SOLUTION.
Pump intake screen
Drain the tank as instructed in the section
COVERED IN CLEANING SOLUTION.
Flow from
Nozzle is clogged
Remove the nozzle and clean with water. If
COVERED IN CLEANING SOLUTION.
White deposits
parts
The white-colored
cleaning solution.
Drain the tank and add fresh water and cleaning
fresh cleaning solution.
The cleaning
The cleaning
support material.
Drain the tank and add fresh water and cleaning
What to Do
temperature
r1 = temperature
sensor failure
r2 = High level
sensor failure
r3 = Low level
appears less
than normal
lowered too quickly
is clogged
cleaning solution too quickly, the solution does not
fill or empty out fast enough to maintain even water
level throughout the tank. This causes a temporary
high or low water level condition. The Alert
indication will go off once the water level reaches
acceptable limits but the unit will not operate again
necessary, clean holes in nozzle with wire brush or
small pin. Verify that all nozzle holes are clear
before replacing.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN WORKING INSIDE THE TANK
OR HANDLING COMPONENTS THAT ARE
“Maintaining your SCA”. Clean the intake screen
with a toothbrush until debris is removed from the
holes.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN WORKING INSIDE THE TANK
OR HANDLING COMPONENTS THAT ARE
nozzle is not
uniform
noticeable on
dark-colored
solution has a
strong odor
support material may
have saturated the
solution is likely
saturated with
SCAHTUM-03 26 SCA 1200HT User Manual
necessary, clean holes in nozzle with wire brush or
small pin. Verify that all nozzle holes are clear
before replacing.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN WORKING INSIDE THE TANK
OR HANDLING COMPONENTS THAT ARE
solution as described in the “Operating your SCA”
portion of this manual. Re-clean the parts using the
solution as described in the “Operating your SCA”
portion of this manual.
Page 27
Problem
Probable
Causes
After several
The cleaning
support material.
Drain the tank and add fresh water and cleaning
Type of cleaning
If using EcoWorks, try switching to WaterWorks.
a more efficient cleaning concentrate.
Liquid Level
from the tank
Sensors or
Contact technical support
System is
Drain strainer is
While wearing gloves, reach into the water and
Do not drain without the strainer.
What to Do
hours parts don’t
appear to be
much cleaner
indicators did
not turn off when
liquid was
added/removed
draining slowly
solution is likely
saturated with
agent.
Indicators have
failed
clogged
solution.
Both are sold by Stratasys resellers. WaterWorks is
remove any large pieces of debris from the drain
strainer. If necessary, clean the drain strainer with
a toothbrush without removing the strainer.
ALWAYS WEAR PROTECTIVE GLOVES AND
EYEWEAR WHEN WORKING INSIDE THE TANK.
Resume draining.
SCAHTUM-03 27 SCA 1200HT User Manual
Page 28
11: Techn ical Specifications
Item
Specification
Model Number
SCA-1200HT
Power Cord
110 North American
220 International
Power Requirements
100-120V~ 50/60 Hz 12A
220-240V~ 50/60 Hz 9A
Main Supply Voltage Fluctuation
up to +/- 10%
Transient Overvoltage
2500V
Electrical Protection
15A/250VAC Fuse
Heater Power
1,200 Watts
1,700 Watts
Regulatory Compliance
CE / UL / CSA / RoHS / WEEE
Tank Capacity
12.2 Gallons / 46.3 Liters
Physical Dimensions
26L x 17.5W x 20.5H inches / 66 x 44.5 x 52 cm
Shipping Package Dimensions
29.5 x 21.8 x 24.8 inches / 75.0 x 55.4 x 63.0 cm
Weight, Shipping Max
NW: 64 lbs / 29 kg
GW: 75 lbs / 34 kg
Large Parts Basket Capacity
10 x 10 x 12 inches / 25 x 25 x 30 cm
Small Parts Basket Capacity
4 x 4 x 4 inches / 10 x 10 x 10 cm
Allowable Liquid Solution
Temperature
Operating Environment Ranges
Temperature: 5ºC - 40ºC
Altitude: 0 M – 2000 M
Temperature Control Accuracy
± 2ºC
Pump Max Flow Rate
10 GPM / 38 LPM
Temperature Display
Digital LED Readout
Timer Display
Digital LED Readout
Liquid Level Checking
Fixed liquid level sensors with separate lamp indicators for
high and low level and audible alarm for both.
Temperature Checking
Over temperature sensor, alerts at 5°C above the set
factory set at 90°C.
Pump and Heater Safety
Lockouts
Low liquid level, high liquid level, over temperature, no time on
timer.
Tank Construction
Deep Drawn Stainless Steel
Ventilation Requirem ents
Must be operated in a well-ventilated space
Required Protective Equip ment
Thermal Gloves
Safety Glasses
Measurement (Installation)
Category:
Installation Category II
Pollution Degree
2
Protection Class
Class I
Marked Degree of protection for
IEC
For Indoor Use Only
10ºC - 85ºC
Humidity: 0% - 80% RH
temperature, audible alarm and lamp indicator. Thermal cutoff
Specifications are subject to change without notice.
SCAHTUM-03 28 SCA 1200HT User Manual
Page 29
12: Technical Support
Technical support for this product is provided by Phoenix Anal ysis & Design Technologies,
Inc. (PADT). Before contacting technical support please do the following:
1. Try the Troubleshooting table in this manual.
2. Visit www.SupportRemoval.com/support and try the troubleshooting found there.
3. Write down your model number, part number, and serial number (found on the back of
the unit).
If the unit is under warranty or covered by a maintenance contract, contact the Authorized
Reseller from whom the unit was purchased.
Otherwise, to receive technical support, visit www.SupportRemoval.com/Resources and
proceed to the “Request Technical Support” page and fill out the support request form there.
If you prefer to e-mail, you can send an e-mail to: sca@padtinc.com. Please include your full
name, company name, phone number and product serial number in all e-mails.
If you prefer to call, dial: 1-800-293-PADT and ask for SCA technical support.
12.1: Replacement Parts
Send email to sca@padtinc.com or call 1-800-293-PADT for information on obtaining
replacement parts.
SCAHTUM-03 29 SCA 1200HT User Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.