Oxford BenchMate E User Manual

Page 1
Page 2
Page 3
Table of Contents :
-2-
1. Introduction 3
2. Volume Range 3
3. Before you start 3
4. Getting Started 3
5. Overview 4
6. Modes 4
6.1 Favourites 4
6.1.1 Edit 5
6.2 Pipetting 5
6.3 Reverse Pipetting 6
6.4 Multi Dispensing 6
6.4.1 Repeat Dispense 6
6.4.2 Auto Dispense 6
6.4.3 Sequence Dispense 7
6.5 Diluting 7
6.5.1 Diluting 7
6.5.2 Diluting + Mixing 8
6.6 Mixing 8
6.6.1 Mixing 8
6.6.2 Pipette + Mixing 9
6.6.3 Diluting + Mixing 9
6.7 Settings 9
6.7.1 Speed 9
6.7.2 Sound 10
6.7.3 User ID 10
6.7.4 Product ID 10
6.8.1 3 Point Calibration 11
6.8.2 2 Point Calibration 11
6.8.3 1 Point Calibration 12
6.8.4 Restore Factory Calibration 12
6.8.5 Last Calibration Date 12
6.8.6 Due Calibration Date 13
6.8.7 Edit Password 13
6.8.8 Forgot Password 13
7. Warning / Safety Precautions 13
8. Battery and Charging 14
10. Maintenance 16
10.1.1 Volume wise dis-assembly procedure 16-18
10.1.2 List of spares 19-21
10.2.1 List of spares 23
10.2.2 List of spares 24
11. Autoclavability 25
12. Troubleshooting 25
Oxford Benchmate E
Electronic Pipette
User Guide
Page 4
-3-
1. Introduction
The Oxford BenchMate E electronic pipette offers you the best ergonomic design in the field of liquid handling devices. The ergonomic design of the electronic pipettes reduces the risk of repetitive strain injuries (RSI). The self explanatory functions of the pipette allow for easy access of each operation. The electronically operated piston movement reduces the manual error and provides high accuracy.
2. Volume Range
Please verify your box includes the following
Ÿ Electronic Pipette Ÿ Li-ion Battery Ÿ Charger Ÿ User Guide Ÿ Calibration Report Ÿ Toolkit
- Grease Container
- Tips
Please insert the Li-ion battery as shown in 8.1.Please charge the battery for 3 hours or unless the bar stops moving in the pipette display. Please follow charging instruction as shown in 8.2.
Please switch on the instrument by pressing the ON / OFF switch. Here you can start working in any modes of the pipette.
8 Channel
3. Before You Start
4. Getting Started
Number of
Channel
1 Channel 1 Channel
1 Channel 1 Channel 1 Channel
8 Channel 8 Channel 8 Channel 8 Channel
12 Channel 12 Channel
12 Channel 12 Channel 12 Channel
15-300
15-300
10-200
10-200
Page 5
8
7
5
6
9
10
1
3
4
Favourites
11
12
13
2
-4-
5. Overview
There are various modes available on the application menu. You can use them according to requirement.
Special Note :
Press “OK” button for selection. Press “ESC” button to come out from the mode any time.
1. Select Favorites from applications.
2. Select Prog1 / Prog2 / Prog3.
3. Select the mode.
4. Select volume by using up / down arrow key.
5. Select Aspiration Speed by using up / down arrow key.
6. Select Dispensing Speed by using up / down arrow key.
7. Select one of the 3 options. (a. Save & Run. b. Save & Exit. c. Do Not Save). a) Save & Run This will save the program and it will be ready for pipetting. b) Save & Exit This will save the program and will exit from menu. c) Do Not Save No program will be saved. You can use ESC key as well if you do not want to save this program.
6. Modes
6.1 Favorites
Steps :
1. Micro USB Charging Port
2. ON/OFF Switch
3. Battery Cover
4. Display
5. OK Key
6. ESC Key
7. Down Arrow Key
8. UP Arrow Key
9. Trigger Key
10. Tip Ejector Pusher
11. Tip Ejector
12. Tip Cone
13. Charging Point for
Auto Charging Stand
Page 6
-5-
This will allow the user to edit the sequence of the saved program. You can change the position of your program from 1st to 2nd or 3rd or vise versa.
This program allows you to edit your saved favorite program. Also you can change the sequence of saved programs.
Edit Program
This program allows you to edit the mode, aspiration speed, dispensing speed and volume. Also you can delete the program from this function.
Steps :
1. Select Favorites.
2. Select Edit.
3. Select Edit Prog.
4. Select any of the following options a) Mode : This will allow you to change the mode. b) Volume : This will allow you the change the volume. c) Speed : This will allow you to change the aspiration and dispensing speed. d) Save & Exit : Come out of the mode after saving the program. e) Do not save : It will not change anything. f) Delete Prog : This will delete the existing program.
1. Select Favorites.
2. Select Edit.
3. Select Edit Seq.
4. Select program position which you want to swap.
5. Press OK button.
6. Select the position with which you want to swap your program by using up / down arrow key.
7. Press OK button. You will be directed to the favorite mode where you can see the changes.
This will allow you to aspirate and dispense the selected volume.
1. Select Pipetting from the applications.
2. Select volume by pressing up / down arrow key.
3. Press OK to finalize the volume.
4. Press trigger key to aspirate the volume.
5. Press trigger key to dispense the volume.
6. Press trigger key to repeat the process or Press ESC key to come out from the mode.
6.1.1 Edit
Edit Sequence
Steps :
6.2 Pipetting
Steps :
Page 7
-6-
6.3 Reverse Pipetting
This mode is generally used for pipetting high viscous liquids accurately. It aspirates little excess liquid than the selected volume and it dispense the exact liquid of the selected volume. The rest of the liquid in the tip will need to be discarded.
Steps :
1. Select Reverse Pipetting from the applications.
2. Select volume by pressing up / down arrow key.
3. Press OK to finalize the volume.
4. Press trigger key to aspirate the volume.
5. Press trigger key to dispense the volume.
The excess liquid will remain in the tip.
6. Press trigger key to discard the excess liquid.
7. Press trigger key to repeat the process or press ESC key to
come out from the mode.
1. Select Multi Dispense from the applications.
2. Select Repeat Dispense from the Multi Dispense menu.
3. Select volume.
4. Select Dispensing Steps, the pipette will calculate the maximum number of steps once you select the volume.
5. Use up/down arrow key to modify the dispensing steps and press OK button.
6. Press trigger key to aspirate the liquid.
7. Press trigger key to dispense the liquid. The excess liquid will remain in the tip.
8. Press trigger key to discard the excess liquid.
9. Press trigger key to repeat the process or press ESC key to come out from the mode.
6.4 Multi Dispense
6.4.1 Repeat Dispense
The repeat dispense mode repeatedly dispenses the volume in selected increments. For better accuracy, it will aspirate the excess volume which will need to be discarded at the end.
6.4.2 Auto Dispense
Auto Dispense has a same function of Repeat Dispense but with an automatic timer setting.
Steps :
Page 8
-7-
Steps :
1. Select Multi Dispense from the applications.
2. Select Auto Dispense from the Multi Dispense menu.
3. Select volume.
4. Select Dispensing Steps. The pipette will calculate the maximum number of steps automatically once you select the volume.
5. Use up/down arrow key to modify the Dispensing Steps and press OK button.
6. Select Time Gap between 1 to 3 sec. and press OK button.
7. Press trigger key to aspirate the liquid.
8. Press trigger key to dispense the liquid. The excess liquid will remain in the tip.
9. Press trigger key to discard the excess liquid.
10. Press trigger key to repeat the process or press ESC key to come out of the mode.
1. Select Multi Dispense from applications.
2. Select Sequence Dispense from Multi Dispense.
3. Select No. of Sequences by pressing up / down arrow key (max. 4 sequences can be selected).
4. Select 1st volume by using up / down arrow key.
5. Repeat this process for 2nd, 3rd and 4th sequence selected.
6. Aspirate the volume by using trigger key.
7. Dispense the 1st sequence of volume.
8. Repeat the process till all the volume.
9. Discard the excess liquid by using trigger key.
This mode is used to dispense the liquid in different sequences.
Note : (If you choose nominal volume in the 1st selection then you won’t be allowed to select other sequences)
6.4.3 Sequence Dispense
Steps :
6.5 Diluting
This technique is used to mix two different liquids in one vessel.
Steps :
1. Select Diluting from applications.
2. Select Diluting.
3. Select No. of Dilutants (Max. 2 dilutants can be selected).
4. Select 1st volume by using up / down arrow key.
5. Select 2nd volume by using up / down arrow key.
6. Aspirate 1st volume by pressing trigger key.
7. Aspirate Air by pressing trigger key.
8. Aspirate 2nd volume by pressing trigger key.
9. Dispense total volume in vessel by using trigger key.
6.5.1 Diluting
Page 9
-8-
Steps :
1. Select Diluting from applications.
2. Select Diluting + Mixing.
3. Select No. of Dilutants (Max. 2 dilutants can be selected).
4. Select 1st volume by using up / down arrow key.
5. Select 2nd volume by using up / down arrow key.
6. Select Mixing Cycle by using up / down arrow key. Max. 10
cycles can be selected.
7. Aspirate 1st volume by pressing trigger key.
8. Aspirate Air by pressing trigger key.
9. Aspirate 2nd volume by pressing trigger key.
10. Dispense total volume in vessel by using trigger key.
11. Mixing will automatically start once you dispense the liquid
into the vessel.
12. You can stop mixing if require by using the ESC key.
13. Press trigger to start the mixing again / Press the ESC key to
completely stop the mixing in between the process. It will discard
the excess liquid remaining in the tip.
6.5.2 Diluting + Mixing
6.6 Mixing
You can mix the liquids using this mode. There are three modes under
mixing.
a) Mixing . b) Pipetting + Mixing. c) Diluting + Mixing.
6.6.1 Mixing
Steps :
1. Select Mixing from the applications.
2. Select volume by using the up/down arrow key.
3. Select Mixing Cycles by using up/down arrow key. Max. 10
cycles can be selected.
4. Aspirate the volume.
5. Mixing will automatically start .
6. You can stop mixing by using the ESC key.
7. Press trigger to start the mixing again / Press the ESC key to
completely stop the mixing in between the process. It will discard
the excess liquid remaining in the tip.
Page 10
-9-
Steps :
1. Select Pipetting + Mixing from the applications.
2. Select volume by using up/down arrow key.
3. Select Mixing Cycles by using up/down arrow key Max. 10 cycles can be selected.
4. Aspirate the selected volume by using trigger key.
5. Dispense the selected volume by using trigger key.
6. Mixing will automatically start .
7. You can stop mixing by using ESC key.
8. Press trigger key to start the mixing again / Press the ESC key to completely stop mixing in between the process. It will discard the excess liquid remaining in the tip.
6.6.3 Diluting + Mixing
Steps :
1. Select Diluting from applications.
3. Select Diluting + Mixing.
4. Select No. of Dilutants. Max. 2 dilutants can be selected.
5. Select 1st volume by using up / down arrow key.
6. Select 2nd volume by using up / down arrow key.
7. Select Mixing Cycle by using up / down arrow key. Max. 10 cycles can be selected.
8. Aspirate 1st volume by pressing trigger key.
9. Aspirate Air by pressing trigger key.
10. Aspirate 2nd volume by pressing trigger key.
11. Dispense total volume in vessel by using trigger key.
12. Mixing will automatically start once you dispense the liquid into the vessel.
13. You can stop mixing by using the ESC key.
14. Press trigger to start the mixing again / Press the ESC key to completely stop mixing in between the process. It will discard the excess liquid remaining in the tip.
6.7 Settings
6.7.1 : Speed
Steps :
1. Select Settings from the applications.
2. Select Speed.
3. Change the Aspiration Speed by using up/down arrow key and press OK.
4. Change the Dispense Speed by using up/down arrow key and press OK.
5. You will find new aspiration and dispense speed on the display.
6. If you do not want to change any speed, press ESC key.
This is a control menu where you can define Speed, Sound, User ID and Product ID of the pipette.
6.6.2 Pipetting + Mixing
Page 11
6.7.3 User ID
Steps :
1. Select Settings from applications.
2. Select User ID.
3. Select 1st character by using up / down arrow key and press OK. You can select letters from alphabets, 1-9 number and few symbols.
4. After pressing OK button, 2nd character will be highlighted. Follow the same process till maximum character.
5. Once you will select the last character, display will show the setting menu with all changes.
6.7.2 Sound
Steps :
1. Select Settings from the applications.
2. Select Sound.
3. Change the sound level by using the up/down arrow key and
press OK.
4. You will find the new selected sound on the display.
5. If you do not want to change sound, press the ESC key.
6.7.4 Product ID
Steps :
1. Select settings from applications.
2. Select Product ID.
3. Select 1st character by using up / down arrow key and press OK
button once you select 1st character. You can select letters from
the alphabet numbers 1-9 and few symbols.
4. After pressing the OK button, the 2nd character will be highlighted.
Follow the same process until maximum character limit is reached.
5. Once you will select the last character, the display will show the
setting menu with all changes.
-10-
6.8 Calibration
The Oxford Benchmate E pipette offers you 3 types of calibration
settings depending on the user’s requirement.
a.) 3 Point Calibration b.) 2 Point Calibration c.) 1 Point Calibration
By default 3 Point Calibration has been adjusted on the pipette for
supreme accuracy at each volume.
Page 12
-11-
6.8.1 3 Point Calibration
Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu by default, press
“0000” if you have not set your own password.
3. Select 3 Point Cal. from the sub menu.
4. Select Expected Volume 1. By default, 10% of the nominal
volume will be on the display, you can change this volume
according to your requirement by using up / down arrow key.
5. Enter Delivered Volume 1.
6. Select Expected Volume 2. By default middle volume will be
on the display, you can change this volume according to
your requirement by using up / down arrow key.
7. Enter Delivered Volume 2.
8. Select Expected Volume 3. By default nominal volume will be on
the display, you can change this volume according to
your requirement by using up / down arrow key.
9. Enter Delivered Volume 3.
10. Select “Yes” to save the changes. It will show Calibration
done on the display.
Select “No” to exit.
6.8.2 2 Point Calibration Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu by default, press
“0000” if you have not set your own password.
3. Select 2 Point Cal. from the sub menu.
4. Select Expected Volume1. By default, 10% of the nominal
volume will be on the display, you can change this volume
according to your requirement by using up / down arrow key.
5. Enter Delivered Volume 1.
6. Select Expected Volume 2. By default nominal volume will
be on the display, you can change this volume according to
your requirement by using up / down arrow key.
7. Enter Delivered Volume 2.
8. Select “Yes” to save the changes. It will show Calibration
done on the display.
9. Select “No” to exit.
Page 13
-12-
6.8.3 1 Point Calibration
Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu by default, press “0000” if you have not set your own password.
3. Select 1 Point Cal. from the sub menu.
4. Select Expected Volume1. By default, 10% of the nominal volume will be on the display, you can change this volume according to your requirement by using up / down arrow key.
5. Enter Delivered Volume1.
6. Select “Yes” to save the changes. It will show Calibration done on the display.
7. Select “No” to exit.
Steps :
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu. By default, press “0000” if you have not set your own password.
3. Select Restore Fact. Cal.
4. Select Yes by pressing up / down arrow key to restore the factory calibration. “Factory Calibration Restored” screen will appear for few seconds and you will directed to the calibration menu.
5. Select No by pressing up / down key to keep the user settings.
Special Note : Whenever the user makes any changes in the calibration, the Applications will show “Calibration User”. Once you restore the factory calibration, Applications will show “Calibration Fact.”.
6.8.4 Restore Factory Calibration
This feature will allow you to restore all the factory generated data of the product. So, all the user changes will be nullified and factory calibration will be restored.
Steps :
6.8.5 Last Calibration Date
Here you can keep a track of your last calibration date. Whenever you calibrate your pipette, you can enter the date of the calibration.
1. Select Calibration from the applications.
2. Select Enter Password from the sub menu. By default, press “0000” if you have not set your own password.
3. Select Last Cal. Date by using up / down arrow key.
4. Enter the date in DD/MM/YY format and you will be directed to the calibration menu.
Page 14
-13-
6.8.6 Due Calibration Date
Here you can save your due date for re-calibration of the product.
Steps :
1. Select Calibration Fact. From the applications.
2. Select Enter Password from the sub menu by default, press
“0000” if you have not set your own password.
3. Select Due Cal. Date by using up / down arrow key.
4. Enter the date in DD/MM/YY format and you will be
directed to the calibration menu.
Steps :
6.8.7 Edit Password
This feature will help user to edit the password. So only authorized persons can modify the calibration.
1. Select Calibration from the applications.
2. Select Edit Password.
3. “Enter Existing Password” by using up / down arrow key. The default password is “0000”.
4. Enter New Password by using up / down arrow key.
5. Repeat the New Password.
6. If everything is ok then “Password Change Successful” window will appear.
6.8.8 Forgot password
In case you forget the password, please contact your supplier. They will be able to reset your password.
7. Warning / Safety Precautions
To prevent fire, electronic shock or any serious injury to persons, follow below safety instruction.
Ÿ Do not submerge the unit and its charger in any liquid. Ÿ Do not operate the charger that has a damaged cord. Ÿ Please use chargers that are compatible with the pipette. Ÿ Do not handle the charger plug with moist hands. Ÿ Do not burn, crush, damage the Lithium battery or expose it
to excessive heat.
Ÿ Always use recommended battery.
Page 15
-14-
8.2 Charging - Power Adaptor
1. Insert the power cable in the pipette as shown in the image (A).
2. Connect the power plug of the power adaptor with the external power source.
8.3 Charging - Charging Pod
1. Insert the power cable in the auto charging pod as shown in the image (B).
2. Connect the power plug of the power adaptor with the external power source.
3. Place the pipette on the auto charging pod.
8.1 Inserting a Battery
1. Please follow the image to open the battery cover.
2. Insert battery supplied along with the instrument in a way that the metal pins inside the pipette establish proper contact with the metal points on the battery.
3. Slide the battery cover down.
8. Battery and Charging
Favourites
1 2
(A)
(B)
Page 16
9.1 Single Channel Pipette Technical Specification
-15-
Cat. No.
OBE-10
Test Volume
(µl)
Inaccuracy
(±) %
Imprecision
(±) %
0.5-10 µl
Volume
Range (µl)
1 5
10
2.5 2 1
1.8 1
0.6
OBE-100
5-100 µl
10 50
100
3 1
0.8
1
0.5
0.2
OBE-200
10-200 µl
20 100 200
2.5
0.7
0.6
0.7
0.3
0.2
OBE-1000
50-1000 µl
100 500
1000
3 1
0.6
0.6
0.4
0.2
OBE-5000
250-5000 µl
500
2500 5000
2.4
1.2
0.6
0.6
0.25
0.15
Color
Cat. No.
Test Volume
(µl)
Inaccuracy
(±) %
Imprecision
(±) %
0.5-10 µl
Volume
Range (µl)
1 5
10
8 4 2
5 2 1
OBE-8-20 OBE-12-20
2-20 µl
2 10 20
7 3 2
3 2
1.6
OBE-8-100 OBE-12-100
5-100 µl
10 50
100
3 1
0.8
2
0.8
0.3
OBE-8-200 OBE-12-200
10-200 µl
20
100 200
5 1
0.7
1.4
0.4
0.25
Color
9.2 Multi Channel (8-12) Pipette Technical Specification
OBE-8-10 OBE-12-10
OBE-8-300 OBE-12-300
15-300 µl
30
150 300
3 1
0.6
1
0.5
0.3
9. Technical Specification
Page 17
10. Maintenance
-16-
You should check the pipette at the beginning of each day for accumulation of dust and dirt on the outside surfaces. Attention should be paid to the tip cone. No other solvent except Iso-propanol should be used to clean the pipette. If the pipette is used daily, it should be inspected every three months.
The servicing procedure starts with the disassembly of the pipette. Please refer to the spare parts lists for better understanding of the components. After disassembling pipette as per mentioned procedure follow below routine maintenance steps :
Pull out the piston and other parts and follow general maintenance. Remember to keep all parts in order for reassembly.
Clean piston, piston spring, Seal and O-ring with a isopropanol and line free tissue. Allow them to dry.
Check the tip cone for foreign particles and remove, if any. Grease the cleaned parts with the approved lubricant provided with each pipette.
Volume 0.5-10µl
Step 1 Step 2
Unscrew the Tip Ejector
as per shown direction
Unscrew the Tip Cone
Cap as per shown direction
Step 3
Pull out the Piston as
per shown direction
10.1.1 Volume-wise Dis-assembly Procedure
Shown here are dis-assembly procedure for different volumes pipettes.
10.1 Single channel pipette maintenance
Page 18
-17-
Step 1 Step 2 Step 3
Unscrew the Tip Ejector
as per shown direction
Unscrew the Tip Cone
as per shown direction
Pull out the Piston as
per shown direction
Volume 50-1000µl
Step 1 Step 2 Step 3
Unscrew the Tip Ejector
as per shown direction
Unscrew the Tip Cone
Cap as per shown direction
Pull out the Piston as
per shown direction
Volume 5-100µl & Volume 10-200µl
Page 19
-18-
Unscrew the Tip Ejector
Cap as per shown direction
then pull out piston
as per shown direction
Volume 250-5000µl
Step 1 Step 2 Step 3
Unscrew the Tip Ejector
as per shown direction
Unscrew the Tip Cone
as per shown direction
Page 20
10.1.2 List of Spares:
Volume 5-100µl
OX AE 50033-3
OX CEP 40111-1
OX SP 00029-3
OX SP 00301-1 OX SP 00161-1
OX SP 00045-3
OX SP 00038-3
OX AE 50017C-1.1
-19-
OX CEP_SP 00041-3
OX AE 50102-1
OX SP 00028-3
OX SP 00158-1
OX SP 00038-3
OX AE 50017B-1.1
OX SP 00048-3
Volume 0.5-10µl
Page 21
OX AE 50017A-1.1
OX AE 50032-3
OX SP 00057-3C
OX SP 00302-1
OX SP 00159-1
OX SP 00044-3
OX SP 00217-1
-20-
Volume 50-1000µlVolume 10-200µl
OX AE 50017C-1.1
OX AE 50251-1
OX CEP 40111-1
OX SP 00029-3
OX SP 00303-1
OX SP 00164-1
OX SP 00046-3
OX SP 00038-3
OX AE 50102-2
Page 22
Volume 250-5000µl
OX AE 50024-1.1
-21-
OX CEP_CM 20126-1
OX AE 50109-1
OX SP 00164-1
OX RP 30156-1
OX SP 00110-1
OX SP 00050-3
OX AE 50018-1.1
Page 23
10.2 Multichannel pipette maintenance
-22-
Move upwards tip ejector ring lock (black part) to remove from bottom assembly. At the time of pipette assembly match two legs of tip ejector ring lock in pipette top half assembly.
Tip ejector ring lock
Page 24
10.2.1 List of Spare Parts
AEM 50112-1
SPM 0011-2L
RPM 30251-1
SPM 00009-2L
SPM 0016-1 (08CH)
SPM 0014-1 (12CH)
SP 00116-1 (FOR 08CH)
SP 00115-1 (FOR 12CH)
SPM 00207-1
SPM 00031-1
AEP 10281-1
AEP 10230-1
AE_SPM 00021-1 (08CH)
AE_SPM 0020-1 (12CH)
AE 50204-1
SP 00381-1
SPM 00321-1
SPM 00275-1
SPM 00276-1
SPM 00277-1
SPM 00158-1
SPM 00279-1 SPM 00280-1
SPM 00027-3 (15-300µl) SPM 00025-3 (10-200µl) SPM 00024-3 (5-100µl) SPM 00028-3 (0.5-10µl) SPM 00028-3 (2-20µl)
SPM 00120-1 (15-300µl 08CH) SPM 00119-1 (10-200µl 08CH) SPM 00118-1 (5-100µl 08CH) SPM 00117-1 (0.5-10µl 08CH) SPM 00117-1 (2-20µl 08CH)
SPM 00124-1 (15-300µl 12CH) SPM 00123-1 (10-200µl 12CH) SPM 00122-1 (5-100µl 12CH) SPM 00121-1 (0.5-10µl 12CH) SPM 00121-1 (2-20µl 12CH)
SPM 00319-1 FOR 15-300µl SPM 00317-1 FOR 10-200µl SPM 00316-1 FOR 5-100µl SPM 00320-1 FOR 0.5-10µl SPM 00320-1 FOR 2-20µl
SPM 00160-1 FOR 15-300µl SPM 00150-1 FOR 10-200µl SPM 00155-1 FOR 5-100µl SPM 00157-1 FOR 0.5-10µl SPM 00157-1 FOR 2-20µl
SPM 00131-3 FOR 15-300µl SPM 00130-3 FOR 10-200µl SPM 00127-3 FOR 5-100µl SPM 00132-3 FOR 0.5-10µl SPM 00132-3 FOR 2-20µl
-23-
RP 30080-3
SPM 00321-2
AEM 50260-1
SPM 00279-1 SPM 00280-1
CPC 00255-1
Page 25
10.2.2 List of Spare Parts
5-100µl 10-200µl 15-300µl
-24-
AEM 50112-1
SPM 0011-2L
RPM 30251-1
SPM 00009-2L
SPM 0017-1 (08CH)
SPM 0015-1 (12CH)
SP 00116-1 (FOR 08CH)
SP 00115-1 (FOR 12CH)
SPM 00207-1
SPM 00031-3
AEP 10281-1
AEP 10230-1
AE_SPM 00021-1 (08CH)
AE_SPM 0020-1 (12CH)
AE 50204-1
SP 00381-1
SPM 00321-1
SPM 00275-1
SPM 00276-1
SPM 00277-1
SPM 00159-1
SPM 00279-1 SPM 00280-1
SPM 00027-3 (15-300µl) SPM 00025-3 (10-200µl) SPM 00024-3 (5-100µl) SPM 00028-3 (0.5-10µl) SPM 00028-3 (2-20µl)
SPM 00120-1 (15-300µl 08CH) SPM 00119-1 (10-200µl 08CH) SPM 00118-1 (5-100µl 08CH) SPM 00117-1 (0.5-10µl 08CH) SPM 00117-1 (2-20µl 08CH)
SPM 00124-1 (15-300µl 12CH) SPM 00123-1 (10-200µl 12CH) SPM 00122-1 (5-100µl 12CH) SPM 00121-1 (0.5-10µl 12CH) SPM 00121-1 (2-20µl 12CH)
SPM 00319-1 FOR 15-300µl SPM 00317-1 FOR 10-200µl SPM 00316-1 FOR 5-100µl SPM 00320-1 FOR 0.5-10µl SPM 00320-1 FOR 2-20µl
SPM 00160-1 FOR 15-300µl SPM 00150-1 FOR 10-200µl SPM 00155-1 FOR 5-100µl SPM 00157-1 FOR 0.5-10µl SPM 00157-1 FOR 2-20µl
SPM 00131-3 FOR 15-300µl SPM 00130-3 FOR 10-200µl SPM 00127-3 FOR 5-100µl SPM 00132-3 FOR 0.5-10µl SPM 00132-3 FOR 2-20µl
-22-
RP 30080-3
SPM 00321-2
AEM 50260-1
SPM 00279-1 SPM 00280-1
CPC 00255-1
AEPM 10314-1 (15-300µl 08CH & 12CH)
Page 26
12. Trouble shooting
Trouble shooting
Issue
Probable Cause
Recommendation
Low battery symbol Battery is low Charge the battery
Blank Display
Low battery / Instrument in Standby mode / No battery / Battery damage / battery connection issue
Charge the battery / Press any key come out from Standby mode / Insert the battery / insert new battery / Remove and Insert the battery again.
Motor speed is set to the lower speed
High vibration of motor
Set the motor speed at highest level
Aspiration / Dispense Motor speed is set to the lower speed or different speed
Aspiration / Dispense time is slow / different
Set the motor speed at highest level
User change calibration / Tip issue
Wrong pipetting volume
Restore Factory calibration. See section 6.8.4 to restore factory calibration / Use recommended standard tips
password issue
Unable to ENTER into Calibration menu
Contact authorized distributor.
Not maintained regularly
Leakage
Open bottom assembly and perform routine maintenance
Excessive usage of pipette
Hard plunger movement
open bottom assembly and perform routine maintenance
Tip fitment issue
Use high quality standard tips
11. Autoclavability
The lower part of the pipette, Tip cone, ejector collar, piston and oring are autoclavable parts. Please autoclave under below condition.
Temperature : 121°C / 250° Fahrenheit Pressure : 1 bar / 15 psi (Pounds) Time : 25 minutes or less
After autoclave, allow parts to cool down for 8 hours in normal condition before reassembling it.
wrong tips attached to the Tip cone
-25-