This symbol is used throughout this document to highlight important safety considerations.
About this Manual
This user manual contains information you need to install and operate the Owlstone V-OVG system.
Additional information and updates are available from www.owlstonenanotech.com under the Vapour
Generator product tag. The website also provides details of permeation source availability as well as
software tools such as concentration calculators (see ‘Calculating Concentrations’ section on page 14).
1 Safety
This user manual contains important information regarding the safe operation of the Owlstone V-OVG. It is
essential that the user manual be read and understood before commencing any work with the system.
Using the V-OVG in a way that is not specified in this manual could be harmful to the health of the operator
and co-workers.
Swagelok® is a registered trademark of the Swagelok Company.
2.2 Disclaimer
Owlstone Ltd makes no representations or warranties, either expressed or implied, with respect to the
contents hereof and specifically disclaims any warranties, merchantability or fitness for any particular
purpose. Furthermore, Owlstone Ltd reserves the right to revise this publication and to make changes from
time to time in the contents hereof without obligation to notify any person of such revision or changes.
2.3 Notice of Proper Use of Owlstone Ltd Instruments
The supplied system is in compliance with international regulations. If this system is used in a manner not
specified by Owlstone Ltd, the protection provided by the system could be impaired
2.4 Contacting Owlstone
Visit the Owlstone website (https://www.owlstonemedical.com) for up to date contact details and service
support:
For general inquires please email support@owlstone.zendesk.com
This product has been designed and manufactured with high quality
materials and components, which can be recycled and reused.
This product is required to comply with the European Union's Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC and\so
should not be disposed of in normal waste.
For users outside the European Union consult local authorities for correct
disposal or contact Owlstone Ltd.
2.5 Recycling and Disposal
2.6 Certificate of Conformity
Owlstone Ltd performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations. When the system is delivered to you, it meets all relevant
electromagnetic compatibility (EMC) and safety standards as described in the declaration below.
Owlstone Ltd declares under its responsibility that the electronic product V-OVG is in conformity with the
following standards:
EMC Directive
The V-OVG system complies with the following standards:
CR47: 2006 Class A Code of Federal Regulations: pt 15 Subpart B – Radio Frequency Devices
– unintentional radiators
EN61326-1:1997 Electrical equipment for measurement, control and laboratory use – EMC
requirements, Group 1, Class B equipment (emission section only)
EN1326-1:1997 Electrical equipment for measurement, control and laboratory use – EMC
requirements, Industrial Location Immunity (immunity section only)
EN61000-3-2:2000 Electromagnetic compatibility (EMC) – part 3-2: Limits – Limits for harmonic
current emissions (equipment input current up to and including 16A per phase)
EN61000-3-3:1995 (+A1/A2) Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limitation
of voltage changes, voltage fluctuations and flicker in public low voltage supply systems for
equipment with rated current <= 16A per phase and not subject to conditional connection
Low Voltage Safety Compliance
This device complies with Low Voltage Directive EN 61010-1:2001.
Changes that you make to your system may void compliance with one or more of these EMC
and safety standards. Changes to your system include replacing a part or adding components,
options, or peripherals not specifically authorized and qualified by Owlstone Ltd. To ensure
continued compliance with EMC and safety standards, replacement parts and additional
components, options, and peripherals must be ordered from Owlstone Ltd or one of its
authorized representatives.
The V-OVG can be used with a wide range of permeation devices many of which could, if
they burst, release toxic or harmful quantities of the material they contain.
For this reason, it is essential that the user conduct a risk assessment for the
substances to be used in the V-OVG and establish safety protocols to cope with the
release of such materials both in the normal operation of the unit and in the case of a
permeation source bursting and releasing its contents all at once.
These protocols must include suitable installation (e.g. in a fume cupboard, provision of
extraction, etc.) and operational procedures to protect the operator.
FCC Compliance Statement
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at his or her own
expense.
3 Introduction
3.1 Overview
The V-OVG system comprises a modified 19” rack that is designed to house a series of instruments to
provide the functionality and adaptability needed to create a wide range of vapour standards.
The system is modular, the user must select the modules which they wish to employ from the range of
standard Owlstone products OVG-4, V-OVG, OFC-1 and OHG-4.
The OVG-4 is the Owlstone Vapour Generator. This unit incubates permeation devices at specified
temperatures in a known flow of gas to create an accurate and precise chemical vapour standard.
The V-OVG is a configuration of the Owlstone Vapour Generator with a large vertically mounted
oven for use with multiple, physically large or specialist permeation devices.
The OHG-4 is the Owlstone Humidity Generator, which can create and monitor a range of humidities
from 1* to 90% RH (* lower humidity depends on the %rh of the users gas supply).
The OFC-1 is the Owlstone Flow Controller, which provides an accurate diluent gas flow that
increases system functionality and compliments OVG-4,V-OVG and OHG-4 sub-units.
The OVG-4 / V-OVG is designed to house and incubate permeation devices at a set temperature, as well as
provide a controlled diluent gas flow to generate the desired chemical concentration of the vapour standard.
At the heart of the OVG-4 / V-OVG is the disposable permeation device (not supplied) which is usually
constructed from 1/4” PTFE tubing as shown in Figure 3. The oven can house a permeation device with a
diameter of up to 8mm and a length of up to 160mm.
In the device is a two-phase system the first phase of which is a liquid or solid reservoir of the desired
chemical generating a stable saturated headspace in the second gaseous phase. It is in this gaseous phase
that the chemical dissolves into, and permeates through, the walls of the tube at a constant rate.
Permeation devices are usually calibrated gravimetrically at a given temperature with the permeation rate
stated in ng min-1. Once the chemical vapour is released from the device it mixes with, and is carried away
by, a known diluent gas flow (ml min-1) and hence the desired concentration (ng ml-1) for the vapour standard
is created.
Calculating concentration with the Split Flow closed
The exhaust flow can be opened or closed dependent on user requirements. When the exhaust valve is
closed, the OVG-4 / V-OVG works as a normal vapour generator where the concentration (Equation 1) can
be altered by adjusting the sample flow.
(Equation 1)
Where [i] = Concentration / ng ml
PR = Permeation rate / ng min-1
FSA = Sample flow / ml min-1
-1
Figure 2 - Theoretical concentrations at different sample flows with a permeation rate of 100ng min-1
Calculating concentration with the Exhaust Flow open
With the exhaust flow open the concentration can be adjusted using the split control whilst the sample flow
remains constant. With the exhaust flow open the concentration is calculated by dividing the permeation
rate by the sum of the split and sample flow, therefore adjusting either flow will alter the concentration
(Equation 2).
Figure 3 outlines the different concentration ranges that can be generated by using the exhaust flow in
conjunction with both high and low sample flows.
Where [i] = Concentration / ng ml-1
PR = Permeation rate / ng min-1
High constant sample flow
Low constant sample flow
F
SA
F
SP
= Sample flow / ml min
= Split flow / ml min-1
-1
Figure 3 - Theoretical high and low constant sample flow concentrations at different exhaust flows.
Concentrations calculated using an analyte permeation rate of 100ng min-1
A concentration calculator is available at www.owlstonenanotech.com to simplify the setting of flows,
temperatures and splits in order to give a particular concentration for a particular source.
Figure 6 is a schematic of the flow path through the V-OVG.
The carrier gas first passes through the permeation oven where the air / analyte mix is achieved. This
gaseous mixture is then split into two separated flows, the Sample Flow (blue) and the Exhaust Flow
(green). The concentration of analyte exiting the Sample Outlet port is dependent on these two flows.
The Sample Flow is controlled by a mass flow controller with an accuracy of 1.5%. The Exhaust Flow is set
manually using the Exhaust control needle valve.
Warning! If power to the unit is interrupted there will be an increase in concentration of
chemical vapour levels within the unit.
If the unit is found in a powered down state it is suggested that the user purge the
permeation oven of excess vapour by opening the split flow fully for 2 hours.
If the gas supply is interrupted whilst a permeation source is being incubated within the VOVG, the concentration of the chemical will increase until it has reached a saturated level.
For this reason, any interruptions in the gas supply will result in the flow controller alarming,
when this happens the permeation oven is switched off to minimise vapour build-up.
If the system is found in this state, please close the sample outlet and purge the oven
through the exhaust (i.e. open the split flow control fully).
4 Pre-installation guidelines
4.1 Location
The V-OVG system is designed to create chemical vapours. To limit the user’s exposure to these vapours it
is strongly recommended that the V-OVG system be operated in a fume hood, or well-ventilated space.
4.2 Environment
Do not place in the following environments:
- Space that is poorly ventilated or confined. Allow at least 50cm clearance from walls and
free flow of air around the system
- Locations with an ambient temperature above 30oC
- Locations where the altitude is greater than 10,000 feet.
- Do not place the unit on fabric or any other soft surface
- Do not cover the unit with a cloth or any other item
- Do not place near flammable materials
- Where maximum relative humidity exceeds 80 %
4.3 Power
The V-OVG is supplied with a power supply that is automatically compatible with all conventional mains
power supplies: 100-240V, 50-60Hz, max 2.5A. It is not necessary to manually select or switch voltages. The
maximum power consumption of the V-OVG is 220 W.
The power supply is supplied with a 5A IEC mains lead with a plug appropriate for the country of use.
4.4 Gas supply
The V-OVG requires a pressure regulated supply of air / nitrogen at 40 - 90 psi. As the V-OVG is primarily
used to validate instrument detection capabilities it is recommended that the gas supply has a dew point
lower then -350C, is free from impurities (Hydrocarbons less than 0.1ppm methane) and particulates (less
then 30μm) . It is also recommended that all gas lines be constructed of refrigeration grade copper or
stainless steel tubing and connected using Swagelok fittings.
A separate exhaust line of at least ¼” OD and of <2 meters in length should be made ready
to connect to the exhaust outlet of the V-OVG system.
This line should be checked for chemical compatibility with the users vapours, it is
recommended that it is exhausted to a fume hood.
Check the condition of the exhaust line at regular intervals for blockages and leaks. The
customer is responsible for ensuring the integrity of any exhaust lines.
Check chemical compatibility: Materials in the flow path include PTFE, copper, stainless
steel, and Viton®. The user must ensure that permeation devices are neither corrosive nor
reactive with materials in the flow path.
Always refer to the Material Safety Data Sheets relevant to the vapour(s) you are handling
and ensure adequate risk controls and COSHH are in place before using potentially
hazardous vapours / gases with the V-OVG.
The V-OVG rack unit can weigh up to 25kg please take care in handling to avoid injury.
4.5 Exhaust
4.6 Analyte(s)
4.7 General Installation safety requirements
1. The V-OVG system will get warm during operation, especially the oven lid.
2. Do not place liquids on or near the V-OVG. Liquid spill may cause instrument failure.
3. Ensure cabling is routed behind the system, at bench level, posing no risk of tripping. Ensure
that all cables are detached from V-OVG before attempting to move the unit.
4. The V-OVG has not been designed for drop tests; any such test or accidental drop will cause
damage to the system.
V-OVG is designed to generate chemical vapours. To limit the user’s exposure to these vapours it is strongly
recommended that the V-OVG be operated in a fume hood, or well-ventilated space.
Ensure adequate space and air-flow around the unit. Do not use where ambient conditions exceed 30°C
Remove the nylon cap from the Air Inlet port on the rear of the system.
Connect a clean dry air supply to the stainless steel ¼” inlet situated on the side of the rack unit. It is
recommended that ¼” analytical grade stainless steel tubing be used; however refrigeration copper tubing or
PTFE may also be used.
Owlstone recommend the user fit a venting toggle valve in this line. It is important that the pressure supply
is removed from the system before attempting to open the V-OVG oven.
3. Exhaust line Installation
Remove the nylon cap from the Exhaust port on the front of each V-OVG sub-unit supplied with the system.
Connect a 2 meter exhaust line to this port venting into a fume-hood, with a ¼” Swagelok fitting. It is
important to note that using longer lengths of exhaust line or an exhaust under negative pressure will affect
the split flow exiting each V-OVG.
Connect the power supply to the back of the V-OVG Rack Unit and plug to your mains power supply. Ensure
that the mains plug is accessible during operation.
5. Remove caps from all Sample Outlet ports
To preserve the cleanliness of the flow path Owlstone ship the V-OVG system with Sample Outlet ports
capped. These caps should be removed before use.
6. Switch the power supply on
Following application of mains power, the user must switch on each V-OVG sub-unit. At this point the LED
displays of the three Eurotherm controllers will light up.
The needle valve controlling the exhaust flow is designed not to close fully. When it is in the
closed position there will still be an exhaust flow of ~30 ml min-1. This prevents vapour build
up in the oven and back diffusion.
Do not attempt to force the needle valve shut as this will damage the thread and the seal. Do
not cap off the exhaust outlet.
Exhaust Flow
displayed here
Exhaust Flow set here
7. Open the Exhaust Flow
To prevent a potentially dangerous build up of vapor in the permeation oven it is essential that a small Split
Flow is set before incubating your first permeation source.
Owlstone recommend at least 100 ml min-1 split flow at all times.
If during the installation procedure the user switches on a sub-unit without first removing the cap from the
Sample Outlet port then an ALARM will indicate on the Eurotherm Sample Flow controller.
In the case of V-OVG, when an alarm occurs on the Sample Flow controller the permeation oven is also
switched off to prevent vapour build-up. Once the temperature within the oven drops by more than 0.2°C
from set-point then the Temperature controller will also alarm.
To reset the alarms
1. First resolve the problem. The most likely cause of a flow alarm is that the Sample Outlet is
capped, or inadequate air pressure has been supplied to the rear of the V-OVG rack.
If this is not the case refer to the Troubleshooting Section of this manual.
2. The Sample Flow will recover to the set-point. Once the set-point is reached reset the alarm by
pressing the two buttons shown in the red circle below at the same time.
3. Once the Sample Flow controller alarm has been acknowledged the temperature of the
permeation oven will start to rise. The alarm on the Temperature controller cannot be
acknowledged until the temperature of the oven is within 0.2degrees of the set-point.
The V-OVG (Figure 8) is a versatile chemical vapour generator that when used in the GEN-SYS rack unit
either on its own or in a combination of two or three units can generate single or multiple gas mixtures.
The V-OVG comprises of two main sections, 1) the permeation oven and 2) the flow control system.
The oven chamber can hold multiple ¼” diameter PTFE permeation sources up to 12cm length, temperature
is digitally controlled from 30 to 100C ±0.1C in 0.1oC increments.
The flow control system comprises of Sample and Exhaust flows. The Sample flow is digitally controlled from
20ml min
-1
to 250ml min-1 in 1ml.min-1 increments, the exhaust flow is adjusted manually using the needle
The concentration of the chemical found in the gaseous phase of the device is dependent on the vapour
pressure of the compound. Given that the vapour pressure of a chemical is directly affected by temperature,
temperature is the main physical parameter controlling the permeation rate of the chemical from the device.
The permeation oven temperature is digitally controlled from 30 to 100oC in 0.1oC increments with 0.2C
variation. This 0.2C variation means a 98% accuracy of the certified permeation rate can be obtained.
As a general rule - if the incubation temperature is increased / decreased by 10C you double / half the
permeation rate of your device.
Using the Up-Down arrows on the temperature controller set your intended incubation temperature. Before
using your first permeation source, wait for the oven temperature to stabilize at set-point.
Temperature Control Alarms
The Eurotherm controller will alarm and display the message “Oven temperature outside set point” if the
temperature deviates by more than ±0.2C from the set point (see Figure 6).
To acknowledge the alarm press the Menu and scroll button together (see blue circle in Figure 5 above).
Using the Up-Down arrows on the temperature controller set your intended sample flow rate.
The total flow across the permeation source is the sum of the sample flow and exhaust flow.
The Sample Flow is set by using the up and down arrow keys (red circle) on the right Eurotherm controller
as shown in figure 8. The maximum and minimum flows are 500 and 50ml min-1 respectively
Figure 10 - Eurotherm Sample Flow controller
Sample Flow control alarms
A “Flow interruption alarm” indicates that the flow to the unit has been interrupted and that the actual flow
has deviated by more than 40ml from the set-point.
To protect the user from any potentially dangerous build up of chemical vapour inside the permeation oven
and flow path the permeation oven is automatically switched off whenever a flow alarm occurs.
To both reset the alarm and restart the permeation oven press both menu and scroll buttons together (refer
Important: Even though each part of the GEN-SYS is leak-tested and Swagelok fittings are
used throughout, the system is not guaranteed leak-proof.
Always refer to the Material Safety Data Sheets relevant to the vapour(s) you are handling
and ensure adequate risk controls and COSHH are in place before using potentially
hazardous vapours / gases with the V-OVG
Important:
It is recommended that each new device loaded into the permeation oven is allowed a 2
day incubation period before use.
When the V-OVG is incubating a chemical permeation source and sampling is not required,
always have the exhaust flow open. This ensures that a gas flow constantly passes over
the permeation source and hence prevents vapour build up.
6.4 Insert a permeation source
To insert a permeation source into the oven, first disconnect the air supply to the V-OVG and wait for the unit
to depressurise. (Sample and Exhaust Flows should read 0 ml min-1). Unscrew the black lid of the oven,
located on the top surface of the V-OVG.
Insert the permeation source into the central well of the oven, check the condition of the sealing O ring and
re-screw the lid.
Important: It is recommended that each new device loaded into the permeation oven is allowed a 2 day
incubation period before use.
When the V-OVG is incubating a chemical permeation source and sampling is not required, always have the
exhaust flow open. This will ensure that a gas flow constantly passes over the permeation source and is
routed to the exhaust port.
It is good practice to apply a label to the front of the V-OVG indicating to other users 1- chemicals present
inside the oven, and 2- the date when the permeation device was loaded
Warning: As a preventative measure to protect against unnecessary build-up of vapour
concentrations inside the oven on occasions when the Sample Outlet is capped, or outlet
pipe work blocked, always ensure the Exhaust Control is set to exhaust at least 100ml min–1.
Exhaust Flow Indicator
Exhaust Control needle
valve
Exhaust outlet port
6.5 Setting the Exhaust Flow
The Exhaust Flow is adjusted using the needle valve. The minimum achievable exhaust flow is
approximately 30ml/min. DO NOT OVERTIGHTEN THIS VALVE.
The Exhaust Flow indicator display gives a digital read-out indicating ml/min exhaust flow. For applications
where accurate exhaust flow measurement is required Owlstone recommend the user check actual exhaust
flow using a flow meter of known accuracy.
Before opening the sample outlet to the outside air verify that the concentration of the
substance you are using does not exceed the Occupational Exposure Limit.
6.6 Sample outlet
If required, the testing device can sample directly from the sample outlet. Otherwise suitable tubing is
recommended to transfer the sample gas flow to the target instrument / experimental set-up.
A 1/8” or ¼” Swagelok compression fitting is provided at the Sample Outlet port depending on the model,
this is intended for use with 1/8” or ¼” tubing. Ensure that the inner diameter of tubing is no less than 1mm.
Do not over tighten. For pre-swaged fittings, a leak tight connection can typically be achieved using
Swagelok fittings by following the procedure “finger tight followed by an additional ¼ turn using a spanner”.
For new fittings, tighten to finger tight then an additional ¾ turn.
To increase the range of concentration that can be generated additional diluent gas can be introduced
downstream of the sample outlet. The Owlstone Flow Controller (OFC-1) is designed for this purpose.
7 Preventative Maintenance
7.1 V-OVG system rack
Maintenance requirements for the V-OVG system rack are limited however Owlstone recommend the
following tasks are performed periodically:
Check that all internal cables are intact with no damaged insulation or frays.
Check that all pipe-work is in good condition, check all Swagelok fittings are tight.
Cleaning: Clean the outside of the equipment carefully using a cloth dampened with water, do not
use chemical cleaning agents. Before using any other cleaning or decontamination method, check
with your local Owlstone representative to make sure that the proposed method will not damage the
equipment.
If potentially hazardous material is spilt onto the equipment, disconnect the power supply and have it
checked by a competent person. It is the user's responsibility to carry out appropriate
decontamination if hazardous material is spilt on the equipment.
Check that the V-OVG rack is situated correctly. The unit should be situated clear of all other
equipment with adequate air flow around the rack. The system should not be used as a shelf.
After removing device, set temperature to
100C and have the exhaust and sample
flow at 250 ml min
-1
Recalibration
Annual
Contact Owlstone. Temperature and flow
controllers will be recalibrated
Warning: Inlet gas line pressure must be vented before following this procedure.
7.2 V-OVG sub-unit
Although the V-OVG is designed to be run continuously at the set point temperature with minimum
maintenance, to ensure prime performance it is recommended that the following steps are taken:
8 User Maintenance Procedures
8.1 Disconnecting power from the GEN-SYS rack
This section describes how to disconnect the GEN-SYS rack from air, exhaust and power services.
1) Vent the gas supply line that provides pressure to the GEN-SYS system. Do not proceed unless
pressure has been vented from this line.
2) Using a 9/16” spanner disconnect the inlet gas line from the rear of the GEN-SYS system.
3) If in use, using a 9/16” spanner disconnect the exhaust line from the rear of the GEN-SYS system
Customers are expected to make their own assessment of COSHH / chemical safety
before following this procedure. Customers should also consider undertaking a risk
assessment before attempting procedures described in this document.
Always physically disconnect Inlet Air, Exhaust and DC power from the rear of the GENSYS rack before following this procedure.
Always ensure sub-units are switched off before removing them from the GEN-SYS rack.
Always allow V-OVG sub-units to cool before removing them from the GEN-SYS rack.
Warning: The user is responsible for considering all chemical safety implications / COSHH
before following this procedure.
Always disconnect DC power, exhaust and gas supply from the rear of the GEN-SYS rack
before following this procedure.
Always consider chemical safety before following this procedure.
Switch off power to
the unit here
8.2 Removing a sub-unit
Note on safety
To improve access, it is often more convenient to remove any pre-existing V-OVG sub-units from the GENSYS rack. This is not essential and may impact the users COSHH evaluation.
1) Switch off power to the unit
Figure 12 – V-OVG front panel
2) Allow the V-OVG to cool down to room temperature
3) Unscrew and remove the 4x screws from the front panel of the V-OVG as shown below
Figure 13 – photograph showing the 4x screws used to mount the V-OVG
4) Remove the ‘roof’ panel from the rack
Unscrew the 4x screws shown below. The roof panel can then be lifted off.
Figure 14 - removing the roof panel
5) Disconnect DC power and gas supply
Both DC power and gas supply are removed by hand.
The DC power connector is removed by pulling gently. The gas supply QC is removed by sliding the tapered
end of the quick-connect towards the rear of the V-OVG
** If your system is not fitted with RS485 communication capability then you can ignore this step **
Figure 30 shows a GEN-SYS system fitted with only 1x V-OVG sub-unit. The 2x unused RS485 cables
along with the terminating connector are strapped to the base of the OVG rack.
To fit additional sub-units the RS485 connectors must be fitted to the back of each OVG as shown below in
figure 31. The terminating connector is not required if 3x sub-units are fitted into the GEN-SYS rack.
Figure 23 –RS485 system with single sub-unit Figure 24 – RS485 cable in orange
6) Replace the roof panel
Replace the roof panel and tighten the 4x screws.
7) Re-connect DC power, Air, Exhaust to the rear of the GEN-SYS system
8) RS485 addressing
Eurotherm controllers must each have a unique address on the RS485 bus. Owlstone will always configure
Eurotherm controllers supplied in a system with unique addresses. If a single sub-unit is supplied then the
addresses will be 1 and 2.
If the user adds further OVG -4 sub-units then each Eurotherm controller must be configured with a unique
address. This address is configured through the front panel of the Eurotherm (ref: Eurotherm user manual).
The fuse is mounted on the rear panel of the V-OVG.
Important notes:
The fuse is a safety device designed to protect V-OVG hardware from internal damage. The user should
never replace a fuse without understanding the root cause of failure.
Procedure
1. Remove the four screws holding the V-OVG in place.
2. Gently remove the unit from the GEN-SYS rack
3. Disconnect Air and Power from the rear of the V-OVG
4. The fuse holder is opened by twisting the cap ¼ turn
Internal to the V-OVG product is a mechanical thermal trip (MTT) that activates if the oven temperature
exceeds 115°C. Once the trip has activated the V-OVG oven will no longer heat up.
Important notes:
The MTT is a safety device designed to protect V-OVG hardware from internal damage. The user should
never reset without understanding the root cause of failure.
Always contact Owlstone for advice before undertaking this procedure.
Procedure
1. Switch off the V-OVG
2. Remove the four screws holding the V-OVG in place.
3. Gently remove the V-OVG from the GEN-SYS rack
4. Disconnect Air and Power from the rear of the V-OVG
5. On the rear of the V-OVG is a red button. Press firmly to reset the MTT
8.6 Resolving Temperature oscillations (AUTO TUNE)
1. Before you start the auto tune sequence the V-OVG temperature should be in the mid-range, Owlstone
recommend 60°. Use up and down arrows on the Eurotherm controller to enter the set point.
2. Depending on the starting temperature set point you may have to wait for an hour for the set point to be
reached before commencing the auto-tune.
3. Once the display shows 60° navigate to the auto tune menu option.
1) Press and hold the Page button until Level 3 is displayed (approximately 5 seconds). When
prompted on the screen enter access code 3 by using the up arrow. Wait a few seconds.
2) Next press the Scroll button until A.Tune (auto tune) is displayed on the controller screen.
3) Press the up arrow to set the A.Tune value to ‘on’. After a few seconds the auto tune sequence
will start, press the Page button to return to the normal view.
4. When the auto tune is complete the auto tune message will stop being displayed on the screen.
In the event of on-going unwanted oscillations contact Owlstone support.
Check that the external power supply has power and is
connected correctly to the rear of the V-OVG rack.
Remove the upper panel from the V-OVG rack and check that
all internal power leads are connected correctly to individual
sub-units (refer to sub-unit installation instructions)
Check sub-unit fuses, replace if necessary. Fuse ratings are
specified on the rating label on the rear of each sub-unit.
Contact Owlstone
Output flow does not reach
set-point
Check pressure supplied to the rear of V-OVG rack is at least
40psi
If sample flow is higher than the set-point:
There may be a problem internal to the unit, contact Owlstone.
If sample flow is lower than the set-point:
Disconnect the Sample Outlet from any downstream flow lines
(in the case of OHG-4 this port is labelled “Wet Air Outlet”). If
the sample flow recovers and is stable at the set-point then the
resistance of the downstream lines may be too great.
Consider increasing pipe diameters, shortening lines,
removing restrictions.
Check that there are no leaks inside the V-OVG rack.
Remove upper panel, check and tighten all Swagelok fittings.
(OHG-4 specific) If the sample flow is still below the set-point
then check that both Wet Air and Dry Air control valves are not
near closed and hence causing an unnecessary restriction.
Try opening both valves a little to reduce the restriction whilst
keeping the same %rh at the sample outlet.
Output flow is unstable
Check pressure supplied to the rear of V-OVG rack is at least
40psi
Run an AUTO-TUNE at your target flow, refer to section xx in
The V-OVG User Manual, as well as more information on the GEN-SYS modular vapour generation system
and permeation sources, is available at our support website:
http://owlstone.zendesk.com/home
For help with any further questions or problems, please contact Owlstone directly on our support address:
support@owlstone.zendesk.com
9.3 Warranty
Owlstone Ltd warrants that equipment supplied will perform to the advertised specifications for a period of 12
months. Should any material failure of the product occur within 12 calendar months of delivery Owlstone
Ltd will repair or replace the equipment free of charge given our satisfaction that sub-standard performance
is genuinely the result of defective material or workmanship and not caused by fair wear and tear
If required, please contact Owlstone Ltd for further detail regarding exact terms of the warranty.
10 Returns
10.1 Chemical Safety
Owlstone takes chemical safety seriously. Customers are not permitted to return hardware to Owlstone
without prior authorisation. Please consult your Owlstone representative regarding the returns procedure.
Before equipment can be returned customers are required to fill out a Decontamination Certificate and
submit to Owlstone for review. After H+S review Owlstone will issue a returns number (#RMA).
The Decontamination Certificate, p/n 90-0027, can be downloaded from our support website.
10.2 Packaging
Customers are encouraged to retain the original equipment packaging in case a future need arises for
equipment return. Owlstone cannot take responsibility for transit damage to poorly packed equipment.
For further advice regarding correct packaging of Owlstone equipment please contact Owlstone.