Introduction
You have just received your Neuro One sound processor, commercialised by Oticon Medical. This manual is intended for
the patient, the caregiver and the audiologist by providing important information and instructions on how to use the
Neuro One sound processor.
• Patient : person implanted with a cochlear implant, non-healthcare professional, without relevant specialized training
• Caregiver: adolescent or adult with a minimum of 8 years of education, non-healthcare professional, without relevant
specialized training
• Audiologist: healthcare professional with at least 3 years of studies in audiology, with specialized training in the
usage of the device
This manual for the Neuro One sound processor explains how to use the device as well as the precautions for use to be
observed. If the information provided in this manual and in the packaging of the Neuro One sound processor is incomplete or illegible, please contact the manufacturer or your local distributor.
Indications
The Neuro One sound processor is designed for adults and children of all ages, diagnosed with unilateral and bilateral
severe to profound sensorineural hearing loss (and hearing aid use does not provide adequate amplification for speech
audiometry), fitted with a Neuro Zti cochlear implant.
Contra-indications
The Neuro One sound processor is not appropriate for patients fitted with an implant other than that specified in the
indications. The Neuro One sound processor is not appropriate for patients with implants whose auricle is too small
and/or too soft to support the device.
Reliability – quality
The cochlear implant system is an active implantable medical device. To ensure your safety, the Neuro One sound
processor is designed and manufactured in accordance with the requirements of European Directive 90/385/EEC.
The device conforms to standard EN 45502-2-3: 2010, on Active Implantable Medical Devices (AIMD) – cochlear and
auditory brainstem implant systems, as well as standards relating to the safety of Medical Devices.
6
Page 7
Implant Compatibility
The Neuro One sound processor is only compatible with the Neuro Zti implant.
Note: it is important to always keep the cochlear implant patient card with you provided by your implant centre and to show it at the time of any examination or medical
treatment.
In the case of a bilateral cochlear implantation (each ear has a cochlear implant system):
patients wearing two cochlear implant systems are strongly advised against using a sound processor programmed for
one ear on the opposite ear. The implant is designed to recognise the sound processor which has been specifically
configured for it (paired). If the sound processor is changed, the implant will not work. To facilitate the recognition of the
left/right sound processor and prevent switching them, different colour foolproof keys (stickers) are provided with the
product. The Left/Right stickers need to be stuck on a part of the sound processor which is not in contact with the skin.
Finally, in case of a fault with one of the sound processors, contact Oticon Medical or your local distributor for a solution.
Neuro One sound processor
compatible with the implants
opposite:
Neuro Zti Cochlear Implant
7
Page 8
How Neuro One Works
The cochlear implant system consists of an external part and an internal part.
External part
• A Neuro One behind-the-ear hearing device (1) with two microphones.
• An external antenna (2) connected to the sound processor via an antenna cable (3).
The sound captured by the microphones is digitally processed by the sound processor.
The stimulation signal is then sent to the implant via the antenna through the skin.
Internal part
The Neuro Zti implant is a small casing (4) placed under the skin. It contains an electronic stimulator which will
distribute the sound to the electrodes (5) placed in the cochlea.
3
Antenna Cable
Antenna
2
4
Implant
Neuro One sound processor
8
1
To the brain
5
Electrodes
Page 9
Packaging contents
1. Neuro One sound processor x 1
2. Antenna with magnet x 1
3. Antenna Cable x 1
4. Case x 1
5. "Perfect Dry" heating desiccation drying system x 1
6. Zinc-Air battery boxes x 3
7. Accessories box x 1 containing:
7.a Additional antenna cable x 1
7.b Additional antenna x 2
7.c Additional magnet x2
7.d Additional ear hook x 4
7.e Holding buckle x 1
7.f Microphone earphones x 1
7.g Screwdriver x 1
8. Carrier x 1 containing:
8.a Cleaning wipes x 2
8.b Cover Clip x 1
N-A 6 Battery blister pack x 1
9. USB memory stick for backing up the settings
that also contains the Accessories user manual
(9.a) x 1
10. Sound processor User Manual x 1
11. Sound processor registration form (activation of
the warranty) x 1
12. Silhouette x1
13. Identification Stickers Left/Right
Pictures are not contractually binding
Note: The references are provided in this IFU in the section “References of Neuro One sound processor, spare parts and accessories”
9
Page 10
Sound processor Description
A. Microphone 1
B. Microphone 2
C. Setting selector wheel with on-off switch
0: Stop
1: Position 1
2: Position 2
3: Position 3
D. Programme button: choice of programmes (press once
to select the following programme P1-> P2-> P3-> P4-> P1
etc.) – checking that the system works (press for
longer)
E. Orange indicator light
F. Antenna socket: connection of the antenna cable
G. Auxiliary socket: connection of external accessories
(FM systems, microphone earphones, etc.)
H. Connection socket for the settings
The sound processor is identified visually by its marking
and serial number.
The serial number of your Neuro One sound processor is indicated on the inside of
the battery compartment: “Pxxxx” (for example P1495/P = sound processor
generation / xxxx = serial number).
It is recommended that you write down and keep this number, as you may need
it in the future.
A
B
C
D
E
F
G
H
Note: All parts indicated above are applied parts.
10
Page 11
Using the Neuro One sound processor
1. Connecting the components
Both ends of the antenna cable are identical to avoid any confusion for the user. It is to be used according to the
following recommendations:
• Connect the antenna to the antenna cable.
• Connect the antenna cable to the sound processor via the antenna socket.
The ends of the antenna cable and of the Neuro One sound processor antenna are equipped with a foolproof key.
The connection of the antenna cable to the sound processor, as well as the antenna cable to the antenna, must ensure
that the foolproof key is facing in the right direction with respect to the socket and the cable (see paragraph "Sound
processor description").
Important: the antenna cable connectors are designed to be used occasionally (change of faulty cable, etc.). To prolong
this accessory's service life, we recommend that you store your Neuro One sound processor in its case without disconnecting the antenna cable.
2. Optimising hold
Magnet
The antenna is held on the skin opposite the implant by means of a magnet.
You can adjust the distance between the skin and the magnet at any time by tightening or unscrewing the magnet.
To screw or unscrew the magnet, here are some instructions you can follow:
• Hold the antenna by grasping the magnet cap by its top and bottom, as shown in photo 1 below.
• To tighten the magnet, turn it in the opposite direction indicated by the arrows under the magnet cap (photo 2).
Photo 1Photo 2
11
Page 12
The underside of the magnet indicates the magnetic force:
top part of the magnetbottom part of the magnet
5
+
If the magnet force is insufficient, the antenna may not be held on the skin properly. It is then recommended to change
the magnet force (higher force).
If it is too strong, an irritation or skin redness may occur. A lower force (delivered in the box) may be advised.
If the magnet does not seem suitable, contact your implant centre which will advise you on the suitable force that
you can order from your local distributor or the Oticon Medical after-sales service.
Ear hook
The sound processor is provided with an ear hook to ensure that it is held on the ear. So that it can be adapted to each
patient, a different length of ear hook is provided in the box. The ear hook can be changed as follows:
• Remove the used ear hook.
• Clip on a new ear hook until you hear a "click".
Holding buckle
For an optimal fit of the sound processor on the ear (notably for paediatric use or active/sport activities),
we recommend that you use the silicon holding buckle provided in the Neuro One accessories box.
It is recommended that you refer to the “Accessories Instructions for use” to see how to fit this.
Note: The sound processor may also be held by using an open ear tip made by a hearing-aid specialist.
12
Page 13
3. Powering the sound processor
Type of batteries that can be used
Your Neuro One sound processor works using 2 non-rechargeable 675 Zinc-Air batteries. It is important to only use "675"
(European designation) or "PR44" (international designation) Zinc-Air batteries, recommended by your implant centre or
Oticon Medical. Using another type of battery risks affecting the proper operation of the Neuro One (reduced battery life,
sound processor power supply fault).
For restocking on batteries, you can directly contact your local distributor or the Oticon Medical after-sales service.
Battery life
The battery life is variable depending on the patient, setting, acoustic environment and use of the device on a daily
basis.
User instructions for batteries
It is recommended that you keep the batteries in a dry place away from temperature variations and observe the expiry
date indicated on the packaging. The two batteries used for powering the Neuro One must be the same type and be
changed at the same time. It is important to remove the batteries when your sound processor is switched off for a
prolonged period (for example overnight).
The Zinc-Air batteries cannot be recharged and are for single-use, and must not be thrown away but returned to a
collection point (point of sale, waste disposal site, etc.) or in the containers provided for this purpose. Refer to current
legislation.
Important: The batteries are small components which could be a hazard if swallowed. Keep the batteries out of the
reach of children.
13
Page 14
Private alarm
The Neuro One sound processor indicates the end of the battery life by consecutive audible beeps (private alarm) that
cannot be heard by those around you. Your clinician must first configure this functionality at the time of an adjustment, so
that you may benefit from this private alarm. Two consecutive series of three beeps notify you of the low battery life 30
minutes before the end of the batteries' life. The following six beeps indicate the end of the batteries' life and the cut-off of
the sound processor. On hearing the first two series of audible beeps or if the sound stops or has deteriorated, it is highly
recommended that you change the two batteries. These discharged batteries could spill corrosive liquid that may damage
your sound processor.
Battery compartment
The Neuro One sound processor is equipped with a battery compartment locking system, to keep the battery compartment closed while wearing or if the sound sound processor is dropped. This closure system is optional and may be
activated or deactivated at any time using the specific screwdriver provided.
• To lock the battery compartment: make sure that the compartment is correctly closed. Use the screwdriver provided in
the box to turn the screw to the locked position (illustration below).
• To unlock the battery compartment: Use the screwdriver to turn the screw to the unlocked position. You can now freely
open and close the battery compartment.
Inserting/changing batteries:
1. Switch off the sound processor (selector wheel on "0")
2. Deactivate the lock, if required.
3. Remove the used batteries and place the two new batteries (remove the protective film) in the compartment observing
the polarity indicated on the compartment (the batteries can only be inserted on one side). Reactivate the lock, if
required.
14
Page 15
You can then switch on your sound processor (turn the selector wheel towards the top to the appropriate position).
21
The lock is not necessary to keep the compartment closed, but comes as an additional feature. This feature does not replace adult supervision. If the sound processor falls
down, verify that the battery compartment is still locked. If necessary, reactivate the lock. Regularly clean the battery compartment so as to prevent the deposit of any
substance.
3
+
Zinc Air 675
4. Switching on/off
Turn the selector wheel towards the top to switch on the sound processor.
Select the programme using the push button (P1, P2, P3, P4).
Set the position of the selector wheel so as to optimise your listening comfort dependent on the acoustic environment
(according to the configuration defined with the clinician).
To stop the device, turn the selector wheel to the bottom so that it is in the “0” position, at which point a click can be felt.
In case the device is stopped for prolonged periods (for example during the night), we recommend observing the
following procedure:
• Switch off the device.
• Open the battery compartment and remove both batteries.
• Place your Neuro One in the desiccation drying system and follow the steps in the related instructions or in the
“Accessories user manual”, and close it until using the sound processor again.
15
Page 16
This operation enables the moisture accumulated during use of the device to be absorbed, the battery contacts to be
dried and therefore prevents possible malfunction of the sound processor.
Warning: This sound processor is equipped with an automatic restart device in case of malfunction. This leads to momentary power loss.
5. Programme button
The Neuro One sound processor enables four different programmes to be saved (P1, P2, P3, P4). The person responsible
for the settings can define these programmes according to your preferences and conditions of use (normal surroundings,
noisy surroundings, use of FM system, Telecoil, etc.).
Each quick press on the programme button enables the following programme to be accessed. The programmes follow
one another in the order P1, P2, P3, P4, P1 etc. The orange indicator light indicates the programme selected. It flashes
once for P1, twice for P2 and so on.
In the case where the clinician does not activate all the programmes, the sound processor will only select the programmes activated (for example, if only P1 and P2 are set, repeatedly pressing on the button will enable you to move
from P1 to P2, and then to P1, etc.).
This function may be locked on request during an adjustment session, according to your preferences or needs, so as to
simplify the use of your device, or prevent a child from changing the programme.
6. Setting selector wheel
This programmable function is used, according to the mode configured, to adjust your listening to the acoustic environment. Your clinician will set the most appropriate mode for your daily routine, and will indicate the suitable setting for
using each of the programmes.
Change the position of the selector wheel in order to find the most comfortable listening position. This function can also
be locked during an adjustment session, notably in the case of young children, to prevent any mishandling, or to
simplify the use of your device.
16
Page 17
7. Indicator light
The orange indicator light is activated during an adjustment session and enables the status of the sound processor to
be easily identified, notably for people who are not familiar with the functioning of the device (teachers, educators, etc.).
Start-up test
At start-up, the light illuminates twice, and then extinguishes.
Stimulation indication
The light regularly flashes depending on the stimulation rate of the implant.
This light configuration can be activated upon request by your programming clinician.
Programme selection
Each quick press on the programme button is followed by one or more consecutive flashes of the light, indicating the
programme selected (once for P1, twice for P2 etc.).
Detecting Battery discharge
The light flashes for 30 seconds to warn of the low battery life. It flashes again 30 minutes later to indicate the full
discharge of the batteries and the cut-off of the sound processor.
This configuration can be deactivated at your request by your clinician.
Detection of any other errors
In case of a significant error that may alter the stimulation, the sound processor restarts (leading to momentary power
loss) and then the light flashes quickly and extinguishes. If the problem persists, contact your local distributor or the
Oticon Medical after-sales service.
17
Page 18
8. List of additional accessories available
For more information about the accessories and their use, refer to the accessories manual available on the USB memory
stick provided with the kit.
Cover ClipThis Nylon and Spandex cover protects the sound
processor from moisture and perspiration, notably during
sporting activities or in case of a damp environment.
Carry caseThe case consists of several compartments that can be
used to store all of the necessary equipment (batteries,
accessories, etc.) for the system during your travels.
ScrewdriverThe screwdriver is used to activate or deactivate the
locking system of the sound processor’s battery compartment.
18
Page 19
Assistive Listening Systems
Assistive listening systems are used to improve the understanding of speech in noisy surroundings by reducing the
effect of distance between the sound source and the user.
1. Integrated telecoil
The Neuro One sound processor integrates a telecoil which enables enhanced listening in public places with a transmitter (theatres, conference rooms, cinemas, airports, etc.). This telecoil is activated by simply selecting the programme
specified beforehand with your implant centre (P1, P2, P3 or P4).
2. Auxiliary Socket
The Neuro One sound processor integrates a direct connection (auxiliary socket) that is compatible with various external
hearing aid accessories (FM system, contralateral microphone, etc.), such as those shown below, that are commercially
available.
The use of such a system requires an adjustment at your implant centre in order to define a new specific programme.
You then just need to connect your system to the auxiliary socket of the Neuro One and select the programme specified
by your clinician (P1, P2, P3 or P4). Contact Oticon Medical to know which FM systems are compatible with the Neuro
One sound processor. For further information, please contact your clinician.
If an accessory is failing, the sound processor stops stimulating and the indicator light flashes slowly for 1 minute, then
the sound processor cuts-off.
Disconnect the auxiliary accessory before starting the sound processor.
19
Page 20
Care and user instructions
1. Care
Precautions
The Neuro One sound processor contains high performance electronic components that can be damaged. It must
therefore be handled with care and not be exposed to extreme temperatures. Secondly, under certain conditions, the
temperature (T°C) of your sound processor may reach 42°C maximum instead of 39°C (temperature when working
normally). Therefore, you may experience a warm sensation. In this case, remove the external part from the ear and
switch it off.
Important: If the device is dropped, check that it then works normally.
If in doubt, do not try to repair it yourself. The device does not contain any component that can be repaired by the user
or the implant centre; any attempt to open the unit will automatically lead to cancellation of the warranty. In case of a
problem, contact the Oticon Medical after-sales service which is alone capable of carrying out maintenance on its
products.
It is important not to immerse your sound processor in water in order to avoid damaging the electronic components.
Therefore, never wear the device in the shower, bath or while swimming. In the event of dropping it in dirty water, switch
off the sound processor, remove the batteries, immediately rinse the equipment in clean water and dry for 72 hours.
Make sure to always carefully disconnect the antenna cable by holding the socket with two fingers. You risk damaging
the cable by pulling directly above. It is highly recommended to keep one or more additional antenna cables. We recommend that your device is serviced annually, which involves checking that it is working properly as well as checking the
calibration of the microphones, so as to optimise its service life.
Important: We strongly advise you to take out insurance to cover loss, theft or irreparable damage. Please contact your
implant centre for information.
20
Page 21
Cleaning
Clean the sound processor very carefully on a regular basis using a cotton cloth or, preferably, a stiff bristled
brush lightly moistened with alcohol. Make sure to protect the microphones from dust, dirt, sweat and
humidity. Cleaning wipes are also provided in this box.
These wipes are saturated with a bactericide and levuricide solution that enables earwax to be effectively
dissolved and dust and perspiration residue to be removed. For further information, refer to the "Accessories
user manual" provided on the backup media included in this box.
Storage
It is recommended to store the entire external equipment away from direct sunlight and heat. The
sound processor must be protected from dust, humidity, mechanical vibrations and impacts. On a
day-to-day basis, you can use your case to store the sound processor when you are not using it. You
just need to remove the batteries before positioning the device in the mould provided for this. For
further information, refer to the "Accessories user manual" provided on the backup media included
in this box.
Desiccation system
In case your sound processor is stopped for prolonged periods (for example during the night), we
recommend removing both batteries and placing the Neuro One in the desiccation box (Perfect Dry®)
provided for absorbing the residual humidity. Refer to the user instructions before use or to the
"Accessories user manual" provided on the backup media included in this box. The batteries should
be stored in a cool (<20°C) and dry place away from direct sunlight and heat.
Recycling
The main objective of Directive 2012/19/EU (WEEE) is the prevention of waste electrical and electronic equipment and in
addition, the re-use, recycling and other forms of recovery of such waste, so as to reduce the disposal of waste. The
"crossed out dustbin" symbol shown in the "Miscellaneous information" section indicates that the equipment cannot be
disposed of with household waste, but is subject to sorted waste collection. The equipment must be delivered to an
appropriate collection point for processing.
21
Page 22
Concerned about the preservation of our environment, Oticon Medical is a member of the environmental body Récylum
(France) for the collection and recycling of our devices. You can send us your system (or via your distributor) so that your
hearing system can be recycled.
2. Risks and user instructions
The sound processor must be removed when in the proximity of a strong magnetic field.
Accessories
The use of accessories that are not compatible may cause uncontrollable electromagnetic disturbance, or leakage
current towards the patient when an accessory, directly connected to the mains and without any electrical insulation,
is connected to the sound processor. It is therefore essential that you only use accessories with the sound processor
provided by Oticon Medical or those which are specified as compatible by Oticon Medical.
As the sound processor is made up of small components (cables, magnets, batteries, etc.), families with young
children must make sure to keep these components out of their reach to avoid any ingestion or inhalation.
Playing sport
- The cochlear implant system allows for participation in most sporting activities. However, precautions must be
taken, wearing a protective helmet, a system for holding the sound processor, or the removal of the sound processor
may be necessary depending on the sport. A severe impact may damage your cochlear implant system. Therefore, it
is not advised to participate in contact sports.
- Deep-sea diving below a depth of 20 m is also not advised.
Medical examinations
- The sound processor must be removed before any medical examination is carried out (radiotherapy, MRI,
ultrasound, scan).
- Your implanted receiver contains a removable magnet that is sensitive to electric currents. You must contact
your implant centre and Oticon Medical before any treatment using electric currents or any exposure to intense
radiation (radiotherapy etc.).
22
Page 23
- Your implanted system is compatible with 1.5 Tesla MRI imaging devices. It is also compatible with 3 Tesla imaging
devices but requires removal of the implant magnet prior to the exam. The MRI exam application form (available on
www.oticonmedical.com) must be completed by the radiology department involved, and returned to Oticon Medical
before any exam is carried out. The MRI exam has to be processed with head first.
Settings
Each sound processor is programmed by a clinician authorised by Oticon Medical for personal use and must not under
any circumstances be lent to or exchanged with another person.
Electrostatic discharge
Electrostatic discharges are the visible sparks when there is contact between two people or with an object. They are
encouraged by cold, dry climates. Contact with certain components can encourage the appearance of such discharges
(clothes made of synthetic material, getting out of the car, plastic toboggan, computer or television screens, carpets,
etc.). The Neuro One sound processor is designed to provide effective protection against these electrostatic discharge
phenomena (compliance with NF EN 60601-1-2: 2007) to prevent any damage to the device or alteration to the listening
programme. However, we advise you to take certain precautions to prevent the exposure of the sound processor to such
discharges: carefully remove a pullover, take care when getting out of the car, etc. Furthermore, any person that may
come into contact with the sound processor must make sure to touch the person wearing the cochlear implant system
before touching the device to remove any residual static electricity.
Passing through security gates
Security gates (airports, shopping centres, etc.) produce powerful electromagnetic fields.
Passing through or closely to these gates may trigger the detector alarm or disrupt the sound received by the person
wearing the implant. It is recommended that you switch off your sound processor and inform the security officers by
showing your patient card.
Travelling by plane
As with any electrical equipment, it is essential to switch off your sound processor during take-off and landing.
23
Page 24
Tools and Troubleshooting
1. Testing tools
Various functionalities and tools are made available to you in order to check that the Neuro One sound processor and
its accessories are working properly:
• Indicator light.
• Private alarm to detect the end of battery life.
• Self Check function to check the integrity of the system.
• Microphone earphones.
It is important to clearly understand the way these tools work in order to best manage your equipment on a
day-to-day basis.
Self Check function
Your sound processor integrates a function that is used to check that it is correctly sending the sound information to the
implant and to thus check that the system and its accessories are working properly (antenna and antenna cable).
The test is carried out as follows:
• Position the antenna on the skin, opposite the implant.
• To carry out the test, press and hold the push button for 6 seconds.
• If the test is successful, an audible beep rings and the orange led flashes for 15 seconds.
• If the test reveals a malfunction, the led will illuminate continually and then briefly flash, twice consecutively.
• In case of an anomaly, refer to the chapter "Problems and Solutions".
24
Page 25
Microphone earphones
The Neuro One sound processor is equipped with an auxiliary socket that enables one to listen to
the audio signals which enter the sound processor. This accessory is used to identify problems on
the microphone: microphone entry blocked, faulty microphone.
Warning: The quality of the sound heard with these earphones is not representative of the sound
heard by the patient with his sound processor. For further information, refer to the "Accessories user
manual" provided on the backup media included in this box.
This test tool is used as follows:
• Switch off the sound processor.
• Carefully connect the earphones to the sound processor's auxiliary socket observing the connection direction.
• Switch on the sound processor on a programme without any auxiliary input activated, to evaluate the sound quality
received.
• Once finished using the sound processor, switch it off again before disconnecting the earphones.
Note: There is a risk of audio feedback when the earphones are too close to the microphone. The sound level received by another listener using the earphones is not
representative of that heard by the patient. When using the microphone earphones, it is important that the sound processor should be turned to a programme without any
auxiliary input activated, this to avoid an overconsumption error on the sound sound processor.
2. Problems and solutions
The Neuro One behind-the-ear sound processor requires numerous adjustments to identify the optimal listening quality.
A check of the adjustment at least once a year is recommended to make sure that the implant system is working properly. For young children, parents and educators are specifically advised to supervise the behaviour of the child during
use of the sound processor. In case of an abnormal feeling or pain during the routine use of the device, it is recommended to change the hook or remove the sound processor from the ear and contact the implant centre.
The combined use of the various testing tools as well as implementing the instructions presented below can assist in
troubleshooting issues with the sound processor. If the problem persists, it is recommended that you contact your
implant centre, local distributor or the Oticon Medical after-sales service.
25
Page 26
No sound is received
1. Check the status of the sound processor (correct connection of the antenna cable, locking of the battery compartment,
position of the selector wheel, programme selected).
2. Change both batteries.
3. Check the quality of the microphone using the microphone earphones.
4. Check that the antenna cable and the antenna are working properly using the integrated Self Check function ("Checking that the system works", "Repair, Tools, Problems and Solutions" section).
5. Change the antenna cable or the antenna.
6. Contact your implant centre, your local distributor or the Oticon Medical after-sales service.
Intermittent sound
1. Check that the position of the selector wheel and the programme selection are suited to the listening conditions.
2. Change both batteries.
3. Check that the care instructions are being observed.
4. Check the condition of the antenna cable and the antenna using the Self Check function.
5. Contact your implant centre, your local distributor or the Oticon Medical after-sales service.
Reduced or distorted sound
1. Check that the position of the selector wheel and the programme selection are suited to the listening conditions.
2. Try another programme.
3. Change the batteries.
4. Check the quality of the microphone using the earphones.
5. Contact your implant centre, your local distributor or the Oticon Medical after-sales service.
Quick flashing indicator light for one minute
1. Change both batteries.
The sound processor restarts and the indicator light flashes quickly then extinguishes
The sound processor detects an error. If the problem remains, contact your local distributor or the Oticon Medical
after-sales service.
26
Page 27
Insufficient battery life (less than a working day)
1. Change both batteries.
2. Check the type of batteries used (type P675 or PR44).
3. Check that the care instructions are being observed ("Care and user instructions" section).
4. Contact your implant centre, your local distributor or the Oticon Medical after-sales service.
Problem of holding the antenna
The antenna is held on the skin opposite the implant by means of a magnet.
If it is held too tightly:
1. Loosen the magnet by turning it in the direction indicated by the arrows (cf. the Optimising hold" paragraph,
in the "Using the Neuro One" section).
2. Ask for advice from your implant centre for changing the magnet (low force magnet provided in the box).
If it is not held tightly enough:
1. Re-tighten the magnet by turning it in the opposite direction to that indicated by the arrows (cf. "Optimising
hold" paragraph, "Using the Neuro One" section).
2. Ask for advice from your implant centre in order to obtain a stronger magnet.
Note: The first time the magnet is unscrewed may prove difficult.
Important: Any equipment sent to your local distributor or the Oticon Medical after-sales service must be
accompanied by a detailed description of the symptoms.
Warning: In case of skin redness or irritation during the routine use of the device, it is recommended to change
the magnet force.
27
Page 28
Miscellaneous Information
1. Technical characteristics
Sound processorMaximum dimensions: 50.5 x 50 x 10.6 mm
Weight (with batteries): 11.4 g
Power supplyOperating range of the device: from 1.8 V to 3.7 V
2 disposable 675 Zinc-Air batteries
Maximum consumption: 60 mW
Maximum transmission power: 35 mW
Storage and Transport
Conditions
Conditions of useOperating temperature: 5°C to 40°C
Storage and transport temperature: -20°C to 50°C
Relative Humidity: 0 to 90%
Atmospheric pressure: 700 hPa to 1060 hPa
Relative Humidity: 0 to 90%
Atmospheric pressure: 700 hPa to 1060 hPa
28
Page 29
0459
2. Symbols
700 hPa
Refer to instruction manual.
Waste electrical and electronic equipment (WEEE)
Type B device
Switching the sound processor on/off
P675: "675" (European designation) or "PR44" (international designation) Zinc Air
batteries
+: Battery positioning direction
Operating instructions
Temperature limit to which the medical device can be exposed
Humidity limit to which the medical device can be exposed
1060 hPa
Atmospheric pressure
Manufacturer
Date of manufacture
Reference
Serial number
CE registration mark with notified body number. Indicates compliance with the
requirements of the Active Implantable Medical Device Directive 90/385/EEC
29
Page 30
3. Profile of the person handling the sound processor
The patient is the operator when he:
• Changes the batteries.
• Switches the sound processor on and off.
• Changes the programme.
• Connects and disconnects the antenna cable.
Children under seven years old must carry out these operations under the supervision of an adult.
4. Manufacturer's declaration
Medical electrical equipment requires special precautions to be taken for EMC (Electromagnetic Compatibility) and it is
necessary to install and start it according to the EMC information provided in the tables below. Portable RF (radio frequency) communication devices, such as mobile telephones, may affect the operation of the Neuro One sound processor. Therefore, these devices should be kept as far away as possible from the Neuro One and its accessories in order to
avoid such effects.
30
Page 31
Guidelines and manufacturer's declaration – Electromagnetic emissions
The Neuro One sound processor is intended for use in the electromagnetic environment specified below.
The Neuro One user should ensure that it is used in such an environment.
Emissions testsCompliance
CISPR 11 RF emissionsGroup 1
CISPR 11 RF emissionsClass B
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations and flicker emissions
IEC 61000-3-3
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Neuro One sound processor is intended for use in the electromagnetic environment specified below.
The Neuro One user should ensure that it is used in such an environment.
Immunity testTest level IEC 60601Level of compliance
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/60 Hertz)
magnetic field
IEC 61000-4-8
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
30 A/m (50/60 hertz)
Not applicable
Not applicable
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
30 A/m
(50/60 hertz)
UT is the A.C. mains voltage prior to application of the test level.
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Neuro One sound processor is intended for use in the electromagnetic environment specified below.
The Neuro One user should ensure that it is used in such an environment.
Immunity testIEC 60601 test level
Radiated RF
IEC 61000-4-3
±10 V/m from 80 MHz to 2.7GHz
80% AM at 1kHz
Level of
compliance
±10 V/m from 80 MHz to 2.7GHz
80% AM at 1kHz
31
Page 32
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
(MHz)
385380 – 390TETRA 400
450430 – 470
710
780
810
870
930
Band
(MHz)
704 – 787LTE Band 13, 17
800 – 960
ServiceModulationMaximum power (W)Distance (m)
GMRS 460,
FRS 460
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation 18 Hz
FM ± 5 kHz deviation
1 kHz
deviation
Pulse
modulation
217 Hz
Pulse
modulation
18 Hz
IMMUNITY
TEST LEVEL
(V/m)
1,80,327
20,328
0,20,39745
20,328
32
Page 33
Test
frequency
(MHz)
Band
(MHz)
ServiceModulationMaximum power (W)Distance (m)
IMMUNITY
TEST LEVEL
(V/m)
1720
1845
1970
24502400 – 2570
5240
5785
Date of last review: 2018-01
1700 – 1990
5100 – 5800
GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE
Band 1, 3, 4, 25; UMTS
Bluetooth, WLAN, 802,11
b/g/n, RFID 2450, LTE
Band 7
WLAN 802,
11 a/n
Pulse
modulation
217 Hz
Pulse
modulation
217 Hz
Pulse
modulation
217 Hz
20,328
20,328
0,20,395500
Essential performance:
The Neuro One sound sound processor should stimulate below the thresholds defined in DigiMap USB within a 10%
tolerance.
33