oticon CI-Link Instruction Manual

Page 1
Neuro – The Cochlear Implant System
CI-Link Programming System
Instructions for Use
0459
(2018)
NEURELEC
Page 2
2
Table of Content
Introduction ..................................................................................................................5
Intended use...................................................................................................................7
Compatibility .................................................................................................................. 7
Description of CI-Link .................................................................................................... 8
CI-Link overview ..............................................................................................................8
CI-Link unit front view ......................................................................................................9
CI-Link unit rear view ..................................................................................................... 10
Packaging content ........................................................................................................12
CI-Link Fitting Room Kit ..................................................................................................12
CI-Link Operating Room Kit (Zti) ......................................................................................13
When receiving CI-Link ................................................................................................ 14
Unpacking and inspection ............................................................................................. 14
Storage ........................................................................................................................ 14
Preparing CI-Link for use .............................................................................................. 15
Installation ....................................................................................................................15
Connecting the programming adaptor to the sound processor ........................................ 16
Page 3
3
Locking and unlocking the programming adaptor ............................................................17
Connecting two sound processors ................................................................................. 18
Wearing the sound processor on clothes ....................................................................... 19
Using CI-Link ............................................................................................................... 20
Disconnecting CI-Link from the PC ..................................................................................21
Visual indications........................................................................................................ 22
Care and maintenance ................................................................................................. 24
Cleaning .......................................................................................................................24
Equipment failure, service and repair ............................................................................ 25
Precautions ................................................................................................................. 26
Troubleshooting ...........................................................................................................27
Disposal ..................................................................................................................... 29
Risks and related instructions for use .......................................................................... 30
Physical and performance characteristics .................................................................... 32
Page 4
4
Symbols...................................................................................................................... 34
Profile of the person handling CI-Link .......................................................................... 36
Manufacturer's declaration ...........................................................................................37
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment .......................................................................................... 39
Page 5
5
Introduction
The CI-Link programming system (also named CI-Link) consists of the CI-Link unit, Genie Medical CI fitting software, and accessories.
The CI-Link programming system allows you as an audiologist or a clinical support, to make the necessary adjustments to sound processors connected to the CI-Link unit.
Audiologist: healthcare professional with at least 3 years of studies in audiology, with
specialized training in the usage of the device
Clinical support: Oticon Medical personnel, who is either speech therapist, audiologist or
engineer (master’s degree), expert and trainer of the device
This manual is your guide to installing and using CI-Link. We strongly recommend that you read this manual carefully before using CI-Link for the first time. Prior to use, we also advise you to consult the following documentation:
Genie Medical CI fitting software Installation guide, and Fitting guide
Neuro 2 sound processor Instructions for use
Note: The Genie Medical CI fitting guide is available on the Oticon Medical website and embed-
ded in the Help menu of the Genie Medical CI fitting software. Contact your local Oticon Medical customer service or Oticon Medical distributor to ask for a paper version to be sent to your address within 7 days at no additional cost.
Note: The patient should be continuously monitored when using the device.
Page 6
6
“Note”: indicating a note/tip
“Important”: indicating important information which should be followed by the user,
to prevent damaging the equipment.
“Warning”: indicating an immediate danger, potential safety hazard.
Action required.
Oticon Medical ( NEURELEC) reserves the right to make changes to the design, characteristics and models without prior notice. The only warranty Oticon Medical ( NEURELEC) makes, is the express written warranty extended on the sale or rental of its products.
Page 7
7
Intended use
The CI-Link programming system is intended, in a healthcare facility, to allow trained audiolo­gists and clinical supports to:
Fit the cochlear implant system
Perform objective measures, including impedance measurement and ECAP (ECAP available
with Neuro Zti implants only).
Compatibility
The CI-Link programming system is compatible with the Neuro Zti cochlear implants and Neuro 2 sound processors.
Note: If you require assistance in setting up, using or maintaining CI-Link please
contact your local Oticon Medical customer service or Oticon Medical distributor.
Warning: Never try to connect CI-Link with elements other than those provided.
Warning: Do not use or connect a sound processor or accessories other than those specified
as being compatible with the CI-Link programming system.
Page 8
8
Description of CI-Link
CI-Link overview
PC (USB 2.0 port) with Genie Medical CI fitting software
Neuro 2 sound processor(s) (without battery module)
CI-Link unit
Programming adaptor left (blue)
USB cable
Programming adaptor right (red)
Note: The programming adaptors indicated above are applied parts.
Page 9
9
CI-Link unit front view
Blue colour code
Indication for left connection
Mini-DIN connector (left)
Connection to left sound processor
“Left” Light indicator
Communication with left sound processor
Mini-DIN connector (right)
Connection to right sound processor
“Right” Light indicator
Communication with right sound processor
“Power” Light indicator
Communication with PC
Manufacturer
Product name
Red colour code
Indication for right connection
“PC Com” Light indicator
Power supply from PC
Page 10
10
PC / USB connector
Connection to PC
BNC trigger output
Connection to third-party equipment* (reserved for later use)
CI-Link unit rear view
* The third-party equipment must be compliant with the IEC 60601-1 standard with 1MOPP isolation, or with the IEC 60950 standard.
Pictures are not contractually binding.
Page 11
11
Important: We recommend not using a USB hub to connect the CI-Link unit to the PC as this
may affect power supply.
Important: PC must be compliant with the IEC 60950-1 standard.
Page 12
12
Packaging content
Different CI-Link packages are available. To ease identification, each package has its own marking.
CI-Link Fitting Room Kit
The CI-Link Fitting Room Kit includes the following:
1 CI-Link unit
1 USB cable (type A->B)
1 Instructions for use
2 Programming adaptors
2 Antenna cables
2 Clips
Page 13
13
CI-Link Operating Room Kit (Zti)
The CI-Link Operating Room Kit (Zti) includes the following:
1 CI-Link unit
1 USB cable (type A->B)
1 Instructions for use
1 Programming adaptor
1 Tamper tool
1 Neuro 2 sound processor
1 Antenna
1 Magnet
2 Antenna cables
1 USB memory stick including the Neuro 2
instructions for use in electronic format
Note: The Neuro 2 instructions for use are available on the Oticon Medical website. Contact your
local Oticon Medical customer service or Oticon Medical distributor to ask for a paper version to be sent to your address within 7 days at no additional cost.
Note: In case of replacement, you may receive the CI-Link unit in a different package.
Note: Several antenna cable lengths are available. Check availability on the order form and
contact your local Oticon Medical customer service or Oticon Medical distributor.
Warning: The use of small parts with infants and young children create a risk of suffocation,
swallowing and ingestion hazard. These parts include: magnet, antenna, antenna cable and tam­per tool. Caregivers must ensure these components are not manipulated by the patient.
Page 14
14
When receiving CI-Link
Unpacking and inspection
Unpack CI-Link carefully.
Keep the packaging material in the case it was delivered. If you need to return CI-Link for
servicing, the original packaging material will protect it against damage during transport.
Visually inspect the equipment for damage. If damage has occurred, do not start using the
CI-Link unit. Contact your local Oticon Medical customer service or Oticon Medical distributor for assistance.
Check that the package includes the items listed above in section on Packaging content.
Storage
If CI-Link is stored, follow the guidelines below:
Store the CI-Link unit and accessories in the packaging material provided to protect the
equipment from damage.
Store the CI-Link unit and accessories in a clean environment.
Page 15
15
Preparing CI-Link for use
Installation
Before using CI-Link, make sure the Genie Medical CI fitting software has been properly installed. For further details on software installation, refer to the Genie Medical CI Installation guide.
Connect CI-Link to the PC, and the sound processor to its accessories (antenna, magnet, antenna cable) as described above in the section on CI-Link overview.
Warning: Training should have been carried out prior to using CI-Link.
Page 16
Connecting the programming adaptor to the sound processor
To connect the programming adaptor to the sound processor:
If needed, deactivate the tamper-proof system with the
tamper tool
If needed, slide the battery module off the sound processor
Slide the programming adaptor onto the sound processor
If needed, reactivate the tamper-proof system with the
tamper tool
Note: During the fitting session, place the battery module in a
safe place to avoid losing it.
Warning: To avoid any risk (ingestion, strangulation) during the
fitting session, especially with infants and young children, provision and usage of the program­ming adaptor must be carried out under professional supervision.
Warning: Do not attempt to connect the programming adaptor to the battery charger as this might damage them.
16
Page 17
17
Locking and unlocking the programming adaptor
The tamper tool secures the connection between the sound processor and the programming adaptor to avoid any unexpected disconnection.
To activate the tamper-proof system keeping the antenna
cable connected: Place the open side of the tamper tool on the lock ring (1) situated around the antenna socket and turn clockwise (2).
To deactivate the tamper-proof system keeping the antenna
cable connected: Place the open side of the tamper tool on the lock ring (1) situated around the antenna socket and turn the lock ring counter clockwise (2) using the tamper tool to deactivate.
Important: In the operating room, we strongly recommend
keeping the programming adaptor connected and locked to the sound processor.
Page 18
18
Connecting two sound processors
Two sound processors can be simultaneously connected to the CI-Link unit by means of pro­gramming adaptors.
The programming adaptor is available in two colour codes: red/blue. These colour-coded pro­gramming adaptors match the CI-link unit red/blue colour indications (top view) to make it eas­ier to use.
When two sound processors are to be connected, make sure you connect each sound processor to the correct side.
When a single sound processor is to be connected, you can connect the sound processor to either side.
Right colour code (red) to connect to the sound processor on the right ear
Left colour code (blue) to connect to the sound processor on the left ear
Page 19
19
Indeed, these programming adaptors are identical and can be exchanged if needed. The red/ blue colour code is only intended for usability.
Note: In the operating room, you can use a single programming adaptor for bilateral implanta-
tion. You just need to swap connectors on the CI-Link unit.
Wearing the sound processor on clothes
The clip accessory is designed for use with children who might get distracted by wearing a sound processor on the ear during fitting.
To use the clip:
Push the sound processor into the clip (1).
Attach the clip to the patient’s clothes, near the ear (2).
To remove the clip:
Release the clip from the patient’s clothes.
Twist out the sound processor (3).
Note: Use a long antenna cable when wearing the clip.
Page 20
20
Using CI-Link
The CI-Link unit powers on (“Power” light indicator turns on) as soon as it is connected to the PC, and the PC is switched on.
Once the sound processor is connected to CI-Link, during the connection process with the Genie Medical CI fitting software, the "Left" and "Right" light indicators will successively light up. This means the software is searching for device connection.
Once the sound processor has been detected, the corresponding “Left”/“Right” light indicator continuously repeats the same blinking sequence (a medium light followed by two short lights). The communication between the software and sound processor has been established.
Start the fitting session by launching the Genie Medical CI fitting software from the desktop shortcut or the Windows programs list. Refer to the Genie Medical CI fitting guide for details on programming the sound processors and performing objective measurements.
Warning: Do not attempt to connect or disconnect a sound processor, or cables while the sound
processor is being fitted, as this might damage it.
Page 21
21
Disconnecting CI-Link from the PC
If you wish to power off CI-Link without switching off the PC, you can disconnect the USB cable from the CI-Link unit or from the PC.
Note: CI-Link automatically powers off together with the PC.
Page 22
22
Visual indications
The CI-Link unit is equipped with four green light indicators to easily identify the status of the device.
"Left" and "Right"
Green light repeats the
following blinking sequence: a medium light followed by two short lights
Green light blinks quickly
Connection established
The green light continuously repeats the same blinking sequence when communication is being established between the sound processor (left/ right) and CI-Link. Conversely, when the communication between the sound processor (left/right) and CI-Link is dis­rupted or when ending the fitting session, the light turns off.
Communication in progress
The green light blinks quickly when CI-Link and sound processor communicate (programming, etc.).
"PC Com"
Green light repeats the
following blinking sequence: a medium light followed by two short lights
Connection established
The green light continuously repeats the same blinking sequence when the communication is being established with software, and remains on once communication is established.
Page 23
23
"PC Com"
Green light blinks quickly
Communication in progress
The green light blinks whenever there is actual communication between CI-Link and software (i.e. firmware upgrade, communication with the sound processor).
"Power"
Green light on
Power supply on
The green light is on whenever CI-Link is connected and powered by the PC.
Page 24
24
Care and maintenance
Cleaning
We recommend regular cleaning of the CI-Link unit and accessories using a damp cloth with water or alcohol. In particular in the operating room, it should be cleaned before each use. For all parts in contact with the patient (i.e. programming adaptor), you can use audiological cleaning wipes, which dissolve earwax and clean perspiration.
Warning: Do not perform servicing and maintenance while CI-Link is in use.
Page 25
25
Equipment failure, service and repair
We recommend regular visual inspection of the CI-Link unit and accessories to check its condi­tion. In the case of a faulty accessory, replace the defective element. If you suspect that the functionality or operating safety of the CI-Link unit may be faulty in any way, disconnect CI-Link from the PC, and make sure that it cannot be used by others until it has been serviced. Contact your local Oticon Medical customer service or Oticon Medical distributor.
Warning: Do not use a defective device or accessory.
Do not return a CI-Link until you have gone through the troubleshooting procedure and agreed with your local Oticon Medical customer service or Oticon Medical distributor to return it. To return, place the CI-Link unit and accessories in the original packaging material, with a detailed description of the problem.
Warning: The CI-Link unit and accessories do not contain any elements which may be repaired by
the user or the implant centre. It is strictly forbidden to modify or open the device for any reason. For all repairs, contact your local Oticon Medical customer service or Oticon Medical distributor which is solely responsible for the maintenance of its products. Any attempt to open the unit and accessories will automatically lead to cancellation of the warranty.
Page 26
26
Precautions
The CI-Link unit and accessories must be handled with care.
Important: Observe the following precautions:
Temperature
Do not expose the instrument to extreme temperatures. Do not store or use the instrument outside its normal operating temperature range. Refer to the technical specifications listed below.
Water and liquids
CI-Link must never be washed or immersed in water or other liquids. It is impor­tant to protect it from dust, dirt and humidity.
Page 27
27
Troubleshooting
You can refer to the Visual Indications section to verify correct CI-Link status.
If you experience issues while using CI-Link, you can try to identify and solve the problem by fol­lowing the steps described below. If the problem persists, contact your local Oticon Medical cus­tomer service or Oticon Medical distributor.
Problem Possible cause Solution
The “Left” and/or “Right” indi­cators light up, then turn off after a few seconds when initiat­ing a connection to a sound pro­cessor plugged in the corre­sponding connector
CI-Link is unable to communicate with the sound processor
Ensure that the programming adaptor is prop­erly connected to both the CI-Link unit and the sound processor. If the problem persists, the sound processor or the programming adaptor might be faulty. Try again with another sound processor and programming adaptor before contacting your local Oticon Medical customer service or Oticon Medical distributor
The “PC Com” indicator does not light up when initiating a con­nection to a sound processor
CI-Link communication problem or self-diag­nostic error
Disconnect the CI-Link unit and then reconnect it. Restart the software and try again. If the problem persists, try connecting it to another computer before contacting your local Oticon Medical customer service or Oticon Medical distributor
Page 28
28
The “Left”, “Right” and “PC Com” green lights blink slowly simultaneously
Unrecoverable error in the CI-Link unit
Disconnect and connect the CI-Link unit again. If the problem persists, contact your local Oticon Medical customer service or Oticon Medical distributor
The “Power” indicator on the interface does not light up
USB cable connection problem
Ensure that the USB cable is properly function­ing and connected to a USB port on the com­puter – and not to a USB hub. Check that the interface itself has been properly connected
The “Power” indicator lights up, then goes out after a few seconds
USB cable connection problem
Restart the software installation process. It may be that the problem is due to the equipment. Try another computer before contacting your local Oticon Medical customer service or Oticon Medical distributor
Page 29
29
Disposal
As electronic equipment, the CI-Link unit and accessories cannot be disposed of with household waste, but are subject to sorted waste collection. They must be delivered to an appropriate col­lection point for processing.
This process must comply with the main objective of Directive 2012/19/EU (WEEE) on the preven­tion of waste electrical and electronic devices and in addition, the re-use, recycling and other forms of recovery of such waste.
Oticon Medical is concerned about the preservation of our environment and is a member of the French environmental body “Récylum” for the collection and recycling of our devices. You can send the CI-Link unit and accessories back to your local Oticon Medical customer service or Oti­con Medical distributor for recycling.
Page 30
30
Risks and related instructions for use
Make sure you always use accessories compatible with the CI-Link unit. Check the Oticon
Medical order form for availability and ordering.
Never try to connect the CI-Link unit and accessories to devices other than those specified by
Oticon Medical. This could lead to a reduction in the overall safety level of the system or cause permanent damage to CI-Link.
Accidental damage (such as dropping) and incorrect handling can have a negative effect on
the functionality of the device. Contact your local Oticon Medical customer service or Oticon Medical distributor for advice.
CI-Link contains electronic components. As such, avoid any partial or total immersion in a liq-
uid, as this may seriously compromise personal safety. In general, avoid storing CI-Link in a humid environment, and store it in a dry location with an ambient temperature.
Do not use the instrument in the presence of flammable anaesthetics (gases).
The CI-Link unit should not be used in close proximity to other radiating systems, such as
X-ray machines, MRI scanners, echography devices, equipment emitting vibrations, ultra­sounds or infrasounds, etc. The CI-Link unit and accessories should not be used adjacent to or stacked with other equipment (except computer, sound processors).
Page 31
31
CI-Link is a non-sterile device. When using the device in an operating room, the sound proces-
sor and the antenna should not be in contact with the sterile field, and should be placed in a sterile covering.
Warning: If you are in any doubt about the functioning of CI-Link, contact your local Oticon
Medical customer service or Oticon Medical distributor for your safety to ensure you are not exposed to any risks.
Page 32
32
Physical and performance characteristics
Essential performance Neuro Cochlear Implant System: Accuracy of the electrical stimulation
(<10% at C-level)
PC interface The serial USB (Universal Serial Bus) port is used for communication
between a PC and the CI-Link unit Communication: USB 2.0 USB Connector Type: "Type B" connector (on the CI-Link unit)
Power supply The CI-Link unit is powered from the PC USB port.
Input Voltage: 5VDC Maximum CI-Link consumption: 2.5 W (with load)
Power output Output Voltage: 1.2VDC up to 4VDC, or 5VDC
Maximum output power (transmitted power): – 400mW per channel (if output voltage is between 1.2VDC and 4VDC) – 900mW per channel (if output voltage is 5VDC)
CI-Link unit Dimensions (L x W x H): 150 x 103.5 x 30mm
Weight: 190g
Cables Programming adaptors
Maximum length of cable: 2.95m
USB cable (type A to type B)
Maximum length of cable: 1.80m
Page 33
33
Output ratings Output for sound processors
These output ratings are valid for left and right Programmable battery supply, common for left and right side: 5.0VDC, or between 1.2VDC and 4.0VDC.
Output for third-party equipment
BNC output provides a TTL signal (5 Volts - 5mA max.)
Operating conditions Operating temperature: +5°C to +45°C
Relative humidity: 10% to 90% Atmospheric pressure: 700 hPa to 1060 hPa
Transport and storage conditions
Temperature: -20°C to 50°C Relative humidity: 0% to 90% Atmospheric pressure: 700 hPa to 1060 hPa
Classification Protection against electric shocks:
Class II ME equipment Type B applied parts
Accessories intended to be used with the CI-Link programming system:
USB cable (181772)
Programming adaptors (159689, 173414)
Tamper tool (185326)
Clip (176669)
Page 34
34
Symbols
0459
CE registration mark with notified body number. Indicates compliance with the requirements of the Active Implantable Medical Device Directive 90/385/EEC.
Warning.
Consult the operating instructions.
www.oticonmedical.com
Consult the electronic operating instructions (eIFU) available at www.oticonmedical.com
Serial number.
Reference.
Batch code.
Date of manufacture.
Manufacturer.
Affected by dampness – store in a dry place.
Type B device. Device for which the parts applied and connected to the user are not conductive.
Refer to instruction manual.
Page 35
35
-20°C
+50°C
Temperature limits to which the medical device can be exposed.
Humidity limits to which the medical device can be exposed.
Atmospheric pressure limits to which the medical device can be exposed.
The electrical device must be disposed of in a separated waste collection. Directive for recycling electrical and electronic equipment (waste electrical and electronic equipment (WEEE)).
5V – 500mA
Input voltage
Class II electrical device with double insulation.
10%
90%
700 hPa
1060 hPa
Page 36
36
Profile of the person handling CI-Link
The professional is the operator when he:
Connects CI-Link to the PC
Connects CI-Link to a programmable sound processor
Performs fittings and objective measurements with the sound processor, through CI-Link and
Genie Medical CI fitting software
Connects and disconnects the accessories
Page 37
37
Manufacturer's declaration
Medical electrical equipment requires special precautions to be taken for EMC (Electromagnetic Compatibility) and it is necessary to install and start it according to the EMC information pro­vided in the tables below. Portable RF (radio frequency) communication devices, such as mobile telephones, may affect the operation of CI-Link. Therefore, these devices should be kept as far away as possible from the CI-Link unit and accessories in order to avoid such effects.
Guidelines and manufacturer's declaration – Electromagnetic emissions
CI-Link is intended for use in the electromagnetic environment specified below. The CI-Link user should ensure that it is used in such an environment.
Emissions tests Compliance
CISPR 11 RF emissions Group 1
CISPR 11 RF emissions Class A
Harmonic emissions IEC 61000-3-2
Not applicable
Voltage fluctuations and flicker emissions IEC 61000-3-3
Not applicable
Page 38
38
Guidelines and manufacturer's declaration – Electromagnetic immunity
CI-Link is intended for use in the electromagnetic environment specified below. The CI-Link user should ensure that it is used in such an environment.
Immunity test Test level IEC 60601 Level of compliance
Electrostatic discharge (ESD) IEC 61000-4-2
±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
± 8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF IEC 61000-4-3
±10 V/m from 80 MHz to 2.7GHz 80% AM at 1kHz
±10 V/m from 80 MHz to 2.7GHz 80% AM at 1kHz
Power frequency (50/60 Hertz) magnetic field IEC 61000-4-8
30 A/m (50/60 hertz)
30 A/m (50/60 hertz)
Warning: CI-Link is intended for use by healthcare professionals only. It may cause radio
interference or disrupt the operation of nearby equipment. It may be necessary to take mitigating actions, such as re-orienting or relocating CI-Link or shielding the location.
Note: UT is the A.C. mains voltage prior to application of the test level.
Page 39
39
Test
frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
power (W)
Distance (m)
IMMUNITY
TEST LEVEL
(V/m)
385 380 – 390 TETRA 400
Pulse
modulation
18 Hz
1,8 0,3 27
450 430 – 470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz
deviation
2 0,3 28
710
704 – 787 LTE Band 13, 17
Pulse
modulation
217 Hz
0,2 0,3 9745
780
810
800 – 960
GSM 800/900,
TETRA 800,
IDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2 0,3 28
870
930
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Page 40
40
Test
frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
power (W)
Distance (m)
IMMUNITY
TEST LEVEL
(V/m)
1720
1700 – 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
217 Hz
2 0,3 28
1845
1970
2450 2400 – 2570
Bluetooth,
WLAN, 802,11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation
217 Hz
2 0,3 28
5240
5100 – 5800
WLAN 802,
11 a/n
Pulse
modulation
217 Hz
0,2 0,3 95500
5785
Page 41
41
Date of last review: 2018-05
Page 42
42
Notes
Page 43
43
Notes
Page 44
www.oticonmedical.com
196840UK - version A
Oticon Medical
NEURELEC
2720 Chemin Saint-Bernard
06220 Vallauris
France
TEL: +33 (0)4 93 95 18 18 FAX: +33 (0)4 93 95 38 01 info@oticonmedical.com
Loading...