OSRAM QUICKTRONIC-M ECG for circular FL User Manual

ECG for T5 fluorescent lamps
Technical Guideline
May 2005
Electronic Control Gear to operate T5/Ø 16 mm fluorescent lamps
Contents
1.
Introduction.......................................................................................... 6
1.1 History.....................................................................................6
1.2 T5/ 16 mm-Fluorescent Lamps............................................7
1.2.1 High Efficiency FH®…HE........................................................8
1.2.2 High Output FQ®…HO............................................................8
1.2.3 Fluorescent Circular FC®........................................................8
1.2.4 Comparison of Lumens between T8/ 26 mm- and
T5/ 16 mm-Fluorescent Lamps............................................8
1.3 Electronic Control Gear ..........................................................8
1.4 Different Principles, Different Behavior................................... 9
1.5 Advantages of Electronic Control Gear ..................................9
1.6 Saving Energy with Electronic Control Gear...........................9
1.7 Ignition of Fluorescent Lamps...............................................10
1.8 Costs and Safety ..................................................................10
1.9 Flexibility upon Request........................................................10
1.10 ECG bring Progress..............................................................10
1.11 The right ECG for every Application .....................................10
1.12 OSRAM ECG Milestones......................................................11
2. Product Features ...............................................................................12
2.1 Lighting Comfort ...................................................................12
2.2 Economy...............................................................................12
2.3 Safety....................................................................................12
2.4 Supply Voltage......................................................................13
2.4.1 Overvoltage and its Reason .................................................13
2.4.2 Possible Implications due to Overvoltage.............................14
2.4.3 Undervoltage and its Reason ...............................................14
2.4.4 Possible Implications due to Undervoltage........................... 14
2.4.5 Supply Voltage QT with 21 mm height ................................14
2.4.6 Supply Voltage for QT with 30 mm height ............................ 15
2.4.7 ECG for 120V / 277V Line Voltage....................................... 15
2.5 Automatic Lamp Restart .......................................................15
2.5.1 Lamp ignition for QTi ............................................................16
2.5.2 Lamp ignition for QT to operate T5-fluorescent lamps .........16
2.5.3 Off- Time for Optimum Preheat Start....................................16
2.6 Behaviour in Response to Lamp Defects .............................16
2.6.1 One-Lamp Operation with Multi-Lamp ECG ......................... 16
2.7 Noise.....................................................................................17
2.8 Power Factor λ .....................................................................18
2.9 ECG Imprint.......................................................................... 19
2.10 Reliability ..............................................................................19
2.11 Resistance to Frequent Switching ........................................19
2.12 ECG Lifetime ........................................................................19
2.13 Thermal influences of the system components ....................20
2.14 cut-off Technology ................................................................20
2.14.1 Advantages for Users ...........................................................21
2.14.2 Physical Background ............................................................21
1
End of Life (EoL acc. to T.2)................................................22
2.15
2.16 U-OUT ..................................................................................22
2.17 Approval Marks.....................................................................23
2.17.1 ENEC-Approval Mark ...........................................................23
2.17.1.1 Safety acc. to EN 61347 ............................................23
2.17.1.2 Performance acc. to EN 60929..................................23
2.17.2 VDE EMC mark ....................................................................23
2.18 Energie Efficiency Index EEI ................................................24
2.19 CE Labelling .........................................................................24
2.20 CCC Approval.......................................................................25
3. ECG installed in Luminaire: Installations and Operation Instructions 26
3.1 Wiring Instructions ................................................................26
3.1.1 Cable Types..........................................................................26
3.1.2 Cable Cross-Sections...........................................................26
3.1.3 Release of Contacts .............................................................27
3.1.4 Insulation ..............................................................................29
3.1.5 Terminals..............................................................................29
3.1.6 Cable routing ........................................................................29
3.2 Electromagnetic Compatibility ..............................................30
3.2.1 Harmonic Content acc. to EN 61000-3-2..............................30
3.2.2 Radio interference suppression............................................31
3.3 Permissible Cable Lengths................................................... 36
3.4 „Hot Wires“............................................................................36
3.5 Switching between Lamp and ECG...................................... 37
3.6 Master-Slave Circuit .............................................................37
3.6.1 Max. length of the connecting cable between 2 luminaries ..38
3.7 PE-Connection for Protection Class I Luminaires.................38
3.8 Functional Earth for Luminaires of Protection Class II.........39
3.8.1 General Information.............................................................. 40
3.8.2 Practical Details....................................................................40
3.9 Temperature Ranges............................................................41
3.9.1 Self heating ECG..................................................................41
3.9.2 Control Gaer Temperatures..................................................42
3.1.2.1 ECG in 30 mm height.................................................27
3.1.2.2 ECG in 21 mm height.................................................27
3.1.3.1 WAGO 250................................................................. 27
3.1.3.2 WAGO 251 – IDC.......................................................27
3.1.3.3 WAGO 251 – horizontal plug .....................................28
3.1.3.4 WAGO 251 mini – IDC...............................................28
3.1.3.5 WAGO 251 mini – horizontal plug.............................. 28
3.2.2.1 Causes of Radio Interference ....................................31
3.2.2.2 Conducted Interferences acc. to EN 55015 ...............31
3.2.2.3 Disturbances due to Fields.........................................32
3.2.2.4 Selective Shielding..................................................... 32
3.2.2.5 Installation Instructions for avoiding Disturbance....... 33
3.2.2.6 Asymmetric installation of ECG..................................34
3.2.2.7 Good wiring arrangement for 2-lamp luminaires........ 35
3.2.2.8 Luminaires with reflector and/or specular louvres...... 35
2
3.9.2.1
3.9.2.2 Ambient Temperature ECG : ta..................................43
3.9.3 Lamp Temperature ...............................................................43
3.9.3.1 Maximum Luminous Flux for T5/ 16 mm-Fluorescent
3.9.4 General Recommendations for Installation...........................44
3.9.5 Measuring the Temperature .................................................44
3.10 Luminaire Wiring Test for Two-lamp Luminaires ..................45
3.10.1 Testing with a Test Adapter and Dummy Lamps.................. 45
3.11 ECG Operation for Luminaires of Protection Classes I and II48
3.12 Insulation Distances in Luminaires .......................................48
3.13 Insulation Test ......................................................................48
3.13.1 Dielectric Resistance in Lighting Systems............................49
3.13.2 Mesuring the Dielectric Resistance between N and PE or L
3.13.3 Three-Phase Operation ........................................................50
3.13.4 Resistance to Overvoltage for QUICKTRONIC for
3.14 Inrush Current / Automatic Circuit Breakers .........................50
3.15 RCDs / Fault Currents ..........................................................51
3.16 Leakage Current...................................................................51
3.17 ECGs in Three-Phase Operation..........................................51
Measuring Point Temperature tc................................42
Lamps ........................................................................43
and PE..................................................................................49
T5/16mm- Fluorescent Lamps...........................................50
4. Lamp Wiring.......................................................................................53
4.1 h = 21 mm............................................................................. 53
4.1.1 QUICKTRONIC® INTELLIGENT 1-lamp version ..................53
4.1.2 QUICKTRONIC ® INTELLIGENT 2-lamp version .................53
4.1.3 QT-FH MULTIWATT F/CW...................................................53
4.1.4 QT-FQ F/CW 1-lamp version................................................53
4.1.5 QT-FQ F/CW 2-lamp version................................................54
4.2 h= 30 mm.............................................................................. 54
4.2.1 QT-FH MULTIWATT 1- and 2-lamp version .........................54
4.2.2 QT-FH 3- and 4-lamp version............................................... 54
4.2.3 QT-FQ 1-lamp version.......................................................... 55
4.2.4 QT-FQ 2-lamp version.......................................................... 55
5. QUICKTRONIC INTELLIGENT...........................................................56
5.1 Definition INTELLIGENT.......................................................56
5.2 Lamp Detection as Fundamental Advantage........................56
5.3 QTi – the High-tech ECG......................................................56
5.4 QTi – Advantages................................................................. 56
5.5 QTi – Practically Applied.......................................................57
5.6 Technical Specialties for non-dimmable QTi ........................57
5.6.1 Inrush current limitation ........................................................57
5.6.2 Resistance to Overvoltage up to 400V .................................58
5.6.3 Lamp-ECG-Combination ......................................................58
5.6.3.1 Straight Fluorescent types .........................................58
5.6.3.2 Compact and Circular lamp types.............................. 59
5.6.4 Wiring ....................................................................................59
3
Dimensions...........................................................................60
5.6.5
5.7 FAQ ......................................................................................60
6. Special Applications .........................................................................61
6.1 Outdoor Application ..............................................................61
6.1.1 Installation Instructions .........................................................61
6.1.2 OUTKIT.................................................................................62
6.2 T5-ECG in Sound Studios ....................................................62
6.2.1 Noise and how to avoid it......................................................62
6.2.2 Recommended minimum distance between lamp and
refelctor................................................................................. 63
6.3 Treatment Rooms, Operating Rooms...................................64
6.3.1 Electromagnetic Interference................................................ 64
6.3.2 Interference from Infrared Transmission Equipment ............65
6.4 Electronic Tagging................................................................65
6.5 Emergency Lighting..............................................................65
6.5.1 Different criteria for lighting................................................... 67
6.5.1.1 Switch-over time for QTi – h=21 mm .........................67
6.5.1.2 Switch-over time for QT-FH…CW – h=30 mm........... 67
6.5.1.3 Switch-over time for QT-FQ…CW – h=30 mm ..........67
6.5.1.4 Switch-over time for QT-…F/CW – h=21 mm ............67
6.5.2 Wiring diagrams for emergency lighting units.......................68
6.5.2.1 Wiring diagram QT-FH 3x14 CW with emergency
lighting component from BAG ....................................68
6.5.2.2 Wiring diagram QT-FH 4x14 CW with emergency
lighting component from BAG ....................................68
6.5.2.3 Wiring diagram QT-FH 3x14 CW with emergency
lighting component from OMNITRONIX..................... 69
6.5.2.4 Wiring diagram QT-FH 4x14 CW with emergency
lighting component from OMNITRONIX..................... 69
6.6 DC supply .............................................................................69
6.7 Portable Luminaires..............................................................70
6.8 Mix-up of FH®- and FQ®-Fluorescent Lamps........................70
7. Appendix ............................................................................................ 72
7.1 Overview of Maximum Cable Lengths..................................72
7.1.1 QUICKTRONIC® INTELLIGENT...........................................72
7.1.2 QT-FH MULTI...CW.............................................................. 72
7.1.3 QT-FQ...CW -30 mm height- ................................................72
7.1.4 QT-FH MULTI…F/CW -21 mm height- .................................72
7.1.5 QT-FQ…F/CW -21 mm height-.............................................73
7.1.6 QT-FC...................................................................................73
7.2 Terminal Types.....................................................................73
7.3 Inrush Currents.....................................................................73
7.4 Lamp/ECG Combinations..................................................... 74
7.5 OSRAM Installation Tips for T5-Systems ............................. 74
7.5.1 Recommended Minimum Distance between Lamp and
Reflector ...............................................................................75
7.5.2 Recommended Minimum Distance between two T5/16mm-
Fluorescent Lamps ...............................................................75
4
Luminaire Optimisation.........................................................76
7.5.3
7.5.4 Maximum luminous flux for FH…HE fluorescent lamps .......76
7.5.5 Verticalness Operation .........................................................76
8. Troubleshooting Tips........................................................................77
8.1 General.................................................................................77
8.2 Equipment Behaviour on Overvoltage.................................. 77
8.3 Equipment Behaviour on Under Voltage ..............................78
8.4 Application faults...................................................................78
8.4.1 Wiring faults on the lamp side...............................................78
8.4.2 Short-to-ground at the output of QUICKTRONIC® ECG.......78
8.4.3 Effects of moisture................................................................78
8.4.4 Installing luminaires in draughty locations ............................78
8.5 Trouble Shooting ..................................................................79
8.5.1 Lamp does not start..............................................................79
8.5.2 Brief Glimmer........................................................................80
8.5.3 Lamp goes out during operation...........................................81
8.5.4 Different brightness levels ....................................................82
8.5.5 Fault in other electrical equipment........................................ 83
8.5.6 Problems at master-slave operation..................................... 83
8.5.7 Humingh or “chirping” from the ECG ....................................83
9. Lamp-ECG Combinations .................................................................84
9.1 FQ®...HO-Fluorescent Lamps...............................................84
9.2 FH®...HE-Fluorescent Lamps ...............................................85
9.3 FC®…Fluorescent Lamps.....................................................85
10. Tender Documents ............................................................................87
10.1 QUICKTRONIC® INTELLIGENT QTi....................................87
10.2 QUICKTRONIC® MULTIWATT for FH…HE h = 30 mm ...... 87
10.3 QUICKTRONIC® for FQ…HO h = 30 mm ............................88
11. Index ................................................................................................... 89
5
1. Introduction
1.1 History
The development of linear fluorescent goes back to the thirties of the 20th century. The diameter of 51mm was very voluminous. However, better efficiency did not come up before the fifties.
1879
Kohlefaden-Glühlampe von Thomas A. Edison Incandescent lamps with carbon filament by Thomas A. Edison
1910
Glühlampen mit Wolfram-Wendel Incandescent lamps with tungsten coils
1925
BILUX®Zweidraht-Scheinwerferlampen
®
two-wire headlight lamps
BILUX
1931
Natriumdampf-Niederdrucklampen Low-pressure sodium vapor lamps
1933
Quecksilberdampf-Hochdrucklampen High-pressure m ercury vapor lamps
1936
Leuchtstofflampen Fluorescent lamps
1954
XBO Xenon-Hochdruck­lampen XBO high-pressure xenon lamps
1968
VIALOX® NAV Standard Natriumdampf-Hochdrucklampen
®
NAV Standard
VIALOX high-pressure sodium vapor lamps
1968
POWERSTAR HQI Halogen-Metalldampflampen POWERSTAR HQI metal halide lamps
1970
HMI METALLOGEN®Lampen HMI METALLOGEN
1971
BILUX®H4 Halogen-Zweidraht-Scheinwerferlampen BILUX two-wire halogen headlight lamps for auromobiles
1973
HALOSTAR Niedervolt-Halogenglühlampen HALOSTAR low-voltage tungsten-halogen lamps
1979
LUMILUX Leuchtstofflampen LUMILUX fluorescent lamps
1980
EVG QUICKTRONIC®DE LUXE / ECG QUICKTRONIC
1982
OSRAM DULUX®L Kompakt-Leuchtstofflampen OSRAM DULUX compact fluorescent lamps
1984
DECOSTAR Niedervolt-Halogenglühlampen mit Kaltlichtreflektor DECOSTAR low-voltage tungsten-halogen lamps with dichroic reflectors
®
lamps
®
H4
®
®
®
DE LUXE
®
L
PL (W)UN (V)fN (Hz)IN (A)lta ( C) 22050600,090,95 C 1xL18 W1x16-2050 2400,0850,93 C
Anwendungsbereich: AC/DC 198 V bis 254 V Geeignet fr Batteriespannungen 154 V bis 276 V Zur Verwendung in Anlagen nach VDE 0108 geeignet Range of application: AC/DC 198 V to 254 V
Temp.-Test
Range of battery voltage: 154 V to 276 V
c = 70 C max.
t
Suitable for emergency installations acc. to VDE 0108
class B 0712T201 OW2 Made in Germany
1985
OSRAM DULUX®EL Kompaktleuchtstoff­lampen mit elektronischem Vorschaltgerät
®
OSRAM DULUX with electronic control gear
1987
POWERSTAR HQI-T Halogen-Metalldampflampen POWERSTAR HQI-T compact metal halide lamps
1991
D1 Gasentladungslampen D1 gas discharge lamps
1993
COLORSTAR DSX-T 80W Natrium-Xenonlampen COLORSTAR DSX-T 80W sodium xenon lamps
1993
FM Mini-Leuchtstofflampen / FM mini fluorescent lamps
1995
FH Hocheffiziente Leuchtstofflampen / FH high­efficient fluorescent lamps
1 2
L
3 4
OSRAM
1996
FQ Lichtstarke Leuchtstofflampe / FQ high power fluorescent lamps
1997
OSRAM ENDURA Die elektrodenlose Hochleistungsleuchtstofflampe OSRAM ENDURA The high-performance electrodeless fluorescent lamp
EL energy-saving lamps
The improvements according to the luminous flux and lifetime with the T12 lamp (38 mm diameter) made an economic and even outdoor application possible. There were continuous improvements for optimizations as for example Amalgamtechnology. In this term fluorescent lamps were operated by conventional control gears (CCG).
The decisive breakthrough was at the end of the 70s and early 80s. In the year 1978, a new T8/ 26 mm- lamp generation started to replace T12/38mm- fluorescent lamps. New phosphors with higher resistance came into the market. Nowadays known under triband-phosphor (LUMILUX light colours). The diameter of the lamps was reduced from 38mm to 26mm however the length was kept with 59, 120 and 150 cm as the sockets G13. The new wattage of 18, 36 and 58 W was advantageous as it was a reduction of at least 10% compared to the T12/38 mm lamps in 20, 40 and 65 W. This was also the hour of birth for the Electronic Control Gears (ECG). First the circuits were in a instant start mode what also was called softstart up to the programmed or preheat start. Together with modern Electronic Control Gear QUICKTRONIC T8/26 mm- fluorescent lamps became even more efficient and longer lasting. In addition, the thermal behavior of the luminous flux was improved. The T8-system was permanently improved as the example of the triband­phosphor shows with a very high service life of the fluorescent lamp.
In 1995, the next milestone of fluorecent lamp development saw the introduction of new FH...HE (H
igh Efficiency) systems to the market. With
the reduced diameter of 16 mm only the lamp was designed. It is available in 14, 21, 28 and 35 W with the G5 sockets. It is 50 mm shorter as the T8 fluorescent lamps.
T5 fluorescent lamps only can be operated by Electronic Control Gear. So the light output and the life time of the lamp were designed from the beginning to an optimum of up to 104 lm/W. The maximum luminous flux of the T5 lamps is at 35 °C compared to 25 °C at T8 or T12 lamps. The reduced lamp diameter of 16 mm as the maximum lumen output at 35 °C are the relevant feature for a higher efficiency of the fixture.
6
In 1996, the T5 lamp family was completed with the types of higher lumens than volume FQ
®
…HO (High Output)-fluorescent lamps. They are available
in the wattage of 24, 39, 49, 54 and 80 W with the identical lengths as the FH…HE types. With up to 7000 lm for FQ 80 W HO this is the lamp family with the highest light output.
In 1999, the third member of the T5/ 16 mm-lamp family was introduced to the market. Away from usual light strips, compact, efficient and unconventional luminaires benefit from this new, circular FC
®
-lamp: 50 %
more light output than with comparable standard circular lamps.
Special Note:
Independent of the lamp diameter of the fluorescent lamp the luminous flux is specified for an ambient temperature of 25 °C. These values are exclusive to be used for light plannings. The value of the luminous flux is for the T5/ 16 mm fluorescent lamps FH
®
...HE and FQ®...HO for 25 °C below the values for 35 °C. The values at 35 °C ambient temperature are only for information. The Circular lamps FC
®
only have value of luminous
flux at 25 °C. The advantages of the T5/ 16 mm fluorescent lamps show their advantages in the improved efficiency of the fixture.
Detailed technical information about QUICKTRONIC
®
are shown in the latest indoor outdoor lighting and can be downloaded under
www.osra.de/ecg
.
QUCKTRONIC
®
for the operation of the T5/ 16 mm fluorescent lamps
have all features of a high quality ECG.
Good radio interference suppression
Good radio interference suppression
Good radio interference suppression
EN 55015
EN 55015
EN 55015 CISPR 15
CISPR 15
CISPR 15
max.Interference voltage[dB µV]
max.Interference voltage[dB µV]
max.Interference voltage[dB µV]
110
110
110
90
90
90
80
80
80 70
70
70
66
66
66
60
60
60
56
56
56
-2
-2
-2
10
10
10
110010-1
110010-1
110010-1
10210
10210
10210
Frequency [MHz]
Frequency [MHz]
Frequency [MHz]
Mains harmonics
Mains harmonics
IEC 61000-3-2
IEC 61000-3-2 EN 61000-3-2
EN 61000-3-2
Mains harmonics [%]
Mains harmonics [%]
30
30
30 25
25
25
Limit values
Limit values
Limit values
20
20
20
to IEC
to IEC
to IEC
15
15
15
10
10
10
5
5
5
0
0
0
3. 5. 7. 9. 11. 13.
3. 5. 7. 9. 11. 13.
3. 5. 7. 9. 11. 13.
Harmonics
Harmonics
Reliability of the ECG
Reliability of the ECG
IEC 61347-2-3
IEC 61347-2-3 EN 61347-2-3
EN 61347-2-3
z All insula ted
z All insula ted z Compliance with creepage and
z Compliance with creepage and
clearance distances
clearance distances
z ECG shutdown in case of failure
z ECG shutdown in case of failure
Lamp operation to standards
Lamp operation to standards
IEC 60929
IEC 60929 EN 60929
EN 60929
rel. system luminous flux [%]
rel. system luminous flux [%]
100
100
90
90
80
80
70
70
60
60
50
50
2
2
ECG
ECG
10. 18.
10. 18.
CCG
CCG
4 6 8 1012141618 222420
4 6 8 1012141618 222420
Time of usage in hours [tsd]
Time of usage in hours [tsd]
1.2 T5/ 16 mm-
Fluorescent Lamps
Immunity
Immunity
IEC 61547
IEC 61547 EN 61547
EN 61547
230 V
230 V 50 Hz
50 Hz
Power regulation
Power regulation
At higher ambient temperatures
At higher ambient temperatures
P Gas [W]
P Gas [W]
50
50
40
40
30
30
20
20
0 1020304050607080
0 1020304050607080
Tube wall temperature [°C]
Tube wall temperature [°C]
Minimum ECG life
Minimum ECG life
z Long-life electrolytic capacitor
z Long-life electrolytic capacitor
(50,000h at t
(50,000h at t
z Optimised circuit
z Optimised circuit z Low self heating
z Low self heating
Functional ECG [%]
Functional ECG [%]
100
100
100
80
80
80
60
60
60
40
40
40
20
20
20
90°C
90°C
90°C
0
0
0
0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140
)
)
c max
c max
60°C
60°C
60°C
70°C
70°C
70°C
Hours of operation [tsd]
Hours of operation [tsd]
Measuring
Measuring point t
point t
50°C
50°C
50°C
c
c
Reliable ignition
Reliable ignition
at low temperatures
at low temperatures
-30 -25 -20 -15 -10 -5
-30 -25 -20 -15 -10 -5
Lamps starting [%]
Lamps starting [%]
EVG
EVG
KVG
KVG
Ambient temperature [°C]
Ambient temperature [°C]
100
100
80
80
60
60
40
40
20
20
0
0
0
0
The diameter and also the description of the new fluorescent lamp family is based on American measures: (1 inch = 25.4 mm) The value is combined with a T (tube). 5/8 of an inch = 16 mm Î T5-fluorescent lamp
Classification:
T2 tube diameter of 7 mm T5 tube diameter of 16 mm T8 tube diameter of 26 mm T12 tube diameter of 38 mm T17 tube diameter of 51 mm (1936)
7
1.2.1 High Efficiency
®
FH
…HE
1.2.2 High Output FQ
®
…HO
1.2.3 Fluorescent Circular
®
FC
1.2.4 Comparison of Lumens between T8/ 26 mm- and T5/ 16 mm­Fluorescent Lamps
Consecutively the important data of the FH…HE, FQ…HO and FC fluorescent lamps are shown.
type length [mm] lumens at
ta=25 °C
FH® 14W HE FH® 21W HE FH® 28W HE FH® 35W HE
549 1200 1350
849 1900 2100 1149 2600 2900 1449 3300 3650
lumens at
ta=35 °C
Values for light colors 827, 830, 840
type length [mm] lumens at
ta=25 °C
FQ® 24W HO FQ® 39W HO FQ® 54W HO FQ® 80W HO FQ® 49W HO
549 1750 2000
849 3100 3500 1149 4300 4900 1449 4450 5000 1449 6150 7000
lumens at
ta=35 °C
Values for light colors 827, 830, 840
type
[mm]
lumens at
ta=25 °C
FC® 22W FC® 40W FC® 55W
225 1800
300 3200
300 4200
Values for light colors 827, 830, 840
Detailed technical data of T5/ 16 mm-fluorescent lamps can be found in the OSRAM product catalogue and under www.OSRAM.com.
Fluorescent lamp innovation: T8 Î T5
T8 (Ø 26 mm)
T8 (Ø 26 mm)
Luminous flux at 25°C
Luminous flux at 25°C
600 mm 900 mm 1.200 mm 1.500 mm
600 mm 900 mm 1.200 mm 1.500 mm
18 W 30 W 36 W 58 W
18 W 30 W 36 W 58 W
1.350 lm 2.400 lm 3.350 lm 5.000 lm
1.350 lm 2.400 lm 3.350 lm 5.000 lm
1.3 Electronic Control Gear
T5 (Ø 16 mm)
T5 (Ø 16 mm)
Luminous flux at 25°C
Luminous flux at 25°C
550 mm 850 mm 1.150 mm 1.450 mm
550 mm 850 mm 1.150 mm 1.450 mm
FH 14 W 21 W 28 W 35 W
FH 14 W 21 W 28 W 35 W
1.200 lm 1.900 lm 2600 lm 3.300 lm
1.200 lm 1.900 lm 2600 lm 3.300 lm
FQ 24 W 39 W 54 W 49 W / 80 W
FQ 24 W 39 W 54 W 49 W / 80 W
1.750 lm 3.100 lm 4.450 lm 4.300 lm / 6.150 lm
1.750 lm 3.100 lm 4.450 lm 4.300 lm / 6.150 lm
Since the early seventies Electronic Control Gear have been used in computers and consumer electronics. As this technology offers substantial advantages, it was only natural to use it also for lighting purposes. Linear and compact fluorescent lamps must be operated with ballasts, as the process of gas discharge requires well defined currents and voltages. The
8
1.4 Different Principles, Different Behavior
1.5 Advantages of Electronic Control Gear
ballast is responsible for preheating the lamp electrodes, for sufficient ignition voltage and for limiting the lamp current.
The basic functions that are mentioned in chapter 1.3, can usually be carried out with electromagnetic (inductive) ballasts. These ballasts are classified into conventional control gear (CCG) and low loss ballasts (LLG). The latter follow the same principle as CCG, however, due to different engineering design they consume less energy. The much better solution is to operate fluorescent lamps with Electronic Control Gear (ECG). Besides the advantages of flicker-free lighting, longer lamp life and higher system efficacy (lamp + ECG), features such as lamp ignition, limitation of the lamp current and compensation are integrated into the ECG. Most Electronic Control Gear are also suitable for DC operation, which means they can be used in emergency lighting installations.
T5/ 16 mm fluorescent lamps FH
®
…HE, FQ®…HO and FC® can only be
operated by Electronic Control Gear.
If fluorescent lamps are operated with magnetic ballasts (principle of magnetic coil, CCG and also low loss ballast), the lamp current equals the frequency of the mains voltage. The resulting stroboscopic effect can be dangerous in cases where people work with rotary machines. Every time, the voltage goes through zero, the lamp current does the same until the lamp is reignited: for every lamp ignition new carriers for the electric charge have to be build up within the gas discharge.
CALM AND
LOWER
LOWER ENERGY
ENERGY CONSUMPTION
CONSUMPTION (25 – 30%)
(25 – 30%)
CALM AND
FLICKERFREE
FLICKERFREE
LIGHT
LIGHT
OPERATION WITHOUT
OPERATION WITHOUT
NOISE
NOISE
LOW MAGNETIC
LOW MAGNETIC
STRAYFIELD
STRAYFIELD
FLICKERFREE
FLICKERFREE
START
START
1.6 Saving Energy with Electronic Control Gear
LONGER
LONGER LAMP LIFE
LAMP LIFE (approx. 50%)
(approx. 50%)
LESS WASTE
LESS WASTE DISPOSAL
DISPOSAL (approx. 30%)
(approx. 30%)
ENERGY SAVING
ENERGY SAVING
(25 – 30%)
(25 – 30%)
LONGER LAMP LIFE
LONGER LAMP LIFE
(approx. + 50%)
(approx. + 50%)
SWITCH OFF AT END
SWITCH OFF AT END
LOW
LOW
WIRING COSTS
WIRING COSTS
DIMMABLE
DIMMABLE
(spec. version)
(spec. version)
AUTOMATIC
AUTOMATIC
OF LAMP LIFE
OF LAMP LIFE
When using Electronic Control Gear the frequency of the lamp voltage is approx. 1000 times higher than the line voltage. The zero of the lamp current are passed so quickly that the average of the value of the electron
density is nearly constant within the discharge plasma. The electrons don’t
have to be built up with every cycle (as it is necessary when using CCG and low loss ballasts). So the limitation of the lamp life due to reignition peaks for CCG operation are avoided with ECG operation. Therefore no stroboscopic effects can occur by using high frequency control gear as there is no longer a gap in the lamp current. Therefore, one lamp type needs less energy to generate the same lumens when operated with high frequency control gear compared to operation with magnetic ballasts. The lower energy consumption reduces the lamp load and increases the lamp
9
1.7 Ignition of Fluorescent Lamps
1.8 Costs and Safety
1.9 Flexibility upon Request
1.10 ECG bring Progress
1.11 The right ECG for every Application
life. Electronic Control Gear improve the efficiency and the lamp life of fluorescent lamps significantly.
Prior to ignition, modern ECG heat the cathode to its optimum temperature for electron emission. After a defined period the lamp is ignited with the required ignition voltage. Only an optimized preheat start can guarantee that the number of switching cycles has only little effect on the lamp life. This is another important feature of ECG which has a positive effect on the cost of operation and which should not be neglected when looking for alternatives to CCG.
At the end of lamp life the emitter paste applied to the lamp electrode is used up. The complete loss of emitter results in an increase of voltage in the vicinity of the electrode. This situation can last over a longer period of time. As an immediate result of the accompanying temperature increase at the lamp ends the lamp sockets may overheat. Modern ECG are able to detect this malfunction and switch the lamps off. Unnecessary attempts to ignite are avoided by an interrupting function and therefore, also overheating is avoided - an important contribution to more safety. Professional ECG control all parameters constantly. A safety shut down at the end of the lamp life is mandatory from January 1
st
, 2007 on for all ECG that operate T4 or T5 tubes as it is included in the IEC 61347 (Omnibusnorm for safety of Electronic Control Gear). For several years now, all OSRAM QUICKTRONIC
®
fulfill the safety requirements acc. the IEC 61347 already. However as there was no Standard for this before, some ECG manufacturer neglect this due to costs.
During past years, we see a clear increase in new, better and more energy efficient lamp systems. Unfortunately, this resulted also in a growing number of various ECG-types. To reduce this large number of types manufacturers of ECG have taken a new direction and have developed new multi-purpose ECG which can be used for a variety of fluorescent lamps of different wattages. New integrated circuits allow the optimum control of lamp features such as lumen output. This type reduction has, of course, a big effect on the customer: ordering, warehousing and installation of only a few ECG-types. The so-called MULTIWATT-ECG reduce all relevant cost drivers.
In addition to the basic tasks of lamp operation which are also fulfilled by magnetic ballasts, Electronic Control Gear have much more to offer: They have better performance and are more reliable, more environmental friendly and more practical than CCG; even more reasons to use professional Electronic Control Gear.
OSRAM offers the right Electronic ballast for every application as shown at the example of T8/ 26 mm fluorescent lamps.
10
1.12 OSRAM ECG Milestones
Burning hours per day
Burning hours per day
24
24
®
QUICKTRONIC
QUICKTRONIC QUICKTRONIC
20
20
Industry,
Industry,
Open space office
Open space office
16
16
®
Railway
Railway
station,
station,
airport
airport
®
QUICKTRONIC
QUICKTRONIC
INSTANT START QTIS e
INSTANT START QTIS e
12
12
Department store,
Department store,
8
8
display
display
4
4
average
average
ECGs
ECGs
0
0
024 6 8
024 6 8
QUICKTRONIC
®
DIMMABLE
DIMMABLE
®
®
PROFESSIONAL
PROFESSIONAL
QUICKTRONIC
QUICKTRONIC
ECONOMIC
ECONOMIC
Daylight
Daylight
linked
linked
illumination
illumination
Switching cycles per day
Switching cycles per day
®
®
Daylight linked
Daylight linked
illumination with
illumination with
presence detectors
presence detectors
For the first time in 1995, T5- fluorescent lamp systems with Cut-
off-technology have been introduced to the market.
Cut-off technology is the cut-off of the permanent filament preheating after lamp ignition. This can be realized due to modifications in the electronic circuit of the ECG. The result of the Cut-off technology are less losses and optimized lamp operation.
Four years later, in 1999, OSRAM sold the first reliable
MULTIWATT-ECG.
This operates all lamps with rated data.
During the following years the trend of miniaturization continued and
the height of Electronic Control Gear was reduced by 30% from 30 mm to 21 mm. In 2002, OSRAM is again the first manufacturer to introduce
MULTIWATT-ECG to operate FQ
®
…HO-fluorescent lamps High
Output in 21 mm height.
In 2003, another novelty is brought to the T5-product segment: As
the first producer, OSRAM offers a 21 mm high 2-lamp ECG for
®
FQ
80 W HO-fluorescent lamps.
In the beginning of 2004, the newest and most innovative member
of the T5-product family has been introduced: micro-controller based ECG capable of operating T5-fluorescent lamps of equal length regardless if it is a FH
®
…HE- or FQ®…HO-type. This micro­controller especially developed under the co-operation with OSRAM is responsible for clear lamp detection and lamp operation with nominal data. QUICKTRONIC INTELLIGENT, QTi make one MULTIWATT-ECG possible for all T5-fluorescent lamps from 14 to 39 W no matter if
®
FH
…HE- or FQ®…HO-types.
11
2. Product Features
2.1 Lighting Comfort
2.2 Economy
2.3 Safety
Flicker-free ignition
Pleasant, flicker-free light with no stroboscopic effects due to high
frequency operation
High comfort level with no distracting choke hum due to fully
electronic operation (see chapter 2.8 noise)
No flickering
No flashing or flickering, electronic defective control for
reliable safety switch-off of defective lamps End-of-life safety shut down
Cut-off of the permanent filament preheating after lamp
ignition
Automatic restart after lamp replacement
High lumen packages for T5 FQ
Very high luminous efficacy for T5 FH
®
High Output system
®
High Efficiency system
Long lamp life due to lamp start with optimum filament pre-heating
and operation with cut-off technology
Low maintenance costs due to long lamp life and reduced relamping
intervals
Lower cooling load of air-conditioning systems due to lower losses
Light engineering with T5 (Ø 16 mm) lamps
Leuchte/ 1xL58 W 1xL58 W 1xFH35 W
Luminaire
Vorschaltgerät/ VVG/ EVG/ EVG/
Control Gear LLG ECG ECG
E [lux] 539 518 500
P
gesamt/total
[W] 260 220 154
% 100 % 85 % 59 %
Büro mit 4 1-flammigen Leuchten
Office with 4 single tube luminaires
4
3
2
1
0
012344,5 m
W/m
2
15 12 9
All Electronic Control Gear QUICKTRONIC
®
for operation of T5/ 16 mm-
fluorescent lamp systems are developed and designed according to all relevant national and international industry standards. Current standard is EN 61347. For Electronic Control Gear for operation of low pressure gas discharge lamps EN 61347-2-3 applies.
12
2.4 Supply Voltage
2.4.1 Overvoltage and its Reason
In detail:
Safe shutdown of the power supply to defective lamps or
at the end-of-life due to End-of-Life detection according to Test 2 Shutdown in the event of broken filaments, no inserted lamp or air leakage
Compliance with European safety standards (EN 61347-2-3)
Protection against short duration voltage surges (DIN VDE 0160)
and transient overvoltages
Low housing temperatures allow the mechanical design of lighting
fittings with F- and FF- as well as M- and MM-approval mark (EN 60598/DIN VDE 0710 and DIN VDE 0711)
Can be used in emergency lighting systems according to
DIN VDE 0108
Electronic Control Gear QUICKTRONIC lamps (FH
®
…HE, FQ®…HO and FC®) can be operated on sinusoidal AC
®
for T5/ 16 mm-fluorescent
voltage and DC voltage The recommended voltage intervals depend on the design of the specific circuits. The following chapters show the recommended voltage ranges and the behaviour of the ECG at overvoltage and undervoltage.
It is called an overvoltage if the ingoing voltage is significantly higher than the nominal value.
In general, we have to differentiate between two overvoltages which also can have different reasons.
1. Transient overvoltage with a typical duration of milliseconds. This overvoltage can be caused by:
- Switching of inductive loads such as welding machines, elevators alternators etc.
- lightening
Quasi-stationary overvoltage with a duration from a few minutes to hours. This overvoltage can be caused by:
- different loads on the mains side (interruption of the neutral conductor in 3-phase installations plus an additional asymmetric
13
2.4.2 Possible Implications due to Overvoltage
2.4.3 Undervoltage and its Reason
2.4.4 Possible Implications due to Undervoltage
2.4.5 Supply Voltage QT with 21 mm height
load distribution)
- unstable power supply (for example some countries in Far East)
It is called overvoltage, if the supply voltage exceeds the specified voltage range of an ECG including tolerances.
In any case, overload means more stress to electronic components. Depending on the magnitude of overvoltage the protective functions of an ECG can come into effect and turn the ballast off.
In extreme situations overvoltages can even destroy electronic components.
Therefore, please pay attention to the design of the mains and tolerances of the Electronic Control Gear when using them.
Supply voltages can not only deviate to higher values but also to lower values. If the supply voltage decreases below the value specified in the technical data of an ECG, we have to deal with undervoltage.
This may be true for the following points:
Different loads on the mains side
Incorrect electric installation
Unstable power supply
In some cases when used with emergency generators
Operating ECG with undervoltage is not as specified. This may result in the following implications:
Lamp operation not according to standards Î affecting lamp life
No safe lamp start, a safe ignition is only guaranteed above supply
voltages of 198 V
Unstable lamp operation meaning the discharge process of a
fluorescent lamp is not stable
In order to keep the lamp wattage constant most ECG types are
controlled on the lamp side. In this case, reduced supply voltages cause much higher currents which may lead to physical stress of components and to failure of the entire ECG. If supply voltages deviate significantly from the nominal values, high switching losses and overload of transistors can occur finally causing ballast failures.
Valid for: QTi and QT…F/CW Recommended voltage range for
continuous operation AC voltage 198 V ... 264 V, 50/60 Hz
DC voltage 176 V ... 264 V Performance at undervoltage
Lamp ignition Voltage drop during operation
U
198 V Î reliable lamp ignition
N
U
176V Î operation possible
N
U
< 176 VÎ damage to ECG
N
possible
14
2.4.6 Supply Voltage for QT with 30 mm height
2.4.7 ECG for 120V / 277V Line Voltage
2.5 Automatic Lamp
Restart
Valid for: QT-FH MULTIWATT and QT-FQ Recommended voltage range for continuous operation AC voltage 198V ... 264V, 50/60 Hz DC voltage 176V ... 264V Performance at undervoltage
Lamp ignition
U
198V Î reliable lamp start
N
Voltage drop during operation
U
176V Î operation possible
N
U
< 176V Î damage to ECG
N
possible
T5/ 16 mm fluorescent lamps are also getting more popular in North America (USA, Canada). Historically in the US-market have been established lamps in 4 ft length besides the types of 240 cm. 4 ft is also known as 48 inch type (1 ft = 30.48 cm) and is acc. to our typical 120 cm types. Related to the straight fluorescent types FH this means 1,149 mm for FH
®
28 W HE and FQ® 54 W HO.
®
…HE and FQ®…HO
OSRAM SYLVANIA offers the complete range for FH
®
…HE and FQ®…HO under PENTRON ECG. The specification there is PENTRON High Performance T5 lamps for FH
®
and PENTRON High Output T5 for FQ®
lamps.
OSRAM SYLVANIA also offers the ECG for the North American line voltages 120 V / 277 V und 60 Hz line frequency as shown at a glance:
MULTIWATT ECG for FH
®
fluorescent lamps: 14, 21, 28 and 35 W HE Types: QTP 1x28T5/UNV PSN suitable for 120-277 V QTP 2x28T5/UNV PSN suitable for 120-277 V
ECG to operate FQ
®
54 W HO Types: QTP 1x54T5UNV/PSN suitable for 120-277 V QTP 2x54T5UNV/PSN suitable for 120-277 V
For a large number of differnt lamp types including T8 OCTRON a variety of dimmable and non dimmable types is available.
Information about availability under:
OSRAM LIGHT CONSULTING (OLC) Hellabrunner Straße 1 81536 München
Tel: +49-89-6213 3076 Fax: +49-89-6213 2020
With all QUICKTRONIC lamps FH
®
…HE, FQ®…HO and FC®, automatic restart takes place after
®
for operation of T5/ 16 mm-fluorescent
a change of lamp provided the power supply is maintained.
Should in the case of a twin-lamp ECG no automatic lamp restart take place after lamp replacement and could an ECG-failure be excluded, please proceed as follows:
15
2.5.1 Lamp ignition for QTi
2.5.2 Lamp ignition for QT to operate T5­fluorescent lamps
2.5.3 Off- Time for Optimum Preheat Start
2.6 Behaviour in Response
to Lamp Defects
2.6.1 One-Lamp Operation with Multi-Lamp ECG
Replace both lamps, take out the lamp replaced first and refit it. Provided lamp and ECG are o.k. both lamps should then light.
Lamp start Preheat
Ignition time < 1 second
Max. number of switching cycles > 100,000 cycles
QT-FH MULTI, QT-FQ, QT…F/CW
Lamp start Preheat
Ignition time < 0.5 second
Max. number of switching cycles > 100,000 cycles
All QUICKTRONIC to operate T5/ 16 mm- fluorescent lamps
®
FH
…HE, FQ®…HO and FC start the lamps at any time with optimum preheat start even after a turn-off followed by an immediate lamp restart. OSRAM QUICKTRONIC ignite the lamp always with optimum preheating of the electrodes. A particular off-time with regards to lamp life is not necessary.
What do we mean by lamp defect or end-of-lamp life?
In most cases, it is not possible to see from outside which lampholders are assigned to which ECG-terminals, so if you insert lamps and they fail to start automatically you should take out the first lamp again and refit it. Both lamps should then light.
Lamp replacement of 2- and multilamp luminaires proceed as follows:
Insert the lamps. If at 2- or multilamp luminaires lamp ignition doesn’t work automatically, take out the lamp replaced first and refit it. Reignition of both lamps works automatically.
What are the requirements?
QTP 2x...
QT - FQ 2x...
QT - FQ 2x...
L
ILI
UU
Start
U
U
Zünd
Start
L
L
I
ILI
L
ILI
UU
Start
L
L
I
ILI
Parallel Circuitry Series Circuitry
16
2.7 Noise
Parallel circuit of lamps operated with multi-lamp ECG in general
single-lamp operation possible
Parallel circuit of lamps, but no single-lamp operation possible because
for example
- the sum of electrodes has to be recognized
For twin- and multi-lamp ECG the question is whether the remaining lamps will continue to operate if one lamp is defect or has been removed.
In the case of twin- or multi-lamp ECG, any lamp fault that causes the safe shutdown circuit to operate will lead to the shutdown of all lamps. This function is called “safety shutdown”. The detection of various “out-of­range” parameters results in a reliable shutdown of the ECG. The ECG do not perform any lamp starts that could cause problems as described under chapter 2.3. In this case, one lamp or the remaining lamps will therefore never continue to burn by itself.
What happens when one lamp is removed from a multi-lamp ECG will depend on the type of circuit. Series circuits always exclude a single-lamp operation. Parallel circuit is one condition for a possible single-lamp operation, however, not the only one. Another important criterion is lamp control during operation of circuit related as well as safety related data.
QUICKTRONIC
®
INTELLIGENT, QTi, are carried out in parallel circuits, but cannot be operated in single-lamp mode. The reason is the sophisticated lamp detection requiring the control of various parameters.
The following table gives a short summary of the different ECG-types:
ECG-type height
QTi
21 mm
QT-FH
30 mm
QT-FQ 30 mm
QT-FH 3x, 4x
30 mm
QT … F/CW
21 mm Series circuit X X X X Parallel circuit. X
For all types shown in the table above a single-lamp operation is not possible.
T5/ 16 mm-fluorescent lamps FH at high frequency with QUICKTRONIC
®
…HE, FQ®…HO and FC® operated
®
control gear are virtually
silent. QUICKTRONIC
®
units themselves are so quiet that even in very quiet surroundings they cannot be heard by the human ear. They are therefore ideal for sound-sensitive areas such as radio and recording studios. The limit of the frequency-dependant sound pressure curve is based on the audibility threshold (in other words, a person with normal hearing will not be able to detect the noise generated by an ECG in the same room).
The factors affecting the sound pressure level are the sound power level of the ECG, the absorption properties of the room, characterised by its volume and reverberation time, and the number of ECGs. In mains supplies with a high level of distortion in which the mains voltage wave form deviates significantly from a sine wave, a „chirping“ may be heard from the reactance coils in the input section of the ECG.
17
2.8 Power Factor λ For all electric loads, the power factor λ is the ratio of effective power (P
voltage x effective current) to apparent power (P
= voltage x apparent
app
current). This value is affected both by the phase displacement cos ϕ between current and voltage by the current wave form distortion ε (non­sinusoidal wave form)
=
eff
λ = P
eff
/ P
= ε cos ϕ
app
In contrast to conventional control gear (CCG, inductive, 50 Hz), there is hardly any phase displacement with Electronic Control Gear (high frequency), which means that capacitor correction is not required. However slight distortions in the current sine-wave curve occur during operation of electronic control gear. In general these distortions are characterized by integer multiples of the mains frequency (harmonics).
The harmonic content of the mains current is controlled by national and international regulations (IEC 61000-3-2, EN 61000-3-2). OSRAM ECG have built-in active electronic harmonic filters for this purpose which guarantee a value for ε of more than 0.95 and hence a power factor λ greater than 0.95.
Exemptions are ECG which apply to the international standard for system power consumption less than 25 W. This standard requires a power factor λ of more than 0.6. These ECG are part of the product segment ECO and are known as QUICKTRONIC
®
ECONOMIC or QT-ECO. They are mainly used to replace conventional control gear, but because of their MULTIWATT-design they can partly operate FH
®
…HE- and FQ®…HO
fluorescent lamps with lower wattages:
®
FH
14W HE
®
FH
21W HE
®
FQ
24W HO
For detailed information about this combination please see
www.OSRAM.de/products/ecg
With regards to their harmonics content all QUICKTRONIC
®
have been tested by VDE according to EN 61000-3-2 and carry the VDE-EMC approval mark.
The confirmation of the ECG’s CE-mark by an independent testing facility reduce the costs and the t.ime necessary for approval of luminaires significantly.
18
2.9 ECG Imprint
2.10 Reliability
2.11 Resistance to Frequent Switching
2.12 ECG Lifetime
Functional Earth terminal
Functional Earth terminal
Functional Earth terminal
End-of-Life Safety-shut-
End-of-Life Safety-shut-
End-of-Life Safety-shut­down
down
down
Cut-Off-Technology
Cut-Off-Technology
Cut-Off-Technology
µProzessor inside
µProzessor inside
µProzessor inside
Thermal Behaviour
Thermal Behaviour
Lamp wiring including
Lamp wiring including
Lamp wiring including max. cable lengths
max. cable lengths
max. cable lengths
Besides component specification and quality their failure rate is significantly related to the operating temperature.
Electronic Control Gear of OSRAM are designed in that way, that a failure rate of less than 2 Promille per 1,000 operating hours is expected if operation takes place at the maximum permitted case temperatur (t
).
c
The resistance to frequent switching of Electronic control gear is based on possible lamp starts per day. Multiplied with the lamp life professional ECG with preheat start reach more than 100.000 switching cycles.
This information is important for combinations with occupancy sensors as this is one of the most popular applications for frequent switching of the lamp-ECG system.
The ECG lifetime depends on the operating temperature and failure rate of the electronic components. Extreme overheating can destroy electronic components in a short period of time and cause the ECG to fail. There exists an exponential relationship between the failure rate of electronic components and their thermal and also electrical behaviour. An indication about the maximum recommended ambient temperature of a luminaire is given by the imprinted measuring point tc at which each ECG reaches its maximum recommended case temperature. The t
-temperature
c
of an OSRAM ECG is closely linked to its temperature of electronic components. For example, when the maximum recommended tc­temperature of 70 °C is reached, a QUICKTRONIC control gear for operation of T5/ 16 mm-fluorescent lamps is expected to last 50,000 hours with a failure rate of max. 10 %. This value equals a failure rate of 2 ‰ per 1,000 operating hours. Due to the exponential dependence on temperature and failure rate of electronic components exceeding the recommended tc-temperature reduces the ECG lifetime dramatically. On the other hand, if the ECG temperature remains below the limit the lifetime is extended. As a rule of thumb, every 10 °C below the imprinted temperature value double the ECG’s lifetime and every 10 °C surpassing the tc-value cut the service life in half.
The measuring temperature t
is an important parameter for both the safety
c
approval for a luminaire according to EN 60598 and the service life of an
19
2.13 Thermal influences of the system components
2.14 cut-off Technology
ECG provided by the manufacturer under consideration of the thermal load of electronic components.
Surviving ECGs [%]
Surviving ECGs [%]
Surviving ECGs [%]
100
100
100
80
80
80
60
60
60
90°C
90°C
40
40
40
20
20
20
90°C
0
0
0
0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140
70°C
70°C
70°C
10°C lower operating temperature at point of
10°C lower operating temperature at point of
10°C lower operating temperature at point of measurement virtually halves ECG failure rate
measurement virtually halves ECG failure rate
measurement virtually halves ECG failure rate
The temperature must be assessed separately for the two system components (ECG and lamp). In the case of the lamp, there are physical laws that restrict the temperature range of an application, whereas in the case of the ECG fixed limits must be set in order to ensure reliable operation.
Apart from this, there are external factors such as the reciprocal influences of ECG, lamp and luminaire and the selected installation site which have an influence. Compliance with the specified limits and hence the guarantee of operational reliability are the responsibility of the relevant luminaire or system manufacturer.
There is a fixed correlation between tc-temperature, the temperatrure of electronic components, the life of each component and hence the complete unit. Thermal contact of an ECG to metallic parts of luminaire housings is very positive due to good thermal conductivity.
The correlation between temperature tc of the measuring point, component life and failure rate is crucial for an objective assessment of the reliability and service life of an ECG. Information about tc-temperature and ECG service life alone are not sufficient.
All QUICKTRONIC T5/ 16 mm-fluorescent lamps FH
are equipped with cut-off technology. After starting the lamp the electrode heating is being switched off. Lamp life increases due to the reduced load of the electrodes. Therefore, cut-off technology increases the lumen output of the luminaire and the lamp life. And for the first time, cut-off technology is included in dimmable ECG thanks to the new intelligent technology of QTi. Compared to Electronic control gear without cut-off the power consumption of ECG with cut-off technology could be reduced by another 5 to 7 %.
10 % loss at 50,000 hours
10 % loss at 50,000 hours
10 % loss at 50,000 hours
60°C
60°C
60°C
Hours of usage [thsd]
Hours of usage [thsd]
Hours of usage [thsd]
®
control gear for operation of
®
…HE and FQ®…HO
50°C
50°C
50°C
Temp. at point of
Temp. at point of
Temp. at point of measurement t
measurement t
measurement t
c
c
c
20
2.14.1 Advantages for Users
The following advantages for users arise from cut-off technology:
6-10 % higher luminaire efficiency highest lamp life 2-3 W lower losses per lamp reduced load of air condition
New circuitry wihtout permanent filament heating
New circuitry wihtout permanent filament heating (cut-off technology)
(cut-off technology)
I
= I
I
= I
Stift
Lampe
Stift
Lampe
Cold Spot
Cold Spot app. 40°C
app. 40°C
Significantly lower temperatures at the
Significantly lower temperatures at the lamp electrodes
lamp electrodes
I
I
Lampe
Lampe
I
I
Stift
Stift
Cut-off advantage for the luminaire
Cut-off advantage for the luminaire
Cut-off advantage for the luminaire
110
110
110
100
100
100
90
90
90
rel.
rel.
rel.
80
80
80
Licht-
Licht-
Licht­strom
strom
strom
70
70
70
60
60
60
10 20 30 40 50
10 20 30 40 50
10 20 30 40 50
Umgebungstemperatur (°C)
Umgebungstemperatur (°C)
Umgebungstemperatur (°C)
Only cut-off technology can fulfill lifetime
Only cut-off technology can fulfill lifetime
100%
100%
90%
90% 80%
80% 70%
70% 60%
60%
Rel.
Rel.
50%
50%
Lum-
Lum-
40%
40%
inous
35°C
35°C
35°C
Cut Offk
Cut Offk
Cut Offk Conventional
Conventional
Conventional ECG circuitry
ECG circuitry
ECG circuitry
inous flux
flux
6 – 10% higher
6 – 10% higher
6 – 10% higher luminaire efficiency for
luminaire efficiency for
luminaire efficiency for direct lighting
direct lighting
direct lighting
5.000h
5.000h
Lifetime [hours]
Lifetime [hours]
Cut-off
Cut-off
Cut-off Conventional
Conventional
Conventional ECG circuitry
ECG circuitry
ECG circuitry
10.000h
10.000h
12.000h
12.000h
13. 16.
13. 16.
16.000h
16.000h
14.000h
14.000h
16. 2 0.
16. 2 0.
20.000h
20.000h
18.000h
18.000h
2.14.2 Physical Background
T5/ 16 mm-fluorescent lamps FH
®
…HE and FQ®…HO are designed to
have their maximum lumen output at 35 °C (compared to 25 °C for T8/ 26 mm). For T5/ 16 mm-fluorescent lamps the so-called cold spot (the point where mercury condensates in a discharge tube, stamped end of the lamp) is located behind the electrode (see graphics) which means near the source of heat.
T8 (Ø 26 mm)
T8 (Ø 26 mm) T12 (Ø 38 mm)
T12 (Ø 38 mm)
Cold Spot T
Cold Spot T
The value of the luminous flux at the ambient temperature of 35 °C is only informative for the luminaire manufacturer. Significant is the value of the cold spot temperature. This value is measured at the socket of the stamped side, approx. in a distance of 2 mm of the glass. For an optimized luminous flux this value should be between 45 °C and 50 °C. This is shown at the so called ‘Horseshoe curves’ where the luminous flux is shown in relation to the lamp ambient temperature.
The cold spot of the T5/ 16 mm fluorescent lamps is influenced by permanent filament heating. This means that higher temperatures reduce the luminous flux. ECG with cut-off technology reduce the losses of the system and are optimized regarding the maximum luminous flux of the
Hg opt.
Hg opt.
40°C
40°C
T5 (Ø 16 mm)
T5 (Ø 16 mm)
21
2.15 End of Life (EoL acc. to T.2)
2.16 U-OUT
system. The cut-off of the permanent filament heating after lamp ignition is an advantage.
Further the values of the horseshoe curves also indicate the arrangement of the lamps within the fixture. To avoid thermal influences of the lamps minimum distances in between have to be kept. The lamps have to be placed that the stamp of all lamps is on the same side. For vertical arrangement the stamp of the lamp always should be placed down. For circular lamps FC the socket has to be placed down.
The measurement of the cold spot temperature is especially important for the luminaire manufacturer. This temperature offers opportunities to optimize the luminaire efficacy.
Fluorescent lamps use up their emitter during operation. The complete loss of emitter on an electrode results in a voltage increase in the vicinity of the electrode filament. As most Electronic control gear have no problem providing this high asymmetric voltage and with it the required additional power the temperature around the electrodes rises significantly.
At the moment the international ECG safety standard IEC 61347-2-3 is under revision. In the future, all ECG must provide a “end-of-life” safety shutdown which is continuously controlled in order to avoid overheating of lamp sockets. The actual status of the standard considers three different test circuits for Electronic control gear. If an ECG complies with one of the three tests, it offers protection against the “end-of-life” behavior of fluorescent lamps.
Asymmetric pulse test (Test 1)
Asymmetric power test (Test 2)
Open filament test (Test 3)
The asymmetric power test (Test 2) is directly simulating the additional load which results from the increased asymmetric voltage in the case of “end-of­life”. In order to pass the test the additional load may nod exceed a specific value depending on the lamp type. Most ECG experts see test 2 (asymmetric power test) as the safest “end-of-life” control, because the direct measurement of the asymmetric additional load mirrors the real lamp behavior at its end of life. OSRAM does not compromise the safety of Electronic control gear and has, for quite some time, specified Test no. 2 as standard test.
U-OUT a binding ECG label according to safety standard EN 61347-2-
3. The former standard EN 60928, still valid until 2006, allows labeling of U-OUT either in the product catalogue or on the ECG housing. U­OUT specifies the largest effective working voltage between
- Lamp terminals
- Each lamp terminal and earth connection, if applicable
This information is important for all components electrically connected on the lamp side of the ECG. All components such as lamp cables, sockets (EN 60061-2), isolating
22
2.17 Approval Marks
2.17.1 ENEC-Approval Mark
material and everything coming in contact with the ECG lamp terminals must be layed out according to U-OUT. OSRAM, as manufacturer, takes care that no higher voltage appears at the lamp terminals than specified by U-OUT. Therefore no additional voltage reserve is needed.
2.17.1.1 Safety acc. to
EN 61347
2.17.1.2 Performance acc. to
EN 60929
2.17.2 VDE EMC mark
stands for E also a conformity mark agreed upon between the testing institutes of the European union. It stands for compliance with the according European standards for safety and performance. Besides sample testing ENEC includes also a permanent control od products and production processes. This certification is testimony of an independent and competent institute testing the safety and performance of Electronic control gear. The number right beside the approval mark identifies the certifying institute. For example 10 is representing VDE in Germany. The ENEC approval mark for ECG to operate fluorescent lamps includes the safety standard EN 61347 and the performance standard EN 60929.
This standard contains the safety requirements of electric units for operation of lamps that are designed for DC- and AC-voltage at 50 or 60 Hz. It is divided into different parts.
The first part EN 61347-1 deals with general and safety requirements.
b) Electronic control gear to operate with AC-voltage at 50 or 60 Hz
This safety standard whose part EN 61347-2-3 together with the general part EN 61347-1 succeeds the former standard EN 60928, is also called „omnibus“ standard.
This standard specifies the performance of Electronic control gear for fluorescent lamps. It defines the operation at AC-voltage, at 50 or 60 Hz and with a supply frequency different from the operating frequency. It is based on performance standards for fluorescent lamps EN 60081 and EN
60901.
uropean Norm Electrical Certification. The ENEC approval is
with an operating frequency different from the frequency of the mains supply and to operate fluorescent lamps according to IEC 60081 and IEC 60901 and other fluorescent lamps for high frequency operation are dealt with in part EN 61347-2-3.
The independent testing institute confirms the compliance of the ECG with the EMC regulation regarding immunity, radio interference suppression and harmonics. At the same time, it is also a confirmation for the CE label that can be applied to ECGs by the manufacturer himself under compliance with the EMC regulation. Luminaire manufacturer can significantly reduce their costs and approval efforts with regards to EMC compliance by using already EMC approved ECGs.
23
2.18 Energie Efficiency Index EEI
2.19 CE Labelling
This label helps consumers identifying the energy consumption of a product. Usually, all Electronic control gear have the best ratings A2 …A3. Dimmable ECG are classified as A1. Magnetic ballasts (CCG) fall under the energy efficiency class C and D and are either already banned from the market or are about to be banned shortly. Low loss ballasts are usually classified in B.
Since January 1996, all products falling under the EU directive of electromagnetic compatibility (EMC) must carry the CE label. The CE label indicates the compliance with the requirements of this directive. From January 1997 all products falling under the Low-Voltage directive must also be CE labeled. There is no question that our products comply with the specific EU directives and therefore are labeled with the CE mark.
Regarding CE labelling here the following explanations:
1. CE-label as basis to introduce product to the market
Since 01.01.1996 manufacturers and importers are obliged to apply CE labels to products that have to comply with EMC regulations either directly on the product, its packaging or the accompanying documents. CE labels are obligatory for the sale of products within the European Union. By applying the CE label manufacturers or importers confirm that their products comply with the “basic requirements” of specific European directives and fulfill their protective goals (for example electromagnetic compatibility). Usually the compliance of particular “basic requirements” is given products are manufactured under consideration of harmonized European standards.
2. CE label is a mark for administrative authorities
The CE label is targeted primarily at administrative authorities. It declares to them that CE labeled products comply with European jurisdiction at the time of sale.
3. No rights for commerce and endusers of examining declarations of conformity issued by manufacturers.
The right to ask for and examine declarations of conformity is reserved to authorities controlling the compliance of electric/electronic products with the legal safety requirements. In Germany it is the “Federal Agency for Post and Telecommunication” BAPT (responsible with regards to the EMC directive) and the trade supervisory boards (responsible with regards to the Low-Voltage directive).
4. CE labels are no quality or approval marks
The only purpose of the CE label is to testify that a product complies with legally specified „basic requirements“ of certain directives. It
24
2.20 CCC Approval
does not make any statement with regards to the quality of labeled products. As a legal administration label without value for endusers it should not be mistaken as an approval mark issued by independent testing institutes (such as ENEC or VDE mark). These testing institutes do also not control the legitimacy of an applied CE label.
Approval mark of the Chinese testing institute CQC (C
C
enter).
hina Qualification
Since 01.08.2003 this approval mark is required in order to sell Electronic control gear for operation of low pressure discharge lamps in the Chinese market.
OSRAM QUICKTRONIC for operation of T5/ 16 mm- fluorescent lamps
®
FH
...HE and FQ®...HO carry this approval mark.
25
3. ECG installed in Luminaire: Installations and Operation Instructions
3.1 Wiring Instructions
3.1.1 Cable Types
3.1.2 Cable Cross-Sections
Please pay attention to the voltage value U-OUT imprinted on the ECG housing when wiring luminaires for FH lamps. This value indicates the possible cable type. For voltages greater than 430V cables with the classification H07 have to be used. U-OUT is the maximum voltage that can occur between the lamp terminals or the lamp terminal and the earth connector. For all OSRAM QUICKTRONIC ECG for operation of T5/∅ 16 mm- fluorescent lamps FH allowing luminaires to be wired with cables H05. Cable types are specified through the terminals in use.
The cable cross-sections are marked on the identification plate of the Electronic control gear.
C
ombi-Wiring (CW) stands for use in automatic or manual wiring. ECG in
30 mm height have the annex CW at the end of the type. ECG types without CW are suitable for manual wiring only in this height. T5-ECG in 21 mm height don’t have this annex as they are equipped with CW terminals only for manual or automatic wiring.
For manual wiring of the IDC a special tool is available for example from WAGO. This tool is listed and can be ordered from WAGO under the order number
0206-0831.
®
…HE and FQ®…HO U-OUT is less than 430V
®
…HE or FQ®…HO fluorescent
Ancillary for manual wiring of the IDC-contact of the CW terminals
26
3.1.2.1 ECG in 30 mm height
3.1.2.2 ECG in 21 mm height
3.1.3 Release of Contacts
3.1.3.1 WAGO 250
Typical values for C
a) Single-core cables
These should have a cross section of 0.5mm² at least and 1.5 mm² at most for the horizontal plug. When using I have a maximum cross section of 0,5 mm²
b) Multi-core cables
horizontal plug
These should have a cross section of 0,5 mm² at least and 1 mm² at most. Multi-core cables can be inserted directly into the horizontal plug terminals. Ferrules may be used but they are not essential. IDC Multi-core cables with a cross section of 0,75 mm² can be used for direct wiring with IDC.
Typical values for C
c) Single-core cables
These should have a cross section of 0,5 mm² at least and 1 mm² at most for the horizontal plug. When using I have a maximum cross section of 0,5 mm²
d) Multi-core cables
horizontal plug
These should have a cross section of 0,5 mm² at least and 1 mm² at most. Multi-core cables can be inserted directly into the horizontal plug terminals. Ferrules may be used but they are not essential. IDC Multi-core cables with a cross section of 0,75 mm² can be used for direct wiring with IDC.
ombi-Wiring terminals of ECG with 30 mm height are:
nsulation Displacement Contacts (IDC) cables should
ombi-Wiring terminals of ECG with 21 mm height are:
nsulation Displacement Contacts (IDC) cables should
3.1.3.2 WAGO 251 – IDC
Release the contact by using the release latch.
Release the contact by pulling the cable upwards. This process can be repeated up to 10 times (depending on the manufacturer) without damaging the terminal. For further details please refer to the data sheets of the manufacturers.
27
3.1.3.3 WAGO 251 – horizontal plug
3.1.3.4 WAGO 251 mini – IDC
WAGO 251 – horizontal plug
The contact can be released with the help of a small screw driver as
Alternatively the contact can be released by simultaneously twisting and pulling the cable.
3.1.3.5 WAGO 251 mini – horizontal plug
Release the contact by pulling the cable upwards. This process can be repeated up to 10 times (depending on the manufacturer) without damaging the terminal. For further details please refer to the data sheets of the manufacturers..
The contact can be released with the help of a special manual tool as shown in the picture below. The tool can be ordered from WAGO under the order number
0206-0830.
Alternatively the contact can be released by simultaneously twisting and pulling the cable.
28
3.1.4 Insulation
3.1.5 Terminals
3.1.6 Cable routing
Depending on the type of terminal the length of insulation to be stripped from the ends of the cables is different. The exact value can be found on the ECG.
WAGO 250 WAGO 251 WAGO 251-mini Stripped Insulation
8-10 8.5 - 11 8.5 - 11
[mm]
h = 30 mm h = 21 mm QT-FQ…CW WAGO 251 QT-FH MULTI…CW WAGO 251 QT-FH…F/CW WAGO 251 mini QT-FQ…F/CW WAGO 251 mini QT-FQ 2x80 WAGO 251 mini QTi WAGO 251 mini
To ensure good radio interference suppression as well as maximum safety and reliability, the following rules for cable routing should be observed:
1.) Cables between ECG and lamp (HF cables) should be kept as short as possible to reduce electromagnetic interference. Please pay attention to the maximum recommended cable lengths as indicated on the ECG (see also overview in 7.1ff)
2.) Mains and lamp cables should never be routed in parallel. Keep HF cables and mains cables as far away from one another as possible (e.g. 5 to 10 cm). This avoids mutual interference between mains and lamp cables.
3.) Lay HF cables away from earthed metal surfaces (if possible several cm away) to reduce capacitive interference.
4.) If long HF cables are unavoidable (e.g. in master-slave circuits) they should be twisted together.
5.) Keep mains cabled in the luminaire as short as possible to reduce interference.
6.) Do not lay mains cables too close to the ECG or the lamps. This applies in particular to through-wiring.
7.) Avoid crossing mains cables and lamp cables; if this is not possible, they should cross at right angles to reduce mains and HF interference.
8.) Lamp cables at high potential (see section 7.1ff „Hot wires“) must be kept as short as possible, particularly with luminaires for tubular fluorescent lamps such as
®
FH
…HE and FQ®…HO.
29
3.2 Electromagnetic Compatibility
3.2.1 Harmonic Content
acc. to EN 61000-3-2
Wiring must comply with the latest versions of the relevant national standards. Cable entry through metal components should never be left unprotected but should be fitted with additional insulation (sleeve, grommet edge protector etc.)
The body of the luminaire or parts thereof must never be employed as a conductor or in any way come into contact with mains or lamp cable conductors (for example as a result of bare cables, too much insulation stripped away, screws protruding through insulation, or sharp metal edges). There is a serious risk that a person may be electrocuted and the control gear damaged beyond repair.
If you are wiring a number of luminaires from a 3 mains supply in a loop (with 5-core cable, for example), you must also ensure that are never connected two phases to the ECG mains terminal instead of the group phase, the neutral conductor and the PE conductor. Otherwise the ECG may fail immediately or within a short time. (see also section 8.4, Troubleshooting)
L and N interchangeable (e.g. for mobile luminaires)?
Yes case labeling or
No case labeling L N
The abbreviation EMC stands for ElectroMagnetic Compatibility. EMC specifications define a series of different test criteria. The most important in connection with electronic control gear are radio interference suppression, harmonic content (up to the 39
th
harmonic) and immunity to interference.
IEC International European
standard
Radio interference
CISPR 15 EN 55015
suppression Harmonic content Immunity to interference
IEC 61000-3-2 EN 61000-3-2 IEC 61547 EN 61547
The CE symbol on OSRAM QUICKTRONIC control gear indicates compliance with immunity to interference, harmonic content and radio interference suppression requirements. By using the CE label OSRAM, as manufacturer of ECG, confirms the compliance with the requirements of the sandards (see also Section 2.17.2 CE labeling).
Lighting equipment is subject to restrictions on harmonics. The maximum permissible threshold values are defined for two classes of equipment:
Class C: Effective input power (system wattage) > 25W
Class D: Effective input power (system wattage) < 25W
The requirements for Class D equipment must be met as of January 1, 2001.
30
3.2.2 Radio interference
suppression
3.2.2.1 Causes of Radio
Interference
3.2.2.2 Conducted
Interferences acc. to EN 55015
Harmonic number
Proportion in % of the mains current of the fundamental wave (50 Hz)
2 2 3
30 x power factor(λ)
5 10 7 7 9 5
11 < n < 39 3
The specified values apply to Class C ECGs. OSRAM QUICKTRONIC
®
control gear typically exhibit values well below
the threshold values.
All QUICKTRONIC
®
units for operation of T5 FH®...HE- and T5 FQ®...HO
fluorescent lamps have a total harmonic distortion (THD) of less than 10%.
Compliance with the limit values for radio interference suppression is also a requirement for the VDE EMC approval.
The ECGs are measured in a reference luminaire.
However, the interference level depends not only on the ECG but also on the arrangement of the lamp and ECG, the luminaire design and in particular the wiring. To obtain approval, compliance with the limit values has to be checked for each luminaire (by the VDE for example).
This is discussed in more detail below.
Radio interference refers to both the radiated and the mains-borne influences of an electrical load on other units connected to the same mains supply and/or in the immediate vicinity. See also Section 4.8.
To ensure that the various electrical loads can operate simultaneously and trouble free, each unit must not exceed certain radio interference values.
A distinction is made here between mains-borne disturbances on the power supply side and atmospheric disturbances due to electromagnetic fields in the vicinity of the luminaires.
These disturbances are caused by non-linear components and the high frequency operation.
31
3.2.2.3 Disturbances due to
Fields
3.2.2.4 Selective Shielding
By using complex input filters it is possible to reduce these disturbances to a level well below the limits prescribed by the relevant standards. Each and every OSRAM ECG complies with these standards. The way in which an ECG is installed in a luminaire, however, can have a considerable influence.
(see Section 7.5 Installation instruction for luminaires)
Owing to the various dynamic electrical and physical processes in the ECG-lamp system, there is always an electromagnetic field around the luminaire. In defining its effects, a distinction is made between electric and magnetic components.
a) Electrical fields
Because of the high-frequency output voltage the lamp and lamp wiring generate an electrical field. This is attenuated to a level well below that prescribed in the relevant standards by filtering on the output side and by suitable circuit design. What applies to an individual ECG does not always automatically apply to the entire system once the ECG has been installed in the luminaire.
b) Magnetic fields
The magnitude of this field is determined solely by the lamp and the geometrical arrangement of the entire system, particularly by the areas enclosed by the lamp, ECG and wiring. The only recommendation that can be made here is to aim for a self­enclosed structure as far as possible and to comply with the wiring instruction.
The following diagrams show the magnetic lines of force for two simple linear luminaires
32
p
3.2.2.5 Installation
Instructions for avoiding Disturbance
a) without reflector b) with a metal reflector
The resulting magnetic field strength in the near field and hence the effect on the environment is reduced in b) by a current induced in a reflector. It is important here for the surface of the reflector to have good electrical conductivity. It is not necessary to earth it.
To shield the electric field, which is always radial around the lamp, it is necessary for the reflector or its surface to be as conductive as possible ant for the connection to earth or to protective earth to be of the lowest possible resistance.
Based on these two requirements, the solution here is to have a reflector, reflector and diffuser or louver with excellent conductivity connected at lowest possible resistance to the ECG earth (PC I) or the PE conductor connection of the luminaire.
The following diagrams show examples of correct and incorrect wiring.
1a) Long-run luminaire with reflector
To avoid interference on the lamp cable, the mains cable shoud be routed to the outside immediately at the luminaire terminal. He lamp cables should be laid in accordance with the criteria specified in Section xx (wiring instructions). The reflector is used here for shielding and should therefore be made of metal and be attached permanently with a high quality plug connector (must have low resistance) to the luminaire body which in turn is connected to control gear earth.
Line voltage
Lam
Hot wires
1b) Bad example
Mains and lamp cables that run in parallel over long distances. This leads inevitably to interaction and therefore to higher energy in the radiated
33
electromagnetic field. Serious problems can occur if, as described in
p
Section xx, wiring instructions, the lamp cables that have high potential with respect to earth (hot wires) are not kept as short as possible by connecting them to the nearest lampholder.
3.2.2.6 Asymmetric
installation of ECG
Lam
Hot wires
The following diagrams apply to both recessed and surface-mounted luminaires:
Hot wires
Cables should be laid close to the body of the luminaire. ECG and reflector need a low resistance earth. Wiring must comply with the recommendations in Section xx (Installation instructions for luminaires). The luminaire design provides effective shielding of the electromagnetic field.
Alternative asymmetric installation
Hot wires
This option is equally suitable.
34
3.2.2.7 Good wiring
arrangement for 2­lamp luminaires
Bad example
The electrical connection between the ECG and the luminaire is poor. Unnecessary crossovers have been created leading to poorer and therefore higher resistance connection with earth. This arrangement is also poor from a thermal point of view.
Übergänge
Hot wires
3.2.2.8 Luminaires with
reflector and/or specular louvres
Hot wires
These parts must be made of metal or at least have a surface (i.e. anodized) with excellent electrical conductivity.
a) The reflector acts as an effective shield Provided the reflector has a very good connection to the central earthing point, the lamp is effectively shielded and there can be no interaction with the ECG and the wiring. Electromagnetic fields are also effectively shielded.
The reflector should be connected by a short cable or screw connection to the body. A poor contact or a loss of contact at this point would have an adverse effect on the EMC behaviour of the complete luminaire and could also impair starting.
35
b) Louvre instead of a reflector
The same applies to louvers as to reflectors. Louvres also have to be good electrical conductors and be connected to luminaire earth.
3.3 Permissible Cable Lengths
3.4 „Hot Wires“
Section 7.1 refers to the maximum recommended cable lengths between the ECG and the lamp. The additional information is discussed elsewhere in this section. These maximum recommended cable lengths must be adhered to in order not to overload the ECG and to ensure that the system will start reliably even under adverse conditions (low ambient temperatures, high humidity levels, aged lamps).
In order to comply with the radio interference suppression limits, the instruction in Section xx must be followed. If the maximum recommended cable lengths for operating an ECG-lamp system are fully exploited, additional radio interference suppression measures may be needed such as shielding or separate filters. Because there are so many different interference factors involved (see Section 3.2), it is not possible to specify a maximum cable length for an ECG-lamp system below which radio interference suppression limits are guaranteed not to be exceeded.
„hot wires“ (high potential). By this we mean the lamp cables which are at the highest potential with respect to circuit earth or protective earth. The other lamp cables are “cold wires” and have a correspondingly lower potential with respect to earth.
36
3.5 Switching between Lamp and ECG
3.6 Master-Slave Circuit
„Hot wires“ are marked on the unit with the shorter cable length. For reasons of radio interference suppression and reliable starting, the “hot wires” must be kept as short as possible. In other words, you should install the ECG to one side in the luminaire, making the low-potential cables longer so that the length of the high-potential cables can be reduced. This type of installation is to be preferred to symmetrical mounting.
In luminaires equipped with more than one ECG (systems with 3, 4 or more lamps), the ECG and its associated lamp(s) should be assigned to one another. For reasons of radio interference suppression and reliable starting we do not recommend splitting the luminaire into a lamp and ECG part. QUICKTRONIC
®
ECG for 3- and 4-lamp operation of FH®…HE fluorecent lamps have been optimally designed to comply with just that recommendation.
For reasons given above we also advise against splitting the ECG into units located in the luminaire and units located outside the luminaire (e.g. on the back of the luminaire) if this means much longer cables between the ECG and the lamps.
In some special applications it may be necessary to disconnect or switch the cables between the ECG and the lamp(s).
If changeover units are used (emergency lighting modules with internal switching) which supply the lamp directly from an emergency supply and interrupt the system circuit between the ECG and the lamp, the following must be observed:
Changeover or disconnection of the lamps from the ECG to the
external unit must be on all terminals.
When switching back from the external supply to ECG operation,
the lamp(s) must first be connected at all terminals to the ECG before the ECG is supplied with power again, otherwise the cutout in the ECG will operate.
Many of these emergency lighting units available on the market do
not comply with the normal operating conditions of the lamp and will therefore damage it. In such cases, OSRAM cannot guarantee that the lamp will last as long as indicated.
Wiring recommendations for OSRAM QUICKTRONIC
®
ECG for multi-lamp
operation of T5/ 16 mm- fluorescent lamps with exemplary emergency units are given in Section 6.5.2ff.
An additional single-lamp „slave luminaire“ can be supplied from a two­lamp ECG installed in a single-lamp „master luminaire“.
This requires a 4-core connecting cable between the two luminaries and, in general, different cable lengths between the ECG and the lamp in the
37
3.6.1 Max. length of the connecting cable between 2 luminaries
“master luminaire” and between the ECG and the lamp in the “slave luminaire”. The following requirements apply to the physical arrangement of the two luminaires:
Max. length of the connecting cable between two luminaries [m]
QTi 2x No master-slave circuit possible QT-FH 2x14-35/230-240 CW 1 m QT-FQ 2x…CW 0.5 m QT-FH 2x…/230-240 F/CW and QT-FQ 2x…/230-240 F/CW
No mater slave circuit possible
The recommended cable lengths are maximum values which must be observed. For information about cable routing in the “slave luminaire” as well as in the “master luminaire” for maximum cable lengths please see the recommendation in Section 7.1. For maximum cable lengths please see the recommendation in Section
8.1.1ff.
3.7 PE-Connection for
Protection Class I Luminaires
max.
0.5 m
ECG with 6-pole terminal
Exposed metal parts of luminaries of protection class I must be reliably and permanently connected to a PE conductor. For all QT-FH... CW, QT­FQ...CW and QT…F/CW one ore both fastening screws are used for grounding. Serrated washers to improve the earth connection are recommended.
QUICKTRONIC
®
INTELLIGENT QTi ECGs have an additional ground
terminal to suppress radio interference (“functional” earth).
UN
UN
DA (–)
DA (–)
DA (+)
DA (+)
1
1
2
2
3
3
4
4
5
5
6
6
7
7
QTi 1x...
QTi 1x...
21
21
22
22
23
23
24
24
25
25
26
26
27
27
L
LL
Wiring 1-lamp QTi
38
p
p
UN
UN
DA (–)
DA (–) DA (+)
DA (+)
Wiring 2-lamp QTi
1
1 2
2 3
3 4
4 5
5 6
6 7
7
QTi 2x...
QTi 2x...
21
21 22
22 23
23 24
24 25
25 26
26 27
27
L
LL
L
LL
To obtain good radio interference suppression, the PE conductor and the line mains cables should not be lais parallel to the lamp cables or alongside the ECG.
3.8 Functional Earth for
Luminaires of Protection Class II
Radio interference: 9 kHz to 300 MHz
9 kHz to 300 MHz
Metallic plate / metal reflector/ metal grid
Lampe
Lampe
Lamp
Low
Low
EVG
EVG
ECG
L N
L N
R
R
Due to the earth connection of the ECG to a metal plate ore the luminaire body, an “internal” short circuit exists in luminaires of protection class I. It means that the interference and leakage currents are redirected into the ECG and, therefore, no so-called interference voltage are generated when measuring conducted disturbances according to CISPR 15. The emitted interference level of such a system is “low”.
Radio interference: 9 kHz to 300 MHz
High
L N
R
Lam
Lam
EVG
ECG
Plasics housing
In this case, the ECG is mounted in a plastic housing. Therefore, the capacitive leakage currents generated by the system lamp and ECG are not circulating within the luminaire, they are fed back into the mains supply via the luminaire environment. The magnitude of these leakage currents depends very much on the mechanical design of a luminaire and its surroundings and on the specific features of the various lamp types
®
(FH
…HE or FQ®…HO) and hence can be very different. They cause interferences on the mains side with regards to measurements of conducted disturbances according to CISPR 15.
39
3.8.1 General Information
The functional earth in this arrangement restores the internal short circuit of the system. Î no interaction of the capacitive currents of the system lamp­ECG with the mains side and therefore no disturbances with regards to CISPR 15 measurements.
3.8.2 Practical Details
Radio interference: 9 kHz to 300 MHz
Lamp
Lamp
Low
ECG
ECG
L N
L N
R
R
Plastics housing
Due to the partly high lamp voltages of T5 fluorescent lamps we recommend to apply a functional earth to T5-lamps for improved radio interference suppression.
Different Electronic Control Gear have to be connected to functional earth for protection class II luminaires (potential equalisation): e.g. QUICKTRONIC There is no coherence between the functional earth and the ground wire. Thus the functional earth can be connected in luminaires of protection class II.
Attention should be paid to:
1.) Luminaire
1.1) Construction
ECG housing and wire of functional earth have to be observed as active parts.
- Requirement regarding double or boosted Insulation against metal parts or the luminaire surface have to be kept.
- the wire for the functional earth connection must not be marked in yellow/green
®
INTELLIGENT QTi)
1.2) Terminal Labelling
Acc. to EN 60445/VDE0197 the functional earth connection has to be marked with the letters „FE“ or the below label. See label in Section
In any case the symbol of the protective earth
3.7.
must not be used for a
40
3.9 Temperature Ranges
3.9.1 Self heating ECG
functional earthing!
2.) Instruction manual
The necessity of the functional earthing of an ECG due to EMI reasons is noted in the technical data or in addition on the type label of the ECG with the appropriate symbol. The regulations for the skinning and the max. length of the wire are valid for L- and N-conductor. The length of the functional earth must not exceed the max. length of the other wires. IEC 60363/VDE V0800-2-548 have to be observed for the functional earthing.
3.) Manufacturing
As the functional earthing isn’t a safety measure of the luminaire, the
test of the ground wire connections can be omitted.
To guarantee a reliable operation temperature ranges indicated on the housing (ambient temperature of the ECG) as measurement temperature tc have to be observed. In general, the lower the operating temperature of an ECG the higher its service life. (See Section
The temperature must be assessed separately for the two system components (ECG and lamp). In the case of the lamp, there are physical laws that restrict the temperature range, whereas in the case of the ECG fixed limits must be set in order to ensure reliable operation. Apart from this, there are external factors such as the reciprocal influences of ECG, lamp and luminaire and the selected installation site which all of which have an influence. Compliance with the specified limits and hence the guarantee of the operational reliability are the responsibility of the relevant luminaire or system manufacturer.
Operation of electronic control gear outside the recommended temperature ranges can result in the following behaviour:
ECG ambient temperatures are too low:
Fluorescent lamps cannot be started reliably. At too low temperatures some electronic components may not be functional or at least limited in their operation.
ECG ambient temperatures are too high:
ECG life shortened or ECG even damaged. Î High ECG failure rate.
Important:
These limit temperatures apply even if the units are not in operation or are in storage. Typical temperature values for storage of ECG are: Storage temperature: -40 °C to max. +80 °C Humidity: 5 % to max. 85 %
Because of their low losses, OSRAM QUICKTRONIC® ECG for operation
2.10, Reliability of ECG)
41
3.9.2 Control Gaer Temperatures
3.9.2.1 Measuring Point
Temperature tc
of T5/ 16 mm- fluorescent lamps FH
®
...HE and FQ®...FO have a very low thermal output, producing typically between 10 °C and 20 °C temperature rise. It allows a wide range of ambient temperatures which is covering almost all applications. If not, adequate measures must be taken in the luminaire or at the site of installation to improve the thermal balance of the luminaire.
If the limit temperature is expected to be exceeded for only short periods (less than one hour, as may be the case in outdoor installations in direct sunlight), but most of the entire time the operating temperature is below the maximum recommended value (at night-time, for example), a certain balance between reduction and extension of service life may be expected. However, there is no guarantee from our side that this will be the case.
The temperature at the tc point must never be exceeded by more than 10 °C, otherwise the unit is very likely to suffer permanent damage.
ECGs may also suffer permanent damage if they are operated below the specified minimum temperature. As already mentioned, if the lamps are also too cold, there will be problems with starting, low luminous flux and a shift towards the red end of the spectrum.
For installing ECGs in luminaires, the measuring point temperature tc on the case is of major importance in any thermal analysis. The maximum recommended value for the unit and marked on the housing must not be exceeded in order to reach the service life as specified in the data sheet.
To obtain a safety approval according to EN 60598 for a luminaire, this temperature limit my be exceeded by up to 5 °C. However, this application has a reducing effect on the ECG service life.
According to EN 60598, tc (c stands for case) is the maximum recommended temperature that may occur at an indicated point on the ECG (Tc measuring point) during normal operation at rated voltage and within the specified temperature range.
In practice, the temperature rise of the housing results from the self-heating of the unit, which in turn results from the power loss and the ambient temperature of the ECG. This is influenced by the position of the lamp and the design of the luminaire and is consequently always higher than the ambient temperature of the luminaire.
Exceeding the maximum recommended tc temperature by a few degrees drastically reduces the expected service life of the unit. If the temperature is exceeded by more than 10 °C, a 50 % reduction in service life can be expected. At 20 °C or more above the maximum recommended temperature the unit is likely to fail very quickly. The limit temperatures of various electronic components ,such as transistors, are primarily responsible for this.
42
3.9.2.2 Ambient Temperature
C
C
C
C
C
C
C
C
C
C
C
C
C
C
C5°C
C0°C
C
C
C
C
C
C5°C
C
C
ECG : ta
3.9.3 Lamp Temperature
If, however, the temperature at the tc point is permanently 10 °C or more below the maximum, the expected service life of the unit will be approximately doubled.
According to EN 60598-1, ta (a stands for ambient) is the maximum value of the steady state temperature at which, during normal operation, limit temperature tc is not exceeded at the measuring point.
Also according to EN 60598-1, there are precisely defined testing and measuring requirements for both surface mounted luminaires (pendant and portable, for example floor lights) and recessed luminaries.
The maximum values specified in the lamp documentation for cold-spot temperature (see Technical Specification T5 Fluorescent Lamps) are important operating criteria for the lamp and must not be exceeded nor fallen below under any circumstances in order to achieve optimum luminous flux. The ambient temperature to reach the maximum lumen output is 35°C for T5/ 16 mm fluorescent lamps FH and 25 °C for FC
®
- circular fluorescent lamps
®
..HE and FQ®...HO,
3.9.3.1 Maximum Luminous Flux for T5/
16 mm-
Fluorescent Lamps
Placement of th filaments for T8 lamps
Placement of the filaments for T5 lamps
Further lamp and ECG should not heat each other. ECG should be dissipated through good thermal connection between ECG and luminaire housing. Usually, T5-fluorescent lamps reach their standard electric properties (rated values) which are used for lighting design purposes at ambient temperatures of 25 °C. Their maximum light outputs, however, are achieved at ambient temperatures of 35 °C.
T5 FH28W (nearly constant power supply): Lumin ous flux / Voltage - Horseshoe
100%
COLD SPOT temperature
90%
80%
70%
60%
50%
40%
relative light output
30%
20%
-25°C
10%
0°C
0%
55% 60% 65% 70% 75% 80% 85% 90% 95% 100%
70°C
75°C
80°C
85°C
90°C
75°C
-5°C
15°C
10°C
5°C
60°C
65°C
45°C
55°C
ambient temperature
5°C
25°C
20°C
relative lamp voltage
50°C 45°C
55°C
35°C
30°C
100%
25°C
40°C
35°C
relative light output
Luminous flux FH®…HE and FQ®…HO fluorescent lamps related to the lamp ambient temperature
T5 FQ54W (nearl y const ant power supply) : Luminous flux / Volt age - Horseshoe
COLD SPOT t emper atur e
90%
80%
70%
60%
50%
40%
30%
20%
-25°
10%
0%
-5°
60% 65% 70% 75% 80% 85% 90% 95% 100%
75°
80°
85°
78°
90°
-5°
15°
10°
70°
20°
65°
55°
relative lamp volt age
60°
45°
ambient temperature
25°
50°C45°
55°
35°
25°
35°
30°
40°
43
3.9.4 General
Recommendations for Installation
3.9.5 Measuring the
Temperature
At significantly lower or higher cold-spot temperatures than the specified temperature the electrical properties of the lamps change drastically and there is a significant reduction in luminous flux.
In normal cases of significant deviations, the shutdown mechanism in the ECG will operate. In extreme cases there may be damage to the electronic control gear.
If the lamp temperature is too low it may be difficult to start and the luminous flux may be too low. Selecting a different site for installation generally helps, or using some kind of outer tube to conserve the heat of the lamp. It is important that this outer tube is installed on the etched lamp side around the electrode (cold spot). In any case, we recommend that luminaire manufacturer informs the electrician by printing the information on the inside of the luminaire.
If the ambient temperatures are too high the ECG can be damaged and the light output is too low. An optimized cooling is necessary. In order to avoid a thermal interference when operating a multi-lamp T5­system, it is required to install the lamps always with the lamp etch on one side.
It is important to ensure that the lamp and the ECG are positioned in the luminaire so that they do not mutually heat one another and that the ECG power loss can be properly dissipated even at the maximum expected ambient temperature and/or supply voltage.
The tc-temperature at the measuring point on the ECG must not be exceeded during operation even at the maximum expected ambient temperature and/or supply voltage. Under “normal” ambient conditions the tc-temperature measured at the measuring point should be at least 5 °C to 10 °C below the specified maximum value so there is a safety margin to allow for extreme situations.
It may be necessary to split lamp and ECG (with, say, the lamp in the luminaire and the ECG in the stand or luminaire support) such that the absence of special measures the lamp and the ECG would not mutually heat each other if arranged in close proximity, leading to excessive temperatures of the lamps and/or the ECG. In such arrangements ensure that the maximum cable length between ECG and lamp(s) is not exceeded and the wiring instructions under Section
The simplest way to measure the relevant temperatures on the lamp (especially at the cold-spot) and on the ECG (tc point) is with thermocouples fixed to the lamp/ECG and a suitable measuring instrument. Make sure the adhesive used is neutral in terms of its thermal, electrical and photometric properties.
To measure the ECG temperature it is convenient to have a thermocouple permanently attached to a housing cover and to exchange this for the original cover.
The temperature values should only be measured when the steady-stae temperature has been reached (in other words, when there has been no
4.1 and 7.1 are followed.
44
3.10 Luminaire Wiring Test
for Two-lamp Luminaires
significant change in temperature for some time). The supply voltage should be held constant at least throughout the entire measuring cycle at the rated voltage of the lamp.
The
following procedure is recommended for the thermal analysis of the
luminaire, taking into account the design requirements specified in EN 60598-1:
1. Thermal situation in the luminaire without contol gear heat.
Luminaire in measurement setup according to EN 60598-1 in standard mounting position, equipped with ECG and lamp and fitted with thermocouples. The lamp is supplied from external control gear, and not from the built-in ballast. In this way the temperature rise in the entire set-up resulting only from the lamp can be measured and the thermal “link” to the environment can be optimized.
Thermal situation in the luminaire with contol gear heat.
2.
Arrangement as described in 1., but the lamp is supplied from internal control gear. By comparison with the measured values obtained already, the additional heat generated by the ECG can now be assessed.
lamp terminal
lamp terminal
N
N L
L
Earth connection via ECG- housing
Earth connection via ECG- housing
QUICKTRONIC
QUICKTRONIC
®
®
1
1 2
2 3
3 4
4
output
output socket 1
socket 1
output
output socket 2
socket 2
3.10.1 Testing with a Test
Adapter and Dummy Lamps
¾Insulation test: 500 VDC acc. to EN 60598 app. 3...5 sec.; R = 2M
¾Insulation test: 500 VDC acc. to EN 60598 app. 3...5 sec.; R = 2M
¾High voltage test (100% for each luminaire)
¾High voltage test (100% for each luminaire)
A more precise wiring test for two-lamp luminaries than the one described in Section 3.10.1 can be performed with a test adapter (own design with the resistors shown in the diagram) and a sample tube (dummy lamp with sockets measuring the resistance). This test can be used for two-lamp luminaires equipped with the before mentioned ECGs.
45
The test is performed on the wired luminaire without mains voltage and without lamps.
46
1. The test adapter is inserted behind the release lever of the 45°-
A
plug-in terminals or, in the case of combi-wiring terminals, in the respective non-wired IDC or horizontal plug-in contact.
2. Instead of the lamps the two dummy lamps are inserted in the lampholders of the empty luminaire.
Dummy 1
Dummy 2
B
E F
Adapter
C D
G H
3. Measure the resistance between A and B and between C and D. The resistance between A and B and between C and D should be 100 .
4. Measure the resistance between E and F and between G and H in the same way.
If the measured resistance is not 100 , the wiring is incorrect.
47
3.11 ECG Operation for Luminaires of Protection Classes I and II
3.12 Insulation Distances in Luminaires
3.13 Insulation Test
In accordance with EN 60598, luminaires are grouped into protection classes according to the measures taken against contact with high voltages.
In the case of protection class I luminaires, all accessible parts which may become live as a result of a fault must have a good conductive connection to the PE conductor. The conductive link between the luminaire and the ECG must not be provided by the PE conductor of the ECG but by appropriate mechanical design features (such as using serrated edge washers or serrated head bolts).
In the case of protection class II luminaires, live parts must be provided with reinforced or double insulation. Protection class II luminaires do not therefore have an earth connection (except protection class II luminaries with functional earth).
ECGs only approved for installation in luminaires are not assigned to a protection class since protection classes are defined only for end products (such as luminaires) and not for components.
All QUICKTRONIC
®
FH
...HE and FQ®...FO are, in principle, suitable for operation in protection
®
units to operate T5/ 16 mm-fluorescent lamps
class I and II, unless otherwise indicated in the latest edition of the Lighting Programme or under www.osram.com/ecg However, radio interference suppression and temperatures must be checked in each case.
As a general rule, we can say that the thermal properties in open metallic luminaires (typically protection class I luminaires) are normally better than in enclosed plastic luminaires (typically protection class II luminaires) because of the good thermal conductivity of metal (heat sink effect) and better convection possibilities in the luminaire.
The use of luminaires is subject of a series of regulations governing electrical safety (shock protection) and operational reliability in wet, dusty, corrosive, flammable and explosive conditions. European standard EN 60598 applies to the electrical safety of luminaires.
To guarantee the electrical safety of luminaires, special attention must be paid to clearances and tracking distances. These terms are defined as follows in EN 60598-1-11 for the mains terminal of the luminaire: „Tracking distances at the mains terminal shall be measured between the active parts in the terminals an any exposed metal part. Clearance shall be measured between the incoming mains cable and exposed metal parts (i.e. from the bare end from which the insulation has been stripped the furthest to the metal part that is exposed). On the side of the terminal to which the internal wires are connected, the clearance shall be measured between the active parts and the exposed metal parts.”
For further information please refer to luminaire standard EN 60598.
Luminaires must be subjected to insulation and high voltage testing (according to EN 60598, VDE 0711, PM 395). Proceed as follows:
Input and output terminals of the ECG – except the PE conductor
terminal – must be connected conductively with one another.
48
3.13.1 Dielectric Resistance
in Lighting Systems
3.13.2 Mesuring the
Dielectric Resistance between N and PE or L and PE
Conduct insulation test at 500 VDC; the leakage current should not
exceed 0.25 mA.
Carry out high voltage test at 1.5 kV AC/50 Hz. This voltage must
be maintained for 1 s without flash-over (i.e. leakage current < 10 mA).
The following are recommended alternatives for the luminaire manufacturers (PM 333, PM 395):
100 % high voltage testing (insulation testing may be omitted) or
100 % insulation testing and 1-2% high voltage testing or
testing by agreement with the testing authority (such as VDE,
KEMA, SEMKO)
Dielectric resistance in lighting systems (> 0.5 M) can be measured in accordance with DIN VDE 0100 Part 600 Section 9 between:
a) The outer conductors (L1, L2, L3) and the protective
earth (PE) conductor
b) The neutral conductor (N) and the protective earth (PE)
conductor c) The outer conductors (L1, L2, L3) among themselves d) The outer conductors (L1, L2, L3) and the neutral
conductor (N)
The insulation test is performed at 500 VDC.
(See picture in Section 3.10) Tests are performed on new and existing installations. In existing installations the test may be carried out without isolating the luminaire approx. every two or three years. There must be no electrical connection between the neutral conductor (N) and the PE conductor. In this insulation measurement (500 VDC with respect to PE), the neutral conductor isolating terminal should only be disconnected if mains voltage is disconnected! Make sure the connection is secure before reapplying mains voltage. Failure to observe these instructions may lead to destruction of all the ECGs in the system due to an unbalanced load and a resultant overvoltage. Recommendation: 500 VDC = max. 1 mA measurement current
Measurement procedure:
The ECG appears momentarily to have low resistance (charging of
the capacitors in the interference suppression filter).
The ECG then appears to have high resistance.
An insulation fault in the lamp circuit does not affect the ECG.
The ECG will not be damaged by insulation tests provided a maximum current of 1 mA is not exceeded (the measuring equipment must be designed as a current source with an internal resistance of 500 k).
Important:
Before using the lighting system, check for correct N conductor connections! While the lighting system is in operation, never interrupt the neutral conductor!
49
3.13.3 Three-Phase Operation
3.13.4 Resistance to Overvoltage for QUICKTRONIC T5/ 16 mm­Fluorescent Lamps
3.14 Inrush Current /
Automatic Circuit Breakers
®
for
Tthe proper connection of the N conductor is very important in an installation for even load distribution in a three-phase operation.
The following diagram shows both the correct and the faulty wiring and ist possible impact:
U
phase-Phase
= UNx 3
(z.B. 400V~)
UN* > U
N
U
N
(z.B. 230V~)
UN* > UN
Theoretical maximum value:
U
*max = UN x 3 (= 400 VAC @ UN = 230 VAC)
N
In practice:
* < 350 V in most cases
U
N
(no complete asymmetrical load distribution)
QUICKTRONIC
®
FH
...HE and FQ®...HO have the following resistance to overvoltage:
®
-ECG for operation of T5/ 16 mm- fluorescent lamps
QUICKTRONIC® INTELLIGENT, QTi
Resistance to overvoltage
350V Î continuous 400V Î 48 hours
QT-FH ... CW (30 mm height)
QT-FQ ... CW (30 mm height)
300V Î continuous 320V Î 48 hours 350V Î 2 hours
QT-FH ... F/CW (21 mm height) QT-FQ ... F/CW (21 mm height)
300V Î continuous 320V Î 48 hours 350V Î 2 hours
When an ECG is switched on, a starting current pulse of very short duration (< 1 ms) occurs as the storage capacitors responsible for internal power supply charge up. If a large number of ECGs are switched on simultaneously (particularly if they are switched on at peak rated voltage) a starting current will flow that will reduce the recommended number of ECGs per automatic circuit breaker below that which would apply if we were to consider only their rated currents. All switching equipment and protection devices must therefore be selected according to their current carrying capacity.
The values mentioned in Section 7.3 refer solely to the automatic circuit breakers type B from Siemens.
50
3.15 RCDs / Fault Currents
3.16 Leakage Current
3.17 ECGs in Three-Phase
Operation
In the case of ECGs with protective earth (PE) connections, both the high short duration starting current and the small continuous current through the interference suppression capacitors in the ECG can trip the residual current detector (RCD).
The following solutions may be considered:
Devide the luminaires into three phases and use three-phase RCDs
Use surge-current-resistant, short-delay
Use 30 mA RCDs (if possible)
In Section
7.3 are the values for QUICKTRONIC
T5/ 16 mm- fluorescent lamps FH
®
...HE and FQ®...HO.
®
ECG for operation of
In protection class I luminaires, the internal HF filter in an ECG with PE conductor connection produces a 50 Hz leakage current through the earth conductor whose value depends on the product type.
This 50 Hz leakage current limits the number of ECGs that can be operated on an RCD.
For all QUICKTRONIC
®
FH
...HE and FQ®...HO the following applies:
®
units to operate T5/ 16 mm- fluorescent lamps
Leakage current < 0.5 mA
When using electronic control gear in a three-phase operation the following points have to be considered:
1. Check whether the mains voltage is within the application range of the ECG (DC/AC range from 198 V to 254 V).
2. The mains connection on the installation side may only be made to the luminaire terminal. For luminaries or luminaire groups in 3­phase circuits.
3. Make absolutely sure that the neutral conductor is correctly connected to all the ECG-luminaires and that it is making proper contact.
4. Cables may only be disconnected or connected when no voltage is present.
5. For 3x230/240 V supply networks in triangular circuit arrangements, protection by way of common disconnection of the phase conductor is necessary.
Important:
In new installations the load must not be connected when the insulation resistance is measured with 500 V DC, since according to VDE 0100 T600 Section 9 the test voltage is also applied between the neutral conductor (N) and all three external lines (L1, L2, L3. In existing installations it is sufficient to carry out an insulation test between the external lines (L1, L2, L3) and the protective earth without disconnecting the loads. The neutral conductor (N) and the protective earth (PE) may not be electrically connected in any way when this is done. For this insulation measurement (500 V DC to earth) the neutral conductor disconnection terminal may only be opened with the mains voltage switched off!
51
Before the equipment is put into operation, make sure that the N conductor is correctly connected!
During operation do not disconnect the N conductor on ist own or first!
Luminaires or luminaire groups can also be wired in 3-phase operation with a common N conductor (neutral conductor) as seen in the diagram
3.14.2
If the common neutral conductor is interrupted in a 3-phase star configuration and voltage is present, then luminaires or groups of luminaires operated with ECG may be exposed to unacceptably high voltages and the ECG itself may be destroyed.
See also Section QUICKTRONIC
®
FH
...HE and FQ®...HO.
®
3.14.3 for high voltage resistance of different
types to operate T5/ 16 mm- fluorescent lamps
52
4. Lamp Wiring
4.1 h = 21 mm
4.1.1 QUICKTRONIC® INTELLIGENT 1-lamp version
4.1.2 QUICKTRONIC INTELLIGENT 2-lamp version
®
4.1.3 QT-FH MULTIWATT F/CW
4.1.4 QT-FQ F/CW 1-lamp version
53
4.1.5 QT-FQ F/CW 2-lamp version
4.2 h= 30 mm
4.2.1 QT-FH MULTIWATT 1- and 2-lamp version
4.2.2 QT-FH 3- and 4-lamp version
54
4.2.3 QT-FQ 1-lamp version
4.2.4 QT-FQ 2-lamp version
General Information:
Technical data edition May 2005 are used for this edition. Generally the ECG imprint is valid. Technical data are subject to change without any notes.
55
5.1 Definition
INTELLIGENT
5.2 Lamp Detection as
Fundamental Advantage
5.3 QTi – the High-tech
ECG
5.4 QTi – Advantages
5. QUICKTRONIC® INTELLIGENT
Electronic Control Gear of OSRAM labelled with this sign are realised in µController technology. The ECG to operate T5/ 16 mm-fluorescent lamps in equal length detect the lamps and operate them with rated data.
Thanks to this different lamp wattages and types can be operated with one ECG type only.
In the past, T5/ 16 mm-fluorescent lamp systems have been devided in two different families, FQ With HO-fluorescent lamp systems high lumen packages (i.e. HO 80 W with up to 7.000 lm) can be realized. FH are not very powerful but extremely efficient (luminous efficacy up to 104 lm/W). Both lamp families have the same lamp lengths, however, different wattages (see Section 2.2) which, up to now, required always a dedicated ECG. This is now a thing of the past due to QTi and their possibility to combine T5 fluorescent lamps of equal length. These T5 lamp types can be operated with just one QTi:
®
FH
14W HE + FQ® 24W HO ( 549 mm)
®
FH
21W HE + FQ® 39W HO ( 849 mm)
®
FH
28W HE + FQ® 54W HO (1,149 mm)
®
FH
35W HE + FQ® 49 W HO + FQ® 80W HO (1,449 mm)
As QUICKTRONIC fluorescent lamps and operate them according to their optimal electric parameters, lamps reach their maximum lamp life. As a result, lamp replacement intervals can be significantly extended. Due to the minimal losses of QUICKTRONIC balance of T5 systems is optimised. The proven OSRAM cut-off technology (turning off the electrode preheating after the lamp started) contributes to the improved energy savings. For the first time, this is also possible for dimmable ECG QTi … DIM thanks to QUICKTRONIC technology.
QUICKTRONIC
®
INTELLIGENT QTi reduce not only the cost of
illumination, they also increase the productivity
Less number of fixture types in the production process
(for example only one basic type for 35/
Less complexity in manufacturing through:
- identical wiring for dimmable and non dimmable QTi
- identical case dimensions (1- or 2-lamp versions)
- cw-terminal (combi wiring) for mechanical and manual wiring
New design options for super flat T5 luminaires due to minimal
height of only 21 mm for QTi.
QUICKTRONIC
®
INTELLIGENT, QTi also reduce the cost through:
®
…HO and FH®…HE.
®
…HE fluorescent lamp systems
®
INTELLIGENT automatically detect T5/ 1 6mm
®
INTELLIGENT the energy
®
INTELLIGENT
49/ 80 Watt)
56
5.5 QTi – Practically
Applied
5.6 Technical Specialties
for non-dimmable QTi
5.6.1 Inrush current limitation
Less stock keeping of luminaires (approx. 50 % less types)
Change of lumen output possible at any time simply by replacing
the lamp
No wrong lamp types by mistake with effect on the lamp life
Lower number of luminaires to be stocked at the user’s facility
In industrial applications different minimum illumination levels have to be applied according to law. Up to now, this required separate luminaire types. From now on, this requirement can be fulfilled with only one basic luminaire type fitted with different lamps.
For example:
Workplace for technical drawing 1000 Lux, R
> 80 Î FQ® 80W HO
a
Workplace for mechanical
manufacturing 500 Lux, R
Factory traffic 300 Lux, R
> 80 Î FQ® 49W HO
a
> 80 Î FH® 35W HE
a
With the new family QUICKTRONIC
®
INTELLIGENT, QTi, OSRAM contributes to reduce the cost of our business partners in terms of logistics or to more flexibility in lighting installations.
Einschaltstrombegrenzung auf max. 1,5 A bei TH= 0,5 ms
16A-Automat
typischerweise
typischerweise
typischerweise
bisher
bisher
bisher
28 Stk.
28 Stk.
28 Stk.
1-lp.EVG
1-lp.EVG
1-lp.EVG
(15 Stk. 2-lp. EVG)
(15 Stk. 2-lp. EVG)
(15 Stk. 2-lp. EVG)
Charakteristik B
41 Stk.
41 Stk.
41 Stk.
QTi
QTi
QTi
nicht dimmbar
nicht dimmbar
nicht dimmbar
1x...
1x...
1x...
(28 Stk. 2x...)
(28 Stk. 2x...)
(28 Stk. 2x...)
The new micro-controller technology allows alternative circuits in the input section of an electronic control gear. Therefore it is now possible to equip non-dimmable versions with a limitation of the starting current.
By limiting the starting currents almost twice as many QTi units can be operated on a single automatic circuit breaker than compared to single­wattage ECGs.
This product feature reduces the wiring efforts in lighting installations.
See Section
7.3 for specific values of starting currents.
57
5.6.2 Resistance to Overvoltage up to 400V
5.6.3 Lamp-ECG­Combination
5.6.3.1 Straight Fluorescent
types
Usually, electronic control gear work with input voltages between 220 V and 240 V in a standard three-phase installation. If the contact of the neutral conductor is missing or faulty, this value can rise – depending on the load distribution – up to a maximum value of
2 * 230 V = 400 V
The resistance to overvoltage of non-dimmable QTi is 300 V for the duration of two hours. Depending on the load distribution in an installation normally values between 280 V and 400 V are achieved. To integrate an optical warning signal without additional stress of components QUICKTRONIC
®
INTELLIGENT units shut down the lamps at voltages greater than 290 V. At the input of the ECG the input voltage is measured in an interval of ms and the lamp is automatically restarted once the voltage drops below 290 V, thus avoiding to influence the lamp life.
Besides the combination of T5/16mm-fluorescent lamps of equal length additional fluorescent lamps can be operated in combination with QTi. All lamp combinations have an ENEC approval, which means that the fluorescent lamps are operated without loss of lumens compared to single­wattage types.
Combinations with FH, FQ, T8
QTi 1x14/24/21/39
QTi 1x14/24/21/39
QTi 1x28/54
QTi 1x28/54
QTi 1x35/49/80
QTi 1x35/49/80
QTi 2x14/24/21/39
QTi 2x14/24/21/39
QTi 2x28/54
QTi 2x28/54
QTi 2x35/49
Q Ti 2x35/49
FQ
FH
FH
HE
HE
FH 21
21 W
W
FH
FH 28
28
HE
HE
W
W
FH 14
14
HE
HE
W
W
FQ
FH
FH
24
24
35
35
HO
HO
HE
HE
W
W
W
W
FQ
FQ
FQ
FQ
49
39
49
39
HO
HO
HO
HO
W
W
W
W
HO
HO
FQ 54
54 W
W
FQ
L
L
80
80
HO
HO
18
18
W
W
W
W
L
L
L 30
30 W
W
L
L 58
58 W
W
L
L 70
70 W
W
L 36
36 W
W
FQ
FQ
58
5.6.3.2 Compact and Circular lamp types
5.6.4 Wiring
Combinations with DL, DF, FC
DF
DF
FC
FC 22
22 W
W
FC
FC 40
40 W
W
FC
FC 55
55 W
W
QTi 1x14/24/21/39
QTi 1x14/24/21/39 QTi 1x28/54
QTi 1x28/54 QTi 1x35/49/80
QTi 1x35/49/80 QTi 2x14/24/21/39
QTi 2x14/24/21/39 QTi 2x28/54
QTi 2x28/54 QTi 2x35/49
QTi 2x35/49
DL
DL 18
18 W
W
DL
DL 24
24 W
W
DL
DL 36
36 W
W
DL
DL 40
40 W
W
DL
DL 55
55 W
W
DL
DF
DF
DL
DF
80
18
80
18
W
W
W
W
24
24 W
W
DF 36
36 W
W
All single-lamp or two-lamp versions of QTi have identical wiring no matter if dimmable with DALI-interface, dimmable with 1-10 V-interface or non­dimmable.
UN
UN
UN
DA ()
DA (–)
DA (–)
DA (+)
DA (+)
DA (+)
1
1
1
2
2
2
3
3
3
4
4
4
5
5
5
6
6
6
7
7
7
QTi 1x...
QTi 1x...
QTi 1x...
21
21
21
22
22
22
23
23
23
24
24
24
25
25
25
26
26
26
27
27
27
L
L
LL
1
1
UN
UN
UN
DA ()
DA (–)
DA (–)
DA (+)
DA (+)
DA (+)
1
2
2
2
3
3
3
4
4
4
QTi 2x...
QTi 2x...
QTi 2x...
5
5
5
6
6
6
7
7
7
Cables which should kept short are always applied to the terminals with the highest numbers. For absolute cable lengths see Section
1-lamp version: #26 and 27 2-lamp version: #24, 25, 26 and 27
This product feature allows the pre-production of luminaires with cable harnesses.
21
21
21
22
22
22
23
23
23
24
24
24
25
25
25
26
26
26
27
27
27
L
L
LL
L
L
LL
7.2.
59
5.6.5 Dimensions
5.7 FAQ
Harmonised and identical dimensions for all 1-lamp and 2-lamp QTi is another product feature increasing the flexibility in luminaire design and production.
Dimensions of all 1-lamp versions (l x b x h) 360 x 30 x 21 mm
Dimensions of all 2-lamp versions (l x b x h) 423 x 30 x 21 mm
Can FH
together with a 2-lamp QTi?
No, because in this case both fluorescent lamps are not operated at their rated data. The ECG will not be damaged, however a significant difference in the luminous flux of both lamps can be seen.
Will the lamp detection be repeated with every restart of the ECG?
Yes, a the clear lamp detection with every restart is guaranteed. The fluorescent lamps start within 1 second.
Can QTi also be operated in emergency lighting installation at DC
voltage?
Yes, QTi are suitable for the operation with DC voltage.
Does the entire approval have to be done for all lamp wattages?
No, according to an international testing institute the entire approval must only be carried out for the system with the highest wattage. All other T5/ reduced testing procedure. However, we recommend to discuss these details with the according national approval authorities. .
®
…HE and FQ®…HO- fluorescent lamps be operated
16 mm-fluorescent lamp combinations are subject to a
60
6. Special Applications
6.1 Outdoor Application
6.1.1 Installation
Instructions
If you intend using electronic control gear for T5/ 16 mm- fluorescent lamps in outdoor luminaires, it is important to remember that depending on the design of the luminaire, it may be exposed to moisture and humidity.
The level of protection of the luminaire (IP ... according to DIN 40050/IEC
529) determines whether standard or special ECG have to be installed.
1) In luminaires of ingress protection type „5“ (protected against water jets: IP 65 for example) standard ECG can be used since dampness cannot penetrate this type of luminaire. So there is no risk of the ECG being corroded.
2) In the case of luminaires of protection type „3“ (protected against splash water, IP 43 for example), it is possible that water droplets will penetrate and cause corrosion. We therefore recommend using a protective housing (OUTKIT) over the ECGs for these luminaires. When using QUICKTRONIC OUTKIT (see Section proper strain relief of the ECG in the OUTKIT cases.
Typical applications include lighting systems for car washes, petrol stations, outdoor advertising, swimming poops etc. For these applications we recommend to use the ECG in combination with OUTKIT.
In addition to general installation and wiring instructions (Section 3) the following information is important:
The mains connection terminals on the ECGs should point
downwards. In other words, the control gear should be installed vertically or upside down with the base of the ECG uppermost, otherwise horizontally with a slight incline (5 to 10°). This will prevent condensate from collecting inside the unit and causing short-circuits on the printed circuit board and tripping the RCD as a result of leakage currents.
All ECG terminals that are not protected by vertical or upside down
installation should be covered by arched metal plates or plastics (better with regards to corrosion) so that spray water and condensate cannot drip into the terminals and therefore into the ECG.
Place ECGs on spacers to protect them from dripping
water/condensation.
To prevent water entering the control gear through the terminals
along the incoming or outgoing cables, it is best to bend the cables ahead of the terminals (to provide a water pocket or drip point). To ensure success, the lowest point of the kink should lie below the level of the terminal inlet.
A small opening at the lowest point on the luminaire is also
recommended so that condensation can escape. However, this opening should be protected so that rain and spray cannot enter.
®
of 21 mm height in combination with
6.1.2) special attention should be paid to the
61
6.1.2 OUTKIT
6.2 T5-ECG in Sound Studios
6.2.1 Noise and how to
avoid it
In summary, we can say that the ECG should be installed in such a way that spray, water drops and condensation cannot enter the ECG and that moisture and condenses inside the ECG can run out. An ECG can withstand condensation for short periods but long-term exposure to moisture should be avoided. The ECG must be operated for at least 30 minutes per day so that condensation inside the ECG can evaporate.
The luminaire casing should not be hermetically sealed. Instead, it should be ventilated so that the condensate that forms during the cool-down period as a result of the change in temperature (a luminaire is switched on, say, at –10 °C, warms up during operation to +30 °C and then cools down again to –10 °C after it has been switched off) is not trapped inside the unit and can evaporate safely.
Technical Data OUT KIT Short OUT KIT Long Voltage Range Temperature Range Lenth Height Distance mouting whole
198 V to 264 V 198 V to 264 V
-25 °C to 50 °C -25 °C to 50 °C 485 mm 550 mm
38 mm 38 mm
452 mm 517 mm
a Standard pack [pcs.]
20 20
OUT KIT Short for ECG with l = 360 mm OUT KIT Long for ECG with l = 423 mm Permitted cable diameter for use of standard cables:
- 2-3 mm for the material attached for 2,3 and 4 wires
- 2-2,7 mm for the material attached for 7 wires
If electronic control gear for T5/ 16 mm-fluorescent lamps are to be used in areas in which noise and electromagnetic interference (see also
2.7
) are important factors there are special requirements that have to be
Section
met when installing the ECG and the luminaires.
Generally speaking, noise is generated in electronic circuits as a „hum“ (at 50 Hz or 100 Hz) or as higher frequency interference in inductances (chokes, transformers) and capacitors.
Compared with conventional control gear (chokes), electronic controlgear with their high-frequency mode of operation (in which inductance values are much lower) generate appreciably lower noise levels and are a problem only in highly sensitive environments such as sound studios for CD quality recordings. The fully electronic control gear for T5/
16 mm-fluorescent
lamps offered by OSRAM is much quieter than conventional ballasts.
T5/ 16 mm-fluorescent lamps can only be operated by electronic
62
6.2.2 Recommended minimum distance between lamp and refelctor
control gear. Therefore, a direct comparison with conventional control gear
(chokes) does not make sense.
In T5 luminaires, the electronic control gear functions as a vibration source and is capable of exciting adjacent metallic or plastic components so that they act as resonators, amplifying the actual noise considerably and help to spread it out. To avoid this, especially the recommended minimum distance between two
16 mm- fluorescent lamps has to be respected so that no noise can
T5/ be generated due to physical contact.
32 mm
1. For high luminaire efficiency the distance should be at least 48 mm between the two lamps (approximately two fingers should fit between the lamps)
2. When designing the luminaire for minimum distance the maximum recommended base temperature of 120 °C may never be exceeded.
The maximum recommended base temperature of 120 °C may never be exceeded. A distance of less than 3 mm between lamp and reflector can result in generation of noise even in luminaires with non-dimmable ECGs. At a distance of less than 6 mm between lamp and reflector the leakage current of the dimmed lamps 35 W, 49 W and 80 W causes visible differences in brightness between the ends of the lamps. In addition, the radio interference suppression is getting worse.
OSRAM recommendation: In general, OSRAM recommends to keep a minimum distance of between lamp and reflector. In certain cases (unfavourable capacitive interference) the minimum distance should be even greater.
To develop luminaires that are as quiet as possible, it is therefore essential to insulate the control gear and chassis or luminaire support. In other words, there should be clearance below the ECG with the ECG mounted on point supports on the luminaire chassis or on rubber absorbers as used for conventional ballasts. Under certain circumstances, this type of mounting may, however, lead to thermal problems since the best way to dissipate the power loss to the environment is to have the ECG in full contact with the chassis.
6 mm
63
6.3 Treatment Rooms, Operating Rooms
6.3.1 Electromagnetic
Interference
Solving this problem with an appropriate housing design and/or type of installation for the luminaire (forced cooling, increased convection effect) has a further advantage in reducing the interference noise level and should therefore seriously considered.
Experiments have shown that the amount of noise generated is closely linked to the operating temperature of the electronic control gear. This is a particularly important factor if the unit has been installed in accordance with the recommendations given above. In extreme cases, it will not be possible to work without additional heat sink.
In addition, the noise level increases disproportionately as the temperature of the ECG rises. It is therefore best to operate the control gear at a temperature below the maximum recommended value. In practice, this means that the amount of noise generated is less, the lower the measuring point temperature tc. A combination of acoustic insulation of the ECG and reduced operating temperature represents the best technical solution.
In general, the following applies:
Electronic control gear for T5/ 16 mm-fluorescent lamps FH®...HE and
®
...HO are so quiet that even in very quiet surroundings they cannot be
FQ discerned by the human ear. They are therefore ideal for sound-sensitive areas such as radio studios with CD quality recordings. If necessary, random samples should be used to determine whether, given the local parameters (volume of the studio, reverberation time and number of ECGs), insulation as described above is needed from an acoustic point of view or whether standard products could be used.
In rooms used for medical treatment electrodes may be placed on a patient’s body to obtain electro-cardiogram or electro-encephalogram recordings. To eliminate interference from magnetic fields, DIN VDE 0107 defines the maximum recommended inductance strengths. Luminaires
®
fitted with QUICKTRONIC
control gear easily fall within this limit values at
distances of 0.75 m and greater.
Because of their magnetic field strengths, conventional control gear that cannot be used for T5/
16 mm-fluorescent lamps but for T8/∅ 26 mm
tubes are often not suitable and have to be placed at least 3 m away.
Fluorescent lamps are not point light sources and cannot be adequately focused, which means they are not considered suitable to light operating tables. Dichroic halogen lamps are used almost exclusively. But even the room lighting has to meet very stringent requirements in terms of radiated magnetic fields. Sensitive patient monitoring systems in the operating room and intensive care wards must not be exposed to leakage magnetic fields.
It is necessary to comply with the maximum recommended interference levels and minimum installation distances for luminaires as defined in VDE 0107/6.81. Important information on this subject can be found in
3.2
, electromagnetic compatibility.
Section
Whereas conventional control gear had to be installed separately from the luminaire in a central switch cabinet far enough away from the treatment area, ECGs for T5/
16 mm fluorescent lamps can, in most cases, be
installed directly in the luminaire without any problem. The actual
64
6.3.2 Interference from
Infrared Transmission Equipment
6.4 Electronic Tagging
6.5 Emergency Lighting
interference levels generated by luminaries fitted with ECGs are generally lower than those generated by the connecting cables between the luminaire and the choke for a conventional separate T8/
26 mm arrangement.
The electrical safety requirements correspond to those for installations in humid locations. In other word, protection class II luminaires should be used. For precise information on the minimum level of protection for the luminaire see DIN 40050/IEC 529.
Because of their low field strengths ECGs are unlikely to affect electronic equipment. There has been no known incidence of a heart pacemaker being affected.
Fluorescent lamps emit energy in wavelength bands that are also used for infrared transmission. These emissions cannot be changed at the lamp. Since IR receivers are often not selective and operate with wide wavebands, the IR equipment may be triggered inadvertently if light from the lighting system enters the receiver. The light emitted from the fluorescent lamp is modulated at twice the operating frequency ( 40 kHz to 120 kHz). Interference may occur if the useful signal also operates in this frequency range.
Interference is likely in cases in which the useful signal falls in the frequency range of the light emitted from the fluorescent lamp. Operating at higher frequencies ( 400 to 1500 kHz) or using optical filters in front of the infrared receivers (absorption filters) may help. Shielding the infrared receiver from direct light (with a tube, for example) may also help.
The carried signal for sound transmission used to be around 95 kHz or higher, which meant that the 3
rd
, 5th and 7th harmonics of the ECG operating frequency ranges (30 to 60 kHz in normal operation and up to 120 kHz in dimmer mode) led to considerable interference in transmission. Headphone manufacturers shifted to higher frequencies, such as 2,3 MHz and 2,8 MHz, to remedy the problem.
For simultaneous interpreting systems, which also operate in the 95 to 250 kHz frequency range, we recommend not to use the first six transmission channels, particularly channel 1, because these are affected by the harmonics of the basic ECG frequencies.
Many department stores and shops now use electronic tagging systems to protect their merchandise (such as clothes, CDs, drugstore articles etc.) against theft. These systems typically operate with resonance frequencies in the kHz-range.
In certain circumstances these systems may malfunction if the operating frequency is between 30 kHz and 130 kHz. It may be possible to eliminate the problem by increasing the distance between the luminaire and the transmitter/receiver.
T5-luminaires with QUICKTRONIC® from OSRAM can be operated on either AC or DC voltage. This means that the same luminaries can be used for both general lighting and emergency lighting in an easy and cost­effective way. Especially safety lighting from installations with high illuminances (i.e. in workplaces with high risks) can be realised
65
economically due to the high luminous efficacy of T5/
A
g
g
lamps operated with QUICKTRONIC
®
control gear.
16 mm-fluorescent
The following electro-technical regulations apply to emergency and safety lighting systems installed in Germany:
VDE 0100
Regulations governing the installation of power systems with rated voltages up to 1000 V
VDE 0107
Installing and testing electrical installations in medical rooms
VDE 0108
Installing and operating power systems in buildings for gatherings of people and safety lighting at work
VDE 0165 VDE 0510
Installing electrical systems in hazardous areas Regulations governing accumulators and battery systems
In addition, there are various lighting standards that have to be observed. In view of the number involved, we have selected just a couple by way of example. The full list can be obtained from Beuth verlag (Publishers) in Berlin.
Lighting requirements for emergency lighting are in included in EN 1838. The classification of EN 1838 is in safety lighting and alternative lighting whereas particular importance is in safety lighting.
Emergency lighting acc. to EN 1838
Safety Lightin
Emergency Lighting
lternative Lightin
Safety Lighting for escape route
Antipanic Lighting
Safety Lighting for workplaces with special danger
Beside the illuminance (lx) the switch-on delay is a very important criterion that covers ECG (time for ignition) and lamp (starting behaviour).
66
6.5.1 Different criteria for lighting
6.5.1.1 Switch-over time for
QTi – h=21 mm
6.5.1.2 Switch-over time for
QT-FH…CW – h=30 mm
6.5.1.3 Switch-over time for
QT-FQ…CW – h=30 mm
6.5.1.4 Switch-over time for
QT-…F/CW – h=21 mm
Illuminance E
Safety lighting for escape route
min
1 lx > 0,5 lx
Antipanic Lighting
horizontal on the ground
Safety Lighting for workplaces with special danger 10 % of the maintenance value of illuminance
absolutely: 15 lx
Switch-on delay
50 % of the required illuminance within 5 s 100 % within 60 s
50 % of the required illuminance within 5 s 100 % within 60 s
0,5 s immediately specified value (10 %) has to be achieved
DIN VDE 0108 includes further information in addition to EN 1838.
In addition to the switch-over times of Electronic Control Gear to operate
16 mm fluorescent lamps , mentioned within the next sections, the
T5/ switch-over time of the relays have to be considered. These data are available in the technical relay data.
Lampstart Ignition time for a) Cold lamp (Stand-by mode) b) warm lamp (i.e.short term
Preheat
< 1 sec.
< 0.5 sec.
interruption of voltage)
Lampstart Ignition time for a) Cold lamp (Stand-by mode) b) warm lamp (i.e.short term
Preheat
< 2 sec.
< 0.5 sec.
interruption of voltage)
Lampstart Ignition time for a) Cold lamp (Stand-by mode) b) warm lamp (i.e.short term
Preheat
< 0.5 sec.
< 0.5 sec.
interruption of voltage)
Lampstart Ignition time for a) Cold lamp (Stand-by mode) b) warm lamp (i.e.short term
Preheat
< 0.5 sec.
< 0.5 sec.
interruption of voltage)
67
6.5.2 Wiring diagrams for emergency lighting units
6.5.2.1 Wiring diagram
QT-FH 3x14 CW with emergency lighting component from BAG
Exemplary wiring diagrams for emergency lighting units
Subject to change without any notes
OSRAM cannot assume warranty for engineering change of the emergency lighting units.
LNPE
+
­7 6 5
L'
4
L
3
L
2
~
N
1
ECG
ELECTRONICS
ECG
ELC ELC
ELC - E
BAG
15 14 13 12 11 10
9 8
6.5.2.2 Wiring diagram
QT-FH 4x14 CW with emergency lighting component from BAG
LNPE
L
N
QT - FH 3x14 CW
3 4 7 8
+
­7 6 5
ECG
L'
4
ELECTRONICS
ECG
L
3
ELC
L
2
~
ELC
N
1
ELC - E
BAG
connected via ELC-E
1 2 5 6
15 14 13 12 11 10
9
8
3
4
7 8
1 2 5
6
L´ N
QT - FH 4x14 CW
3 4 7 8
1 2 5 6 9
10
connected via ELC-E
3
4
7 8
1
2
5
6
68
6.5.2.3 Wiring diagram QT-FH 3x14 CW with emergency lighting component from OMNITRONIX
6.5.2.4 Wiring diagram QT-FH 4x14 CW with emergency lighting component from OMNITRONIX
L N
3 4 7 8
QT-FH 3x14 CW
OMNITRONIX
1
2
5
6
connected via MCME
3 4
7 8
1 2
5
6
connected via MCME
L
N
3
4
7
8
QT-FH 4x14 CW
1
2
5
6
9
10
3
4 7
8
1 2
5
6 9
10
6.6 DC supply
OMNITRONIX
Luminaires for emergency lighting are switched to battery supply only in the event of a power failure. In mains operation, the luminaires are powered by the normal supply. The mains and emergency lighting switchover arrangement must reliably separate mains operation from emergency lighting operation; it must be a break-before-make arrangement.
A deep discharge protection system must be provided for battery systems. This effectively prevents the batteries discharging too much and suffering
69
6.7 Portable Luminaires
_
V
damage as a result and also prevents possible damage to the electronic control gear.
General
Switchover from mains supply to emergency supply and vice versa must take place in a break-before-make arrangement (See
Section 6.5). In this
discrete switching sequence there is a period – the length of which depends on the design of the emergency monitoring system – in which current does not flow or at least the supply voltage falls considerably below its minimum recommended value. These switching times must comply with the limits already mentioned in DIN 5035.
In accordance with DIN VDE 0108, the battery units must be designed for rated operation of at least 3 hours. If the ECG is supplied with a rectified AC voltage, this voltage should have as small a residual ripple as possible. The AC voltage component must be less than 5 %.
If changeover units are used (emergency lighting fixtures with internal switching) which supply the lamp directly from an emergency supply and interrupt the system circuit between the ECG and the lamp, the following must be observed:
Changeover or disconnection of the lamps from the ECG to the
external unit must be on all terminals
When switching back from the external supply to ECG operation,
the lamp(s) must first be connected at all terminals to the ECG before the ECG is supplied with power again (for example by using a time delay relay), otherwise the shutdown mechanism in the ECG will operate.
Many of these emergency lighting units available on the market do
not comply with the normal operating conditions for the lamps and therefore damage them. In such cases, OSRAM cannot guarantee the life of the lamps.
Portable ECG luminaires of protection class I (i.e. cable and plug with earth terminals), require a fuse in both the L conductor and on the N conductors of the mains supply. If VDE approval documents for ECGs state “for permanently installed luminaires” then only the L side is fuse-protected. The N side must be fused in the luminaire if the ECG is to be used also suitable for portable luminaries. The additional fuse on the N side must be designed for mains voltage, suitable for the system input current and be of the “anti-surge” type.
®
All OSRAM QUICKTRONIC
ECG for T5/ 16 mm- fluorescent lamps are
equipped with two internal fused on the circuit board so that there is no need for additional measures as described above.
~
230­240
The units are generally labelled with L and N and not with the symbol „
“ for
AC voltage.
6.8 Mix-up of FH®- and
®
-Fluorescent
FQ
Except QUICKTRONIC® INTELLIGENT, QTi, all other ECGs for T5/
16 mm- fluorescent lamps can only operate the according FH
®
…HE-
70
Lamps
or FQ®…HO-lamp in one length.
If there is a mix-up of fluorescent lamps in the luminaire, this can cause problems.
Effects on the system:
The lamps will generally be started. However, there will be lamp blackening and early failure after a short period of time and (lamp life much less than 1000 hours). The ECG will not be damaged at any time.
We recommend to label the reflector of the T5-luminaire with a small sticker showing the exact lamp description.
71
7. Appendix
7.1 Overview of Maximum
Cable Lengths
7.1.1 QUICKTRONIC®
INTELLIGENT
7.1.2 QT-FH MULTI...CW
-30 mm height-
7.1.3 QT-FQ...CW
-30 mm height-
7.1.4 QT-FH MULTI…F/CW
-21 mm height-
The following tables give an overview of the maximum cable lengths of
®
QUICKTRONIC
®
FQ
...HO.
ECGs for T5 16 mm-fluorescent lamps FH®...HE and
The sequence of fixing the wires to the terminals goes from the upper right downwards.
ECG-type Sequence PIN
QTi 1x14/24/21/39 QTi 1x28/54 QTi 1x35/49/80 QTi 2x14/24/21/39 QTi 2x28/54 QTi 2x35/49
21-27 21-27 21-27 21-27 21-27 21-27
PIN
PIN
PIN
PIN
PIN
PIN
21
22
23
24
25
26
27
2 2 1 1 2 2 1 1 2 2 1 1 2 2 2 1 1 1 1 2 2 2 1 1 1 1
ECG-type Sequence PIN
QT-FH 1x14-35 CW QT-FH 2x14.35 CW
6-1 6-1
PIN 5 PIN 4 PIN 3 PIN 2 PIN
6
1
2 2 1 1 2 2 2 2 1 1
ECG-type Sequence PIN
QT-FH 1x14 QT-FH 1x21
1-4 1-4
PIN 2 PIN 3 PIN
1
4
2 2 1 1 2 2 1 1
ECG-type Sequence PIN 1 PIN 2 PIN 3 PIN 4 PIN 5 PIN 6 PIN 7 PIN
8
QT-FH 3x14 CW
1-6 re
3-8 li
1,5 1,5 1,5 1,5 1 1 1 1
ECG-type Sequence
QT-FH 4x14 CW
1-10 re 1,5 1,5 1,5 1,5 1,5 1,5 1 1 1 1
3-8 li
PIN 1 PIN 2 PIN 3 PIN 4 PIN 5 PIN 6 PIN 7 PIN 8 PIN 9 PIN
10
ECG-type Sequence PIN
QT-FQ 1x24 CW QT-FQ 1x39 CW QT-FQ 1x49 CW QT-FQ 1x54 CW QT-FQ 1x80 CW QT-FQ 2x24 CW QT-FQ 2x39 CW QT-FQ 2x49 CW QT-FQ 2x54 CW
ECG-type Seque
QT-FH 1x14-35 F/CW QT-FH 2x14-28 F/CW
6-1 6-1 6-1 6-1 6-1 6-1 6-1 6-1 6-1
nce
1-7 1-7
PIN 5 PIN 4 PIN 3 PIN 2 PIN
6
1
1 1 2 2 1 1 2 2 1 1 2 2 1 1 2 2 1 1 2 2 2 2 2 2 1 1 2 2 2 2 1 1 2 2 2 2 1 1 2 2 2 2 1 1
PIN 1 PIN 2 PIN 3 PIN 4 PIN 5 PIN 6 PIN
7
2 2 1 1 1 1 2 2 2 2
72
7.1.5 QT-FQ…F/CW
-21 mm height-
7.1.6 QT-FC
7.2 Terminal Types
7.3 Inrush Currents
ECG-type Seque
nce
QT-FQ 1x24-39 F/CW QT-FQ 1x54 F/CW QT-FQ 1x80 F/CW QT-FQ 2x24-39 F/CW QT-FQ 2x54 F/CW QT-FQ 2x80
ECG-type Seque
QT-FC 1x55/230-240 S
1-7 1-7 1-7 1-7 1-7 1-7
nce
1-4
PIN 1 PIN 2 PIN 3 PIN 4 PIN 5 PIN 6 PIN
7
2 2 1 1 2 2 1 1 2 2 1 1 1 1 2 2 2 2 1 1 2 2 2 2 1 1 2 2 2 2
PIN 1 PIN 2 PIN 3 PIN 4 PIN 5 PIN 6 PIN
7
2 2 1 1
WAGO 250 WAGO 251 WAGO 251-
mini QT-FH MULTI CW X QT-FQ … CW X QT-FH 1x14, 21 X QT-FH 3, 4x14 X QT-FH...F/CW X QT-FQ...F/CW X QTi X
Special features of particular terminals: see Section 3.1ff
QUICKTRONIC® INTELLIGENT
Ip[A] TH [µs] 10A 16A QTi 1x14/24/21/39 QTi 1x28/54 QTi 1x35/49/80
QTi 2x14/24/21/39 QTi 2x28/54 QTi 2x35/49 QT-FQ 2x80
1 155 26 41 1 155 26 41 1 155 26 41
1 200 19 31 1 200 19 31 1 200 19 31
60 230 5 9
Max. number of ECGs with automatic circuit
breakers
QUICKTRONIC® 21 mm height
Ip[A] TH [µs] 10A 16A QT-FH 1x14-35 F/CW QT-FH 2x14-28 F/CW
QT-FQ 1x24-39 F/CW QT-FQ 1x54 F/CW QT-FQ 1x80 F/CW QT-FQ 2x24-39 F/CW QT-FQ 2x54 F/CW QT-FQ 2x80 F/CW
17 155 25 41 20 210 17 28
17 155 25 41 27 170 17 28 27 170 17 28 27 170 17 28 32 210 10 17 39 260 8 14
Max. number of ECGs with automatic circuit
breakers
73
7.4 Lamp/ECG Combinations
7.5 OSRAM Installation Tips for T5-Systems
QUICKTRONIC® 30mm height
Max. number of ECGs
with automatic circuit
breakers
Ip[A] TH [µs] 10A 16A QT-FH 1x14-35 CW QT-FH 2x14-35 CW
QT-FQ 1x24 CW QT-FQ 1x39 CW QT-FQ 1x49 CW QT-FQ 1x54 CW QT-FQ 1x80 CW QT-FQ 2x24 CW QT-FQ 2x39 CW QT-FQ 2x49 CW QT-FQ 2x54 CW
QT-FH 1x14 QT-FH 1x21 QT-FH 3x14 CW QT-FH 4x14 CW
20 210 17 28 20 210 17 28
17 155 25 41 17 155 25 41 20 210 17 28 20 210 17 28 28 230 8 13 20 210 17 28 28 230 8 13 28 230 8 13 28 230 8 13
17 155 25 41 17 155 25 41 20 230 17 28 20 230 17 28
QUICKTRONIC for FC
®
Max. number of ECGs
with automatic circuit
breakers
Ip[A] TH [µs] 10A 16A QT-FC 1x55/230-240 S
28 230 8 13
Valid Lamp-ECG combinations are available in the Lighting Programme.
Edition January 2005 also available in
www.osram.com/ecg
Section 9 or under
Maximum lamp temperatures for T5/ 16 mm fluorescent lamps compared
26 mm-fluorescent lamps
to T8/
T8/L36W 120 45...55 100...120 T8/L58W 140 50...65 110...130 T5/FH28W 120 40...60 160 (200...250*) T5/FQ54W 120 50...70 160 (200...250*)
Max. temp. at lamp caps
°C °C °C
Lamp centre Electrode area
*) in FH - E C G - Luminaires
the lamp could be incorrectly used,
e
. g . FH35W with FQ80W
74
7.5.1 Recommended Minimum Distance between Lamp and Reflector
7.5.2 Recommended Minimum Distance between two T5/
16mm-
Fluorescent Lamps
The maximum recommended base temperature of 120 °C may never be exceeded (see A distance of less than 3 mm between lamp and reflector can result in generation of noise even in luminaires with non-dimmable ECGs. At a distance of less than 6 mm between lamp and reflector the leakage current of the dimmed lamps 35 W, 49 W and 80 W causes visible differences in brightness between the ends of the lamps. In addition, the radio interference suppression is getting worse.
OSRAM recommendation:
In general, OSRAM recommends to keep a between lamp and reflector.
In certain cases (unfavourable capacitive interference) the minimum distance should be even greater.
OSRAM recommendation:
Section 7.5).
minimum distance of 6 mm
1. For high luminaire efficiency the distance should be at least 48 mm between the two lamps (approximately two fingers should fit between the lamps)
2. When designing the luminaire for minimum distance the maximum recommended base temperature of 120 °C
never be exceeded.
may
32 mm
Exemptions:
Interference between dimmable ECG-lamp circuits in parallel configuration can cause flickering. Therefore, we recommend a minimum distance of 120 mm (from lamp axis to lamp axis) between two lamps in a dimmable two-lamp luminaire where two dimmable single-lamp ECGs are used. Or in other applications where multiple dimmable ECGs are used in parallel such as RGB-applications in illuminated ceilings.
If the wiring is done very carefully a minimum distance of 50 mm (between the two lamp axis) can be achieved.
75
7.5.3 Luminaire Optimisation
With the mentioned measuring principle the best correlation between ambient temperature and cold spot temperature of the T5-lamps is achieved.
Attach the thermocouples in the area of the lamp caps within a distance of 2mm from the glass tube
x
7.5.4 Maximum luminous flux for FH
®
…HE
fluorescent lamps
Measuring point with the best correlation between cold-spot temperature and ambient temperature
T5 FH28W (nearly constant power supply): Luminous flux / Voltage - Horseshoe
100%
COLD SPOT temperature
90%
80%
70%
60%
50%
40%
relative light output
30%
20%
-25°C
10%
0°C
0%
55% 60% 65% 70% 75% 80% 85% 90% 95% 100%
80°C
85°C
90°C
-5°C
10°C
5°C
75°C
75°C
70°C
15°C
65°C
55°C
20°C
60°C
45°C
ambient temperature
5°C
25°C
relative lamp voltage
Lamp etch
55°C
50°C 45°C
35°C
30°C
25°C
40°C
35°C
7.5.5 Verticalness Operation
So called Horseshoe Curve shows the correlation in between the cold spot temperature and the ambient temperature ta.
If luminares are operated in horizontal position the lamp stamp of the
®
FH
…HE und FQ®…HO fluorescent lamps have to be placed down.
Luminaires with circular FC
®
lamps have to have the lamp socket
positioned down.
76
8. Troubleshooting Tips
8.1 General 1) ECG in constant operation (24 hours)
Recommendation:
Installations with ECGs operating 24 hours a day should be switched of each day for a few minutes.
Reason
When a lamp comes to the end of its life there is an increase in lamp voltage causing an asymmetrical additional load in the ECG­circuit. When exceeding a certain value this additional load shuts down the ECG (EoL, T.2). When the ECG is switched on again all the relevant values of the lamps are checked and “critical” values can be detected.
2) Wiring of multiple ECG Recommendation:
Lamp cables from different ECGs should not be routed together.
Raeson:
Mutual interference may lead to problems with starting and/or normal operation (lamps may automatically disconnect)
8.2 Equipment Behaviour on Overvoltage
3) Coding of the lamp cables
To speed up troubleshooting in luminaires and luminaire systems it is extremely useful for the lamp cables to be uniquely coded (colour coded or numbered). This applies especially to two-lamp ECGs and even more so to 3- and 4-lamp ECGs since the large number of cables increases the risk of connecting incorrectly.
4) Terminal blocks (2, 3, 4, 6 and 7 positions)
These terminal blocks are used with the majority of OSRAM ECGs. Note that only single core cables with a cross-section of 0.75 mm² to 1.5 mm² can be inserted without pressing the release tab. Larger cross-sections are not recommended; the release tab has to be pressed down for smaller cross-sections.
Appearance of the lighting system:
The lamps are of different brightness (from time to time).
Depending on the ECG type, the internal shutdown circuit will operate at voltages above 280 V. In the event of a fault, the lamp generator will be disconnected. The mains input circuit and various other sub-circuits, however, continue to be supplied with this high voltage. If the mains voltage increases further there will be changes in the operating characteristics of the unit.
In most cases, this will lead to damage or destruction of the overvoltage­sensitive varistor/protective diode, which in turn will cause the unit fuse to trigger and damage to the unit.
If the system fault is corrected and the mains voltage has returned to the specified tolerance range without damage to the unit, the unit can continue to operate normally after an interruption in the mains supply and a recovery time of typically 20 seconds. If the mains voltage continues to be high, the unit will once again go into protection mode as soon as power flows to it. Over short periods of time, this state does not usually damage the unit.
77
If a QUICKTRONIC
®
-ECG for T5/ 16m m-fluorescent lamps is operated
over long periods on a supply voltage greater than 280 V it may fail as a result of overheating (with the exception of QUICKTRONIC
®
INTELLIGENT).
8.3 Equipment Behaviour on Under Voltage
Important:
Significant under voltage can cause ECG failure for all ECG in constant wattage circuitry. Due to the characteristics of arrangement the line current increases for decreasing voltage. Below the specified voltage range this causes an overload of the filter chokes. First the colour of the copper wire and board below are changing. For a longer overload both windings of one or both filter chokes fail due to melting or swelling in the windings.
Compact and straight fluorescent lamps go out below the specified value of the ECG or even don’t ignite if voltage is too low. The ECG will not be affected while the lamp is still burning. If the lamp is disabled at the end of its life (i.e. emitter paste is wasted), the safety shut-down of the ECG cannot work below the specified voltage range and a damage of the ECG is possible.
8.4 Application faults
8.4.1 Wiring faults on the
lamp side
All single-lamp QUICKTRONIC terminals:
If there is a wiring fault the lamp will not operate or will cold start. In particularly serious cases there is a risk that the ECG will be destroyed.
®
ECGs and units with 6-pole output
8.4.2 Short-to-ground at the
output of
®
QUICKTRONIC
ECG
8.4.3 Effects of moisture
8.4.4 Installing luminaires
in draughty locations
If there is a short-circuit-to earth at one or more of the connecting cables between the ECG and the lamp, the ECG will fail.
Damage to the ECG:
Melting and swelling of only one winding on one or both input filter
chokes
Breakdown of one or two rectifier diodes, irrespective of whether
they are discrete or integrated components.
There are, however, other causes of these major faults. Generally, as a consequence, consequential damage occurs at other components and this damage need not be unique in character.
Rust at the edges of metal housings may indicate the effects of moisture. If the printed circuit board and/or various components are tarnished, it can be assumed that there has been a serious moisture ingress. There will be short-circuits between neighbouring component connections or solder points with high potential difference, resulting in failure of the ECG. Symptoms of serious faults are indicated by a “tidermark” inside the housing.
Depending on the location of the fluorescent lamp in the draught, the lamp will cool down in certain areas of the bulb. This leads to local depletion of mercury and to a darkening of the lamp in these areas because there is no
78
mercury available to generate UV radiation. This can be remedied by installing a thermal tube in the luminaire. The
effect always or nearly always occurs in the area of the luminaire where the ECG is not located. This is due to the power loss of the ECG side of the lumiaire will always be slightly warmer for the same reason.
Caution: Mercury depletion leads to a reduction in lamp voltage and to an increase in discharge current. This may result in damage to the ECG or, in extreme cases to failure of the ECG.
8.5 Trouble Shooting
8.5.1 Lamp does not start Problem:
Lamp does not start (with two-lamp ECGs both lamps fail to start), no visible glow shortly after start-up. Same behaviour even after being off for 1 minute (internal reset) and restart.
Possible cause:
a) RSD or other protective device in the installation has operated
Check the wiring on the mains side or insulation resistance. Has the max. recommended number of ECGs on one phase in a 3-phase system been exceeded= Make sure that the neutral conductor is connected properly to all the luminaires and makes good contact. Check that moisture has not penetrated the luminaire?
Remedy:
Possible cause:
b) Fault in the wiring on the mains side
Check whether the mains voltage is in the required range for the ECG. Make sure that the neutral conductor is connected properly to all the luminaires and makes good contact. Check that the cables sit correctly in the terminals.
Remedy:
Possible cause:
c) At least one lamp has reached the end of its life owing to a cathode break or increase in lamp voltage.
Replace the lamp (for two-lamp ECGs we recommend replacing both lamps at the same time to keep maintenance time and costs to a minimum).
Remedy:
Possible cause:
d) The “fail-safe” overload protection device in the ECG has responded (the ECG is permanently damaged)
Check whether the lamp(s) operate at other positions. If not, check whether the mains voltage is within the permissible range. Make sure that the neutral conductor is connected properly to all the l uminaires and makes good contact. Replace the ECG and lamp.
Remedy:
79
8.5.2 Brief Glimmer Problem:
Lamp does not start but there is a brief glimmer from one or both lamp (i.e. the protective circuit in the ECG has responded at start-up). Same behaviour even after being off for one minute (internal reset) and restart
Possible cause:
a) At least one lamp has reached the end of its life owing to a cathode break or increase in lamp voltage
Replace the lamp (for two-lamp ECGs we recommend replacing both lamps at the same time to keep maintenance time and costs to a minimum).
Remedy:
Possible cause:
b) Wiring faulty between ECG and lamp (output terminals not used or reversed: contact problems in the holder or the terminals (e.g. wire cross-section too small or too large)
Check the lamp-side wiring fo rcorrect contact. Has the lamp connection been wired according to the wiring diagram on the ECG? For two-lamp ECGs in particular, check that the common or external connection is correctly wired.
Remedy:
Possible cause:
c) A lamp with the wrong wattage has been installed or, in the case of two-lamp ECGs, only one lamp is installed or there are lamps of different wattages
The wattage and type of lamp must correspond to the wattage and type indicated on the ECG or the lamp/ECG combination should be in line with OSRAM’s Indoor and Outdoor Lighting brochure. ECGs designed for two-lamp operation must be operated with two lamps.
Remedy:
Possible cause:
d) The mains voltage lies outside the limit values indicated on the ECG or in the data sheet
Check the voltage at the ECG and at source; check the wiring on the mains side.
Remedy:
Possible cause:
e) The temperature at the measuring point of the ECG is too high (for a brief description of the temperature at the ECG and lamp see end of text)
The luminaire or site of installation should be modified to ensure that the maximum recommended temperature is not exceeded even in onerous conditions (high ambient temperature and/or high supply voltage)
Remedy:
80
8.5.3 Lamp goes out during
operation
Possible cause:
f) Changeover times and voltage levels are outside recommended tolerances for emergency lighting systems with changeover between AC and DC
Remedy: Measure the DC supply voltge and check the switchover properties, or consult the equipment manufacturers.
Possible cause:
g) Non-sinusoidal mains voltage or DC voltage with high residual ripple (e.g. operation with fully electronic leading edge phase control dimmer switch or artificial mains network/emergency generating set)
Remedy:
Check whether the mains voltage is in the required range for the ECG and the wave form or residual ripple in DC operation is within the specified limit values. Dimming is not recommended unless the ECG is expressly approved for dimming, in which case the prescribed controls (special accessories) must be used.
Possible cause:
h) At least one lamp has reached the end of its life owing to a increase in lamp voltage
Remedy: Replace the lamp (for two-lamp ECGs we recommend replacing both lamps at the same time to keep maintenance time and costs to a minimum)
Problem:
The lamp goes out during operation (at least one lamp in the case of two­lamp units)
Possible cause:
a) The reversible protection device in the ECG has responded during operation. The reason may be an intensive transient mains fault (mains voltage falls below the minimum voltage specified on the ECG for longer than permissible). Pulses of exceptional energy (transients) are superimposed on the mains voltage. The value of the mains voltage exceeds the maximum recommended value (e.g. because of a fault in the supply unit). Slow increase in mains voltage following neutral conductor break (unbalanced load, depending, among other things, on the mains load).
Remedy: Disconnect the ECG or luminaire from the mains then check the supply voltage. If such problems occur sporadically, we recommend recording the mains voltage and/or using an oscilloscope or memory voltmeter. The electricity supplier may have to be consulted. Make sure that the neutral conductor is connected properly to this luminaire and makes good contact.
81
8.5.4 Different brightness
levels
Possible cause:
b) The temperatures at the measuring point on the ECG or at the cool spot on the lamp(s) are exceeded (for a brief description of the temperature at the ECG and lamp see end of text).
Remedy: The luminaire or site of installation should be modified to ensure that the maximum recommended temperature is not exceeded even in onerous conditions circumstances (high ambient temperature and/or high supply voltage).
Problem:
Luminous output too low compared with other luminaires. Different brightness levels for the two lamps in two-lamp luminaires. Different brightness levels at the lamp ends
Possible cause:
a) Typical maintenance behaviour of a fluorescent lamp at the end of its life
Replace the lamp (for two-lamp ECGs we recommend replacing both lamps at the same time to keep maintenance time and costs to a minimum).
Remedy:
Possible cause:
b) Lamps of different wattages colour appearance or incorrect wattage
The lamp wattage must match the wattage indicated on the ECG. The colour appearande should be homogenous within an application.
Remedy:
Possible cause:
c) Incorrect wiring between ECG and lamp (output terminals not used or reversed; contact problems)
Check the lamp-side wiring for correct contact. Has the lamp connection been wired according to the wiring diagram on the EG? For two-lamp ECGs in particular, check that the common or external connection is correctly wired. Pay particular attention in the case of special combinations.
Remedy:
Possible cause:
d) Lamps are “force cooled” by draughts
Find the cause of the draught and either eliminate the draught or protect the lamps accordingly.
Remedy:
82
8.5.5 Fault in other
electrical equipment
8.5.6 Problems at master-
slave operation
8.5.7 Humingh or
“chirping” from the ECG
Problem:
Fault in other electrical equipment, particularly radio and television receivers
Possible cause:
a) Wiring problems
Lamp cables should be short, far enough away ( > 5 cm) from earthed metallic parts and, if possible, not laid parallel to mains cables (particularly in the luminaire). If cross-overs are needed they must be at right angles. The mains cables must also be as short as possible.
Remedy
Possible cause:
b) Electrical equipment, radios and televisions are inadequately immunity to interference
Increase the distance between the luminaire and the equipment, if necessary, contact the manufacturer.
Remedy:
Possible cause:
c) The IR remote control signals for TV operate at a similar frequency to the ECG
Move the IR receiver ont the TV out of the readiation field of the lamp or disable it.
Remedy:
Problem:
Problems on master-slave arrangements for 2lamp ECGs
Possible cause:
Wiring problems
Remedy:
Lamp cables should be short, far enough away ( > 5 cm) from earthed metallic parts and, if possible, not laid parallel to mains cables (particularly in the luminaire). If cross-overs are needed they must be at right angles. The mains cables must also be as short as possible. In master-slave arrangements the maximum length of the cable to the daughter luminaire must not be exceeded.
Problem:
Huming or “chirping” from the ECG
Possible cause:
Non-sinusoidal AC voltage
Remedy:
Eliminate sources of interference if necessari in consultation with the electricity supplire.
83
9. Lamp-ECG Combinations
9.1 FQ®...HO-Fluorescent Lamps
Lamp ECG type LxWxH [mm] P
Sys
lm
[W]
1-lp
FQ
®
24W
QTi 1x14/24/21/39 360 x 30 x 21 27 1750
QT-FQ 1x24… CW 360 x 30 x 30 27 1750 QT-FQ 1x24-39…F/CW 360 x 30 x 21 28 1750 QT-M 1x26-42…S 103 x 67 x 31 27 1750 QT 1x24/230-240 237 x 30 x 30 25 1750 QT-ECO 1x18-24…S 80 x 40 x 22 22 1600 QT-ECO 1x18-24…L 150 x 22 x 22 22 1600
2-lp
FQ
®
24W
QTi 2x14/24/21/39 423 x 30 x 21 54 2x1750
QT-FQ 2x25…CW 360 x 30 x 30 51 2x1750 QT-FQ 2x54… F/CW 423 x 30 x 21 53 2x1750 QT-M 2x26-32 … S 123 x 79 x 33 54 2x1750
Lamp ECG type LxWxH [mm] P
1-lp
FQ
®
39W
QTi 1x14/24/21/39 360 x 30 x 21 43 3100
QT-FQ 1x39 … CW 360 x 30 x 30 42 3100 QT-FQ 1x24-39…F/CW 360 x 30 x 21 41 3100 QT-M 1x26-42 … S 103 x 67 x 31 40 3000
2-lp
FQ
®
39W
QTi 2x14/24/21/39 423 x 30 x 21 88 2x3100
QT-FQ 2x39…CW 360 x 30 x 30 85 2x3100 QT-FQ 2x24-39…F/CW 423 x 30 x 21 82 2x3100
Sys
[W]
lm
Lamp ECG type LxWxH [mm] P
1-lp
FQ
®
49W
QTi 1x35/49/80 360 x 30 x 21 55 4300
QT-FQ 1x49…CW 360 x 30 x 30 54 4300
2-lp
FQ
®
49W
QTi 2x35/49 423 x 30 x 21 110 2x4300
QT-FQ 2x49…CW 360 x 30 x 30 110 2x4300
Sys
[W]
lm
Lamp ECG type LxWxH [mm] P
1-lp
FQ
®
54W
QTi 1x28/54 360 x 30 x 21 61 4450
QT-FQ 1x54…CW 360 x 30 x 30 61 4450 QT-FQ 1x54…F/CW 360 x 30 x 21 59 4450
2-lp
FQ
®
54W
QTi 2x28/54 423 x 30 x 21 119 2x4450
QT-FQ 2x54…CW 360 x 30 x 30 118 2x4450 QT-FQ 2x54…F/CW 423 x 30 x 21 122 2x4450
Sys
[W]
lm
Lamp ECG type LxWxH [mm] P
1-lp
FQ
®
80W
QTi 1x35/49/80 360 x 30 x 21 89 6150
QT-FQ 1x80…CW 360 x 30 x 30 86 6150 QT-FQ 1x80…F/CW 360 x 30 x 21 87 6150
2-lp
FQ
®
80W
QT-FQ 2x80 423 x 30 x 21 176 2x6150
Sys
[W]
lm
84
9.2 FH®...HE-Fluorescent Lamps
9.3 FC®…Fluorescent Lamps
Lamp ECG type LxWxH [mm] P
Sys
lm
[W]
1-lp
FH
®
14W
QTi 1x14/24/21/39 360 x 30 x 21 18 1200
QT-FH 1x14-35…CW 360 x 30 x 30 18 1200 QT-FH 1x14-35…F/CW 360 x 30 x 21 18 1200 QT-FH 1x14 237 x 30 x 30 16 1200 QT-ECO 1x4-16...S 80 x 44 x 22 15 1200 QT-ECO 1x4-16…L 150 x 22 x 22 15 1200
2-lp
FH
®
14W
QTi 2x14/24/21/39 423 x 30 x 21 32 2x1200
QT-FQ 2x14-35…CW 360 x 30 x 30 31 2x1200 QT-FH 2x14-28…F/CW 423 x 30 x 21 32 2x1200
3-lp 4-lp
FH FH
®
14W
®
14W
QT-FH 3x14…CW 425 x 40 x 30 50 3x1200 QT-FH 4x14…CW 425 x 40 x 30 65 4x1200
Lamp ECG type LxWxH [mm] P
1-lp
FH
®
21W
QTi 1x14/24/21/39 360 x 30 x 21 25 1900
QT-FH 1x14-35…CW 360 x 30 x 30 24 1900 QT-FH 1x14-35…F/CW 360 x 30 x 21 23 1900 QT-FH 1x21 237 x 30 30 23 1900 QT-ECO 1x18-21...S 80 x 40 x 22 23 1800
2-lp
FH
®
21W
QTi 2x14/24/21/39 423 x 30 x 21 47 2x1900
QT-FH 2x14-35…CW 360 x 30 x 30 46 2x1900 QT-FH 2x14-28…F/CW 423 x 30 x 21 46 2x1900
Sys
[W]
lm
Lamp ECG type LxWxH [mm] P
1-lp
FH
®
28W
QTi 1x28/54 360 x 30 x 21 32 2600
QT-FH 1x14-35…CW 360 x 30 x 30 31 2600 QT-FH 1x14-35…F/CW 360 x 30 x 21 31 2600
2-lp
FH
®
28W
QTi 2x28/54 423 x 30 x 21 63 2x2600
QT-FH 2x14-35…CW 360 x 30 x 30 61 2x2600 QT-FH 2x14-28…F/CW 423 x 30 x 21 61 2x2600
Sys
[W]
lm
Lampe ECG type LxBxH [mm] P
1-lp
FH
®
35W
QTi 1x35/49/80 360 x 30 x 21 39 3300
QT-FH 1x14-35…CW 360 x 30 x 30 38 3300 QT-FH 1x14-35…F/CW 360 x 30 x 21 38 3300
2-lp
FH
®
35W
QTi 2x35/49 423 x 30 x 21 79 2x3300
QT-FH 2x14-35…CW 360 x 30 x 30 77 2x3300
Lamp ECG type LxBxH [mm] P
1-lp
FC
®
22W
QT-FQ 1x24…CW 360 x 30 x 30 27 1800
QT-M 1x26-42…S 103 x 67 x 31 26 1800 QT-ECO 1x18-24…S 80 x 40 x 22 22,5 1650
2-lp
FC
®
22W
QT-ECO 1x18-24…L 150 x 22 x 22 22,5 1650 QT-FQ 2x24…CW 360 x 30 x 30 51 2x1800
QT-M 2x26-32…S 123 x 79 x 33 54 2x1800 QT 2x24… 280 x 42 x 30 51 2x1800
Sys
[W]
Sys
[W]
lm
lm
85
Lamp ECG type LxBxH [mm] P
1-lp
FC
®
40W
QT-FQ 1x39…CW 360 x 30 x 30 42 3200
QT-M 1x26-42…S 103 x 67 x 31 44 3200
2-lp
FC
®
40W
QT-FQ 2x39…CW 360 x 30 x 30 85 2x3200
Sys
[W]
lm
Lamp ECG type LxBxH [mm] P
1-lp
FC
®
55W
QT-FQ 1x55…S 123 x 79 x 33 60 4000
Sys
[W]
lm
Lamp ECG type LxBxH [mm] P
FC® 22 + 40
QT-M 2x26-32…S 123 x 79 x 33 70
Sys
[W]
lm
1800 +
3200
86
10. Tender Documents
10.1 QUICKTRONIC® INTELLIGENT QTi
See www.osram.com/ecg/tender-documents
µProzessor controlled ECG to operate T5/ 16 mm fluorescent lamps
®
FQ
...HO and FH®...HE in equal length. Automatic lamp detection during lamp starting Optimized Operation of all approved lamps with rated data Lamp operation acc. to EN 60929 and IEC 60929
Range :
QTi 1x35/49/80 QTi 1x28/54 Geometry: 360 x 30 x 21 mm³ QTi 1x14/24/21/39
QTi 2x35/49 QTi 2x28/54 Geometry: 423 x 30 x 21 mm³ QTi 2x14/24/21/39
Lamp Preheat Start within 1 Second Ambient Temperature: -20 °C up to+50 °C Emergency Lighting acc. to . DIN VDE 0108 (EN 61347-2-3) possible DC Voltage Range: 154 V to 276 V AC Voltage Range:: 198 V to 254 V
Approval Marks:
10.2 QUICKTRONIC® MULTIWATT for FH…HE h = 30 mm
- Circuitry
-Circuitry
Automatic restart after lamp replacement ECG-lifetime: 50.000 h with a max. 10 % failure rate (at tc = 70 °C)
Fully electronical, digital control gear to operate all FH®...HE-fluorescent lamps Lamp preheat start within 1 Second Combi Wiring Terminal for automatic and manual wiring
- Circuitry
Ambient Temperature: -20 °C to +50 °C
Emergency lighting acc. to. DIN VDE 0108 (EN 61347-2-3) possible DC Voltage Range: 154 V to 276 V AC Voltage Range: 198 V to 254 V
-Circuitry
Range:
QUICKTRONIC® MULTIWATT
87
10.3 QUICKTRONIC® for FQ…HO h = 30 mm
QT-FH 1x14-35/230-240 CW QT-FH 2x14-35/230-240 CW Geometry: 360 x 30 x 30 mm³
Automatic restart after lamp replacement ECG-lifetime: 50.000 h with a max. 10 % failure rate (at tc = 70 °C)
Fully electronical, digital control gear to operate all FQ®...HO fluorescent lamps Lamp preheat start within 1 Second Combi Wiring Terminal for automatic and manual wiring
- Circuitry
Ambient Temperature: -20 °C to +50 °C
Emergency Lighting acc. to DIN VDE 0108 (EN 61347-2-3) possible DC Voltage Range: 154 V to 276 V AC Voltage Range: 198 V to 254 V
-Circuitry
Range:
QUICKTRONIC® for FQ®...HO fluorescent lamps QT-FQ 1x24/230-240 CW QT-FQ 2x24/230-240 CW QT-FQ 1x39/230-240 CW QT-FQ 2x39/230-240 CW QT-FQ 1x49/230-240 CW QT-FQ 2x49/230-240 CW QT-FQ 1x54/230-240 CW QT-FQ 2x54/230-240 CW QT-FQ 1x80/230-240 CW Geometry: 360 x 30 x 30 mm³
Automatic restart after lamp replacement ECG-lifetime: 50.000 h with a max. 10 % failure rate (at tc = 70 °C)
88
11. Index
Technical data are subject to change without any notes. Printed data in this edition replace previous.
Ambient Temperature 3.9.2 Approval Marks 2.17 Automatic Relamping 2.5
Cable Cross Section 3.1.2 Cable Length 3.3; 7.1 Cable Routing 3.1.6 Cable Type 3.1.1 CCC-Mark 2.20 CE-Mark 2.17.2; 2.19 Circuit breaker 3.15 Cold Spot 3.9.3; 3.9.5; 7.5.4 Conducted Interferences 3.2.2.2 cut-off Technology 2.14
DC-Voltage 6.6
ECG in Sound Studios 6.5 ECG Milestones 1.12 ECG Temperature 3.9.2 Economy 2.2 EEI 2.18 Electromagnetic Compatibility EMC 2.19; 3.2 Electromagnetic Disturbances 6.3.1 Emergency Lighting 6.5 End-of-Life 2.17.1 Energy Saving 1.6 Exterior Applications 6.1
Failure Rate 2.10 FAQ for QTi 5.7 Fault Currents 3.16; 6.1.1 Fluorescent Circline 1.2.3 Functional Earth 3.7 ff
Harmonic Content 3.2 ff High Efficiency FH High Output FQ High Voltage Test 3.13 Hot Wires 3.4 Humidity 3.3; 3.9; 6.1.1
Ignition Time 6.5.1 ff Immunity 3.2 ; 8.5.5 Inrush Current 3.14 Installation Instructions 3.2.2.5; 3.9.4; 6.1.1 Insulation Distances 3.12 Insulation Test 3.13 Insulation 3.1.4
®
…HE 1.2.1
®
…HO 1.2.2
89
Lamp Failure 2.6 Lamp Temperature 3.9.3 Lamp-ECG-Combination 7.4 ; 9 ff Leakage Current 3.16 Lifetime 2.12; 3.9.1 Lighting Comfort 2.1 Line Voltage 120 V/277 V 2.4.7 Line Voltage 2.4 ff Luminaire Wiring Test 3.10
Magnetic Field 3.2.2.4 Master-Slave-Operation 8.5.6 Measurementpoint Temperature 3.9.2.1
Noise 2.7; 6.2.1
OUTKIT 6.1; 6.1.2 Overvoltage 2.3; 2.4; 1 f; 8.2
PE Connection 3.7 PE-Connection 3.7; 3.8.2; 3.11 Permissible Cable Lengths 3.3; 7.1 ff Power Factor λ 2.8 Protection Class I or II 3.11; 6.7
QUICKTRONIC
®
INTELLIGENT QTi 5 ff
Radio Interference Suppression 2.17; 3.2.2; 3.1.6; 3.4 Reliability 2.10 Resistance to Frequent Switching 2.11
Safety Lighting 6.5 Safety 2.3 Selective Shielding 3.2.2.4 Self heating 3.9.1 Storage Temperature 3.9 Switching between Lamp and ECG 3.5
ta Ambient Temperature 3.9 ff tc Measurement Point 3.9.2.1 Temperature Measurement 3.9.5 Temperature Range 3.9 Tender Documents 10f Terminals 7.2 Test Adapter 3.10.1 Thermal Influences 2.13 Three-Phase-System 3.13.3 Treatment Rooms 6.3 Troubleshooting Tips 8ff
Undervoltage 2.4.3 f; 8.3 U-OUT 2.16
90
VDE-EMC sign 2.8; 2.17.2 VDE-sign 2.19 Voltage Range 2.4 Voltage Resistance 3.13.4 ; 3.17
Wiring Diagrams 3.1 ff; 6.1.1
91
Head office Germany
OSRAM GmbH Hellabrunner Straße 1 81543 München Tel.: +49-89-62 13-0 Fax: +49-89-62 13-20 20
International addresses
Albania
(supported by OSRAM Greece)
Argentina
OSRAM Argentina S.A.C.I. Ramos Mejia 2456 B 1643 ADN Beccar Pcia. De Buenos Aires Tel.: +54-11-6333-8000 Fax: +54-11-6333-8001
Australia
OSRAM Australia Pty. Ltd. Sydney 11th Floor, Building 1 423 Pennant Hills Road 2120 Pennant Hills, N.S.W. P.O. Box 673 1715 Pennant Hills Tel.: +61-29-4 81-83 99 Fax: +61-29-4 81-91 27
Austria
OSRAM GmbH Lemböckgasse 49/C/5 1230 Wien Postfach 1 62 1231 Wien Tel.: +43-1-6 80 68 -0 Fax: +43-1-6 80 68 -7
Azores
(supported by OSRAM Portugal)
Benelux
OSRAM Benelux B.V. Klaverbaan 102 2908 KD Capelle a/d Ijssel Netherlands Tel.: NL +31-10-750 14 14
BE +32-78-55 08 20
Fax: NL +31-10-750 14 06
BE +32-78-55 08 28
Bosnia-Herzegovina
(supported by OSRAM Croatia)
Brazil
OSRAM do Brasil Lâmpadas Elétricas Ltda. Av. Dos Autonomistas, 4229 06090-901 Osasco-SP/Brazil Tel.: +55-11-36 84 74 08 Fax: +55-11-36 85 94 95
Bulgaria
OSRAM EOOD Nikola Obreschkov 1 Wh. A., App. 1 1113 Sofia Tel.: +359-2-9 71 22 62 Fax: +359-2-9 71 45 46
Canada
OSRAM SYLVANIA Ltd./Lte. 2001 Drew Road Mississauga Ontario L5S 1S4 Tel.: +1-905-6 73 61 71 Fax: +1-905-6 71 55 84
Chile
OSRAM Chile Ltda. Santa Elena de Huechuraba 1135 B Comunade Huechuraba Santiago de Chile Tel.: +56-2-7 40-09 39 Fax: +56-2-7 40-04 66
China
OSRAM China Lighting Ltd. No.1 North Industrial Road, Postal Code 528000 Foshan, Guangdong Tel.: +86-757-864 82-111 Fax: +86-757-864 82-222
OSRAM Shanghai Rep. Office Harbour Ring Plaza No. 18 Xi Zang Middle Road Room 2802, 2803 A Shanghai, 200001 P.R.C. Tel.: +86-21-53 85 28 (48) Fax: +86-21-64 82 12 19
Colombia
OSRAM de Colombia Diagonal 109 No. 21-05 Oficina 607, 608 Bogotá Tel.: +57-1-6 19 24 07 Fax: +57-1-6 37 18 55
Croatia
OSRAM d.o.o. Majstora Radonje 10 10000 Zagreb Tel.: +385-1-303-20 00 Fax: +385-1-303-20 01
Czechia
OSRAM spol. s.r.o. Kodanska 1441/46 100 10 Praha 10 Tel.: +420-234 06 60 00 Fax: +420-234 06 60 20
Denmark
OSRAM A/S Dybendalsvænget 3 2630 Tåstrup Postboks 259 2630 Tåstrup Tel.: +45-44-77 50-00 Fax: +45-44-77 50-55
Egypt
OSRAM Rep. Office Cairo 5th Floor, Unit No. 507 57 Giza Street Cairo, Giza Tel.: +20-2-7 48 66 46 Fax: +20-2-7 48 66 46
Ecuador
OSRAM del Ecuador S.A. Casilla 09-01-8410 Guayaquil Tel.: +593-4-2 89 36 09 Fax: +593-4-2 89 35 58
Estonia
(supported by OSRAM Finland)
Finland
Oy OSRAM AB, Helsinki Vanha Porvoontie 229 01380 Vantaa Box 91 01301 Vantaa Tel.: +358-9-74 22 33 00 Fax: +358-9-74 22 33 74
France
OSRAM SASU 5, Rue d’Altorf 67124 Molsheim BP 1 09 67124 Molsheim Tel.: +33-388-49 75 99 Fax: +33-388-49 75 975
Great Britain
OSRAM Ltd., London OSRAM House Waterside Drive Langley, Berkshire SL3 6EZ Tel.: +44-17 53 48 4 (100) Fax: +44-17 53 48 42 22
Greece
OSRAM A.E. Frantzi 6 & Ag. Pavlou 12132 Peristeri Tel.: +30-2 10-5 20 18 00 Fax: +30-2 10-5 22 72 00
Hong Kong
OSRAM Prosperity Co. Ltd. Rm 4007-09 Office Tower Convention Plaza 1 Harbour Road, Wanchai Tel.: +852-25 11 22 68 Fax: +852-25 11 20 38
Hungary
OSRAM KFT. Alkotas utca 41. 1123 Budapest Tel.: +36-1-2 25-30 55 Fax: +36-1-2 25-30 54
India
OSRAM India Private Ltd. Signature Towers, 11th Floor, Tower-B South City-I 122001 Gurgaon Haryana/India Tel.: +91-124-238 31-80 Fax: +91-124-238 31-82
Indonesia
PT. OSRAM Indonesia Jalan Siliwangi KM 1 Desa Keroncong Jatiuwung 15134 Tangerang Tel.: +62-21-5 90 01 27 Fax: +62-21-5 90 05 59
Iran
OSRAM Lamps OSRAM PJS Co. Bokharest Ave, Str. 6, No. 13 Tehran Tel.: +98-21-8 73 84 76 Fax: +98-21-8 73 24 13
Italy
OSRAM Società Riunite OSRAM Edison-Clerici Spa Via Savona 105 20144 Milano Tel.: +39-02-42 49-1 Fax: +39-02-42 49-380
Japan
OSRAM MELCO Ltd. Tobu Yokohama Bldg.No. 3 (4F) 8-29 Kita-Saiwai 2-chome, Nishi-Ku 220-0004 Yokohama Tel.: +81-45-3 23 51-29-0 Fax: +81-45-3 23 51-55
OSRAM Ltd. Tobu Yokohama Bldg.No. 3 (6F) 8-29 Kita-Saiwai 2-chome, Nishi-Ku 220-0004 Yokohama Tel.: +81-45-3 23 51-00 Fax: +81-45-3 23 51-10
Korea
OSRAM Korea Co. Ltd. 3rd. Fl. Ye-Sung Bldg. 150-30 Samsung-dong, Kangnam-Ku Seoul 135-090 Tel.: +82-2-5 54 41 12 Fax: +82-2-5 56 16 44
Latvia
(supported by OSRAM Finland)
Lithuania
(supported by OSRAM Finland)
Macedonia
(supported by OSRAM Greece)
Madeira
(supported by OSRAM Portugal)
Malaysia
OSRAM Sdn Bhd
7.05-7.06 Amoda Building 22 Jalan Imbi 55100 Kuala Lumpur Tel.: +60-3-21 45 95-33 Fax: +60-3-21 45 95-35
Mexico
OSRAM de México, S.A. de C.V. Camino a Tepalcapa No. 8 Col. San Martin 54900 Tultitlán Edo. de México Tel.: +52-55-58 99-18 00 Fax: +52-55-58 84-70 00
Norway
OSRAM AS Strandveien 50 1366 Lysaker Tel.: +47-67 83 67-00 Fax: +47-67 83 67-40
Philippines
OSRAM Philippines Ltd. Corp. Unit 2002–2003 Antel Global Corporate Center Julia Vargas Avenue Ortigas Center Pasig City Tel.: +632-687 60 48-50 Fax: +632-687 60 57
Poland
OSRAM sp. z o.o. ul. Wiertnicza 117 02-952 Warszawa Tel.: +48-22-550 23 00 Fax: +48-22-550 23 19
Portugal
OSRAM Empresa de Aparelhagem Eléctrica Lda. Rua do Alto do Montijo Nr. 15-4 andar 2794-069 Carnaxide Tel.: +351-2 14 16 58 60 Fax: +351-2 14 17 12 59
Romania
OSRAM Romania S.R.L. Calea Plevnei nr. 139 corp B, sector 6 060011 Bucaresti Tel.: +40-21-2077-386 Fax: +40-21-2077-389
Russia
OSRAM Moscow Ul. Malaja Kaluschskaja 15/4 119071 Moscow Tel.: +7-095-9 35 70-70 Fax: +7-095-9 35 70-76
Serbia and Montenegro
OSRAM d.o.o., Beograd Cika Ljubina 15/V YU-11000 Beograd Tel: +381 (0)11-30 30-860 Fax: +381 (0)11-30 30-853
Singapore
OSRAM Pte. Ltd. 159 Sin Ming Road #05-04 Amtech Building 575625 Singapore Tel.: +65-65 52 01 10 Fax: +65-65 52 71 17
Slovakia
OSRAM Nové Zámky Komárnanská cesta 7 94093 Nové Zámky Tel.: +42-1-35 64 64-0 Fax: +42-1-35 64 64-880
Slovenia
(supported by OSRAM Austria)
South Africa
OSRAM (Pty.) Ltd., 260, 15th Road 1683 Randjespark/Midrand Private BAG X 206 1685 Halfway House/Midrand Tel.: +27-11-2 07 56 00 Fax: +27-11-8 05 17 11
Spain
OSRAM, S.A. Calle de la Solana, 47 28850 Torrejón de Ardoz (Madrid) Tel.: +34-91-6 55 52 00 Fax: +34-91-6 56 82 16
Sweden
OSRAM AB Rudanvägen 1 13625 Haninge Box 5 04 13650 Haninge Tel.: +46-8-7 07 44-00 Fax: +46-8-7 07 44-40
Switzerland
OSRAM AG, Winterthur In der Au 6 8401 Winterthur/Töss Postfach 2179 8401 Winterthur/Töss Tel.: +41-52-2 09 91 91 Fax: +41-52-2 09 92 75
Taiwan
OSRAM Taiwan Company Ltd. Sung Chiang Road, 7th Floor, No. 87 Sung Chiang Road P.O. Box 46304 Taipei – Taiwan, R.O.C. Tel.: +886-2-25 08 35 02 Fax: +886-2-25 09 67 82
Thailand
OSRAM Thailand Co. Ltd. 100/45, 24th Floor Sathorn Nakorn Tower North Sathorn Road Khwaeng Silom Khet Bangrak, Bangkok 10500 Tel.: +66-2-6 36 74 75 Fax: +66-2-6 36 74 77
Turkey
OSRAM AMPUL TIC. A.S. Meclisi Mebusan Caddesi 125 80400 Findikli, Istanbul/TR Tel.: +90-212-334-1334 Fax: +90-212-334-1142
Ukraine
OSRAM Ukraine Podil Plasa Business Center 30-A Spaska Str., office 2-3B Kiev 04070 Tel.: +38-044-467 66 67 Fax: +38-044-467 69 58
United Arab Emirates
OSRAM Middle East FZE P.O. Box 17476 Room #602-603, LOB #16 Jebel Ali Free Zone Dubai United Arab Emirates Tel.: +971-4-88 13-767 Fax: +971-4-88 13-769
USA
OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923 Tel.: +1-978-777-19 00 Fax: +1-978-750-21 52
Vietnam
OSRAM Singapore Pte. Ltd. Rep. Office Vietnam 59A Ly Thai To Street, Hanoi Press Club Hoan Kiem District Hanoi Tel.: +84-4-93 49-801 Fax: +84-4-93 49-803
Internet
http://www.osram.de http://www.osram.com http://catalog.myosram.com/DE http://catalog.myosram.com/EN
130T015GB 05/05 PC-P Subject to change without notice. Errors and omissions excepted.
z
Printed on paper treated with chlorine-free bleach.
Loading...