10.2 QUICKTRONIC® MULTIWATT for FH…HE h = 30 mm ...... 87
10.3 QUICKTRONIC® for FQ…HO h = 30 mm ............................88
11. Index ................................................................................................... 89
5
1. Introduction
1.1 History
The development of linear fluorescent goes back to the thirties of the 20th
century. The diameter of 51mm was very voluminous. However, better
efficiency did not come up before the fifties.
1879
Kohlefaden-Glühlampe
von Thomas A. Edison
Incandescent lamps with carbon filament
by Thomas A. Edison
1910
Glühlampen mit Wolfram-Wendel
Incandescent lamps with tungsten coils
DECOSTAR Niedervolt-Halogenglühlampen
mit Kaltlichtreflektor
DECOSTAR low-voltage tungsten-halogen
lamps with dichroic reflectors
®
lamps
®
H4
®
®
®
DE LUXE
®
L
PL(W)UN (V)fN (Hz)IN (A)lta ( C)
22050600,090,95 C
1xL18 W1x16-2050
2400,0850,93 C
Anwendungsbereich: AC/DC 198 V bis 254 V
Geeignet fr Batteriespannungen 154 V bis 276 V
Zur Verwendung in Anlagen nach VDE 0108 geeignet
Range of application: AC/DC 198 V to 254 V
Temp.-Test
Range of battery voltage: 154 V to 276 V
c = 70 C max.
t
Suitable for emergency installations acc. to VDE 0108
class B
0712T201 OW2 Made in Germany
1985
OSRAM DULUX®EL Kompaktleuchtstofflampen mit elektronischem Vorschaltgerät
®
OSRAM DULUX
with electronic control gear
1987
POWERSTAR HQI-T
Halogen-Metalldampflampen
POWERSTAR HQI-T
compact metal halide lamps
FQ Lichtstarke Leuchtstofflampe / FQ high power
fluorescent lamps
1997
OSRAM ENDURA Die elektrodenlose
Hochleistungsleuchtstofflampe
OSRAM ENDURA The high-performance
electrodeless fluorescent lamp
EL energy-saving lamps
The improvements according to the luminous flux and lifetime with the T12
lamp (38 mm diameter) made an economic and even outdoor application
possible. There were continuous improvements for optimizations as for
example Amalgamtechnology. In this term fluorescent lamps were operated
by conventional control gears (CCG).
The decisive breakthrough was at the end of the 70s and early 80s. In the
year 1978, a new T8/∅ 26 mm- lamp generation started to replace
T12/∅38mm- fluorescent lamps. New phosphors with higher resistance
came into the market. Nowadays known under triband-phosphor (LUMILUX
light colours). The diameter of the lamps was reduced from 38mm to 26mm
however the length was kept with 59, 120 and 150 cm as the sockets G13.
The new wattage of 18, 36 and 58 W was advantageous as it was a
reduction of at least 10% compared to the T12/∅38 mm lamps in 20, 40
and 65 W.
This was also the hour of birth for the Electronic Control Gears (ECG). First
the circuits were in a instant start mode what also was called softstart up to
the programmed or preheat start. Together with modern Electronic Control
Gear QUICKTRONIC T8/∅26 mm- fluorescent lamps became even more
efficient and longer lasting. In addition, the thermal behavior of the
luminous flux was improved.
The T8-system was permanently improved as the example of the tribandphosphor shows with a very high service life of the fluorescent lamp.
In 1995, the next milestone of fluorecent lamp development saw the
introduction of new FH...HE (H
igh Efficiency) systems to the market. With
the reduced diameter of 16 mm only the lamp was designed. It is available
in 14, 21, 28 and 35 W with the G5 sockets. It is 50 mm shorter as the T8
fluorescent lamps.
T5 fluorescent lamps only can be operated by Electronic Control Gear. So
the light output and the life time of the lamp were designed from the
beginning to an optimum of up to 104 lm/W. The maximum luminous flux of
the T5 lamps is at 35 °C compared to 25 °C at T8 or T12 lamps. The
reduced lamp diameter of 16 mm as the maximum lumen output at 35 °C
are the relevant feature for a higher efficiency of the fixture.
6
In 1996, the T5 lamp family was completed with the types of higher lumens
than volume FQ
®
…HO (High Output)-fluorescent lamps. They are available
in the wattage of 24, 39, 49, 54 and 80 W with the identical lengths as the
FH…HE types. With up to 7000 lm for FQ 80 W HO this is the lamp family
with the highest light output.
In 1999, the third member of the T5/∅ 16 mm-lamp family was introduced
to the market. Away from usual light strips, compact, efficient and
unconventional luminaires benefit from this new, circular FC
®
-lamp: 50 %
more light output than with comparable standard circular lamps.
Special Note:
Independent of the lamp diameter of the fluorescent lamp the luminous flux
is specified for an ambient temperature of 25 °C. These values are
exclusive to be used for light plannings. The value of the luminous flux is
for the T5/∅ 16 mm fluorescent lamps FH
®
...HE and FQ®...HO for 25 °C
below the values for 35 °C. The values at 35 °C ambient temperature are
only for information. The Circular lamps FC
®
only have value of luminous
flux at 25 °C. The advantages of the T5/∅ 16 mm fluorescent lamps show
their advantages in the improved efficiency of the fixture.
Detailed technical information about QUICKTRONIC
®
are shown in the
latest indoor outdoor lighting and can be downloaded under
www.osra.de/ecg
.
QUCKTRONIC
®
for the operation of the T5/∅ 16 mm fluorescent lamps
have all features of a high quality ECG.
Good radio interference suppression
Good radio interference suppression
Good radio interference suppression
EN 55015
EN 55015
EN 55015
CISPR 15
CISPR 15
CISPR 15
max.Interference voltage[dB µV]
max.Interference voltage[dB µV]
max.Interference voltage[dB µV]
110
110
110
90
90
90
80
80
8070
70
70
66
66
66
60
60
60
56
56
56
-2
-2
-2
10
10
10
110010-1
110010-1
110010-1
10210
10210
10210
Frequency[MHz]
Frequency[MHz]
Frequency [MHz]
Mains harmonics
Mains harmonics
IEC 61000-3-2
IEC 61000-3-2
EN 61000-3-2
EN 61000-3-2
Mains harmonics[%]
Mains harmonics [%]
30
30
30
25
25
25
Limit values
Limit values
Limit values
20
20
20
to IEC
to IEC
to IEC
15
15
15
10
10
10
5
5
5
0
0
0
3.5.7.9.11.13.
3. 5. 7. 9. 11. 13.
3. 5. 7. 9. 11. 13.
Harmonics
Harmonics
Reliability of the ECG
Reliability of the ECG
IEC 61347-2-3
IEC 61347-2-3
EN 61347-2-3
EN 61347-2-3
zAll insulated
z All insula tedzCompliance with creepage and
z Compliance with creepage and
clearance distances
clearance distances
zECG shutdown in case of failure
z ECG shutdown in case of failure
Lamp operation to standards
Lamp operation to standards
IEC 60929
IEC 60929
EN 60929
EN 60929
rel. system luminous flux[%]
rel. system luminous flux [%]
100
100
90
90
80
80
70
70
60
60
50
50
2
2
ECG
ECG
10.18.
10.18.
CCG
CCG
4681012141618222420
4 6 8 1012141618 222420
Time of usage in hours[tsd]
Time of usage in hours [tsd]
1.2 T5/∅ 16 mm-
Fluorescent Lamps
Immunity
Immunity
IEC 61547
IEC 61547
EN 61547
EN 61547
230 V
230 V
50 Hz
50 Hz
Power regulation
Power regulation
At higher ambient temperatures
At higher ambient temperatures
P Gas [W]
P Gas [W]
50
50
40
40
30
30
20
20
01020304050607080
0 1020304050607080
Tube walltemperature[°C]
Tube wall temperature [°C]
Minimum ECG life
Minimum ECG life
zLong-life electrolytic capacitor
z Long-life electrolytic capacitor
(50,000h att
(50,000h at t
zOptimised circuit
z Optimised circuitzLow selfheating
z Low self heating
Functional ECG [%]
Functional ECG [%]
100
100
100
80
80
80
60
60
60
40
40
40
20
20
20
90°C
90°C
90°C
0
0
0
0 20 40 60 80 100 120 140
0 20 40 60 80 100 120 140
020406080100120140
)
)
c max
c max
60°C
60°C
60°C
70°C
70°C
70°C
Hours of operation[tsd]
Hours of operation [tsd]
Measuring
Measuring
point t
point t
50°C
50°C
50°C
c
c
Reliable ignition
Reliable ignition
at low temperatures
at low temperatures
-30-25-20-15-10-5
-30 -25 -20 -15 -10 -5
Lamps starting[%]
Lamps starting [%]
EVG
EVG
KVG
KVG
Ambient temperature[°C]
Ambient temperature [°C]
100
100
80
80
60
60
40
40
20
20
0
0
0
0
The diameter and also the description of the new fluorescent lamp family is
based on American measures: (1 inch = 25.4 mm)
The value is combined with a T (tube).
5/8 of an inch = 16 mm Î T5-fluorescent lamp
Classification:
T2 tube diameter of 7 mm
T5 tube diameter of 16 mm
T8 tube diameter of 26 mm
T12 tube diameter of 38 mm
T17 tube diameter of 51 mm (1936)
7
1.2.1 High Efficiency
®
FH
…HE
1.2.2 High Output FQ
®
…HO
1.2.3 Fluorescent Circular
®
FC
1.2.4 Comparison of
Lumens between
T8/∅ 26 mm- and
T5/∅ 16 mmFluorescent Lamps
Consecutively the important data of the FH…HE, FQ…HO and FC
fluorescent lamps are shown.
type length [mm]lumens at
ta=25 °C
FH® 14W HE
FH® 21W HE
FH® 28W HE
FH® 35W HE
549 1200 1350
849 1900 2100
1149 2600 2900
1449 3300 3650
lumens at
ta=35 °C
Values for light colors 827, 830, 840
type length [mm]lumens at
ta=25 °C
FQ® 24W HO
FQ® 39W HO
FQ® 54W HO
FQ® 80W HO
FQ® 49W HO
Detailed technical data of T5/∅ 16 mm-fluorescent lamps can be found in
the OSRAM product catalogue and under www.OSRAM.com.
Fluorescent lamp innovation: T8 Î T5
T8 (Ø26 mm)
T8 (Ø 26 mm)
Luminous fluxat 25°C
Luminous flux at 25°C
600 mm 900 mm 1.200 mm 1.500 mm
600 mm 900 mm 1.200 mm 1.500 mm
18 W 30 W 36 W 58 W
18 W 30 W 36 W 58 W
1.350 lm 2.400 lm 3.350 lm 5.000 lm
1.350 lm 2.400 lm 3.350 lm 5.000 lm
1.3 Electronic Control Gear
T5 (Ø16 mm)
T5 (Ø 16 mm)
Luminous fluxat 25°C
Luminous flux at 25°C
550 mm 850 mm 1.150 mm 1.450 mm
550 mm 850 mm 1.150 mm 1.450 mm
FH14 W 21 W 28 W 35 W
FH14 W 21 W 28 W 35 W
1.200 lm 1.900 lm 2600 lm 3.300 lm
1.200 lm 1.900 lm 2600 lm 3.300 lm
FQ24 W 39 W 54 W 49 W / 80 W
FQ24 W 39 W 54 W 49 W / 80 W
1.750 lm 3.100 lm 4.450 lm 4.300 lm / 6.150 lm
1.750 lm 3.100 lm 4.450 lm 4.300 lm / 6.150 lm
Since the early seventies Electronic Control Gear have been used in
computers and consumer electronics. As this technology offers substantial
advantages, it was only natural to use it also for lighting purposes. Linear
and compact fluorescent lamps must be operated with ballasts, as the
process of gas discharge requires well defined currents and voltages. The
8
1.4 Different Principles,
Different Behavior
1.5 Advantages of
Electronic Control Gear
ballast is responsible for preheating the lamp electrodes, for sufficient
ignition voltage and for limiting the lamp current.
The basic functions that are mentioned in chapter 1.3, can usually be
carried out with electromagnetic (inductive) ballasts. These ballasts are
classified into conventional control gear (CCG) and low loss ballasts (LLG).
The latter follow the same principle as CCG, however, due to different
engineering design they consume less energy. The much better solution is
to operate fluorescent lamps with Electronic Control Gear (ECG). Besides
the advantages of flicker-free lighting, longer lamp life and higher system
efficacy (lamp + ECG), features such as lamp ignition, limitation of the lamp
current and compensation are integrated into the ECG. Most Electronic
Control Gear are also suitable for DC operation, which means they can be
used in emergency lighting installations.
T5/ ∅ 16 mm fluorescent lamps FH
®
…HE, FQ®…HO and FC® can only be
operated by Electronic Control Gear.
If fluorescent lamps are operated with magnetic ballasts (principle of
magnetic coil, CCG and also low loss ballast), the lamp current equals the
frequency of the mains voltage. The resulting stroboscopic effect can be
dangerous in cases where people work with rotary machines. Every time,
the voltage goes through zero, the lamp current does the same until the
lamp is reignited: for every lamp ignition new carriers for the electric charge
have to be build up within the gas discharge.
CALM AND
LOWER
LOWER
ENERGY
ENERGY
CONSUMPTION
CONSUMPTION
(25 –30%)
(25 – 30%)
CALM AND
FLICKERFREE
FLICKERFREE
LIGHT
LIGHT
OPERATION WITHOUT
OPERATION WITHOUT
NOISE
NOISE
LOW MAGNETIC
LOW MAGNETIC
STRAYFIELD
STRAYFIELD
FLICKERFREE
FLICKERFREE
START
START
1.6 Saving Energy with
Electronic Control Gear
LONGER
LONGER
LAMP LIFE
LAMP LIFE
(approx. 50%)
(approx. 50%)
LESS WASTE
LESS WASTE
DISPOSAL
DISPOSAL
(approx. 30%)
(approx. 30%)
ENERGY SAVING
ENERGY SAVING
(25 –30%)
(25 – 30%)
LONGER LAMP LIFE
LONGER LAMP LIFE
(approx. + 50%)
(approx. + 50%)
SWITCH OFF AT END
SWITCH OFF AT END
LOW
LOW
WIRING COSTS
WIRING COSTS
DIMMABLE
DIMMABLE
(spec. version)
(spec. version)
AUTOMATIC
AUTOMATIC
OF LAMP LIFE
OF LAMP LIFE
When using Electronic Control Gear the frequency of the lamp voltage is
approx. 1000 times higher than the line voltage. The zero of the lamp
current are passed so quickly that the average of the value of the electron
density is nearly constant within the discharge plasma. The electrons don’t
have to be built up with every cycle (as it is necessary when using CCG
and low loss ballasts). So the limitation of the lamp life due to reignition
peaks for CCG operation are avoided with ECG operation. Therefore no
stroboscopic effects can occur by using high frequency control gear as
there is no longer a gap in the lamp current. Therefore, one lamp type
needs less energy to generate the same lumens when operated with high
frequency control gear compared to operation with magnetic ballasts. The
lower energy consumption reduces the lamp load and increases the lamp
9
1.7 Ignition of Fluorescent
Lamps
1.8 Costs and Safety
1.9 Flexibility upon
Request
1.10 ECG bring Progress
1.11 The right ECG for every
Application
life. Electronic Control Gear improve the efficiency and the lamp life of
fluorescent lamps significantly.
Prior to ignition, modern ECG heat the cathode to its optimum temperature
for electron emission. After a defined period the lamp is ignited with the
required ignition voltage. Only an optimized preheat start can guarantee
that the number of switching cycles has only little effect on the lamp life.
This is another important feature of ECG which has a positive effect on the
cost of operation and which should not be neglected when looking for
alternatives to CCG.
At the end of lamp life the emitter paste applied to the lamp electrode is
used up. The complete loss of emitter results in an increase of voltage in
the vicinity of the electrode. This situation can last over a longer period of
time. As an immediate result of the accompanying temperature increase at
the lamp ends the lamp sockets may overheat. Modern ECG are able to
detect this malfunction and switch the lamps off. Unnecessary attempts to
ignite are avoided by an interrupting function and therefore, also
overheating is avoided - an important contribution to more safety.
Professional ECG control all parameters constantly. A safety shut down at
the end of the lamp life is mandatory from January 1
st
, 2007 on for all ECG
that operate T4 or T5 tubes as it is included in the IEC 61347
(Omnibusnorm for safety of Electronic Control Gear). For several years
now, all OSRAM QUICKTRONIC
®
fulfill the safety requirements acc. the
IEC 61347 already.
However as there was no Standard for this before, some ECG
manufacturer neglect this due to costs.
During past years, we see a clear increase in new, better and more energy
efficient lamp systems. Unfortunately, this resulted also in a growing
number of various ECG-types. To reduce this large number of types
manufacturers of ECG have taken a new direction and have developed
new multi-purpose ECG which can be used for a variety of fluorescent
lamps of different wattages. New integrated circuits allow the optimum
control of lamp features such as lumen output. This type reduction has, of
course, a big effect on the customer: ordering, warehousing and installation
of only a few ECG-types. The so-called MULTIWATT-ECG reduce all
relevant cost drivers.
In addition to the basic tasks of lamp operation which are also fulfilled by
magnetic ballasts, Electronic Control Gear have much more to offer: They
have better performance and are more reliable, more environmental
friendly and more practical than CCG; even more reasons to use
professional Electronic Control Gear.
OSRAM offers the right Electronic ballast for every application as shown at
the example of T8/ ∅ 26 mm fluorescent lamps.
10
1.12 OSRAM ECG
Milestones
Burning hours per day
Burning hours per day
24
24
®
QUICKTRONIC
QUICKTRONIC
QUICKTRONIC
20
20
Industry,
Industry,
Openspace office
Open space office
16
16
®
Railway
Railway
station,
station,
airport
airport
®
QUICKTRONIC
QUICKTRONIC
INSTANT START QTISe
INSTANT START QTIS e
12
12
Departmentstore,
Department store,
8
8
display
display
4
4
average
average
ECGs
ECGs
0
0
02468
024 6 8
QUICKTRONIC
®
DIMMABLE
DIMMABLE
®
®
PROFESSIONAL
PROFESSIONAL
QUICKTRONIC
QUICKTRONIC
ECONOMIC
ECONOMIC
Daylight
Daylight
linked
linked
illumination
illumination
Switching cycles per day
Switching cycles per day
®
®
Daylight linked
Daylight linked
illumination with
illumination with
presence detectors
presence detectors
• For the first time in 1995, T5- fluorescent lamp systems with Cut-
off-technology have been introduced to the market.
Cut-off technology is the cut-off of the permanent filament
preheating after lamp ignition. This can be realized due to
modifications in the electronic circuit of the ECG. The result of the
Cut-off technology are less losses and optimized lamp operation.
• Four years later, in 1999, OSRAM sold the first reliable
MULTIWATT-ECG.
This operates all lamps with rated data.
• During the following years the trend of miniaturization continued and
the height of Electronic Control Gear was reduced by 30% from
30 mm to 21 mm.
In 2002, OSRAM is again the first manufacturer to introduce
MULTIWATT-ECG to operate FQ
®
…HO-fluorescent lamps High
Output in 21 mm height.
• In 2003, another novelty is brought to the T5-product segment: As
the first producer, OSRAM offers a 21 mm high 2-lamp ECG for
®
FQ
80 W HO-fluorescent lamps.
• In the beginning of 2004, the newest and most innovative member
of the T5-product family has been introduced: micro-controller
based ECG capable of operating T5-fluorescent lamps of equal
length regardless if it is a FH
®
…HE- or FQ®…HO-type. This microcontroller especially developed under the co-operation with OSRAM
is responsible for clear lamp detection and lamp operation with
nominal data.
QUICKTRONIC INTELLIGENT, QTi make one MULTIWATT-ECG
possible for all T5-fluorescent lamps from 14 to 39 W no matter if
®
FH
…HE- or FQ®…HO-types.
11
2. Product Features
2.1 Lighting Comfort
2.2 Economy
2.3 Safety
• Flicker-free ignition
• Pleasant, flicker-free light with no stroboscopic effects due to high
frequency operation
• High comfort level with no distracting choke hum due to fully
electronic operation (see chapter 2.8 noise)
• No flickering
• No flashing or flickering, electronic defective control for
reliable safety switch-off of defective lamps
End-of-life safety shut down
• Cut-off of the permanent filament preheating after lamp
ignition
• Automatic restart after lamp replacement
• High lumen packages for T5 FQ
• Very high luminous efficacy for T5 FH
®
High Output system
®
High Efficiency system
• Long lamp life due to lamp start with optimum filament pre-heating
and operation with cut-off technology
• Low maintenance costs due to long lamp life and reduced relamping
intervals
• Lower cooling load of air-conditioning systems due to lower losses
Light engineering with T5 (Ø 16 mm) lamps
Leuchte/1xL58 W1xL58 W1xFH35 W
Luminaire
Vorschaltgerät/VVG/EVG/EVG/
Control GearLLGECGECG
E [lux]539518500
P
gesamt/total
[W]260220154
%100 %85 %59 %
Büro mit 4 1-flammigen Leuchten
Office with 4 single tube luminaires
4
3
2
1
0
012344,5 m
W/m
2
15129
All Electronic Control Gear QUICKTRONIC
®
for operation of T5/∅ 16 mm-
fluorescent lamp systems are developed and designed according to all
relevant national and international industry standards.
Current standard is EN 61347. For Electronic Control Gear for operation of
low pressure gas discharge lamps EN 61347-2-3 applies.
12
2.4 Supply Voltage
2.4.1 Overvoltage and its
Reason
In detail:
• Safe shutdown of the power supply to defective lamps or
at the end-of-life due to End-of-Life detection according to Test 2
Shutdown in the event of broken filaments, no inserted lamp or air
leakage
• Compliance with European safety standards (EN 61347-2-3)
• Protection against short duration voltage surges (DIN VDE 0160)
and transient overvoltages
• Low housing temperatures allow the mechanical design of lighting
fittings with F- and FF- as well as M- and MM-approval mark
(EN 60598/DIN VDE 0710 and DIN VDE 0711)
• Can be used in emergency lighting systems according to
DIN VDE 0108
Electronic Control Gear QUICKTRONIC
lamps (FH
®
…HE, FQ®…HO and FC®) can be operated on sinusoidal AC
®
for T5/∅ 16 mm-fluorescent
voltage and DC voltage The recommended voltage intervals depend on the
design of the specific circuits.
The following chapters show the recommended voltage ranges and the
behaviour of the ECG at overvoltage and undervoltage.
It is called an overvoltage if the ingoing voltage is significantly higher
than the nominal value.
In general, we have to differentiate between two overvoltages which
also can have different reasons.
1. Transient overvoltage with a typical duration of milliseconds.
This overvoltage can be caused by:
- Switching of inductive loads such as welding machines, elevators
alternators etc.
- lightening
Quasi-stationary overvoltage with a duration from a few minutes to hours.
This overvoltage can be caused by:
- different loads on the mains side (interruption of the neutral
conductor in 3-phase installations plus an additional asymmetric
13
2.4.2 Possible Implications
due to Overvoltage
2.4.3 Undervoltage and its
Reason
2.4.4 Possible Implications
due to Undervoltage
2.4.5 Supply Voltage
QT with 21 mm height
load distribution)
- unstable power supply (for example some countries in Far East)
It is called overvoltage, if the supply voltage exceeds the specified voltage
range of an ECG including tolerances.
In any case, overload means more stress to electronic components.
Depending on the magnitude of overvoltage the protective functions of an
ECG can come into effect and turn the ballast off.
In extreme situations overvoltages can even destroy electronic
components.
Therefore, please pay attention to the design of the mains and tolerances
of the Electronic Control Gear when using them.
Supply voltages can not only deviate to higher values but also to
lower values. If the supply voltage decreases below the value
specified in the technical data of an ECG, we have to deal with
undervoltage.
This may be true for the following points:
• Different loads on the mains side
• Incorrect electric installation
• Unstable power supply
• In some cases when used with emergency generators
Operating ECG with undervoltage is not as specified. This may result
in the following implications:
• Lamp operation not according to standards Î affecting lamp life
• No safe lamp start, a safe ignition is only guaranteed above supply
voltages of 198 V
• Unstable lamp operation meaning the discharge process of a
fluorescent lamp is not stable
• In order to keep the lamp wattage constant most ECG types are
controlled on the lamp side. In this case, reduced supply voltages
cause much higher currents which may lead to physical stress of
components and to failure of the entire ECG. If supply voltages
deviate significantly from the nominal values, high switching losses
and overload of transistors can occur finally causing ballast failures.
Valid for: QTi and QT…F/CW
Recommended voltage range for
continuous operation
AC voltage 198 V ... 264 V, 50/60 Hz
DC voltage 176 V ... 264 V
Performance at undervoltage
Lamp ignition
Voltage drop during operation
U
≥ 198 V Î reliable lamp ignition
N
U
≥ 176V Î operation possible
N
U
< 176 VÎ damage to ECG
N
possible
14
2.4.6 Supply Voltage for QT
with 30 mm height
2.4.7 ECG for 120V / 277V
Line Voltage
2.5 Automatic Lamp
Restart
Valid for: QT-FH MULTIWATT and QT-FQ
Recommended voltage range for continuous operation
AC voltage 198V ... 264V, 50/60 Hz
DC voltage 176V ... 264V
Performance at undervoltage
Lamp ignition
U
≥ 198V Î reliable lamp start
N
Voltage drop during operation
U
≥ 176V Î operation possible
N
U
< 176V Î damage to ECG
N
possible
T5/∅ 16 mm fluorescent lamps are also getting more popular in North
America (USA, Canada). Historically in the US-market have been
established lamps in 4 ft length besides the types of 240 cm. 4 ft is also
known as 48 inch type (1 ft = 30.48 cm) and is acc. to our typical 120 cm
types. Related to the straight fluorescent types FH
this means 1,149 mm for FH
®
28 W HE and FQ® 54 W HO.
®
…HE and FQ®…HO
OSRAM SYLVANIA offers the complete range for FH
®
…HE and FQ®…HO
under PENTRON ECG. The specification there is PENTRON High
Performance T5 lamps for FH
®
and PENTRON High Output T5 for FQ®
lamps.
OSRAM SYLVANIA also offers the ECG for the North American line
voltages 120 V / 277 V und 60 Hz line frequency as shown at a glance:
MULTIWATT ECG for FH
®
fluorescent lamps: 14, 21, 28 and 35 W HE
Types: QTP 1x28T5/UNV PSN suitable for 120-277 V
QTP 2x28T5/UNV PSN suitable for 120-277 V
ECG to operate FQ
®
54 W HO
Types: QTP 1x54T5UNV/PSN suitable for 120-277 V
QTP 2x54T5UNV/PSN suitable for 120-277 V
For a large number of differnt lamp types including T8 OCTRON a variety
of dimmable and non dimmable types is available.
Information about availability under:
OSRAM LIGHT CONSULTING (OLC)
Hellabrunner Straße 1
81536 München
Tel: +49-89-6213 3076 Fax: +49-89-6213 2020
With all QUICKTRONIC
lamps FH
®
…HE, FQ®…HO and FC®, automatic restart takes place after
®
for operation of T5/∅ 16 mm-fluorescent
a change of lamp provided the power supply is maintained.
Should in the case of a twin-lamp ECG no automatic lamp restart take
place after lamp replacement and could an ECG-failure be excluded,
please proceed as follows:
15
2.5.1 Lamp ignition for QTi
2.5.2 Lamp ignition for QT
to operate T5fluorescent lamps
2.5.3 Off- Time for
Optimum Preheat
Start
2.6 Behaviour in Response
to Lamp Defects
2.6.1 One-Lamp Operation
with Multi-Lamp ECG
Replace both lamps, take out the lamp replaced first and refit it. Provided
lamp and ECG are o.k. both lamps should then light.
Lamp start Preheat
Ignition time < 1 second
Max. number of switching cycles > 100,000 cycles
QT-FH MULTI, QT-FQ, QT…F/CW
Lamp start Preheat
Ignition time < 0.5 second
Max. number of switching cycles > 100,000 cycles
All QUICKTRONIC to operate T5/∅ 16 mm- fluorescent lamps
®
FH
…HE, FQ®…HO and FC start the lamps at any time with optimum
preheat start even after a turn-off followed by an immediate lamp
restart. OSRAM QUICKTRONIC ignite the lamp always with optimum
preheating of the electrodes. A particular off-time with regards to
lamp life is not necessary.
What do we mean by lamp defect or end-of-lamp life?
In most cases, it is not possible to see from outside which lampholders are
assigned to which ECG-terminals, so if you insert lamps and they fail to
start automatically you should take out the first lamp again and refit it. Both
lamps should then light.
Lamp replacement of 2- and multilamp luminaires proceed as follows:
Insert the lamps. If at 2- or multilamp luminaires lamp ignition doesn’t work
automatically, take out the lamp replaced first and refit it. Reignition of both
lamps works automatically.
What are the requirements?
QTP 2x...
QT - FQ 2x...
QT - FQ 2x...
L
ILI
UU
Start
U
U
Zünd
Start
L
L
I
ILI
L
ILI
UU
Start
L
L
I
ILI
Parallel Circuitry Series Circuitry
16
2.7 Noise
• Parallel circuit of lamps operated with multi-lamp ECG ≠in general
single-lamp operation possible
• Parallel circuit of lamps, but no single-lamp operation possible because
for example
- the sum of electrodes has to be recognized
For twin- and multi-lamp ECG the question is whether the remaining lamps
will continue to operate if one lamp is defect or has been removed.
In the case of twin- or multi-lamp ECG, any lamp fault that causes the safe
shutdown circuit to operate will lead to the shutdown of all lamps.
This function is called “safety shutdown”. The detection of various “out-ofrange” parameters results in a reliable shutdown of the ECG. The ECG do
not perform any lamp starts that could cause problems as described under
chapter 2.3.
In this case, one lamp or the remaining lamps will therefore never continue
to burn by itself.
What happens when one lamp is removed from a multi-lamp ECG will
depend on the type of circuit. Series circuits always exclude a single-lamp
operation. Parallel circuit is one condition for a possible single-lamp
operation, however, not the only one. Another important criterion is lamp
control during operation of circuit related as well as safety related data.
QUICKTRONIC
®
INTELLIGENT, QTi, are carried out in parallel circuits, but
cannot be operated in single-lamp mode. The reason is the sophisticated
lamp detection requiring the control of various parameters.
The following table gives a short summary of the different ECG-types:
ECG-type
height
QTi
21 mm
QT-FH
30 mm
QT-FQ
30 mm
QT-FH 3x, 4x
30 mm
QT … F/CW
21 mm
Series circuit X X X X
Parallel circuit. X
For all types shown in the table above a single-lamp operation is not
possible.
T5/∅ 16 mm-fluorescent lamps FH
at high frequency with QUICKTRONIC
®
…HE, FQ®…HO and FC® operated
®
control gear are virtually
silent.
QUICKTRONIC
®
units themselves are so quiet that even in very quiet
surroundings they cannot be heard by the human ear. They are
therefore ideal for sound-sensitive areas such as radio and recording
studios. The limit of the frequency-dependant sound pressure curve
is based on the audibility threshold (in other words, a person with
normal hearing will not be able to detect the noise generated by an
ECG in the same room).
The factors affecting the sound pressure level are the sound power level of
the ECG, the absorption properties of the room, characterised by its volume
and reverberation time, and the number of ECGs.
In mains supplies with a high level of distortion in which the mains voltage
wave form deviates significantly from a sine wave, a „chirping“ may be
heard from the reactance coils in the input section of the ECG.
17
2.8 Power Factor λ For all electric loads, the power factor λ is the ratio of effective power (P
voltage x effective current) to apparent power (P
= voltage x apparent
app
current). This value is affected both by the phase displacement cos ϕ
between current and voltage by the current wave form distortion ε (nonsinusoidal wave form)
=
eff
λ = P
eff
/ P
= ε cos ϕ
app
In contrast to conventional control gear (CCG, inductive, 50 Hz), there is
hardly any phase displacement with Electronic Control Gear (high
frequency), which means that capacitor correction is not required. However
slight distortions in the current sine-wave curve occur during operation of
electronic control gear. In general these distortions are characterized by
integer multiples of the mains frequency (harmonics).
The harmonic content of the mains current is controlled by national and
international regulations (IEC 61000-3-2, EN 61000-3-2). OSRAM ECG
have built-in active electronic harmonic filters for this purpose which
guarantee a value for ε of more than 0.95 and hence a power factor λ
greater than 0.95.
Exemptions are ECG which apply to the international standard for system
power consumption less than 25 W. This standard requires a power factor
λ of more than 0.6. These ECG are part of the product segment ECO and
are known as QUICKTRONIC
®
ECONOMIC or QT-ECO. They are mainly
used to replace conventional control gear, but because of their
MULTIWATT-design they can partly operate FH
®
…HE- and FQ®…HO
fluorescent lamps with lower wattages:
®
FH
14W HE
®
FH
21W HE
®
FQ
24W HO
For detailed information about this combination please see
www.OSRAM.de/products/ecg
With regards to their harmonics content all QUICKTRONIC
®
have been
tested by VDE according to EN 61000-3-2 and carry the VDE-EMC
approval mark.
The confirmation of the ECG’s CE-mark by an independent testing facility
reduce the costs and the t.ime necessary for approval of luminaires
significantly.
18
2.9 ECG Imprint
2.10 Reliability
2.11 Resistance to Frequent
Switching
2.12 ECG Lifetime
Functional Earth terminal
Functional Earth terminal
Functional Earth terminal
End-of-Life Safety-shut-
End-of-Life Safety-shut-
End-of-Life Safety-shutdown
down
down
Cut-Off-Technology
Cut-Off-Technology
Cut-Off-Technology
µProzessor inside
µProzessor inside
µProzessor inside
Thermal Behaviour
Thermal Behaviour
Lamp wiring including
Lamp wiring including
Lamp wiring including
max. cable lengths
max. cable lengths
max. cable lengths
Besides component specification and quality their failure rate is significantly
related to the operating temperature.
Electronic Control Gear of OSRAM are designed in that way, that a failure
rate of less than 2 Promille per 1,000 operating hours is expected if
operation takes place at the maximum permitted case temperatur (t
).
c
The resistance to frequent switching of Electronic control gear is based on
possible lamp starts per day. Multiplied with the lamp life professional ECG
with preheat start reach more than 100.000 switching cycles.
This information is important for combinations with occupancy sensors as
this is one of the most popular applications for frequent switching of the
lamp-ECG system.
The ECG lifetime depends on the operating temperature and failure rate of
the electronic components. Extreme overheating can destroy electronic
components in a short period of time and cause the ECG to fail. There
exists an exponential relationship between the failure rate of electronic
components and their thermal and also electrical behaviour.
An indication about the maximum recommended ambient temperature of a
luminaire is given by the imprinted measuring point tc at which each ECG
reaches its maximum recommended case temperature. The t
-temperature
c
of an OSRAM ECG is closely linked to its temperature of electronic
components. For example, when the maximum recommended tctemperature of 70 °C is reached, a QUICKTRONIC control gear for
operation of T5/∅ 16 mm-fluorescent lamps is expected to last
50,000 hours with a failure rate of max. 10 %. This value equals a failure
rate of 2 ‰ per 1,000 operating hours. Due to the exponential dependence
on temperature and failure rate of electronic components exceeding the
recommended tc-temperature reduces the ECG lifetime dramatically. On
the other hand, if the ECG temperature remains below the limit the lifetime
is extended. As a rule of thumb, every 10 °C below the imprinted
temperature value double the ECG’s lifetime and every 10 °C surpassing
the tc-value cut the service life in half.
The measuring temperature t
is an important parameter for both the safety
c
approval for a luminaire according to EN 60598 and the service life of an
19
2.13 Thermal influences of
the system
components
2.14 cut-off Technology
ECG provided by the manufacturer under consideration of the thermal load
of electronic components.
Surviving ECGs[%]
Surviving ECGs [%]
Surviving ECGs [%]
100
100
100
80
80
80
60
60
60
90°C
90°C
40
40
40
20
20
20
90°C
0
0
0
020406080100120140
020406080100120140
020406080100120140
70°C
70°C
70°C
10°Cloweroperating temperatureat point of
10°Cloweroperating temperatureat point of
10°C lower operating temperature at point of
measurement virtually halvesECGfailurerate
measurement virtually halvesECGfailurerate
measurement virtually halves ECG failure rate
The temperature must be assessed separately for the two system
components (ECG and lamp). In the case of the lamp, there are physical
laws that restrict the temperature range of an application, whereas in the
case of the ECG fixed limits must be set in order to ensure reliable
operation.
Apart from this, there are external factors such as the reciprocal influences
of ECG, lamp and luminaire and the selected installation site which have an
influence. Compliance with the specified limits and hence the guarantee of
operational reliability are the responsibility of the relevant luminaire or
system manufacturer.
There is a fixed correlation between tc-temperature, the temperatrure of
electronic components, the life of each component and hence the complete
unit. Thermal contact of an ECG to metallic parts of luminaire housings is
very positive due to good thermal conductivity.
The correlation between temperature tc of the measuring point, component
life and failure rate is crucial for an objective assessment of the reliability
and service life of an ECG. Information about tc-temperature and ECG
service life alone are not sufficient.
All QUICKTRONIC
T5/∅ 16 mm-fluorescent lamps FH
are equipped with cut-off technology.
After starting the lamp the electrode heating is being switched off. Lamp life
increases due to the reduced load of the electrodes. Therefore, cut-off
technology increases the lumen output of the luminaire and the lamp life.
And for the first time, cut-off technology is included in dimmable ECG
thanks to the new intelligent technology of QTi. Compared to Electronic
control gear without cut-off the power consumption of ECG with cut-off
technology could be reduced by another 5 to 7 %.
10 %lossat 50,000hours
10 % loss at 50,000 hours
10 % loss at 50,000 hours
60°C
60°C
60°C
Hoursofusage[thsd]
Hours of usage [thsd]
Hours of usage [thsd]
®
control gear for operation of
®
…HE and FQ®…HO
50°C
50°C
50°C
Temp. at point of
Temp. at point of
Temp. at point of
measurementt
measurement t
measurement t
c
c
c
20
2.14.1 Advantages for Users
The following advantages for users arise from cut-off technology:
6-10 % higher luminaire efficiency
highest lamp life
2-3 W lower losses per lamp
reduced load of air condition
New circuitrywihtoutpermanent filamentheating
New circuitry wihtout permanent filament heating(cut-off technology)
(cut-off technology)
I
= I
I
= I
Stift
Lampe
Stift
Lampe
Cold Spot
Cold Spot
app. 40°C
app. 40°C
Significantlylowertemperaturesat the
Significantly lower temperatures at the
lampelectrodes
lamp electrodes
I
I
Lampe
Lampe
I
I
Stift
Stift
Cut-off advantage for the luminaire
Cut-off advantage for the luminaire
Cut-off advantage for the luminaire
110
110
110
100
100
100
90
90
90
rel.
rel.
rel.
80
80
80
Licht-
Licht-
Lichtstrom
strom
strom
70
70
70
60
60
60
1020304050
1020304050
10 20 30 40 50
Umgebungstemperatur (°C)
Umgebungstemperatur (°C)
Umgebungstemperatur (°C)
Onlycut-off technology canfulfilllifetime
Only cut-off technology can fulfill lifetime
100%
100%
90%
90%80%
80%70%
70%60%
60%
Rel.
Rel.
50%
50%
Lum-
Lum-
40%
40%
inous
35°C
35°C
35°C
Cut Offk
Cut Offk
Cut Offk
Conventional
Conventional
Conventional
ECG circuitry
ECG circuitry
ECG circuitry
inous
flux
flux
6 –10% higher
6 –10% higher
6 – 10% higher
luminaire efficiencyfor
luminaire efficiencyfor
luminaire efficiency for
directlighting
directlighting
direct lighting
5.000h
5.000h
Lifetime [hours]
Lifetime [hours]
Cut-off
Cut-off
Cut-off
Conventional
Conventional
Conventional
ECG circuitry
ECG circuitry
ECG circuitry
10.000h
10.000h
12.000h
12.000h
13.16.
13.16.
16.000h
16.000h
14.000h
14.000h
16.20.
16.2 0.
20.000h
20.000h
18.000h
18.000h
2.14.2 Physical Background
T5/∅ 16 mm-fluorescent lamps FH
®
…HE and FQ®…HO are designed to
have their maximum lumen output at 35 °C (compared to 25 °C for
T8/∅ 26 mm). For T5/∅ 16 mm-fluorescent lamps the so-called cold spot
(the point where mercury condensates in a discharge tube, stamped end of
the lamp) is located behind the electrode (see graphics) which means near
the source of heat.
T8 (Ø 26 mm)
T8 (Ø 26 mm)
T12 (Ø 38 mm)
T12 (Ø 38 mm)
Cold Spot T
Cold Spot T
The value of the luminous flux at the ambient temperature of 35 °C is only
informative for the luminaire manufacturer. Significant is the value of the
cold spot temperature. This value is measured at the socket of the stamped
side, approx. in a distance of 2 mm of the glass. For an optimized luminous
flux this value should be between 45 °C and 50 °C. This is shown at the so
called ‘Horseshoe curves’ where the luminous flux is shown in relation to
the lamp ambient temperature.
The cold spot of the T5/∅ 16 mm fluorescent lamps is influenced by
permanent filament heating. This means that higher temperatures reduce
the luminous flux. ECG with cut-off technology reduce the losses of the
system and are optimized regarding the maximum luminous flux of the
Hg opt.
Hg opt.
≈ 40°C
≈ 40°C
T5 (Ø 16 mm)
T5 (Ø 16 mm)
21
2.15 End of Life
(EoL acc. to T.2)
2.16 U-OUT
system. The cut-off of the permanent filament heating after lamp ignition is
an advantage.
Further the values of the horseshoe curves also indicate the arrangement
of the lamps within the fixture. To avoid thermal influences of the lamps
minimum distances in between have to be kept. The lamps have to be
placed that the stamp of all lamps is on the same side. For vertical
arrangement the stamp of the lamp always should be placed down. For
circular lamps FC the socket has to be placed down.
The measurement of the cold spot temperature is especially important for
the luminaire manufacturer. This temperature offers opportunities to
optimize the luminaire efficacy.
Fluorescent lamps use up their emitter during operation. The complete loss
of emitter on an electrode results in a voltage increase in the vicinity of the
electrode filament. As most Electronic control gear have no problem
providing this high asymmetric voltage and with it the required additional
power the temperature around the electrodes rises significantly.
At the moment the international ECG safety standard IEC 61347-2-3 is
under revision. In the future, all ECG must provide a “end-of-life” safety
shutdown which is continuously controlled in order to avoid overheating of
lamp sockets.
The actual status of the standard considers three different test circuits for
Electronic control gear. If an ECG complies with one of the three tests, it
offers protection against the “end-of-life” behavior of fluorescent lamps.
• Asymmetric pulse test (Test 1)
• Asymmetric power test (Test 2)
• Open filament test (Test 3)
The asymmetric power test (Test 2) is directly simulating the additional load
which results from the increased asymmetric voltage in the case of “end-oflife”. In order to pass the test the additional load may nod exceed a specific
value depending on the lamp type. Most ECG experts see test 2
(asymmetric power test) as the safest “end-of-life” control, because the
direct measurement of the asymmetric additional load mirrors the real lamp
behavior at its end of life. OSRAM does not compromise the safety of
Electronic control gear and has, for quite some time, specified Test no. 2 as
standard test.
U-OUT a binding ECG label according to safety standard EN 61347-2-
3. The former standard EN 60928, still valid until 2006, allows labeling
of U-OUT either in the product catalogue or on the ECG housing. UOUT specifies the largest effective working voltage between
- Lamp terminals
- Each lamp terminal and earth connection, if applicable
This information is important for all components electrically connected on
the lamp side of the ECG.
All components such as lamp cables, sockets (EN 60061-2), isolating
22
2.17 Approval Marks
2.17.1 ENEC-Approval Mark
material and everything coming in contact with the ECG lamp terminals
must be layed out according to U-OUT.
OSRAM, as manufacturer, takes care that no higher voltage appears at the
lamp terminals than specified by U-OUT. Therefore no additional voltage
reserve is needed.
2.17.1.1 Safety acc. to
EN 61347
2.17.1.2 Performance acc. to
EN 60929
2.17.2 VDE EMC mark
stands for E
also a conformity mark agreed upon between the testing institutes of the
European union.
It stands for compliance with the according European standards for safety
and performance. Besides sample testing ENEC includes also a permanent
control od products and production processes. This certification is
testimony of an independent and competent institute testing the safety and
performance of Electronic control gear. The number right beside the
approval mark identifies the certifying institute. For example 10 is
representing VDE in Germany.
The ENEC approval mark for ECG to operate fluorescent lamps includes
the safety standard EN 61347 and the performance standard EN 60929.
This standard contains the safety requirements of electric units for
operation of lamps that are designed for DC- and AC-voltage at 50 or 60
Hz. It is divided into different parts.
The first part EN 61347-1 deals with general and safety requirements.
b) Electronic control gear to operate with AC-voltage at 50 or 60 Hz
This safety standard whose part EN 61347-2-3 together with the general
part EN 61347-1 succeeds the former standard EN 60928, is also called
„omnibus“ standard.
This standard specifies the performance of Electronic control gear for
fluorescent lamps. It defines the operation at AC-voltage, at 50 or 60 Hz
and with a supply frequency different from the operating frequency. It is
based on performance standards for fluorescent lamps EN 60081 and EN
60901.
uropean Norm Electrical Certification. The ENEC approval is
with an operating frequency different from the frequency of the
mains supply and to operate fluorescent lamps according to IEC
60081 and IEC 60901 and other fluorescent lamps for high
frequency operation are dealt with in part EN 61347-2-3.
The independent testing institute confirms the compliance of the ECG with
the EMC regulation regarding immunity, radio interference suppression and
harmonics. At the same time, it is also a confirmation for the CE label that
can be applied to ECGs by the manufacturer himself under compliance with
the EMC regulation. Luminaire manufacturer can significantly reduce their
costs and approval efforts with regards to EMC compliance by using
already EMC approved ECGs.
23
2.18 Energie Efficiency
Index EEI
2.19 CE Labelling
This label helps consumers identifying the energy consumption of a
product. Usually, all Electronic control gear have the best ratings A2 …A3.
Dimmable ECG are classified as A1. Magnetic ballasts (CCG) fall under the
energy efficiency class C and D and are either already banned from the
market or are about to be banned shortly.
Low loss ballasts are usually classified in B.
Since January 1996, all products falling under the EU directive of
electromagnetic compatibility (EMC) must carry the CE label. The CE label
indicates the compliance with the requirements of this directive. From
January 1997 all products falling under the Low-Voltage directive must also
be CE labeled. There is no question that our products comply with the
specific EU directives and therefore are labeled with the CE mark.
Regarding CE labelling here the following explanations:
1. CE-label as basis to introduce product to the market
Since 01.01.1996 manufacturers and importers are obliged to apply
CE labels to products that have to comply with EMC regulations
either directly on the product, its packaging or the accompanying
documents. CE labels are obligatory for the sale of products within
the European Union. By applying the CE label manufacturers or
importers confirm that their products comply with the “basic
requirements” of specific European directives and fulfill their
protective goals (for example electromagnetic compatibility). Usually
the compliance of particular “basic requirements” is given products
are manufactured under consideration of harmonized European
standards.
2. CE label is a mark for administrative authorities
The CE label is targeted primarily at administrative authorities. It
declares to them that CE labeled products comply with European
jurisdiction at the time of sale.
3. No rights for commerce and endusers of examining declarations of
conformity issued by manufacturers.
The right to ask for and examine declarations of conformity is
reserved to authorities controlling the compliance of
electric/electronic products with the legal safety requirements. In
Germany it is the “Federal Agency for Post and Telecommunication”
BAPT (responsible with regards to the EMC directive) and the trade
supervisory boards (responsible with regards to the Low-Voltage
directive).
4. CE labels are no quality or approval marks
The only purpose of the CE label is to testify that a product complies
with legally specified „basic requirements“ of certain directives. It
24
2.20 CCC Approval
does not make any statement with regards to the quality of labeled
products. As a legal administration label without value for endusers
it should not be mistaken as an approval mark issued by
independent testing institutes (such as ENEC or VDE mark). These
testing institutes do also not control the legitimacy of an applied CE
label.
Approval mark of the Chinese testing institute CQC (C
C
enter).
hina Qualification
Since 01.08.2003 this approval mark is required in order to sell Electronic
control gear for operation of low pressure discharge lamps in the Chinese
market.
OSRAM QUICKTRONIC for operation of T5/∅ 16 mm- fluorescent lamps
®
FH
...HE and FQ®...HO carry this approval mark.
25
3. ECG installed in Luminaire:
Installations and Operation
Instructions
3.1 Wiring Instructions
3.1.1 Cable Types
3.1.2 Cable Cross-Sections
Please pay attention to the voltage value U-OUT imprinted on the ECG
housing when wiring luminaires for FH
lamps. This value indicates the possible cable type.
For voltages greater than 430V cables with the classification H07 have to
be used.
U-OUT is the maximum voltage that can occur between the lamp terminals
or the lamp terminal and the earth connector.
For all OSRAM QUICKTRONIC ECG for operation of T5/∅ 16 mm-
fluorescent lamps FH
allowing luminaires to be wired with cables H05.
Cable types are specified through the terminals in use.
The cable cross-sections are marked on the identification plate of the
Electronic control gear.
C
ombi-Wiring (CW) stands for use in automatic or manual wiring. ECG in
30 mm height have the annex CW at the end of the type. ECG types
without CW are suitable for manual wiring only in this height. T5-ECG in
21 mm height don’t have this annex as they are equipped with CW
terminals only for manual or automatic wiring.
For manual wiring of the IDC a special tool is available for example from
WAGO. This tool is listed and can be ordered from WAGO under the order
number
0206-0831.
®
…HE and FQ®…HO U-OUT is less than 430V
®
…HE or FQ®…HO fluorescent
Ancillary for manual wiring of the IDC-contact of the CW terminals
26
3.1.2.1 ECG in 30 mm height
3.1.2.2 ECG in 21 mm height
3.1.3 Release of Contacts
3.1.3.1 WAGO 250
Typical values for C
a) Single-core cables
These should have a cross section of 0.5mm² at least and 1.5 mm²
at most for the horizontal plug.
When using I
have a maximum cross section of 0,5 mm²
b) Multi-core cables
horizontal plug
These should have a cross section of 0,5 mm² at least and 1 mm² at
most.
Multi-core cables can be inserted directly into the horizontal plug
terminals.
Ferrules may be used but they are not essential.
IDC
Multi-core cables with a cross section of 0,75 mm² can be used for
direct wiring with IDC.
Typical values for C
c) Single-core cables
These should have a cross section of 0,5 mm² at least and 1 mm² at
most for the horizontal plug.
When using I
have a maximum cross section of 0,5 mm²
d) Multi-core cables
horizontal plug
These should have a cross section of 0,5 mm² at least and 1 mm² at
most.
Multi-core cables can be inserted directly into the horizontal plug
terminals.
Ferrules may be used but they are not essential.
IDC
Multi-core cables with a cross section of 0,75 mm² can be used for
direct wiring with IDC.
ombi-Wiring terminals of ECG with 30 mm height are:
nsulation Displacement Contacts (IDC) cables should
ombi-Wiring terminals of ECG with 21 mm height are:
nsulation Displacement Contacts (IDC) cables should
3.1.3.2 WAGO 251 – IDC
Release the contact by using the release latch.
Release the contact by pulling the cable upwards.
This process can be repeated up to 10 times (depending on the
manufacturer) without damaging the terminal. For further details please
refer to the data sheets of the manufacturers.
27
Loading...
+ 65 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.