• 12 •
LipidPro
TM
T est Strip Information LipidPro
TM
T est Strip Information
LIMITATIONS OF SYSTEM (for lipid profile test):
Lipid Profile Test Strips provide accurate results when the following constraints
are observed:
• Do not use neonate samples
Physicians -
Please note the following interferences that may affect the results:
• Extremes in hematocrit may affect test results. Hematocrit levels less than 30
% may cause falsely high reading and hematocrit levels greater than 55 %
may cause falsely low reading. If you do not know your hematocrit level, consult your healthcare professionals.
• Interferences: Acetaminophen, uric acid, ascorbic acid (vitamin C), and other
reducing substances (when occurring in normal blood or normal therapeutic
concentrations) do not significantly affect results. However, abnormally high
concentrations in blood may cause inaccurately high results.
• Antiglycolysis and anticoagulants in blood samples may affect the test results
• Total cholesterol may be decreased by dopamine and gentisic acid.
• HDL-cholesterol may be decreased by dopamine.
• Triglyceride may be decreased by dopamine and Methyldopa.
Performance Characteristics:
The performance of the lipid profile test strips has been evaluated in laboratory
and in clinical tests (Please refer to strip instructions for use in for more details).
Testing Range: The test range of LipidPro
TM
‚ is 100 ~ 400 mg/dL (2.59-
10.36mmol/L) for total cholesterol, 25~80mg/dL (0.65- 2.08mmol/L) for HDL-
cholesterol and 70~600 mg/dL (0.80-0.86mmol/L) for Triglyceride.
LIMITATIONS OF SYSTEM (for Glucose test):
The LipidPro
TM
Glucose Test Strips provide accurate results when the following
constraints are observed:
• Do not use neonate samples.
• Inaccurate results may occur when in shock, hypotensive, hyperglycemic, or
hyperosmolar state, with or without ketosis.
• Glucose test strips used above altitudes of 10,000 feet will have an effect on
test results.
Physicians
- Please note the following interferences that may affect test results:
• Extremes in hematocrit may affect test results. Hematocrit levels less than 20
% may cause falsely high reading. Hematocrit levels greater than 60 % may
cause falsely low readings. If you do not know your hematocrit level, consult
your healthcare professionals.
• Interferences: Acetaminophen, Uric and, Ascorbic acid (Vitamin C), and other
reducing substances when occurring in normal blood or normal therapeutic
concentrations do not significantly affect results. However, abnormally high
concentrations in may cause inaccurately high results.
• Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 3000
mg/dl do not significantly affect the results. Values beyond these levels
should be interpreted with caution.
• Blood samples that contain a high concentration of dissolved oxygen may
lower the test result.
• EDTA containing tube is recommended as an anticoagulant tube.
Performance Characteristics:
The performance of the glucose test strips has been evaluated in laboratory and
in clinical tests (Please refer to strip instructions for use for more details).
Measurement Range: The measurement range of the LipidPro
TM
Glucose
Testing is 10 to 600 mg/dL (0.6-33.3mmol/L).