OSANG Healthcare ILM-0001A-RF User Manual

03 •
Dear LipidProTM lipid profile and glucose measuring system Owner
Thank you for your choosing the LipidProTM lipid profile and glucose measuring system. This instruction for use includes important informa- tion you need know about the LipidPro
TM
lipid profile and glucose measuring system. Please read itcarefully. Our goal for LipidProTM lipid profile and glucose measuring system is attributed to provide the best quality healthcare products coupled with superior customer service.
Indications For Use
LipidProTM system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only). LipidProTM system which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipopro- tein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL choles- terol values are reported only when triglycerides are ≤ 350 mg/dL; when triglycerides are > 350 mg/dL, calculated LDL-cholesterol are not reported.
Glucose measurement is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients or health care professionals as an aid in the management of diabetes. Glucose measurment is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternate site testing should be done during steady- state times when glucose is not changing rapidly.
LP-CL1-1B0-01 Rev. 2010-02
Catalog number: CL-LP1-1A0-01
TEL : 1-888-446-3246
• 02 • • 03 •
Dear LipidProTM lipid profile and glucose measuring system Owner
Thank you for your choosing the LipidProTM lipid profile and glucose measuring system. This instruction for use includes important informa­tion you need know about the LipidPro
TM
lipid profile and glucose measuring system. Please read itcarefully. Our goal for LipidProTM lipid profile and glucose measuring system is attributed to provide the best quality healthcare products coupled with superior customer service.
Indications For Use
LipidProTM system is intended for in home (self-testing) or health care professionals and for testing outside the body (in vitro diagnostic use only). LipidProTM system which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipopro­tein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL choles­terol values are reported only when triglycerides are ≤ 350 mg/dL; when triglycerides are > 350 mg/dL, calculated LDL-cholesterol are not reported.
Glucose measurement is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients or health care professionals as an aid in the management of diabetes. Glucose measurment is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Alternate site testing should be done during steady- state times when glucose is not changing rapidly.
• 04 • • 05 •
How to use Instructions for use
This guide includes all the information you need to use LipidProTM for self-testing use. For additional information and specific information for each individual test. Please make sure you read the strip instructions for use included with the test strips.
If you experience any difficulty or have any question, while using LipidPro
TM
Lipid
profile and glucose measuring system please, visit our website at
www.infopiausa.com
Customer Service is available 24 hours a day, 7 days a week, 365 days a year. Please Call Customer Support toll free : 1-888-446-3246
Test Principle
LipidProTM lipid profile and glucose measuring system has two kinds of test principle according to test items: one for Lipid profile and the other for glucose.
Lipid profile test result is based on a reading of reflection density. When the blood is applied, the color changes in test area through an enzyme reaction. The meter records this change in color and converts the measurement signal to the displayed result using the data previously entered via the code. The deeper the color is, the higher the lipid level is.
Glucose in the blood sample will react to the electrodes in the glucose test strip, generating an electrical current that will stimulate a chemical reaction. This reaction is measured by LipidPro
TM
meter and displayed as your blood glucose result. The enzyme glucose oxidase on the blood glucose test strip reads specifically with the blood glucse. The current generated is converted and displayed as blood glucose value. The LipidPro glucose test System is plasma-calibrated to allow easy comparison of results with laboratory methods. Blood glucose meter which is calibrated against a whole blood method may have different results in compari­son to LipidPro glucose test System. The laboratory system used for calibration of the LipidPro glucose test System is YSI 2300 STAT plus which is equipped with a glucuse oxidase system.
• 06 • • 07 •
LipidProTM lipid profile and glucose measuring system Kit Contents
1. LipidProTM Meter 5. Two AAA batteries
2. Instructions for Use 6. Lancing Device
3. Warranty Card 7. Lancet
4. Carrying Case
Your LipidProTM lipid profile and glucose measuring system has been sealed to protect the contents. If you find your seal has been broken, please return it to the place of purchase. Required but not provided
- LipidProTM Brand Test Strips
- LipidProTM Brand Control Solutions
- Capillary Rods
- Control Solution Range Cards
Contents
1. Learning the System
LipidProTM Testing System Kit Contents ----------------------------------- 7 LipidPro
TM
Testing Meter ------------------------------------------------------ 8 LipidProTM Testing Meter Display ------------------------------------------- 9 LipidProTM Test Strip Information ------------------------------------------- 10
2. Setting Your Meter
Battery Use --------------------------------------------------------------------- 18 Setting Mode ------------------------------------------------------------------- 19 Before Testing – Coding the meter ---------------------------------------- 21 hecking the System with Control Solution ------------------------------- 24
3. Testing Your Blood
Instruction for Lancing Device ---------------------------------------------- 29 Testing Your Blood – Preparation ------------------------------------------32 Quick reference of test procedure ----------------------------------------- 34 Test procedure -----------------------------------------------------------------36 Glucose test strip Discard Function ---------------------------------------44 Alternate site testing ----------------------------------------------------------45 Performing a Blood Test using an Alternate Site -----------------------46
4. Reviewing Your Results ---------------------------------------------------47
5. Deleting Your Results ------------------------------------------------------49
6. Printing Your Results ------------------------------------------------------- 49
7. LipidPro™ Software and Thermal printer ----------------------------50
8. Cleaning Your Meter and Maintenance -------------------------------51
9. Troubleshooting -------------------------------------------------------------52
10. Product Specifications ---------------------------------------------------55
11. Warranty ----------------------------------------------------------------------58
12. Declaration of Conformity -----------------------------------------------59
TC
HD
L
TG
MEM
• 08 • • 09 •
LipidProTM Testing Meter DisplayLipiProTM Testing Meter
1. Do not use LipidProTM meter in a dry environment, especially if synthetic materials are present. Synthetic clothes, carpets, etc., may cause damaging static discharges in a dry environment.
2. Do not use LipidProTM meter near cellular or cordless telephones, walkie talkies, garage door openers, radio transmitters, or other electronic or electrical equipment that are sources of electromag­netic radiation, as these may interfere with the proper operation of the meter.
Strip Connector
Glucose Strip measurement connector
Strip discard function
Strip discard function by push toward outside
LCD Screen
Display measurement level of glucose, TC, HDL, TG, inclukding time and message
Up button(ƞ(
Maneuvering button for choice of function in each mode
Down(Ơ( button
Maneuvering button for choice of function in each mode&RFID( ) operation button & Printer operation button
Mode button( ) Memory mode, Setting mode, Cholesterol measurement mode, On and Off button
Cholesterol Testing part
Cholesterol strip inserting part
Buzzer sound print part
Printing part for Warning sound, Fixed sound and buzzer sound
Battery Cover
Purpose for Battery replacement and fixed cover
Blood glucose
Total cholesterol
Triglyceride
HDL-cholesterol
LDL-cholesterol
Blood insertion for glucose test
Blood insertion for lipid profile test
Sample blood
Serum
Control Solution Symbol
Displaying code number of Test Strips
Short of batteries
Alarm
Testing unit
Temperature & date of test
Hour & Date of test
Average Test Result(only for glucose)
Month & Day
MTH
• 10 • • 11 •
LipiProTM Test Strip Information
Target to apply blood
This test strip is for glucose.
color coding label
This test strip colored black enables to test TC, HDL, TG and calculated LDL.
This test strip colored red is for Total Cholesterol.
This test strip colored blue is for HDLCholesterol.
This test strip colored green is for Triglyceride.
Confirmation Window
Contact Bar
Front -
Back -
Use only LipidProTM Test Strips with your LipidProTM meter. *Note: Please read the individual Package insert of LipidPro test strip for complete detail
information.
5. Glucose test strip
Caution
1. Store the LipidProTM test strip vials in a cool and dry place.Keep out of direct sunlight . Do not freeze.
2. Store test strips in their original vial only. Do not mix the test strips in new vials or in any other container.
3. Immediately replace the vial cap and close tightly after removing the LipidPro
TM
test strip
4. Make a note of the discard date which is three months from the date you first open a new vial of strips. Throw LipidPro
TM
test strip and vial away
after the discard date.
5. Avoid getting dirt, food, and water on the test strip. Do not handle test strips
with wet hands.
6. Do not use the test strips after the expiry date printed on the package or
vial since it may cause inaccurate results.
7. Do not bend, cut, or alter the test strip.
8. LipidPro™test strips are for single use only. Do not re-use.
9. Do not test at the following temperature condition:
- LipidPro™test strips below 64ºF (18ºC) or above 86ºF (30ºC).
10. Do not test with humidity below 10% or above 90%.
11. Refer to additional information in the LipidPro
TM
test strip package.
12. Avoid getting dirt, food, and water on the color-coding label (backside of
glucose test strip).
13. Discard the used test strips carefully according to the local regulations.
LipidProTM Test Strip Information
Front -
Back -
Front -
Back -
Front -
Back -
Front -
Back -
1. Lipid Profile test strip(Black)
2. Total Cholesterol test strip((Red)
3. HDL-Cholesterol test strip(Blue)
4. Triglyceride test strip(Green)
Ribs that guide the strip into the meter
Total Cholesterol
test area
HDL-Cholesterol
test area
Triglyceride
test area
The part for holding strip
Blood application and Test area
15 •
Accuracy:
The accuracy results obtained with the LipidPro
TM
lipid profile and glucose measuring system were compared to glucose results obtained with the Hitachi Glucose Auto meter 747, a laboratory instrument. Glucose levels were measured on 160 patients at three different clinical centers.
System accuracy results for glucose concentration <75 mg/dL
System accuracy results for glucose concentration ≥ 75 mg/dL
Alternate site :
Alternate site glucose test results obtained from alternative sites were comparend to the glucose results obtained with the Hitachi Glucose Auto meter 747(reperence method), a laboratory instrument. Glucose levels were measured at three different clinical centers.
Test results of DORSAL HAND were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
24/27 (93%)
Within ± 10mg/dL
27/27 (100%)
Within ± 15mg/dL
27/27 (100%)
Within ± 5mg/dL
4/5 (80%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
Within ± 5mg/dL
84/133 (63%)
Within ± 10mg/dL
119/133 (89%)
Within ± 15mg/dL
129/133 (97%)
Within ± 20mg/dL
133/133 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
84/133 (63%)
Within ± 10mg/dL
119/133 (89%)
Within ± 15mg/dL
129/133 (97%)
Within ± 20mg/dL
133/133 (100%)
Test results of VENTRAL HAND were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
4/5 (60%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
111/145 (77%)
Within ± 10mg/dL
143/145 (99%)
Within ± 15mg/dL
145/145 (100%)
Within ± 20mg/dL
145/145 (100%)
• 12 • • 13 •
LipiProTM Test Strip Information
Total Cholesterol
*NOTE: Please read individual Package of LipidProTM test strip for complete detail information.
LIMITATIONS OF SYSTEM (for Lipid profile test):
Lipid Profile Test Strips provide accurate when the following constraints are observed :
• Do not use neonate samples
1) Physicians - Please note the following interferences that may affect the results:
- For is with capillary whole blood.
- Extremes in hematocrit may affect test results. Hemartocrit levels less than 30% may cause falsely high reading and hematocrit level greater than 55% may cause falsely low reading.
- Interference: Acetaminophen, Uric acid, ascorbic acid (vitamin C), and other reducing substances(when occurring in normal blood or normal therapeutic concentrations) do not significantly affect results. However, abnormally high concentrations in blood may cause inaccurately high results.
- Use LIpidPro system
TM
at the frequency your doctor recommends testing for total
cholesterol, HDL-cholesterol, and triglyceride.
Performance Characteristics:
The performance or the Lipid Profile test strips has been evaluated in laboratory and in clinical tests(Please refer to strip instructions for use in for more details). Testing Range: The test range of Lipid Pro™ is 100 ~ 400mg/dL for Total Cholesterol, 25~80mg/dL for HDL­Cholesterol and 70 ~ 600mg/dL for Triglyceride.
1) Accurracy
The accuracy lipid profile results obtained with the LipidPro™‚ test system were compared to results obtained with Reference device. Lipid profile levels were measured on 1200 capillary whole blood specimens at three different centers.
Precision
- The precision Total Cholesterol, HDL-cholesterol and Triglyceride test results were measured with Three level venous whole blood samples for 20 days.
- The precision Total cholesterol, HDL-cholesterol and Triglyceride test results were measured with two control solutions for 20 days.
Sample Type
Capillary
slope
1.0013
y-interceptor
2.2025
R
2
0.9933
Total Cholesterol
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
120.0
1.9
1.6
231.8
3.2
1.4
349.7
4.0
1.1
Triglyceride
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
130.0
1.9
1.4
357.2
5.0
1.4
530.1
6.1
1.2
Total Cholesterol
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
150.3
2.2
1.5
250.6
3.6
1.4
HDL-Cholesterol
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
59.9
1.3
2.1
29.9
0.8
2.7
Triglucerode
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
120.5
2.9
2.4
250.5
2.8
1.1
HDL-Cholesterol
Lipid Profile
Mean Conc. (mg/dL)
SD (mg/dL)
CV (%)
30.9
0.8
2.5
51.9
1.1
2.1
73.0
1.6
2.2
HDL-Cholesterol
Sample Type
Capillary
slope
0.9935
y-interceptor
-1.2036
R
2
0.9938
Triglyceride
Sample Type
Capillary
slope
1.0169
y-interceptor
-0.2810
R
2
0.9938
Mean
SD CV
TC
118 .9
2.1
1.8
TG
128.5
2.2
1.7
Level 1 Level 2 Level 3
HDL
31.5
0.7
2.3
TC
236.0
3.4
1.4
TG
357.6
4.3
1.2
HDL
50.2
1.1
2.2
TC
348.0
5.
1.4
TG
529.0
6.0
1.1
HDL
71.2
1.5
2.1
• 14 • • 15 •
LipiProTM Test Strip Information
*NOTE: Please read individual Package of LipidProTM test strip for complete detail information.
LIMITATIONS OF SYSTEM (for Glucose test):
The LipidPro
TM
Glucose Test Strips provide accurate results when the following constraints
are observed:
• Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low results low results may occur for individuals experiencing a hyperglycemic hyperosmolar state, with or without ketosis. Critically ill patients should not be tested with blood glucose meters.
• Use only the LipidPro
TM
glucose test strips with the LipidProTM meter.
• Use fresh capillary whole blood only.
• Do not use neonate samples.
• The test strips are for siingle use only. Do not reuse.
• Dehydration may lower test results.
• Glucose test strips used above altitudes of 10,000 feet will have an effect on test results.
Physicians - Please note the following interferences that may affect test results:
• Extremes in hematocrit may affect test results. Hematocrit levels less than 30 % may cause falsely high reading. Hematocrit levels greater than 55 % may cause falsely low readings.
• Interferences: Acetaminophen, Uric and, Ascorbic acid (Vitamin C), and other reducing substances when occurring in normal blood or normal therapeutic concentrations do not significantly affect results. However, abnormally high concentrations may cause inaccurately high results.
• Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 3000 mg/dl do not significantly affect the results. Values beyond these levels should be interpreted with caution.
• Blood samples that contain a high concentration of dissolved oxygen may lower the test result.
• EDTA containing tube is recommended as an anticoagulant tube.
Performance Characteristics:
The performance of the glucose test strips has been evaluated in laboratory and in clinical tests. Measurement Range: The measurement range of the LipidPro
TM
Glucose Testing is 20 to
600 mg/dL.
Accuracy:
The accuracy results obtained with the LipidPro
TM
lipid profile and glucose measuring system were compared to glucose results obtained with the Hitachi Glucose Auto meter 747, a laboratory instrument. Glucose levels were measured on 160 patients at three different clinical centers.
System accuracy results for glucose concentration <75 mg/dL
System accuracy results for glucose concentration ≥ 75 mg/dL
Alternate site :
Alternate site glucose test results obtained from alternative sites were comparend to the glucose results obtained with the Hitachi Glucose Auto meter 747(reperence method), a laboratory instrument. Glucose levels were measured at three different clinical centers.
Test results of DORSAL HAND were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
24/27 (93%)
Within ± 10mg/dL
27/27 (100%)
Within ± 15mg/dL
27/27 (100%)
Within ± 5mg/dL
4/5 (80%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
Within ± 5mg/dL
84/133 (63%)
Within ± 10mg/dL
119/133 (89%)
Within ± 15mg/dL
129/133 (97%)
Within ± 20mg/dL
133/133 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
84/133 (63%)
Within ± 10mg/dL
119/133 (89%)
Within ± 15mg/dL
129/133 (97%)
Within ± 20mg/dL
133/133 (100%)
Test results of VENTRAL HAND were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
4/5 (60%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
111/145 (77%)
Within ± 10mg/dL
143/145 (99%)
Within ± 15mg/dL
145/145 (100%)
Within ± 20mg/dL
145/145 (100%)
• 16 • • 17 •
Precision:
Repeatability evaluation for venous blood samples.
Intermediate evaluation for control solutions.
Mean (mg/dL)
CV (%)
45.0
3.1
84.7
2.1
130.8
2.2
206.2
2.6
341.8
2.4
Mean (mg/dL)
CV (%)
49.7
2.3
100.5
1.7
300.3
0.7
Test results of VENTRAL HAND were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
4/5 (60%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
111/145(77%)
Within ± 10mg/dL
143/145 (99%)
Within ± 15mg/dL
145/145 (99%)
Within ± 20mg/dL
145/145(100%)
Test results of UPPER ARM were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
4/5 (80%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ±5%
1.9/145(75%)
Within ±10%
143/145 (99%)
Within ±15%
145/145 (100%)
Within ±20%
145/145(100%)
Test results of FORE ARM were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
4/5(80%)
Within ± 10mg/dL
5/5 (100 %)
Within ± 15mg/dL
5/5 (100 %)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ±5%
95/145(66%)
Within ±10%
137/145 (94%)
Within ±15%
143/145 (99%)
Within ±20%
145/145(100%)
Test results of THIGH were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
3/5(60%)
Within ± 10mg/dL
5/5 (100%)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ± 5mg/dL
103/145(71%)
Within ±10%
137/145 (94%)
Within ±15%
144/145 (99%)
Within ±20%
145/145(100%)
Test results of CALF were compared to the reference method.
System accuracy results for glucose concentration<75mg/dL
Within ± 5mg/dL
2/5(40%)
Within ± 10mg/dL
5/5 (100 %)
Within ± 15mg/dL
5/5 (100%)
System accuracy results for glucose concentration ≥ 75 mg/dL
Within ±5%
100/145(69%)
Within ±10%
138/145 (95%)
Within ±15%
145/145 (100%)
Within ±20%
145/145(100%)