Oriflow Rhino User Manual

Page 1
Operating Instructions for
Rhino
Duct Leakage Tester
(Metric)
© Copyright 2015 ORIFLOW LLC
Page 2
Table of Contents
Introduction
Introduction ……………………………………….…….. 2
Safety Precautions
Safety Precautions ………………………………….…. 2
Prior to Testing
System Preparation …………………….………………. 3 Power Requirements …………………………………... 3 Extension Cord Requirements ………………………... 4 Flex-Duct Length ………………………………………... 4 Determining Maximum Allowable Leakage ….……... 4 Determining Which Orifice Plate to Use ……………... 5 - 7
Duct Leakage Test Setup and Procedure
Install Orifice Plate and Upper Tube Section………… 8 - 9 Connect Flexible Duct to Orifice Tube ……………….. 10 Connect Flexible Duct to Duct System ………………. 10 System Test Pressure Connection …………………… 11 Duct Leakage Test Setup (SUPPLY) ………………… 12 Duct Leakage Test Setup (RETURN/EXHAUST) ...….. 13 Inlet Transition Required for Return/Exhaust Testing . 14 Zeroing Pressure Gauges ……………………………… 15 Avoiding Over-Pressurization ………………………… 16 Obtaining System Test Pressure ……………………… 16 - 17 Determining Leakage Rate……………………………… 17 - 19
Troubleshooting
Zero reading on “DUCT SYSTEM” gauge ………….... 19
Zero reading on “ORIFICE PLATE” gauge …………….. 20 Can’t obtain system test pressure ……………………… 21
“ORIFICE PLATE” gauge reading maxed out………...... 21
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 3
Introduction
These operating instructions are for ORIFLOW model RHINO air leakage tester. Take pride in knowing that you have purchased the best air leakage tester on the market. It is no secret that Oriflow manufactures the highest quality testers, and has unparalleled customer service and technical support.
The orifice plates available for your duct leakage tester are constructed from laser­cut 3.2 mm stainless steel and do not require recalibration for 10 years. Note the authority having jurisdiction may override this requirement, and if so, Oriflow has very reasonable rates for calibrating orifice plates manufactured by Oriflow.
Safety Precautions
Before operating your tester, read the following safety precautions:
DO NOT operate the tester in the rain,
DO NOT operate the tester while it is near or in water,
DO NOT operate the tester with a damaged electrical cord or plug,
DO NOT remove the inlet safety screen,
DO NOT touch the blower wheel when the unit is plugged in,
DO NOT look into the discharge end of the tester when the unit is plugged in,
DO NOT use the tester as a ladder or step stool,
DO NOT allow children near the tester,
DO use an extension cord of the proper gauge (see Table 2),
DO use the proper voltage and line frequency listed on the motor nameplate,
DO lock the caster when the tester is positioned horizontally,
DO use the tester on level ground,
DO secure the tester when transporting it,
DO wear proper hearing protection, safety glasses and work gloves,
DO seek assistance when lifting the tester (e.g., loading onto truck, going up
or down stairs).
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 4
Prior to Testing
System Preparation
Cap off all ends of system using clear plastic and duct tape or sheetmetal duct end caps. Make sure you test the part of the duct system that leakage testing is required. Usually, this is from the system fan up to, but not including, the VAV boxes (terminal units). Refer to the engineer's specifications and all applicable codes and test standards.
Power Requirements
Do you have an adequate power supply for your tester (see Table 1)?
Table 1 – Tester Amp Draw
Tester Model
Rhino
Flow
Voltage
208 Slide Gate 3 13
208 VFD 3 14
230 VFD 1 21
230 Slide Gate 3 12
230 VFD 3 13
380 Slide Gate 3 7
380 VFD 3 8
460 Slide Gate 3 6
460 VFD 3 7
Control
Option
Phase
Full Load
Amps
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 5
Extension Cord Requirements
If you need an extension cord, is it the proper gauge? See Table 2 below for
extension cord requirements.
Table 2 – Required Extension Cord Wire Gauge
Rhino Model
Flow
Control
Option
VFD 230 1
Slide Gate 230 3
VFD 230 3
Slide Gate 380 3
VFD 380 3
Slide Gate 460 3
VFD 460 3
Flex-Duct Length
Voltage Phase
Cord Length
Extension
(meters)
3 to 15 10
15 to 30 10
3 to 15 14
15 to 30 14
3 to 15 14
15 to 30 14
3 to 15 14
15 to 30 14
3 to 15 14
15 to 30 14
3 to 15 14
15 to 30 14
3 to 15 14
15 to 30 14
Gauge
Wire
Make sure you have enough flexible-duct with your tester for the job. Each tester includes 3.8 m, which is enough for most applications. Extra lengths of flexible-duct are available at www.oriflow.com/products.
Determining Maximum Allowable Leakage
To determine the maximum allowable leakage for the project, use ORIFLOW’s free online programs to make these calculations (www.oriflow.com/programs).
Calculations are done for either of the two typical specifications: percentage of system flow, or leakage class. If the specification uses leakage class, you will need to calculate the total duct system surface area. Oriflow has a free Adobe form available to our customers that will calculate duct surface area given the duct shape, dimensions and length.
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 6
Determining Which Orifice Plate to Use
After determining the allowable leakage at the system test pressure, refer to the
following tables for the capacities of each orifice plate available for the RHINO model
tester. Make sure you have an orifice plate that can measure the maximum allowable leakage. The proper plate is the one where the maximum allowable leakage falls between the minimum and maximum leakage at the system static pressure. For example, if you are testing a system at 2500 Pa pressure and the maximum allowable leakage is 283 L/s, you will need a 125 mm orifice plate since the 125 mm plate can be used for up to 380 L/s of leakage (see Table 7).
It is a good idea to have the next larger orifice plate size since it is common for duct systems to leak more than the maximum allowable. Smaller diameter plates are use when leakage is relatively low.
Table 3 – Tester Capacities using the 25 mm Orifice Plate
System Static
Pressure
(Pa)
Minimum Leakage*
(L/s)
Maximum
Leakage
(L/s)
500 4 21 1000 4 19 1500 4 18 2000 4 17 2500 4 14 3000 4 12 3500 4 9 4000 4 6
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 7
Determining Which Orifice Plate to Use (continued)
Table 4 – Tester Capacities using the 50 mm Orifice Plate
System Static
Pressure
(Pa)
500 17 87 1000 17 78 1500 17 73 2000 17 66 2500 17 57 3000 17 47 3500 17 35 4000 17 26
Table 5 – Tester Capacities using the 75 mm Orifice Plate
System Static
Pressure
(Pa)
Minimum Leakage*
(L/s)
Minimum Leakage*
(L/s)
Maximum
Leakage
(L/s)
Maximum
Leakage
(L/s)
500 40 195 1000 40 180 1500 40 165 2000 40 145 2500 40 130 3000 40 105 3500 40 80 4000 40 60
Table 6 – Tester Capacities using the 100 mm Orifice Plate
System Static
Pressure
(Pa)
Minimum Leakage*
(L/s)
Maximum
Leakage
(L/s)
500 75 370 1000 75 340 1500 75 310 2000 75 280 2500 75 245 3000 75 195 3500 75 155 4000 75 105
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 8
Determining Which Orifice Plate to Use (continued)
Table 7 – Tester Capacities using the 125 mm Orifice Plate
System Static
Pressure
(Pa)
500 125 560 1000 125 520 1500 125 480 2000 125 430 2500 125 380 3000 125 310 3500 125 240 4000 125 175
Table 8 – Tester Capacities using the 159 mm Orifice Plate
System Static
Pressure
(Pa)
Minimum Leakage*
(L/s)
Minimum Leakage*
(L/s)
Maximum
Leakage
(L/s)
Maximum
Leakage
(L/s)
500 215 730 1000 215 670 1500 215 630 2000 215 560 2500 215 510 3000 215 425 3500 215 310 4000 215 230
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 9
Install Orifice Plate and Upper Tube Section
The orifice plate that was determined from the previous section should now be installed with the serial number facing upwards, so the corresponding calibration
certificate may be referenced after installation. Refer to Figures 1 through 4.
Figure 1 – Install Orifice Plate and then Rotate Clockwise
Figure 2 – Rotate Orifice Plate Clockwise until it Locks in Place
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 10
Install Orifice Plate and Upper Tube Section (continued)
After the orifice plate is locked in place, place the upper tube section over the bolt/wing nut set, rotate clockwise, and tighten wing nuts (Figures 3 and 4).
Figure 3 – Install Upper Tube Section and Rotate Clockwise
Figure 4 – Tighten Wing Nuts after Rotation
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
Page 11
Connect Flexible Duct to Orifice Tube
See the figures below for securing one end of the flexible duct to the orifice tube.
Note that the worm-gear clamp should be slid over the flex-duct before connecting it
to the orifice tube. Slide the flexible duct so that it overlaps the orifice tube 25 to 50 mm, and tighten clamp using a 8 mm nut driver.
Figure 5 – Installing Flex-Duct on Tube Figure 6 – Tightening Clamp
Connect Flex-Duct to Duct System
Find a convenient location in the duct system where the tester has easy access. Make sure that the connection where you connect the flexible duct is a sturdy, sealed tight connection. You don’t want to create a leaky connection where the tester is supplying air into the system.
Figure 7 – Connection to Duct System
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
10
Page 12
System Test Pressure Connection
So that you can monitor the system static pressure, you need to drill a hole at least 1 meter away from the flex-duct connection. Drill a 8 mm diameter hole and insert the pressure tubing from the “DUCT SYSTEM” gauge so that 150 to 300 mm of tubing is inside the duct system. Using putty or duct tape, seal the connection. If your pressure tubing has a plastic static pressure tap at the end of the pressure tubing,
insert the tap in the hole and seal it to the duct wall using duct tape. See Figure 8
below.
Figure 8 - Static Pressure Tubing Connection
See Figure 9 showing the test setup for SUPPLY system testing. Refer to Figure 10
for the test setup required to measure air leakage of a RETURN/EXHAUST system. Note you will need an inlet transition to attach to the blower inlet when performing a return/exhaust test. The dimensions of the transition required depends on whether
you have a Rhino model with an inlet slide gate or VFD for flow control. See Figure 11 and 12 for inlet transition dimensions required.
DUCT SYSTEM pressure tubing inserted into duct system
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
11
Page 13
Figure 9 – Duct Leakage Test Setup for a SUPPLY System
Tap into existing outlet if possible
Cap and seal all outlets
1. Drill 9.5 mm hole
2. Insert tubing 150 to 300 mm
3. Seal penetration with putty
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
12
Page 14
Figure 10 – Duct Leakage Test Setup for a RETURN/EXHAUST System
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
13
Page 15
Figure 11 - Inlet Transition Required for Rhino with Inlet Slide Gate
Figure 12 - Inlet Transition Required for Rhino with VFD Speed Controller
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
14
Page 16
Zeroing Pressure Gauges
Before you run the tester, make sure both gauges have been zeroed.
Zeroing Analog Gauges
Using a very small screwdriver, turn the zero-adjustment screw on the gauge until the needle is aligned with the zero reading. Turning the screw clockwise increases the pressure reading; turning it counterclockwise, decreases the
reading. See Figure 13.
Figure 13 – Zeroing Analog Gauges
(zero BOTH gauges before testing)
Zeroing Digital Gauges
Use the following steps to zero each digital gauge:
1. Press MENU button once.
2. Press
3. Press
4. Press
5. Press
6. Press MENU button two times to get back to the pressure reading.
7. Pressure should be reading zero or within +/- 5 Pa If not, repeat steps.
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
arrow button until Adu shows on screen.
E button once to go into auto-zero mode E button again and AUTO will be blinking on screen E button a third time to complete the zeroing process.
15
Page 17
Avoid Over-Pressurization
Prior to starting the blower, shut the inlet slide gate on models without the speed controller option. For models with the VFD speed controller, rotate the speed control knob completely counter-clockwise.
Figure 14 – Avoid Over-Pressurizing by Minimizing Air Delivery upon Startup
Obtaining System Test Pressure
Turn the blower on and slowly open the inlet slide gate for models without the
speed control option. If your tester has the VFD speed controller, rotate speed control knob slowly clockwise to obtain system test pressure.
Figure 15 – Obtaining System Test Pressure
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
16
Page 18
Obtaining System Test Pressure (continued)
Regardless of tester model used, pay attention to the “DUCT SYSTEM” gauge while increasing airflow. When you have reached the required system static pressure, tighten the set-screw on the inlet damper (or STOP rotating speed control knob).
Figure 16 – Duct System Gauge Reading (analog on left, digital on right)
Determining the Leakage Rate
Now that you have obtained the system test pressure, note the pressure reading of
the ORIFICE PLATE gauge (Figure 17). Refer to your calibration certificate to
determine the leakage rate that corresponds to the gauge reading. An example of a
calibration certificate is shown in Figure 18; it is for informational purposes only and used in the example shown on the next page. Always refer to the calibration
certificate(s) that came with your calibrated orifice plate(s).
Figure 17 –ORIFICE PLATE Gauge Reading (analog on left, digital on right)
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
17
Page 19
Determining the Leakage Rate (continued)
SAMPLE CALIBRATION CERTIFICATE
USE YOUR CERTIFICATE
203.2 mm
Figure 18 Example Calibration Certificate. Use the calibration certificate that came with your orifice
plate
Example (refer to Figures 17 and 18) For an ORIFICE PLATE gauge reading of 725 Pa (Figure 16), the leakage
rate is approximately 194 L/s using the calibration certificate shown in Figure 18 (see arrow).
For an exact value, use the equation that comes with every calibration certificate. For the ORIFICE PLATE gauge reading shown in Figure 17:
Leakage = 7.206 x
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
18
Page 20
Determining the Leakage Rate (continued)
where the value of “7.206” is a constant and will be specific to the orifice plate used (every plate has its own value; yours will be different). Using a calculator, take the square root of the gauge reading first, and then multiply by the constant for your orifice plate (for this example, the constant is "7.206"). The exact leakage is 7.206 x (square root of 725) = 194.0 L/s.
Oriflow has FREE Adobe forms available that will calculate the leakage, leakage factor, allowable leakage, Pass/Fail criteria and duct surface area among other
variables (www.oriflow.com/freeform).
Troubleshooting
Zero reading on “DUCT SYSTEM” gauge
If the gauge reading is zero, refer to the table below to fix this problem. Perform
troubleshooting steps in the order shown in the Table 9 below.
Table 9 –Troubleshooting steps when “DUCT SYSTEM” gauge is zero
Step Scenario What to do
1
2
3
4
Inlet damper was left shut,
cutting off air to the system.
Pressure tubing connected
to gauge incorrectly
Plugged fitting or
malfunctioning gauge.
System is leaking
too much air.
Open inlet damper slowly.
Look at the P1 and P2 stickers that are located on the orifice tube and gauges. Make sure tubing from P1 tap matches the P1 tap on the “ORIFICE PLATE” gauge.
Gently blow through other end of pressure tubing from “DUCT SYSTEM” gauge, and watch needle increase pressure reading.
Check for these other sources of leakage:
Rectangular duct joints (check and seal corners),  Fire or smoke dampers,  Duct joints (pay particular attention to flex-duct
joints if they are part of the leak test),
VAV boxes (pay particular attention to parallel box
back draft dampers),
Built-up air handlers,  Plenums,  Uncured duct sealant blow-thru (follow
manufacturer instructions for cure time),
Improperly sealed or un-sealed joints,  Hot water coils and electric heaters,  Open duct end that was supposed to be
sealed/capped-off for the leak test.
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
19
Page 21
Zero reading on “ORIFICE PLATE” gauge
If the gauge reading is zero, refer to the table below to fix this problem. Perform
troubleshooting steps in the order shown in Table 10 below.
Table 10 –Troubleshooting steps when “ORIFICE PLATE” gauge is zero
Step Scenario What to do
1
2 Plugged pressure taps.
3
4
4
5 Malfunctioning gauge.
6 Not sure
Inlet damper was left shut,
cutting off air to the system.
Orifice plate too large for
application.
System is leaking very little
air.
Pressure tubing connected
incorrectly.
Open inlet damper slowly.
Remove pressure tubing from “ORIFICE PLATE” gauge, and blow through each end to make sure the pressure taps on the orifice tube are clear.
If you are using a 100 mm ID orifice plate, and the system is leaking 20 L/s, you probably won’t even notice the gauge needle moving. This can happen to other orifice plates too. You need to use an orifice plate with a smaller bore diameter so that a small amount of airflow/leakage will result in a bigger pressure drop.
You’ll know if this is true if the fan inlet is almost shut. Not likely unless system is small. Solution is to use a smaller ID plate.
Look at the P1 and P2 stickers that are located on the orifice tube and gauges. Make sure they match.
Remove the clear pressure tubing from the lower pressure tap on the orifice tube and blow through the end of the tubing. The needle should move in response. If not, the gauge is faulty.
The best way to make sure that your tester is working properly is to disconnect the flex duct from the system and turn the blower on, and slowly open the inlet damper. If you feel a lot of air coming out of the tube end, you should see a pressure reading on the gauge “ORIFICE PLATE”.
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
20
Page 22
Can’t obtain system test pressure
This happens when the system is leaking too much air. Make sure all outlets are sealed. Check corners of rectangular duct for excessive leakage. Inspect all duct and fitting joints for leakage. Make sure you seal all suspect joints and allow to cure 24 to 48 hours. Always refer to duct sealant manufacturer’s instructions.
Check for these other sources of leakage:
Rectangular duct joints (check and seal corners), Fire or smoke dampers, Duct joints (pay particular attention to flex-duct joints if they are part of the
leak test),
VAV boxes (pay particular attention to parallel box back draft dampers), Built-up air handlers, Plenums, Uncured duct sealant blow-thru (follow manufacturer instructions for cure
time),
Improperly sealed or un-sealed joints, Hot water coils and electric heaters, Open duct end that was supposed to be sealed/capped-off for the leak test.
A non-toxic smoke machine is an excellent tool for locating significant sources of leakage. Call ORIFLOW at 727-400-4881 or online at www.oriflow.com for more information.
“ORIFICE PLATE” gauge reading maxed out
Scenario 1: The system is leaking too much air. Make sure all outlets are
sealed. Check corners of rectangular duct for excessive leakage. Inspect all duct and fitting joints for leakage. Make sure you seal all suspect joints and allow curing time of 24 to 48 hours. Always refer to duct sealant manufacturer’s instructions.
Scenario 2: The orifice plate bore could be too small, causing a high
pressure drop at low to moderate flows. You need an orifice plate with a larger bore.
ORIFLOW LLC www.oriflow.com 2125 Range Rd., Suite B Clearwater, Florida 33765
727-400-4881 (phone) 8am to 5pm PST 877-420-7091 (fax) sales@oriflow.com engineering@oriflow.com
© Copyright 2018 ORIFLOW LLC Rev: 03-2018
21
Loading...