orfit AIO SOLUTION P-LD-PED, AIO SOLUTION P-LD Instruction Manual

INSTRUCTIONS FOR USE
THE AIO SOLUTION®
IN CARBON FIBRE LAMINATE
drivenbycare 1 / 2
(+32) 3 326 20 26
USA tollfree 1-888-ORFIT-US (888-672-4887)
AIO SOLUTION® code:
P-LD-PED
Article No. : 32301-PED
A. GENERAL PRODUCT INFORMATION The products referred to in these instructions are medical
devices, used for patient positioning and immobilisation in radiation therapy.
B. PRODUCT DESCRIPTION This carbon fibre laminate BASE PLATE, used for the
treatment of both children and adults, is the basic element of "THE TOTAL AIO SOLUTION® TREATMENT PACKAGE”. It is used in combination with THE AIO SOLUTION® CUSHIONS, the RAYCAST® HIGH PRECISION HEAD SUPPORTS and BLOCKS & WEDGES, and with the EFFICAST® PRE-CUTS to form a reproducible patient positioning and immobilisation device. Information on these other parts can be found in the respective ‘instructions for use’.
C. PRECAUTIONS FOR USE This is a fragile product, please handle with care. Always
place the base plate on a flat surface. Always fix the base plate securely to the treatment couch before positioning a patient. More information on how to fix this base plate to a couch top can be found on our website
http://www.orfit.com/en/fixation-devices/.
If the base plate is used in overhang over the couch a Table Fix (art. no. 35744/6) or iBeam couch fixation (art. no.
32813) needs to be used at the caudal end to secure the base plate to the couch top. This prevents the base plate from falling off the table when the patient moves.
The CFL base plate is constructed to be light in weight. When handled roughly, it may get damaged and fibres may come off. When this happens, stop using the base plate to
prevent fibres from getting in contact with the patient’s or
user’s skin. Contact your distributor.
D. STORAGE Always store the product in a safe place to prevent it from
getting damaged. Take care not to damage the edges of the plate when storing it in an upright position. Prevent hard objects from falling onto the plate. Store the base plate between +10°C (50°F) and 40°C (122°F).
E. PROPERTIES
E.1. Physical Properties
Dimensions: L 1350 mm x W 544 mm x H 30 mm Weight : 3,630 kg
E.2. Mechanical Properties
Bending:
The bending (sag) Δ Lw of the AIO base plate (measured
between the two positioning holes of the head support, at
323 mm from the ‘maximum overhang line) can be
calculated using the next equation: Δ Lw = 0,07695W + 0,04889, where Δ Lw is the displacement of the plate (mm)
W is the vertical force (weight), exerted at the measuring point (kg)
SD: ± 0,017 mm
drivenbycare 2 / 2
(+32) 3 326 20 26
USA tollfree 1-888-ORFIT-US (888-672-4887)
E.3. Dosimetric Properties
Attenuation factor (± 0,15 %)
6 MV
1.4%
10 MV
1.2%
H2O equivalence
4 mm
Note: Use these numbers as a guidance only. Perform the measurements again in your department to verify these results.
F. MAINTENANCE AND WASTE MANAGEMENT This product can be cleaned and disinfected by means of
soapy water or an isopropanol based disinfectant, applied with a soft cloth. If unsure about the cleaning fluid, do not use. Never use aerosol sprays, corrosive cleaning
agents, solvents or abrasive detergents.
Periodic checks of the product should be done to insure the parts are not worn and require repair or replacement. Do not attempt to make repairs yourself. Contact your distributor if there are any questions or concerns. The product can be disposed of with household waste.
G. ADDITIONAL INFORMATION For additional information such as distributor contact
information, product brochures, Safety Data Sheets and regulatory information, please visit our website www.orfit.com.
The instructions were written in accordance with the European Directive 93/42/EEC for Medical Devices. It is prohibited to make alterations to this text without prior approval from ORFIT Industries.
THE AIO SOLUTION® is a registered trademark of ORFIT Industries
Ref. No. 31954 - 32301-PED AIO.doc
VERSION 3 - LAST UPDATE: 01/12/2014 - REVISION DATE: 01/12/2016