ORFIT STRIPS are cut from a low temperature thermoplastic sheet of coated material. The
stripsaredesignedforthequickandeasyproductionoforthosis,externalimmobilizationdevices
and rehabilitation aids.
The pre-cut ORFIT STRIPS are specifically designed to facilitate the fabrication of smaller
orthosis with more detail, and for creative input in functional assistive devices.
ORFIT STRIPS are applied directly to the patient once activated.
!ORFIT STRIPS are not suitable for internal use. This product may not be used on open
wounds or in the mouth.
B. PRODUCT RANGE
STRIPS
ORFIT STRIPS are available in different colours, widths, and thicknesses. They all have a
standard length of 45 cm. Each package of ORFIT STRIPS contains 10 pieces of the same
width and thickness either in the same colour or a mix of both (5 pieces of each colour).
Art. No.Pieces / Colour
8113MG.1/NS10 x Sonic Silver2.045 cm x 0.5 cmnon perforated
8113GO.1/NS10 x Gold2.045 cm x 0.5 cmnon perforated
8113GOMG.1/NS
8124GO.1/NS10 x Gold3.445 cm x 2 cmnon perforated
8124MB.1/NS10 x Atomic Blue3.445 cm x 2 cmnon perforated
8124MBGO.1/NS
C. PRECAUTIONS BEFORE USE
5 x Sonic Silver
5 x Gold
5 x Gold
5 x Atomic Blue
Thickness
(mm)
2.045 cm x 0.5 cmnon-perforated
3.445 cm x 2 cmnon perforated
Length x Width
(cm)
Perforation type
1. Make sure the workplace is well-ventilated to avoid overheating.
2. The necessary tools shouldin no way put the patient at risk.
3. Position the patient comfortably and ensure that you have easy access to their involved
extremity.
by
1 / 5
(+32) 3 326 20 26
www.orfit.com
driven
care
D. TECHNICALACTIVATION
1. ORFIT STRIPS are activated and softened by heating at a minimum temperature of 65C
(149F). Possible activationsourcesare: waterbath, dry heater, heatgun,heatingplate,hotair
oven. The activation time depends onthe heat sourceand variesfrom 2 to 5 minutes.
2. When using a heatgun, do not exceed250C(482F) to avoid breakdown of the material.
Whenusing a heating plate oran oven, the hotsurface must becovered with a Teflonfilm.
!3. Please note: temperatures of 65°C (149°F) or more can also be reachedin thepatient's
daily life. Think of a closed car in the summer, the surfaceof a hotradiator, a sauna or
the proximity of an open fireplace.
4. High temperatures up to a maximum of 120C (248F) do not damage ORFIT STRIPS, but
are not user-friendly. High temperatures are allowed provided that the activation time is
reduced accordingly and that the product is sufficiently rubbed with talcum powder. Wear
gloves and do not apply ORFIT STRIPS directly to the patient at these high activation
temperatures.
!5. Never use an open flame to activate ORFIT STRIPS.
E. WORKING PROPERTIES
Cutting
Cut the ORFIT STRIP into the estimated length prior to activation, using a suitable pair of
scissors or cutter.
!Be careful of possible cuts when using a cutter. Always keep the assisting hand away
from the cutting line.
Applying
1. Activate the ORFIT STRIP until it is completely soft. Take it out of the water and dry briefly
on a towelfor a few seconds.
Make sure that the temperature of the activated material will not burn the patient.
!ORFIT STRIPS are:
-non-adhesive when wet heated, or when dry heated and extra powdered with
talcum.
-slightly adhesive when dry heated at a low temperature.
-fully adhesive when dry heated at a high temperature.
2. Several application techniques are possible:
- gravity technique: use gravity to assist the moulding process.
- closed technique: mould the material around the body part and overlap the edges.
!As with all thermoplastics, take advantage of the stretch and the elasticity of ORFIT
STRIPS while moulding but be aware that the product may tear if overstretched.
3. In case of accidental bonding, let ORFIT STRIPS harden completely so that parts that are
stuck can be separated. Reactivation in hot water is then safe again.
4. Remove the orthosis from the patient when the moulded ORFIT STRIPS have hardened.
Excessive material can be trimmedbefore complete hardening.To do so, use a suitable pair
of bandage scissors.
The coolingtime can be shortenedby means of cold water, coldair,a coldbandageor a cold
spray.
5. To attach strapsto the completed orthosis, the NS film (coating) must be REMOVED locally.
It can be scratched off with a knife, sandpaper or with a grinding tool. Use hot dry heat to
make the spots sticky.
1. Take one piece of ORFITSTRIPS(Gold
or Atomic Blue 2 cm wide) and cut in
half.
2. Activatethe ORFIT STRIPS in hotwater
of minimum 65°C (149°F) and drybriefly
on a towel.
3. Place the middle of the ORFIT STRIPS
over the dorsal part of the proximal
phalanges.
4. Crisscross the ORFIT STRIPS over the
volar surface of the palm at the level of
the distal palmarcrease.
5. Mould the crossed part into the palm to
support the arch.
6. Pull the rest of one strip through the first
webspaceandthe otherpart aroundthe
ulnar border of the hand to meet
together on the back of the hand at the
levelof the metacarpals.
7. Hold the ORFIT STRIPS in place while
allowing the material to conform to the
dorsalfingers.
8. When cooled, remove the orthosis from
the patient and trim both ends of the
ORFIT STRIP on the ulnar and radial
borders of the hand. Round of all the
corners.
9. Usea heat guntoactivatetheglueofthe
self-adhesive hook tape and to activate
both ends of the ORFITSTRIPS.
1
3
4
7
8
9
11
10
10. Attach a piece of hook tape on the two
ends of the orthosis.
11. Apply one Velcro loop strap from the
ulnar to the radial border.
12. This Anti Claw orthosis allows full finger
flexion while preventing hyperextension
of the digits (MCP joints).
by
3 / 5
12
(+32) 3 326 20 26
www.orfit.com
driven
care
B.Anti-Swan Orthosis
4
Instructions:
1. Take one pieceof ORFITSTRIPS(Gold
or Sonic Silver 0.5 cm wide).
2. Cutit into two pieces:one pieceof 10cm
and onepiece of 5 cm.
3. Activate both pieces of the ORFIT
STRIPS in hot water of minimum 65°C
(149°F) and dry briefly on a towel.
4. Place the middle of the largest piece
over the dorsal part of the proximal
phalanx and bring both ends to overlap
overthedistalpartof themiddle phalanx
whiletheProximalInterphalangeal (PIP)
joint is flexed.
5. Let this oval shaped ORFIT STRIP
hardenon the patient.
6. Take the second piece of the ORFIT
STRIPS and place it directly under the
PIP joint so that it overlaps on the outer
borders of the hardened oval.
2
6
8
9
7. When both pieces have hardened,
carefully remove the shaped ORFIT
STRIP.
8. Trim the edges of the ovalshaped part.
9. Trim the edges of the overlapping parts
as well.
10. Use a heatgun tomake both ends ofthe
oval shape sticky.
11. Useaheatguntomakeboth
overlapping endssticky as well.
12. Attach the sticky overlapping edges to
the sticky parts of the oval shape.
13. Dip into cold water to let the orthosis
completely harden and place over the
flexedfinger.
14. TheAnti-Swan orthosisallowsfull
flexion while preventing hyperextension
of the PIPjoint.
10
11
12
15. Manyadditionalorthosisand
rehabilitationaidscanbe fabricatedfrom
ORFIT STRIPS. Use your creativity to
come up withyour own designs!
by
4 / 5
14
(+32) 3 326 20 26
www.orfit.com
driven
care
F. FINISHING
The edgesof an ORFITSTRIPS orthosis can be smoothed by rubbing with a wet finger.
G. MAINTENANCE ANDWASTEMANAGEMENT
OrthosesmadeofORFITSTRIPSshouldbecleaneddaily.Uselukewarmwater and liquidsoap,
biological detergent or toothpaste.Rinse welland dry thoroughly.
!Never use solvents. Avoid acid detergents.
Sterilisationof ORFIT STRIPS orthosis in an autoclave is impossible.
Disinfection is possible with alcohol, quaternary ammonium or a solution of commercial
disinfecting soaps (HAC®, Sterillium®, etc.).
!Avoid prolonged contact with detergents and acids which may affect the NS film.
After use, an orthosis can be disposed of with normal household waste without harming the
environment. ORFIT STRIPSis biodegradable.
H. ADVICE FOR THE PATIENT
!Make sure the patientunderstands the wearing schedule of their orthosis.
I. STORAGE
- ORFITSTRIPS should be stored vertically, if supported andhorizontally if not.
- Stock should bekept in a dark, cool,dryplaceat a temperatureof min. 10°C(50°F)andmax.
30°C (86°F) and in the original packaging.
Once removed from the packaging, left-overs should be placed back in the packaging for
storageto avoid adhesion of the NS film and biodegradation.
Low temperature thermoplastics can only be kept for a limited period of time and must be
protected as much as possible from light, heat and humidity. The material ages in relation to
storagecircumstances. When aged, it becomes brittle and often very stretchy when activated.
J. GENERAL SAFETY ADVICE
!* ORFIT STRIPS are not suitablefor internal use.This product may not be used on open
wounds or in the mouth.
!* Never use an open flame to activate ORFIT STRIPS.
!* To make orthosis and rehabilitationaids,ORFIT STRIPS may only be used by qualified
These instructions are written in accordance with the European Directive 93/42/EEC for Medical Devices. It is prohibited to make
alterations to thistext withoutprior approval from ORFIT Industries.
ORFIT®is a registered trademark of ORFIT INDUSTRIES N.V.