Please, read the enclosed safety instructions carefully and observe them!
These instruments should be safeguarded by isolated or common fuses (breakers)!
For safety information the EN 61 010-1 + A2 standard must be observed.
This instrument is not explosion-safe!
TECHNICAL DATA
Measuring instruments of the OMB 450 series conform to the European regulation 89/336/EWG.
The instruments are up to the following European standards:
EN 55 022, class B
EN 61000-4-2, -4, -5, -6, -8, -9, -10, -11
Seismic capacity:
IEC 980: 1993, chapter 6
The instruments are applicable for unlimited use in agricultural and industrial areas.
CONNECTION
Supply of energy from the main line has to be isolated from the measuring leads.
The OMB 451/452 model series are programmable, three-color panel bargraphs with auxiliary display and adjustable
LCD scale. The instruments are designed as dimensional replacement of the ZEPAKOMP instruments. Available are types
UNI, PWR and UQC.
Type OM 402UNI is a multifunction instrument with the option of configuration for 8 various input options, easily configurable
in the instrumentmenu. By further options of input modules it is feasible to measure larger ranges of DC voltage and current
or increase the number of inputs up to 4 (applies for PM).
The instrument is based on an 8-bit microcontroller with a multichannel 24-bit sigma-delta converter, which secures high
accuracy, stability and easy operation of the instrument.
The OMB 451/452 is a multifunction instrument available in following types and ranges
type UNI
DC:0…60/150/300/1 200 mV
PM: 0…5 mA/0…20 mA/4…20 mA/±2 V/±5 V/±10 V/±40 V
OHM: 0…100 Ϊ/0…1 kΪ/0…10 kΪ/0…100 kΪ/Auto
RTD-Pt: Pt 50/100/Pt 500/Pt 1 000
RTD-Cu: Cu 50/Cu 100
RTD-Ni: Ni 1000/Ni 10000
T/C: J/K/T/E/B/S/R/N/L
DU:Linear potentiometer (min. 500 Ϊ)
type UNI, option A
DC: 0…1 A/0…5 A/120 V/±250 V/±500 V
type UNI, option B (expansion by 3 more inputs)
PM:
3x 0…5 mA/0…20 mA/4…20 mA/±2 V/±5 V/±10 V/±40 V
PROGRAMMABLE PROJECTION
Selection: of type of input and measuring range
Measuring range: adjustable as fixed or with automatic change
Setting: manual, optional projection on the display may be set in the menu for both limit values of the input
signal , e.g. input 0…20 mA> 0…850,0
LCD scale: illuminated and freely programmable
Projection: -9999…9999 (-99999…999999)
COMPENSATION
of conduct: in the menu it is possible to perform compensation for 2-wire connection
of conduct in probe: internal connection (conduct resistance in measuring head)
of CJC (T/C): manual or automatic, in the menu it is possible to perform selection of the type of thermocouple and
compensation of cold junctions, which is adjustable or automatic(temperature at the brackets)
LINEARIZATION
Linearization:* by linear interpolation in50 points (solely via OM Link)
DIGITAL FILTERS
Floating average: from2…30 measurements
Exponen.average: from2…100 measurements
Rounding: setting the projection step for display
MATHEMATIC FUNCTIONS
Min/max. value: registration of min./max. value reached during measurement
Tare: designed to reset display upon non-zero input signal
|
4
INSTRUCTIONS FOR USE OMB 450UNI
* only for types DC, PM, DU
Peak value: the display shows only max. or min. value
Mat. operations: polynome, 1/x, logarithm, exponential, power, root, sin x
EXTERNAL CONTROL
Lock : control keys blocking
Hold : display/instrument blocking
Tare : tare activation/resetting tare to zero
Resetting MM : resetting min/max value
Memory: data storage into instrument memory
2.2 OPERATION
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument
are performed in three adjusting modes:
LIGHTSimple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code
PROFIComplete programming menu
- contains complete instrument menu and is protected by optional number code
USER User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine
the right (see or change)
- access without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface,
which is a standard equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the
purchase of OML cable to connect the instrument to PC. It is manufactured in version RS 232 andUSB and is compatible
with all ORBIT MERRET instruments. Another option for connection is with the aid of data output RS 232 or RS 485 (without
the need of the OML cable).
The program OM LINK in „Basic“ version will enable you to connect one instrument with the option of visualization and
archiving in PC. The OM Link „Standard“ version has no limitation of the number of instruments connected.
2INSTRUMENT DESCRIPTION
2.3 OPTIONS
Excitation is suitable for supplying power to sensors and transmitters. It has a galvanic separation.
Comparators are assigned to monitor one, two, three or four limit values with relay output. The user may select limits regime:
LIMIT/DOSING/FROM-TO. The limits have adjustable hysteresis within the full range of the display as well as selectable
delay of the switch-on. Reaching the preset limits is signalled by LED and simultaneously by the switch-on of the relevant relay.
Data outputs are for their rate and accuracy suitable for transmission of the measured data for further projection or directly
into the control systems. We offer an isolated RS232 and RS485 with the ASCII or DIN MessBus protocol.
Analog outputs will find their place in applications where further evaluating or processing of measured data is required in
external devices. We offer universal analog output with the option of selection of the type of output - voltage/current. The
value of analog output corresponds with the displayed data and its type and range are selectable in Menu.
Measured data record is an internal time control of data collection. It is suitable where it is necessary to register measured
values. Two modes may be used. FAST is designed for fast storage (40 records/s) of all measured values up to 8000
records. Second mode is RTC, where data record is governed by Real Time with data storage in a selected time segment
and cycle. Up to 250 000 values may be stored in the instrument memory. Data transmission into PC via serial interface
RS232/485 andOM Link.
INSTRUCTIONS FOR USE OMB 450UNI
|
5
INSTRUMENT CONNECTION3
The instrument supply leads should not be in proximity of the incoming low-potential signals. Contactors, motors with larger
input power should not be in proximity of the instrument. The leads into the instrument input (measured quantity) should be
in sufficient distance from all power leads andappliances. Provided this cannot be secured it is necessary to use shielded
leads with connection to ground (bracket E).
The instruments are tested in compliance with standards for use in industrial area, yet we recommend to abide by the above
mentioned principles.
MEASURING RANGES
TypeInput IInput U
DC
PM
OHM
RTD-Pt
RTD-Cu
RTD-Ni
T/C
DU
0…5/20 mA/4…20 mA±2/±5/±10/±40 V
0…0,1/1/10/100 kΪ/Autorange
Pt 100/Pt 500/ Pt 1 000
Cu 50/100
Ni 1 000/10 000
J/K/T/E/B/S/R/N/L
Linear potentiometer (min. 500 Ϊ)
OPTION “A”
TypeInput IInput U
DC
±0,1 A/±0,25 A/±0,5 A to GND (C)
±2 A/±5 A to GND (B)
0…60/150/300/1 200 mV
±100 V/±250 V/±500 V to GND (C)
|
6
INSTRUCTIONS FOR USE OMB 450UNI
123
-
+
L
N
POWER SUPPLY
3INSTRUMENT CONNECTION
OMB 451UQC
INPUT C2/Reset
+
INPUT B2
+
INPUT A2
+
INPUT C1/Reset
+
+
INPUT B1
+
INPUT A1
-
GND
ABCD EF GH
RxD/L+
TxD/L-
IV V VI
GND
III II I
GND
Analog output - I
Analog output - U
C 1
789
L2
E 2
C 2
10 11 12
L3
E 3
C 3
13 14 15
L4
E 4
C 4
Relays
Open collector
16 17 18 19 20 21 22 23
EXT. 3
456
E
L1
E 1
DB
CEA
!
-
+
+
GND
Excitation
INPUT - I
INPUT - U
EXT. 1
EXT. 2
S+ S- E--E+
S+
E+
ES+
-
+
-
2524
ES-ES-
+
+
INPUT - U
+
*GND - U/I0,5
-
*GND - I5
-
INPUT - I
+
CJC
DC, PM
RTD, OHM, Ni
T/C
DU
Excitation
Option A
INPUT - U
INPUT - U
GND - U
GND - I
INPUT - I
PWR
!
Excitation has the minus pole common with the
input - the bracket no. 22 - GND and you may set its
value by trimmer above the bracket no. 16
Maximum of 250 mA may be connected to “INPUT - I” (bracket no. 23) , i.e. 10-times range overload.
Mind the correct connection/mistaking of current - voltage input.
!
Destruction of measuring resistance in current input (15R) may occur.
INSTRUCTIONS FOR USE OMB 450UNI
|
7
INSTRUMENT SETTING4
Setting PROFISetting LIGHT
• For expert users
• Complete instrument menu
• Access is password protected
• Possibility to arrange items of the
„User“ menu
• Tree menu structure
• For trained users
• Only items necessary for instrument
setting
• Access is password protected
• Possibility to arrange items of the
„User“ menu
• Linear menu structure
Setting USER
|
8
INSTRUCTIONS FOR USE OMB 450UNI
• For user operation
• Menu items are set by the user
(Profi/Light) as per request
• Access is not password protected
• Optional menu structure either tree
(PROFI) or linear (LIGHT)
4.1 SETTING
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument
are performed in three adjusting modes:
LIGHT Simple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code
PROFI Complete programming menu
- contains complete instrument menu and is protected by optional number code
USER User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine
the right (see or change)
- acces without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface, which is a standard
equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable
to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET
instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
Scheme of processing the measured signal
4INSTRUMENT SETTING
Input
CHAN. A
Linearization
FILTER
MAT.FCE.
Linearization
Min/Max
Relays
Analog
Data
Memory
Display
INSTRUCTIONS FOR USE OMB 450UNI
|
9
INSTRUMENT SETTING4
Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of
these keys it is possble to browse through the operation menu and to select and set required values.
Indication of measured input (red/green/orange LED)
Relay status
(red/green/orange LED)
ON the digit is lit
OFF the digit is not lit
OFF the digit is flashing
limits withrestriction
(hysteresis, delay)
Measured value (red/green LED)
the last two places may display
measuring units
0500200400100300
LCD scale
Keys control
372.4~C
12T34
M
Indication of measured input
input number
(only for option “A”)
Symbols used in the instructions
DC PM DU OHM RTD T/C
DEF
values preset from manufacture
42
symbol indicates a flashing light (symbol)
MIN
inverted triangle indicates the item that can be placed in USER menu
CONNEC.
broken line indicates a dynamic item, i.e. it is displayed only in particular selection/version
(green LED)
Indicates the setting for given type of instrument
Function (green LED)
M Min/max. value
T Tare
increasing the value by rotating the ke to the right (UP)decreasing the value by rotating the ke to the left (DOWN)
pressing the key shortly
pressing the key for longer than 2 seconds (> 2 s.),
after pressing the key the set value will not be stored
after pressing the key the set value will be stored
30
continues on page 30
|
10
INSTRUCTIONS FOR USE OMB 450UNI
Setting the decimal point and the minus sign
DECIMAL POINT
Its selection in the menu, upon modification of the number to be adjusted it is performed by the short push key
transition beyond the highest decade, when the decimal point starts flashing. Positioning is performed by / .
THE MINUS SIGN
Setting the minus sign is performed by the key
the current number (e.g..: 013 >
Control keys functions
KeyMeasurementMenuSetting numbers/selection
access into USER menuexit menuquit editing
programmable key functionback to previous levelmove to higher decade
programmable key functionmove to previous itemmove down
programmable key functionmove to next itemmove up
, on class 100 > -87)
+
programmable key functionmove to next levelconfi rm setting/selection
+
+
+
+
direct access into PROFI menu
access into LIGHT/PROFI menu
on higher decade. When editing the item substraction must be made from
confi rm setting after selection
numeric value
numeric value is set to zero
confi guration of an item for
“USER” menu
determine the sequence of items
in “USER - LIGHT” menu
4INSTRUMENT SETTING
with
Setting items into „USER“ menu
• inLIGHT or PROFI menu
• no items permitted in USER menu from manufacture
• on items marked by inverted triangle
---
legend is flashing - current setting is displayed
+
NOYESSHOW
return to item
INSTRUCTIONS FOR USE OMB 450UNI
---
|
11
light
5
SETTING
5.0 SETTING “LIGHT”
LIGHT Simple programming menu
- contains only items necessary for instrument setting and is protected by optional number code
Setting LIGHT
light
• For capable users
• Only items necessary for instrument
setting
• Access is password protected
• Possibility to arrange items of the
„User“ menu
• Linear menu structure
Preset from manufacture
Password “0”
Menu LIGHT
USER menu of f
Setting the items
DEF
!
Upon delay exceeding 60 s the programming mode
is automatically discontinued and the instrument
itself restores the measuring mode
|
12
INSTRUCTIONS FOR USE OMB 450UNI
light
light
5SETTING
142.8PASSW.0
Type of instruments
OHMRTD
T/C
PMDCDUOHM
Limit 1Limit 4Limit 3Limit 2
LIM . L.120LIM . L.240LIM . L.360LIM . L.480
Setting type A.O.End A.O.Beginnig A.O.
TYP . A .O.MIN. A .O.MAX. A.O.10004-20
Bargraph projection - beginning
Colour
DU
C. MIN.YESC. MAX.YES
Access password
+
Measuring range
PMTYP E
2-WIRECONNEC.00000. oFO RM. A
EXT. 1TCCONNEC.23C.J .TEM .FOR M. A000000
0MIN A100MAX AFO RM. A
Bargraph projection - end
0MIN. BG .
Menu type
CERV.CO LOR
4-20m AMODE
100MAX. BG.
MENULIG HT
0000. oo
LCD scale division
Return to manufacture calibration
RE. CAL .YES
Calibration - only for “DU”
Selecting projection and connection
Level of LCD backlight
5 DIV.DIVIS.
Return to manufacture setting
RE. SET.YES
Option - comparator
Option - Analog output
100%BACKL.
Language selection
LANG.CZECH
Identifi cation
New password
PAS. LI.0
Type of instruments
OMB 452UNI
Input
SW: version
65-0014-20
Return to measur-
142.8IDENT .YES
ing mode
INSTRUCTIONS FOR USE OMB 450UNI
|
13
5
light
SETTING
142.8
light
PASSW .0
TYP E
Type „DC“ 16
18
Type “PM”
20
Type “OHM“
22
Type “RTD-Pt“
24
Type “RTD-Ni“
26
Type “T/C“
28
Type “DU”
30
Type “RTD-Cu“
|
14
INSTRUCTIONS FOR USE OMB 450UNI
Entering access password
for access into the menu
Access into instrument
PASSW .
menu
PAS = 0
- access into menu is unrestricted, after
releasing keys you automaticaly move to first
item of the menu
Set „Pas sword” = 42 Example
0120212
32TYP E
DCPMOHM RTD-P t RTD-NiTC
DU
TYP E
- primary selection of the type of instrument
- performs default setting
from manufacture, incl. calibration