Orbit Merret OM 652UC Instruction Manual

OM 652UC
6 DIGIT PROGRAMMABLE
IMPULSE COUNTER/FREQUENCY METER
STOPWATCH/TIMER
GUARANTEE
YEARS
|
INSTRUCTIONS FOR USE OM 652UC
SAFETY INSTRUCTIONS
Please, read the enclosed safety instructions carefully and observe them! These instruments should be safeguarded by isolated or common fuses (breakers)! For safety information the EN 61 010-1 + A2 standard must be observed. This instrument is not explosion-safe!
TECHNICAL DATA
Measuring instruments of the OM 652 series conform to the European regulation 89/336/EWG and the Ordinance 168/1997 Coll.
They are up to the following European: EN 55 022, class B EN 61000-4-2, -4, -5, -6, -8, -9, -10, -11
The instruments are applicable for unlimited use in agricultural and industrial areas.
CONNECTION
Supply of energy from the main line has to be isolated from the measuring leads.
ORBIT MERRET, spol. s r.o.
Vodňanská 675/30 198 00 Prague 9 Czech Republic
Tel: +420 - 281 040 200 Fax: +420 - 281 040 299 e-mail: orbit@merret.cz www.orbit.merret.cz
INSTRUCTIONS FOR USE OM 652UC
|
3
1. Contens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. Instrument description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3. Instrument connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4. Instrument setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Symbols used in the instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Setting the DP andthe (-) sing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Control keys functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Configuration of „User“ menu items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5. Setting “LIGHT” menu
5.0 Description “LIGHT” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Access to menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting initial value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Selection of instrument measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Selection of control START
H
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Selection of control STOP
H
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Selection of digital input filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Selection of type of input A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting level of input A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selection of type of input C (Resetting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting level of input C (Resetting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting multiplying constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting dividing constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting offset - PRESET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Selection of projection format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting the menu type (LIGHT/PROFI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Restoration of manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6. Setting “PROFI” menu
6.0 Description “PROFI” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.1 “PROFI” menu - INPUT
6.1.1 Resetting internal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1.2 Instrument configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
6.1.3 Setting external control input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.1.4 Setting function of the control key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.2 “PROFI” menu - CHANNELS
6.2.1 Setting calibration constants and offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2.2 Setting digital filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2.3 Projection format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
6.3 “PROFI” menu - OUTPUTS
6.3.1 Configuration and setting the limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.2 Setting data output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.3.3 Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3.4 Setting display brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.4 “PROFI” menu - SERVICE
6.4.1 Selection of the type of programming menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.4.2 Restoration of manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4.3 Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4.4 Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7. Setting “USER” menu
7.0 Configuration “USER” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
8. Data protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
9. Error statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
10. Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
11. Instrument dimension and installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12. Certificate of guarantee . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
1CONTENS
|
INSTRUCTIONS FOR USE OM 652UC
The OM 652UC model is a universal 6 digit panel programmable impulse counter/frequencymeter andstopwatch/timer. The instrument is based on an 8-bit microprocessor, which secures high accuracy, stability and easy operation of the
instrument.
Measuring modes
COUNTER Single-channel counter
C
FREQUENCY Frequencymeter
F
STOPWATCH Stopwatch
H
TIMER Timer
H
PROGRAMMABLE PROJECTION
Calibration calibration coefficient may be set in„CM“ Projection -99999...999999 with fixed or floating DP, for measuring modes STOPWATCH/TIMER with the option of setting in format 10/24/60 Time base: 0,5/1/5/10 s
DIGITAL FILTERS
Input filter the instrument enables filtering the input signal and thus suppress unwatnted interfering signals (e.g. relay backswings). The parameter set gives maximum feasible measured frequency processed by the instrument, 5/40/100/1 000 Hz Exponen.average: from 2…100 measurements
LINEARIZATION
Linearization:* by linear interpolation in 25 points (solely via OM Link)
FUCTIONS
Preset initial non-zero value, unloaded always after instrument resetting Rounding setting projection step for display Tare*: designed to reset display upon non-zero input signal
OM Link company interface for control, setting and instrument update
EXTERNAL CONTROL
Hold display/instrument blocking Lock locking the control keys for access into Configuration menu Resettting resetting/pre-setting the counter Tare* tare activation Start/Stop stopwatch/timer control
2 INSTRUMENT DESCRIPTION
INSTRUCTIONS FOR USE OM 652UC
|
5
2.2 Operationn
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instru­ment are realized in two adjusting modes:
LIGHT Simple programming menu
- contains only items necessary for instrument setting and is protected by an optional numeral code
PROFI Complete programming menu
- contains complete instrument menu and is protected by an optional numeral code
USER User programmable menu
- may contain arbitrary items selected from programmable menu (LIGHT/PROFI), which determines the authorization (see or change)
- access is without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete operation and setting of the instrument may be performed via communication interface OM Link, which is a standard equipment of every instrument.
The operation program is freely available (www.orbit.merret.cz) and the only requirement is the purchase of OML cable for connecting the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need for OML cable). The OM LINK program version „Standard“ allows you to connect an unlimited number of instruments with the option of
visualizatiion and storage in PC.
2.3 Extension
Excitation is suitable for feeding sensors and converters. It has a galvanic isolation. Comparators are assigned to control two limit values with relay output. The limits have adjustable hysteresis as well as
selectable delay of the switch-on. Reaching the preset limits is signalled by LED and simultaneously by the switch-on of the relevant relay.
Data outputs are for their rate and accuracy suitable for transmission of measured data for further projection or directly into the control systems. We offer an isolated RS232 and RS485 with the ASCII protocol.
Analog outputs will find their place in applications where further evaluating or processing of measured data in external devices is required. We offer a universal analog output with the option of selection of output type - voltage/current. The value of analog output corresponds with the displayed data and its type and range are selectable in the programming mode.
Time backup by means of RTC circuit is designed for the „TIMER“ measuring mode and secures time measuring even if the instrument is switched-off (without display projection).
2INSTRUMENT DESCRIPTION
|
INSTRUCTIONS FOR USE OM 652UC
The instrument supply leads should not be in proximity of the incoming low-potential signals. Contactors, motors with larger input power should not be in proximity of the instrument. The leads into the instrument input (measured quantity) should be in sufficient distance from all power leads and appliances.
Provided this cannot be secured it is necessary to use shielded leads with connection to ground (bracket E). The instruments are tested in compliance with standards for use in industrial area, yet we recommend to abide by the above
mentioned principles.
CONNECTION
Description Connection
Input A (< 30 V)
input signal < 43 V(absolute 60 V) GND + Input A(< 30 V)
Input A (< 300 V)
input signal < 300 V GND + Input A(< 300 V)
Resetting (< 30 V)
input signal < 43 V GND + Resetting (< 30 V)
Resetting (< 300 V)
input signal < 300 V GND + Resetting (< 300 V)
Function Description Control Hold
Blocking display andinstrument outputs upon contact, bracket (no. 11/12)
Lock
Control keys blocking upon contact, bracket (no. 11/12)
Tare
Tare activation upon contact, bracket (no. 11/12)
+
11 12 13 14 15 16 17 18
GND
Reset (< 300 V)
INPUT A (< 300 V)
-
Excitation
Hold/Lock
+
-
54
L
N
POWER SUPPLY
E
123 567
8910
L2
L1
+
+
INPUT A (< 30 V)
232
485
GND
TxD
RxD
GND
Rx/Tx+
Rx/Tx-
GND
AO - I
AO - U
+
+
-
+
-
Reset (< 30 V)
+
3 INSTRUMENT CONNECTION
Grounding on bracket „E“ has to connected at all times
!
INSTRUCTIONS FOR USE OM 652UC
|
7
Table of comparation levels
Input
Type of
input
Maximum input
voltage
(Level A, C)
Maximum comparation levels
L > H H > L
Input A
Resetting
(< 30 V)
NPN, Contact xxx 0,5 V 4,5 V
PNP 9,7 V 0,5 V 4,5 V PNP 14,4 V 1,0 V 9,0 V PNP 19,2 V 1,5 V 13,3 V PNP 23,9 V 2,0 V 17,8 V PNP 28,7 V 2,5 V 22,1 V PNP 33,5 V 3,0 V 26,6 V PNP 38,3 V 3,4 V 31,0 V PNP 43,0 V 3,9 V 35,5 V
Input A
Resetting
(< 300 V)
NPN, Contact !!! prohibited !!!
PNP 84 V 4,9 V 39,8 V PNP 128 V 9,2 V 78,0 V PNP 170 V 13,6 V 117,8 V PNP 211 V 17,8 V 156,0 V PNP 253 V 22,3 V 195,8 V PNP 295 V 26,5 V 234,1 V PNP 301 V 30,9 V 273,9 V
3INSTRUMENT CONNECTION
|
INSTRUCTIONS FOR USE OM 652UC
SETTING PROFISETTING LIGHTSETTING USER
INSTRUMENT SETTING3
• For trained users
• Only items necessary for instrument setting
• Access is password protected
• Possibility to arrange items of the „User“ menu
• Linear menu structure
• For expert users
• Complete instrument menu
• Access is password protected
• Possibility to arrange items of the „User“ menu
• Tree menu structure
• For user operation
• Menu items are set by the user (Profi/Light) as per request
• Access is not password protected
• Optional menu structure either tree (PROFI) or linear (LIGHT)
INSTRUCTIONS FOR USE OM 652UC
|
9
4.1 Setting
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument are performed in three adjusting modes:
LIGHT Simple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code PROFI Complete programming menu
- contains complete instrument menu and is protected by optional number code USER User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine the right (see or change)
- acces without password All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface, which is a standard equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable to con­nect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
3INSTRUMENT SETTING
10
|
INSTRUCTIONS FOR USE OM 652UC
Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of these keys it is possble to browse through the operation menu and to select and set required values.
Symbols used in the instructions
H
Indicates the setting for given type of instrument
DEF
values preset from manufacture
42
symbol indicates a flashing light (symbol)
MI N
inverted triangle indicates the item that can be placed in USER menu
CONECT.
broken line indicates a dynamic item, i.e. it is displayed only in particular selection/version
after pressing the key the set value will not be stored
after pressing the key the set value will be stored
30
continues on page 30
Setting the decimal point and the minus sign
DECIMAL POINT
Its selection in the menu, upon modification of the number to be adjusted it is performed by the control key with transition beyond the highest decade, when the decimal point starts flashing . Positioning is performed by / .
THE MINUS SIGN
Setting the minus sign is performed by the key on higher decade. When editing the item substraction must be made from the current number (e.g..: 013 >
, on class 100 > -87)
12
Relay status
ON the digit is lit OFF the digit is not lit OFF the digit is flashing
limits with restriction
(hysteresis, delay)
Measured value
4 INSTRUMENT SETTING
INSTRUCTIONS FOR USE OM 652UC
|
11
Control keys functions
Setting items into „USER“ menu
• in LIGHT or PROFI menu
• no items permitted in USER menu from manufacture
• on items marked by inverted triangle
NO
item will not be displayed in USER menu
YES
item will be displayed in USER menu with the option of setting
SHO
item will be solely displayed in USER menu
4INSTRUMENT SETTING
Key Measurement Menu Setting numbers/Selection
access into USER menu exit menu w/o saving
transition to next item w/o saving
tare value return to previous level move to higher decade
cancel Tare move to previous item move down
cancel Tare move to next item move up
Tare confi rm selection setting/selection confi rmation
+
numeric value is set to zero
+
access into LIGHT/PROFI menu
+
direct access into PROFI menu ­temporary (remains LIGHT)
+
confi guration of an item for “USER” menu
+
determine the sequence of items in “USER - LIGHT” menu
---
+
NO YES SHO
---
return to item
legend is flashing - current setting is displayed
12
|
INSTRUCTIONS FOR USE OM 652UC
5.0 Setting “LIGHT”
LIGHT Simple programming menu
- contains only items necessary for instrument setting and is protected by optional numeral code
SETTING LIGHT
5 SETTING
lightlight
• For capable users
• Only items necessary for instrument setting
• Password protected access
• Possibility to arrange items of the „User“ menu
• Linear menu structure
Password “0” Menu LIGHT USR menu off Setting the items
DEF
Preset from manufacture
INSTRUCTIONS FOR USE OM 652UC
|
13
Return to previous measuring mode
TYP. A. O . MI N . A . O. MAX. A .O. 10004-20mA
LI M . L.1 25 LI M . L. 2 75
MENU LI G H T
IDENT.
142. 8
OM 652. . . .
N. PASS. 0FI R M . YES
YES
142. 8
PASSW.
0
+
FREQV .
MODE
OFFFI LTER NPN.CO N .TYPE A 10LEV ELA
NPN.CON .TYPE C 10LEV ELC 1SCALE
0OFFSET
SET V .
0
CLEA RM . STOPCONTAC.M. START
000000FORM . A1DI VI D .
5SETTING
lightlight
Option - Analog output
Option - Comparator
Menu type
Identifi cation
New password
Return to manufacture setting
Access code
Only for Mode > TIME and RTC
Upon delay exceeding 60 s the programming mode is automatically discontinued and the instrument itself restores the measuring mode
!
14
|
INSTRUCTIONS FOR USE OM 652UC
142. 8
PASS. 0
Entering access password for access into the menu
PASS.
Access into instrument menu
PAS = 0
- access into menu is unrestricted, after releasing keys you automaticaly move to first item of the menu
PAS > 0
- access into menu is protected ny numeric code
Set “PASS.” = 42 Example
0 1 2 02 12
42
22
32 SET V.
SET V .
Setting initial value
- the function allows the user a single-time setting of initial value of display projection
- the instrument is preset from manufacture into “FREQU.” measuring mode and unless another mode is set the item remains hidden
- if you need to set initial value for another mode it is necessary to do so upon next access to programming menu > after
change of measuring mode
- setting “SET V.” is a one-time operation unlike the “OFFSET” option, i.e. after resetting the display value is “0”, provided there is no other value set in the “OFFSET” item
DEF
= 0
SET V .
0
Setting initial value
The item „SET V.“ is not projected for measuring mode „FREQU.“
!
Set “S ET V.” = 233 Example
0 1 2 3 03
33
13
23 MODE233133033
+
5 SETTING
lightlight
INSTRUCTIONS FOR USE OM 652UC
|
15
COUN t . FREQV .
MODE
Selection of instrument measuring mode
- elementary selection of instrument type
DEF
= FREQU.
MODE
Menu Instrument mode
COUNT. Counter FREQU. Frequencymeter TIME Stopwatch/timer RTC Stopwatch/backup timer
MODE
RTCTI M E
Selection of “COUNT.” mode Example
FILTE RFREQV .
COU Nt .
M. START
Selection of stopwatch/ timer control
- menu for time setting is accessible only in the stopwatch/timer mode
CONTIN.
Stopwatch/timer is running constantly if the instrument
is turned on
CONTAC.
Stopwatch/timer is running upon contact making
EDGE
Stopwatch/timer is controlled by the priming
signal edge
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
RU n . ST.C.
Stopwatch/timer is controlled and reset by the
edge of the priming signal
- time is set off by the edge (by the signal
passing across the comparing level) and stopped by the next edge
C.RU n . ST.
Stopwatch/timer is controlled and reset by the
edge of the priming signal
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
CLR.RU N .
Stopwatch/timer is reset and set off by the edge of
the priming signal
CL. RU .RE .
Stopwatch/timer is reset
and set off by the edge of the priming signal, the cycle is repeated with every other edge
RUN
Stopwatch/timer is only set
off by the edge
DEF
= COntAC.
M. START
CONTIN. CONTAC. EDGE RU N .ST.C.
C.RUN .ST. CLR.RU N CL.RU .RE . RU N
>
>
Selection of stopwatch control > EdGE Example
M. STOPCONTAC. EDGE
Only for measuring mode
TIME • RTC
5SETTING
lightlight
16
|
INSTRUCTIONS FOR USE OM 652UC
Only for measuring mode
TIME • RTC
Only for measuring mode
COUNT. • TIME • RTC
5 SETTING
lightlight
M. STOP
Selection of stopwatch resetting
- menu of the resetting option is accessible only in the stopwatch/timer regime
DEF
= CLEAR
CLEA R
Stopwatch/timer is reset through input „Clear“
ST. CLR.
Stopwatch/timer is stopped and reset through input
„Clear“
STOP
Stopwatch/timer is stopped through input „Clear“
Selection of type of stopwatch resetting > St. CLr. Example
CLEAR ST. CLR. FILTE R
M. STOP
CLEA R ST. CLR. STOP
OFF 1000 100 40 5
FI LTER
Selection of digital filter
- digital filter may suppress unwanted interfering impulses (e.g. relay backswings) on the input signal. The set parameter gives maximum possible frequency, which the instrument processes w/o limitation
DEF
= OFF
FI LTER
Maximum input frequency 100 Hz > 100 Example
TYP E A .OFF 1000 100
When accessing upon contact and available maximum input frequency we recommend using filter
!
INSTRUCTIONS FOR USE OM 652UC
|
17
5SETTING
lightlight
10
Setting input level
LEV EL .A
Setting input level for Input A
- setting applies for Input A
- setting level (only for type PNP) of the input voltage, the instrument subsequently automatically selects divider and thus the comparing levels
- range of setting 0…43 V(Input < 30 V)
- range of setting 43…300 V(Input < 300 V)
- table of comparing levels is on page 7
Setting the level = 34 V Example
10 11 12 13 14
34
14
24 TYP E C.
LEV EL .A
NPN.CON . PNP
TYP E A
Selection of type of input
- setting applies for Input A
DEF
= NPN.CON.
TYPE
Menu Type of input
NPN.CON. NPN or contact PNP PNP
Selection of type of inpu t > NPN Example
LEV EL. ANPN.CO N.
TYP E A
Input levels (Level A) have to be set after this option
!
18
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
NPN.C PNP
TYP E C
Selection of type of input
- setting applies for resetting input
DEF
= NPN.CON.
TYPE
Menu Type of input
NPN.CON. NPN or contact PNP PNP
Selection of type of inpu t > NPN Example
LEV EL. CNPN.CO N.
TYP E C
Input levels (Level C) have to be set after this option
!
10
Setting input level
LEV EL .C
Setting input level for resetting input
- setting the level (only for PNP type) of input voltage, the instrument subsequently automatically selects divider and thus the comparing levels
- range of setting 0…43 V(Input < 30 V)
- range of setting 43…300 V(Input < 300 V)
- table of comparing levels is on page 7
Setting levels = 34 V Example
10 11 12 13 14
34
14
24 SCALE
LEV EL .C
INSTRUCTIONS FOR USE OM 652UC
|
19
5SETTING
lightlight
1
Setting division constant
DI VI D .
Setting division constant
- calibration constant is for calculation of the input value to required display value
- range: -0,00001...999999
-
DEF
= 1
Division constant = 22 Example
1 2 02 12 22 OFFSET
DI VI D .
1
Setting multiplying­constant
SCALE
Setting multiplying constant
- calibration constant serves for calculation of the input value to required display value
- by entering minus value direction of the calculation is changed, i.e. we count down
- range: -0,00001...999999
-
DEF
= 1
Cali bration consta nt = 3,12 Example
1 2 02 12 012 112
212
DI VI D .
312 312 .0312 31 . 2 3 .12
SCALE
20
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
0
Setting PRESET
OFFSET
Setting additive constant
- PRESET
- offset of the measuring by a set value, which shall be loaded always upon instrument resetting
- range: -99999...999999
(+ time formats)
-
DEF
= 0
Setting „PRESE T“ = 24 Example
0 1 2 3 4
24
04
14 FORM . A
OFFSET
FORM . A
000000 00000. o 0000. oo 000. ooo 00. oooo 0. ooooo FLO A . P.
FORM . A
Selection of projection format
- instrument enables classical projection of number with fixed position of decimal point as well as projection with floating allowing fo projection of number in its most precise form „FLOA. P.“
- for measuring modes „TIME“ and „RTC“ special time formats are preset
DEF
= 000000
DEF
= HH.MM.SS
H
Projection of DP on display > 00000.o Example
LI M . L. 1
* subsequent menu item depends on instrument equipment
000000 00000. o
99. MM.SS HH. MM HHHH. MM MMMM.SS MM.SS.CC 99.SS . CC H . MM.SS.CHH. MM. SS
F C
H
Loading...
+ 44 hidden pages