Orbit Merret OM 652UC Instruction Manual

OM 652UC
6 DIGIT PROGRAMMABLE
IMPULSE COUNTER/FREQUENCY METER
STOPWATCH/TIMER
GUARANTEE
YEARS
|
INSTRUCTIONS FOR USE OM 652UC
SAFETY INSTRUCTIONS
Please, read the enclosed safety instructions carefully and observe them! These instruments should be safeguarded by isolated or common fuses (breakers)! For safety information the EN 61 010-1 + A2 standard must be observed. This instrument is not explosion-safe!
TECHNICAL DATA
Measuring instruments of the OM 652 series conform to the European regulation 89/336/EWG and the Ordinance 168/1997 Coll.
They are up to the following European: EN 55 022, class B EN 61000-4-2, -4, -5, -6, -8, -9, -10, -11
The instruments are applicable for unlimited use in agricultural and industrial areas.
CONNECTION
Supply of energy from the main line has to be isolated from the measuring leads.
ORBIT MERRET, spol. s r.o.
Vodňanská 675/30 198 00 Prague 9 Czech Republic
Tel: +420 - 281 040 200 Fax: +420 - 281 040 299 e-mail: orbit@merret.cz www.orbit.merret.cz
INSTRUCTIONS FOR USE OM 652UC
|
3
1. Contens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. Instrument description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3. Instrument connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4. Instrument setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Symbols used in the instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Setting the DP andthe (-) sing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Control keys functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Configuration of „User“ menu items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5. Setting “LIGHT” menu
5.0 Description “LIGHT” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Access to menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting initial value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Selection of instrument measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Selection of control START
H
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Selection of control STOP
H
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Selection of digital input filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Selection of type of input A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting level of input A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selection of type of input C (Resetting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting level of input C (Resetting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting multiplying constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting dividing constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting offset - PRESET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Selection of projection format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting the menu type (LIGHT/PROFI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Restoration of manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6. Setting “PROFI” menu
6.0 Description “PROFI” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.1 “PROFI” menu - INPUT
6.1.1 Resetting internal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1.2 Instrument configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
6.1.3 Setting external control input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.1.4 Setting function of the control key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.2 “PROFI” menu - CHANNELS
6.2.1 Setting calibration constants and offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2.2 Setting digital filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2.3 Projection format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
6.3 “PROFI” menu - OUTPUTS
6.3.1 Configuration and setting the limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.2 Setting data output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.3.3 Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3.4 Setting display brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.4 “PROFI” menu - SERVICE
6.4.1 Selection of the type of programming menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.4.2 Restoration of manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4.3 Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4.4 Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7. Setting “USER” menu
7.0 Configuration “USER” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
8. Data protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
9. Error statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
10. Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
11. Instrument dimension and installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12. Certificate of guarantee . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
1CONTENS
|
INSTRUCTIONS FOR USE OM 652UC
The OM 652UC model is a universal 6 digit panel programmable impulse counter/frequencymeter andstopwatch/timer. The instrument is based on an 8-bit microprocessor, which secures high accuracy, stability and easy operation of the
instrument.
Measuring modes
COUNTER Single-channel counter
C
FREQUENCY Frequencymeter
F
STOPWATCH Stopwatch
H
TIMER Timer
H
PROGRAMMABLE PROJECTION
Calibration calibration coefficient may be set in„CM“ Projection -99999...999999 with fixed or floating DP, for measuring modes STOPWATCH/TIMER with the option of setting in format 10/24/60 Time base: 0,5/1/5/10 s
DIGITAL FILTERS
Input filter the instrument enables filtering the input signal and thus suppress unwatnted interfering signals (e.g. relay backswings). The parameter set gives maximum feasible measured frequency processed by the instrument, 5/40/100/1 000 Hz Exponen.average: from 2…100 measurements
LINEARIZATION
Linearization:* by linear interpolation in 25 points (solely via OM Link)
FUCTIONS
Preset initial non-zero value, unloaded always after instrument resetting Rounding setting projection step for display Tare*: designed to reset display upon non-zero input signal
OM Link company interface for control, setting and instrument update
EXTERNAL CONTROL
Hold display/instrument blocking Lock locking the control keys for access into Configuration menu Resettting resetting/pre-setting the counter Tare* tare activation Start/Stop stopwatch/timer control
2 INSTRUMENT DESCRIPTION
INSTRUCTIONS FOR USE OM 652UC
|
5
2.2 Operationn
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instru­ment are realized in two adjusting modes:
LIGHT Simple programming menu
- contains only items necessary for instrument setting and is protected by an optional numeral code
PROFI Complete programming menu
- contains complete instrument menu and is protected by an optional numeral code
USER User programmable menu
- may contain arbitrary items selected from programmable menu (LIGHT/PROFI), which determines the authorization (see or change)
- access is without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete operation and setting of the instrument may be performed via communication interface OM Link, which is a standard equipment of every instrument.
The operation program is freely available (www.orbit.merret.cz) and the only requirement is the purchase of OML cable for connecting the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need for OML cable). The OM LINK program version „Standard“ allows you to connect an unlimited number of instruments with the option of
visualizatiion and storage in PC.
2.3 Extension
Excitation is suitable for feeding sensors and converters. It has a galvanic isolation. Comparators are assigned to control two limit values with relay output. The limits have adjustable hysteresis as well as
selectable delay of the switch-on. Reaching the preset limits is signalled by LED and simultaneously by the switch-on of the relevant relay.
Data outputs are for their rate and accuracy suitable for transmission of measured data for further projection or directly into the control systems. We offer an isolated RS232 and RS485 with the ASCII protocol.
Analog outputs will find their place in applications where further evaluating or processing of measured data in external devices is required. We offer a universal analog output with the option of selection of output type - voltage/current. The value of analog output corresponds with the displayed data and its type and range are selectable in the programming mode.
Time backup by means of RTC circuit is designed for the „TIMER“ measuring mode and secures time measuring even if the instrument is switched-off (without display projection).
2INSTRUMENT DESCRIPTION
|
INSTRUCTIONS FOR USE OM 652UC
The instrument supply leads should not be in proximity of the incoming low-potential signals. Contactors, motors with larger input power should not be in proximity of the instrument. The leads into the instrument input (measured quantity) should be in sufficient distance from all power leads and appliances.
Provided this cannot be secured it is necessary to use shielded leads with connection to ground (bracket E). The instruments are tested in compliance with standards for use in industrial area, yet we recommend to abide by the above
mentioned principles.
CONNECTION
Description Connection
Input A (< 30 V)
input signal < 43 V(absolute 60 V) GND + Input A(< 30 V)
Input A (< 300 V)
input signal < 300 V GND + Input A(< 300 V)
Resetting (< 30 V)
input signal < 43 V GND + Resetting (< 30 V)
Resetting (< 300 V)
input signal < 300 V GND + Resetting (< 300 V)
Function Description Control Hold
Blocking display andinstrument outputs upon contact, bracket (no. 11/12)
Lock
Control keys blocking upon contact, bracket (no. 11/12)
Tare
Tare activation upon contact, bracket (no. 11/12)
+
11 12 13 14 15 16 17 18
GND
Reset (< 300 V)
INPUT A (< 300 V)
-
Excitation
Hold/Lock
+
-
54
L
N
POWER SUPPLY
E
123 567
8910
L2
L1
+
+
INPUT A (< 30 V)
232
485
GND
TxD
RxD
GND
Rx/Tx+
Rx/Tx-
GND
AO - I
AO - U
+
+
-
+
-
Reset (< 30 V)
+
3 INSTRUMENT CONNECTION
Grounding on bracket „E“ has to connected at all times
!
INSTRUCTIONS FOR USE OM 652UC
|
7
Table of comparation levels
Input
Type of
input
Maximum input
voltage
(Level A, C)
Maximum comparation levels
L > H H > L
Input A
Resetting
(< 30 V)
NPN, Contact xxx 0,5 V 4,5 V
PNP 9,7 V 0,5 V 4,5 V PNP 14,4 V 1,0 V 9,0 V PNP 19,2 V 1,5 V 13,3 V PNP 23,9 V 2,0 V 17,8 V PNP 28,7 V 2,5 V 22,1 V PNP 33,5 V 3,0 V 26,6 V PNP 38,3 V 3,4 V 31,0 V PNP 43,0 V 3,9 V 35,5 V
Input A
Resetting
(< 300 V)
NPN, Contact !!! prohibited !!!
PNP 84 V 4,9 V 39,8 V PNP 128 V 9,2 V 78,0 V PNP 170 V 13,6 V 117,8 V PNP 211 V 17,8 V 156,0 V PNP 253 V 22,3 V 195,8 V PNP 295 V 26,5 V 234,1 V PNP 301 V 30,9 V 273,9 V
3INSTRUMENT CONNECTION
|
INSTRUCTIONS FOR USE OM 652UC
SETTING PROFISETTING LIGHTSETTING USER
INSTRUMENT SETTING3
• For trained users
• Only items necessary for instrument setting
• Access is password protected
• Possibility to arrange items of the „User“ menu
• Linear menu structure
• For expert users
• Complete instrument menu
• Access is password protected
• Possibility to arrange items of the „User“ menu
• Tree menu structure
• For user operation
• Menu items are set by the user (Profi/Light) as per request
• Access is not password protected
• Optional menu structure either tree (PROFI) or linear (LIGHT)
INSTRUCTIONS FOR USE OM 652UC
|
9
4.1 Setting
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument are performed in three adjusting modes:
LIGHT Simple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code PROFI Complete programming menu
- contains complete instrument menu and is protected by optional number code USER User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine the right (see or change)
- acces without password All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface, which is a standard equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable to con­nect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
3INSTRUMENT SETTING
10
|
INSTRUCTIONS FOR USE OM 652UC
Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of these keys it is possble to browse through the operation menu and to select and set required values.
Symbols used in the instructions
H
Indicates the setting for given type of instrument
DEF
values preset from manufacture
42
symbol indicates a flashing light (symbol)
MI N
inverted triangle indicates the item that can be placed in USER menu
CONECT.
broken line indicates a dynamic item, i.e. it is displayed only in particular selection/version
after pressing the key the set value will not be stored
after pressing the key the set value will be stored
30
continues on page 30
Setting the decimal point and the minus sign
DECIMAL POINT
Its selection in the menu, upon modification of the number to be adjusted it is performed by the control key with transition beyond the highest decade, when the decimal point starts flashing . Positioning is performed by / .
THE MINUS SIGN
Setting the minus sign is performed by the key on higher decade. When editing the item substraction must be made from the current number (e.g..: 013 >
, on class 100 > -87)
12
Relay status
ON the digit is lit OFF the digit is not lit OFF the digit is flashing
limits with restriction
(hysteresis, delay)
Measured value
4 INSTRUMENT SETTING
INSTRUCTIONS FOR USE OM 652UC
|
11
Control keys functions
Setting items into „USER“ menu
• in LIGHT or PROFI menu
• no items permitted in USER menu from manufacture
• on items marked by inverted triangle
NO
item will not be displayed in USER menu
YES
item will be displayed in USER menu with the option of setting
SHO
item will be solely displayed in USER menu
4INSTRUMENT SETTING
Key Measurement Menu Setting numbers/Selection
access into USER menu exit menu w/o saving
transition to next item w/o saving
tare value return to previous level move to higher decade
cancel Tare move to previous item move down
cancel Tare move to next item move up
Tare confi rm selection setting/selection confi rmation
+
numeric value is set to zero
+
access into LIGHT/PROFI menu
+
direct access into PROFI menu ­temporary (remains LIGHT)
+
confi guration of an item for “USER” menu
+
determine the sequence of items in “USER - LIGHT” menu
---
+
NO YES SHO
---
return to item
legend is flashing - current setting is displayed
12
|
INSTRUCTIONS FOR USE OM 652UC
5.0 Setting “LIGHT”
LIGHT Simple programming menu
- contains only items necessary for instrument setting and is protected by optional numeral code
SETTING LIGHT
5 SETTING
lightlight
• For capable users
• Only items necessary for instrument setting
• Password protected access
• Possibility to arrange items of the „User“ menu
• Linear menu structure
Password “0” Menu LIGHT USR menu off Setting the items
DEF
Preset from manufacture
INSTRUCTIONS FOR USE OM 652UC
|
13
Return to previous measuring mode
TYP. A. O . MI N . A . O. MAX. A .O. 10004-20mA
LI M . L.1 25 LI M . L. 2 75
MENU LI G H T
IDENT.
142. 8
OM 652. . . .
N. PASS. 0FI R M . YES
YES
142. 8
PASSW.
0
+
FREQV .
MODE
OFFFI LTER NPN.CO N .TYPE A 10LEV ELA
NPN.CON .TYPE C 10LEV ELC 1SCALE
0OFFSET
SET V .
0
CLEA RM . STOPCONTAC.M. START
000000FORM . A1DI VI D .
5SETTING
lightlight
Option - Analog output
Option - Comparator
Menu type
Identifi cation
New password
Return to manufacture setting
Access code
Only for Mode > TIME and RTC
Upon delay exceeding 60 s the programming mode is automatically discontinued and the instrument itself restores the measuring mode
!
14
|
INSTRUCTIONS FOR USE OM 652UC
142. 8
PASS. 0
Entering access password for access into the menu
PASS.
Access into instrument menu
PAS = 0
- access into menu is unrestricted, after releasing keys you automaticaly move to first item of the menu
PAS > 0
- access into menu is protected ny numeric code
Set “PASS.” = 42 Example
0 1 2 02 12
42
22
32 SET V.
SET V .
Setting initial value
- the function allows the user a single-time setting of initial value of display projection
- the instrument is preset from manufacture into “FREQU.” measuring mode and unless another mode is set the item remains hidden
- if you need to set initial value for another mode it is necessary to do so upon next access to programming menu > after
change of measuring mode
- setting “SET V.” is a one-time operation unlike the “OFFSET” option, i.e. after resetting the display value is “0”, provided there is no other value set in the “OFFSET” item
DEF
= 0
SET V .
0
Setting initial value
The item „SET V.“ is not projected for measuring mode „FREQU.“
!
Set “S ET V.” = 233 Example
0 1 2 3 03
33
13
23 MODE233133033
+
5 SETTING
lightlight
INSTRUCTIONS FOR USE OM 652UC
|
15
COUN t . FREQV .
MODE
Selection of instrument measuring mode
- elementary selection of instrument type
DEF
= FREQU.
MODE
Menu Instrument mode
COUNT. Counter FREQU. Frequencymeter TIME Stopwatch/timer RTC Stopwatch/backup timer
MODE
RTCTI M E
Selection of “COUNT.” mode Example
FILTE RFREQV .
COU Nt .
M. START
Selection of stopwatch/ timer control
- menu for time setting is accessible only in the stopwatch/timer mode
CONTIN.
Stopwatch/timer is running constantly if the instrument
is turned on
CONTAC.
Stopwatch/timer is running upon contact making
EDGE
Stopwatch/timer is controlled by the priming
signal edge
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
RU n . ST.C.
Stopwatch/timer is controlled and reset by the
edge of the priming signal
- time is set off by the edge (by the signal
passing across the comparing level) and stopped by the next edge
C.RU n . ST.
Stopwatch/timer is controlled and reset by the
edge of the priming signal
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
CLR.RU N .
Stopwatch/timer is reset and set off by the edge of
the priming signal
CL. RU .RE .
Stopwatch/timer is reset
and set off by the edge of the priming signal, the cycle is repeated with every other edge
RUN
Stopwatch/timer is only set
off by the edge
DEF
= COntAC.
M. START
CONTIN. CONTAC. EDGE RU N .ST.C.
C.RUN .ST. CLR.RU N CL.RU .RE . RU N
>
>
Selection of stopwatch control > EdGE Example
M. STOPCONTAC. EDGE
Only for measuring mode
TIME • RTC
5SETTING
lightlight
16
|
INSTRUCTIONS FOR USE OM 652UC
Only for measuring mode
TIME • RTC
Only for measuring mode
COUNT. • TIME • RTC
5 SETTING
lightlight
M. STOP
Selection of stopwatch resetting
- menu of the resetting option is accessible only in the stopwatch/timer regime
DEF
= CLEAR
CLEA R
Stopwatch/timer is reset through input „Clear“
ST. CLR.
Stopwatch/timer is stopped and reset through input
„Clear“
STOP
Stopwatch/timer is stopped through input „Clear“
Selection of type of stopwatch resetting > St. CLr. Example
CLEAR ST. CLR. FILTE R
M. STOP
CLEA R ST. CLR. STOP
OFF 1000 100 40 5
FI LTER
Selection of digital filter
- digital filter may suppress unwanted interfering impulses (e.g. relay backswings) on the input signal. The set parameter gives maximum possible frequency, which the instrument processes w/o limitation
DEF
= OFF
FI LTER
Maximum input frequency 100 Hz > 100 Example
TYP E A .OFF 1000 100
When accessing upon contact and available maximum input frequency we recommend using filter
!
INSTRUCTIONS FOR USE OM 652UC
|
17
5SETTING
lightlight
10
Setting input level
LEV EL .A
Setting input level for Input A
- setting applies for Input A
- setting level (only for type PNP) of the input voltage, the instrument subsequently automatically selects divider and thus the comparing levels
- range of setting 0…43 V(Input < 30 V)
- range of setting 43…300 V(Input < 300 V)
- table of comparing levels is on page 7
Setting the level = 34 V Example
10 11 12 13 14
34
14
24 TYP E C.
LEV EL .A
NPN.CON . PNP
TYP E A
Selection of type of input
- setting applies for Input A
DEF
= NPN.CON.
TYPE
Menu Type of input
NPN.CON. NPN or contact PNP PNP
Selection of type of inpu t > NPN Example
LEV EL. ANPN.CO N.
TYP E A
Input levels (Level A) have to be set after this option
!
18
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
NPN.C PNP
TYP E C
Selection of type of input
- setting applies for resetting input
DEF
= NPN.CON.
TYPE
Menu Type of input
NPN.CON. NPN or contact PNP PNP
Selection of type of inpu t > NPN Example
LEV EL. CNPN.CO N.
TYP E C
Input levels (Level C) have to be set after this option
!
10
Setting input level
LEV EL .C
Setting input level for resetting input
- setting the level (only for PNP type) of input voltage, the instrument subsequently automatically selects divider and thus the comparing levels
- range of setting 0…43 V(Input < 30 V)
- range of setting 43…300 V(Input < 300 V)
- table of comparing levels is on page 7
Setting levels = 34 V Example
10 11 12 13 14
34
14
24 SCALE
LEV EL .C
INSTRUCTIONS FOR USE OM 652UC
|
19
5SETTING
lightlight
1
Setting division constant
DI VI D .
Setting division constant
- calibration constant is for calculation of the input value to required display value
- range: -0,00001...999999
-
DEF
= 1
Division constant = 22 Example
1 2 02 12 22 OFFSET
DI VI D .
1
Setting multiplying­constant
SCALE
Setting multiplying constant
- calibration constant serves for calculation of the input value to required display value
- by entering minus value direction of the calculation is changed, i.e. we count down
- range: -0,00001...999999
-
DEF
= 1
Cali bration consta nt = 3,12 Example
1 2 02 12 012 112
212
DI VI D .
312 312 .0312 31 . 2 3 .12
SCALE
20
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
0
Setting PRESET
OFFSET
Setting additive constant
- PRESET
- offset of the measuring by a set value, which shall be loaded always upon instrument resetting
- range: -99999...999999
(+ time formats)
-
DEF
= 0
Setting „PRESE T“ = 24 Example
0 1 2 3 4
24
04
14 FORM . A
OFFSET
FORM . A
000000 00000. o 0000. oo 000. ooo 00. oooo 0. ooooo FLO A . P.
FORM . A
Selection of projection format
- instrument enables classical projection of number with fixed position of decimal point as well as projection with floating allowing fo projection of number in its most precise form „FLOA. P.“
- for measuring modes „TIME“ and „RTC“ special time formats are preset
DEF
= 000000
DEF
= HH.MM.SS
H
Projection of DP on display > 00000.o Example
LI M . L. 1
* subsequent menu item depends on instrument equipment
000000 00000. o
99. MM.SS HH. MM HHHH. MM MMMM.SS MM.SS.CC 99.SS . CC H . MM.SS.CHH. MM. SS
F C
H
INSTRUCTIONS FOR USE OM 652UC
|
21
5SETTING
lightlight
22
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
LI M . L. 2
LI M . L.1
25
LI M . L.1
Setting the boundary for limit 1
- range of the setting is -99999…999999 (+ time formats)
- presetting “Hysteresis”=0 “Delay”=0
- contingent change of hysteresis or offset of the switch-on may be performed in the PROFI menu
DEF
= 25
Setting the boundary for limit 1
LI M . L. 2
Setting the boundary for limit 2
- range of the setting is -99999…999999 (+ time formats)
- presetting “Hysteresis”=0 “Delay”=0
- contingent change of hysteresis or offset of the switch-on may be performed in the PROFI menu
DEF
= 75
75
Setting the boundary for limit 2
Setting limit 2 > L 2 = 230 Example
100 100 130 130 120 110
230 MENU
* following item of the menu depends on instrument equipment, provided it has an analog output the following item is „Type“
Setting limit 1 > L 1 = 30 Example
25 26 20 29 28 27
20 30 LI M . L .2
Items for “Limits” and “Analog output” are acces­sible only if the instrument contains them.
!
Only with option > Dual Comparator
INSTRUCTIONS FOR USE OM 652UC
|
23
5SETTING
lightlight
MAX. A .O.
Assigning the display value to the end of the
AO range
- range of the setting is -99999…999999
DEF
= 100
MI N . A . O.
Assigning the display value to the beginning of
the AO range
- range of the setting is -99999…999999
DEF
= 0
TYP. A. O .
Setting the type of analog output
Menu Range Description
0-20nA 0…20 mA Er 4-20 4…20 mA with indication of error statement (<3,6 mA) 4-20nA 4…20 mA 0-5nA 0…5 mA 0-2 u 0…2 V 0-5 u 0…5 V 0-10 u 0…10 V
DEF
= E 4
Display value for the beginning o f the AO range > M In.A.O. = 0 Example
MAX. A .O.
TY .A. O .
Type of analog output - 0...10 V > tY.A.O. = U 10 Example
4-20m A 0-2 V0-5m A 0-10 V 0-5 V MI N . A .O .
0-20mA Er4-20 4-20m A 0-5mA 0-2 V 0-5 V 0-10 V
0
Assigning the display value to the beginning of the AO range
0 MAX. A .O .
100
Assigning the display value to the end of the AO range
MI N . A . O.
Display value for the end of the AO range > MAX. A.O. = 120 Příklad
100 100 120 110 MENU
Only with option > Analog output
24
|
INSTRUCTIONS FOR USE OM 652UC
5 SETTING
lightlight
Menu LIGHT > M ENU = LIGHT Example
FI R M .
Restoration of the instrument manufacture
setting
- in case of incorrect setting or calibration it
is possible to return to manufacture setting. Prior execution of the changes you will be asked to confirm your selection (YES)
- reading the manufacture calibration and original setting of items in the menu
MENU
MENU
Setting the menu type LIGHT/PROFI
LIGHT > LIGHT menu, a simple menu,
which contains only items necessary for instrument setting
> linear structure of the menu
PROFI > PROFI menu, a complete menu for
entire instrument setting
> tree structure of the menu
DEF
= LIGHT
LIGHT FI R M . .
LI G H T PROFI
FI R M .
Restoration of manufacture setting > FIRM . Example
FI R M . . n . PASS.
YES
YES
Do not perform restoration of user setting (USER) prior to its saving in Profi menu
!
INSTRUCTIONS FOR USE OM 652UC
|
25
5SETTING
lightlight
0
Setting new access password
n . PASS.
n . PASS.
Setting new access password
- access password for LIGHT/PROFI menu
- range of the numeral code 0…9999
- when setting password to “0000” the
access into LIGHT/PROFI menu is accessible without call for entering it
- in case of loss of password universal password “8177” may be used
DEF
= 0
IDENT.
IDENT.
SW version of the instrument
- the display shows the type identification of the instrument, SW number, SW version and current input setting (Mode)
- if the SW version reads a letter on the first position, then it is a customer SW
- after the identification is completed the menu automatically quits the display and measuring mode is restored
YES
OM 652 61 - 001
setting the input typeSW number SW version
FREQV .
Instrument type
142. 8
Return to measuring mode
New password - 341 > n.PASS. = 341 Example
0
IDENT.
1 01 11 21 31
41 041 341 241 141
26
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
6.0 Setting “PROFI”
PROFI Complete programming menu
• contains complete instrument menu and is protected by optional number code
• designed for expert users
• preset from manufacture is menu LIGHT
Switching over to “PROFI” menu
+
• temporary switch-over to PROFI menu, which is suitable to edit a few items
• after quitting PROFI menu the instrument automatically switches to LIGHT menu
• access is password protected (if it was not set under item N. PASS. =0)
+ • access into LIGHT menu and transition to item „MENU“ with subsequent selection of „PROFI“
and confirmation
• after re-entering the menu the PROFI type is active
• access is password protected (if it was not set under item N. PASS. =0)
• For expert users
• Complete instrument menu
• Access is password protected
• Possibility to arrange items of the „User“ menu
• Tree menu structure
SETTING PROFI
INSTRUCTIONS FOR USE OM 652UC
|
27
6SETTING
profi profi
28
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
6.1 Setting “PROFI” - INPUT
6.1.1 Resetting internal values
INPUT
SERVIC.
OUTPUT
CHANNE.
CLEA R
EXT . IN.
CONFIG
KEYS
The basic instrument parameters are set in this menu
CLEA R
Resetting internal values
CONFIG
Primary instrument setting
EXT . IN.
Setting the external input function
KEYS
Setting the ENTER key function
INPUT
SERVIC.
OUTPUT
CHANNE.
CLEAR
EXT . IN.
CONFIG
KEYS
CL. TA R.
CL. CN T .
CLEA R
Resetting internal values
CL. CN T .
Counter resetting
CL. TA R.
Tare resetting
INSTRUCTIONS FOR USE OM 652UC
|
29
6.1.2 Instrument configurtion
6.1.2a Setting the initial value
C
H
M. TI M E
MODE
INPUT
SERVIC.
OUTPUT
CHANNE.
EXT . IN.
CONFIG
KEYS
M. START
M. STOP
TYP E A .
FI LTER
LEV EL .C
TYP E C .
LEV EL .A
POLAR.
BACKUP
SET. V.CLEA R
CONFIG
Primary instrument setting
SET. V.
Setting the initial value
MODE
Setting the instrument measuring mode
M. TI M E
Setting the time base
M. START
Setting the stopwatch control
M. STOP
Setting stopwatch resetting
FI LTER
Setting the input filtration constant
TYP E -
Setting the type of input
LEV EL . -
Setting the input level
POLAR.
Selection of active level/ edge
BACKUP
Setting data backup/ time
INPUT
SERVIC.
OUTPUT
CHANNE.
CLEA R
EXT . IN.
CONFIG
SET V.
KEYS
M. TI M E
MODE
M. STOP
M. START
TYP E A
FI LTER
LEV EL .C
TYP E C
LEV EL .A
POLAR.
BACKUP
0
SET V .
Setting initial value
- function allows the user a one-time setting of the display initial value
6SETTING
profi profi
30
|
INSTRUCTIONS FOR USE OM 652UC
6.1.2b Selection of measuring mode
6.1.2c Selection of measuring period/time base
CLEA RINPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
COUNT.
FREQV .
KEYS
RTC
TI M E
DEF
M. START
M. STOP
M. TI M E
MODE
TYP E A
LEV EL .C
TYP E C
LEV EL .A
FI LTER
BACKUP
POLAR.
SET. v .
MODE
Selection of instrument measuring mode
COUNT.
Mode „Counter“
- counts on input A
FREQU .
Mode „Frequencymeter“
- measures frequency on input A
TI M E
Mode „Stopwatch/ timer“
RTC
Mode „Stopwatch/ timer“ with RTC backup
- not in standard equipment
DEF
0.5
5. 0
1. 0
10. 0
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
M. TIME
MODE
M. START
M. STOP
TYP E A
LEV EL .C
TYP E C
LEV EL .A
FI LTER
BACKUP
POLAR.
SET. v .CLEAR
M. TI M E
Selection of measuring period/time base
- if you set measuring period e.g. for 1 s, the measuring runs approximtely from 1 sto 2 s(1 s+ maximum one cycle of measured signal). If no signal arrives within 2 sit is taken that the signal has zero frequency
- range of setting of the time base is 0,5 sto 10 s
- in the „RTC“ regime with data projection the set time defines the cycle of switching between time/date, min. is 5 s, datue is displayed for approx. 2,5 s
6 SETTING
profi profi
INSTRUCTIONS FOR USE OM 652UC
|
31
6SETTING
profi profi
6.1.2d Selection of stopwatch/timer control
H
CLEA R
RUN
CL. RU .RE .
CLR.RU N .
MODE
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
TYP E A
LEV EL .C
TYP E C
LEV EL .A
FI LTER
BACKUP
POLAR.
SET. V.
M. START
M. STOP
DEF
CONTIN.
CONTAC.
EDGE
RUN .ST. C.
C.RUN .ST.
M. TI M E
M. START
Selection of stopwatch/ timer control
- time setting menu is accessible only in the stopwatch/timer regime
CONTIN.
Stopwatch/timer is running constantly if the
instrument is turned on
CONTAC.
Stopwatch/timer is running upon contact
making
EDGE
Stopwatch/timer is controlled by the priming
signal edge
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
RU n . ST.C.
Stopwatch/timer is controlled and reset by
the edge of the priming signal
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
C.RU n . ST.
Stopwatch/timer is controlled and reset by
the edge of the priming signal
- time is set off by the edge (by the signal passing across the comparing level) and stopped by the next edge
CLR.RU N .
Stopwatch/timer is reset and set off by the edge
of the priming signal
CL. RU .RE .
Stopwatch/timer is
reset and set off by the edge of the priming signal, the cycle is repeated with every other edge
RUN
Stopwatch/timer is only
set off by the edge
32
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
142. 8
+
PASS. 0
INPUT CHANNE.
SET. A
SCALE 1
FORM . A
000000 00000.o  0.ooooo FLOA. P.
FI LTER MOD. F. A
CON . F.A 0
NO EXPON.. ROUND.
CLEAR
CL. CNT. CL. TAR ...
CONFIG DI VI D . 1
EXT . IN. EX T. 1
OFF LOCK. K.  HOLD  TARE  CLEAR  CLR. ST.
KEYS ALL
NO YES
M. TI M E
0.5 1.0 5.0 10.0
MODE
COUNT. FREQV.. TIME RTC
FI LTER
OFF 1000 100 40 5
LEV EL. A
TYP E A .
NPN.CON. PNP
LEV EL. C
TYP E C.
NPN.CON. PNP
POLAR.
Lo \ HI /
10
10
BACKUP
YES NO
OFFSET 0
Programming schem
HH.MM.SS 99.MM.SS MM.SS HHHH.MM  H.MM.S
S
SET V .
0
M. STOP
CLEAR ST. CLR... STOP
M. START
CONTIN. CONTAC... EDGE  RUN
BACKUP
NO VALUE TIME
INSTRUCTIONS FOR USE OM 652UC
|
33
6SETTING
profi profi
IDENT. OM 652UC. . . .
SERVIC.
OUTPUT
LI M I T S LI M . 1 .
RESTO R.
SAVE
MENU
LIGHT PROFI
TI M . L.1 0
HYS. L.1 0
LI M . L.1 25
LI M . 2.
TI M . L.2 0
HYS. L. 2 0
LI M . L. 2 75
TYP. L.2
CLOSE OPEn
USER
FI R M .
MOD. L.1
HYSTER. C.-PULS.
TYP. L.1
CLOSE OPEN
DI S P.
100MAX. A .O .
0MI N . A .O .
AN. OUT. TYP. A. O .
0-20mA Er4-20 4-20mA 0-5mA 0-2 V 0-5 V 0-10 V
ADDR. 0
DATA BAUD
300 600 1200 2400 4800 9600 19200 38400 57600 115200 230400
BRIGHT.
25% 50% 75% 100%
0N . PASS.
MOD. L.2
HYSTER. ON RUN
Upon delay exceeding 60 s the programming mode is automatically discontinued and the instrument itself restores the measuring mode
!
34
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
6.1.2e Selection of stopwatch/timer resetting
H
6.1.2f Selection of input filter parameters
C
H
MODE
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
TYP E A
LEV EL .C
TYP E C
LEV EL .A
FI LTER
BACKUP
POLAR.
SET. V.
M. START
M. STOP
M. TI M E
DEF
CLEA R
ST. CLR.
ST. CLR.
CLEA R
M. STOP
Selection of stopwatch resetting
- menu of the resetting option is accessible only in the stopwatch/timer regime
CLEA R
Stopwatch/timer is reset through input „Clear“
ST. CLR.
Stopwatch/timer is stopped and reset
through input „Clear“
STOP
Stopwatch/timer is stopped through input
„Clear“
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
DEF
OFF
1000
5
40
100
MODE
TYP E A
LEV EL .C
TYP E C
LEV EL .A
M. STOP
BACKUP
POLAR.
M. START
FI LTER
M. TI M E
SET. v .CLEAR
FI LTER
Selection of digital input filter
- digital filter may suppress unwanted interfering impulses (e.g. relay backswings) on the input signal. The set parameter gives maximum possible frequency (Hz) of the instrument, which the instrument w/o limitation
When accessing upon contact and available maximum input frequency we recommend using filter
!
INSTRUCTIONS FOR USE OM 652UC
|
35
6.1.2g Selection of the type of input
6.1.2h Setting input level
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
DEF
NPN.CON
PNPMODE
LEV EL .C
TYP E C
LEV EL .A
M. STOP
BACKUP
POLAR.
M. START
M. TI M E
FI LTER
TYP E A
SET. v .CLEAR
TYP E A
Selection of type of input
- setting applies for Input A
NPN.CON
Type of input NPN and upon contact
PNP
Type of input PNP
Input levels (Level A) must be set after this selection
!
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
10
MODE
LEV EL .C
TYP E C
M. STOP
BACKUP
POLAR.
M. START
M. TI M E
FI LTER
TYP E A
LEV EL .A
SET. v .CLEAR
LEV EL .A
Setting input level
- setting applies for Input A
- setting level (only for type PNP) of the input voltage, the instrument subsequently automatically selects divider and thus comparing levels
- range of setting 0…43 V
(Input A< 30 V, bracket No. 13) (Input C < 30 V, bracket No. 14)
- range of setting 43…300 V
(Input A<300 V, bracket No. 18) (Input C <300 V, bracket No. 16)
- table of comparing levels is on page 7
6SETTING
profi profi
Nastavení pro vstup Nulovani (type C) je shodné s nastavením Vstupu A
!
Nastavení pro vstup Nulovani (Level. C) je shodné s nastavením Vstupu A
!
Setting for input Resetting (type C) is identical with setting for Input A
!
Setting for input Resetting (Level. C) is identical with setting for Input A
!
36
|
INSTRUCTIONS FOR USE OM 652UC
6.1.2i Selection of active level or edge
6.1.2j Selection of display status backup
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
DEF
Lo \
Hi /MODE
TYP E C
M. STOP
LEV EL .C
M. START
M. TI M E
FI LTER
TYP E A
BACKUP
POLAR.
LEV EL .A
SET. v .CLEAR
POLAR.
Selection of active level or edge
Lo \
Active upon change of declining edge Lo >Hi
- upon entering the contact > active on switch-on
Hi /
Active upon change of entering edge Hi > Lo
- upon entering the contact > active on switch-off
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
DEF
NO
YESMODE
M. TI M E
TYP E C
LEV EL .C
FI LTER
TYP E A
POLAR.
BACKUP
LEV EL .A
SET. V.CLEAR
BACKUP
Selection of display status backup
- setting display value restoration after power failure or instrument switch-off
NO
After switch-on the instrument loads the
display status from the memory
YES
Instrument resets itself after switch-on
6 SETTING
profi profi
INSTRUCTIONS FOR USE OM 652UC
|
37
6.1.2k Setting the display status backup
H
6.1.3 External input function selection
INPUT
SERVIC.
OUTPUT.
CHANNE.
EXT .IN.
CONFIG
KEYS
TI M E
VALUE
NO
DEF
MODE
TYP E C
M. STOP
LEV EL .C
M. START
M. TI M E
FI LTER
TYP E A
POLAR.
BACKUP
LEV EL .A
SET. V.CLEAR
BACKUP
Selection of display status backup
- time setting menu is accessible only in the stopwatch/timer regime
- setting display value restoration after power failure or instrument switch-off
NO
Instrument resets itself after every switch-on
VALUE
After switch-on the instrument loads the
display status from the memory
TI M E
Instrument downloads „running“ time fromRTC
- item accessible only with extension „Time backup“
INPUT
SERVIC.
OUTPUT
CHANNE.
CLEA R
CONFIG
KEYS
OFF
LOCK. K.
DEF
HOLD
EXT. IN.
EXT. 1
TARE
CLR. ST.
CLEA R
EXT . IN.
External input function selection
OFF
Input is off
LOCK. K.
Auxiliary input governs the „LOCK“ function
- the input governs the blocking of control keys on front panel
HOLD
Auxiliary input governs the „HOLD“ function
- the input governs the HOLD function, which blocks all instrument functions
TARE
Auxiliary input governs the „TARE“ function
- the TARE function is activated through the input, only in the “Frequency” mode
CLEA R
Auxiliary input governs the „Clear“ function
- stopwatch/counter is cleared (preset) through the input
CLR. St.
Auxiliary input governs the „Clear“ function
- stopwatch/counter is cleared (preset) through the input, Stopwatcg stops altogether
6SETTING
profi profi
38
|
INSTRUCTIONS FOR USE OM 652UC
6.1.4 Optional accessory functions of the keys
INPUT
SERVIC.
OUTPUT
CHANNE.
CLEA R
EXT. IN.
CONFIG
YES
NO
DEF
KEYS
ALL
ALL
Assigning accessory functions of control keys
- owing to limited space in the instrument’s memory it is not feasible to set the keys’ functions one by one
YES
Accessory functions are on
Counter Frequency Stopwatch
- Resetting tare Resetting
- Resetting tare Display hold
- Tare projection Start
Resetting Tare Stop
NO
Accessory functions are off
6 SETTING
profi profi
Control keys functions are described on page 11
!
INSTRUCTIONS FOR USE OM 652UC
|
39
6SETTING
profi profi
40
|
INSTRUCTIONS FOR USE OM 652UC
6.2 Setting “PROFI” - CHANNEL
6.2.1a Setting multiplying constant
6.2.1b Setting division constant
INPUT
SERVIC.
OUTPUT
CHANNE.
SET. A
FORM . A
FILTER
CHAN . A
In this menu the instrument input param­eters are set
SET. A
Setting calibration constant
FI LTER
Setting the digital filters
FORM . A
Selection of projection format
INPUT
SERVIC.
OUTPUT.
CHANNE.
SET. A
FORM . A
FILTER
CHAN . A SCALE
OFFSET
DIVID.
1
SCALE
Setting multiplying constant
- multiplying constant serves for calculation of input value to required display value
- by entering minus value the direction of calculation is changed, i.e. we count down
- range: -0,00001...999999
-
DEF
= 1
INPUT
SERVIC.
OUTPUT.
CHANNE.
SET. A
FORM . A
FILTER
CHAN . A
OFFSET
DI VI D .
SCALE 1
DI VI D
Setting division constant
- division constant serves for calculation of input value to required display value
- range: 0,00001...999999
-
DEF
= 1
6 SETTING
profi profi
If non-zero value is set in the “TIME” or “RTC” mode in the “OFFSET” item, it applies that the multiplying constant “SCALE” is negative
!
H
INSTRUCTIONS FOR USE OM 652UC
|
41
6.2.1c Setting additive constant - PRESET
6.2.2 Setting the digital filters
INPUT
SERVIC.
OUTPUT.
CHANNE.
SET. A
FORM . A
FILTER
CHAN . A
DIVID.
OFFSET
SCALE 0
OFFSET
Setting PRESET constant
- offset of the measuring by a set value, which shall be loaded always upon instrument resetting
- range: -99999...999999
-
DEF
= 0
MOD. F.A.INPUT
SERVIC.
OUTPUT.
CHANNE.
SET. A
FORM . A
FILTER
CHAN . A NO
EXPOn.
ROUND.
CON . F.A.
MODE
Setting the digital filters
CONS t .
Setting the constant
- this menu item is always displayed after selection of a particular type of filter
-
DEF
= 2
NO
Filters are switched off
EXP ON .
Selection of exponential filter
- calculation of value from the number of measurements selected in„CONST“
ROUND.
Selection of value round-up
- it is set by …arbitrary number, which determines the projection step (e.g.: “Con. F.A.”=2,5 > display 0,
2.5, 5,...)
6SETTING
profi profi
If non-zero value is set in the “TIME” or “RTC” mode in the “OFFSET” item, it applies that the multiplying constant “SCALE” is negative
!
H
42
|
INSTRUCTIONS FOR USE OM 652UC
6.2.3 Selection of projection format
FORM . A
H
DEF
000000
00000. 0
0000. 00
000. 000
00. 0000
FLOA . P.
HH. MM.SS
H. MM.SS . C
99. SS .CC
MMMM.SS
HHHH. MM
HH. MM
DEF
INPUT
SERVIC.
OUTPUT.
CHANNE.
SET.In
FILTER
CHAN . A
0. 00000
99. MM.SS
MM.SS.CC
FORM . A
Selection of projection format
- the instrument enables projection of number withdecade positioning of decimal point
- for projection of time there are also other projection forms available
6 SETTING
profi profi
In mode “TIME” or “RTC” the time base is preset according to projection format
in seconds > 000000…0.00000, Floa.P., HH.MM.SS, 99.MM.SS, MMMM.SS
in minutes > HH.MM, HHHH.MM
in tenths of seconds > H.MM.SS.C
in hundredths of seconds > MM.SS.CC, 99.SS.CC
!
H
INSTRUCTIONS FOR USE OM 652UC
|
43
6SETTING
profi profi
44
|
INSTRUCTIONS FOR USE OM 652UC
6.3 Setting „PROFI“ - OUTPUTS
6.3.1a Selection of mode of output L 1
6.3.1b Selection of mode of output L 2
CHANNE.
SERVIC.
OUTPUT.
LI M I T S
AN. OUT.
DATA
DI S P .
INPUT
It is possible to set the parameters of the instrument output signals in this menu
LI M I T S
Setting the type and the switching of limits
DATA
Setting the type and the parameters ot data
output
AN. OUT.
Setting the type and parameters of analog
output
DI S P .
Setting the display brightness
INPUT
SERVIC.
OUTPUT
CHANNE.
LIMITS
AN. OUT.
DATA
DI S P .
LIM. 1
LI M . 2
HYSTER.
C. -P U LS .
DEF
HYS. L.1
LI M . L.1
TI M . L.1
MOD. L.1
TYP. L.1
MOD. L.1
Mode of limit 1
HYSTER.
Standard mode - limit, hysteresis and delay
C. -P U LS .
Automatic clearing of
counter to preset value and generate an impulse of the length set in “TIM.L.1”
Setting is available only for LIM 1
!
INPUT
SERVIC.
OUTPUT
CHANNE.
LIMITS
AN. OUT.
DATA
DI S P .
LIM. 2
LI M . 1 HYSTER.
ON RUN
DEF
HYS. L.2
LI M . L. 2
TI M . L. 2
MOD. L.2
TYP. L.2
MOD. L.2
Mode of limit 2
HYSTER.
Standard mode - limit,
hysteresis and delay
ON RUN
Relay is switched on/
off if the stopwatch is running
Setting is available only for LIM 2
!
6 SETTING
profi profi
INSTRUCTIONS FOR USE OM 652UC
|
45
6.3.1c Selectionof type of output
6.3.1d Setting values for limits evaluation
INPUT
SERVIC.
OUTPUT
CHANNE.
LIMITS
AN. OUT.
DATA
DI S P .
LIM. 1
LI M . 2
CLOSE
OPEN
DEF
TYP. L.1
HYS. L.1
LI M . L.1
TI M . L.1
MOD. L.1
TYP. L.1
Setting the type of relay function
CLOSE
Relay switches on when the condition is met
OPEN
Relay switches off when the condition is met
Setting is identical for LIM 1 and LIM 2
!
INPUT
SERVIC.
OUTPUT
CHANNE.
LIMITS
AN. OUT.
DATA
DI S P .
LIM. 1
LI M . 2
HYS. L.1
LI M . L.1
TI M . L.1
MOD. L.1
TYP. L.1
25
LI M . L.1
Setting the boundary for relay switch-on
- within the full display range
-
DEF
= 25 (L 1), 75 (L 2)
HYS. L.1
Setting hysteresis
- defines the band around the limit (on both sides, LIM. ±1/2 HYS.)
- within the full display range
-
DEF
= 0
TI M . L.1
Setting the offset of the relay switch-on
- within the range 0…99,9 s
-
DEF
= 0
Setting is identical for LIM 1 and LIM 2
!
6SETTING
profi profi
46
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
6.3.2a Selection of transmission rate of data output
6.3.2b Setting the instrument address
CHANNE.
SERVIC.
OUTPUT.
LI M I T S
AN. OUT.
DATA
DI S P .
INPUT
ADDR.
BAUD 600
1200
2400
4800
9600
19200
38400
5760 0
11 5 2 0 0
DEF
2304 0 0
600
BAUD
Setting the data output rate
300
Rate - 300 Baud
600
Rate - 600 Baud
1200
Rate - 1200 Baud
2400
Rate - 2400 Baud
4800
Rate - 4800 Baud
9600
Rate - 9600 Baud
19200
Rate - 19200 Baud
38400
Rate - 38400 Baud
5760 0
Rate - 57600 Baud
11 5 2 0 0
Rate - 115200 Baud
2304 0 0
Rate - 230 400 Baud
CHANNE.
SERVIC.
OUTPUT.
LI M I T S
AN. OUT.
DATA
DI S P .
INPUT
ADDR.
BAUD 00
ADDR.
Setting the instrument address
- setting within the range 0…31
-
DEF
= 00
INSTRUCTIONS FOR USE OM 652UC
|
47
6SETTING
profi profi
6.3.3a Selection of type of analog output
6.3.3b Selection of analog output range
DI S P .
CHANNE.
SERVIC.
OUTPUT.
LI M I T S
AN. OUT.
DATA
INPUT TYP. A.O.
MI N . A . O.
MAX. A .O.
0-20mA
4-20mA
0-5 mA
Er4-20
DEF
0-5 V
0-2 V
0-10 V
TYP. A. O .
Setting the type of analog output
0-20mA
Type - 0…20 mA
Er4-20
Type - 4…20 mA
- with indication of error statement (<3,6 mA)
4-20mA
Type - 4…20 mA
0-5 mA
Type - 0…5 mA
0-2 V
Type - 0…2 V
0-5 V
Type - 0…5 V
0-10 V
Type - 0…10 V
DI S P .
CHANNE.
SERVIC.
OUTPUT.
LI M I T S
AN. OUT.
DATA
INPUT TYP. A.O.
MI N . A . O.
MAX. A .O.
0
AN. OUT.
Setting the analog output range
- analog output is isolated and its value corresponds with the displayed data. It is fully programmable, i.e. it allows to assign the AO limit points to any two arbitrary points of the entire measuring range
MI N . A . O.
Assigning the displayed value to the beginning of
the analog output range
- range of the setting is -99999…999999
-
DEF
= 0
MAX. A .O.
Assigning the displayed value to the end of the
analog output range
- range of the setting is -99999…999999
-
DEF
= 100
48
|
INSTRUCTIONS FOR USE OM 652UC
6 SETTING
profi profi
6.3.4 Selection of display brightness
CHANNE.
SERVIC.
OUTPUT
INPUT. LI M I T S
AN. OUT.
DATA
DI S P .
BRIGHT. 100%
50%
25%
75%
DEF
BRIGHT.
Setting the display brightness
- by selecting the display brightness we may react properly to light conditions in place of location of the instrument
25%
Display brightness - 25 %
50%
Display brightness - 50 %
75%
Display brightness - 75 %
100%
Display brightness - 100 %
INSTRUCTIONS FOR USE OM 652UC
|
49
6SETTING
profi profi
50
|
INSTRUCTIONS FOR USE OM 652UC
6.4 Setting “PROFI” - SERVICE
6.4.1 Selection of the type of programming menu
CHANNE.
SERVIC.
OUTPUT
MENUINPUT
N. PASS.
RESTO R.
IDENT.
The instrument‘s service functions are set in this menu
MENU
Selection of menu type LIGHT/PROFI
RESTO R.
Restoration of the manufacture setting and
instrument calibration
N. PASS.
Setting new access password
IDENT.
Instrument identification
CHANNE.
SERVIC.
OUTPUT
MENU
RESTO R.
INPUT
N. PASS.
LI G H T
PROFI
DEF
IDENT.
MENU
Selection of menu type LiGHT/PROFI
- allows to set the menu complexity as per user needs and abilities
LI G H T
Active LIGHT menu
- simple programming menu, contains only items necessary for instrument configuraction and setting
- linear menu structure > items in succession
PROFI
Active PROFI menu
- complete programming menu for expert users
- tree menu
6 SETTING
profi profi
Change of setting is valid with next access into menu
!
INSTRUCTIONS FOR USE OM 652UC
|
51
6.4.2 Restoration of the manufacture setting
6.4.3 Setting new access password
6.4.4 Instrument identification
CHANNE.
SERVIC.
OUTPUT
MENU
RESTOR.
INPUT
N. PASS.
FI R M .
IDENT.
SAV E
USER
RESTO R.
Restoration of the instrument manufacture
setting
FI R M .
Return to manufacture setting of the instrument
- downloading manufacture setting for currently selected type of instrument (items described DEF)
USER
Return to user setting of the instrument
- downloading user setting of the instrument, i.e. setting which was stored under item SERVIC./RESTOR./SAVE
SAV E
Storing user setting of the instrument
- storing the setting enables the operator its future contingent restoration
CHANNE.
SERVIC.
OUTPUT
MENU
RESTO R.
N. PASS.
INPUT
IDENT.
0
N. PASS.
Setting new password for access into the LIGHT
and PROFI menu
- this option allows to change the numeral code, which protects the access into the LIGHT and PROFI Menu.
- numeral code range is 0…9999
- universal password in case of loss „8177“
CHANNE.
SERVIC.
OUTPUT
MENU
RESTO R.
N. PASS.
INPUT
IDENT.
IDENT.
Projection of instrument SW version
- the display shows the type identification of the instrument, SW number, SW version and current input setting (Mode)
- if the SW version reads a letter on the first position, then it is a customer SW
6SETTING
profi profi
After restoration of setting the instrument switches off for several seconds
!
52
|
INSTRUCTIONS FOR USE OM 652UC
7.0 “USER” menu configuration
USER menu is designed for users who need to change only several items of the setting without the option to change the basic instrument setting (e.g. repeated change of limit setting)
• there are no default items from manufacture in USER menu
• menu configuration possible on items indicated by inverse triangle
MOD. L.1
• setting may be performed in LIGHT or PROFI menu, with the USER menu then overtaking the given menu structure
SETTING
---
+
NO YES SHO
---
return to item
flashing sign - current setting is displayed
NO
item will not be displayed in USER menu
YES
item will be displayed in USER menu with the chance of editing
SHO
item will be solely displayed in USER menu
7 SETTING
useruser
• For user operation
• Menu items are set by the user (Profi/Light) as per request
• Access is not password protected
SETTING USER
INSTRUCTIONS FOR USE OM 652UC
|
53
Setting sequence of items in “USER” menu
In compiling USER menu from active LIGHT menu the items (max. 10) may be assigned a sequence, in which they will be projected in the menu
Example:
Into USER menu were selected these items: (keys
+
) > CL. Cnt., LIM. L 1, LIM. L 2, for which we have preset this sequence:
(tlačítky
+
):
CL. Cnt. 5 LIM. L 1 0 (sequence not determined) LIM. L 2 1
Upon entering USER menu (key
) items will be projected in the following sequence: LIM. L 2 > CL. Cnt. > LIM. L 1
7SETTING
useruser
---
+
0
setting projection sequence
54
|
INSTRUCTIONS FOR USE OM 652UC
The instruments communicate via serial line RS232 or RS485. For communication they use the ASCII protocol. Communica­tion runs in the following format:
ASCII: 8 bit, no parity, one stop bit
The transfer rate is adjustable in the instrument menu and depends on the control processor used. The instrument address is set in the instrument menu in the range of 0 ÷ 31. The manufacture setting always presets the ASCII protocol, rate of 9600 Baud, address 00. The type of line used - RS232 / RS485 - is determined by an exchangeable card automatically identified by the instrument.
COMMANDS FOR INSTRUMENT OPERATION
The commands are described in specification you can find at www.orbit.merret.cz/rs. A command consists of a number and a letter. The size of the letters have a significance.
DETAILED DESCRIPTION OF COMMUNICATION VIA SERIAL LINE
Activity Data transferred
Data solicitation (PC) # A A <CR>
Data transfer (Instrument) > R <SP> D D D D D (D) (D) <CR>
Command corfi rmation (Instrument) - OK ! A A <CR>
Command corfi rmation (Instrument) - Bad ? A A <CR>
Instrument identifi cation # A A 1Y <CR>
HW identifi cation # A A 1Z <CR>
One-time mesasurement # A A 7X <CR>
Repeated mesasurement # A A 8X <CR>
Setting to transmit display + relay value
#AA1X<CR>
Setting to transmit measured value
#AA1x<CR>
Setting limit1
# A A 1L D (D) (D) (D) (D) (D) <CR>
Setting limit 2
# A A 2L D (D) (D) (D) (D) (D) <CR>
LEGENDA
#3523HBeginning of the command
AA 031
Two signs of the inst. address (sending in ASCII - decades and
units, ex. “01“, “99“ universal
<CR> 13 0D
H
Carriage return
<SP> 32 20
H
Space
D
Data - usually signs “0“…“9“, “-“, “.“; (D) - DP and (-) may prolong data
R50
H
…57
H
Relay and Tare status
!3321
H
Positive command corfi rmation (ok)
?633F
H
Negative command corfi rmation (bad)
>623E
H
Beginning of the transmitted data
8 DATA PROTOCOL
RELAY, TARE
Signs Relay 1 Relay 2 Tare
P000
Q10 0
R01 0 S11 0
T00 1
U10 1
V01 1
W1 1 1
INSTRUCTIONS FOR USE OM 652UC
|
55
8DATA PROTOCOL
56
|
INSTRUCTIONS FOR USE OM 652UC
ERROR CAUSE ELIMINATION
E. D. Un.
Number is too small (large negative) to be displayed
change DP setting, channel constant
E. D. Ou.
Number is too large to be displayed change DP setting, channel constant
E. T. Un.
Number is outside the table range increase the table values, change input setting
(channel constant)
E. T. Ow.
Number is outside the table range increase the table values, change input setting
(channel constant)
E.I. Un.
Input quantity is smaller than permitted input quantity range
change input signal value or input (range) setting
E.I. OW.
Input quantity is larger than permitted input quantity range
change input signal value or input (range) setting
E. HW.
A part of the instrument does not work properly
send the instrument for repair
E. EE
Data in EEPROM corrupted perform restoration of manufacture setting, upon
repeated error statement send instrument for repair
E.. DATA
Data in EEPROM outside the range perform restoration of manufacture setting, upon
repeated error statement send instrument for repair
E. CLR
Memory was empty (presetting carried out)
upon repeated error statement send instrument for repair, possible failure in calibration
9 ERROR STATEMENTS
INSTRUCTIONS FOR USE OM 652UC
|
57
58
|
INSTRUCTIONS FOR USE OM 652UC
INPUT Type: upon contact, TTL, NPN/PNP
Measuring: 1x counter/frequency 1x stopwatch/timer
- measuring range isadjustable
Input frequency: 0,1...50 kHz
PROJECTION
Display: 999999, intensive red or green 7-segment LED, digit height 14 mm Projection: -99999…999999 Decimal point: adjustable - in programming mode Brightness: adjustable - in programming mode
INSTRUMENT ACCURACY
Temperature coef.: 50 ppm/°C Accuracy: ±0,05 % of the range + 1 digit (frequency) Time base: 0,5/1/5/10 s
Multiplying constant: ±0,00001....999999
Division constant: ±0,00001....999999
Filtration constant: allows for setting max. valid frequency, which is processed
(OFF/5...1000 Hz) Type of fi lter: digital Preset: -99999...999999 Data backup: preservation of measured data even after instrument switch-
off (EEPROM) Functions: Tare - display resetting Hold - stop measuring (upon contact) Lock - control keys locking RTC: the course of time is backed up by battery upon discon-
nection from the instrument supply (may be turned off -
jumper inside the instrument) minimum lifetime 1 year Battery: Lithium cell CR 2032RV, 3V/220 mAh OM Link: Company communication interface for instrument operaion,
setting and update Watch-dog: reset after 540 ms Calibration: při 25°C a40 % r.v.
COMPARATOR
Type: digital, adjustable in the menu, contact switch-on < 50 ms Limits: -99999…999999 Hysteresis: 0…999999 Delay: 0…99,9 s Outputs: 2x relays with switch-on contact (Form A) (230 VAC/30 VDC, 3 A)* Relay: 1/8 HP 277 VAC, 1/10 HP 125 V, Pilot Duty D300
DATA OUTPUTS
Protocols: ASCII Data format: 8 bit + no parity + 1 stop bit Rate: 600…230 400 Baud RS 232: isolated, two-way communication RS 485: isolated, two-way communication, addressing (max. 31 instruments)
ANALOG OUTPUTS
Type: isolated, programmable with resolution of max. 4 000
points, analog output corresponds with the displayed data,
type and range are adjustable Non-linearity: 0,2 % of the range TC: 100 ppm/°C Rate: response to change of value < 250 ms Voltage: 0…2 V/5 V/10 V Current: 0…5/20 mA/4…20 mA
- compensation of conduct up to 450 Ohm
EXCITATION
Adjustbale: 5…24 VDC/max. 1,2 W, isolated
- cannot be combined with data/analog output
POWER SUPPLY
Options: 10…30 V AC/DC, 10 VA, isolated,
- fuse inside (T 4000 mA) 80…250 V AC/DC, 10 VA, isolated
- fuse inside (T 630 mA)
MECHANIC PROPERTIES
Material: Noryl GFN2 SE1, incombustible UL 94 V-I Dimensions: 96 x 48 x 120 mm Panel cut-out: 90,5 x 45 mm
OPERATING CONDITIONS
Connection: connector terminal board, conductor cross-section <1,5 mm
2
/<2,5 mm
2
Stabilisation period: within 15 minutes after switch-on Working temp.: 0°…60°C Storage temp.: -10°…85°C Cover: IP65 (front panel only) Construction: safety class I Overvoltage category: EN 61010-1, A2 Insulation resistance: for pollution degree II, measurement category III Instrument power supply > 670 V (PI), 300 V (DI) Input/output > 300 V (PI), 150 (DI) EMC: EN 61000-3-2+A12; EN 61000-4-2, 3, 4, 5, 8, 11; EN 550222, A1, A2
* values apply for resistance load
10 TECHNICAL DATA
INSTRUCTIONS FOR USE OM 652UC
|
59
10TECHNICAL DATA
60
|
INSTRUCTIONS FOR USE OM 652UC
11 INSTR. DIMENSIONS AND INSTALLATION
Panel cut
Panel thickness: 0,5...20 mm
Side view
Front view
90,5 mm
45 mm
96 mm
48 mm
13,5 mm
119 mm
12
Instrument installation
1. insert the instrument into the panel cut-out
2. fit both travellers on the box
3. press the travellers close to the panel
Instrument disassembly
1. slide a screw driver under the traveller wing
2. turn the screw driver and remove the traveller
3. take the instrument out of the panel
INSTRUCTIONS FOR USE OM 652UC
|
61
12CERTIFICATE OF GUARANTEE
Product OM 652UC
Type . . . . . . . . . . . . . . . . . . . .
Manufacturing No. . . . . . . . . . . . . . . . . . . . .
Date of sale . . . . . . . . . . . . . . . . . . . .
A guarantee period of 60 months from the date of sale to the user applies to this instrument. Defects occuring during this period due to manufacture error or due to material faults shall be eliminated free of charge.
For quality, function and construction of the instrument the guarantee shall apply provided that the instrument was connected and used in compliance with the instructions for use.
The guarantee shall not apply to defects caused by:
- mechanic damage
- transportation
- intervention of unqualified person incl. the user
- unavoidable event
- other unprofessional interventions
The manufacturer performs guarantee and post.guarantee repairs unless provided for otherwise.
YEAR
S
hall not apply
the user
Stamp, signature
62
|
INSTRUCTIONS FOR USE OM 652UC
NOTE
INSTRUCTIONS FOR USE OM 652UC
|
63
NOTE
64
|
INSTRUCTIONS FOR USE OM 652UC
TECHDOK - OM 652UC - 2009 - 1v4 - en - V
DECLARATION OF CONFORMITY
Company: ORBIT MERRET, spol. s r.o. Klánova 81/141, 142 00 Prague 4, Czech Republic, IDNo: 00551309
Manufactured: ORBIT MERRET, spol. s r.o. Vodňanská 675/30, 198 00 Prague 9, Czech Republic
declares at its full responsibility that the product presented hereunder meets all technical requirements, is safe for use when utilised under the terms and conditions determined by ORBIT MERRET, spol.s r.o. and that our company has taken all measures to ensure conformity of all products of the type listed hereunder, which are being brought out to the market, with technical documentation and requirements of the appurtenant statutory orders.
Product: Programmable panel instrument
Type: OM 352/652
Version: UC
Conformity is assessed pursuant to the following standards: Electrical safety: EN 61010-1
EMC: EN 50131-1, chapter 14 and chapter 15 EN 50130-4, chapter 7 EN 61000-4-11 EN 50130-4, chapter 8 EN 61000-4-11 EN 50130-4, chapter 9 EN 61000-4-2 EN 50130-4, chapter 10 EN 61000-4-3 EN 50130-4, chapter 11 EN 61000-4-6 EN 50130-4, chapter 12 EN 61000-4-4 EN 50130-4, chapter 13 EN 61000-4-5 EN 50130-5, chapter 20 prEN 50131-2-1, par. 9.3.1 EN 61000-4-8 EN 61000-4-9 EN 61000-3-2 ed. 2:2001 EN 61000-3-3: 1997, Cor. 1:1998, Z1:2002 EN 55022, chapter 5 and chapter 6 and government ordinance:
Electrical safety: No. 168/1997 Sb. EMC: No. 169/1997 Sb.
The evidence are the protocols of authorized and accredited organization: VTÚE Praha, experimental laboratory No. 1158, accredited by ČIA
VTÚPV Vyškov, experimental laboratory No. 1103, accredited by ČIA
Place and date of issue: Prague, 18. March 2003 Miroslav Hackl Company representative
Mode of asses. of conformity: §12, par. 4 b, d of Act No. 22/1997 Sb.
Loading...