Reagents are formulated exclusively for chemical analysis and must not be used for
any other purpose. Keep reagents out of reach of children. Some reagents contain
compounds which are not entirely harmless environmentally. Be aware of the
ingredients and take proper care when disposing of the test solution.
CAUTION
Please read this instruction manual before unpacking, setting up or using the
colorimeter. Please read the method description completely before performing the test.
Be aware of the risks of using the required reagents by reading the MSDS (Material Safety
Data Sheets). Failure could result in serious injury to the operator or damage to the
MSDS: www.orbeco.com
The accuracy of the instrument is only valid if the instrument is used in an
environment with controlled electromagnetic disturbances according to DIN 61326.
Wireless devices, e.g. wireless phones, should not be used near the instrument.
instrument.
CAUTION
Revision_2 07/2009
MC500 Colorimeter 07-2009
1
Table of contents
Part 1 METHODS
1.1 Table of Methods ............................................................................................................................................... 8
Acid demand to pH 4.3.........................................................................................................................................12
Aluminum with tablets.........................................................................................................................................18
Ammonia with tablets ..........................................................................................................................................22
Ammonia, low range (LR) ....................................................................................................................................26
Ammonia, high range (HR) .................................................................................................................................28
total Chlorine ......................................................................................................................................................45
total Chlorine ......................................................................................................................................................49
total Chlorine ......................................................................................................................................................53
in presence of Chlorine ................................................................................................................................... 56
in absence of Chlorine .....................................................................................................................................59
COD, low range (LR) ...............................................................................................................................................62
COD, middle range (MR) ...................................................................................................................................... 64
COD, high range (HR) ............................................................................................................................................66
Color ............................................................................................................................................................................68
Copper with tablet .................................................................................................................................................70
total Copper ........................................................................................................................................................ 73
DEHA T ........................................................................................................................................................................80
Hardness, total .........................................................................................................................................................90
Hardness, total HR ..................................................................................................................................................92
Hydrazine with liquid reagent ...........................................................................................................................96
Hydrazine with Vacu-vials ................................................................................................................................... 98
Iron ........................................................................................................................................................................... 104
Iron with tablet................................................................................................................................................ 106
Iron (powder packs)....................................................................................................................................... 108
Iron (TPTZ) (powder packs)......................................................................................................................... 110
Mangenese with tablet .....................................................................................................................................112
Manganese LR (powder packs) ...................................................................................................................... 114
Molybdate with tablet ....................................................................................................................................... 118
Nitrite with tablet ................................................................................................................................................ 124
Nitrite LR (powder pack) ................................................................................................................................... 126
Nitrogen, total LR ................................................................................................................................................ 128
Nitrogen, total HR................................................................................................................................................ 130
Oxygen, active ...................................................................................................................................................... 132
in presence of Chlorine ................................................................................................................................ 138
in absence of Chlorine .................................................................................................................................. 140
pH-Value LR with tablet .................................................................................................................................... 166
pH-Value with tablet .......................................................................................................................................... 168
pH-Value with liquid reagent .......................................................................................................................... 170
pH-Value HR with tablet.................................................................................................................................... 172
Silica LR PP .............................................................................................................................................................178
Sulfate T ..................................................................................................................................................................184
2.3.4 Performing Zero ............................................................................................................................211
2.3.5 Performing Test .............................................................................................................................. 212
2.3.6 Ensuring reaction periods (countdown) ............................................................................... 212
2.3.7 Changing chemical species ....................................................................................................... 213
2.4.6 Lab function .................................................................................................................................... 239
2.4.7 User operations .............................................................................................................................. 240
2.4.8 Special functions ........................................................................................................................... 250
2.4.10 Instrument special functions / Service .................................................................................. 252
2.5 Data transfer ...................................................................................................................................253
2.5.1 Print of data ..................................................................................................................................... 253
2.5.2 Data transfer to a personal computer (PC) .......................................................................... 253
* = free, combined, total; PP = powder pack; T = tablet;
L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range;
Vacu-vial® is a registered trade mark of CHEMetrics Inc.
272 Nitrite LR PPPP0.01-0.3mg/l NDiazotization530126
280 Nitrogen,
total
LR TT
281 Nitrogen,
total
HR TT
Oxygen, active T
290
292 Oxygen, dissolved Vacu-vial10-800 μg/l O
Ozone (DPD) T
300
tube test0.5-25mg/l NPersulfate
digestion method
tube test5-150mg/l NPersulfate
digestion method
tablet0.1-10mg/l O
tablet0.02-2mg/l O
DPD530132
2
Rhodazine D
2
DPD/Glycine
3
TM
5
430128
430130
530134
530136
70 PHMB Ttablet2-60mg/l PHMBBuer/Indicator560142
Phosphate, T ortho LRtablet0.05-4mg/l PO
320
4
Ammoniummolybdate
2,3
660144, 146
* = free, combined, total; PP = powder pack; T = tablet;
L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range;
Vacu-vial® is a registered trade mark of CHEMetrics Inc.
MC500 Colorimeter 07-2009
9
1.1 Methods
1.1 Table of Methods
No. AnalysisReagentRangeDisplayed asMethod
321 Phosphate, ortho
HR T
323 Phosphate, PP
ortho
324 Phosphate,
tablet1-80mg/l PO
PP0.06-2.5mg/l PO
ortho TTtube test0.06-5mg/l PO
4
4
4
Vanando
-
molybdate
Ascorbic acid
Ascorbic acid
2
2
2
Page
[nm]
430144, 148
660144, 150
660144, 152
327 Phosphate 1 C,
Vacu-vial5-40mg/l PO
ortho
328 Phosphate 2 C,
Vacu-vial0.05-5mg/l PO
ortho
325 Phosphate, hydr. TTtube test0.02-1.6mg/l P
4
4
Vanadomolybdate
Stannous
2
chloride
Acid digestion,
Ascorbic acid
326 Phosphate, total TTtube test0.02-1.1mg/l P
Acid persulf digestion,
Ascorbic acid
316 Phosphonate PPPP0-125mg/l
Persulfate UVOxidation
pH-Value LR T
329
330 pH-Value Ttablet6.5-8.4—Phenolred
331 pH-Value Lliquid6.5-8.4—Phenolred
pH-Value HR T
332
340 Potassium Ttablet0.7-12mg/l K
350 Silica Ttablet0.05-4mg/l SiO
351 Silica LR PPPP0.1-1.6mg/l SiO
352 Silica HR PPPP1-90mg/l SiO
212 Sodium
tablet5.2-6.8—
Bromocresolpurple5560166
5
5
tablet8.0-9.6—Thymolblue
Tetraphenylborate-
4
Turbidity
Silicomolybdate
2
Heteropolyblue2660178
2
Silicomolybdate2430180
2
tablet0.2-16% NaOClPotassium iodide
hypochlorite T
355 Sulfate Ttablet5-100mg/l SO
360 Sulfate PPPP5-100mg/l SO
4
4
BariumsulfateTurbidity
Bariumsulfate-
2
Turbidity
365 Suldetablet0.04-0.5mg/l SDPD/Catalyst
370 Sulte Ttablet0.1-5mg/l SO
384 Suspended Solids Direct
0-750mg/l TSSphotometric660192
DTNB430190
3
Reading
386 TurbidityDirect
Reading
390 Urea Ttablet +
liquid
400 Zinc Ttablet0.02 -1mg/l ZnZincon
0-1000FAUAttenuated
Radiation Method
0.1-2.5mg/l UreaIndophenol/
Urease
3
2
430144, 154
660144, 156
660144, 158
2
660144, 160
2
660162
560168
560170
5
560172
430174
2,3
660176
5
530182
610184
530186
3,4
660188
530194
610196
610198
* = free, combined, total; PP = powder pack; T = tablet;
L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range;
Vacu-vial® is a registered trade mark of CHEMetrics Inc.
10
MC500 Colorimeter 07-2009
1.1 Methods
Literature
The reagent formulations are based on internationally recognized test methods. Some are
described in national and/or international guidelines.
1) Unied German Procedures for the Analysis of Water and Wastewater
2) Standard Methods for the Examination of Water and Wastewater; 18th Edition, 1992
5) Colorimetric Chemical Analytical Methods, 9th Edition, London
Notes for searching:
Active Oxygen -> Oxygen, active
Alkalinity-m -> Alkalinity, total
Alkalinity, total -> Alkalinity, total
Biguanide -> PHMB
Calcium Hardness -> Hardness, Calcium
Monochloramine -> Chloramine, Mono
Total Hardness -> Hardness, total
m-Value -> Alkalinity, total
p-Value -> Alkalinity-p
Silicon dioxide -> Silica
total Alkalinity -> Alkalinity, total
total Hardness -> Hardness, total
Langelier Saturation -> Mode function 70
Index (Water Balance)
Important Notes for use of this manual - PLEASE READ:
1) The Method number for each test is indicated in the large circles at the beginning of each
procedure. Here is an example:
2) The Symbol "Ø" signies the OUTSIDE Diameter of the Sample Vial
3) Please note the appropriate reagent platform (tablet, powder, vial, etc.) for each test.
4) Be sure to observe the proper reagent reaction time indicated for each procedure for accurate
results.
MC500 Colorimeter 07-2009
2
0
11
1.1 Methods
Acid demand to pH 4.3
2
0
with Tablet
0.1 – 4 mmol/l
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-PHOTOMETER tablet straig ht
from the foil to the water sample and crush the tablet
using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
The result is shown in the display as Acid demand to
pH 4.3 in mmol/l.
marks are aligned.
marks are aligned.
12
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Acid demand to pH 4.3 are
identical.
2. For accurate results, be sure to test on exactly 10 ml of water sample .
MC500 Colorimeter 07-2009
13
1.1 Methods
Alkalinity, total = Alkalinity-m =
3
0
m-Value
with Tablet
5 – 200 mg/l CaCO
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Ad d one ALKA-M-PHOTOMETER tablet straight
from the foil to the water sample and crush the tablet
using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
The result is shown in the display as total Alkalinity.
marks are aligned.
marks are aligned.
14
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Alkalinity to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
Fill a clean vial (24 mm Ø) with 10 ml of the water
close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-P-PHOTOMETER tablet straight
from the foil to the water sample and crush the tablet
using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure
that the
8. Press TEST key.
marks are aligned.
marks are aligned.
The result is shown in the display as Alkalinity-p.
16
MC500 Colorimeter 07-2009
1.1 Methods
Notes
1. The terms Alkalinity-p, p-Value and Alkalinity to pH 8.2 are identical.
2. For accurate test results exactly 10 ml of water sample must be taken for the test.
3. This method was developed from a volumetric procedure for the determination of
Alkalinity-p. Due to undened conditions, the deviations from the standardized method
may be greater.
4. Conversion table:
1 mg/l CaCO
mg/l CaCO
3
3
----0.0560.100.07
1 °dH17.8----1.781.25
1 °fH10.00.56----0.70
1 °eH14.30.801.43----
°dH°fH°eH
CaCO
3
°dH
°eH
°fH
°aH
5. By determining Alkalinity-p and Alkalinity-m it is possible to classify the alkalinity as
Hydroxide, Carbonate and Hydrogencarbonate.
The following dierentiation is only valid if:
a) no other alkalis are present and
b) Hydroxide and Hydrogen are not present in the same water sample.
Case 1: Alkalinity-p = 0
Hydrogen carbonate = m
Carbonate = 0
Hydroxide = 0
Case 2: Alkalinity-p > 0 and Alkalinity-m > 2p
Hydrogen carbonate = m – 2p
Carbonate = 2p
Hydroxide = 0
Case 3: Alkalinity-p > 0 and Alkalinity-m < 2p
Hydrogen carbonate = 0
Carbonate = 2m – 2p
Hydroxide = 2p – m
MC500 Colorimeter 07-2009
17
1.1 Methods
Aluminum
4
0
with Tablet
0.01 – 0.3 mg/l Al
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALUMINUM No. 1 tablet straight from the
foil to the water sample and crush the tablet using a
clean stirring rod (dissolve the tablet).
6. Add one ALUMINUM No. 2 tablet straight from the
foil to the same water sample and crush the tablet using
a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several
times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is nished the measurement starts
automatically.
marks are aligned.
marks are aligned.
18
The result is shown in the display in mg/l Aluminum.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Before using clean the vials and the measuring beaker with Hydrochloric acid
(approx. 20%). Rinse then thoroughly with deionized water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates.
The eect of this is generally insignicant unless the water has uoride added articially.
In this case. the following table should be used:
Fluoride
[mg/l F]
0.20.050.110.160.210.270.32
0.40.060.110.170.230.280.34
0.60.060.120.180.240.300.37
0.80.060.130.200.260.320.40
1.00.070.130.210.280.360.45
1.50.090.200.290.370.48---
Example: If the result of Aluminum determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F. the true concentration of Aluminum is
0.17 mg/l Al.
4.
Al
Al2O
3
Displayed value: Aluminum [mg/l Al]
0.050.100.150.200.250.30
MC500 Colorimeter 07-2009
19
1.1 Methods
12
34
Aluminum
5
0
with Powder Pack
0.01 – 0.25 mg/l Al
Use two clean vials (24 mm Ø) and mark one as blank
for zeroing.
1. Fill 20 ml of the water sample in a 100 ml beaker.
2.
Add the contents of one Aluminum ECR F20
Powder Pack straight from the foil to the water sample.
3. Dissolve the powder using a clean stirring rod.
Countdown 1
0:30
start:
Countdown 2
5:00
start:
4. Press [Wait for a reaction period of 30 seconds.
5.
Pack straight
6. Dissolve the powder using a clean stirring rod.
7. A d d 1 drop of Aluminum ECR Masking Reagent in the vial marked as blank.
8. Add 10 ml of the prepared water sample to the vial (this
is the blank).
9. Add the remaining 10 ml of the prepared water sample
in the second clean vial (this is the sample).
10. Close the vials tightly with the caps and swirl several
times to mix the contents.
11. Press [
Wait for a reaction period of 5 minutes.
] key.
After the reaction period is nished proceed as follows:
Add
the contents of
from the foil to the same water sample.
] key.
one Hexamine F20 Powder
20
MC500 Colorimeter 07-2009
1.1 Methods
After the reaction period is nished proceed as follows:
12. Place the vi al (the blan k) in the sample chamber
making sure that the marks are aligned.
prepare Zero
press ZERO
13. Press ZERO key.
14. Remove the vial from the sample chamber.
15. Place the vial (the sample) in the sample chamber
making sure that the
marks are aligned.
Zero accepted
prepare Test
16. Press TEST key.
press TEST
The result is shown in the display in mg/l Aluminum.
Notes:
1. Before using clean the vials and the measuring beaker with Hydrochloric acid
(approx. 20%). Rinse then thoroughly with deionized water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates.
The eect of this is generally insignicant unless the water has uoride added articially.
In this case, the following table should be used:
Fluoride
[mg/l F]
Displayed value: Aluminum [mg/l Al]
0.050.100.150.200.250.30
0.20.050.110.160.210.270.32
0.40.060.110.170.230.280.34
0.60.060.120.180.240.300.37
0.80.060.130.200.260.320.40
1.00.070.130.210.280.360.45
1.50.090.200.290.370.48---
Example: If the result of Aluminum determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F, the true concentration of Aluminum is
0.17 mg/l Al.
4.
Al
Al2O
3
MC500 Colorimeter 07-2009
21
1.1 Methods
Ammonium
6
0
with Tablet
0.02 – 1 mg/l N
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one AMMONIA No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean
stirring rod.
6. Add one AMMONIA No. 2 tablet straight from the foil to the same water sample and crush the tablet using a
clean stirring rod.
7. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
Wait for a reaction period of 10 minutes.
After the reaction period is nished the measurement
starts automatically.
marks are aligned.
marks are aligned.
22
The result is shown in the display in mg/l Ammonium.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. The tablets must be added in the correct sequence.
2. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2
tablet has been added.
3. The temperature of the sample is important for full color development.
At a temperature below 20°C the reaction period is 15 minutes.
4. Sea water samples:
Ammonia conditioning reagent is required when testing sea water or brackish water
samples to prevent precipitations of salts.
Fill the test tube with the sample to the 10 ml mark and add one level spoonful of
Conditioning Powder. Mix to dissolve, then continue as described in the test instructions.
5. Conversion:
mg/l NH4 = mg/l N x 1.29
mg/l NH3 = mg/l N x 1.22
6.
N
NH
NH
4
3
MC500 Colorimeter 07-2009
23
1.1 Methods
12
34
Ammonia
6
2
with Powder Pack
0.01 – 0.8 mg/l N
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill a clean vial (24 mm Ø) with 10 ml of deionized water
(this is the blank).
2. Fill the other clean vial (24 mm Ø) with 10 ml of the water sample (this is the sample).
3. Add the contents of one Ammonia Salicylate F10 Powder Pack straight from the foil to each vial.
4. Close the vials with the caps and shake to mix the contents.
Countdown 1
3:00
start:
Countdown 2
15:00
start:
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
5. Press [
Wait for a reaction period of 3 minutes.
6. Add the contents of one Ammonia Cyanurate F10
7. Close the vials tightly with the caps and shake to mix the
8. Press [
Wait for a reaction period of 15 minutes.
9. Place the vial (the blank) in the sample chamber making
10. Press ZERO key.
11. Remove the vial from the sample chamber.
12. Place the vial (the sample) in the sample chamber
13. Press TEST key.
The result is shown in the display in mg/l Ammonium.
] key.
After the reaction period is nished proceed as follows:
Powder Pack straight from the foil to each sample.
contents.
] key.
After the reaction period is nished proceed as follows:
sure that the marks are aligned.
making sure that the
marks are aligned.
24
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Extremely basic or acidic water samples should be adjusted with 0.5 mol/l (1 N) Sulfuric
acid solution or 1 mol/l (1 N) Sodium hydroxide solution to pH 7.
2. Interferences:
Interfering substanceInterference levels and treatments
Calciumgreater than 1000 mg/l CaCO
IronInterferes at all levels. Correct as follows:
a) determine the concentration of iron present in the
sample by performing a total Iron test
b) add the same iron concentration as determined
to the deionized water (step 1).
The interference will be blanked out successfully.
Magnesiumgreater than 6000 mg/l CaCO
Nitrategreater than 100 mg/l NO3-N
Nitritegreater than 12 mg/l NO2-N
Phosphategreater than 100 mg/l PO4-P
Sulfategreater than 300 mg/l SO
Suldeintensies the color
Glycine, Hydrazine,
Color, Turbidity
Less common interferences such as Hydrazine and
Glycine will cause intensied colors in the prepared
sample. Turbidity and color will give erroneous high
values. Samples with severe interferences require
distillation.
3
3
4
3.
N
NH
NH
4
3
MC500 Colorimeter 07-2009
25
1.1 Methods
12
34
Ammonia LR
6
5
with Vario Tube Test
0.02 – 2.5 mg/l N
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml
deionized water (this is the blank).
Countdown 1
20:00
start:
prepare Zero
press ZERO
Ø 16 mm
2. Open another white capped reaction vial and add
2 ml of the water sample (this is the sample).
3. Add the contents of one AMMONIA Salicy late F5 Powder Pack straight from the foil into each vial.
4. Add th e con tents of one AMM ONIA Cyanurate F5 Powder Pack straight from the foil into each vial.
5. Close the vials tightly with the caps and swirl several
times to dissolve the powder.
6. Press [Wait for a reaction period of 20 minutes.
7. Place the vial (the blank) in the sample chamber making
sure that the marks are l aligned.
8. Press ZERO key.
9. Remove the vial from the sample chamber.
] key.
After the reaction period is nished proceed as follows:
Zero accepted
prepare Test
press TEST
26
10. Place the vial (the sample) in the sample chamber
making sure that the marks are
11. Press TEST key.
The result is shown in the display in mg/l Ammonium.
l
aligned.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted to approx. pH 7 before analysis
(use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. If chlorine is known to be present, add one drop of 0.1 mol/l Sodium thiosulfate for
each 0.3 mg/l Cl2 in a one liter (1000 mL) water sample.
3. Iron interferes with the test. The interferences will be eliminated as follows:
Determine the amount of total iron present in the water sample. To produce the blank
add an iron standard solution with the same iron concentration to the vial (point 1)
instead of deionized water
4. Conversion:
mg/l NH4 = mg/l N x 1.29
mg/l NH3 = mg/l N x 1.22
5.
N
NH
NH
4
3
MC500 Colorimeter 07-2009
27
1.1 Methods
12
34
Ammonia HR
6
6
with Vario Tube Test
1 – 50 mg/l N
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 0.1 ml
deionized water (this is the blank).
Countdown 1
20:00
start:
prepare Zero
press ZERO
Ø 16 mm
2. Open anoth er white capped reaction vial and add
0.1 ml of the water sample (this is the sample).
3. Add the contents of one Vario AMMONIA Salicylate F5 Powder Pack straight from the foil into each vial.
4. Add the contents of one Vario AMMONIA Cyanurate F5 Powder Pack straight from the foil into each vial.
5. Close the vials tightly with the caps and swirl several
times to dissolve the powder.
6. Press [Wait for a reaction period of 20 minutes.
7. Place the vial (the blank) in the sample chamber making
sure that the marks are l aligned.
8. Press ZERO key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber
making sure that the marks are
] key.
After the reaction period is nished proceed as follows:
l
aligned.
Zero accepted
prepare Test
press TEST
28
11. Press TEST key.
The result is shown in the display
MC500 Colorimeter 07-2009
in mg/l
Ammonium.
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted to approx. pH 7 before analysis
(use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. If chlorine is known to be present, add one drop of 0.1 mol/l Sodium thiosulfate for each
0.3 mg/l Cl2 in a one litre water sample.
3. Iron interferes with the test. The interferences will be eliminated as follows:
Determine the amount of total iron present in the water sample. Add an iron standard
solution with the same concentration to the vial (point 1) instead of deionized water to
produce the blank.
4. Conversion:
mg/l NH4 = mg/l N x 1.29
mg/l NH3 = mg/l N x 1.22
5.
N
NH
NH
4
3
MC500 Colorimeter 07-2009
29
1.1 Methods
Boron
8
5
with Tablet
0.1 – 2 mg/l B
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample,
close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
20:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one BORON No. 1 tablet straight from the foil to the
6. Add one BORON No. 2 tablet straight from the foil to
7. Close the vial tightly with the cap and swirl several times
8. Place the vial in the sample chamber making sure that
9. Press TEST key.
Wait for a reaction period of 20 minutes.
After the reaction period is nished the measurement
marks are aligned.
the
water sample and crush the tablet using a clean stirring
rod and dissolve the tablet.
the same water sample and crush the tablet using a clean
stirring rod.
until the tablets are dissolved.
marks are aligned.
the
starts automatically.
30
The result is shown in the display in mg/l Boron.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. The tablets must added in the correct sequence.
2. The sample solution should have a pH value between 6 and 7.
3. Interferences are prevented by the presence of EDTA in the tablets.
4. The rate of color development depends on the temperature. The temperature of the
sample must be 20°C ± 1°C.
5.
B
H3BO
3
MC500 Colorimeter 07-2009
31
1.1 Methods
Bromine
8
0
with Tablet
0.05 – 13 mg/l Br
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and
crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
marks are aligned.
marks are aligned.
32
The result is shown in the display in mg/l Bromine.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Bromine may show lower results. To avoid any
measurement errors, only use glassware free of Chlorine consumption.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2. Preparing the sample:
When preparing the sample, the escape of Bromine gases, e.g. by pipetting or shaking,
must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD color development is carried out at a pH value of 6.3 to 6.5. The reagent
tablet therefore contains a buer for the pH adjustment. Strong alkaline or acidic water
samples must be adjusted between pH 6 and pH 7 before the reagent is added
(use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding of the measuring range:
Concentrations above 22 mg/l Bromine can produce results within the measuring range
up to 0 mg/l. In this event, the water sample must be diluted with water free of
Bromine. 10 ml of the diluted sample will be mixed with the reagent and the
measurement repeated.
Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Bromine.
MC500 Colorimeter 07-2009
33
1.1 Methods
63
Chloramine (Mono) and Free Ammonia
with Powder Pack and Liquid Reagent
0.04 – 4.50 mg/l Cl
2
Chloramine (Mono)
>> with NH4
without NH4
>> with NH4
>> without NH4
Notes:
1. Full color development – temperature
The reaction periods indicated in the manual refer to a sample temperature between
64.4° and 68°F (18° and 20°C). Due to the fact that the reaction period is strongly
influenced by sample temperature, you have to adjust both reaction periods according to
the following table:
Sample temperature in
°F / °C
41 / 510
50 / 108
60.8 / 166
68 / 205
73.4 / 232.5
77 / 252
The following selection is shown in the display:
for the determination of Monochloramine and free
Ammonia
for the determination of Monochloramine
Select the desired determination with the arrow keys [
and [ ]. Confirm with [ ] key.
Reaction period in min
]
34
MC500 Colorimeter 07-2009
1.1 Methods
12
34
Chloramine (Mono)
6
3
with Powder Pack
0.04 – 4.50 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. A dd the contents of one Mon och l or F RG T
6. Close the vial tightly with the cap and swirl several times
7. Place the vial in the sample chamber making sure that
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
The result is shown in the display in mg/l
mg/l NH2Cl
mg/l N
marks are aligned.
the
Powder Pack straight from the foil into the water
sample.
to mix the contents (approx. 20 seconds).
marks are aligned.
the
starts automatically.
Monochloramine.
mg/l Cl
2
MC500 Colorimeter 07-2009
Notes:see previous page
35
1.1 Methods
12
34
prepare Zero
press ZERO
63
Chloramine (Mono) and Free Ammonia
with Powder Pack and Liquid Reagent
0.04 – 4.50 mg/l Cl
0.01 – 0.50 mg/l NH4-N
Use two clean vials (24 mm Ø) and mark one as the
chloramine vial, the other as the ammonia vial.
1. Fill both vials (24 mm Ø) with 10 ml of water sample,
close tightly with the cap.
2. Place the chloramine vial in the sample chamber making
sure that the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one Monochlor F RGT Powder Pack straight from the foil into the chloramine vial.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Add 1 drop of Free Ammonia Reagent Solution into
the ammonia vial (Note 1).
2
marks are aligned.
Zero accepted
prepare T1
press TEST
Countdown
5:00
36
8. Close the vial tightly with the cap and invert several
times to mix the contents.
9. Place the chloramine vial in the sample chamber making
sure that the
10. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
11. Remove the vial from the sample chamber.
12. Add the contents of one Monochlor F RGT Powder Pack straight from the foil into the ammonia vial.
marks are aligned.
MC500 Colorimeter 07-2009
1.1 Methods
T1 accepted
prepare T2
press TEST
Countdown
5:00
13. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
14. Place the ammonia vial in the sample chamber making
sure that the
marks are aligned.
15. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l
Monochloramine and mg/l free Ammonia.
*,** mg/l Cl2
*,** mg/l NH2Cl
*,** mg/l N [NH2Cl]
The reading of Monochloramine is shown as:
mg/l Cl
2
mg/l NH2Cl
mg/l N
The reading of Ammonia is shown as N.
Conversion:
*,** mg/l N [NH4]
mg/l NH
mg/l NH3 = mg/l N · 1.22
= mg/l N · 1.29
4
Notes:
1. Hold the bottle vertically and squeeze slowly.
2. To determine the ammonia concentration the difference between the chloramine (T1)
and the sum of chloramine and ammonia (T2) is calculated. If T2 exceeds the range limit
the following message is displayed:
NH2CL + NH4 > 0.5 mg/l
In this case the sample has to be diluted and the measurement repeated.
3. see also page 34
MC500 Colorimeter 07-2009
37
1.1 Methods
Chloride
9
0
with Tablet
0.5 – 25 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
2:00
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CHLORIDE T1 tablet straight from the foil
to the water sample, crush the tablet using a clean
stirring rod and dissolve the tablet.
6. Add one CHLORIDE T2 tablet straight from the foil
to the same water sample and crush the tablet using a
clean stirring rod.
7. Close the vial tightly with the cap and swirl gently
several times until the tablet is dissolved (Note 1).
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is nished the measurement
starts automatically.
The result is shown in the display in mg/l Chloride.
marks are aligned.
marks are aligned.
38
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Ensure that all particles of the tablet are dissolved – Chloride causes an extremely ne
distributed turbidity with a milky appearance. Heavy shaking leads to bigger sized particles which can cause false readings.
2. High concentrations of electrolytes and organic compounds have dierent eects on the
precipitation reaction.
3. Ions which also form deposits with Silver nitrate in acidic media, such as Bromides,
Iodides and Thiocyanates, interfere with the analysis.
4. Highly alkaline water should - if necessary - be neutralized using Nitric acid before
analysis.
MC500 Colorimeter 07-2009
39
1.1 Methods
1
1
1
Chlorine
>> di
free
total
>> di
Chlorine
0
0
with Tablet
0.01 – 6 mg/l Cl
2
Chlorine
0
1
1
with Liquid Reagent
0.02 – 4 mg/l Cl
Chlorine
0
with Powder Pack
0.02 – 2 mg/l Cl
The following selection is shown in the display:
for the dierentiated determination of free, combined and
total Chlorine.
2
2
>> free
>> total
40
for the determination of free Chlorine.
for the determination of total Chlorine.
Select the desired determination with the arrow
] and [ ]. Conrm with [ ] key.
keys [
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Chlorine may show lower results. To avoid any
measurement errors, only use glassware free of Chlorine consumption.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2. For individual testing of free and total Chlorine, the use of dierent sets of glassware is
recommend (EN ISO 7393-2, 5.3)
3. Preparing the sample:
When preparing the sample, the escape of Chlorine gases, e.g. by pipetting or shaking,
must be avoided. The analysis must take place immediately after taking the sample.
4. The DPD color development is carried out at a pH value of 6.3 to 6.5. The reagents
therefore contain a buer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7
before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
5. Exceeding of the measuring range:
Concentrations above
10 mg/l Chlorine using tablets
4 mg/l Chlorine using liquid reagents
2 mg/l using powder packs
can produce results within the measuring range up to 0 mg/l. In this event, the water
sample must be diluted with water free of Chlorine. 10 ml of the diluted sample will be
mixed with the reagent and the measurement repeated.
6. Turbidity (lead to errors):
The use of the DPD No. 1 tablet (method 100) in samples with high Calcium ion
content* and/or high conductivity* can lead to turbidity of the sample and therefore
incorrect measurements. In this event, the reagent tablet DPD No. 1 High Calcium
should be used as an alternative. Even if the turbidity does occur after the DPD No. 3
tablet has been added, this can be prevented by using the DPD No. 1 HIGH CALCIUM
tablet.
* it is not possible to give exactly values, because the development of turbidity depends
on nature and ingredients of the sample.
7. If ??? is displayed at a dierentiated test result see page 260.
Oxidizing agents such as Bromine, Ozone etc. interfere as they react like Chlorine.
MC500 Colorimeter 07-2009
41
1.1 Methods
Chlorine, dierentiated determination
01
0
with Tablet
0.01 – 6 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare T1
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
marks are aligned.
marks are aligned.
42
10. Remove the vial from the sample chamber.
11. Add one DPD No. 3 tablet straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
12. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
MC500 Colorimeter 07-2009
1.1 Methods
13. Place the vial in the sample chamber making sure that
12. Place the vial in the sample chamber making sure that
the marks are aligned.
13. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is nished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
See page 41.
MC500 Colorimeter 07-2009
51
1.1 Methods
12
34
1
Chlorine, free
1
0
with Powder Pack
0.02 – 2 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Ad d th e c on te nts of on e Chlorine FREE -DPD / F10 Powder Pack straight from the foil to the water
sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
The result is shown in the display in mg/l free
Chlorine.
marks are aligned.
marks are aligned.
52
Notes:
See page 41.
MC500 Colorimeter 07-2009
1.1 Methods
12
34
1
Chlorine, total
1
0
with Powder Pack
0.02 – 2 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
3:00
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add th e content s of one Chlorin e TOTAL-DP D /
F10 Powder Pack straight from the foil to the water
sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is nished the measurement
starts automatically.
marks are aligned.
marks are aligned.
MC500 Colorimeter 07-2009
The result is shown in the display in mg/l total
Chlorine.
Notes:
See page 41.
53
1.1 Methods
1
Chlorine dioxide
>> with Cl
without Cl
2
Chlorine dioxide
0
with Tablet
0.05 – 11 mg/l ClO
The following selection is shown in the display:
2
>> with Cl
>> without Cl
for the determination of Chlorine dioxide in the presence
of Chlorine.
for the determination of Chlorine dioxide in the absence of
Chlorine.
Select the desired determination with the arrow keys
] and [ ]. Conrm with [ ] key.
[
54
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Chlorine dioxide may show lower results. To avoid
any measurement errors, only use glassware free of Chlorine consumption.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2. Preparing the sample:
When preparing the sample, the escape of Chlorine dioxide gases, e.g. by pipetting or
shaking, must be avoided. The analysis must take place immediately after taking the
sample.
3. The DPD color development is carried out at a pH value of 6.3 to 6.5. The reagent
tablet therefore contains a buer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7
before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding of the measuring range:
Concentrations above 19 mg/l Chlorine dioxide can produce results within the
measuring range up to 0 mg/l. In this event, the water sample must be diluted with
water free of Chlorine dioxide. 10 ml of the diluted sample will be mixed with the
reagent and the measurement repeated.
5. If ??? is displayed at a dierentiated test result see page 260.
Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Chlorine dioxide.
MC500 Colorimeter 07-2009
55
1.1 Methods
1
Chlorine dioxide
2
0
in the presence of Chlorine
with Tablet
0.05 – 11 mg/l ClO
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6.
sample.
7. Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod.
8. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
9. Transfer the contents of the second vial into the
prepared vial.
marks are aligned.
Fill a second clean vial with 10 ml of the water
Zero accepted
prepare T1
press TEST
56
10. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
11. Place the vial in the sample chamber making sure that
the
12. Press TEST key.
marks are aligned.
MC500 Colorimeter 07-2009
1.1 Methods
13. Remove the vial from the sample chamber, empty the
vial, rinse vial and cap several times. Fill with a few drops of the water sample.
14. Add one DPD No. 1 tablet straight from the foil and
crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times
17. Place the vial in the sample chamber making sure that
the
18. Press TEST key.
19. Remove the vial from the sample chamber.
20. Add one DPD No. 3 tablet straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
21. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that
the
23. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is nished the measurement
starts automatically.
The result is shown in the display in:
as Chlorine dioxide in mg/l Chlorine,
or
as Chlorine dioxide in mg/l ClO2.
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
marks are aligned.
marks are aligned.
MC500 Colorimeter 07-2009
Notes:
See next page.
57
1.1 Methods
Notes: (Chlorine dioxide in the presence of Chlorine)
1. The conversion factor to convert Chlorine dioxide as Chlorine to Chlorine dioxide as ClO2
is approximately 0.4 (more exactly 0.38).
mg/l ClO2 = mg/l ClO2 [Cl] x 0.38
ClO2[Cl]
ClO
(Chlorine dioxide displayed as Chlorine units ClO2 [Cl] has its origin out of the swimming
poolwater treatment according to DIN 19643.)
2. The total Chlorine result given includes the contribution by the Chlorine dioxide (as
Chlorine) reading. For true total Chlorine value subtract the Chlorine dioxide (as
Chlorine) reading from the quoted total Chlorine reading.
3. Also see page 55.
2
58
MC500 Colorimeter 07-2009
1.1 Methods
1
Chlorine dioxide
2
0
in absence of Chlorine
with Tablet
0.05 – 11 mg/l ClO
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it,
leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and
crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
marks are aligned.
marks are aligned.
*,** mg/l ClO2 [Cl]
*,** mg/l ClO2
MC500 Colorimeter 07-2009
The result is shown in the display
as Chlorine dioxide in mg/l Chlorine,
or
as Chlorine dioxide in mg/l ClO
Notes:
See page 55.
.
2
59
1.1 Methods
1
Chlorine HR (Kl)
0
5
with Tablet
5 – 200 mg/l Cl
Insert the adapter for 16 mm Ø vials.
1. Fill a clean vial (16 mm Ø) with 8 ml of the water sample, close tightly with the cap.
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
2. Place the vial in the sample chamber making sure that
the marks are
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CHLORINE HR (KI) tablet str aight from the foil to the water sample and crush the tablet using
a clean stirring rod.
6. Add one ACIDIFYING GP tablet straight from the
foil to the same water sample and crush the tablet using a
clean stirring rod.
7. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
8. Place the vial in the sample chamber making sure
that the marks are
9. Press TEST key.
l
aligned.
l
aligned.
60
The result is shown in the display in mg/l Chlorine.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Oxidizing agents interfere as they react like Chlorine.
MC500 Colorimeter 07-2009
61
1.1 Methods
1
COD LR
3
0
with Vario Tube Test
0 – 150 mg/l O
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml deionized water (this is the blank (Note 1)).
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
2. Open anoth er white capped reaction vial and add
2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently
several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor
at a temperature of 150°C.
5. (CAUTION: The vials are hot!) Remove the tubes from the heating block and allow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are
10. Press TEST key.
The result is shown in the display in mg/l COD.
l
aligned.
l
aligned.
62
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Run samples and blanks with the same batch of vials. The blank is stable when stored in
the dark and can be used for further measurements with vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for
nal measurements.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is
important to place the vials carefully in the sample chamber. The precipitate at the
bottom of the sample should be not suspended.
4.
Clean the outside of the vials with a towel to remove nger prints or other marks.
5. Samples can be measured when the Chloride contents does not exceed 1000 mg/l.
6. Be aware that low readings can result if the water contains compounds that cannot be
oxidized completely, as compared to the reference method.
MC500 Colorimeter 07-2009
63
1.1 Methods
1
COD MR
3
1
with Vario Tube Test
0 – 1500 mg/l O
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml deionized water (this is the blank (Note 1)).
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
2. Open anoth er white capped reaction vial and add
2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently
several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor
at a temperature of 150°C.
5. (CAUTION: The vials are hot!) Remove the tubes from the heating block and allow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are
10. Press TEST key.
The result is shown in the display in mg/l COD.
l
aligned.
l
aligned.
64
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Run samples and blanks with the same batch of vials. The blank is stable when stored in
the dark and can be used for further measurements with vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for
nal measurements.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is
important to place the vials carefully in the sample chamber. The precipitate at the
bottom of the sample should be not suspended.
4.
Clean the outside of the vials with a towel to remove nger prints or other marks.
5. Samples can be measured when the Chloride contents does not exceed 1000 mg/l.
6. Be aware that low readings can result if the water contains compounds that cannot be
oxidized completely, as compared to the reference method.
7. For samples under 100 mg/l COD it is recommended to repeat the test with the tube
test for COD LR.
MC500 Colorimeter 07-2009
65
1.1 Methods
31
Ø 16 mm
COD HR
2
with Vario Tube Test
0 – 15 g/l O2 (=^ 0 – 15 ,000 mg/l O2)
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 0.2 ml deionized water (this is the blank (Note 1)).
2. Open anoth er white capped reaction vial and add
0.2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently
several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor
at a temperature of 150°C.
5. (CAUTION: The vials are hot!) Remove the tubes from the heating block and allow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
66
6. Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are
10. Press TEST key.
The result is shown in the display in g/l COD.
MC500 Colorimeter 07-2009
l
aligned.
l
aligned.
1.1 Methods
Notes:
1. Run samples and blanks with the same batch of vials. The blank is stable when stored in
the dark and can be used for further measurements with vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for
nal measurements.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is
important to place the vials carefully in the sample chamber. The precipitate at the
bottom of the sample should be not suspended.
4.
Clean the outside of the vials with a towel to remove nger prints or other marks.
5. Samples can be measured when the Chloride contents does not exceed 1000 mg/l.
6. Be aware that low readings can result if the water contains compounds that cannot be
oxidized completely, as compared to the reference method.
7. For samples under 1 g/l COD it is recommended to repeat the test with the test kit for
COD MR or for samples under 0,1 g/l COD with the tube test COD LR.
MC500 Colorimeter 07-2009
67
1.1 Methods
2
prepare Zero
press ZERO
04
Color, true and apparent
(APHA Platinum-Cobalt
Standard Method)
0 – 500 Pt-Co units (mg/l Pt)
Sample preparation (Note 4):
Step A
Filter approx. 50 ml deionized water through a membrane
filter with a pore width of 0.45 µm.
Discard the filtrate. Filter another 50 ml deionized water
and keep it for zeroing.
Step B
Filter approx. 50 ml water sample using the same filter. Keep
this filtrate for sample measurement.
1. Fill a clean vial (24 mm Ø) with 10 ml of the filtrated deionized water (from Step A), close tightly with the
cap.
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
marks are aligned.
Zero accepted
prepare Test
press TEST
68
4. Remove the vial from the sample chamber and empty
it completely.
5. Rinse the vial with the filtrated water sample and fill
with 10 ml ltrated water sample (from Step B).
6. Place the vial in the sample chamber making sure that
the
7. Press TEST key.
marks are aligned.
The result is shown in the display in Pt-Co units (mg/l Pt).
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. This color scale was originally developed by A. Hazen as a visual comparison scale.
It is therefore necessary to ascertain whether the extinction maximum of the water
sample is in the range from 420 to 470 nm, as this method is only suitable for water
samples with yellowish to yellowish-brown coloration. Where applicable, a decision
should be made based on visual inspection of the water sample.
2. This method 204 – Hazen – is calibrated on the basis of the standards specified by
”Standard Methods for the Examination of Water and Wastewater” (also see EN ISO
7887:1994).
1 Pt-Co color unit = 1 mg/L of platinum as chloroplatinate ion
3. The estimated detection limit is 15 mg/L Pt.
4. Color may be expressed as “apparent” or “true” color. The apparent color is defined
as the color of a solution due to dissolved substances and suspended particles in the
sample. This manual describes the determination of true color by filtration of the water
sample. To determine the apparent color, non-filtrated deionized water and sample are
measured.
5. Sample collection, preservation and storage:
Pour the water sample into clean glass or plastic containers and analyse as soon as
possible after the sample is taken. If this is not possible, fill the container right up to the
top and seal tightly. Do not stir the sample; avoid lengthy contact with the air.
The sample may be stored in a dark place at a temperature of 39.2°F (4°C) for 24 hours.
Before performing measurements, the water sample must be brought up to room
temperature.
MC500 Colorimeter 07-2009
69
1.1 Methods
1
Copper
5
0
with Tablet
0.05 – 5 mg/l Cu
Copper
>> di
free
total
>> di
>> free
>> total
Note:
1. If ??? is displayed at the dientiated test result see page 260.
The following selection is shown in the display:
for the dierentiated determination of free, combined and
total Copper.
for the determination of free Copper.
for the determination of total Copper.
Select the desired determination with the arrow keys
] and [ ]. Conrm with [ ] key.
[
70
MC500 Colorimeter 07-2009
1.1 Methods
1
Copper, dierentiated determination
5
0
with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare T1
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
9. Remove the vial from the sample chamber.
10. Add one COPPER No. 2 tablet straight from the foil
to the same water sample and crush the tablet using a
clean stirring rod.
11. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
marks are aligned.
marks are aligned.
T1 accepted
prepare T2
press TEST
*,** mg/l free Cu
*,** mg/l comb Cu
*,** mg/l total Cu
MC500 Colorimeter 07-2009
12. Place the vial in the sample chamber making sure that
the
13. Press TEST key.
The result is shown in the display in:
mg/l free Copper
mg/l combined Copper
mg/l total Copper
marks are aligned.
71
1.1 Methods
1
Copper, free
5
0
with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
The result is shown in the display in mg/l free Copper.
marks are aligned.
marks are aligned.
72
MC500 Colorimeter 07-2009
1.1 Methods
1
Copper, total
5
0
with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tabl et and one COP PER
No. 2 tablet straight from the foil to the water sample
and crush the tablets using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
marks are aligned.
marks are aligned.
The result is shown in the display in mg/l total Copper.
MC500 Colorimeter 07-2009
73
1.1 Methods
12
34
1
Copper, free (Note 1)
5
3
with Powder Pack
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample,
close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
2:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one Cu 1 F10 Powder Pack straight
6. Close the vial tightly with the cap and swirl several times
7. Place the vial in the sample chamber making sure that
8. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is nished the measurement
marks are aligned.
the
from the foil to the water sample.
to mix the contents (Note 3).
marks are aligned.
the
starts automatically.
74
The result is shown in the display in mg/l Copper
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. For determination of Total Copper, a digestion is required.
2. Extremely acid water samples (pH 2 or less) must be adjusted between pH 4 and
pH 6 before the reagent is added (with 8 mol/l Potassium hydroxide solution KOH).
3. Accuracy is not aected by undissolved powder.
4. Interferences:
Cyanide, CN-Cyanide prevents full color development.
Add 0.2 ml Formaldehyde to 10 ml water sample and wait for a reaction
time of 4 minutes (Cyanide is masked). After this perform test as
described. Multiply the result by 1.02 to correct the sample dilution by
Formaldehyde.
+
Silver, Ag
If a turbidity remains and turns black, silver interferences is likely.
Add 10 drops of saturated Potassium chloride solution to 75 ml of water
sample. Filtrate through a ne lter. Use 10 ml of the ltered water
sample to perform test.
MC500 Colorimeter 07-2009
75
1.1 Methods
1
Cyanide
5
7
with Reagent Test
0.01 – 0.5 mg/l CN
1. Fill a clean vial (24 mm Ø) with 2 ml of the water sam-ple and 8 ml of deionized water, close tightly with the
cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add two level spoons No. 4 (grey) of Cyanide-11 into
6. Add two level spoons No. 4 (grey) of Cyanide-12,
7. Fill the vial with drops of the same size by holding the
3 drops of Cyanide-13
8. Close the vial tightly with the cap and invert several times
9. Place the vial in the sample chamber making sure that
10. Press TEST key.
Wait for a reaction period of 10 minutes.
After the reaction period is nished the measurement
marks are aligned.
the
the prepared water sample, replace the cap tightly and
invert the vial several times to mix the contents.
replace the cap tightly and invert the vial several times
to mix the contents.
bottle vertically and squeeze slowly:
to mix the contents.
the
marks are aligned.
starts automatically.
76
The result is shown in the display in mg/l Cyanide.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Only free Cyanide and Cyanides that can be destroyed by Chlorine are determined by
this test.
2. In the present of Thiocyanate, heavy metal complexes, colorants or aromatic amines, the
cyanide must be separated out by distillation before analysis is performed.
3. Store the reagents in closed containers at a temperature of + 15°C to + 25°C.
MC500 Colorimeter 07-2009
77
1.1 Methods
1
Cyanuric acid
6
0
with Tablet
2 – 160 mg/l Cys
1. Fill a clean vial (24 mm Ø) with 5 ml of the water
sample and 5 ml deionized water (Note 1), close
tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CYANURIC ACID tablet straight from the
foil to the prepared water sample and crush the tab let using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved (Note 2, 3).
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
The result is shown in the display in mg/l Cyanuric acid.
marks are aligned.
marks are aligned.
78
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Use deionized water or tap water free of Cyanuric acid.
2. If Cyanuric acid is present a cloudy solution will result.
3. Dissolve the tablet completely (therefore swirl the vial approx. 1 minute).
Undissolved particles of the tablet can cause too high results.
MC500 Colorimeter 07-2009
79
1.1 Methods
1
DEHA (N,N-Diethylhydroxylamine)
6
5
with Tablet and Liquid Reagent
20 – 500 μg/l DEHA / 0.02 – 0.5 mg/l DEHA
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample,
close tightly with the cap (Note 2).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Fill the vial with drops of the same size by holding the
6 drops (0.25ml) of DEHA solution
6. Close the vial tightly with the cap and swirl several times
7. Add one DEHA tablet straight from the foil to the same
8. Close the vial tightly with the cap and swirl several times
9. Place the vial in the sample chamber making sure that
10. Press TEST key.
Wait for a reaction period of 10 minutes.
After the reaction period is nished the measurement
The result is shown in the display as DEHA.
marks are aligned.
the
bottle vertically and squeeze slowly:
to mix the contents.
water sample and crush the tablet using a clean stirring
rod.
until the tablet is dissolved.
marks are aligned.
the
starts automatically.
80
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Application: Testing of residual corrosion inhibitors (Oxygen scavengers) in boiler feed
water or condensate.
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionized water.
3. Keep the sample dark during color development time. UV-light (sunlight) causes high
measurement results.
4. Ideal temperature for full color development is 20°C ± 2°C.
5. Interferences:
• Iron (II) interferes at all concentrations:
Repeat the test procedure but without adding the DEHA solution. If the
displayed result is above 20 μg/l subtract this value from the DEHA test result.
• Substances which reduce Iron (III) or complex iron may interfere.
• Substances which may interfere when present in concentrations at:
Borate (as Na2B4O7)500 mg/l
Cobalt0.025 mg/l
Copper8.0 mg/l
Hardness (as CaCO
Lignosulfonates0.05 mg/l
Manganese0.8 mg/l
Molybdenum80 mg/l
Nickel0.8 mg/l
Phosphate10 mg/l
Phosphonates10 mg/l
Sulfate1000 mg/l
Zinc50 mg/l
6. There is an option to change the unit from mg/l to μg/l.
mg/l
µg/l
)1000 mg/l
3
MC500 Colorimeter 07-2009
81
1.1 Methods
12
34
1
DEHA (N,N-Diethylhydroxylamine)
6
7
with Powder Pack
and Liquid Reagent
20 – 500 μg/l DEHA / 0.02 – 0.5 mg/l DEHA
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing (Note 2).
1. Fill a clean vial with 10 ml deionized water (this is the
blank).
2. Fill the second clean vial with 10 ml of the water sample
(this is the sample).
Countdown 1
10:00
start:
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
3. Add
4. Close the vials tightly with the caps and swirl several times
5. Add 0.20 ml DEHA 2 Rgt Solution to each vial (Note
6. Close the vials tightly with the caps and swirl several times
7. Press [
Wait for a reaction period of 10 minutes (Note 5).
8. Place the vial (the blank) in the sample chamber making
9. Press ZERO key.
10. Remove the vial from the sample chamber.
11. Place the vial (the sample) in the sample chamber making
12. Press TEST key.
The result is shown in the display as DEHA.
the contents of
straight from the foil into each vial.
to mix the contents.
4).
to mix the contents.
] key.
After the reaction period is nished proceed as follows:
sure that the marks are aligned.
sure that the
one OXYSCAV 1 Rgt Powder Pack
marks are aligned.
82
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Application: Testing of residual corrosion inhibitors (Oxygen scavengers) in boiler feed
water or condensate.
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionized water.
3. Ideally temperature for full color development is 25°C ± 3 °C.
4. Volume should always be metered by using suitable pipette (class A).
5. Keep blank and sample dark during color development time. UV-light (sunlight) causes
too high measurement results.
6. Interferences:
• Iron (II) interferes at all concentrations:
Repeat the test procedure but without adding the DEHA Rgt 2 solution. If the
displayed result is above 20 μg/l subtract this value from the DEHA test result.
• Substances which reduce Iron (III) interfere. Substances which complex iron strongly
may interfere also.
• Substances who may interfere when present in concentrations at:
Borate (as Na2B4O7)500 mg/l
Cobalt0.025 mg/l
Copper8.0 mg/l
Hardness (as CaCO
Lignosulfonates0.05 mg/l
Manganese0.8 mg/l
Molybdenum80 mg/l
Nickel0.8 mg/l
Phosphate10 mg/l
Phosphonates10 mg/l
Sulfate1000 mg/l
Zinc50 mg/l
)1000 mg/l
3
7. There is an option to change the unit from mg/l to μg/l.
mg/l
µg/l
MC500 Colorimeter 07-2009
83
1.1 Methods
1
Fluoride
7
0
with Liquid Reagent
0.05 – 2 mg/l F
Regard notes!
1. Fill a clean vial (24 mm Ø) with exactly 10 ml of water
sample (Note 4), close tightly with the cap.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add exactly 2 ml SPADNS reagent solution (Note 4)
to the water sample. Caution: Vial is lled up to the top! (Note 8)
6. Close the vial tightly with the cap and swirl several times
to mix the contents.
7. Place the vial in the sample chamber making sure that
the
Press TEST key.
The result is shown in the display in mg/l Fluoride.
marks are aligned.
marks are aligned.
84
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. The same batch of SPADNS reagent solution must be used for adjustment and test.
The adjustment process needs to be performed for each new batch of SPADNS reagent
solution (see Standard Methods 20th, 1998, APHA, AWWA, WEF 4500 F D., S. 4-82).
The procedure is described in chapter 2.4.5 “Calibration" Mode 40“ on page 233.
2. During adjustment and test the same vial should be used for zeroing and test, as dierent
vials may exhibit minor tolerances.
3. The calibration solution and the water samples to be tested should have the same
temperature (± 1°C).
4. As the test result is highly dependent on exact sample and reagent volumes, the sample
and reagent volumes should always be metered by using a 10 ml resp. 2 ml volumetric
pipette (class A).
5. The accuracy of the test methods decreases above a level of 1.2 mg/l Fluoride. Although
the results are suciently accurate for most applications, even more exact results can be
achieved by 1:1 dilution of the sample prior to use and subsequent multiplication of the
result by 2.
6. SPADNS reagent solution contains Arsenite.
Chlorine concentrations up to 5 mg/l do not interfere.
7. Seawater and wastewater samples must be distilled.
8. It is convenient to use larger volume vials for this test.
MC500 Colorimeter 07-2009
85
1.1 Methods
1
Hardness, Calcium
9
0
with tablet reagent
50 – 900 mg/l CaCO
1. Fill a clean vial (24 mm Ø) with 10 ml deionized water.
2. Add one CALCHECK tablet straight from the foil to
the deionized water and crush the tablet using a clean
stirring rod.
3. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
3
prepare Zero
press ZERO
Countdown
2:00
Zero accepted
prepare Test
press TEST
4. Place the vial in the sample chamber making sure that
the
5. Press ZERO key.
Wait for a reaction period of 2 minutes.
After the reaction period is nished the measurement
starts automatically.
6. Remove the vial from the sample chamber.
7. Add 2 ml of the water sample to the prepared vial.
Caution: Vial is lled up to the top! (Note 4)
8. Close the vial tightly with the cap and swirl several times
(5x) to mix the contents.
9. Place the vial in the sample chamber making sure that
the
10. Press TEST key.
marks are aligned.
marks are aligned.
The result is shown in the display as Calcium Hardness.
86
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10
before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2. The tolerance of the method is increased with higher concentrations. When diluting
samples, this should be taken into account, always measuring in the rst third of the range.
3. This method was developed from a volumetric procedure for the determination of
calcium. Due to undened conditions, the deviations from the standardized method may
be greater.
4. It is convenient to use larger volume vials for this test.
5.
CaCO
°dH
°eH
°fH
°aH
3
MC500 Colorimeter 07-2009
87
1.1 Methods
1
prepare Zero
press ZERO
91
Hardness, Calcium 2T
with Tablet
0 – 500 mg/l CaCO
1. Fill a clean vial (24 mm Ø) with 10 ml of water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CALCIO H No. 1 tablet straight from the
6. Add one CALCIO H No. 2 tablet straight from the foil
7. Close the vial tightly with the cap and swirl gently several
marks are aligned.
foil to the 10 ml water sample, crush the tablet using
a clean stirring rod and dissolve the tablet completely.
to the same water sample and crush the tablet using a
clean stirring rod.
times until the tablet is completely dissolved.
3
Zero accepted
prepare Test
press TEST
Countdown
2:00
88
8. Place the vial in the sample chamber making sure that
the
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as Calcium
Hardness
marks are aligned.
.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. To optimize the readings an optional batch related calibration can be performed using
Mode 40, see page 231.
2. Strong alkaline or acidic water samples must be adjusted to a pH-value between
pH 4 and 10 before the tablets are added (use 1 mol/l Hydrochloride acid resp.
1 mol/l Sodium hydroxide).
3. For accurate test results exactly 10 ml of water sample must be taken for the test.
4. This method was developed from a volumetric procedure for the determination of
Calcium Hardness. Due to undefined conditions, the deviations from the standardised
method may be greater.
5. The tolerance of the method is increasing with higher concentrations. When diluting
samples, this should be taken in account, always measuring in the first third of the
range.
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one HARDCHECK P tablet straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is nished the measurement
starts automatically.
marks are aligned.
marks are aligned.
90
The result is shown in the display as total Hardness.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10
before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2. Conversion table:
1 mg/l CaCO
mg/l CaCO
3
3
----0.0560.100.07
1 °dH17.8----1.781.25
1 °fH10.00.56----0.70
1 °eH14.30.801.43----
CaCO
3.
3
°dH
°eH
°fH
°aH
°dH°fH°eH
MC500 Colorimeter 07-2009
91
1.1 Methods
2
Hardness, total HR
0
1
with Tablet
20 – 500 mg/l CaCO
1. Fill a clean vial (24 mm Ø) with 1 ml of the water sam-ple and 9 ml of deionized water, close tightly with the
cap.
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2. Place the vial in the sample chamber making sure that
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one HARDCHECK P tablet straight from the foil
6. Close the vial tightly with the cap and swirl several times
7. Place the vial in the sample chamber making sure that
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is nished the measurement
marks are aligned.
the
to the water sample and crush the tablet using a clean
stirring rod.
until the tablet is dissolved.
marks are aligned.
the
starts automatically.
92
The result is shown in the display as total Hardness.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10
before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2. Conversion table:
1 mg/l CaCO
mg/l CaCO
3
3
----0.0560.100.07
1 °dH17.8----1.781.25
1 °fH10.00.56----0.70
1 °eH14.30.801.43----
CaCO
3.
3
°dH
°eH
°fH
°aH
°dH°fH°eH
MC500 Colorimeter 07-2009
93
1.1 Methods
2
Hydrazine
0
5
with powder reagent
0.05 – 0.5 mg/l N2H4 / 50 – 500 µg/l N2H
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample (Note 1, 2), close tightly with the cap.
4
prepare Zero
press ZERO
Countdown
10:00
start:
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add 1 g HYDRAZINE test powder (Note 3) to the
water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents.
7. Press [
Wait for a reaction period of 10 minutes.
8. The slight turbidity that occurs when the reagent is added
must be removed by ltration (Note 4).
9. Place the vial in the sample chamber making sure that
the
marks are aligned.
] key.
After the reaction period is nished proceed as follows:
marks are aligned.
Zero accepted
prepare Test
press TEST
94
10. Press TEST key.
The result is shown in the display as Hydrazine.
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. If the water sample is cloudy, it should be ltered it before performing the zero calibration.
2. The temperature of the water sample should not exceed 21°C.
3. Using the Hydrazine spoon: 1 g is equivalent to one level spoon.
4. Qualitative folded lter papers for medium precipitates are recommended.
5. In order to check whether the reagent has aged (if it has been stored for a lengthy period),
perform the test as described above using tap water. If the result is above the
detection limit of 0.05 mg/l, you should use new reagents, as it could result in major result
deviations.
6. There is an option to change the unit from mg/l to μg/l.
mg/l
µg/l
MC500 Colorimeter 07-2009
95
1.1 Methods
2
Hydrazine
0
6
with Liquid Reagent
0.005 – 0.6 mg/l N2H4 / 5 – 600 µg/l N2H
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill a clean vial with 10 ml deionized water (this is the
blank).
2. Add 1 ml Hydra 2 Rgt Solution into the vial (Note 3).
3. Close the vial tightly with the cap and swirl several times
to mix the contents.
4
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
12:00
4. Place the vial (the blank) in the sample chamber making
sure that the
5. Press ZERO key.
6. Remove the vial from the sample chamber.
7. Fill the second clean vial with 10 ml of the water sample
(this is the sample).
8. Add 1 ml Hydra 2 Rgt Solution into the vial.
9. Close the vial tightly with the cap and swirl several times
to mix the contents.
10. Place the vial (the blank) in the sample chamber making
sure that the
11. Press TEST key.
Wait for a reaction period of 12 minutes.
After the reaction period is nished the measurement
starts automatically.
The result is shown in the display as Hydrazine.
marks are aligned.
marks are aligned.
96
MC500 Colorimeter 07-2009
1.1 Methods
Notes:
1. Samples cannot be preserved and must be analyzed immediately.
2. Sample temperature should be 21°C ± 4 °C.
3. Caused by the reagent itself the blank may develop a faint yellow color.
4. Interferences:
• Ammonium causes no interferences up to 10 mg/l.
At a concentration of 20 mg/l it is possible. that the test result increase up to 20 %.
• Morpholine does not interfere up to 10 mg/l.
• Highly colored or turbid samples:
Mix 1 part deionized water with 1 part household bleach. Add 1 drop of this
mixture into 25 ml water sample and mix. Use 10 ml prepared sample in place of
deionized water in point 1.
Note: at point 7 use the unprepared water sample.
Principle: Hydrazine is oxidized by the household bleach. The interference by color
will be eliminated by zeroing.
5. There is an option to change the unit from mg/L to μg/L.
mg/l
µg/l
MC500 Colorimeter 07-2009
97
1.1 Methods
2
Hydrazine
0
7
with Vacu-vials
0.01 – 0.7 mg/l N2H4 / 10 – 700 µg/l N2H
Insert the adapter for 13 mm Ø vials.
1. Place the blank in the sample chamber. The blank is part
of the test kit.
®
K-5003 (see Notes)
4
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
2. Press ZERO key.
3. Remove the blank from the sample chamber.
4. Fill the sample container to the 25 ml mark with the water
sample.
®
5. Place one Vacu-vial
tip by pressing the vial against the side of the sample
container. The Vacu-vial® breaks at the neck and the vial
lls automatically. A small volume of inert gas remains in
the Vacu-vial®.
6. Mix the contents of the Vacu-vial
times, allowing the bubble to move from one end to the
other. Dry the outside of the vial.
7. Place the Vacu-vial
8. Press TEST key.
Wait for a reaction period of 10 minutes.
After the reaction period is nished the measurement
starts automatically.
in the sample container. Snap the
®
by inverting it several
®
in the sample chamber.
98
The result is shown in the display as Hydrazine.
MC500 Colorimeter 07-2009
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.