Optris CT LT Specifications

optris
Separate electronics with easy accessible programming keys and LCD
Selectable analog output: 0/4 20 mA, 0 5 V, 0 10 V,
®
CT LT
TECHNICAL DATA
Precise non-contact temperature measurement from –50 °C to 975 °C
Features:
• Rugged and usable up to 180 °C ambient temperature without cooling
• Separate electronics with easy accessible programming keys and LCD backlit display
• Selectable analog output: 0/4 – 20 mA, 0 – 5 V, 0 – 10 V, thermocouple type K or J
• Optional USB, RS485, RS232 interface, relay outputs (2 x optically isolated),
CAN-Bus, Profi bus DP, Ethernet
Installation of up to 32 sensing heads
CTex: Explosion proof version (ATEX)
General specifi cations
Environmental rating IP 65 (NEMA-4)
Ambient temperature –20 °C ... 180 °C (130°C to LT02) (sensing head)
Storage temperature – 40 °C ... 130 °C (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 40 g (sensing head) / 420 g (electronics)
0 °C ... 85 °C (electronics)
– 40 °C ... 85 °C (electronics)
Electrical Specifi cations
Outputs / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 1 m (standard), 3 m, 8 m, 15 m
Power Supply 8 – 36 V DC
Channel 1:0 / 4 Channel 2: sensind head temperature (–20 °C ... 180 °C as 0 – 5 V or 0 – 10 V), alarm output
optically isolated
(optional)
mV min. 100 kΩ load impedance thermocouple 20 Ω
emissivity adjustment, ambient temperature compensation, trigger (reset of hold functions)
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
eff
Measurement specifi cations
Temperature range (scalable via programming keys or software)
Spectral range 8 – 14 μm
Optical resolution (90 % energy)
CF-lens (optional) 0.6 mm @ 10 mm (with LT22)
System accuracy (at ambient temp. 23 ±5 °C)
Repeatability (at ambient temp. 23 ±5 °C)
Temperature resolution (display)
2), 3)
NETD
Response time 150 ms (95 %)
Emissivity/ Gain (adjustable via programming keys or software)
Transmissivity/ Gain (adjustable via programming keys or software)
Signal processing (parameter adjustable via programming keys or software, respectively)
Software optris
1)
Whichever is greater
2)
At object temperatures >0 °C, ε = 1
3)
At time constant 200 ms and T
1), 2)
1), 2)
–50 °C ... 975 °C (LT22) –50 °C ... 600 °C (LT15) –50 °C ... 600 °C (LT02)
22:1 (precision glass optics) 15:1 (precision glass optics) 2:1 (with flat front window)
0.8 mm @ 10 mm (with LT15)
2.5 mm @ 23 mm (with LT02)
±1 % or ±1 °C
±0.5 % or ±0.5 °C
0.1 K
0.05 K (LT22 / LT15)
0.1 K (LT02)
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold function with threshold and hysteresis
®
Compact Connect
25 °C
Obj
Current draw Max. 100 mA
Optris GmbH · Ferdinand-Buisson-Str. 14 · 13127 Berlin · Germany Phone: +49 (0)30 500 197-0 · Fax: +49 (0)30 500 197-10 · E-mail: sales@optris.com · www.optris.com
optris
Specifi cations are subject to change without notice · CT LT-E2017-10-A
Optics CT LT CF, D:S = 22:1 (far eld = 1.5:1)
Optics CT LT SF, D:S = 15:1
(mm)
Optical specifi cations
®
CT LT
Optics CT LT CF, D:S = 2:1 (far fi eld = 2.5:1)Optics CT LT SF, D:S = 2:1
S 5 53.8 102.5 151.3 200 251.3 302.5 353.8 405
D 0 100 200 300 400 500 600 700 800
Optics CT LT SF, D:S = 15:1
S 6.5 11.6 16.6 21.7 26.7 35 43.3 51.6 59.9 68.2 76.5
D 0 100 200 300 400 500 600 700 800 900 1000
Optics CT LT SF, D:S = 22:1
S 6.5 10.9 15.2 19.6 23.9 28.3 32.6 37 41.3 45.7 50
D 0 110 220 330 440 550 660 770 880 990 1100
Versions with built-in CF lenses
(mm)
(mm)
S 7 5 3.1 2.5 5.4 9.5 13.7 17.8
D 0 10 20 23 30 40 50 60
Optics CT LT CF, D:S = 15:1 (far fi eld = 1.5:1)
(mm)
(mm)
S 7 3.9 0.8 4.7 8.6 12.5 16.4 20.3 24.2
D 0 5 10 15 20 25 30 35 40
Optics CT LT CF, D:S = 22:1 (far fi eld = 1.5:1)
(mm)
(mm)
S 7 3.8 0.6 4.4 8.2 12 15.8 19.6 23.4
D 0 5 10 15 20 25 30 35 40
Optics CT LT CF, D:S = 22:1 (far fi eld = 1.5:1)Optics CT LT SF, D:S = 22:1
(mm)
(mm)
(mm)
(mm)
(mm)
(mm)
S 6.5 5 4 3.4 5.4 8.4 13.3 18.3 23.2 28.2 33.1
D 0 25 40 50 60 75 100 125 150 175 200
Dimensions
Sensing head (standard)
Sensing head (built-in CF lenses)
(mm)
(mm)
S 6.5 4.4 3.2 2.3 4.1 6.7 11.1 15.5 19.9 24.3 28.7
D 0 25 40 50 60 75 100 125 150 175 200
Electronics
(mm)
(mm)
Loading...