TECHNICAL DATA
Non-contact temperature measurement through ames from 200 °C to 1650 °C
Accurate temperature measurement through ames in the range from 200 °C to 1650 °C (392 °F to 2912 °F) to monitor workpieces inside ovens (heated with re), to measure inside chemical reactors and to observe the brick temperature in combustion chambers
Double laser aiming marks real spot location and spot size up from 1.6 mm (0.06 in) at any distance
Optics 45:1 with selectable focus, compact sensor head size
Usable up to 85 °C (185 °F) ambient temperature without cooling and automatic laser switch o󰀨 at 50 °C (122 °F)
Cooling and protection accessories for harsh environmental conditions
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature
Storage temperature – 40 °C ... 125 °C (–4 °F ... 257 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
1)
20 °C ... 85 °C (–4 °F ... 185 °F) (sensing head) (50 °C [122 °F] with laser ON) 20 °C ... 85 °C (–4 °F ... 185 °F) (electronics)
– 40 °C ... 85 °C (–40 °F ... 185 °F) (electronics)
420 g (14.8 oz) (electronics)
Electrical specications
Output / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Output / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 5 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 3 m (standard), 8 m, 15 m
0 / 4
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
emissivity adjustment, ambient temperature compensation, trigger (reset of hold functions)
(9.8 ft [standard], 26.2 ft, 49.2 ft)
eff
Measurement specications
Temperature range (scalable via programming keys or software)
Spectral range 3.9 μm
Optical resolution (90 % energy)
System accuracy (at ambient temp. 23 ±5 °C) (at ambient tem. 73 ±9 °F)
Repeatability (at ambient temp. 23 ±5 °C) (at ambient tem. 73 ±9 °F)
Temperature resolution (digital) 0.1 K
Exposure time
Emissivity/ Gain (adjustable via programming keys or software)
IR window correction (adjustable via programming keys or software)
Signal processing (parameter adjus­table via software)
Software optris
1)
The functioning of the LCD display may be limited in ambient temperatures
below 0 °C
2)
For object temperatures above >300 °C
3)
ε = 1, response time 1 s
4)
With dynamic adaptation at low signal levels
4)
(90 % signal) 10 ms
200 °C ... 1450 °C (MT) (392 °F ... 2642 °F [MT]) 400 °C ... 1650 °C (MTH) (752 °F ... 2912 °F [MTH])
45:1
2) 3)
±1 %
±0.5 % or +0.5 °C (±0.5 % or +0.9 °F
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold function with threshold and hysteresis
®
Compact Connect
3)
3)
Power supply 8 – 36 V DC
Current draw max. 160 mA
Laser 635 nm 1 mW, ON/OFF via electronic box or software
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
®
CTlaser MT
Chart SF optics, D:S = 45:1
SF
27 mm @ 1200 mm (1.06 in @ 47.2 in)
45:1
S
(0.79)
D 0
20
21.7
23.4
25
27
38.4
50
(0.85)
(0.92)
(0.98)
(1.06)
(1.51)
300
600
900
1200
(0)
(11.8)
(23.6)
(35.4)
(47.2)
1500 (59.0)
(1.97)
1800
(70.8)
61.7
(2.43)
2100
(82.7)
(2.89)
(94.5)
Dimensions
Sensing head Electronics
73.4
2400
Further optics, D:S = 45:1
... SF 27 mm @ 1250 mm
... CF1 1.6 mm @ 70 mm
... CF2 3.4 mm @ 150 mm
... CF3 4.5 mm @ 200 mm
mm (in)
mm (in)
... CF4 10 mm @ 450 mm
(1.06 in @ 49.2 in)
(0.06 in @ 2.76 in)
(0.13 in @ 5.91 in)
(0.18 in @ 7.87 in)
(0.39 in @ 17.7 in)
Accessories
Mounting angle, adjustable in two axes (ACCTLAB)
Water cooling and air purge for sensing head (ACCTLW + ACCTLAP)
Cooling housing
(ACCJCTL)
Mounting angle for cooling housing, adjustable in two axes (ACCJAB)
Mounting device for cooling housing
(ACCTLRM)
Specications are subject to change without notice · CTlaser MT-US2018-08-A