Optris CTlaser LT, CTlaser LTF Specifications

optris
®
CTlaser LT
TECHNICAL DATA
Precise focussing and non- contact temperature measure- ment from –50 °C to 975 °C (–58 °F to 1787 °F)
Low and high temperature measurements of smallest spots up from 0.9 mm (0.04 in)
Double laser aiming marks real spot location and spot size at any distance
Optics 75:1 and 50:1 with selectable focus
CT laser F (fast) for scanning of fast moving low temperature objects up from 9 ms response time
Usable up to 85 °C (185 °F) ambient temperature without cooling and
automatic laser switch o󰀨 at 50 °C (122 °F)
Selectable analog outputs 0/4 – 20 mA, 0 – 5/ 10 V, thermocouple type K or J
Optional plug in digital interfaces USB, RS232, RS485, CAN or Probus DP
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature
Storage temperature 40 °C ... 125 °C (–4 °F ... 257 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
1)
20 °C ... 85 °C (–4 °F ... 185 °F) (sensing head) (50 °C [122 °F] with laser ON) 20 °C ... 85 °C (–4 °F ... 185 °F) (electronics)
40 °C ... 85 °C (40 °F ... 185 °F) (electronics)
420 g (14.8 oz) (electronics)
Electrical specications
Output / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Output / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 5 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 3 m (standard), 8 m, 15 m
Power supply 8 – 36 V DC
Current draw (laser) Max. 160 mA
Laser 635 nm 1 mW, ON/OFF via electronic box or software
Channel 1: 0 / 4 Channel 2: sensing head temperature (– 40 °C ... 85 °C [–40 °F ... 185 °F) as 0 – 5 V or 0 – 10 V), alarm output
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
emissivity adjustment, ambient temperature compensation, trigger (reset of hold functions)
(9.8 ft [standard], 26.2 ft, 49.2 ft)
20 mA, 0
5/ 10 V, thermocouple J, K
; 0.4 A;
eff
Measurement specications
Temperature range (scalable via programming keys or software)
Spectral range 8 – 14 μm
Optical resolution (90 % energy)
Selectable focus
1)
(CTlaser)
System accuracy (at ambient temp. 23 ±5 °C) (at ambient tem. 73 ±9 °F)
Repeatability (at ambient temp. 23 ±5 °C) (at ambient tem. 73 ±9 °F)
Temperature resolution (NETD)
Response time
Emissivity/ Gain (adjustable via sensor or software)
IR window correction (adjustable via software)
Signal processing (parameter adjus­table via software)
Software optris
1)
The functioning of the LCD display may be limited in ambient temperatures
below 0 °C
2)
Different spotsizes for CTlaser F (D:S = 50:1)
3)
Whichever is greater
4)
At object temperatures >0 °C, ε = 1
5)
With dynamic adaption at low signal levels
2)
5)
(90 % signal) 9 ms CTlaser F / 120 ms CTlaser
–50 °C ... 975 °C (–58 °F ... 1787 °F)
75:1 CTlaser 50:1 CTlaser F
CF1: 0.9 mm @ 70 mm (0.04 in @ 2.76 in) CF2: 1.9 mm @ 150 mm (0.07 in @ 5.91 in) CF3: 2.75 mm @ 200 mm (0.11 in @ 7.87 in) CF4: 5.9 mm @ 450 mm (0.23 in @ 17.7 in) SF: 16 mm @ 1200 mm (0.63 in @ 47.2 in)
±1 % or ±1 °C ±1.5 % or ±1.5 °C (±1 % or ±1.8 °F (±1.5 % or ±2.7 °F
±0.5 % or ±0.5 °C ±1 % or ± 1 °C (±0.5 % or ±0.9 °F) (±1 % or ±1.8 °F)
0.1 K / 0.5 K with CTlaser F
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold function with threshold and hysteresis
3),4)
(CTlaser)
3),4)
3),4)
[CTlaser])
3),4)
2),3)
2),3)
(CTlaser F)
2),3)
2),3)
®
Compact Connect
(CTlaser F)
[CTlaser F])
(CTlaser)
(CTlaser)
(CTlaser F)
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
20
S
(079)
D 0
(0)
®
CTlaser LT
19
18
17
(0.75)
300
(11.8)
(0.71)
600
(23.6)
(0.67)
900
(35.4)
16
(0.63)
1200
(47.2)
25
(0.98)
1500
(59.0)
34
(1.34)
1800
(70.8)
43
(1.69)
2100
(82.7)
52
(2.05)
2400
(94.5)
mm (in)
mm (in)
CF1 optics 75:1SF optics 75:1
20
0.9
25
55
70
100
130
160
190
220
S
(079)
D 0
(0.04)
(0.98)
(2.17)
(2.76)
(3.94)
(5.12)
(6.30)
(7.48)
70
150
250
300
400
500
600
(98.4)
(11.8)
(15.7)
(2.8)
(0)
(5.91)
(19.7)
(23.6)
700
(27.6)
(8.66)
800
(31.5)
mm (in)
mm (in)
CF3 optics 75:1CF2 optics 75:1
20
S
(079)8 (0.31)9 (0.35)
D 0
100
(0)
(3.94)
200
(7.87)
24
(0.94)
300
(11.8)
38
(1.50)
400
(15.7)
53
(2.09)
500
(19.7)
68
(2.68)
600
(23.6)
82
(3.23)
700
(27.6)
97
(3.82)
800
(31.5)
mm (in)
mm (in)
CF4 optics 75:1
20
17
14
11
S
D 0
(079)
(0.67)
(0.55)
100
(3.94)
200
(7.87)
(0)
(0.47)
300
(11.8)
7.5
(0.30)9 (0.35)
400
(15.7)
500
(19.7)
15
(0.59)
600
(23.6)
20
(079)
700
(27.6)
26
(1.02)
800
(31.5)
mm (in)
mm (in)
Dimensions
Sensing head Electronics
S
(079)
D 0
20
11
2.75
14
25,5
37
48
60
71
(0.47)
(0.11)
(0.55)
(1.00)
(1.47)
(1.89)
(2.36)
400
500
600
100
200
300
(15.7)
(0)
(3.94)
(7.87)
(11.8)
(19.7)
(23.6)
700
(27.6)
(2.80)
800
(31.5)
mm (in)
mm (in)
Specications are subject to change without notice · CTlaser LT-US2018-10-A