![](/html/c4/c431/c4317e9c3057a58b4c73cf6de227d18cffe278ca9e661af5dd2e4c20863a9f3b/bg1.png)
optris CTlaser G7
TECHNICAL DATA
Non-contact temperature
measurement of ultra-thin
glass sheets from 100 °C
to 1200 °C
Features:
• Accurate temperature measurement of ultra-thin flat glass used for touch displays
of smartphones, tablets etc.
• Double laser aiming for exact marking of measurement spot
• Optics 45:1 with selectable focus, compact sensor head size
• Suitable for ambient temperatures of up to 85 °C without additional cooling
• Cooling and protection accessories for rugged environments
General specifications
Environmental rating IP 65 (NEMA-4)
Ambient temperature
Storage temperature – 40 ... 85 °C (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (sensing head) / 420 g (electronics)
1)
–20 ... 85 °C (sensing head, 50 °C with laser ON)
–20 ... 85 °C (electronics)
– 40 ... 85 °C (electronics)
Electrical specifications
Outputs / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Inputs Programmable functional inputs for external
0 / 4
–
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance
thermocouple 20 Ω
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
eff
Measurement specifications
Temperature range
(scalable via programming
keys or software)
Spectral range 7.9 μm
Optical resolution
(90 % energy)
System accuracy
(at ambient temp. 23 ±5 °C)
Repeatability
(at ambient temp. 23 ±5 °C)
Temperature resolution
(digital)
Response time
Emissivity/ Gain
(adjustable via programming
keys or software)
Transmissivity/ Gain
(adjustable via programming
keys or software)
Signal processing (parameter
adjustable via programming
keys or software, respectively)
Software optris Compact Connect
1)
The functioning of the LCD display may be limited in ambient temperatures
below 0 °C
2)
ε = 1, response time 1 s
3)
Whichever is greater
4)
With dynamic adaptation at low signal levels
2)
4)
(90 % signal) 150 ms
100 ... 1200 °C
45:1
±1 % or ±1.5 °C
±0.5 %
0.5 K
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
3)
Cable length 3 m (standard), 8 m, 15 m
Power Supply 8 – 36 V DC
Current draw Max. 160 mA
Laser 635 nm 1 mW, ON/OFF via electronic box or software
Optris GmbH · Ferdinand-Buisson-Str. 14 · 13127 Berlin · Germany
Phone: +49 (0)30 500 197-0 · Fax: +49 (0)30 500 197-10 · E-mail: info@optris.global · www.optris.global
![](/html/c4/c431/c4317e9c3057a58b4c73cf6de227d18cffe278ca9e661af5dd2e4c20863a9f3b/bg2.png)
optris CTlaser G7
Optical specications
Chart SF optics, D:S = 45:1 Further optics, D:S = 45:1
... SF 27 mm @ 1250 mm
... CF1 1.6 mm @ 70 mm
... CF2 3.4 mm @ 150 mm
... CF3 4.5 mm @ 200 mm
SF
27 mm @
1200 mm
45:1
S 20 20.8 21.7 22.5 23.4 24.2 25 25.9 27 32.5 38.4 50 61.7 73.4
D 0 150 300 450 600 750 900 1050 1200 1350 1500 1800 2100 2400
Dimensions
(mm)
(mm)
... CF4 10 mm @ 450 mm
Sensing head Electronics
Accessories (examples)
Mounting angle, adjustable
in two axes (ACCTLAB)
Cooling housing
(ACCJCTL)
Mounting angle for cooling housing,
adjustable in two axes (ACCJAB)
Water cooling and air purge for
sensing head (ACCTLW + ACCTLAP)
Mounting device for cooling housing
(ACCTLRM)
Specifications are subject to change without notice · CTlaser G7-EN2019-04-A