optris
®
CTlaser G5
TECHNICAL DATA
Infrared thermometer with
laser visor for temperature
measurements on glass
from 100 °C to 1650 °C
(212 °F to 2912 °F)
Features:
• Precise temperature measurements on glat glass, container glass, light bulbs, car
glass and solar pannels
• Temperature ranges from 100 °C to 1650 °C (212 °F to 2912 °F), measurement
elds from 1.0 mm (0.04 in) and exposure time 10 ms
• Innovative double laser visor for exact marking of measuring eld
• Compact size of sensing head
• Suitable for ambient temperatures of up to 85 °C (185 °F) without additional
cooling
• Cooling and protection accessories for rugged environments
General specications
Environmental rating IP 65 (NEMA-4) front mountable
Ambient temperature
Storage temperature – 40 °C ... 125 °C (–4 °F ... 257 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
1)
at vacuum processes (up to 10
– 20 °C ... 85 °C (–4 °F ... 185 °F) (sensing head)
(50 °C [122 °F] with laser ON)
– 20 °C ... 85 °C (–4 °F ... 185 °F) (electronics)
– 40 °C ... 85 °C (–40 °F ... 185 °F) (electronics)
420 g (14.8 oz) (electronics)
–3
mbar)
Electrical specications
Output / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Output / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 5 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 3 m (standard), 8 m, 15 m
0 / 4
–
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
(9.8 ft [standard], 26.2 ft, 49.2 ft)
eff
Measurement specications
Temperature range
(scalable via programming keys or
software)
Spectral range 5.0 μm
Optical resolution
(90 % energy)
System accuracy
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Repeatability
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Temperature resolution (digital) 0.1 K
Exposure time
Emissivity/ Gain (adjustable
via sensor or software)
IR window correction
(adjustable via software)
Signal processing (parameter adjustable via software)
Software optris
1)
The functioning of the LCD display may be limited in ambient temperatures
below 0 °C
2)
ε = 1, response time 1 s
3)
Whichever is greater
4)
With dynamic adaptation at low signal levels
2)
4)
(90 % signal) 10 ms (G5HF)
100 °C ... 1200 °C (212 °F ... 2192 °F) (G5L)
250 °C ... 1650 °C (482 °F ... 2912 °F) (G5H)
200 °C ... 1650 °C (392 °F ... 2912 °F) (G5HF)
45:1 (G5L, G5HF, G5H1F)
70:1 (G5H)
±1 % or ±1.5 °C
(
±1 % or ±2.7 °F3))
±0.5 % or +0.5 °C
(
±0.5 % or +1.8 °F
80 ms (G5H)
120 ms (G5L)
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
3)
3) 4)
3) 4)
)
Power supply 8 – 36 V DC
Current draw max. 160 mA
Laser 635 nm 1 mW, ON/OFF via electronic box or software
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
Chart SF optics, D:S = 45:1
®
CTlaser G5
Additional optics, D:S = 45:1
20
20.8
21.7
22.5
(0.85)
300
(11.8)
(0.89)
450
(17.7)
23.4
(0.92)
600
(23.6)
S
(0.79)
(0.82)
D
0
150
(0)
(5.91)
Chart SF optics, D:S = 70:1
19
S
D 0
20
(0.79)
(0)
19.6
(0.77)
150
(5.91)
19.3
(0.76)
300
(11.8)
(0.75)
450
(17.7)
18.5
(0.73)
600
(23.6)
24.2
(0.95)
750
(29.5)
18.2
(0.72)
(29.5)
... CF1 1.6 mm @ 70 mm
(0.06 in @ 2.76 in)
... CF2 3.4 mm @ 150 mm
(0.13 in @ 5.91 in)
... CF3 4.5 mm @ 200 mm
(0.18 in @ 7.87 in)
... CF4 10 mm @ 450 mm
(0.39 in @ 17.7 in)
1800
(70.9)
61.7
(2.43)
2100
(82.7)
73.4
(2.89)
2400
(94.5)
mm
(in)
mm
(in)
25
(0.98)
900
(35.4)
25.9
(1.02)
1050
(41.3)
27
(1.06)
1200
(47.2)
32.5
38.4
(1.28)
(1.51)50(1.97)
1350
1500
(53.2)
(59.1)
Additional optics, D:S = 70:1
... CF1 1 mm @ 70 mm
(0.04 in @ 2.76 in)
... CF2 2.2 mm @ 150 mm
(0.09 in @ 5.91 in)
... CF3 2.9 mm @ 200 mm
(0.11 in @ 7.87 in)
... CF4 6.5 mm @ 450 mm
(0.26 in @ 17.7 in)
17.8
17.4
17
(0.69)
1050
(41.3)
(0.67)
1200
(47.2)
21.6
(0.85)
1350
(53.2)
(0.7)
900
750
(35.4)
26.3
(1.04)
1500
(59.1)
35.5
(1.40)
1800
(70.9)
44.8
(1.76)54(2.13)
2100
(82.7)
2400
(94.5)
mm
(in)
mm
(in)
Dimensions
Sensing head
Accessories (examples)
Mounting angle, adjustable
in two axes (ACCTLAB)
Electronics
Water cooling and air purge for
sensing head (ACCTLW + ACCTLAP)
Mounting device for cooling housing
(ACCTLRM)
Specications are subject to change without notice · CTlaser G5-US2018-08-A