optris
®
CTlaser 1M/2M
TECHNICAL DATA
Precise non-contact temperature measurement
with precise aiming from
250 °C to 2200 °C
(482 °F to 3992 °F)
Features:
• Accurate temperature measurements of metals, secondary metal processing and
ceramic materials
• Double laser aiming marks real spot location at any distance
• Optical resolution up to 300:1 with selectable focus
• Temperature ranges from 250 °C to 2200 °C (482 °F to 3992 °F), measuring spots
up from 0,45 mm (0.02 in) and response times up from 1 ms
• Usable up to 85 °C (185 °F) ambient temperature without cooling and automatic
laser switch o at 50 °C (122 °F)
• Short measuring wavelength of 1.0 μm or 1.6 μm
General specications
Environmental rating IP 65 (NEMA-4) front mountable
Ambient temperature
Storage temperature – 40 °C ... 125 °C (–4 °F ... 257 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
1)
at vacuum processes (up to 10
– 20 °C ... 85 °C (–4 °F ... 185 °F) (sensing head)
(50 °C [122 °F] with laser ON)
– 20 °C ... 85 °C (–4 °F ... 185 °F) (electronics)
– 40 °C ... 85 °C (–40 °F ... 185 °F) (electronics)
420 g (14.8 oz) (electronics)
–3
mbar)
Electrical specications
Output / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Output / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 3 m (standard), 8 m, 15 m
0 / 4
–
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
(9.8 ft [standard], 26.2 ft, 49.2 ft)
eff
Measurement specications
Temperature range
(scalable via programming keys or
software)
Spectral range 1.0 μm (1M) / 1.6 μm (2M)
Optical resolution
(90 % energy)
System accuracy
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Repeatability
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Temperature resolution 0.1 K (1ML, 2ML)
Exposure time
Emissivity/ Gain (adjustable
via sensor or software)
IR window correction
(adjustable via software)
Signal processing (parameter adjus-
table via software)
Software optris
1)
The functioning of the LCD Display may be limited in ambient temperatures
below 0 °C
2)
ε = 1, Exposure time 1 s
3)
With dynamic adaptation at low signal levels
2)
3)
485 °C ... 1050 °C (905 °F ... 1841 °F) (1ML)
650 °C ... 1800 °C (1202 °F ... 3272 °F) (1MH)
800 °C ... 2200 °C (1472 °F ... 3992 °F) (1MH1)
250 °C ... 800 °C (482 °F ... 1472 °F) (2ML)
385 °C ... 1600 °C (725 °F ... 2912 °F) (2MH)
490 °C ... 2000 °C (914 °F ... 3632 °F) (2MH1)
150:1 (1ML, 2ML)
300:1 (1MH, 1MH1, 2MH, 2MH1)
±(0.3 % of reading +2 °C)
(±[0.3 % of reading +3.6 °F])
±(0.1 % of reading +1 °C)
(±[0.1 % of reading +1.8 °F])
0.1 K (1MH, 1MH1, 2MH, 2MH1)
1 ms (90 %)
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
Current draw (laser) Max. 100 mA
Power supply 8 – 36 V DC
Laser 635 nm 1 mW, ON/OFF via electronic box or software
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
12
S
(0.47)
1MH/2MH SF
300:1
3.7 mm @ 1100 mm
(0.15 in @ 43.3 in)
®
CTlaser 1M/2M
9.8
7.5
5.2
3.7
(0.39)
(0.30)
(0.20)
(0.15)
7.3
(0.29)
9.4
(0.37)
13
(0.51)
16.6
(0.65)
mm
(in)
1MH/2MH CF4
300:1
1.5 mm @ 450 mm
12
9.7
(0.38)
7.4
(0.30)5 (0.20)
(0.47)
S
2.7
(0.11)3 (0.12)6 (0.24)9 (0.35)
12
(0.47)
mm
(in)
7.3 mm @ 1100 mm
(0.29 in @ 43.3 in)
1ML/2ML SF
150:1
1MH/2MH CF2
300:1
0.5 mm @ 150 mm
(0.02 in @ 5.91 in)
1 mm @ 550 mm
(0.04 in @ 21.7 in)
1ML/2ML CF2
150:1
1MH/2MH CF3
300:1
0.7 mm @ 200 mm
20
16.5
13
9.6
7.3
13.5
17.3
23.5
S
D 0
S
(0.79)
(0)
12
(0.47)
4.4
(0.17)
(0.65)
300
(11.8)
4.5
(0.18)
(0.51)
600
(23.6)
13
(0.51)
(0.38)
900
(35.4)
21.4
(0.84)
(0.29)
1100
(43.4)
30
(1.18)
(0.53)
1350
(53.1)
(0.68)
1500
(59.0)
(1.50)
38
(0.93)
1750
(68.9)
46.4
(1.83)
30
(1.18)
2000
(78.7)
54.8
(2.16)
20
S
D 0
S
(
0.79
(0)
12
(
0.47
7.3
)
(0.29)8 (0.31)
100
(3.94)
6.4
)
(0.25)
200
(7.87)
(0.03)
22
36
50
64
58
(0.87)
(1.42)
(1.97)
(2.52)
300
400
500
(11.8)
(15.7)
0.7
7.1
13.4
(0.28)
(0.53)
(19.7)
19.8
(0.78)
600
(23.6)
26.1
(1.0)
(2.28)
700
(27.6)
32.5
(1.03)
92
(3.62)
800
(31.5)
38.8
(1.53)
mm
(in)
mm
(in)
mm
(in)
mm
(in)
mm
(in)
mm
(in)
3 mm @ 450 mm
1ML/2ML CF4
150:1
1MH/2MH FF
300:1
12 mm @ 3600 mm
24 mm @ 3600 mm
1ML/2ML FF
150:1
S
D 0
S
S
D 0
20
(0.79)
(0)
12
(0.47)
20
(0.79)
(0)
16.3
(0.64)
100
(3.94)
12
(
0.47
(0.83)
(35.4)
12.5
8.7
4.9
5.6
10.7
15.8
21
700
28
53.4
(2.10)
6000
(0.83)
800
(31.5)
33
(1.30)
(2.46)
(264.6)
62.5
6730
mm
(in)
mm
(in)
mm
(in)
mm
(in)
mm
(in)
(0.49)
(0.34)
(0.19)
(0.22)
(0.42)
(0.62)
200
300
400
500
(7.87)
(11.8)
(15.7)
12
(
0.47
(0.87)
1800
(70.8)
12
)
(
0.47
22
23
(0.91)
2700
(106.3)
)
21
900
12
)
(
0.47
)
24
(0.94)
3600
(141.7)
(19.7)
600
(23.6)
21.3
(0.84)
(1.61)
(196.8)
41
5000
(27.6)
(1.10)
(236.2)
1.3 mm @ 200 mm
1ML/2ML CF3
150:1
20
10.7
1.3
12
22.6
33.3
44
55
65
S
D 0
(0.79)
(0)
(0.42)
100
(3.94)
(0.05)
200
(7.87)
(0.47)
300
(11.8)
(0.89)
400
(15.7)
(1.31)
500
(19.7)
(1.73)
600
(23.6)
(2.17)
700
(27.6)
(2.56)
800
(31.5)
mm
(in)
mm
(in)
Dimensions
Sensing head Electronics
Specications are subject to change without notice · CTlaser 1M 2M-US2018-08-A