Precise non-contact
temperature measurement
of liquid metals from
1000 °C to 2000 °C
(1832 °F to 3632 °F)
Features:
• Precise temperature measurement of molten metal
• Decrease of measurement errors at emissivity changes or wrong settings thanks to
short wavelength of 525 nm
• Temperature range from 1000 °C to 2000 °C (1832 °F to 3632 °F), measuring
elds from 1 mm (0.04 in) and exposure times from 1 ms
• Suited for ambient temperatures of 85 °C (185 °F) without additional cooling,
laser switches of automatically at 50 °C (122 °F)
• Optical resolution 150:1, focus settings scalable
• Double laser visor with 2 laser beams for exact measuring eld marking and
focussing
General specications
Environmental rating IP 65 (NEMA-4) front mountable
Ambient temperature
Storage temperature – 40 °C ... 125 °C (–4 °F ... 257 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.2 oz) (sensing head)
1)
at vacuum processes (up to 10
– 20 °C ... 85 °C (–4 °F ... 185 °F) (sensing head)
(50 °C [122 °F] with laser ON)
– 20 °C ... 85 °C (–4 °F ... 185 °F) (electronics)
– 40 °C ... 85 °C (–40 °F ... 185 °F) (electronics)
420 g (14.8 oz) (electronics)
–3
mbar)
Electrical specications
Output / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Output / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Inputs Programmable functional inputs for external
0 / 4
–
20 mA, 0
– 5
/ 10 V, thermocouple J, K
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
eff
Measurement specications
Temperature range
(scalable via programming keys or
software)
Spectral range 525 nm
Optical resolution
(90 % energy)
System accuracy
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Repeatability
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±9 °F)
Temperature resolution (digital) 0.2 K
Exposure time
Emissivity/ Gain (adjustable
via sensor or software)
IR window correction
(adjustable via software)
Signal processing (parameter adjustable via software)
Software optris
1)
The functioning of the LCD display may be limited in ambient temperatures
below 0 °C
2)
ε = 1, Exposure time 1 s
3)
With dynamic adaptation at low signal levels
2)
3)
(90 % signal) 1 ms
1000 °C ... 2000 °C
(1832 °F ... 3632 °F)
150:1 (3ML)
±1 % of reading (≤1100 °C)
±(0.3 % of reading +2 °F) (>1100 °C)
±1 % of reading (≤2012 °F)
±(0.3 % of reading +3.6 °F) (>2012 °F)
±0.5 % of reading (≤1100 °C)
±(0.1 % of reading +1 °C) (>1100 °C)
±0.5 % of reading (≤2012 °F)
±(0.1 % of reading +1.9 °F) (>2012 °F)
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
Cable length 3 m (standard), 8 m, 15 m
Current draw (laser) Max. 160 mA
Power supply 8 – 36 V DC
Laser 635 nm 1 mW, ON/OFF via electronic box or software
(9.8 ft [standard], 26.2 ft, 49.2 ft)
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
®
CTlaser 05M
7.3 mm @ 1100 mm
(0.29 in @ 43.3 in)
05M SF
150:1
S
D 0
20
(0.79)
(0)
16.5
(0.65)
300
(11.8)
13
(0.51)
600
(23.6)
9.6
(0.38)
900
(35.4)
7.3
(0.29)
1100
(43.3)
13.5
(0.53)
1350
(53.1)
17.3
(0.68)
1500
(59.0)
23.5
(0.93)
1750
(68.9)
30
(1.18)
2000
(78.7)
mm
(in)
mm
(in)
Dimensions
Sensing head Electronics
24 mm @ 3600 mm
(0.94 in @ 142 in)
05M FF
150:1
S
D 0
20
(0.79)
(0)
21
(0.83)
900
(35.4)
22
(0.87)
1800
(70.8)
23
(0.91)
2700
(106)
24
(0.94)
3600
(142)
29
41
53.4
62.5
(1.14)
(1.61)
(2.10)
4000
5000
(158)
(197)
6000
(236)
(2.46)
6730
(265)
mm
(in)
mm
(in)
Accessories
Mounting angle, adjustable
in two axes (ACCTLAB)
Water cooling and air purge for
sensing head (ACCTLW + ACCTLAP)
Cooling housing
(ACCJCTL)
Mounting angle for cooling housing,
adjustable in two axes (ACCJAB)
Mounting device for cooling housing
(ACCTLRM)
Specications are subject to change without notice · CTlaser 05M-US2018-08-A