optris
®
CT G5
TECHNICAL DATA
Precise noncontact
temperature measurement of
glass from 100 °C to 1650 °C
(212 °F to 3002 °F)
Features:
• Accurate temperature measurement of at glass, container glass,
light bulb manufacturing, car glass production and manufacturing of
photovoltaic cells from 100 °C to 1650 °C (212 °F to 3002 °F)
• Applicable up to 85 °C (176 °F) ambient temperature without additional cooling
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature –20 °C ... 85 °C (–4 °F ... 176 °F) (sensing head)
Storage temperature – 40 °C ... 85 °C (–40 °F ... 176 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 42 g (1.5 oz) (sensing head)
0 °C ... 85 °C (32 °F ... 176 °F) (electronics)
– 40 °C ... 85 °C (–40 °F ... 176 °F) (electronics)
420 g (14.8 oz) (electronics)
Electrical Specications
Outputs / analog
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Channel 1: 0 / 4 – 20 mA, 0 – 5/ 10 V, thermocouple J, K
Channel 2: sensing head temperature (–20 °C ...
85 °C [–4 °F ... 176 °F] as 0 – 5 V or 0 – 10 V),
alarm output
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance, thermocouple 20 Ω
eff
Measurement specications
Temperature range
(scalable via programming
keys or software)
Spectral range 5 μm
Optical resolution
(90 % energy)
System accuracy (at ambient
temp. 23 ±5 °C) (23 ±41 °F)
Repeatability (at ambient temp.
23 ±5 °C) (23 ±0.5 °F)
Temperature resolution
(NETD)
Response time (90 % signal) 80 ms (G5H)
Emissivity/ Gain
(adjustable via programming
keys or software)
Transmissivity/ Gain
(adjustable via programming
keys or software)
Signal processing (parameter
adjustable via programming
keys or software, respectively)
Software optris
1)
Whichever is greater
100 °C ... 1200 °C (212 °F ... 2192 °F) (G5L)
250 °C ... 1650 °C (482 °F ... 3002 °F) (G5H)
10:1 (G5L)
20:1 (G5H)
±1 % or ±2 °C
(±1 % or ±3.6 °F1))
±0.5 % or ±0.5 °C
(±0.5 % or ±0.9 °F1))
0.1 K / 0.2 K (G5H)
120 ms (G5L)
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
1)
1)
Inputs Programmable functional inputs for external
Cable length 3 m (standard), 8 m, 15 m
Power Supply 8 – 36 V DC
Current draw Max. 100 mA
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
(9.8 ft [standard], 26.2 ft, 49.2 ft)
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-Mail: sales@optris-ir.com · www.optris.com

optris
®
CT G5
Optical specications
10:1 optics 20:1 optics
6.5
14.9
23.3
31.6
40
51.6
63.3
74.9
86.5
S
D 0
(0.26)
(0)
(0.59)
100
(3.94)
(0.93)
200
(7.87)
(1.24)
300
(11.8)
(1.57)
400
(15.7)
(2.03)
500
(19.7)
(2.37)
600
(23.6)
(2.95)
700
(27.6)
(3.41)
800
(31.5)
mm
(in)
mm
(in)
Dimensions
S
D 0
6.5
(0.26)
(0)
9.9
(0.39)
100
(3.94)
13.3
(0.52)
200
(7.87)
16.6
(0.65)
300
(11.8)
20
(0.79)
400
(15.7)
26.6
(1.05)
500
(19.7)
33.3
(1.31)
600
(23.6)
39.9
(1.57)
700
(27.6)
46.5
(1.83)
800
(31.5)
mm
(in)
mm
(in)
Sensing head
Compact Connect software
Electronics
• Software for easy sensor setup and remote controlling, supports
multi tasking
• Graphic display for temperature trends and automatic data logging for
analysis and documentation with 1 ms response time
• Adjustment of signal processing functions and programming of outputs
and functional inputs of the sensor
• Automatic emissivity adjustment
• The software CompactConnect allows to customize the sensor to
application needs of the user
Specications are subject to change without notice · CT G5-US2017-10-A