Optris CTex Specifications

optris
®
CT LT
TECHNICAL DATA
Precise non-contact temperature measurement from –50 °C to 975 °C (–58 to 1787 °F)
Features:
• One of the smallest infrared sensors worldwide with 22:1 optical resolution
• Rugged and usable up to 180 °C (356 °F) ambient temperature without cooling
Separate electronics with easy accessible programming keys and LCD backlit display
Selectable analog output: 0/4 – 20 mA, 0 – 5 V, 0 – 10 V, thermocouple type K or J
Optional USB, RS485, RS232 interface, relay outputs (2 x optically isolated),
CAN-Bus, Probus DP, Ethernet
Installation of up to 32 sensing heads
CTex: Explosion proof version (ATEX)
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature –20 °C to 180 °C (–4 °F to 356 °F)
Storage temperature – 40 °C to 130 °C (–40 °F to 266 °F) (sensing head)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 40 g (1.4 oz) (sensing head) /
(130 °C (266 °F) to LT02) (sensing head) 0 °C to 85 °C (32 °F to 185 °F) (electronics)
– 40 °C to 85 °C (–40 °F to 185 °F) (electronics)
420 g (14.8 oz) (electronics)
Electrical Specications
Outputs / analog Channel 1: 0 / 4 – 20 mA, 0 – 5 / 10 V, thermocouple J, K
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
Inputs Programmable functional inputs for external
Cable length 1 m (standard), 3 m, 8 m, 15 m
Channel
2: sensind head temperature (–20 °C to 180 °C (–4 °F to 356 °F) as 0 – 5 V or 0 – 10 V), alarm output
; 0.4 A;
optically isolated
(optional)
mV min. 100 kΩ load impedance thermocouple 20 Ω
emissivity adjustment, ambient temperature compensation, trigger (reset of hold functions)
(3.3 ft (standard), 9.8 ft, 26.2 ft, 49.2 ft)
eff
Measurement specications
Temperature range (scalable via programming keys or software)
Spectral range 8 – 14 μm
Optical resolution (90 % energy)
CF-lens (optional) 0.6 mm @ 10 mm (with LT22) (0.02 in @ 0.4 in)
System accuracy ent temp. 23 ±5 °C) (73 ±9 °F)
Repeatability temp. 23 ±5 °C) (73 ±9 °F)
Temperature resolution (display) 0.1 K
2), 3)
NETD
Response time 150 ms (95 %)
Emissivity/ Gain (adjustable via programming keys or software)
Transmissivity/ Gain (adjustable via programming keys or software)
Signal processing (parameter adjustable via programming keys or software, respectively)
Software optris
1)
Whichever is greater
2)
At object temperatures >0 °C (> 32 °F), ε = 1
3)
At time constant 200 ms and T
1), 2)
1), 2)
(at ambient
(at ambi-
–50 °C to 975 °C (LT22) (–58 °F to 1787 °F) –50 °C to 600 °C (LT15) (–58 °F to 1112 °F) –50 °C to 600 °C (LT02) (–58°F to 1112 °F)
22:1 (precision glass optics) 15:1 (precision glass optics) 2:1 (with flat front window)
0.8 mm @ 10 mm (with LT15) (0.03 in @ 0.4 in)
2.5 mm @ 23 mm (with LT02) (0.1 in @ 0.9 in)
±1 % or ±1 °C (±1 % or ±1.8 °F)
±0.5 % or ±0.5 °C (±0.5 % or ±0.9 °F)
0.05 K (LT22 / LT15)
0.1 K (LT02)
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold function with threshold and hysteresis
®
Compact Connect
25 °C (77 °F)
Obj
Power Supply 8 – 36 V DC
Current draw Max. 100 mA
Optris GmbH · Ferdinand-Buisson-Str. 14 · 13127 Berlin · Germany Tel.: +49 (0)30 500 197-0 · Fax: +49 (0)30 500 197-10 · E-mail: info@optris.com · www.optris.com
optris
Optical specications
53.8
(2.12)
(3.94)
11.6
(0,46)
100
(3.94)
100
16.6
(0.65)
200
(7.87)
102.5
(4.04)
200
(7.87)
S5 (0.20)
D 0
(0)
Optics CT LT SF, D:S = 15:1
6.5
S
(0.26)
D 0
(0)
Optics CT LT SF, D:S = 22:1
®
CT LT
151.3 (5.96)
300
(11.8)
21.7
26.7
(0.85)
(1.05)
300
400
(11.8)
(15.7)
200
(7.87)
400
(15.6)
35
(1.38)
500
(19.7)
251.3 (9.89)
500
(19.7)
43.3
(1.70)
600
(23.6)
302.5 (11.9)
600
(23.6)
51.6
(2.04)
700
(27.6)
353.8 (13.9)
700
(27.6)
59.9
(2.36)
800
(31.5)
405
(15.9)
800
(31.5)
(2.69)
(35.4)
68.2
900
mm (in)
mm (in)
Optics CT LT CF, D:S = 2:1 (far eld = 2.5:1)Optics CT LT SF, D:S = 2:1
S7 (0.28)5 (0.20)
D 0
(0)
Optics CT LT CF, D:S = 15:1 (far eld = 1.5:1)
76.5
(3.01)
1000
(39.4)
(mm)
(mm)
S7 (0.28)4 (0.16)
D 0
(0.0)5 (0.20)
Optics CT LT CF, D:S = 22:1 (far eld = 1.5:1)
10
(0.39)
0.8
(0.03)
10
(0.39)
3.1
(0.12)
20
(0.79)
(0.10)
(0.91)
4.7
(0.19)
15
(0.59)
2.5
5.4
9.5
13.7
17.8
20.3
(0.80)
35
(1.38)
(0.70)
60
(2.36)
24.2
(0.95)
40
(1.57)
mm (in)
mm (in)
mm (in)
mm (in)
(0.21)
(0.37)
(0.54)
23
30
40
(1.18)
(0.34)
(0.79)
(1.57)
8.6
12.5
(0.49)
20
(0.98)
50
(1.97)
16.4
(0.65)
25
30
(1.18)
6.5
10.9
15.2
19.6
S
D 0
(0.26)
(0)
(0.43)
110
(4.33)
(15.2)
220
(8.66)
(0.77)
330
(13.0)
23.9
(0.94)
440
(17.3)
Versions with built-in CF lenses
6.5
S
(0.26)5 (0.20)4 (0.16)
D 0
(0)
25
(0.98)
40
(1.57)
3.4
(0.13)
50
(1.97)
5.4
(0.21)
60
(2.36)
8.4
(0.33)
75
(2.95)
13.3
(0.52)
100
(3.94)
Dimensions
Sensing head (standard)
Sensing head (built-in CF lenses)
28.3
(1.11)
550
(21.7)
18.3
(0.72)
125
(4.92)
32.6
(1.28)
660
(26.0)
23.2
(0.92)
150
(5.91)
37
(1.46)
770
(30.3)
28.2
(1.11)
175
(6.89)
41.3
(1.63)
880
(34.6)
33.1
(1.30)
200
(7.87)
45.7
(1.80)
990
(39.0)
mm (in)
mm (in)
50
(1.97)
1100
(43.3)
mm (in)
mm (in)
Optics CT LT CF, D:S = 22:1 (far eld = 1.5:1)Optics CT LT CF, D:S = 22:1
Electronics
S
(0.26)
D 0
S7 (0.28)
D 0
(0)5 (0.20)
6.5
(0)
4.4
(0.17)
25
(0.98)
3.8
(0.15)
3.2
(0.13)
40
(1.57)
0.6
(0.02)
(0.39)
2.3
(0.09)
50
(1.97)
4.4
8.2
12
15.8
19.6
23.4
(0.17)
(0.32)
(0.47)
(0.62)
(0.77)
10
15
20
25
(0.98)
15.5
(0.61)
125
(4.92)
30
(1.18)
19.9
(0.78)
150
(5.91)
(1.38)
4.1
(0.16)
60
(2.36)
(0.59)
(0.26)
(2.95)
(0.79)
6.7
11.1
(0.44)
75
100
(3.94)
mm
(0.92)
(in)
mm
35
40
(in)
(1.57)
24.3
28.7
(1.13)
200
(7.87)
mm (in)
mm (in)
(0.96)
175
(6.89)
Specications are subject to change without notice · CT LT-E2018-08-A