optris
®
CT 1M/2M
TECHNICAL DATA
Precise non-contact
temperature measurement
from 250 °C to 2200 °C
(483 °F to 3992 °F)
Features:
• Miniaturized Infrared Thermometer with 1.0 μm respectively 1.6 μm wave length
range for measurements of metals, of secondary metal processing, metal oxides
and ceramic materials
• Very small sensing head of 14 mm (0.6 in) diameter and 28 mm (1.1 in) length
ts everywhere and is usable up to 125 °C (257 °F) ambient temperature
without cooling
• Temperature ranges from 250 °C to 2200 °C (483 °F to 3992 °F), measuring
spots up from 1.8 mm (0.07 in) and exposure times up from 1 ms
• Short measuring wave length of 1.0 μm respectively 1.6 μm reduces error
of temperature readings on surfaces with low or unknown emissivity
General specications
Environmental rating IP 65 (NEMA-4)
Ambient temperature –20 °C ... 100 °C (1M) to 125 °C (2M)
Storage temperature – 40 °C ... 100 °C (1M) to 125 °C (2M)
Relative humidity 10 – 95 %, non condensing
Vibration (sensor) IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock (sensor) IEC 68-2-27: 50 G, 11 ms, any axis
Weight 40 g (1.4 oz) (sensing head) /
(–4 °F ... 212 °F [1M] to 257 °F [2M]) (sensing head)
0 °C ... 85 °C (32 °F ... 185 °F) (electronics)
(–40 °F ... 212 °F [1M] to 257 °F [2M]) (sensing head)
– 40 °C to 85 °C (–40 °F to 185 °F) (electronics)
420 g (14.8 oz) (electronics)
Electrical Specications
Outputs / analog 0 / 4 – 20 mA, 0 – 5 / 10 V, thermocouple J, K, alarm
Output / alarm 24 V / 50 mA (open collector)
Optional Relay: 2 x 60 V DC/ 42 V AC
optically isolated
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet
(optional)
Output impedances mA max. 500 Ω (with 8 – 36 V DC)
mV min. 100 kΩ load impedance thermocouple 20 Ω
Inputs Programmable functional inputs for external
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)
Cable length 3 m (standard), 8 m, 15 m
(9.8 ft [standard], 26.2 ft, 49.2 ft)
; 0.4 A;
eff
Measurement specications
Temperature range
(scalable via programming
keys or software)
Spectral ranges 1.0 μm (1M)/ 1.6 μm (2M)
Optical resolution CT
1ML / 2ML (90 % energy)
Optical resolution CT
1MH / 1MH1 / 2MH / 2MH1
(90 % energy)
System accuracy
temp. 23 ±5 °C) (73 ±9 °F)
Repeatability (at ambient temp.
23 ±5 °C) (73 ±9 °F)
Temperature resolution 0.1 K
Exposure time
Emissivity/ Gain
(adjustable via programming
keys or software)
Transmissivity/ Gain
(adjustable via programming
keys or software)
Signal processing (parameter
adjustable via programming
keys or software, respectively)
Software optris
1)
ε = 1, Exposure time 1 s
2)
With dynamic adaptation at low signal levels
1)
(at ambient
2)
485 °C ... 1050 °C (905 °F ... 1929 °F) (1ML)
650 °C ... 1800 °C (1202 °F ... 3272 °F) (1MH)
800 °C ... 2200 °C (1472 °F ... 3992 °F) (1MH1)
250 °C ... 800 °C (482 °F ... 1472 °F) (2ML)
385 °C ... 1600 °C (725 °F ... 2912 °F) (2MH)
490 °C ... 2000 °C (914 °F ... 3632 °F) (2MH1)
40:1 (2.7 mm @ 110 mm)
75:1 (1.5 mm @ 110 mm)
±(0.3 % of reading +2 °C)
(±[0.3 % of reading +3.6 °F])
±(0.1 % of reading +1 °C)
(±[0.1 % of reading +1.8 °F])
1 ms (90 %)
0.100 – 1.100
0.100 – 1.100
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
Power Supply 8 – 36 V DC
Current draw Max. 100 mA
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone.: 603-766-6060 · E-Mail: sales@optris-ir.com · www.optris.com
optris
®
CT 1M/2M
Optical specications
CT 1ML/2ML SF, D:S = 40:1 CT 1ML/2ML CF, D:S = 40:1 (far eld = 12:1)
6.5
10.7
14.9
19.1
23.3
27.4
31.6
35.8
40
45.8
51.6
S
D 0
(0.26)
(0)
(0.42)
200
(7.9)
(0.59)
400
(15.7)
(0.75)
600
(23.6)
(0.92)
800
(31.5)
(1.08)
1000
(39.4)
(1.24)
1200
(47.2)
(1.41)
1400
(55.1)
(1.57)
1600
63.0)
(1.80)
1800
(70.9)
(2.03)
2000
(78.7)
mm
(in)
mm
(in)
6.5
4.5
S
D 0
(0.26)
(0)
(0.18)
60
(2.36)
2.7
(0.11)6 (0.24)
110
150
(4.33)
(5.91)
CT 1MH / 1MH1 / 2MH / 2MH1 CF, D:S = 75:1 (far eld = 14:1)
10.3
(0.41)
200
(7.87)
14.5
(0.57)
250
(9.84)
18.6
(0.73)
300
(11.8)
22.8
(0.90)
350
(13.8)
27
(1.06)
400
(15.7)
mm
(in)
mm
(in)
6.5
8.4
10.2
S
D 0
(0.26)
(0)
(0.33)
200
(7.9)
(0.40)
400
(15.7)
12.5
(0.49)
600
(23.6)
Dimensions
Sensing head
(standard)
Accessories (examples)
13.9
(0.55)
800
(31.5)
15.8
(0.62)
1000
(39.4)
17.6
(0.69)
1200
(47.2)
19.5
(0.77)
1400
(55.1)
21.3
(0.84)
1600
(63.0)
24.8
(0.98)
1800
(70.9)
28.3
(1.11)
2000
(78.7)
mm
(in)
mm
(in)
6.5
S
(0.26)
D 0
(0)
Electronics
3.8
(0.15)
60
(2.36)
1.5
(0.06)
110
(4.33)
4.4
(0.17)
150
(5.91)
8.1
(0.32)
200
(7.87)
11.7
(0.46)
250
(9.84)
15.4
(0.61)
300
(11.8)
19
(0.75)
350
(13.8)
22.6
(0.89)
400
(15.7)
mm
(in)
mm
(in)
CF-lens
(ACCTCFHT)
Mounting bracket, xed
(ACCTFB)
Air purge collar with integrated
CF-lens (ACCTAPLCFHT)
Specications are subject to change without notice · CT 1M 2M-US2018-08-A