Optris CSmicro 2W LT, CSmicro 3M, CSmicro LT, CSmicro LT HS, CX Operating Manual

optris
®
CSmicro
Infrared-thermometer
LT/ LTH/ LTHS/ 2M/ 3M
Operator’s Manual
Optris GmbH
Ferdinand-Buisson-Str. 14 13127 Berlin Germany
Tel.: +49 30 500 197-0 Fax: +49 30 500 197-10
E-mail: info@optris.global Internet: www.optris.global
-Table of Contents 3-
Table of Contents
Table of Contents ............................................................................................................................................. 3
1 General Information ................................................................................................................................. 7
1.1 Description ....................................................................................................................................... 7
1.2 Warranty ........................................................................................................................................... 8
1.3 Scope of Supply ............................................................................................................................... 8
1.4 Maintenance ..................................................................................................................................... 9
1.5 Model Overview .............................................................................................................................. 10
1.6 Factory Default Settings ................................................................................................................. 11
2 Technical Data ........................................................................................................................................ 15
2.1 General Specifications ................................................................................................................... 15
2.2 Electrical Specifications .................................................................................................................. 17
2.3 Pin Configuration ............................................................................................................................ 19
2.4 Measurement Specifications .......................................................................................................... 20
-4 -
2.5 Optical Charts ................................................................................................................................. 23
2.6 CF Lens and Protective Window .................................................................................................... 33
3 LED Functions ........................................................................................................................................ 35
3.1 Automatic Aiming Support .............................................................................................................. 35
3.2 Self Diagnostic ............................................................................................................................... 36
3.3 Temperature Code Indication ......................................................................................................... 37
4 Mechanical Installation .......................................................................................................................... 38
4.1 Mounting Accessories [LT/ 2M/ 3M] ............................................................................................... 39
4.2 Mounting Accessories [LT15HS] .................................................................................................... 40
4.3 Air Purge Collars [LT/ 2M/ 3M] ....................................................................................................... 41
4.4 Air Purge Collar [LT15HS] .............................................................................................................. 42
4.5 Further Accessories ....................................................................................................................... 43
5 Electrical Installation ............................................................................................................................. 45
5.1 Analog Mode .................................................................................................................................. 45
-Table of Contents 5-
5.2 Maximum Loop Impedance [CSMA models] .................................................................................. 48
5.3 Digital Mode ................................................................................................................................... 49
5.4 Alarm Output .................................................................................................................................. 51
6 IRmobile app ........................................................................................................................................... 52
7 Software CompactConnect ................................................................................................................... 54
7.1 Installation ...................................................................................................................................... 54
7.2 Communication Settings ................................................................................................................ 55
8 Basics of Infrared Thermometry........................................................................................................... 56
9 Emissivity ............................................................................................................................................... 57
9.1 Definition......................................................................................................................................... 57
9.2 Determination of unknown Emissivity ............................................................................................ 57
9.3 Characteristic Emissivity ................................................................................................................ 58
Appendix A – Emissivity Table Metals ......................................................................................................... 59
Appendix B – Emissivity Table Non Metals ................................................................................................. 61
-6 -
Appendix C – Smart Averaging ..................................................................................................................... 62
Appendix D – Direct Connection to an RS232 Interface ............................................................................. 63
Appendix E – Declarations of Conformity.................................................................................................... 64
-General Information 7-
The CSmicro sensing head is a sensitive optical system. Please use only the thread for mechanical installation.
Avoid abrupt changes of the ambient temperature.  Avoid mechanical violence on the head – this may destroy the system (expiry of warranty).
If you have any problems or questions, please contact our service department.
Read the manual carefully before the initial start-up. The producer reserves the right to change the herein described specifications in case of technical advance of the product.
1 General Information
1.1 Description
Thank you for choosing the optris® CSmicro infrared thermometer. The sensors of the optris CSmicro series are noncontact infrared thermometers. They calculate the surface temperature based on the emitted infrared energy of objects [►8 Basics of Infrared Thermometry]. The sensor housing of the optris CSmicro is made of stainless steel (IP65/ NEMA-4 rating) – the sensor electronics is integrated inside the connection cable.
-8 -
► All accessories can be ordered according to the referred part numbers in brackets [ ].
1.2 Warranty
Each single product passes through a quality process. Nevertheless, if failures occur please contact the customer service at once. The warranty period covers 24 months starting on the delivery date. After the warranty is expired the manufacturer guarantees additional 6 months warranty for all repaired or substituted product components. Warranty does not apply to damages, which result from misuse or neglect. The warranty also expires if you open the product. The manufacturer is not liable for consequential damage or in case of a non-intended use of the product. If a failure occurs during the warranty period the product will be replaced, calibrated or repaired without further charges. The freight costs will be paid by the sender. The manufacturer reserves the right to exchange components of the product instead of repairing it. If the failure results from misuse or neglect the user has to pay for the repair. In that case you may ask for a cost estimate beforehand.
1.3 Scope of Supply
CSmicro incl. connection cable Mounting nut Isolated mounting bracket (only included for LTH devices) Operators manual
-General Information 9-
Never use cleaning compounds which contain solvents (neither for the lens nor for the housing).
1.4 Maintenance
Lens cleaning: Blow off loose particles using clean compressed air. The lens surface can be cleaned with a
soft, humid tissue (moistened with water) or a lens cleaner (e.g. Purosol or B+W Lens Cleaner).
-10 -
Series
Model
Measurement range
Spectral response
Output
Optics
Specialty
LT
LT02
-50 to 1030 °C
8-14 µm
0/5-10 V
(CSMV)
or
4-20 mA
(CSMA)
2:1 LT15
15:1 LT15H
T
amb
max. 180 °C
LT22H
22:1
LT HS
-20 to 150 °C
15:1
0,025 K resolution
2M 2ML
250 to 800 °C
1,6 µm 40:1 2MH
385 to 1600 °C
75:1
3M 3ML
50 to 350 °C
2,3 µm 22:1
3MH
100 to 600 °C
33:1
1.5 Model Overview
The sensors of the CSmicro series (CSMV and CSMA) are available in the following versions:
-General Information 11-
CSmicro
LT02
LT15
LT15HS
LT15H
LT22H
Temperature range:
0...350 °C
-20…150 °C
0...500 °C
Output:
0...3,5 V or 4…20 mA
0…5 V or 4…20 mA
Emissivity:
0,950
Transmission:
1,000
Average time:
0,3 s
Smart Averaging:
active
Smart Averaging hysteresis:
2 °C
Ambient temperature source:
internal (head)
Status-LED function:
Self diagnostic
Input (IN/ OUT/ green):
Inactive (mV version) or Communication input (mA version)
Output (OUT/ yellow):
mV output (mV version) or Communication output (mA version)
Vcc adjust:
inactive
Signal processing:
Hold mode: off
Calibration:
Gain 1,000/ Offset 0,0
Failsafe:
inactive
1.6 Factory Default Settings
The units have the following presetting at time of delivery:
-12 -
CSmicro
2ML
2MH
3ML
3MH
Temperature range:
250...800 °C
385...1600 °C
50…350 °C
100...600 °C
Output:
0…10 V or 4...20 mA
Emissivity:
0,950
Transmission:
1,000
Average time:
0,3 s
Smart Averaging:
active
Smart Averaging hysteresis:
2 °C
Ambient temperature source:
internal (head)
Status-LED function:
Self diagnostic
Input (IN/ OUT/ green):
Inactive (mV version) or Communication input (mA version)
Output (OUT/ yellow):
mV output (mV version) or Communication output (mA version)
Vcc adjust:
inactive
Signal processing:
Hold mode: off
Calibration:
Gain 1,000/ Offset 0,0
Failsafe:
inactive
Smart Averaging means a dynamic average adaptation at high signal edges. [Activation via software only]. Appendix C Smart Averaging
-General Information 13-
OUT
At 3-state output the following settings are default:
Pre-alarm difference:
2 °C
No alarm level:
8 V Pre-alarm level:
5 V Alarm level:
0 V Service voltage:
10 V
IN/ OUT:
At Alarm output (open collector) the following settings are default:
Mode:
normally closed
Temp code output:
activated (for values above alarm level)
Range settings:
0 °C = 0 %/ 100 °C = 100 %
Vcc adjust:
If activated the following settings are default:
Output voltage range:
0 - 10 V
Difference mode:
activated
For a usage of the CSmicro LT (mV version) for online maintenance applications (in electrical cabinets e.g.) the following recommend settings are already included in the factory default setting (but not active):
-14 -
Vcc adjust:
Alarm level Alarm value (IN/ OUT pin) Vcc
1 40 °C 11 V 2 45 °C 12 V 3 50 °C 13 V 4 55 °C 14 V 5 60 °C 15 V 6 65 °C 16 V 7 70 °C 17 V 8 75 °C 18 V 9 80 °C 19 V
10 85 °C 20 V
-Technical Data 15-
Environmental rating
IP65 (NEMA-4)
Ambient temperature
Sensing head: see: Measurement Specifications Electronics (inside cable): -20...80 °C [CSMV]
-20...75 °C 1) [CSMA]
Storage temperature
-40…85 °C
Relative humidity
10...95 %, non condensing
Material
Stainless steel
Dimensions
28 mm x 14 mm (head) [LT/ 2M/ 3M] 32 mm x 14 mm (head) [3M CF1] 55 mm x 29,5 mm (head incl. massive housing) [LT15HS] 35 mm x 12 mm (electronics)
Weight
42 g [LT/ 2M/ 3M] 200 g [LT15HS]
Cable length
head – electronics: 0,5 m (standard), 3 m, 6 m [LT/ 2M/ 3M 2)] after electronics: 0,5 m (standard), 3 m, [LT/ 2M/ 3M] head – electronics: 0,5 m [LT15HS] after electronics: 0,5 m (standard), 3 m, 6 m [LT15HS]
Cable diameter
2,8 mm (head – electronics) 4,3 mm (electronics – end of cable)
Vibration
IEC 60068-2-6 / -64
2 Technical Data
2.1 General Specifications
-16 -
Shock
IEC 60068-2-27 (25G and 50G
Pressure resistance (head)
8 bar
Software
optional
1)
for Vcc (supply voltage) 5-12 VDC/ at Vcc > 12 VDC the max. ambient temperature of the electronics is 65 °C
2)
6 m cable length not available for 3M version
-Technical Data 17-
Used Pin
Function
CSMV
CSMA
OUT
IN/ OUT
x
Analog
0-5 V 1) or 0-10 V 2)/ scalable
4-20 mA/ scalable (current loop between Power and GND pin)
x
Alarm
output voltage adjustable; N/O or N/C
output current adjustable; N/O or N/C (current loop between Power and GND pin)
x
Alarm
3-state alarm output (three voltage level for no alarm, pre-alarm, alarm)
-
x
Alarm
programmable open collector output (NPN type) [0-30 V DC/ 50 mA] 4)
programmable open collector (NPN type) [0-30 V DC/ 500 mA]
x
Temp. Code
Temp. Code Output (open collector (NPN type)) [0-30 V DC/ 50 mA] 4)
Temp. Code Output (open collector (NPN type)) [0-30 V DC/ 500 mA]
x
Input
programmable functions:
external emissivity adjustment ambient temperature compensation triggered signal output and peak hold
function
5)
Reset of hold function 6)
programmable functions: triggered signal output and peak hold
function
5)
Reset of hold function 7)
x
x
Serial digital 3)
uni- (burst mode) or bidirectional
uni- (burst mode) or bidirectional Output impedances
min. 10 kΩ load impedance
max. 1 kΩ loop impedance
2.2 Electrical Specifications
-18 -
Current draw
9 mA
4-20 mA
Power supply
5...30 VDC
5...30 VDC
Status-LED
green LED with programmable functions:
alarm indication (threshold independent from alarm outputs) automatic aiming support self diagnostics
temperature code indication
Vcc adjust mode
10 adjustable emissivity and alarm values by variation of supply voltage/ Service mode for analog output [LT only]
1)
0...4,6 V at supply voltage 5 VDC; also valid for alarm output
2)
only at supply voltage ≥ 11 V
3)
inverted RS232, TTL, 9,6 kBaud
4)
500 mA if the mV output is not used
5)
High level: > 0,8 V/ Low level: < 0,8 V
6)
Reset of peak or valley hold by High level at IN/ OUT pin (Low: open or GND / High: >2,4 V...11 V)
7)
Reset of peak or valley hold by Low level at IN/ OUT pin (Low: GND / High: open or >1 V...11 V)
-Technical Data 19-
white
Power
Power supply
yellow
OUT
Analog output/ TxD/ Alarm output
green
IN/ OUT
Analog input/ RxD/ Open collector output
brown
GND
Ground ()
black
Shield
Shield
white
Power
Current loop (+)/ Alarm output
yellow
OUT
TxD
green
IN/ OUT
Trigger input/ RxD/ Open collector output
brown
GND
Current loop (-)/ Ground ()
black
Shield
Shield
2.3 Pin Configuration
You will find a detailed description of the different sensor connections in chapter 5 Electrical Installation.
-20 -
LT02/ LT15
LT15H/LT22H
LT15HS
Temperature range IR (scalable via software)
-50...1030 °C
-20...150 °C
Ambient temperature (sensing head)
-20...120 °C
-20...180 °C
--20...75 °C
Spectral range
8...14 µm
Optical resolution
2:1/ 15:1
15:1/ 22:1
15:1
CF-lens (optional)
2,5 mm@ 23 mm/ 0,8 mm@ 10 mm
0,8 mm@ 10 mm/ 0,6 mm@ 10 mm
0,8 mm@ 10 mm Accuracy
1),2)
±1,0 °C or ±1,0 %
Repeatability
1),2)
±0,5 °C or ±0,5 %
±0,3 °C or ±0,3 %
Temperature coefficient 3)
±0,05 K/ K or ±0,05 %/ K (whichever is greater)
NETD
50 mK 4)
25 mK
5)
Response time
14 ms (90 % signal)
150 ms (90 % signal)
Warm-up time
10 min
Emissivity/ Gain
0,100...1,100 (adjustable via software)
Transmissivity
0,100...1,100 (adjustable via software)
Interface (optional)
USB (programming interface)
Signal processing
Average, Peak hold, Valley hold (adjustable via software)
2.4 Measurement Specifications
1)
at ambient temperature 235 °C, whichever is greater; Epsilon = 1; Response time 1 s
2)
at object temperatures > 23 °C
3)
for ambient temperatures <18 °C and >28 °C
4)
T
= 200 °C, response time 200 ms
Object
5)
T
= 20 °C, response time 150 ms
Object