optris
®
CSlaser hs LT
TECHNICAL DATA
Two-wire IR thermometer
for smallest dierences
in temperature of 0.025 K
Features:
• New performance standard for IR measurement of smallest temperature
dierences of 25 mK
• Single-piece, robust design for easy installation into your process systeme
• Standard two-wire interface for reliable data transfer and easy integration
into a PLC
• Innovative double-laser sighting for exact marking of measurement targets
• Fast sensor parameterisation and real-time measurement via USB
• Dependable in ambient temperatures up to 85 °C (185 °F) without additional
cooling
• Wide power range: 5 – 28 V DC
General specications
Environmental rating IP 65 (NEMA-4) front mountable
at vacuum processes (up to 10
Ambient temperature –20 °C ... 85 °C (–4 °F ... 185 °F)
(50 °C [122 °F] with laser ON)
Storage temperature –40 °C ... 85 °C (–40 °F ... 185 °F)
Relative humidity 10 – 95 %, non condensing
Vibration IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Shock IEC 68-2-27: 50 G, 11 ms, any axis
Weight 600 g (21.7 oz)
–3
mbar)
Electrical specications
Output / analog 4 – 20 mA
Loop resistance Max. 1000 Ω
Output / alarm 0 – 30 V / 500 mA
(open collector)
1)
Measurement specications
Temperature range
(scalable via software)
Spectral range 8 – 14 μm
Optical resolution
(90 % energy)
System accuracy
(at T
= 23 ±5 °C and T
amb
(at T
73 ±9 °F and T
amb
Repeatability
(at T
= 23 ±5 °C and T
amb
(at T
73 ±9 °F and T
amb
Temperature resolution 0.025 K (at T
Response time (90 % signal) 150 ms
Emissivity/ Gain (adjustable
via sensor or software)
IR window correction
(adjustable via software)
Signal processing (parameter adjustable via software)
Software optris
1)
In dependence on supply voltage
2)
Whichever is greater
Obj
>68 °F)
Obj
Obj
>68 °F)
Obj
>20 °C)
>20 °C)
–20 °C ... 150 °C
(–4 °F ... 302 °F)
50:1
±1 % or ±1 °C
(
±1 % or ±34 °F2))
±0.3 % or ±0.3 °C
(±0.3 or ±32.5 °F2))
(0.025 K [at T
0.100 – 1.100
0.100 – 1.000
Peak hold, valley hold, average; extended hold
function with threshold and hysteresis
®
Compact Connect
2)
2)
>20 °C and time constant >0.2 s)
obj
>68 °F and time constant >0.2 s])
obj
Output / digital Uni-/ bidirectional, 9.6 kBaud,
Cable length
(connector version only)
Current draw (laser) 45 mA at 5 V
Power supply 5 – 30 V DC
0/3 V Pegel, USB optional
3 m / 8 m / 15 m
(9.8 ft / 26.2 ft / 49.2 ft)
20 mA at 12 V
12 mA at 24 V
Optris Infrared Sensing, LLC · 200 International Drive, Suite 170 · Portsmouth · NH 03801 USA
Phone: 603-766-6060 · E-mail: sales@optris-ir.com · www.optris.com
optris
Optical parameter
®
CSlaser hs LT
CSlaser hs LT opticsSF optics, D:S = 50:1, 24 mm @ 1200 mm
S
D 0
Connections
Analog mode
of operation
(0.8)
... SF 24 mm @ 1200 mm
(0.9 in @ 47.2 in)
... CF1 1.4 mm @ 70 mm
(0.0 in @ 2.8 in)
20
20.5
21
21.5
22
22.5
23
23.5
24
29.5
35
46
57
68
(0.8)
(0.8)
(0.8)
(0.9)
(0.9)
(0.9)
(0.9)
(0.9)
(1.2)
(1.4)
(1.8)
(2.2)
300
450
600
750
900
1050
1200
1350
1500
1800
150
(11.8)
(17,7
)
(23.6)
(29.5)
(35.4)
(41.3)
(47.2)
(53.1)
(0)
(5.9)
(59.0)
(70.8)
2100
(82.7)
(2.7)
2400
(94.5)
mm
(in)
mm
(in)
... CF2 3 mm @ 150 mm
(0.1 in @ 5.9 in)
... CF3 4 mm @ 200 mm
(0.2 in @ 7.9 in)
... CF4 9 mm @ 450 mm
(0.4 in @ 17.7 in)
Loop +
Laser +
RxD
4 – 20 mA
0 – 30 V DC / 500 mA
5 – 28 V DC
+
–
Loop –
Laser –
Shield
Digital mode
of operation
Dimensions
Dimensions CSlaser LT
Loop +
Laser +
TxD
RxD
Loop –
Laser –
Shield
5 V DC [white]
IN [yellow]
OUT [green]
GND [brown]
Shield [black]
PC
Electrical connections /
emissivityadjustment
(sensor back side)
Specications are subject to change without notice · CSlaser hs LT-US2018-08-A