OPTO TECH CORPORATION TP339U Datasheet

THERMOPILE
Part No : 80-01-00007
Type : TP339U
Microelectronics Business Division 2003.01 1 / 8
Tel: 886-3-5638951 Fax: 886-3-5783696 http://www.opto.com.tw
Since 1983
OPTO TECH
Features
The thermopile sensor consists of a series of 116 thermoelements, forming a sensitive region size of 545μm (diameter). The sensor is hermetically sealed into a TO-18 metal housing, with an optical filter. This standard filter allows measurements to be made in the spectral range above 5
μ
m wavelength. The thermosensor exhibits an almost white noise, comparable to an ohmic resistance. It has a constant signal versus frequency up to its frequency limit, and is directly proportional to incident radiation. The thermopile sensors are featured with an additional temperature reference resistor on the same chip. The standard version of temperature reference resistor is housing connected to ground.
Applications
* Ear thermometers; clinic thermometers
* Infrared thermometers
* Consumer applications: hair dryer, micro-wave oven, air conditioner, refrigerator
* Continuous temperature control of manufacturing
* Security system
* Radiation monitor switch system
* Absorbing measurement for gas analysis
* Thermoelectric converter
* Heat flux flowmeter
THERMOPILE
Part No : 80-01-00007
Type : TP339U
Microelectronics Business Division 2003.01 2 / 8
Tel: 886-3-5638951 Fax: 886-3-5783696 http://www.opto.com.tw
Since 1983
OPTO TECH
Electrical Characteristics
Parameter Condition Min. Typ . Max. Unit
Thermopile
Number of thermojunctions --- 116 ---
Chip size --- 1740*1740 ---
μ
m2
Active region size
Interference layer
---
Diameter
545
---
μ
m
Thickness of substrate
Silicon-substrat e
600 625 650 μm
Resistance of thermopile 25℃ 50 65 80
Sensitivity
With 5-14μm filter
70 85 100 V/W
Detecctivity 1.0*10
8
1.3*108 1.7*108 cm*Hz
1/2
/W
Time constant --- 16 --- ms
Noise voltage 28 32 36 nV/Hz
1/2
NEP 0.28 0.36 0.48 nW/Hz
1/2
Temperature range Operation -20 --- 100
Temperature reference resistor
---
Resistance (2) 25℃ 97 100 103
β
value (2) 0
o
C/50℃ 3950 3970 3990 K
Measured at 1 Hz chopper frequency, within spectral range 5-14μm, using a blackbody radiator
of 500K temperature.
Note : Thermistor should be operated under 1mA.
THERMOPILE
Part No : 80-01-00007
Type : TP339U
Microelectronics Business Division 2003.01 3 / 8
Tel: 886-3-5638951 Fax: 886-3-5783696 http://www.opto.com.tw
Since 1983
OPTO TECH
Thermopile voltage vs. blackbody temperature
Frequency response
Loading...
+ 5 hidden pages