Optima Scale OP-900-LD User Manual

OP-900-LD INDICATOR
USER’S MANUAL
optimascale.com
TABLE OF CONTENTS
Safety Precautions 1
Preparations and Set Up
Power Supply
Specications
Remote Control
2
2
3
4
1
Display and Key Descriptions
Operating Instructions
Calibration
7-8
5-6
4
Indicator Parameter Settings
Helpful Denitions
Connectors
13
Troubleshooting
Q&A
Contact Us
15
16
12
14
9-11
SAFETY PRECAUTIONS
For safe operation of the weighing indicator, please follow these instructions:
● Calibration inspection and maintenance of the indicator are prohibited by non-pro- fessional staff
● Please ensure that the indicator rests on a stable surface
● The indicator is a piece of static sensitive equipment; Please cut off power during electrical connections
● Touching the internal components by hand is prohibited
● DO NOT exceed the rated load limit of the unit
● DO NOT step on the unit
● DO NOT jump on the scale
● DO NOT use this product if any of the components are cracked
● DO NOT use for purposes other then weight taking
● To avoid damaging the battery do not keep charger plugged in once battery is fully charged
● Make sure the weight is not over the Max capacity as it could damage the load cell inside
● Material that has a static electric charge could inuence the weighing. Discharge the static electricity of the samples, if possible. Another solution to the problem is to wipe both sides of the pan and the top of the case with an anti-static agent
Please take anti-static prevention measures
Any accumulated charge on the body of the human operator should be discharged rst
before opening the protective container with ESDS devices inside. The discharge can be accomplished by:
● Putting a hand on a grounded surface or, ideally, by wearing a grounded Anti-static Wrist Strap and an Anti-static Mat
PREPARATION & SET UP
● Plug into a wall outlet to avoid interference with other wirings
● Turn on the indicator while there is no load
● Calibration may be required before weighing when the scale is initially installed or moved from a location
1
FEATURES
● Large 3” Ultra bright LED display
● Multiple weighing units: kg/lb
● Gross/Tare/Zero
● Accumulation weighing
● Power saving mode
● Remote with antenna
● Mounting bracket
Technical Parameters
● Accuracy class: 5000 e
● Resolution - Display: 30,000 ; ADC: 2,000,000
● Zero stability error: TK0 < 0.1μV//K
● Span stability error: TKspn < ± 6 ppm//K
● Sensitivity (internal): 0.3 μV / d
● Input voltage: -30 to +30mV DC
● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm max
● AC power: AC 100-250V (use only the included 9V adapter supplied)
● Operation temperature: -10 °C ~ +40 °C
● Operation humidity: ≤90%RH
● Storage temperature: -40 °C ~ +70 °C (32-104°F)
AC Adapter
The indicator is powered by an adapter, plug the adapter directly into the AC110V outlet located at the bottom of the indicator. We recommend to plug into a wall outlet to avoid interference with other wirings.
2
OP-900-LD
SPECIFICATIONS
FIGURE 1: INDICATOR MEASUREMENTS
20.5 in.
NET STABLE ZERO
3.25 in.
lb kg
15.25 in.
6 in.
23 in.
4.75 in.
8.5 in.
2.5 in.
3
REMOTE CONTROL
Button Key Function Sign
1 ON/OFF
2 Total
3 Unit
4 Tare
5 Zero
6 Enter/Set
DISPLAY AND KEY DESCRIPTION
Net Shows you are in Net weight mode (weight without tared weight) Stable Shows the weight on the scale is stable Zero Shows you have zeroed the scale
lb The weight is shown in pounds kg The weight is shown in kilograms
Power
Arrow keys
Return/Enter
4
OPERATING INSTRUCTIONS
Power On
Turn on the power by pressing the power button (1) for 2 seconds. Once on, the scale will begin to auto-check and count down from 0-9 sequentially before entering
the weighing mode
Note: Anything on the scale before powering on will automatically be tared out.
Zeroing
● The zero function is used only when the scale is empty and is not at gross zero due to material build up
● Pressing the ZERO button (5) will reset your scale to 0
● Depending on what your manual zero range parameter is set to, you can zero out any number within your set selection, after that you will receive an error and will need to tare out the weight
Unit Selection
To switch between measuring units kg/lb press the UNIT (3) button
Tare Function
● The Tare function is used when you only wish to see the current change in weight, not the entire amount of weight that is on the scale
● When the indicator is in gross mode (normal weighing) pressing the TARE button (4) will Tare the current weight on the scale and enter the net mode (net light shown)
● For example if you are using a container add the container to the scale, press tare to reset back to 0
● Add your product to the scale to weigh without the weight of the container
● To exit Tare mode press the TARE button (4) again to enter gross mode and you will see the total weight of the container and the product
Note: If you remove the container the scale will show the minus weight of the container
5
Accumulation
● The accumulation function is used to add multiple weights and total them together
● In weighing mode load the rst weight, once stable press the TOTAL button (2) to enter the accumulation mode.
● The screen will ash “
● Remove weight so the scale is at 0 before adding the second weight to the scale
● Once stable press the TOTAL button (2) to add the second weight to the accumulated total
n001
” followed by the rst weight
● The screen will ash “
● Continue this process until all the weights you want to accumulate are stored
● When you are ready to see the accumulated weight press and hold the TOTAL button (2) and the ENTER button (6) at the same time
● The accumulated number “
together) will ash on the display followed by the total
● The total will display by ashing between 2 sets of numbers
● There are 8 digits in total, the display will ash 4 at a time, the rst 4 on the left and
the last 4 on the right. For example if the rst 4 digits are “0012” and the last 4 digits are “3456” the actual weight is 001234.56 or 1234.56 lbs/kg
● To exit accumulation mode, wait for the last 4 digits on the right of the screen to appear, and then press and hold the TOTAL button (2) for one second
● “
CLr n
● If NO you do not want to clear the accumulated total, then keep “
● If YES you do want to clear the accumulated total, then use the arrow button (4) to change to “
● Finally, press the ENTER button (6) to select exit accumulation mode and return to back weighing mode
” will be displayed, asking you if you want to keep the data?
CLr y
n002
”.
” followed by the second weight
n00
#” (the number of weights you are adding
CLr n
”.
6
REMOTE CONTROL
Button Sign Function
1 ON/OFF
2
3
4
5
6
Total
Unit
Tare
Zero
Enter/Set
CALIBRATION PROCEDURE
1. Turn on the scale by holding ON/OFF for 2 seconds.
2. Press ON/OFF and ENTER together to access the setup menu.
3. If done correctly, the display should now show
4. Press ENTER to access the C1 channel. The display should show [
5. Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).
6. Press ENTER to set the value. The display will now show
7. Press ENTER to access the C2 channel. The display should show [
8. Press ZERO to change the setting to the decimal places desired (The C2 channel is
C01
.
#].
#].
C02
C1
.
C2
used to adjust the decimal point on the scale. A value of 1 means there is one digit behind the decimal point.)
9. Press ENTER to set the value. The display will now show
10. Press ENTER to access the C3 channel. The display should show [
11. Press ZERO to cycle through the values until the desired graduation appears. (The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set to 1, the scale will read in 0.1 lb. increments.)
12. Press ENTER to set the value. The display will now show
13. Press ENTER to access the C4 channel. The display will show [######].
14. Enter in the maximum capacity you want to use for this scale by using TOTAL and UNIT
ues down and up. (The C4 channel is used to enter in the max capacity of the scale;
Make sure this doesn’t exceed the max capacity of the scale; Max capacity divided by the increment set in C02 and C03 above cannot exceed 5000.)
15. Press ENTER to set the value. The display will now show
16. Press ENTER to access the C5 channel. The display should show [
17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
18. Press ZERO to change the value to 1.
19. Press ENTER . The display will count down from 10-1 while the scale is
to move the cursor left and right, and TARE and ZERO to move the val-
C03
C04
C05
.
C3
.
.
C5 0
#].
].
calibrating zero. When the display shows 0 the zero calibration is complete.
7
REMOTE CONTROL
Button Sign Function
1 ON/OFF
2
3
4
5
6
20. Press ENTER to continue. The display will now show
21. Press ENTER to access the C06 channel. The display will show [
22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to set the value of C6 to [ and then show [######].
23. Enter the calibration weight value you will use (at least 10% of max capacity you set
in C04 by using TOTAL
and ZERO
24. Place the calibration weight you have on the empty scale and press ENTER .
25. The scale will count down from 10 to 0. Once 0 has been reached, the display will show
26. Press ENTER to continue. The display will now show
27. Press ON/OFF quicly to save and exit the setup menu.
28. The scale has now been calibrated. The display will show the value of the calibration
CALEnd
move the values down and up.
.
C6 1
and UNIT to move the cursor left and right, and TARE
]. Press ENTER . The display will ash
Total
Unit
Tare
Zero
Enter/Set
C06
C07
.
C6 0
.
].
SPAN
,
weight on the scale.
29. If the scale does not show the value of the calibration weight, check that the feet on the platform are not screwed in too tightly, and verify that the platform is level.
30. Unload the scale; the display should read
31. If the scale does not display 000000, check that the feet on the platform are not screwed in too tightly, and verify that the platform is level.
8
000000
.
INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07) and a
parameter settings section (C08 and up).
To enter calibration/parameter settings, follow the procedure below:
1. Press and hold the POWER button (1) and the ENTER button (6) at the same time
2. Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow
buttons (2-5) and the ENTER button (6)
3. Press the 6 button to enter/edit the parameter setting
Press the 1
Table 1. Indicator Parameter Settings
Function Parameter Settings/Options Default
Weighing Unit
Decimal Setting
Graduation Setting (readability of the least
signicant digit)
Maximum Capacity
Zero Calibration
Calibration
Restore Default Settings
Warning Tone
Automatic Power Off
Power Saving Mode
button to save and exit settings at any time
C01 C02
C03
1 = kg 2 = lb
0 = no decimal 1 = 0.0 2 = 0.00 3 = 0.000
4 = 0.0000 options: 1/2/4/10/20/50
Example with no decimal places (ie. C02=0) 1 = 1 lb 2 = 2 lb 5 = 5 lb 10 = 10 lb 20 = 20 lb 50 = 50 lb
set max capacity ex. 100kg = 0100.00 1000
C04 C05
C06 C07 C08 C09
C10
0 = zero calibration not needed 1 = set the zero calibration (Please be sure scale is empty and the stable light is on)
0 = calibration not needed 1 = Ready to calibrate with a calibration weight
0 = do not restore 1 = restore to default settings
0 = turn off warning tone 1 = turn on warning tone
0 = turn off auto power off 10 = power off automatically if no change within 10 min. 30 = power off automatically if no change within 30 min. 60 = power off automatically if no change within 60 min.
LED Version OP900A:
0 = turn off power saving setting 3 = turn off display if no change within 3 minutes 5 = turn off display if no change within 5 minutes
1
0
1
0
0
0
1
0
0
9
Function Parameter Settings/Options Default
Hold Function
C11
Unit Conversion
C12
Upper Limit Alarm
0 = turn off hold function 1 = Peak hold - Grabs the highest weight 2 = Manual hold - Grabs the current weight 3 = Auto hold - Automatically holds data when stable
0 = kg/lb convert on 1 = kg/lb convert off
Set upper limit within the max. capacity 000000
0
0
C13
Lower Limit Alarm
Set lower limit within the max. capacity 000000
C14
Inner Code Display
check the inner code (raw data)
C15
Set Date
Set date from left to right: year/month/day
C16
Set Time
Set the time from left to right: hour/minute/second
C17
Communication Setting
Baud Rate
Manual Zero Range
Initial Zero Range
Zero Tracking
Zero Tracking Time
C18
C19
C20
C21
C22
C23
Set the serial interface data output method: 0 = Turn off serial interface data output 1 = Continuous sending mode, connect remote display 2 = Print mode, connect printer 3 = Command request mode, connect computer.
4 = PC continuous sending mode, connect computer
0=1200 (for remote display)
1=2400 2=4800 3=9600
0 = turn off manually zero setting
1 = ±1% max capacity 2 = ±2% max capacity 4 = ±4% max capacity 10 = ±10% max capacity 20 = ±20% max capacity 100 = ±100% max capacity
0 = no initial zero setting
1 = ±1% max capacity 2 = ±2% max capacity 5 = ±5% max capacity 10 = ±10% max capacity 20 = ±20% max capacity 100 = ±100% max capacity
0= turn off zero tracking
0.5 = ±0.5d d = division
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d Note: the zero tracking range can not be bigger than manual zero range
0 = turn off zero tracking time 1 = 1 second 2 = 2 seconds 3 = 3 seconds
0
3
10
10
.05
1
10
Function Parameter Settings/Options Default
Overload Range
C24
00 = turn off overload range
01-99d = overload range setting d = division
9
Negative Display
Standstill Time
Standstill Range
Digital Filter (for ltering moving
weight such as animals)
Noise Filter
Print Time and Date
C25
C26
C27
C28
C29
C30
0 = -9d 10 = -10% max. capacity 20 = -20% max. capacity
0 = quick 1 = medium 2 = slow
1 = 1d d = division 2 = 2d 5 = 5d 10 = 10d
0 = turn off dynamic lter 1 = 1 digital lter strength 2 = 2 digital lter strength 3 = 3 digital lter strength 4 = 4 digital lter strength 5 = 5 digital lter strength 6 = 6 digital lter strength Note: The higher the number, the higher the lter
strength
0 = turn off noise lter 1 = 1 digital lter strength 2 = 2 digital lter strength 3 = 3 digital lter strength
0 = yy.mm.dd 1 = mm.dd.yy 2 = dd.mm.yy 3 = yy.mm.dd
10
1
2
0
2
11
HELPFUL DEFINITIONS
Division: The amount of increments a scale offers. How accurate the scale can be
Capacity: the maximum amount the scale can contain
Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered
on that will automatically zero.
example: If initial zero range is set to 10% of the max. capacity and your max. capacity is
100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it will automatically zero out the weight.
Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let you manually zero (anything above this percent will be tared)
Zero Tracking Range: A subset to the manual zero range; if the weight on the scale is not stable, the zero tracking range still allows you to zero within a set division of the scale
Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero’d
Overload Range: Weight allowance that is out of the set calibrated range. Adds a tolerance to the calibrated max. capacity without having to recalibrate. example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set the overload range to 60, you can add 1060lbs of weight to the scale without it displaying an error code
Negative Display: How far you can go in the negative direction before displaying an error code
Standstill Time: How fast the scale will stabilize
Standstill Range: How much the scale can uctuate before being determined stable
Digital Filter: For ltering moving weight, such as animals, It changes how sensitive
the scale is to variations in movement.
Noise Filter: A lter for how susceptible the scale is to general variations
Baud Rate: The rate at which information is transferred in a communication channel.
example: In the serial port context, “9600 baud” means that the serial port is capable of transferring a maximum of 9600 bits per second.
12
CONNECTORS
Connecting load cells to the indicator
● The indicator can connect with 6 load cells of 350Ω at most
● 4 wire or 6 wire load cell connections are both okay
● Please contact us directly if you have other special needs for your application
● There are two connection methods between the load cell and indicator
Quick Disconnect as shown below: Figure 2: Quick Disconnect connection diagram
Figure 2: Quick Disconnect connection diagram
Hardwire (Using Inner Terminal Block Connection:
Note: Make sure you follow all the anti-static rules to avoid damage to your indicator
● Excitation voltage: 5V DC
● Largest output current: 120 mA
● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm maximum
● Open the back cover of the weighing indicator, and insert signal cable to the terminal block (see gure 3); Make sure the screw on terminal block is xed tightly
FIGURE 3:
INNER TERMINAL BLOCK
CONNECTION DIAGRAM
Table 4. Wiring Color Code
Signal Name Color Code Description
+Exe/ +EX RED Positive excitation voltage to load cell
+IN / +SIG GREEN Positive output signal from load cell
HD / SHLD YELLOW/THICK BLACK Shield Wire
-IN / -SIG WHITE Negative output signal from load cell
-EXC / -EX BLACK Negative excitation voltage to load cell
13
Error Codes
Error Reason Solution
UUUUUU
nnnnnnn
ERR1
ERR2
ERR3
ERR4 ERR5 ERR6
TROUBLESHOOTING
1. Overload
2. Wrong connection with load cell
3. Load cell has quality problem
1. Calibration is no good
2. Wrong connection with load cell
3. Load cell has quality problem
During calibration, weight is not used or the weight is above the max. capacity
During calibration, the weight is below the minimum required weight
During calibration, the input signal is negative
During calibration signal is unstable After the platform is stable, start calibration
EEPROM Error Change PCB
Exceed Zero Range See Q&A section
1. Reduce the weight
2. Check load cell connection
3. Inspect load cell; Check the input/output
4. See Q&A section
1. Make sure scale is level
2. Check load cell connection
3. Check load cell input and output resistance
4. See Q&A section Use correct weight within the dened range
The calibration weight minimum is 10% of the max. capacity set in C04. Recommended to use 60%-80% of max.
capacity if possible
1. Check all wire connections
2. Check load cell
3. Recalibrate
4. PCB replacement needed if steps 1-3 fail
14
Q&A
Q: The scale does not turn on
A: Make sure the power cord is plugged in, and that there is power. One easy way to
test this is by connecting another appliance to the same outlet and see if it’s operational
Q: The reading goes negative when a load is applied
A: Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or
junction box (if one is used)
Q: How do I resolve ERR6 error?
A: Please follow the procedure below:
1) Turn on the indicator and make sure nothing is on the scale, and that the scale is level and not wobbling
2) Press and hold the “POWER and ENTER” buttons (1+6) simultaneously for 2 seconds
3) The screen will read “C01”
4) Using the arrow buttons, change C01 to C20. You have to change the 1st digit from 0 to 2 rst before you can change the 2nd digit 1 to a 0.
5) Press the “ENTER” button to enter C20 parameter
6) Change the value of C20 on the right to 100 if using the up arrow key
7) Press the “ENTER” button to enter your selection
8) The screen will read “C21” now
9) Press “ENTER” key to enter C21 parameter
10) Change the value on the right of C21 to 100 if available, 20 if not
11) Press “ENTER” key to enter your selection
12) he screen will read “C22” now
13) Press “POWER” key to save and exit
14) Power the indicator off and then on, and see if this resolves the ERR 6 issue. If not, then following the Q&A answers below for resolving “nnnnnn”and “uuuuuu”
errors Q: How do I resolve “nnnnnn” and “uuuuuu” error? A: 1) Check to see if the cable that runs from the indicator to the junction box is
damaged. If it is, replace the cable.
2) Open up the junction box (if available) and check to see if there is any water
damage. If so, replace the junction box
3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block)
are not loose. Re-tighten the screws even if the wires seem to be connected
4) Recalibrate
5) If steps 1-4 do not work, there is a possibility one or more load cells are
defective (consult with support@optimascale.com for further instructions)
15
CONTACT US
Please e-mail sales@optimascale.com for any sales related questions.
Please e-mail support@optimascale.com for any support related questions.
Don’t forget to visit our website at:
optimascale.com
Below are remote guides to print out
REMOTE CONTROL
Button Sign Function
1 ON/OFF
2
3
4
5
Total
Unit
Tare
Zero
REMOTE CONTROL
Button Sign Function
1 ON/OFF
2
3
4
5
Total
Unit
Tare
Zero
6
Enter/Set
6
Enter/Set
Loading...