For safe operation of the weighing indicator, please follow these
instructions:
● Calibration inspection and maintenance of the indicator are prohibited by non-pro-
fessional staff
● Please ensure that the indicator rests on a stable surface
● The indicator is a piece of static sensitive equipment; Please cut off power during
electrical connections
● Touching the internal components by hand is prohibited
● DO NOT exceed the rated load limit of the unit
● DO NOT step on the unit
● DO NOT jump on the scale
● DO NOT use this product if any of the components are cracked
● DO NOT use for purposes other then weight taking
● To avoid damaging the battery do not keep charger plugged in once battery is fully
charged
● Make sure the weight is not over the Max capacity as it could damage the load cell
inside
● Material that has a static electric charge could inuence the weighing. Discharge
the static electricity of the samples, if possible. Another
solution to the problem is to wipe both sides of the pan and the top of the case with
an anti-static agent
Please take anti-static prevention measures
Any accumulated charge on the body of the human operator should be discharged rst
before opening the protective container with ESDS devices inside. The discharge can
be accomplished by:
● Putting a hand on a grounded surface or, ideally, by wearing a grounded Anti-static
Wrist Strap and an Anti-static Mat
PREPARATION & SET UP
● Plug into a wall outlet to avoid interference with other wirings
● Turn on the indicator while there is no load
● Calibration may be required before weighing when the scale is initially installed or
moved from a location
1
FEATURES
● LED or LCD display option
● Multiple weighing units: kg/lb (g/oz/lb:oz offered on LCD versions)
● Gross/Tare/Pre-Set Tare/Zero
● Multiple Hold functions
● Count weighing
● Accumulation weighing
● Overload / Underload indication
● Connects to multiple printers
● Splash proof keyboard and display
● Connects to a Remote Display/Scoreboard
● Power saving mode
● Ability to modify gravity based on different geographical locations
● RS232 output
● NTEP approved for 5,000 divisions
● Relay output (optional) and 4-20mA analog output (optional)
● Can connect to a PC or printer for data logging (optional)
● Wireless capability (optional)
● Rechargeable battery (optional)
Indicator Model Options
The OP-900 series consists of the OP-900A (LED) series and OP-900B (LCD)
series. Options for rechargeable battery, stainless steel enclosure, washdown enclosure,
computer connection, relay output and analog output can be added.
Please contact sales@optimascale.com for a list of possible options.
Technical Parameters
● Accuracy class: 5000 e
● Resolution - Display: 30,000 ; ADC: 2,000,000
● Zero stability error: TK0 < 0.1μV//K
● Span stability error: TKspn < ± 6 ppm//K
● Sensitivity (internal): 0.3 μV / d
● Input voltage: -30 to +30mV DC
● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm max
● AC power: AC 100-250V (use only the included 9V adapter supplied)
● Operation temperature: -10 °C ~ +40 °C
● Operation humidity: ≤90%RH
● Storage temperature: -40 °C ~ +70 °C (32-104°F)
2
SPECIFICATIONS
FIGURE 1: INDICATOR MEASUREMENTS
3
POWER SUPPLY
AC Adapter
If the indicator is powered by an adapter, plug the adapter directly into the “DC” pin
located at the bottom of the indicator. We recommend to plug into a wall outlet to avoid
interference with other wirings.
Battery (Optional)
If you have an indicator with the rechargeable battery option, please charge the
internal battery fully before rst time use. A 110 to 220V AC adapter should be
provided with your indicator. Please use only the AC adapter provided to prevent damage to your indicator. To keep the battery in best condition, fully discharge the battery
every month by leaving the indicator on until the indicator powers off, and then recharge
fully. If the battery is not going to be used for a long period of time it is recommended to
remove it to avoid leakage.
On OP-900A
● When the Battery is low the battery indicator light ashes red
● During charging the red light will stay lit
● The light will turn green once fully charged
ON OP-900B
● symbol will indicate battery’s charge
● symbol indicates that the battery needs to be charged
4
OP-900A (LED)
OP 900B (LCD)
5
DISPLAY AND KEY DESCRIPTION
ON/OFF Powers the Indicator On or Off if held for 2 seconds
HOLDHolds the weight (5 Hold options; can be changed in parameter settings)
TOTAL1. Accumulates weights
2. Works with “Print” to perform the accumulation function and check the
accumulation result
UNITSShifts between weighing units
COUNTUse the scale to count product based on a sample weight
TARE1. Resets the scale to zero when there is something on the scale
(ex. Tare out the weight of a pallet to weigh only the product on it)
2. Clears the tare to see the gross weight (pallet + product)
ZEROZero’s the scale
PRINTPrint data
The scale is at zero
The scale is stable
GrossShows you are in Gross weight mode (includes tare); default mode
NetShows you are in Net weight mode (weight without tared weight)
pcsShows you are in Counting mode
totalShows you are in Accumulation mode
holdShows you are in Hold mode
lbThe weight is shown in pounds
kgThe weight is shown in kilograms
batteryFlashes red = low battery, Solid red = charging, Green = fully charged
OverFlashes when weight is higher than set alarm parameter
AcceptFlashes when weight is within the set alarm parameters
UnderFlashes when weight is lower than set alarm parameter
Power
Back
Save and Exit
Arrow keys
Return/Enter
6
OPERATING INSTRUCTIONS
Power On
● Turn on the power by pressing the power button for 2 seconds. Once on, the scale
will ash the voltage and then begin to auto-check and count down from 0-9
sequentially before entering the weighing mode
Note: Anything on the scale before powering on will automatically be tared out.
Zeroing
● The zero function is used only when the scale is empty and is not at gross zero due
to material build up
● Pressing the ZERO key will reset your scale to 0
● Depending on what your manual zero range parameter is set to, you can zero out
any number within your set selection, after that you will receive an error and will
need to tare out the weight
Unit Selection
● To switch between measuring units ((kg/g/lb/oz/lb:oz) press the UNITS key
Tare Function
● The Tare function is used when you only wish to see the current change in weight,
not the entire amount of weight that is on the scale
● When the indicator is in gross mode (gross light is shown) pressing the TARE key
will Tare the current weight on the scale and enter the net mode (net light shown)
● For example if you are using a container add the container to the scale, press tare
and the display will show the tare symbol
● Add your product to the scale to weigh without the weight of the container
● To exit Tare mode press the TARE key again to enter gross mode and you will see
the total weight of the container and the product
Note: If you remove the container the scale will show the minus weight of the container
To use a pre-set tare weight
● Press and hold the TARE key for 2 seconds
● Input the tare weight using the arrow keys
● Press print key to conrm
and reset back to 0
7
Counting Function
● The counting function is used to count a high volume of identical parts. You can do
this by setting a sample and then either adding to the sample or taking away from
the sample to count the number of objects on the scale
● In weighing mode: Put a sample weight on the scale (Sample options are 5, 10, 20,
50, 100, 200, 500)
● Then press the COUNT key to go to the counting mode (“pcs” will light up)
● It will then display the sample number (ex.
the up or down arrows
● Press the PRINT key to conrm your sample number
● The scale is now ready to start counting, load your product on the scale and the
indicator will show the quantity
● To exit counting mode press the COUNT key
● If you want to count a different product hold the PRINT and COUNT key together
and the sample pieces will reset back to zero
PCS 0
) which you can change using
Accumulation
● The accumulation function is used to add multiple weights and total them together
● In weighing mode load the rst weight, once stable press the ACCUM key to enter
the accumulation mode. The “total” light will display
● Remove weight and press ZERO key before adding the second weight to the scale
● Once stable press ACCUM key to add the weight to the accumulated total
● Press ZERO Key before adding the next weight to the scale
● Repeat previous step until all desired weights have been added to the total (max 999
times)
● When you are done and want to display the accumulated total, press the ACCUM
and PRINT key together. The accumulated number “
you are adding together) will ash on the display followed by the total
● The total will display by ashing between 2 sets of numbers
● There are 8 digits in total, the display will ash 4 at a time, the rst 4 on the left and
the last 4 on the right. For example if the rst 4 digits are “0012” and the last 4 digits
are “3456” the actual weight is 001234.56 or 1234.56 lbs/kg
● If you want to print the accumulated total, hold the PRINT key for one second while
the last 4 digits of the total are shown
● To exit accumulation mode, wait for the last 4 digits to the right of the screen to
appear, and then press and hold the ACCUM key for one second
● “
CLr n
● If NO you do not want to clear the accumulated total, then keep “
● If YES you do want to clear the accumulated total, then use the arrow key to change
to “
● Finally, press the PRINT key to select exit accumulation mode
” will be displayed, asking you if you want to keep the data?
CLr y
”.
###” (the number of weights
n
CLr n
”.
8
Hold
There are 4 different hold functions you can choose from in the C11 parameter
1. Peak Hold: Grabs the highest weight
force)
● Press the HOLD key then add weight to the scale
● The indicator will show the highest weight it recorded and hold it on the screen until
a higher weight is placed on the scale
2. Manual Hold: Grabs the current weight and holds it so it will not change/uctuate
● While weighing, press HOLD and the indicator will hold the current weight on the
screen until HOLD is pressed again
3. Auto Hold: If the weight on the scale is above 20d (20 x division) and is stable,
the indicator will hold that weight on the screen for 3 seconds then go back to general
weighing
● Pressing the hold key is unnecessary, holding is done automatically when the scale
is stable
4. Average Hold: Used for animal weighing, the indicator will display the average
weight sampled from 3 seconds
● Add livestock to scale and press HOLD
● Indicator screen will show “
from those 3 seconds
● Press HOLD again to exit holding mode
LOC
(for materials testing, ie. tension and pulling
” for 3 seconds, then display the average weight
Print
● If the indicator is connected to a printer and the weight on the scale is stable press the
PRINT key to print the current weight
● In accumulation mode hold the PRINT key for one second when the last 4 digits of the
total weight are shown to print the total weight
Note: In tare mode the printer can not print if negative weight is shown
9
CALIBRATION PROCEDURE
1. Turn on the scale by holding ON/OFF for 2 seconds.
2. Press HOLD and PRINTtogether to access the setup menu.
3. If done correctly, the display should now show
4. Press PRINT to access the C1 channel. The display should show [
5. Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).
6. Press PRINT to set the value. The display will now show
7. Press PRINT to access the C2 channel. The display should show [
8. Press ZERO to change the setting to the decimal places desired (The C2 channel is
used to adjust the decimal point on the scale. A value of 1 means there is one digit
behind the decimal point.)
9. Press PRINT to set the value. The display will now show
10. Press PRINT to access the C3 channel. The display should show [
11. Press ZERO to cycle through the values until the desired graduation appears.
(The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set
to 1, the scale will read in 0.1 lb. increments.)
12. Press PRINT to set the value. The display will now show
13. Press PRINT to access the C4 channel. The display will show [######].
14. Enter in the maximum capacity you want to use for this scale by using UNIT and
COUNT
values down and up. (The C4 channel is used to enter in the max capacity of the
to move the cursor left and right, and TARE and ZERO to move the
C01
.
#].
#].
#].
C02
C03
C04
C1
.
C2
.
C3
.
scale; Make sure this doesn’t exceed the max capacity of the scale; Max capacity
divided by the increment set in C02 and C03 above cannot exceed 5000.)
15. Press PRINT to set the value. The display will now show
16. Press PRINT to access the C5 channel. The display should show [
17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
18. Press ZERO to change the value to 1.
19. Press PRINT. The display will count down from 10-1 while the scale is
calibrating zero. When the display shows 0 the zero calibration is complete.
OP-900 offers 2 calibration methods, Single Point which uses one weight to calibrate or
Linear Calibration, which uses multiple (2-7) weights for a more accurate calibration.
To Calibrate using only 1 calibration weight (Single Point Calibration)
20. Press PRINT to continue. The display will now show
21. Press PRINT to access the C06 channel. The display will show [
22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to
set the value of C6 to [
then show [######].
C6 1
]. Press PRINT. The display will ash
C05
C06
.
C5 0
.
C6 0
].
].
SPAN
, and
10
23. Enter the calibration weight value you will use (at least 10% of max capacity you set
in C04 by using UNIT
and ZERO
24. Place the calibration weight you have on the empty scale and press PRINT.
25. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show
26. Press PRINT to continue. The display will now show
27. Press ACCUM to save and exit the setup menu.
28. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale.
29. If the scale does not show the value of the calibration weight, check that the feet on
the platform are not screwed in too tightly, and verify that the platform is level.
30. Unload the scale; the display should read
31. If the scale does not display 000000, check that the feet on the platform are not
screwed in too tightly, and verify that the platform is level.
CALEnd
move the values down and up.
.
and COUNT to move the cursor left and right, and TARE
.
000000
C07
.
To Calibrate using only multiple calibration weights (Linear Calibration)
1. Press PRINT to continue. The display will now show
2. Press PRINT to access the C06 channel. The display will show [
3. The C6 channel is used to calibrate the scale with a known weight. Press ZERO
to set the value of C6 to [
and then show [
4. Press ZERO to enter the number of weights you want to use (2-7); the more
you use, the more accurate the calibration will be. (an example of 2 will be used)
5. Press PRINT to set your value. The screen will ash [
6. Enter the lowest calibration weight value you will use (at least 10% of max
capacity you set in C04) by using UNIT
and right, and TARE
7. Place the calibration weight you have on the empty scale and press PRINT.
8. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show he screen will ash [
9. Enter the next calibration weight value you will use by using UNIT and COUNT
to move the cursor left and right, and TARE
and up.
10. Place the calibration weight you have on the empty scale and press PRINT.
11. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show he screen will show
12. Then follow steps 26-31 on the single point calibration instructions
LnE 2
C6 2
].
and ZERO move the values down and up.
dbno02
CALEnd
]. Press PRINT. The display will ash
and COUNT to move the cursor left
] then [
.
000200
and ZERO move the values down
C06
dbno01
]
.
C6 0
] then [
].
SPAN
000100
]
,
11
INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07 explained above) and a
parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be
OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08 and up
can be accessed by the user. If you break the ofcial seal by opening the back of the indicator to
access the seal switch, you may need to have the indicator recertied. Be sure to adjust the seal
switch back to the original setting after calibration/conguration has been performed.
To enter calibration/parameter settings, follow the procedure below:
1. Make sure the unit is set to either kg or lb
2. Press and hold the HOLD and PRINT key at the same time for 2 seconds
3. Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow
keys and return keys as labeled under each indicator button
4. Press the PRINT key to enter/edit the parameter setting
Press the ACCUM key to save and exit settings at any time
Table 1. Indicator Parameter Settings
FunctionParameter Settings/Options
Weighing Unit
Decimal Setting
Graduation Setting
(readability of the least
signicant digit)
Maximum Capacity
C01
C02
C03
1 = kg
2 = lb
3 = gram
4 = oz
Note: for calibration only kg or lb are allowed
0 = zero calibration
1 = set the zero calibration (Please ensure scale is empty and the
stable light is on)
0 = calibration not needed
1 = Ready to calibrate with one calibration weight
2 = Ready to calibrate using multiple calibration weights (Linear)
3 = Sensitivity Output
12
FunctionParameter Settings/Options
Restore Default
Settings
Warning Tone
Automatic
Power Off
Power Saving Mode
Hold Function
Unit Conversion
C07
C08
C09
C10
C11
0 = do not restore
1 = restore to default settings
0 = turn off warning tone
1 = turn on warning tone
0 = turn off auto power off
10 = power off automatically if no change within 10 minutes
30 = power off automatically if no change within 30 minutes
60 = power off automatically if no change within 60 minutes
LED Version OP900A:
0 = turn off power saving setting
3 = turn off display if no change within 3 minutes
5 = turn off display if no change within 5 minutes
LCD Version OP900B:
0 = turn off the backlight
1 = backlight only when the weight changes or keyboard is pressed
2 = constant backlight
0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
3 = Auto hold - Automatically holds data when stable
4 = Average hold - for animal weighing, averages the weight from a
sample of 3 seconds
5 = Auto Average hold - Average hold without the need to press the
hold key
* See table 2
C12
Upper Limit Alarm
Set upper limit within the max. capacity
C13
Lower Limit Alarm
Set lower limit within the max. capacity
C14
Inner Code Display
check the inner code (raw data)
C15
Set Date
Set date from left to right: year/month/day
C16
Set Time
Set the time from left to right: hour/minute/second
C17
Communication
Setting
Baud Rate
C18
C19
Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Continuous sending mode, connect remote display
2 = Print mode, connect printer
3 = Command request mode, connect computer.
4 = PC continuous sending mode, connect computer
5 = PC/big display, continuous sending mode
6 = Print to adhesive label printer
7 = Print to Zebra adhesive label printer
0=1200 (for remote display)
1=2400
2=4800
3=9600
13
FunctionParameter Settings/Options
Manual
Zero Range
Initial Zero Range
Zero Tracking
Zero Tracking Time
Overload Range
C20
C21
C22
C23
C24
0 = turn off manually zero setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
0= turn off zero tracking
0.5 = ±0.5d d = division
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note: the zero tracking range can not be bigger than manual zero
range
0 = turn off zero tracking time
1 = 1 second
2 = 2 seconds
3 = 3 seconds
00 = turn off overload range
01-99d = overload range setting d = division
0 = turn off dynamic lter
1 = 1 digital lter strength
2 = 2 digital lter strength
3 = 3 digital lter strength
4 = 4 digital lter strength
5 = 5 digital lter strength
6 = 6 digital lter strength
Note: The higher the number, the higher the lter strength
14
FunctionParameter Settings/Options
Noise Filter
C29
Print Time and Date
C30
Analog Output Setting
C31
Calibrate Current
0 = turn off noise lter
1 = 1 digital lter strength
2 = 2 digital lter strength
3 = 3 digital lter strength
0 = turn off relay output
1 = turn on relay output function 1
2 = turn on relay output function 2
3 = Reserved menu
9.7000 - 9.9999
9.7000 - 9.9999
C37
Version No.
C38
Print Mode
C41
Print Carriage Return
0 = auto mode
1 = gross mode
2 = tare mode
0 - 9 (How much space between print outs)
C42
Space Print
0 - 9 (Where the data prints on the paper: 0 = left ; 9 = right)
C43
Date Print
C44
Time Print
C45
0 = do not print the date
1 = print the date
0 = do not print the time
1 = print the time
Table 2. Unit Conversion Parameter Settings
Parameter SettingsUnits Available
C01= 3 & C12= 0gram only
C01= 4 & C12= 0oz only
C01= 1 & C12= 0kg only
C01= 1 & C12= 1kg/lb
C01= 1 & C12= 2kg/lb/oz
C01= 1 & C12= 3kg/lb/lb:oz/oz
C01= 1 & C12= 4kg only
C01= 2 & C12= 0lb only
C01= 2 & C12= 1lb/kg
C01= 2 & C12= 2kg/lb/oz
C01= 2 & C12= 3kg/lb/lb:oz/oz
C01= 3 & C12= 4lb only
15
Table 3. Default Parameter Settings
FunctionParameter Default Setting
Weighing UnitC011
Decimal SettingC020
Graduation SettingC031
Maximum CapacityC041000
Zero CalibrationC050
CalibrationC060
Restore DefaultC070
Warning ToneC081
Automatic
Power Off
Power Saving ModeC100
Hold FunctionC110
Unit ConversionC121
Upper Limit AlarmC13000000
Lower Limit AlarmC14000000
Inner Code DisplayC15
Set DateC16
Set TimeC17
Communication SettingC180
Baud RateC193 (9600)
Manual Zero RangeC2010
Initial Zero RangeC2110
C090
Zero TrackingC220.5
Zero Tracking TimeC231
Overload RangeC249
Negative DisplayC2510
Standstill TimeC261
Standstill RangeC272
Digital FilterC280
Noise FilterC292
Print Time and DateC300
Analog Output SettingC311
Calibrate CurrentC324
Relay Output SettingC331
Multi-connection add.C340
Wireless Communication
Gravity of Calibration
Location
Gravity of DestinationC379.7936
C356
C369.7936
16
HELPFUL DEFINITIONS
Division: The amount of increments a scale offers. How accurate the scale can be
Capacity: the maximum amount the scale can contain
Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered
on that will automatically zero.
example: If initial zero range is set to 10% of the max. capacity and your max. capacity is
100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it
will automatically zero out the weight.
Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let
you manually zero (anything above this percent will be tared)
Zero Tracking Range: A subset to the manual zero range; if the weight on the scale
is not stable, the zero tracking range still allows you to zero within a set division of the
scale
Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the
scale to fall within the zero tracking range tolerance and still qualify to be zero’d
Overload Range: Weight allowance that is out of the set calibrated range. Adds a
tolerance to the calibrated max. capacity without having to recalibrate.
example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set
the overload range to 60, you can add 1060lbs of weight to the scale without it
displaying an error code
Negative Display: How far you can go in the negative direction before displaying an
error code
Standstill Time: How fast the scale will stabilize
Standstill Range: How much the scale can uctuate before being determined stable
Digital Filter: For ltering moving weight, such as animals, It changes how sensitive
the scale is to variations in movement.
Noise Filter: A lter for how susceptible the scale is to general variations
Baud Rate: The rate at which information is transferred in a communication channel.
example: In the serial port context, “9600 baud” means that the serial port is capable of
transferring a maximum of 9600 bits per second.
17
CONNECTORS
Connecting load cells to the indicator
● The indicator can connect with 6 load cells of 350Ω at most
● 4 wire or 6 wire load cell connections are both okay
● Please contact us directly if you have other special needs for your application
● There are two connection methods between the load cell and indicator
Quick Disconnect as shown below:
FIGURE 2: QUICK DISCONNECT CONNECTION DIAGRAM
Hardwire (Using Inner Terminal Block Connection:
Note: Make sure you follow all the anti-static rules to avoid damage to your indicator
● Excitation voltage: 5V DC
● Largest output current: 120 mA
● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm maximum
● Open the back cover of the weighing indicator, and insert signal cable to the
terminal block (see gure 3); Make sure the screw on terminal block is xed tightly
18
FIGURE 3: INNER TERMINAL BLOCK CONNECTION DIAGRAM
Table 4. Wiring Color Code
Signal NameColor CodeDescription
+Exe/ +EXREDPositive excitation voltage to load cell
+IN / +SIGGREENPositive output signal from load cell
HD / SHLDYELLOW/THICK BLACKShield Wire
-IN / -SIGWHITENegative output signal from load cell
-EXC / -EXBLACKNegative excitation voltage to load cell
DB9 Connection (9 pin Serial Connector)
The DB9 9 pin serial connector is used for different purposes depending on the
indicator model
● Figure 4 shows the pin assignment on the DB9 9 pin connector
FIGURE 4: DB9 SERIAL CONNECTOR PINOUT
There are 3 Output formats to choose from
1. RS232 Serial Output Format (Standard)
2. 4-20 mA Analog Output (Optional)
3. Relay Output (Optional)
19
RS232 SERIAL OUTPUT FORMAT
Follow the pin out of Table 5 below to connect the indicator the RS-232 Serial device
Table 5. DB9 Pin Description
DB9 PinDenitionFunction
2TXTTransmit Data
3RXDReceive Data
5GNDGround Interface
The serial output format depends on the settings for parameter C18. The serial output
consists of a string of ASCII characters. Here is a list of the serial parameters
● 8 data bits
● 1 stop bits
● No parity
● No handshaking
Note: With the RS232 Output option we have data logging software available as seen in
Figure 5.
FIGURE 5. DATA LOGGING SOFTWARE
Below are the formats of the serial output
● C18=0 - Turn off serial interface data output
● C18=1 - Continuous sending mode, connect 2nd big display
● C18=2 - Print mode, connect printer
● C18=3 - Command request mode, connect computer.
● C18=4 - PC continuous sending mode, connect computer
● C18=5 - PC/big display, continuous sending mode
● C18=6 - Print to adhesive label printer
● C18=7 - Print to Zebra adhesive label printer
20
Remote Display Continuous Sending Mode (C18=1)
For use with a Scoreboard/Remote Display
Note: Baud Rate must be set to 1200 (C19 = 0)
Output Continuous Format
S
T
X
123456
S
S
S
W
W
W
A
B
012Decimal point position
100XXXXXX0
010XXXXXXX
110XXXXX.X
001XXXX.XX
101XXX.XXX
Bit 3printing=1
Bit 4Extend display=1
Bit 5Constant 1
Bit 6Constant 0
21
Print Mode (C18 = 2)
For printing on a non-adhesive ticket printer. Parameters 16, 17, 30, & 42-45 all effect
your ticket print out.
Normal weighing ticket printout example:
The 4-20 mA analog output of the OP-900
scale is a voltage sourcing sensor that will
output current which is proportional to the
calibrated scale’s weight range (i.e. 4 mA = 0
LBS and 20 mA = 10,000 LBS). It is
important to note, that this is a sourcing
output sensor (i.e. the sensor will source
9VDC with an output current range of 4-20
mA). Unlike many other “loop powered” type 4-20 mA
sensors, an external supply voltage should not be connected
to the unit’s 4-20 mA circuit, however, the ground connection
(pin 6) of the DB-9 connector will need to be connected to
the same ground as the data acquisition device which will be
responsible for interoperating the 4-20 mA signal. This ground
connection is imperative, as both the data acquisition device’s
power supply and the scale’s internal 9VDC power supply will
need to be on the same ground plane for the output current to
FIGURE 6. CONNECTION DIAGRAM
be synced and measured correctly. Please reference Figure 6 as a visual clarication on
how to connect your scales 4-20 mA output to a data acquisition device.
Below is a list of important notes when using the indicator with the 4-20 mA option
● Resolution: 1/1000
● Outside Load: 100-350ohms
● Inside connection: load input port pin “1” of J2, ground port pin “GND” of J2
● Outside connection: load input port pin 1 of DB9, ground port pin 6 of DB9
● To test the connection, connect a 250 ohm load; Locate a volt meter, and probe
across the 250 ohm load. As the weight input to the indicator varies, the voltage of
the volt meter will change accordingly
● Pressing the TARE key will reset to output current to 4mA
● 0-20mA output can be set by setting parameter C31 to 0
● Please note that this option will disable the RS232 weigh data output (serial ASCII
data) that comes in the standard indicator
Calibration:
● Press PRINT and HOLD key to go into conguration mode
● Go to C32 and press PRINT key
● The display should show [out-4] and output should be at 4mA
● Press the up/down arrow keys and the [out-#] will increase/decrease
Note: X corresponds to the output current. For example if out-12 is displayed, then
12mA should be the output. If not, press the left and right arrow key to adjust/calibrate
24
RELAY OUTPUT (OPTIONAL)
● The indicator can output 4 signals, which when connected to outside equipment,
can perform an automatic control function and an upper/lower limit alarm function.
● Change parameter setting C33 following Table 6 below:
Table 6: Relay Output Parameter Setting
Output PortPort DenitionFunction
Out1Turn off output functionNo Output Signal
C33=0
C33=1
C33=2,3Preserved, no function
Out2Turn off output functionNo Output Signal
Out3Turn off output functionNo Output Signal
Out4Turn off output functionNo Output Signal
Out1Turn on overload control functionOutput overload control signal
Out2Turn on compliance control functionOutput compliance control signal
Out3Turn on under-load control functionOutput under-load control signal
Out4Turn on stable control functionOutput stable control signal
Table 7 below shows the DB9 port pinout for the relay output option. Please note that
this is optional and only available for the OP-900 indicator ordered with this option.
The relay output option will disable the RS232 weigh data output (serial ASCII data) that
comes in the standard indicator.
Table 7: Relay Output Pin Denition
DP9 PinDenitionPort
11st output signal pinOut1
st
61
22
72
33
83
44
94
output signal pinOut1
nd
output signal pinOut2
nd
output signal pinOut2
rd
output signal pinOut3
rd
output signal pinOut3
th
output signal pinOut4
th
output signal pinOut4
Rating:
AC 250V 3A
DC 30V 3A
25
Error Codes
ErrorReasonSolution
UUUUUU
nnnnnnn
ERR1
ERR2
ERR3
ERR4
ERR5
ERR6
TROUBLESHOOTING
1. Overload
2. Wrong connection with load cell
3. Load cell has quality problem
1. Calibration is no good
2. Wrong connection with load cell
3. Load cell has quality problem
During calibration, weight is not used or
the weight is above the max. capacity
During calibration, the weight is below
the minimum required weight
During calibration, the input signal is
negative
During calibration signal is unstableAfter the platform is stable, start calibration
EEPROM ErrorChange PCB
Exceed Zero RangeSee Q&A section
1. Reduce the weight
2. Check load cell connection
3. Inspect load cell; Check the input/output
4. See Q&A section
1. Make sure scale is level
2. Check load cell connection
3. Check load cell input and output resistance
4. See Q&A section
Use correct weight within the dened range
The calibration weight minimum is 10% of
the max. capacity set in C04.
Recommended to use 60%-80% of max.
capacity if possible
1. Check all wire connections
2. Check load cell
3. Recalibrate
4. PCB replacement needed if steps 1-3 fail
26
Q&A
Q:The scale does not turn on
A:Make sure the power cord is plugged in, and that there is power. One easy way to
test this is by connecting another appliance to the same outlet and see if it’s
operational
Q:The reading goes negative when a load is applied
A:Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or
junction box (if one is used)
Q:How do I resolve ERR6 error?
A:Please follow the procedure below:
1) Turn on the indicator and make sure nothing is on the scale, and that the scale
is level and not wobbling
2) Press and hold the “PRINT and HOLD” key simultaneously for a few of seconds
3) The screen will read “C01”
4) Using the arrow keys, change C01 to C20. You have to change the 1st digit
from 0 to 2 rst before you can change the 2nd digit 1 to a 0.
5) Press “PRINT” key to enter C20 parameter
6) Change the value of C20 on the right to 100 if possible using the up arrow key.
If 100 is not available change to 20
7) Press “PRINT” key to enter your selection
8) The screen will read “C21” now
9) Press “PRINT” key to enter C21 parameter
10) Change the value on the right of C21 to 100 if available, 20 if not
11) Press “PRINT” key to enter your selection
12) he screen will read “C22” now
13) Press “TOTAL” key to save and exit
14) Power the indicator off and then on, and see if this resolves the ERR 6 issue.
If not, then following the Q&A answers below for resolving “nnnnnn”and “uuuuuu”
errors
Q:How do I resolve “nnnnnn” and “uuuuuu” error?
A:1) Check to see if the cable that runs from the indicator to the junction box is
damaged. If it is, replace the cable.
2) Open up the junction box (if available) and check to see if there is any water
damage. If so, replace the junction box
3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block)
are not loose. Re-tighten the screws even if the wires seem to be connected
4) Recalibrate
5) If steps 1-4 do not work, there is a possibility one or more load cells are
defective (consult with support@optimascale.com for further instructions)
27
CONTACT US
Please e-mail sales@optimascale.com for any sales related questions.
Please e-mail support@optimascale.com for any support related questions.
Don’t forget to visit our website at:
optimascale.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.