Page 8
4.1 Adjust the observation head
Loosen the lock-screw, turn the observation head to a comfortable position for observation,
and then lock the lock-screw.
4.2 Place the specimen on the stage
Put the specimen on the mechanical stage (transparent glass plate). Ensure that the specimen is centred over the stage opening by adjusting the coaxial knobs of the stage.
Brighteld vision
4.3 Lamp settings
The microscope is tted with a 3.5W X-LED3 illuminator. Insert the plug of the cable into
the power socket and turn on the switch on the back of the main body. Turn the brightness
adjustment knob (on the right of the body) until you hear a “click”, meaning that the lamp is
on, then keep turning the knob to a brightness level suitable for observation.
4.4 Adjust interpupillary distance
Hold the right and left parts of the observation head by both hands and adjust the interpupillary distance by turning the two parts until one circle of light can be seen. After nishing
the interpupillary adjustment, turn the dioptric adjustment ring on the left eye tube to the “0”
position at the graduation line on the eyetube.
4.5 Focus and diopter adjustment
Loosen the focus-stop knob (near the left focus knob), rotate the coarse focusing knob to
bring the slide into focus with an objective with a low magnication, and then lock the focusstop knob. Adjust the ne focusing knob to get the image sharp and clear while observing
with your right eye, then turn the left dioptric adjustment ring to a sharp and clear image
also with the other eye. Turn the tension-adjust-knob (near the right focus knob) to get a
suitable tension for the focus system. When the image is in focus, turn the nosepiece to
choose the objective you need.
4.6 Condenser adjustment
Turn the condenser adjustment knob to move the condenser up or down for a clear and
uniform sample illumination. The two condenser-centring screws can be used to centre the
condenser.
4.7 Set the numerical aperture
Adjust the aperture of the iris diaphragm to set the numerical aperture of the illuminator,
thus controlling image contrast and resolution.
4.8 Field diaphragm
Adjust the eld diaphragm to align the system and to control stray light.
4.9 Additional lters
The chromatic correction lter (blue) can be inserted into the holder (just above the ring of
the eld diaphragm) for denition of specimen parts.
4.0 USING THE MICROSCOPE