16 optics11.com
2.3 Calibrating the optical signal
For each unique medium/probe combination, the optical signal has to be calibrated. Once this is
done, the next step is to calibrate the geometrical factor (see paragraph 2.4).
Please note that for correct operation, the optical signal calibration should always be
performed before the geometrical factor calibration
The calibration of the optical signal is done by the OP1550 interferometer and controlled via by the
Piuma software. First ensure the optical probe is inserted in the Piuma indenter head and the
optical fiber connector is properly connected to the OP1550 interferometer ‘Signal’ input
connector. When preparing the probe to measure samples in a solution, please insert few
millimeters of solution in a clean petri-dish and use the manual stage to lower the probe into the
solution. Confirm the cantilever and a part of the glass probe is fully submersed and the liquid
tension cannot affect the cantilever anymore.
Please note that glass is hydrophobic, and that it is possible that an air bubble is trapped
between the cantilever and the glass base. This air bubble will affect the measurement, and
needs to be removed (see Chapter 4.3).
Automated optical signal calibration
Once the probe is connected and adapted to the correct medium, select the mode of operation in
main Piuma software window to ‘Displacement control’ (Figure 9) and be sure to turn off the DMA
mode in the DMA tab of the configure ‘Profile’ menu (Figure 10, left). Then click on ‘Calibration’ in
the Chiaro Software Menu to enter the calibration menu (Figure 9). Press ‘Scan OP1550 Wavelength’
to start the optical calibration (Figure 10, right). You can observe the wavelength scan process, after
closing the calibration window (Figure 11).