Opsytec UVpad E User Manual

Opsytec Dr. Gröbel GmbH
Am Hardtwald 6-8
Tel.: 07243 / 94783 - 50
Fax: 07243 / 94 83 - 65
info@uv-groebel.de
UVpad E
Manual
Version: 1.1.2
1 Content 2

1 Content

1 Content................................................................................................................................. 2
2 Change history ..................................................................................................................... 3
3 Symbols Overview .............................................................................................................. 3
4 Function Overview .............................................................................................................. 4
5 Technical Data ..................................................................................................................... 5
5.1 General technical data ............................................................................................... 5
5.2 Radiometric measurements ....................................................................................... 5
5.3 spectral ranges ........................................................................................................... 5
USB ........................................................................................................................... 6
5.4 / hardware-software-requirements ............................................................................. 6
5.5 Effect related measurements ..................................................................................... 6
6 Intended Use ........................................................................................................................ 7
7 Conventional Use ................................................................................................................ 7
8 Important Information ......................................................................................................... 8
9 Operation ............................................................................................................................. 9
9.1 Switching the device on and off ................................................................................ 9
9.2 Standby Menu ......................................................................................................... 10
9.2.1 Start Measurement ............................................................................................... 11
9.2.2 Start manual dark measurement .......................................................................... 13
9.2.3 Show last Measurement ....................................................................................... 15
9.2.4 Setting integration time during measurements .................................................... 15
9.2.5 Recording time .................................................................................................... 17
9.2.6 Settings ................................................................................................................ 17
9.2.7 Information .......................................................................................................... 17
9.3 Settings .................................................................................................................... 18
9.3.1 Sensitivity ............................................................................................................ 18
9.3.2 Averaging ............................................................................................................ 19
9.3.3 Effective Function* ............................................................................................. 19
9.3.4 Language ............................................................................................................. 21
9.3.5 Date / Time .......................................................................................................... 21
9.3.6 Factory Settings ................................................................................................... 21
10 Data Export / PC-Software ................................................................................................ 22
10.1 Installation ............................................................................................................... 22
10.2 Connecting with the PC ........................................................................................... 22
10.3 Viewer Software ...................................................................................................... 23
10.4 Export of measurements .......................................................................................... 25
10.5 Displaying and comparing Measurements .............................................................. 26
10.6 Displaying the Irradiance profile ............................................................................. 26
10.7 Deleting Measurements ........................................................................................... 27
11 Remote control / Live measurement .................................................................................. 28
12 Notes for optimal measurement results ............................................................................. 31
12.1 The UVpad E displays „MAX“ ............................................................................... 31
12.2 The UVpad E displays too little irradiance values .................................................. 31
12.3 The UVpad E displays too high or too low UVC Values ....................................... 31
12.4 The UVpad E displays “ - “ ..................................................................................... 32
12.5 The measured value is not reproducible .................................................................. 32
13 Cleaning, Maintenance & Troubleshooting....................................................................... 33
13.1 Calibration ............................................................................................................... 33
13.2 Cleaning .................................................................................................................. 33
2 Change history 3
Date
Version
Changed by
Comment
01.04.2015
1.0.7
Paravia
first Version
29.09.2015
1.0.8
Paravia
setting “fixed (Res. 0.01)” added
23.03.2017
1.0.9
Uhl
Upgrade measurement time; UVA/VIS range adjusted
10.11.2017
1.1.0
Paravia
Manual dark measurement
10.01.2018
1.1.1
Uhl
Remote control / LIVE measurement
19.02.2018
1.1.2
Paravia
Editorial modifications
13.3 Battery Change ........................................................................................................ 34
13.4 Troubleshooting ...................................................................................................... 35
14 Declaration of Conformity ................................................................................................. 36
15 Notes .................................................................................................................................. 37

2 Change history

3 Symbols Over view

Meaning: Failure to observe the mentioned instructions can re­sult in an injury of the user.
Meaning: Failure to observe the mentioned instructions can re­sult in a damage of the device.
Meaning: Instructions are to be observed for the regular opera­tion.
*
Optional functions, not available for every UVpad E
4 Function Overview 4

4 Function Overview

The UVpad E combines the benefits of the spectral measuring technology with a handheld and easy to operate radio meter. Use the spect roradiometer UVpad E for an accurate meas urement of the UV irradiance and for comparisons of different UV lamps and UV LEDs. 512 photodi­odes measure the spectrum in the wavelength range of 200 – 440 nm. The categorization in UVA, UVB and UVC is taken according to international standards and traceable. The calibra­tion to only one light source is omitted.
During the measurement, the spectrum is shown on the graphic display. The irradiances for UVA, UVB, UVC and VIS are available by the click of a button. In the background, the spec­trum is already record ed. Whi le measuri ng, the do se an d the ch ronolo gical i rradian ce seq uence can additionally be documented. Spectra and measurement data can be exported and evaluated with the provided software.
The UVpad E offers these functions/features:
Spectral measurements:
Real-time of the spectrum
Real-time of irradiance
Radiometric measurements:
Spectrum at max irradiance
Irradiance (UVA, UVB, UVC, VIS)
Profile of irradiance (UVA, UVB, UVC, VIS)
Dose of irradiance (UVA, UVB, UVC, VIS)
Measurements related to cau se and eff ect s:
Photo initiator spectrums (optional)
Sensitivity functions of broadband radiometers (optional)
5 Technical Data 5
Size (L x W x H)
160 x 100 x 20 mm³
Weight
~500 g
Power
battery powered, 3 x CR2032
Max. operational temperature
60°C
Display
graphical, 128 x 64 px
Optically active surface
Diameter: 10 mm
Optically active surface position
external sensor
internal memory
50 measurements with export option
Spectral range
200 - 440 nm ±3nm
Spectral bandwidth
~ 2 nm
Irradiance measurement range
2 - 15000 mW/cm²
Irradiance dose measurement range
1 mJ/cm² - 600 J/cm²
Irradiance profile
4 spectral ranges (UVA, UVB, UVC, VIS)
Measurement time
0 to 120 s, adjustable to ~ (rounded)
5 s (100 Hz)
Time resolution
10 ms to 250 ms, adjustable
Sampling rate
10 ms - 1000 ms,
storage rate for export: see time resolution
Averaging
1 - 32
Gain
normal / LowGain for 10 to 30 ms
Cosine correction
yes
Calibration
yes, traceable to PTB / NIST
frequency of recalibration, recommended
12 months
according
CIE
UV-C
240 - 280 nm
UV-B
280 - 315 nm
UV-A
315 - 400 nm
VIS
400 - 480 nm
based customer specification or CIE
optional

5 Technical Data

5.1 General technical data

5.2 Radiometric measurements

5.3 spectral ranges

120 s (4 Hz) 48 s (10 Hz) 24 s (20 Hz) 10 s (50 Hz)
automatically set
5 Technical Data 6
USB connector
MiniUSB
PC requirements
min Intel CORE i3, 2 GB Ram, >40 Gb HDD
operating system
Windows 7 with .NET Framework > 4.0
Photoinitiator / Sensitivity curve
optional
5.4 USB / hardware-software-requirements

5.5 Effect related measurements

6 Intended Use 7
optically active surface

6 Intended Use

The UVpad E determines the irradiance in the UV and VIS spectral range and indicates these on the integ r ated display. Typic al Applic ati ons:
- Measurement of UV-LEDs and ultraviol et lamps
- UV dose measureme nt s
- Universal irrad iance measurement
- GMP and IQ/OQ docum ent ation

7 Conventional Use

The UVpad E is solely intended for the determination and visualization of irradiances, which are generated in UV curing systems.
The UVpa d E is battery operated and suitable for usage with lithium batteries, type CR2032. In case other batteries, than the supplied ones, are used, an operating range of up to 70 °C has to be observed. Spare batteries are to be obtained from your supplier.
Operation only allowed in dry environments. The assembly position is arbitrary.
When using the UVp ad E, light and UV radiation can be reflected and distributed at the housing. Appropriate protection measures for the protection against radiation need to be applied, if neces­sary. Note: The UVpad E itself does not emit any hazardous radiation.
The housing, with the exception of the battery compartment, must not be opened by the user.
The optically active surface must not be touched.
Gloves need to be worn when cleaning the optically active surface.
The UVpad E is a highly sensitive optical measurement device. It must not be exposed to shocks.
Any other usage than the one described above leads to a damage of the product. Furthermore, this involves risks, such as short-circuits, fire and electric shocks. The entire device must not be modi­fied and/or remodeled! The safety indications are to be observed without fail.
8 Important Information 8
Skin oil and contamination are absorbing in the UV spectral range. Finger
are to be avoided. If
ATTENTION: Opening of the housing (except for the battery compartment)
is a highly sensitive optical measurement device. It must not be
to cool

8 Important Information

Please read the instruction manual completely before start-up and operati on. The latest version of the instruction manual applies.
prints / contamination on the optically active surface necessary, clean thoroughly with Isopropanol. Use gloves when cleaning.
The UVpad E is calibrated.
leads to a loss of calibration.
Only use batteries, which are suitable for up to 60 °C.
The UVpa d E exposed to shocks.
When using the UVpad E, light and UV radiation can be reflected and distributed at the housing. Appropriate protection measures for the protection against radi- ation need to be applied, if necessary.
The maximum housing temperature of the sensor can re ach > 60 °C and there is the risk of skin burns on contact. Please allow for the UVpad E down and use protective gloves to avoid skin burns.
9 Operation 9
Key
Function
Back / Cancel / Off
Down
Up  Confirm / Continue

9 Operation

The UVpad E is operated via four keys. These are arranged on the front panel below the dis­play. The keys and the functions are illustrated as follows:
The keys  and are addit ionally available as softke ys. Softkey functions are directly dis- played above the key. Example:

9.1 Switching the device on and off

The UVpad E is activated via any key. During the initialization (ca. 2s), the welcome screen is displayed:
After the initialization, the standby menu is displayed:
9 Operation 10
, the device switches off
In this example, the language “English” is set.
In the standby menu, the UVpad E is switched off via the key .
After 2 minutes of not using the UVpad E automatically.

9.2 Standby Menu

Via the standby menu, all functions of the UVpad E are available for you. These are:
Start measurement
Show last measurement
Scope measurement
Settings
Information
In the standby menu and in all other menus, select the active menu item via the keys  and . Confirm the active menu item with the key . With , you return to the previous menu with- out saving any changes.
The selected menu item is highlighted in black and displayed inverted. If several settings are selectable, then the selected setting is marked with. Example:
“German” is selected, but not set.
With the key -key, the setting is accepted and the language is being changed.
9 Operation 11
From the standby menu, a measurement is
First, the spectrum is displayed.
Switch between spectral and radiometric view
matically recorded if the radiometric view is

9.2.1 Start Measurement

started with the key .
Use ‘”Auto” mode to get the best integration time or, in fixed mode use  and  ke ys to adjust integration time.
Notes: The optimal integration time, gain (L/H) and the saturation of the measurement range are dis­played at the upper display edge. The integration time is displayed in ms. Example:
T 100: The integration time is 100 ms
Hint:
For optimum measurement results, the measuring range should be fully con­trolled. Very good measuring results are achieved when the measuring range is at least 50%.
The modulation is linear to the set integration time. Example:
55/5000 T10 M20%: Measuring range is 20% used => Bad T40 M80%: Measuring range is 80% used => Very good
by pressing .
If storing measurements is sel ected via menu „recording time“, the measurement is auto-
shown.
9 Operation 12
First, the dose of irradiance is displayed.
With the key , the irradiance profile for
Stop the measurement by pressing the key for at least 0.5s.
The measurement results are displayed automatically. For a better overview, these are parti­tioned to four screens.
1. Irradiance (only when viewing saved measurements)
2. Dose of irradiance (UVA, UVB, UVC, VIS)
3. Spectrum at maximum irradiance (only when viewing saved measurements)
4. Irradiance profile over the duration of the measurement
These can be displayed with the keys or .
With the key , you are respectively being redirected to the previous menu. With the keys or  you can return to the standby screen.
Note: If the irradiance is too low for the measurement range, then the UVpad E dis­plays no measured values. The irradiance and the dosage values are then di s­played with “-“.
UVA, UVB, UVC, VIS and total are dis­played.
9 Operation 13
Note:
integration time, the UVpad
1. Set the correct integration time with
3. Cover the sensor, wait 1-2 seconds or

9.2.2 Start manual dark measurement

For manual darkness measurement, select the menu item "dark+radiometric" under Settings, Measurement mode. If your UVpad E is equipped with several effective functions, manual dark measurement may be available for further effective functions.
Manual darkness measurement is only active in the sensitivity mode "fixed". Therefore, set a fixed integration time!
If the darkness measurement is not done for the E will not displa y any values. In this case th e irradiance and dose val ues are displayed with "-".
Manual darkness measurement is important if the irradiance at the end of the short-wave measuring range (e. g. 240 nm) is not "zero", otherwise th e meas­ured values are too low.
For the measurement with manual dark measurement, start the measurement and set the integration time with the button  or .
the button or .
2. Start irradiance display with key or
start dark measurement with key .
3-4 measurements and press to switch to the irradiance display.
9 Operation 14
4. The irradiance is displayed.
Note:
measurement is not done or the integration
5. To save the measurem en t, p ress th e
button for at least two seconds. After release, the irradi ation values are dis­played.
The irradiance is not displayed if the darkness
time has been changed.
The note DARK is displayed.
Note: The optimum integration time, the gain (L/H) and the control of the measuring range are dis­played at the top of the display. The integration time is displayed in ms.
Example: T 100: The integration time is 100 ms
9 Operation 15
Select the measurement with the buttons
If 50 measurements have alread y been sav ed, then a no te comes up pri or to a
stating that the memory is full and that the measurement
Confirm with Yes (). Return with or .

9.2.3 Show last Measurement

In the standby menu select the item “Sh ow las t measu remen t” to d ispl ay the l ast m easurem ent. Select the values from the list of measurements, which you want to display. The navigation in the display screens is taken out, as described in chapter 9.2.1.
and . Confirm with . Return with .
new measurement, can be taken out without saving it, or that alternatively, the oldest measure­ment result is being overwritten.
See Delete all Measurements See Data Export
Delete all measurements:
The internal mem ory can be deleted directl y on the UVpad E or via a PC. For this purpose, please select the prior menu item “Clear memory”.

9.2.4 Setting integration time during meas ur ements

If using fixed integration time, during a measurement you can set up the sensitivity of the UVpad E with the keys and . Select for a lower sensitivity and for a higher sensitiv­ity. The sensitivity is not saved.
9 Operation 16
s not fill the
), if single lines of the spectrum are cut on top.
Sensitivity too low
Sensitivity correct
Sensitivity too high
Select a higher sensitivity (), if the spectrum is noisy or doe display completely.
Select a lower sensitivity ( The measurement range is then exceeded and is shown in the display MAX.
Examples:
Risk of UV radiation.
Protect your skin and eyes during the work at and with UV radiation sources.
9 Operation 17
Set recording time via the buttons
With any key, you are being redirected to the

9.2.5 Recording time

If storing measurements is selected via settings (120 s to 5 s), the measurement is automatical­ly recorded if the radiometric view is shown.
and . Confirm with or cancel the s et- ting with .

9.2.6 Settings

In the menu „Settings“, you can set up the sensitivity, the effective function, the menu lan­guage and the date / time. Furthermore, you can reset the UVpad E to the factory settings.

9.2.7 Information

Select the menu item “Information” from the standby menu, to display the serial number, the firmware version, as well as the date of calibration.
menu.
9 Operation 18
In this menu, the following functions are
In the setting „automatically“, the UVpad E
diance is existent. This setting is suitable for
measuring range then is 0 .. 650 mW/cm².

9.3 Settings

By selecting the menu item „Settings“, you are being directed to the settings menu of the UVpad E.
available:
- Settings of the sensitivity
- Selection of the effective function
- Language settings
- Settings of date and time
- Reset to factory settings

9.3.1 Sensitivity

The UVp ad E adjusts the sensitivity automatically, in order to ensure the optimal signal-noise ratio. The sensitivity is hereby set via the sampling rate. With high irradiances, the sam pling rate is increased in ord er to not ex ceed the measurement ran ge. With low irradi ances the sam­pling rate is dec reased. As a result in belt conveyor s ystems erroneous m easurements may be recorded. Please follow the instructions on the display and / or set a fixed integration time as sensitivity. Five pre-defined sensitivities are available.
adjusts the sensitivity, once a minimum irra-
most applications. In the setting „low“, the sampling rate is al­ways between 10 ms and 100ms. The UVpad E can measur e UV-systems with a very high irradiance or significant irradiance modifica­tions (e.g. with shutter operation or with very high tape speed). In the setting „high“, the UVpad E measures with a sampling rate of 1000 ms. This is ideal for applications with a low irradiance. In the settings „fixed“ and „fixed (Res 0.01)“, the integration time is predetermined. The UVpad E can be used optimally for known systems with shutters
Note: In irradiation profile resolution is enhanced from 1 mW/cm² to 0.01 mW/ m² for the set­tings "high" and "Fixed (Res. 0.01). The
9 Operation 19
The menu item “Sample rate” is displayed, if
Select „radiometric“, for traceable and inter-
optimization. The international comparability
you have selected a fixed sensitivity.
The integration time is set via the buttons and . Confirm with or cancel the s et- ting with .
Pressing  on 10 ms setting switches to low sensitivity. This reduces sensitivity by factor of 5.

9.3.2 Averaging

The UVpad E can calculate the irradiance from the moving average. Hereby are signal-to-noise ratio, and thus measurement accuracy can be enhanced. For integration times below 500 ms we recommend averaging over 16-32 measurements, while for integration times over 500 ms smaller settings may be useful.

9.3.3 Effective Function*

Effective functions serve for the process-oriented measurement. This is often also called effect­related or actinic measurement.
In that case, effective functions can be for example:
- Absorption spectrums of photo initiators
- Spectral sensitivity function
- Filter function for instrument comparisons
In the Eff. function menu, the desired function can be selected using the -key or -ke y and confirmed with the -key.
nationally comparable measurements.
Select an effective function for a process­oriented measurement and a targeted system
is not given.
9 Operation 20
Example: radiometric
Example: with effective function
If measurements are t aken out effect-related, then this i s displayed in the irradiance and in the dosage display by the addition “(ef)” for “effective”. Rad iometric and effect-r elated measure­ments are illustrated subsequently.
The function “Effective Function” is only available, if your UVpad E has been programmed with a customized effective function at factory.
9 Operation 21
Select the desired l anguage with the keys
First set the day with the keys  and . Se-
Confirm the restoration of the factory settings

9.3.4 Language

The UVpad E comes with a menu in the German and English language.
and and confirm with .

9.3.5 Date / Time

The UVpad E has a real-time clock to check the validity of the calibration. You can set up the real-time clock in the menu „Date / Time“.
lect  for the month, th e year, the hours and minutes. Confirm your settings with . With  you can cancel the settings at any time.

9.3.6 Factory Settings

Reset the UVpad E to the factory settings and confirm.
with .
With  you can cancel the restoration of the factory settings.
10 Data Export / PC-Software 22
4.0 or higher is required. This is found on the installation CD or free of

10 Data Export / PC-Software

10.1 Installation

For the installation, please proceed as follows:
1.) If necessary, disconnect the connection of the UVpad E with the PC.
2.) Start the installation with “setup.exe” in th e master data of the Software CD. Follow
the instructions of the installation program.
The driver installation is executed as an independent application TUSB3410.
For the software UVpad E Viewer, the Microsoft .NET Framework Version
charge at: www.microsoft.de.
3.) After completing the inst allation, connect the UVpad E with the PC. The UVpad E is
integrated as a virtual, serial interface. The driver installation is taken out automatical­ly under Windows 7.

10.2 Connecting with the PC

Connect the UVpad E with the PC and switch it on. The screen displays the message “USB connected”.
Once the UVpad E is connected with the PC, it does not react to any keystroke. Disconnect the connection with the PC and the UVpad E continues working autonomously.
10 Data Export / PC-Software 23

10.3 Viewer Software

The UVpad E software is divided into two areas. The are a measurement overview displays the executed measurements with date / time, irradiation and dosage, as well as the integration time, the saturation und the measurement duration.
Individual measurem ents can be com pared in th e di spla y area bas ed on sp ectr a, ir radiance pro­file or by means of measurement data.
The buttons EN and DE let you change the language (German / English).
When starting the so ftware, the measuremen ts are automaticall y imported, when an UVpad E is being detected. Importing can take a few seconds. During this time, a progress bar displays the import:
If the UVpad E is not connected or not ready yet, then an error message pops up. Restart the import after a few seconds via the button „Connect“.
10 Data Export / PC-Software 24
Viewer displays measurement settings and results as
In the spectrum view and the irradiance profile graph, you can zoom with the right mouse but­ton. Use left mouse button to zoom out again. Example:
The max. irradiance and dose will then be calculated in active view (irradiance profile).
Working with different measurements: You can select and compare d iffer ent measur ement s. You can also add a comm ent in t he table before exporting data.
Note: The UVpad E sensitivity, saturation and errors. Use mouse over function in table view this.
10 Data Export / PC-Software 25

10.4 Export of measurements

Measurements can be exported individually, in a group or completely. For this purpose, mark up to 10 measurements.
Button „Deselect all“ deselect marked measurements.
Select “Export selected measurements” or “Export all measuremen ts” for the d ata export. The measurements are ex ported as a text file, separated by a comma, and can thus be import ed to a table calculation program, such as Microsoft EXCEL.
The exported file contains all information to the measurement, the spectrum and the irradiation profile.
Marked measurements appear in the display area, as illustrated by the following LED spec­trum:
10 Data Export / PC-Software 26

10.5 Displaying and compari ng Me a sureme nt s

Select the desired display via the tab:
Below is an example with two irradiance measurements.
Up to 10 measurem ents are s electabl e at t he s am e time an d ar e label ed i n t he dis pla y area wi th date / time.
Each display can be saved via the button “save chart” as an image file (*.jpg, *.bmp, *.tiff, *.wmf).

10.6 Displaying the Irradia nc e profile

The irradiance profil es are separately available for all fou r spectral ranges. For this purpose, mark the desired measurements and select the desired spectral range on the tab “Irradiance profile”.
The following example illustrates an irradiance profile for all spectral ranges.
The data logging tak es place adju stable b etween 1 0 and 250 ms. With shorter integration times, the maximum irradiance is being saved. The calculation of the dosage is taken out independently in real time.
10 Data Export / PC-Software 27
, the serial number and the assignment of memory
The screen can be adjusted via the zoom function. For this purpose, change the X and the Y intercept and confirm with „Zoom Fit“. The maxi mum irradiance Emax and th e dosage D in the display area are displayed on the right.

10.7 Deleting Measurements

You can delete th e entire memory of t he UVpad E via the button “Delete all measu rements”. Confirm the confirmation prompt with OK. Individual measurements cannot be deleted.
Tip: Via t he button „Info “ space of the UVpad E are displayed.
11 Remote control / Live measurement 28
Export area

11 Remote control / Live measurement

In addition to the export functions, the UVpad E software also has a remote control called "Live measurement ". Th e button with the s ame n ame takes you to a separ ate user interf ace for remote control of the UVpad E.
The remote control is available for firmware version 1.2.4 and higher. Ask for an update if necessary!
The user interface is divided into sections. All measurements received by the UVpad E are displayed in the dis play area. The paramet ers that must b e adjusted for a measuremen t can be found in the parameter area. Once a measurement has been completed, the user has the option of exporting a "Single measuremen t" or "All measuremen ts" in the export ar ea. In contrast to the single measurement, when exporting all measurements not only the averaged spectrum is written to a *. csv file. All single dark current or spectra measurements are also written to the file.
Display area
Parameter area
11 Remote control / Live measurement 29
Tip:
Before starting a measurement, integration time, dark current or another effective function can be set in the parameter area. Basically, two different operating modes are available for one measurement.
In the calibrated mode, radiometric weighted measured values are displayed. In the "RAW data" mode, only unweighted count values can be recorded and exported. In con­trast to the calibrated mode, there is no automatic dark current correction and calibration.
Three methods are available for the dark current measurement, depending on the selected mode. In the "Auto" mode a dark current measurement is done directly in the device. In "Auto" mode, dark current correction is carried out as usual in the UVpad E device.
“Calibration + Auto” mode correspo nds to the measurement result of the UVpad E.
(All other modes are suitable for special applications and are calculated in the PC.)
In "Manual" mode, an entire dark current spectrum is recorded which is later calculated with the spectrum. If no dark current correction is to be carried out, select the mode "Without".
After starting a dark measurement or spectrum, the selected number of averages is imported from the device. The progress of a measurement series is displayed by a progress bar.
For best measurement results, the measuring range should be almost com­pletely controlled. Good measurement results are achieved if the measuring range is used by at least 50%. The modulation is linear to the set integration time. Example:
Integration time 10 ms; modulation 20%: Measuring range is used up to 20%  Bad
Integration time 40 ms; modulation 80%: Measuring range is used to 80% Very good
After the measurement has been completed, there are two options in the export area (single measurement / all measurements) for saving the measurement results in a file.
11 Remote control / Live measurement 30
Tip:
When exporting all measurements, in addition to the averaged measurement result, all individ­ual measurements as well as the measurement results for irradiance are exported to a *. csv file. The export of a "single measurement" only provides the averaged measurement result. All measurement settings are also saved in the export file.
Note:
The subsequent modification of a parameter blocks both export options!
To switch between German and English use the key combination:
Ctrl + Shift + d ~ german
Ctrl + Shift + e ~ english
12 Notes for optimal measurement results 31
If the measuring range is over modulated in setting “fixed sensitivity” a hint

12 Notes for optimal measurement results

The UVpad E achieves op ti m al m easu remen t res u l t s, i f t he m easu remen t ran ge is w ell adj us ted to the irradiance and thus to the UV system. This is generally taken out automatically. The following notes are there to support you with the setting of the UVpad E, so that optimal measurement results can be achieved.

12.1 The UVpad E displays „MAX“

The UVpad E saves the maximum irradiance during the measurement. Due to the individual lines in the spect rum , t he m easu remen t s i gnal c an p artly exceed the measur emen t range. In this case, the UVpad E displays the message “MAX”. Select low sensitivity or a fixed integration time. Reduce the lamp power, if this is insufficient.
will be shown after a measurement. Note that measurement dat a will not be stored.

12.2 The UVpad E displays too little irradiance values

Check if the sensit ivity has been set correctl y. Optimal measurement results are achieved, if the measurement ran ge is used to at least 50%. After a measurement, the value M should be higher than 50%.
Select a low or automatic sensitivity or set a short integration time.
When "automatic sensitivity" is set, the spectrum and irradiance will be stored as the optimal measuring range is automatically found. In the case of band systems, it may happen that a maximum is recorded prior to the lamp position and that UVpad E is over modulated below the lamp. In this case the shown irradiance may be too low. However, the dose is cor­rectly. In this case a fixed integration time may be helpful.

12.3 The UVpad E displays too high or t oo low UV C Values

The UVpad E measures the entire UV spectrum range and divides it standardized into UV spectrum ranges (UVA, UVB, UVC). For comparison: Broadband radiometers possibly only measure a part of this range, if these do not operate with standardized spectral ranges. Thus, high deviations can arise especially in the UVC. Furthermore, the UV lamps generally emit visible light and IR radiation apart from the UV radiation. Both can generate scattered light in the UVpad E, which is then detected. It is also possible that the UVpad E heats up during the measurement and thus the reading noise in-
12 Notes for optimal measurement results 32
Reduce the integration time to 10ms and not saturation. Then set the
creases. Therefore, the UVpad E has an automatic correction function. However, low UVC values can possibly not be avoided, even if the UV lamp does not emit UVC.

12.4 The UVpad E displays “ - “

Just like each m easuremen t device, t he UVpad E has a fixed meas urement ran ge. If th is range is fallen below, then the UV radiation cannot be captured correctly. In that case, the UVp ad E displays “ – “.
Select a higher, lower or automatic sensitivity or set a long integration time.

12.5 The measured value is not reproducible

Measurements are only stored and displayed in UVpad E when the saturation is between 10% and 100%. This can cause to saturation direct under the light source, however correct meas­urement may be recorded before or after that. Especi all y UV LEDs ca n le ad to a co rrect meas­urement switching the LED on within one measurement, while all other are saturated.
Select a fixed integration time.
Note: highest possible integration time
13 Cleaning, Maintenance & Troubleshooting 33
Confirm with the key .
calibration. This can depend on the individual use, the environment and

13 Cleaning, Maintenance & Troubleshooting

13.1 Calibration

The calibration of the UVpad Es is checked during the initialization. If the UVpad E is not calibrated properly anymore, then this is indicated by a note after the initialization process. Measurements are also possible at any time with an expired calibration.
The UVpad E can only be maintained and calibrated by the manufacturer. For this purpose, please consult your supplier.
UV spectroradiometer are in principle subject to a high high-energy load by UV radiation. This can affect the calibr ation and l ead to an earl y due date for a re­the radiation sources.

13.2 Cleaning

Skin oil and contamination are abs orbing in the UV spectral range and t hus affect the meas­urement result. Finger prints or contamination on the optically active entrance opening are therefore generally to be avoided. Should a cleaning become necessary, please use a lintless cloth and Isopropanol. Clean the optically active surf ace with circul ar movements. Avo id a strong force. If n ecessary, leave th e UVpad E to dry completely after cleaning it. Please note:
For cleaning the optically active surface, please wear gloves.
Do not use Acetone or other chemical cleaning agents to clean the UVpad E.
13 Cleaning, Maintenance & Troubleshooting 34
The housing, with the exception of the battery compartment, is not to be opened
When the batteries are emty a measurement cannot be started and the
The measured data will be stored. Date and time should be adjusted.
except for the battery compartment, must not be opened by the
ATTENTION: Opening of the housing (except for the battery compartment)
by the user.

13.3 Battery Change

UVpad E reinitializes. The message "empty battery" is displayed.
Change the batteries.
The UVpad E is operated with lithium batteries. For a battery change, switch the UVpad E off and turn it with the rear panel facing towards you. Loosen the three marked screws with a screwdriver (Torx T6) and remove the cover.
Remove the old lithium batteries carefully and replace them by new ones. Close the cover again and tighten the screws carefully.
Please note:
Only use batteries, which are suitable for up to 70 °C. Appropriate batteri es can be obtained from your supplier.
The housing, user.
The UVpad E is calibrated.
leads to a loss of calibration.
13 Cleaning, Maintenance & Troubleshooting 35
Error / Error Message
Cause
Remedy
The UVpad E cannot be
The batteries are empty
Change batteries.
delivery status.
The measured value is not
Automatic, low or high sen-
Use fixed sensitivity
The UVpad E displays too
Effect-related measurement
Check the settings for effect-
The UVpad E displays a sat-
Measurement range exce ed ed
Use the sensitivity setting
The UVpad E displays a
Sensitivity too low
Increase the sensitivity to The calibration is expired
Calibration interval expired
Have the UVpad E re-
The UVpad E is not detected
The UVpad E is switched off
Check USB cable
Switch the UVpad E on and
Use delivered cable.

13.4 Troubleshooting

The subsequent notes and error messages address the user. The information should help to en­sure the release and the appropriate operation of the reference spectrometer. For this reason,
possible causes and remedies are listed below:
switched on
reproducible
high / too low results
Interruption
sitivity selected
The optically active surface is contaminated
The calibration is invalid
wrong settings
Restart the device. For this purpose, remove all batteries and replace them after 5 minutes. The UVpad E is in
related measurements.
Clean the optically active surface.
Have the UVpad E re­calibrated.
check integration time
uration “MAX”
strong noise
by the PC
„low“.
„high“.
calibrated.
wait a few seconds for the initialization.
14 Declaration of Conformity 36

14 Declaration of Conformity

Hereby we,
Opsytec Dr. Gröbel GmbH
Am Hardtwald 6-8
76275 Ettlingen
declare that the following product
spectral Radiometer
Type designation:
UVpad E
have been developed, constructed and produced under our sole responsibility and that the product is conform with the following standard(s) or guideline(s) in this declaration:
2014/35/EU
“Directive of th e European Parliament and of the Council on the harmonization of the laws of the Member States regarding the provision of electrical equipment for use within certain voltage limits on the market (Low-Voltage Directive) “.
The conformity of the designated product with the provisions of the directive is verified by full compliance with the following standards:
DIN EN 60204-1: Safety of machinery - Electrical equipment of machines - Part 1: General requirements (IEC 60204-1:2005, modified); German version EN 60204-1:2006
Ettlingen, 04.11.2016
Dr. Mark Paravia Executive Board
Electronically created and valid without signature.
15 Notes 37

15 Notes

Loading...