Onset UX100-014M Operating Manual

HOBO® Single Channel Thermocouple Data Logger (UX100-014M) Manual
The HOBO Single Channel Thermocouple data logger records temperature in indoor environments with J, K, T, E, R, S, B, or N type thermocouple sensors (sold separately). The logger also has an internal 10K thermistor for cold-junction compensation of the thermocouple output. The thermocouple channel has a 20-bit resolution while the internal 10K thermistor has 12-bit resolution. Using HOBOware®, you can easily configure the logger alarm to trip for specific high or low temperatures. Or, you can set up burst logging in which the logger records data at a different interval during certain conditions. The logger can also calculate minimum, maximum, average, and standard deviation statistics. This compact data logger has a built-in LCD screen to monitor temperature, logging status, battery use, and memory consumption in between readouts.
Specifications
Thermocouple
Type Range Accuracy Resolution
HOBO Single Channel Thermocouple Data Logger
UX100-014M
Included Items:
Command™ strip
Double-sided tape
Hook & loop strap
Required Items:
HOBOware 3.4 or later
USB cable (included with
software)
Thermocouple sensor type J, K, T, E, R, S, B, or N
J -210° to 760°C
K -260° to 1,370°C
T -260° to 400°C
E -260° to 950°C
R -50° to 1,550°C
S -50° to 1,720°C
B 550° to 1,820°C
N -260° to 1,300°C
Internal 10K Thermistor (Temperature)
Range -20° to 70°C (-4° to 158°F)
Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution 0.024°C at 25°C (0.04°F at 77°F); see Plot A
Drift <0.1°C (0.18°F) per year
Logger
Logger Operating Range Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing);
Logging Rate 1 second to 18 hours, 12 minutes, 15 seconds
Logging Modes Fixed interval (normal), burst, or statistics
Memory Modes Wrap when full or stop when full
Start Modes Immediate, push button, date & time, or next interval
Stop Modes When memory full, push button, or date & time
Restart Mode Push button
Time Accuracy ±1 minute per month at 25°C (77°F), see Plot B
Power Source One 3V CR2032 lithium battery and USB cable
Battery Life 1 year, typical with logging rate of 1 minute and sampling interval of
Memory 512 KB (208,076 measurements, maximum)
(-346° to 1,400°F)
(-436° to 2,498°F)
(-436° to 752°F)
(-436° to 1,742°F)
(-58° to 2,822°F)
(-58° to 3,128°F)
(1,022° to 3,308°F)
(-436° to 2,372°F)
±0.6°C (±1.08°F) ± thermocouple probe accuracy
±0.7°C (±1.26°F) ± thermocouple probe accuracy
±0.6°C (±1.08°F) ± thermocouple probe accuracy
±0.6°C (±1.08°F) ± thermocouple probe accuracy
±2.2°C (±3.96°F) ± thermocouple probe accuracy
±2.2°C (±3.96°F) ± thermocouple probe accuracy
±2.5°C (±4.5°F) ± thermocouple probe accuracy
±1.0°C (±1.8°F) ± thermocouple probe accuracy
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
15 seconds or greater
0.03°C (0.06°F)
0.04°C (0.07°F)
0.02°C (0.03°F)
0.03°C at (0.05°F)
0.08°C (0.15°F)
0.08°C (0.15°F)
0.1°C (0.18F)
0.06°C (0.11°F)
16211-G
HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual
Specifications (continued)
Logger (continued)
Download Type USB 2.0 interface
Full Memory Download Time 1 minute, 30 seconds
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly
Size 3.66 x 5.94 x 1.52 cm (1.44 x 2.34 x 0.6 in.)
Weight 30 g (1.06 oz)
Environmental Rating IP50
or go blank in temperatures outside this range
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
Plot A: Internal Temperature Accuracy & Resolution Plot B: Time Accuracy
Logger Components and Operation
Start/Stop Button
Mounting
Loop
LCD Screen
Battery Tray
USB Port
Start/Stop Button: Press this button for 3 seconds to start or stop logging data, or to resume logging on the next even logging interval. This requires configuring the logger in HOBOware with a push button start or stop, and with “Resume logging on next button push” selected (see Setting up the Logger). You can also press this button for 1 second to record an internal event (see Recording Internal Logger Events) or to turn the LCD screen on if the option to turn off the LCD has been enabled (see Setting up the Logger).
Battery Tray: Remove the battery tray on the top of the logger to access the logger battery (see Battery Information).
Alarm/Stats Button
Internal 10K Thermistor (Temperature)
Mounting
Loop
Thermocouple Port
Alarm/Stats Button: Use this button to clear a tripped alarm (see Setting up Alarms) or to switch between statistics, alarm readings, and the current temperature reading.
Internal 10K Thermistor (Temperature): The internal 10K thermistor (not visible in the diagram) measures ambient temperature, if the channel is configured to log in HOBOware (see Setting up the Logger).
Mounting Loops: Use the two mounting loops to mount the logger with the hook-and-loop strap (see Mounting the Logger).
Thermocouple Port: Use this port to connect the thermocouple sensor (see Connecting the Thermocouple Sensor).
1-800-LOGGERS 2 www.onsetcomp.com
HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual
USB Port: Use this port (not visible in the diagram) to connect the logger to the computer or the HOBO U-Shuttle via USB cable (see Setting up the Logger and Reading Out the Logger).
LCD Screen: This logger is equipped with an LCD screen that displays details about the current status. This example shows all symbols illuminated on the LCD screen followed by definitions of each symbol in the table below.
LCD Symbol Description
The logger is waiting to be launched. Press and hold the Start/Stop button for 3 seconds to launch the logger.
The logger has been launched with push button stop enabled; press and hold the Start/Stop button for 3 seconds to stop the logger. Note: If you also launched the logger with a push button start, this symbol will not appear on the display for 30 seconds.
The battery indicator shows the approximate battery power remaining.
The logger has been configured to stop logging when memory fills. The memory bar indicates the approximate space remaining in the logger to record data. When first launched, all five segments in the bar will be empty. In this example, the logger memory is almost full (only one segment in the memory bar is empty).
The logger has been configured to never stop logging (wrapping). The logger will continue recording data indefinitely, with newest data overwriting the oldest data. When first launched, all five segments in the memory bar will be empty. In this example, the memory is full (all five segments are filled in) and new data is now overwriting the oldest data. This will continue until the logger is stopped or the battery runs out.
The logger is currently logging.
A sensor reading is above or below the high or low alarm that you configured. Press and release the Alarm/Stats button until the “alm” symbol (described below) is displayed on the screen. This symbol at left will clear depending on how alarms were configured in HOBOware. If the alarm was configured to clear when the logger is relaunched, this symbol will remain on the LCD. Otherwise, it will clear when the sensor reading is back within the alarm limits or by pressing the Alarm/Stats button for 3 seconds.
An alarm is ready to be cleared. This will only appear if “Cleared with button press” was selected in the HOBOware alarm settings. Press the Alarm/Stats button for 3 seconds to clear the alarm.
These symbols show the maximum, minimum, average, and standard deviation values most recently calculated for the logger (if the logging
LCD Symbol Description
mode has been set to Statistics in HOBOware). Press the Alarm/Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature (or to the alarm value if applicable).
This is the sensor reading that tripped the alarm. Press the Alarm/Stats button to view this reading. Press the Alarm/Stats button again to cycle through any statistics (defined above) and ultimately back to the current temperature.
This is an example of a temperature reading. Temperature units are determined by the settings in HOBOware. To switch between Celsius and Fahrenheit, change the Display Preferences in HOBOware before launching the logger.
The logger has been configured to start logging on a particular date/time. The display will count down in days, hours, minutes, and seconds until logging begins. In this example, 5 minutes and 38 seconds remain until logging will begin.
The launch settings are being loaded onto the logger from HOBOware. Do not disconnect the USB cable during this process.
An error occurred while loading the launch configurations onto the logger from HOBOware. Make sure the USB cable is connected to both the logger and the computer and try launching again.
The logger has been stopped with HOBOware or because the memory is full.
Notes:
You can disable the LCD screen when logging. Select “Turn LCD off” when setting up the logger as described in the next section. When this option is enabled, you can still temporarily view the LCD screen by pushing the Start/Stop button for 1 second. The LCD will then remain on for 10 minutes.
The LCD screen refreshes every 15 seconds regardless of the logging interval selected in HOBOware. If you choose a logging interval less than 15 seconds, the data will be recorded at the faster interval, but the sensor readings will only be updated on the screen every 15 seconds.
When the logger has stopped logging, the LCD screen will remain on until the logger is offloaded to a computer or HOBO U-Shuttle (unless launched with the “Turn LCD off” option). Once the logger has been offloaded and disconnected from the computer, the LCD will turn off automatically after 2 hours. The LCD will turn back on the next time the logger is connected to the computer.
Setting up the Logger
Use HOBOware to set up the logger, including setting alarms, selecting the options to start and stop logging, and choosing a logging mode.
1. Connect the logger and open the Launch Logger window.
To connect the logger to a computer, use the USB cable provided. Click the Launch icon on the HOBOware toolbar or select Launch from the Device menu.
Important: USB 2.0 specifications do not guarantee operation outside the range of 0°C (32°F) to 50°C (122°F).
1-800-LOGGERS 3 www.onsetcomp.com