Onset T-ION-TVOC Connection Instructions

Ion Science VOC Monitor Connection Instructions
For use with HOBO® H22, U30, U12, UX120-006M, and RX3000 data loggers and HOBO ZW data nodes
Applies to this Ion Science VOC Monitor:
T-ION-TVOC 4-20mA representing 0 to 10/100/1000 ppm TVOC
WARNING! All Onset devices listed in this document are NOT intrinsically safe and should NOT be used in
hazardous areas!
This document provides instructions on connecting the Ion Science VOC Monitor listed above to each of the following:
FlexSmart
Analog Sensor Port/Module option used with HOBO U30 and RX3000 series loggers.
4-20mA cable used with the HOBO U12 and UX120-006M data loggers and HOBO ZW data
nodes
It also lists configuration values used by HOBOware Note: For VOC monitor details, refer to the documentation provided by Ion Science.
Required
Ion Science VOC Monitor listed above
HOBO H22, U12, U30, UX120-006M, or RX3000 data logger, or HOBO ZW data node
FlexSmart Analog Module, Onset Part No: S-FS-CVIA (for H22 series); Analog Sensor Port/Module option (for U30 and
RX300 series); 4-20mA cable, Onset Part No. CABLE-4-20mA or voltage adapter, Onset Part No. CABLE-ADAP5 (for U12, UX120-006M, or ZW series
HOBOware software, version 2.2.1 or higher (2.4.0 or higher for U30 series; 3.6 or higher for UX120-006M); HOBOware
Pro 3.0 or higher for ZW series
HOBOlink (for RX3000 series data loggers)
Customer-supplied lengths of hook-up wire
Analog Module used with HOBO H22 series data loggers
®
software to configure the logger for the monitor.
Ion Science VOC Monitor
Connecting the VOC Monitor to the Analog Module or Port
H22, U30, or RX3000 Connection
HOBO H22, U30, or RX3000 Series Data
HOBO H22 or U30 Series Data Logger
S-FS-CVIA
FlexSmart Analog Module
S-FS-CVIA FlexSmart Analog Module
Terminal Block
S1-
Shld
S1+
Exc+
Exc-
S2+
S2-
+
Power
-
+
4-20mA
Logger (H22-001 shown)
(H22-001 shown)
All hook-up wire customer-supplied
Power and 4-20mA Terminal Blocks
(inside VOC monitor housing)
Ion Science VOC Monitor
-
S-FS-CVIA Module Wiring Diagram
(for H22 series only)
U30 Analog Sensor Port
Terminal Blocks
Exc Out
Ch1 Signal
Ch1 Return
Shield
Ch2 Signal
Shield
Analog Sensor Port
Ch2 Return
Ext Return
(installed in U30)
+
Power
-
+
4-20mA
Power and 4-20mA Terminal Blocks
(inside VOC monitor housing)
-
Analog Sensor Port/Module Wiring Diagram
Analog Sensor Port Wiring Diagram
(U30 shown)
(for U30 series only)
© 2009–2015 Onset Computer Corporation. All rights reserved. Onset, HOBO, HOBOware, and HOBOlink are registered trademarks and FlexSmart is a trademark of Onset Computer Corporation. Other products and brand names may be trademarks or registered trademarks of their respective owners.
13043-D MAN-ION_SCIENCE
Ion Science VOC Monitor Connection Instructions
U12, ZW, or UX120-006M Connection
HOBO ZW Series Wireless Data Node; U12 or
UX120-006M Data Logger (UX120-006M shown)
Connect to analog port
CABLE-4-20MA
AC-SENS-1*
CABLE-4-20MA
(+)
(-)
Blue (-)
Yellow (+)
AC-SENS-1*
+
Power
-
+
4-20mA
-
Ion Science VOC Monitor
*CAUTION: Polarity is very important!
Before plugging AC-SENS-1 into AC power, make sure the correct polarity lead
connections to device are made as shown!
Power and 4-20mA Terminal Blocks
(inside VOC monitor housing)
Configuring the Data Logger for the VOC Monitor using HOBOware Software
Note: HOBOware software provides a configuration file for the monitor. The table below lists the recommended configuration values that this file contains. For information on loading configuration files, refer to the software documentation.
Channel Name Warm-Up* Measurement
Type
VOC Continuous Current 4 20 mA 0 10 (or 100 or 1000)
Raw Value 1
Raw Value 2
Raw Units
Scaled Value 1
Scaled Value 2 Scaled Units
ppm
*Loop power of 22 mA can be provided by the FlexSmart Analog Module and H22 (or Analog Sensor Port/Module and U30 or RX3000). Lamp and electronics require 65 mA. Note: Providing continuous warm-up excitation power will accelerate data logger battery depletion. Therefore, we recommend using the appropriate Onset AC power adapter to power the logger.
Page 2 of 2
Loading...