Onset SD-MA-420 Operating Manual

Self-Describing 4-20 mA Input Sensor (SD-MA-420) Manual
For use with HOBO® MX1104 and MX1105 data loggers
Specifications
Measurement Range 0 to 20.1 mA
Accuracy ±0.001 mA ±0.2.% of absolute current
Resolution 0.002% of full scale
Cable Length 47 cm (18.5 inches)
This cable measures currents from 0 to 20.1 mA and must be connected so that the current flows through, and with the proper polarity, as shown below. Do not expose to current above 20 mA or negative current. Do not cut off the end of the black cable where it connects to the blue and yellow wires as that contains the precision resistor required for current measurements.
Blue wire (-)
Yellow wire (+)
0-20 mA
Using the Sensor with the HOBO MX Analog/Temp/RH/Light (MX1104) or HOBO MX 4-Channel Analog (MX1105) Data Logger
1. Insert the plug into an analog sensor port on the logger. Turn the sensor plug one-quarter clockwise until it stops.
Using the diagram on the logger housing shown above as reference, insert the plug into the analog sensor port.
Turn the plug one­quarter clockwise.
2. In HOBOmobile, connect to the logger and tap Configure. The sensor type will be selected automatically. Set any other logger settings and tap Start. See the HOBOmobile User’s Guide at https://www.onsetcomp.com/manuals/17841-hobomobile-users-guide for details on configuring the logger.
The flat part of the plug will be facing up when properly installed.
1-800-LOGGERS (564-4377) • 508-759-9500 www.onsetcomp.com/support/contact
© 2019 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOmobile are registered trademarks of Onset Computer Corporation. All other trademarks are the property of their respective companies.
23983-A
Loading...