Onset CX402-VFC215, CX402-T205, CX402-VFC205, CX402-T215, CX402-T230 Series Manual

...
InTemp® CX400 Series Temperature Logger Manual
InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed to meet World Health Organization (WHO) and Centers for Disease Control and Prevention (CDC) guidelines, this logger is ideal for clinical applications, such as vaccine storage and pharmaceutical manufacturing. This Bluetooth® Low Energy-enabled logger is designed for wireless communication with a mobile device. Using the InTemp app, you can easily configure the logger with one of four preset configurations designed for ambient storage, clinical refrigerator, freezer, or refrigerator monitoring or you can set up a custom configuration for other applications. You can also quickly perform daily logger checks, download reports, and monitor tripped alarms. Use the built-in LCD screen to check the current temperature, daily maximum or minimum temperatures, logging status, battery use, and more. You also have the option to use InTempConnect™ to track logger deployments and automatically upload logger data to build custom reports for further analysis. The CX402 model uses a 2- or 4-meter probe and is available with a 5, 15, or 30 mL glycol bottle (bottle holder included). It also contains an internal sensor for ambient temperature monitoring. The CX403 model is available with an internal sensor only.
InTemp CX400 Temperature Loggers
Models:
CX402-T205 & CX402-VFC205, 2 meter probe and 5 mL glycol bottle
CX402-T215 & CX402-VFC215, 2 meter probe and 15 mL glycol bottle
CX402-T230 & CX402-VFC230, 2 meter probe and 30 mL glycol bottle
CX402-T405 & CX402-VFC405,
4 meter probe and 5 mL glycol bottle
CX402-T415 & CX402-VFC415, 4 meter probe and 15 mL glycol bottle
CX402-T430 & CX402-VFC430 4 meter probe and 30 mL glycol bottle
CX402-T2M, CX402-B2M, &
CX402-VFC2M, 2 meter probe
CX402-T4M, CX402-B4M, &
CX402-VFC4M, 4 meter probe
CX403
Included Items:
Double-sided tape (for models with glycol bottles)
Two AAA 1.5 V alkaline batteries
Battery door and screw
NIST Certificate of Calibration
Required Items:
InTemp app
Device with iOS or Android™
and Bluetooth
Specifications
External Temperature Probe
Range -40° to 100°C (-40° to 212°F)
Accuracy CX40x-Txxx and CX402-BxM:
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
Response Time Probe only: 1 minute 30 seconds typical to 90% in air moving 1 m/s
NIST Calibration CX40x-Txx and CX402-BxM: Single point NIST calibration, probe and
Cable Length 2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
Probe Dimensions Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches)
Ambient Temperature Sensor
Range -30° to 70°C (-22° to 158°F)
Accuracy CX40x-Txxx and CX402-BxM:
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
Response Time 15 minutes typical to 90% in air moving 1 m/s
Logger
Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F) ±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F) ±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)
CX40x-VFCxxx: ±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F) ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) ±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F)
With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s
logger body CX40x-VFCxxx: Single point NIST calibration, probe only
long, 3.18 mm (0.125 inches) diameter
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F) ±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
CX40x-VFCxxx: ±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F) ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F) ±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F)
19996-H
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
InTemp CX400 Series Temperature Logger Manual
Specifications (continued)
Logger Operating Range -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate 1 second to 18 hours
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery Type Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life 1 year, typical with logging interval of 1 minute. Faster logging intervals,
Memory 128 KB (84,650 measurements, maximum)
Full Memory Download Time
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or
Dimensions 9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
Weight 90.2 g (3.18 oz)
Environmental Rating IP54
Passed RTCA DO160G, part 21H
remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.
Approximately 60 seconds; may take longer the farther the device is from the logger
go blank in temperatures outside this range
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
See last page
Logger Components and Operation
Alarm LED
LCD
Audible alarm speaker
Start button
External temperature probe (CX402 models only)
Start Button: Press this button for 1 second to start the logger when it is configured to start “on button push.”
Mute or Next Button: Press this button for 1 second to mute a beeping alarm (see Logger Alarms). For CX402 loggers, press this button for 1 second to switch between the external probe and ambient internal sensor temperature readings.
Magnets: There are four magnets on the back of the logger (not shown in diagram) for mounting.
Temperature Sensor: This internal sensor measures ambient temperature.
External Temperature Probe: This is the probe that is plugged into the logger for measuring temperature (CX402 models only).
Audible Alarm Speaker: This is the speaker for the audible alarm that beeps when an alarm is tripped or the external probe is removed (if applicable).
Temperature sensor (behind rectangular plug)
One of four magnets (on back of logger)
Mute or Next button
Alarm LED: This LED blinks every 5 seconds when an alarm is tripped or the external probe is removed (if applicable). See Logger Alarms.
LCD: This screen shows the latest temperature reading and other status information. The LCD screen refreshes at the same rate as the logging interval. The example shows all symbols illuminated on the LCD screen followed by a table with descriptions of each symbol.
LCD Symbol Description
An alarm has tripped because the temperature reading is outside the specified range
The logger has been configured to perform daily or twice daily checks (twice daily is shown in this example), but no checks have been performed yet.
The once daily or twice daily logger check (twice in this example) has been performed.
This indicates how much memory has been used for the current deployment. In this example, approximately 40 percent of the memory has been used.
This shows the approximate battery power remaining.
1-800-LOGGERS 2