InTemp™ CX Temperature Logger (CX40x) Manual
InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed to
meet World Health Organization (WHO) and Centers for Disease Control and Prevention (CDC)
guidelines, this logger is ideal for clinical applications, such as vaccine storage and pharmaceutical
manufacturing. This Bluetooth® Low Energy-enabled logger is designed for wireless
communication with a mobile device. Using the InTemp app, you can easily configure the logger
with one of four preset configurations designed for ambient storage, clinical refrigerator, freezer,
or refrigerator monitoring or you can set up a custom configuration for other applications. You
can also quickly perform daily logger checks, download reports, and monitor tripped alarms. Use
the built-in LCD screen to check the current temperature, daily maximum or minimum
temperatures, logging status, battery use, and more. The CX402 model uses a 2- or 4-meter probe
and is available with a 5, 15, or 30 mL glycol bottle (bottle holder included). It also contains an
internal sensor for ambient temperature monitoring. The CX403 model is available with an
internal sensor only.
Specifications
InTemp CX Temperature
Logger
Models:
• CX402-T205, 2 meter probe
and 5 mL glycol bottle
• CX402-T215, 2 meter probe
and 15 mL glycol bottle
• CX402-T230, 2 meter probe
and 30 mL glycol bottle
• CX402-T405, 4 meter probe
and 5 mL glycol bottle
• CX402-T415, 4 meter probe
and 15 mL glycol bottle
• CX402-T430, 4 meter probe
and 30 mL glycol bottle
• CX402-T2M, 2 meter probe
• CX402-T4M, 4 meter probe
• CX403
Included Items:
• Double-sided tape (for
models with glycol bottles)
• Two AAA 1.5 V alkaline
batteries
• Battery door and screw
• NIST Certificate of
Calibration
Required Items:
• InTemp app
• Device with iOS 8.3–9.2 or
Android™ 4.4, 5.0, or 5.1
and Bluetooth 4.0
External Temperature Probe
Range -40° to 100°C (-40° to 212°F)
Accuracy ±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
Response Time Probe only: 1 minute 30 seconds typical to 90% in air moving 1 m/s
Cable Length 2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
Probe Dimensions Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches)
Ambient Temperature Sensor
Range -30° to 70°C (-22° to 158°F)
Accuracy ±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
Response Time 15 minutes typical to 90% in air moving 1 m/s
Logger
Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate 1 second to 18 hours
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery Type Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life 1 year, typical with logging interval of 1 minute. Faster logging intervals,
Memory 128 KB (84,650 measurements, maximum)
Full Memory Download
Time
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F)
±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)
With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s
With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s
long, 3.18 mm (0.125 inches) diameter
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
remaining connected with the InTemp app, excessive report generation,
numerous audible alarms, and paging all impact battery life.
Approximately 60 seconds; may take longer the farther the device is
from the logger
19996-C

InTemp CX Temperature Logger (CX40x) Manual
Specifications (continued)
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or
Dimensions 9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
Weight 90.2 g (3.18 oz)
Environmental Rating IP54
Logger Components and Operation
Alarm LED
LCD
Audible alarm
speaker
Start button
External
temperature
probe (CX402
models only)
Start Button: Press this button for 1 second to start the logger
when it is configured to start “on button push” (see Configuring
the Logger).
Mute or Next Button: Press this button for 1 second to mute a
beeping alarm (see Logger Alarms). For CX402 loggers, press
this button for 1 second to switch between the external probe
and ambient internal sensor temperature readings.
Magnets: There are four magnets on the back of the logger (not
shown in diagram) for mounting.
Temperature Sensor: This internal sensor measures ambient
temperature.
External Temperature Probe: This is the probe that is plugged
into the logger for measuring temperature (CX402 models
only).
Audible Alarm Speaker: This is the speaker for the audible
alarm that beeps when an alarm is tripped or the external
probe is removed (if applicable).
Alarm LED: This LED blinks every 5 seconds when an alarm is
tripped or the external probe is removed (if applicable). See
Logger Alarms.
LCD: This screen shows the latest temperature reading and
other status information. The LCD screen refreshes at the same
rate as the logging interval. The example shows all symbols
illuminated on the LCD screen followed by a table with
descriptions of each symbol.
Temperature
sensor (behind
rectangular
plug)
One of four
magnets (on
back of logger)
Mute or Next
button
go blank in temperatures outside this range
The CE Marking identifies this product as complying with all relevant
directives in the European Union (EU).
See last page
LCD Symbol Description
An alarm has tripped because the temperature
reading is outside the specified range
The logger has been configured to perform daily or
twice daily checks (twice daily is shown in this
example), but no checks have been performed yet.
The once daily or twice daily logger check (twice in
this example) has been performed.
This indicates how much memory has been used
for the current deployment. In this example,
approximately 40 percent of the memory has been
used.
The shows the approximate battery power
remaining.
The logger is currently logging.
The logger is currently connected to a phone or
tablet via BlueTooth. The more bars there are, the
stronger the signal.
The logger is waiting to be started. Press and hold
the Start button for 3 seconds to start the logger.
This is an example of a temperature reading from
an external probe.
This is an example of a temperature reading from
the internal sensor.
This is an example of a minimum temperature,
which is the lowest probe temperature reading
from the day (CX402 models) or the lowest
ambient temperature reading (CX403 model).
This is an example of a maximum temperature,
which is the highest probe temperature reading
from the day (CX402 models) or the lowest
ambient temperature reading (CX403 model).
The external probe is not connected to the logger
(if applicable).
and
1-800-LOGGERS 2 www.onsetcomp.com