Onicon FT-3100 Programming Manual

FT-3100 Transmitter
Inline Electromagnetic Flow Meter
Programming Guide
ONICON
Flow and Energy Measurement
TABLE OF CONTENTS
1.0 INTRODUCTION .........................................................................................................................5
1.1 Purpose of this Guide ...................................................................................................................................... 5
1.2 Display and User Interface ............................................................................................................................. 5
1.3 Accessing the Programming Menus .......................................................................................................... 6
2.0 QUICK START MENU ..................................................................................................................6
2.1 MENU - QUICK START ..................................................................................................................................... 7
Empty Pipe Detector [E.P.Detect=ON] .................................................................................................... 7
Empty Pipe Threshold [R max=kohm XXXX] ......................................................................................... 7
Scaled Pulse Output 2 [Pls2=Gal X.XXXX] ............................................................................................. 7
Output 2 Pulse Duration [Tpls2=ms X.XXXX] ........................................................................................ 8
Analog Output 1 Full Scale Value [A1S=Gal/m XXXXXX] ................................................................. 8
3.0 MAIN MENU ............................................................................................................................... 9
3.1 MENU 1 - SENSOR ............................................................................................................................................ 9
Sensor Body Model [S.Model=XXX] ......................................................................................................... 9
Liner Material [Lining=XXXXXX] ...............................................................................................................10
Unit Type [U.type=XXXXXX] .......................................................................................................................10
Diameter [Diam.=XXX] .................................................................................................................................10
KA [KA=±XX.XXX] ..........................................................................................................................................10
KZ [KZ=±XXXX] ..............................................................................................................................................10
Empty Pipe Detector [E.P.Detect=ON] ..................................................................................................11
Empty Pipe Threshold [R max=kohm XXXX] .......................................................................................11
Electrode Cleaning [El. Cleaning=MIN] .................................................................................................11
Sensor Cable Length [S.cable=ft XXX] ...................................................................................................12
Alarm Delay Interval [S.err.delay=m XXX] ............................................................................................12
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.2 MENU 2 - UNITS .............................................................................................................................................13
Diameter Unit Type [Diam.=in] .................................................................................................................13
Remote Mount Cable Unit Type [S.cable=ft] ......................................................................................13
Flow Rate Unit Type [FR unit=IMPERIAL] .............................................................................................13
Pulse 1 Unit Type [Pl1 unit=IMPERIAL] ..................................................................................................14
Pulse 2 Unit Type [Pl2 unit=IMPERIAL] ...................................................................................................14
Total Forward Flow Totalizer Unit Type [T+ unit=IMPERIAL] ..........................................................14
Total Forward Flow Totalizer Units [T+ unit=Gal] ...............................................................................14
Total Forward Flow Totalizer Decimal Point Position [T+ D.P.=X] ................................................14
Partial Forward Flow Totalizer Unit Type [P+ unit=IMPERIAL] .......................................................14
Partial Forward Flow Totalizer Units [P+ unit=Gal] ............................................................................15
Partial Forward Flow Totalizer Decimal Point Position [P+ D.P.=X] .............................................15
Total Reverse Flow Totalizer Unit Type [T- unit=IMPERIAL] ............................................................15
Total Reverse Flow Totalizer Units [T- unit=Gal] .................................................................................15
Total Reverse Flow Totalizer Decimal Point Positiion [T- D.P.=X] .................................................15
Partial Reverse Flow Totalizer Unit Type [P- unit=IMPERIAL] .........................................................15
Partial Reverse Flow Totalizer Units [P- unit=Gal] ..............................................................................15
Partial Reverse Flow Totalizer Decimal Point Position [P- D.P.=X] ...............................................16
Temperature Unit Type [Temp. unit= °F] ................................................................................................ 16
Mass Units Enable [Mass units=OFF] ......................................................................................................16
Specic Gravity Coecient [Sg=kg/dm3 X.XXXX] ..............................................................................16
ONICON Incorporated 727.447.6140 Page 2 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.3 MENU 3 - SCALES ...........................................................................................................................................17
Full Scale Flow Rate 1 [FS1= Gal/m XXX.X] ...........................................................................................17
Full Scale Flow Rate 2 [FS2 = Gal/m XXX.X] ..........................................................................................18
Scaled Pulse Output 1 [Pls1 = Gal X.XXXX] ...........................................................................................18
Output 1 Pulse Duration [Tpls1 = ms X.XXXX] .....................................................................................19
Output 1 Full Scale Frequency [Frq1 = Hz X.XXXXX] ........................................................................19
Scaled Pulse Output 2 [Pls2 = Gal X.XXXX] ...........................................................................................19
Output 2 Pulse Duration [Tpls2 = ms X.XXXX] .....................................................................................20
Output 2 Full Scale Frequency [Frq2 = Hz X.XXXXX] ........................................................................20
3.4 MENU 4 - MEASURE.......................................................................................................................................21
Flow Signal Damping [Damping=SMART] ............................................................................................21
Low Flow Cut-o [Cut-o = % XXX] ........................................................................................................22
Autorange [Autorange=OFF]......................................................................................................................22
High Immunity Input Filter [H.im.imp.=OFF] ........................................................................................22
3.5 MENU 5 - ALARMS ........................................................................................................................................23
Maximum Forward Flow Alarm Threshold [Max+ = Gal/m OFF] ................................................23
Maximum Reverse Flow Alarm Threshold [Max - = Gal/m OFF] .................................................23
Minimum Forward Flow Alarm Threshold [Min + = Gal/m OFF] ................................................23
Minimum Reverse Flow Alarm Threshold [Min - = Gal/m OFF] ..................................................24
Alarm Hysteresis Threshold [Hysteresis=%XXX] ................................................................................24
mA Output Value When in Alarm State [mA v.alarm=%XXX] .......................................................24
Frequency Output Value When in Alarm State [Hz v.alarm = %XXX} ........................................25
3.6 MENU 6 - INPUTS...........................................................................................................................................26
Total Forward Flow Totalizer Reset Enable [T+reset=ON] .............................................................26
Partial Forward Flow Totalizer Reset Enable [P+reset=OFF] .........................................................26
Total Reverse Flow Totalizer Reset Enable [T-reset=OFF] ...............................................................26
Partial Reverse Flow Totalizer Reset Enable [P-reset=OFF] ............................................................26
Disable Totalizers via Pulse Input [Meas. Lock=OFF] ....................................................................... 26
Activate Autozero Calibration via Pulse Input [Calibration=OFF] ...............................................27
Change Flow Range via Pulse Input [Range change=OFF] ...........................................................27
3.7 MENU 7 - OUTPUTS ......................................................................................................................................28
Digital Output 1 Function Select [Out1=XXXXXX] ............................................................................28
Digital Output 2 Function Select [Out2=XXXXXX] ............................................................................28
mA Output 1 Function & Scaling [Out mA1=4_20+/-] ...................................................................29
Analog Output 1 Full Scale Value [A1S=Gal/m XXXXXX] ...............................................................29
3.8 MENU 9 - DISPLAY .........................................................................................................................................30
Display Language [Language=GB] ..........................................................................................................30
Contrast Adjustment [Contrast=X] .........................................................................................................30
Display/User Interface Inactivity Time Interval [Disp. times=sXXX] ...........................................30
Display Refresh Frequency [D.rate=Hz X].............................................................................................30
Display Lock [Disp. lock=OFF]...................................................................................................................31
Enable Partial Totalizers [Part. tot.=OFF] ...............................................................................................31
Enable Negative Totalizers [Neg. tot. = OFF] ....................................................................................... 31
Enable Net Totalizers [Net tot.=OFF] .....................................................................................................32
Enable Quick Start Menu [Quick start=ON] ........................................................................................32
Default Display Page [Disp. fn = X] ..........................................................................................................33
ONICON Incorporated 727.447.6140 Page 3 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.9 MENU 11 - FUNCTIONS ..............................................................................................................................34
Reset Total Forward (+) Totalizer [T + reset] .......................................................................................34
Reset Partial Forward (+) Totalizer [P + reset] ....................................................................................34
Reset Total Reverse (-) Totalizer [T - reset] ........................................................................................... 34
Reset Partial Reverse (-) Totalizer [P - reset] ........................................................................................34
Load Factory Default Sensor Body Settings [Load sens.f.def] ......................................................34
Load Factory Default Transmitter Settings [Load Conv.f.def] .......................................................34
3.10 MENU 12 - DIAGNOSTIC ............................................................................................................................35
Self Diagnostic Test [Self test] ................................................................................................................... 35
Test Display [Display test] ........................................................................................................................... 35
Simulate Flow Rate [Flow sim. = OFF] ...................................................................................................35
Firmware Information [Firmware info] ...................................................................................................35
Serial Number [S/N = XXXXXX] ................................................................................................................35
Working Time [WT = XXXX:XX:XX:XX] ...................................................................................................35
ONICON Incorporated 727.447.6140 Page 4 onicon.com
SECTION 1.0: INTRODUCTION
FT-3100 TRANSMITTER PROGRAMMING GUIDE
1.1 PURPOSE OF THIS GUIDE
1.2 DISPLAY AND USER INTERFACE
Gal/m
2 ALARM(S)
ONICON
Flow and Energy Measurement
+0.00
SHORT PUSH LONG PUSH
The purpose of this guide is to provide programming information for the FT-3100
transmitter congured with the following options:
• One (1) analog output
• Two (2) digital outputs
• One (1) digital input
It does not include information on serial communications or data logging.
The FT-3100 transmitter is equipped with a lighted graphic display and 3-button user interface as shown below.
0
Briefly press the middle button to advance to the next menu page.
ENTER
ESC
The ENTER button is used in programming.
STATUS ICONS
Icon Description Icon Description
Empty Pipe Signal Error
Low Flow Alarm Excitation Error
High Flow Alarm General Alarm (Flashing)
Flow Rate Overow Flow Rate Simulation (Flashing)
Pulse 1 Overow Pulse 2 Overow
Generic Alarm (Flashing) Calibration (Flashing)
A multicolored LED inside the eld wiring compartment provides
addition information on the operating status.
Red LED: Alarm
Blue LED: USB comm enabled
Green LED: Normal Operation
ONICON Incorporated 727.447.6140 Page 5 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
ACCESS CODE Code : 4
999999990
*******
QUICK START
E. P. Detect=
R max=kohm Pls2=Gal Tpls2=ms A1S=Gal/m Main Menu
ON
0220
1000.0
0500.0
004500
2 ALARM(S)
0
+0.00
Gal/m
1.3 ACCESSING THE PROGRAMMING MENUS
Access to the programming menus is password protected. The factory default access code is 4********. The three user interface push-button functions described below change when operating in the program mode.
Push Button Short Press
(<1 second)
• Increments the numeric value or selected parameter
• Returns to the previous menu item
Moves the cursor to the right on input elds
• Advances to the next menu item
• Changes the process data display
• Enters or leaves the selected function
ENTER
ESC
SECTION 2.0: QUICK START MENU
• Enters the program mode
• Cancels the selected function in progress
The most commonly used programming functions are available in the Quick Start
menu. The menu can be accessed by briey pressing the
the access code 4*******. Additional programming options are accessed via the Main Menu.
Long Press
(>1 second)
• Decrements the numeric value
• Advances to the next menu item
Moves the cursor to the left on input elds
• Returns to the previous menu item
• Exits the current menu
Conrms the selected function to enable the
totalizer reset request, when enabled.
key and entering
ENTER
ESC
1. From the main screen, quickly press the Enter key.
2. Enter the access code.
3. Enter into the Quick Start menu.
ONICON Incorporated 727.447.6140 Page 6 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
2.1 MENU - QUICK START
Empty Pipe Detector [E.P.Detect = ON]
Empty Pipe Threshold [R max = kohm XXXX]
Empty Pipe Detector [E.P.Detect = ON]
Empty Pipe Threshold [R max = kohm XXXX]
Scaled Pulse Out 2 Volume [Pls2 = Gal XXX.XXX]
Output 2 Pulse Duration [Tpls2 = XXXX.X]
Analog Output 1 Full Scale Value [A1S = Gal/m XXXXXX]
MAIN MENU
This setting enables or disables the empty pipe detector.
The Empty Pipe Detector senses when uid is drained from the pipe and the ow sensing electrodes are exposed to air. When this occurs, the displayed ow rate
will drop to zero and the empty pipe icon will be displayed in the upper left­hand side of the display window. If the current (mA) output alarm is active, the milliamp output will also indicate an alarm condition.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or o. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
The empty pipe threshold function is used to set the maximum allowable
resistance value between the ow sensing electrodes in the transmitter. The
transmitter will go into empty pipe alarm when the value is exceeded. The default value is 500. The maximum allowable value is 9,999.
Scaled Pulse Output 2 [Pls2 = Gal X.XXXX]
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the resistance. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
This function is only available when output pulse 2 in OUTPUTS menu (menu 7) [Out2 = XXXXX] is set to Pulses+, Pulses- or Pulses±. When active, Scaled Output Pulse 2 sets the unit of measure and the volume or mass that equates to one
pulse. This pulse weight setting is limited by the diameter of the ow sensor.
The list of units of measure available for selection can be in US/IMPERIAL or SI units depending on the pulse 2 unit type [Pl2 unit = IMPERIAL] setting in UNITS menu (menu 2). In the same way, the list will show mass units, if mass units are enabled [Mass units = OFF] in menu 2. The tables below show the available units of measure.
ONICON Incorporated 727.447.6140 Page 7 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
US/IMPERIAL Volume Units
3
in
Cubic Inch
Gal US Gallon
ttG US Gallon x 10,000
3
ft
Cubic Foot
bbl Standard Barrel
BBL Barrel of Oil
3
hf
Kf
3
Cubic Foot x 100
Cubic Foot x 1,000
KGL US Gallon x 1,000
IGL Imperial Gallon
IKG Imperial Gallon x 1,000
Aft Acre Foot
MGL Mega Gallon (US)
IMG Imperial Mega Gallon
SI/METRIC Volume Units
3
cm
Cubic Centimeter
ml Mililiter
l Liter
3
dm
Cubic Decimeter
dal Decaliter
hl Hectoliter
3
m
Cubic Meter
ML Megaliter
US/IMPERIAL Mass Units
Oz Ounce
Lb Pound
Ton Short Ton (2,000 lbs)
SI/METRIC Mass Units
g Gram
kg Kilogram
t Metric Ton (1,000 kg)
Output 2 Pulse Duration [Tpls2 = ms X.XXXX]
Analog Output 1 Full Scale Value [A1S = Gal/m XXXXXX]
To enter a new value, press enter to access the function and use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
Output 2 pulse duration is only active when scaled pulse output 2 [Pls2= Gal X.XXXX] is active. It is used to set the duration of each pulse. The duration is set in milliseconds with a range of 0.4 to 9999.99.
Note: Pulse durations must match the requirements of the input they are connected to. Very short pulse times may not be counted by the input and long duration pulsed may damage electromechanical registers.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the pulse duration. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the full scale ow value for the analog output.
The maximum allowable ow rate is a function of the diameter of the ow sensor. Regardless of the diameter, the maximum ow rate setting cannot exceed a
velocity of 10 m/s.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 8 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
SECTION 3.0: MAIN MENU
3.1 MENU 1 - SENSOR
The menu can be accessed by briey pressing the
access code 4*******.
1. From Quick Start menu, push
repeatedly
until Main Menu is highlighted.
2. Enter into the main menu.
Sensor Body Model [S Model = XXX]
Liner Material [Lining = XXXXXX]
Unit Type [U type = XXXXXX]
Diameter [Diam. = in XXX]
KA [KA = ± XX.XXX]
KZ [KZ = ± XXXX]
Empty Pipe Detector [E.P.Detect = ON]
Empty Pipe Threshold [R max = kohm XXXX]
Elecrode Cleaning [El. Cleaning = MIN]
Sensor Cable Length [S.Cable = ft XXX]
Alarm Delay Interval [S.err.delay = XXX]
key and entering the
ENTER
ESC
QUICK START
E. P. Detect=
R max=kohm Pls2=Gal Tpls2=ms A1S=Gal/m
Main Menu
MAIN MENU
1-Sensor
2-Units 3-Scales 4-Measure 5-Alarms 6-Inputs
ON
0220
1000.0
0500.0
004500
Sensor Body Model [S.Model = XXX]
Enter the numeric code for the type of sensor body connected to the transmitter.
Body Type Sensor Code
Threaded Steel 011
Threaded Polypropylene 013
Wafer 002
Flanged 004
To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the code as per the table above. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 9 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Liner Material [Lining = XXXXXX]
Unit Type [U.type = XXXXXX]
Diameter [Diam. = in XXX]
Select the sensor body liner material from the following list.
Liner Material* Code
PTFE PTFE
Polypropylene PP
Ebonite HR
*Disregard all liner material options that are not shown in this list.
To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the code as per the table above. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This parameter congures the transmitter for U.S. (imperial) or SI (metric) units.
To change the existing setting, press enter to access the function and use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This must be set to match the nominal diameter of the sensor body connected to the transmitter. If the Unit Type is set to U.S. units, the nominal sizes will be in inches. For SI units, the nominal sizes will be in millimeters.
KA [KA = ± XX.XXX]
KZ [KZ = ± XXXX]
To enter a new diameter, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
KA is one of two coecients of calibration that are derived during the sensor
body factory calibration. Both variables must be programmed in to the
transmitter to ensure accurate ow measurement. The KA value is provided on the certicate of calibration and on a label on the sensor body.
To enter a new coecient, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
KZ is the second of two coecients of calibration that are derived during the
sensor body factory calibration. Both variables must be programmed in to the
transmitter to ensure accurate ow measurement. The KZ value is provided on the certicate of calibration and on a label on the sensor body.
To enter a new coecient, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 10 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Empty Pipe Detector [E.P.Detect = ON]
Empty Pipe Threshold [R max = kohm XXXX]
Electrode Cleaning [El. Cleaning = MIN]
This setting enables or disables the empty pipe detector.
The Empty Pipe Detector senses when uid is drained from the pipe and the ow sensing electrodes are exposed to air. When this occurs, the displayed ow rate
will drop to zero and the empty pipe icon will be displayed in the upper left-hand side of the display window. If the current (mA) output alarm is active the milliamp output will also indicate an alarm condition.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
The empty pipe threshold function is used to set the maximum allowable
resistance value between the ow sensing electrodes in the transmitter. The
transmitter will go into empty pipe alarm when the value is exceeded. The default value is 500. The maximum allowable value is 9,999.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the resistance. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
When enabled, the electrode cleaning function transmits a depolarizing signal to
the electrodes in between ow measurement cycles. This function is enabled by
default whenever the empty pipe alarm is active. The default setting, when active,
is AVG. It should only be used with uid conductivities >100µS/cm.
Electrode Cleaning Settings Code
O OFF
Minimum MIN
Average AVG
Maximum MAX
To change the existing setting, press enter to access the function and use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 11 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Sensor Cable Length [S.cable = ft XXX]
Enter the remote transmitter cable length. The allowable cable length is a
function of the uid conductivity with a maximum allowable cable length of 164ft
(50m). Use of the remote mount transmitter option with conductivity levels of
less the 15µS/cm is not recommended. Use the table below to determine the
maximum cable length for your installation.
Alarm Delay Interval [S.err.delay = m XXX]
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the length. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
This function delays the activation of empty pipe, excitation error and signal error alarms. The range of the delay interval is from 0 to 500 measurement cycles. For meters up to 10” (250mm) in diameter this equates to 0 to 10 seconds. For 12” to 20” (300 to 500mm) meters, the interval range is from 0 to 25 seconds.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the interval. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 12 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.2 MENU 2 - UNITS
Diameter Unit Type [Diam = in]
Remote Mount Cable Unit Type [S.cable = ft]
Flow Rate Unit Type [FR unit = IMPERIAL]
Pulse 1 Unit Type [Pl1 unit = IMPERIAL]
Pulse 2 Unit Type [Pl2 unit = IMPERIAL]
Total Forward Flow Totalizer Unit Type [T+unit = IMPERIAL]
Total Forward Flow Totalizer Units [T+unit = Gal]
Total Forward Flow Totalizer Decimal Point Position [T+D.P. = X]
Partial Forward Flow Totaliazer Unit Type [P+unit = IMPERIAL]
Partial Forward Flow Totalizer Units [P+unit = Gal]
Partial Forward Flow Totalizer Decimal Point Position [P+D.P. = X]
Total Reverse Flow Totalizer Unit Type [T-unit = IMPERIAL]
Total Reverse Flow Totalizer Units [T-unit = Gal]
Total Reverse Flow Totalizer Decimal Point Position [T-D.P. = X]
Partial Reverse Flow Totalizer Unit Type [P-unit = IMPERIAL]
Partial Reverse Flow Totalizer Units [P-unit = Gal]
Partial Reverse Flow Totalizer Decimal Point Position [P-D.P. = X]
Temperature Unit Type [Temp.unit = °F]
Mass Units Enable [Mass units = OFF]
Specic Gravity Coecient [Sg = kg/dm
3
X.XXXX]
Diameter Unit Type [Diam. = in]
Remote Mount Cable Unit Type [S.cable = ft]
Flow Rate Unit Type [FR unit = IMPERIAL]
This function sets the nominal diameter of the sensor body for U.S. (imperial) or SI (metric) units. This setting changes the list of nominal diameters in SENSOR menu (menu 1) [Diam. = ft XXX] from U.S to SI units.
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the cable length measurement units for U.S. (imperial) or SI (metric). This setting changes the way cable lengths are entered in SENSOR menu (menu 1) [S.cable = ft XXX].
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of ow measurement units for U.S (imperial) or SI
(metric).
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 13 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Pulse 1 Unit Type [Pl1 unit = IMPERIAL]
Pulse 2 Unit Type [Pl2 unit = IMPERIAL]
Total Forward Flow Totalizer Unit Type [T+ unit = IMPERIAL]
This function sets the list of totalizer measurement units for U.S. (imperial) or SI (metric) for output 1. The function is only active when the output 1 function selection [Out1 = XXXXXX] is set for scaled pulse
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of totalizer measurement units for U.S. (imperial) or SI (metric) for output 2. The function is only active when the output 2 function selection [Out2 = XXXXXX] is set for scaled pulse
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of forward ow totalizer measurement units for U.S
(imperial) or SI (metric).
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Total Forward Flow Totalizer Units [T+ unit = Gal]
Total Forward Flow Totalizer Decimal Point Position [T+ D.P. = X]
Partial Forward Flow Totalizer Unit Type [P+ unit = IMPERIAL]
This function sets measurement unit for the forward ow totalizer.
To change the existing setting, press enter to access the function. Use the up arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the position of the decimal point when displaying the forward
ow total.
To move the decimal point, press enter to access the function. Use the up and/ or right arrow keys to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of partial forward ow totalizer measurement units for
U.S (imperial) or SI (metric).
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 14 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Partial Forward Flow Totalizer Units [P+ unit = Gal]
Partial Forward Flow Totalizer Decimal Point Position [P+ D.P. = X]
Total Reverse Flow Totalizer Unit Type [T- unit = IMPERIAL]
Total Reverse Flow Totalizer Units [T- unit = Gal]
This function sets measurement unit for the partial forward ow totalizer.
To change the existing setting, press enter to access the function. Use the up arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the position of the decimal point when displaying the forward
ow total.
To move the decimal point, press enter to access the function. Use the up and/ or right arrow keys to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of reverse ow totalizer measurement units for U.S
(imperial) or SI (metric).
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets measurement unit for the reverse ow totalizer.
To change the existing setting, press enter to access the function. Use the up arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Total Reverse Flow Totalizer Decimal Point Position [T- D.P. = X]
Partial Reverse Flow Totalizer Unit Type [P- unit = IMPERIAL]
Partial Reverse Flow Totalizer Units [P- unit = Gal]
This function sets the position of the decimal point when displaying the reverse
ow total.
To move the decimal point, press enter to access the function. Use the up and/ or right arrow keys to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the list of partial reverse ow totalizer measurement units for
U.S (imperial) or SI (metric).
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets measurement unit for the partial reverse ow totalizer.
To change the existing setting, press enter to access the function. Use the up arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 15 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Partial Reverse Flow Totalizer Decimal Point Position [P- D.P. = X]
Temperature Unit Type [Temp. unit = °F]
Mass Units Enable [Mass units = OFF]
Specic Gravity Coecient
[Sg = kg/dm3 X.XXXX]
This function sets the position of the decimal point when displaying the reverse
ow total.
To move the decimal point, press enter to access the function. Use the up and/ or right arrow keys to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the measurement units for the on-board temperature sensor.
To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Enables or disables the mass measurement function. When enabled,
measurement unit lists for ow rates and totals will be in mass units.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
This function is only enabled when mass units are enabled. The static value
entered here will be used to convert volumetric ow into mass ow.
To enter a new value, press enter to access the function. Use the right arrow key
to move the cursor and the up arrow key to change the coecient. Press enter to
accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 16 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.3 MENU 3 - SCALES
Full Scale Flow Rate 1 [FS1= Gal/m XXX.X]
Full Scale Flow Rate 1 [FS1 = Gal/m XXX.X]
Full Scale Flow Rate 2 [FS2 = Gal X.XX.X]
Pulse Output 1 Unit and Multiplier [Pls1 = X.XXXXX]
Pulse Output 1 Duration [Tpls1 = ms X.XXXXX]
Pulse Output 2 Unit and Multiplier [Pls2 = Gal X.XXXXX]
Pulse Output 2 Duration [Tpls2 = ms X.XXXXX
Frequency Output 1 Full Scale Frequency [Frq1 = Hz X.XXXXX]
Frequency Output 2 Full Scale Frequency [Frq2 = Hz X.XXXXX]
The full scale ow rate 1 function sets the unit of measure and time base for the meter and the maximum ow rate for range 1. This eects the ow rate display
and the alarm function. It also sets the unit of measure and time base for the analog output.
The maximum allowable ow rate is a function of the diameter of the ow sensor. Regardless of the diameter, the maximum ow rate setting cannot exceed a velocity of 10 m/s. The following rules apply to the maximum ow rate setting. These may also aect the selection of the unit of measure and the time base.
• The largest numeric value representing ow is 9999. Larger values use units
with multipliers.
• The full scale ow rate setting cannot exceed 10 m/s
• The full scale ow rate must be greater than 0.4 m/s
US/IMPERIAL Volume Units
3
in
Cubic Inch
Gal US Gallon
ttG US Gallon x 10,000
3
ft
Cubic Foot
bbl Standard Barrel
BBL Barrel of Oil
3
hf
Kf
3
Cubic Foot x 100
Cubic Foot x 1,000
KGL US Gallon x 1,000
IGL Imperial Gallon
IKG Imperial Gallon x 1,000
Aft Acre Foot
MGL Mega Gallon (US)
IMG Imperial Mega Gallon
The list of units of measure available for selection can be in US/IMPERIAL or
SI units depending on the ow rate unit type [FR unit = IMPERIAL] selected in
UNITS menu (menu 2). In the same way, the list will show mass units, if mass units are enabled [Mass units= OFF] in menu 2.
The tables below show the available units of measure.
Time Base
s Seconds
m Minutes
h Hours
US/IMPERIAL Mass Units
Oz Ounce
Lb Pound
Ton Short Ton (2,000 lbs)
d Days
SI/METRIC Volume Units
3
cm
Cubic Centimeter
ml Mililiter
l Liter
3
dm
Cubic Decimeter
dal Decaliter
hl Hectoliter
3
m
Cubic Meter
ML Megaliter
SI/METRIC Mass Units
g Gram
kg Kilogram
t Metric Ton (1,000 kg)
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 17 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Full Scale Flow Rate 2 [FS2 = Gal/m XXX.X]
Scaled Pulse Output 1 [Pls1 = Gal X.XXXX]
The full scale ow rate 2 function sets the maximum ow rate for range 2, when
the autorange function [Autorange = ON/OFF] in MEASURE menu (menu 4) is turned on. Range 2 uses the unit of measure and time base setting from full scale
ow rate 1.
The maximum allowable ow rate is a function of the diameter of the ow sensor. Regardless of the diameter, the maximum ow rate setting cannot exceed a velocity of 10 m/s. The following rules apply to the maximum ow rate setting. These may also aect the selection of the unit of measure and the time base.
• The largest numeric value representing ow is 9999. Larger values use units
with multipliers.
• The full scale ow rate setting cannot exceed 10 m/s
• The full scale ow rate must be greater than 0.4 m/s
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
This function is only available when output pulse 1 in OUTPUTS menu (menu 7) [Out1 = XXXXX] is set to Pulses+, Pulses- or Pulses±. When active, scaled output pulse 1 sets the unit of measure and the volume or mass that equates to one
pulse. This pulse weight setting is limited by the diameter of the ow sensor.
The list of units of measure available for selection can be in US/IMPERIAL or SI units depending on the pulse 1 unit type [Pl1 unit = IMPERIAL] setting in UNITS menu (menu 2). In the same way, the list will show mass units, if mass units are enabled [Mass units= OFF] in menu 2. The tables below show the available units of measure.
US/IMPERIAL Volume Units
3
in
Cubic Inch
Gal US Gallon
ttG US Gallon x 10,000
3
ft
Cubic Foot
bbl Standard Barrel
BBL Barrel of Oil
3
hf
Kf
3
Cubic Foot x 100
Cubic Foot x 1,000
KGL US Gallon x 1,000
IGL Imperial Gallon
IKG Imperial Gallon x 1,000
Aft Acre Foot
MGL Mega Gallon (US)
IMG Imperial Mega Gallon
SI/METRIC Volume Units
3
cm
Cubic Centimeter
ml Mililiter
l Liter
3
dm
Cubic Decimeter
dal Decaliter
hl Hectoliter
3
m
Cubic Meter
ML Megaliter
US/IMPERIAL Mass Units
Oz Ounce
Lb Pound
Ton Short Ton (2,000 lbs)
SI/METRIC Mass Units
g Gram
kg Kilogram
t Metric Ton (1,000 kg)
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 18 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Output 1 Pulse Duration [Tpls1 = ms X.XXXX]
Output 1 Full Scale Frequency [Frq1 = Hz X.XXXXX]
Output 1 pulse duration is only active when scaled pulse output 1 [Pls1 = Gal X.XXXX] is active. It is used to set the duration of each pulse. The duration is set in milliseconds with a range of 0.4 to 9999.99.
Note: Pulse durations must match the requirements of the input they are connected to. Very short pulse times may not be counted by the input and long duration pulsed may damage electromechanical registers.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the pulse length. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function is only available when output pulse 1 in OUTPUTS menu (menu 7) [Out1 = XXXXX] is set to Freq+, Freq- or Freq±. When active, output 1 full scale frequency sets maximum frequency for the output. This maximum frequency
equals the maximum ow rate set by full scale ow rate 1 [FS1 = Gal/m XXX.X] in
SCALES menu (menu 3).
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the pulse length. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Scaled Pulse Output 2 [Pls2 = Gal X.XXXX]
This function is only available when output pulse 2 in OUTPUTS menu (menu 7) [Out2 = XXXXX] is set to Pulses+, Pulses- or Pulses±. When active, scaled output pulse 2 sets the unit of measure and the volume or mass that equates to one
pulse. This pulse weight setting is limited by the diameter of the ow sensor.
The list of units of measure available for selection can be in US/IMPERIAL or SI units depending on the pulse 2 unit type [Pl2 unit= IMPERIAL] setting in UNITS menu (menu 2). In the same way, the list will show mass units, if mass units are enabled [Mass units = OFF] in menu 2. The tables below show the available units of measure.
The tables shown for scaled pulse output 1 units of measure also apply to this output.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 19 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Output 2 Pulse Duration [Tpls2 = ms X.XXXX]
Output 2 Full Scale Frequency [Frq2 = Hz X.XXXXX]
Output 2 Pulse Duration is only active when scaled pulse output 2 [Pls2 = Gal X.XXXX] is active. It is used to set the duration of each pulse. The duration is set in milliseconds with a range of 0.4 to 9999.99.
Note: Pulse durations must match the requirements of the input they are connected to. Very short pulse times may not be counted by the input and long duration pulsed may damage electromechanical registers.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the pulse length. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function is only available when output pulse 2 in OUTPUTS menu (menu 7) [Out2 = XXXXX] is set to Freq+, Freq- or Freq±. When active, output 2 full scale frequency sets maximum frequency for the output. This maximum frequency
equals the maximum ow rate set by full scale ow rate 2 [FS2 = Gal/m XXX.X] in
SCALES menu (menu 3).
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the pulse length. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 20 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
FLOWRATE(%)
Example wtih damping to OFF
FILTERS
Damping = OFF
60
50
40
30
20
10
0
0 2 4 6 8 10 12 14 16 18 20 22
IN
>>>>>
.FLOWRATE
OUT
>>>>>
TIME (Sec)
.MEASURE
3.4 MENU 4 - MEASURE
Flow Signal Damping [Damping = SMART]
Flow Signal Damping [Damping=SMART]
Low Flow Cut-o [Cut-o=%XXX]
Autorange [Autorange= OFF]
High Immunity Input Filter [H.im.inp = OFF]
The ow signal damping function allows the user to smooth out the eects of sudden changes in the ow readings on the display and the output. Damping may be applied using the “Smart” lter or by setting xed time interval ltering. It
is recommended to start with no damping.
Smart ltering automatically adapts to changing ow conditions. This lter will provide signicant ltering when large, rapid changes are detected and minimal ltering when ow is more stable. This type of ltering is useful when large changes in ow occur at random intervals.
Time based ltering applies a xed time constant to the ltering algorithm. Longer time constants increase the damping eect, but they also reduce response time. See examples below. This type of ltering is useful when the variations in ow are relatively constant and repeatable. Damping time interval
are 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 & 1000 seconds.
60
Example wtih damping to OFF
50
40
30
FLOWRATE(%)
20
10
0
0 2 4 6 8 10 12 14 16 18 20 22
.FLOWRATE
>>>>>
Damping = OFF
FILTERS
IN
TIME (Sec)
OUT
>>>>>
60
Example wtih damping to 5,0sec.
50
40
30
FLOWRATE(%)
20
.MEASURE
IN
10
0
0 2 4 6 8 10 12 14 16 18 20 22
.FLOWRATE
>>>>>
FILTERS
Damping = from 0.2s to 1000s
TIME (Sec)
OUT
>>>>>
.MEASURE
To enter a new value, press enter to access the function. Use the right arrow key
to move the cursor and the up arrow key to change the lter setting. Press enter
to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 21 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Low Flow Cut-o [Cut-o = % XXX]
Autorange [Autorange = OFF]
This function sets the low ow cut-o for ow measurement. The meter will report all ow measurements below the threshold as zero ow. This cut­o function is necessary to prevent small random noise signals from being reported as ow. It is based on a percentage of the full scale ow rate 1 [FS1=
Gal/m XXX.X] in SCALES menu (menu 3). The allowable settings are between
0 and 25% of full scale. For most applications a value between 0.1 and 1.0% is
recommended.
To enter a new value, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the cut-o percentage.
Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
When autorange is enabled the meter will operate with two dierent ow ranges.
In order to get the best results, range 1 [FS1= Gal/m XXX.X] in SCALES menu (menu 3) must be larger than range 2 [FS2= Gal/m XXX.X].
When the ow rate increases and reaches the 100% of the full scale 1, then the meter automatically switches to full scale 2. When the ow rate decreases to a value on full scale 2 equal to the 90% of full scale.1, then the meter reverts to full
scale 1.
When autorange is enabled, the manual range change function [Range change= OFF] in INPUTS menu (menu 6) cannot be used.
High Immunity Input Filter [H. im. inp. = OFF]
Note: This function does NOT increase the accuracy of the measurement; it only increases the resolution of the 4-20 mA output when the meter is operating at very
low ow rates (e.g. The ow rate of a domestic water distribution with much higher daytime ow versus nighttime ow).
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and exit
the function. Press escape to exit the program mode. T
able of Contents
When enabled, the high immunity input lter will increase ltering at the input circuitry. This ltering is only eective when conductivity levels are greater than 500 µS/cm. It should only be used after all other methods of minimizing noise have been employed as this will aect the ow measurement. The default setting is o.
Note: The use of this lter will reduce ow measurement accuracy.
To change the existing setting, press enter to access the function. Use the up
arrow key to turn the function on or off. Press enter to accept the change and exit
the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 22 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.5 MENU 5 - ALARMS
Maximum Forward Flow Alarm Threshold [Max+ = Gal/m OFF]
Maximum Forward Flow Alarm Threshold [Max + = Gal/m OFF]
Maximum Reverse Flow Alarm Threshold [Max - = Gal/m OFF]
Minimum Forward Flow Alarm Threshold [Min + = Gal/m OFF]
Minimum Reverse Flow Alarm Threshold [Min - = Gal/m OFF]
Hysteresis Threshold [Hysteresis = % XX]
Current Output Alarm State Value [mA v.alarm = % XXX]
Frequency Output Alarm State Value [Hz v.alarm = % XXX]
This function sets the alarm threshold for maximum forward ow. Flow rates
above this threshold will trigger the alarm. The threshold value is entered as a
ow rate in the engineering units shown. The minimum value must be ≥1% of the full scale maximum ow rate of the meter (10m/s or 32.8ft/s velocity). Alarm threshold values are entered in 1% increments of the full scale limit up to 100%.
Example: In a 6” pipe 32.8 ft/s is equal to 2853 GPM. 1% of this value would be
approximately 28.5GPM. Alarm threshold values would be entered in increments of 28.5 GPM.
To enter a value, press enter to access the function. Use the up arrow and right arrow keys change the threshold value. Press enter to accept the change and exit the function. Press escape to exit the program mode. T
able of Contents
Maximum Reverse Flow Alarm Threshold [Max - = Gal/m OFF]
Minimum Forward Flow Alarm Threshold [Min + = Gal/m OFF]
This function sets the alarm threshold for maximum reverse ow. Flow rates
above this threshold will trigger the alarm. The threshold value is entered as a
ow rate in the engineering units shown. The minimum value must be ≥1% of the full scale maximum ow rate of the meter (10m/s or 32.8ft/s velocity). Alarm threshold values are entered in 1% increments of the full scale limit up to 100%.
Example: In a 6” pipe 32.8 ft/s is equal to 2853 GPM. 1% of this value would be
approximately 28.5GPM. Alarm threshold values would be entered in increments of 28.5 GPM.
To enter a value, press enter to access the function. Use the up arrow and right arrow keys change the threshold value. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function sets the alarm threshold for minimum forward ow. Flow rates
below this threshold will trigger the alarm. The threshold value is entered as a
ow rate in the engineering units shown. The minimum value must be ≥1% of the full scale maximum ow rate of the meter (10m/s or 32.8ft/s velocity). Alarm threshold values are entered in 1% increments of the full scale limit up to 100%.
Example: In a 6” pipe 32.8 ft/s is equal to 2853 GPM. 1% of this value would be
approximately 28.5GPM. Alarm threshold values would be entered in increments of 28.5 GPM.
To enter a value, press enter to access the function. Use the up arrow and right arrow keys change the threshold value. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 23 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Minimum Reverse Flow Alarm Threshold [Min - = Gal/m OFF]
Alarm Hysteresis Threshold
[Hysteresis = % XXX]
This function sets the alarm threshold for minimum reverse ow. Flow rates
below this threshold will trigger the alarm. The threshold value is entered as a
ow rate in the engineering units shown. The minimum value must be ≥1% of the full scale maximum ow rate of the meter (10m/s or 32.8ft/s velocity). Alarm threshold values are entered in 1% increments of the full scale limit up to 100%.
Example: In a 6” pipe 32.8 ft/s is equal to 2853 GPM. 1% of this value would be
approximately 28.5GPM. Alarm threshold values would be entered in increments of 28.5 GPM.
To enter a value, press enter to access the function. Use the up arrow and right arrow keys change the threshold value. Press enter to accept the change and exit the function. Press escape to exit the program mode. T
able of Contents
This function is used to dene the ow value where the meter exits the minimum or maximum ow alarm state. When ows drop below the threshold for the maximum ow alarm or rises above the minimum ow alarm by more than
the hysteresis percentage value, the meter exits the alarm state. It is set as a
percentage of the full scale ow rate 1 [FS1 = Gal/m XXX.X] in SCALES menu (menu 3. The range of settings is 0-25% of full scale.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the threshold percentage. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
mA Output Value When in Alarm State
[mA v.alarm = % XXX]
This function establishes the 0/4-20mA output level when the meter is in an alarm condition caused by an empty pipe, ADC error or an open coil indication.
The alarm value is set as a percentage of the 0/4 – 20mA output. The valid range
is 0-125% of 20mA. For example, a setting of 10% would cause the output to go
to 2mA when the meter is in the alarm state. One or more alarm icons will also be displayed in the upper left hand side of the display when this alarm is active.
The default setting is 20% (4mA.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 24 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Frequency Output Value When in Alarm State
[Hz v.alarm = % XXX]
This function establishes the frequency output level for output 1 or output 2 when the meter is in an alarm condition caused by an empty pipe, ADC error or an open coil indication.
This function is only available when either output pulse 1or output pulse 2 in OUTPUTS menu (menu 7) [Out1 = XXXXX] [Out2 = XXXXX] is set to Freq+, Freq­or Freq±.
The alarm value is set as a percentage of the full scale frequency. The valid range
is 0 – 125%. The default setting is 0%. One or more alarm icons will also be
displayed in the upper left hand side of the display when this alarm is active.
The default setting will cause the frequency output to indicate 0 Hz when in the alarm state.
To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 25 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.6 MENU 6 - INPUTS
Total Forward Flow Totalizer Reset Enable [T+ reset = ON]
Partial Forward Flow Totalizer Reset Enable [P+ reset = OFF]
Total Forward Flow Totalizer Reset Enable [T+ reset = ON]
Partial Forward Flow Totalizer Reset Enable [P+ reset = OFF]
Total Reverse Flow Totalizer Reset Enable [T- reset = OFF]
Partial Reverse Flow Totalizer Reset Enable [P- reset = OFF]
Disable Flow Measurement via Pulse Input [Meas. lock = OFF]
Activate Autozero Calibration via Pulse Input [Calibration = OFF]
Change Flow Range via Pulse Input [Range change = OFF]
When enabled, this function allows the forward ow totalizer to be remotely
reset via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
When enabled, this function allows the partial forward ow totalizer to be
remotely reset via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
Total Reverse Flow Totalizer Reset Enable [T- reset = OFF]
Partial Reverse Flow Totalizer Reset Enable [P- reset = OFF]
Disable Totalizers via Pulse Input [Meas. Lock = OFF]
When enabled, this function allows the total reverse ow totalizer to be remotely
reset via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
When enabled, this function allows the partial reverse ow totalizer to be
remotely reset via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
When enabled, this function allows the all ow totalizers to be remotely disabled
via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 26 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
Activate Autozero Calibration via Pulse Input [Calibration = OFF]
Change Flow Range via Pulse Input [Range change = OFF]
When enabled, this function allows the autozero calibration function to be remotely activated via the transmitter’s pulse input.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
When enabled, this function allows the pulse input to be used to remotely
change the ow range to full scale ow rate 2 [FS2 = Gal/m XXX.X] in SCALES
menu (menu 3).
Note: this function cannot be used when the autorange [Autorange = OFF] function
in MEASURE menu (menu 4) is enabled.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 27 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.7 MENU 7 - OUTPUTS
Digital Output 1 Function Select [Out1 = XXXXXX]
Digital Output 2 Function Select [Out2 = XXXXXX]
OFF: Disables output
MAX AL+: Maximum forward ow rate alarm (Energized = no alarm)
MIN AL+: Minimum forward ow rate alarm (Energized = no alarm)
MAX/MIN+: Maximum or minimum forward ow rate alarms (Energized = no alarm)
MAX AL-: Maximum reverse ow rate alarm (Energized = no alarm)
MIN AL-: Minimum reverse ow rate alarm (Energized = no alarm)
MAX/MIN-: Maximum or minimum reverse ow rate alarms (Energized = no alarm)
MAX/MIN +/-: Maximum and minimum ow rate alarm, either direction (Energized = no alarm)
P. EMPTY: Empty pipe alarm output (Energized = no alarm)
HARDW.AL.: Master alarm for open coil, empty pipe or measurement error (Energized = no alarm)
OVERFLOW: Flow out of range alarm (Energized = no alarm)
ALL ALARMS: Activates when any alarm is active (Energized = no alarm)
EXT. COMM.: Allows the pulse output state (high or low) to be set via MODBUS
F.R. SIGN: Flow direction indication (Energized = reverse (-) ow)
*SCALE: Indicates which scale is active (Energized = Scale2)
FREQ+: Frequency output proportional to forward (+) direction ow rate
FREQ-: Frequency output proportional to reverse (-) direction ow rate
FREQ+/-: Frequency output to forward (+) / reverse (-) direction ow rate
PULSES.+: Scaled pulse output for forward (+) direction ow rate
PULSES-: Scaled pulse output for reverse (-) direction ow rate
PULSES+/-: Scaled pulse output for forward (+) / reverse (-) direction ow rate
Digital Output 1 Function Select [Out1 = XXXXXX]
Digital Output 2 Function Select [Out2 = XXXXXX]
mA Output 1 Function & Scaling [Out mA1 = 4_20 +/-]
Analog Output 1 Full Scale Value [A1S = Gal/m XXXXXX]
The available functions for digital output 1 are listed in the table below.
The available functions for digital output 2 are listed in the table below.
Digital Output Functions
*This function is only available when either the autorange [Autorange= OFF] function in menu 4 or the
change ow range via pulse input [Range change= OFF] function INPUTS (MENU 6 are active.
To change the setting, press enter to access the function. Use the up and right arrow keys to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 28 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
mA Output 1 Function & Scaling [Out mA1 = 4_20 +/-]
OUTPUT CURRENT IN mA FOR THE ASSOCIATED % OF FULL SCALE FLOW
mA Output Conguration Options Reverse Flow Value Zero Direct Flow Value
The available functions for current (mA) output 1 are listed in the table below.
There are three elds to modify when setting up an output:
• Zero scale value: 0 or 4mA
• Full scale value: 20 or 22mA
Output function: + = forward (+) ow, - = reverse (-) ow, blank = absolute
ow, -0+ = zero ow center scale
The values corresponding to the scale points are shown in the following chart:
≥ -110% -100% 0% +100% ≥ +110%
0 - 20 + 0 0 0 20 20
0 - 22 + 0 0 0 20 22
4 - 20 + 4 4 4 20 20
4 - 22 + 4 4 4 20 22
0 - 20 - 20 20 0 0 0
0 - 22 - 22 20 0 0 0
4 - 20 - 20 20 4 4 4
4 - 22 - 22 20 4 4 4
0 - 20 20 20 0 20 20
0 - 22 22 20 0 20 22
4 - 20 20 20 4 20 20
4 - 22 22 20 4 20 22
0 - 20 - 0+ 0 0 10 20 20
0 - 22 - 0+ 0 1 11 21 22
4 - 20 - 0+ 4 4 12 20 20
4 - 22 - 0+ 4 4 12 20 22
To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
Analog Output 1 Full
This function sets the full scale ow value for the analog output.
Scale Value [A1S = Gal/m XXXXXX]
The maximum allowable
ow rate is a function of the diameter of the ow sensor.
Regardless of the diameter, the maximum ow rate setting cannot exceed a
velocity of 10 m/s.
To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
ONICON Incorporated 727.447.6140 Page 29 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.8 MENU 9 - DISPLAY
Display Language [Language = GB]
Contrast Adjustment [Contrast = X]
Display Language [Language = GB]
Contrast Adjustment [Contrast = X]
Display/User Interface Inactivity Time Interval [Disp. times = s XXX]
Display Refresh Frequency [D. rate = Hz X]
Enable Partial Totalizers [Part. tot. = OFF]
Enable Negative Totalizers [Neg. tot. = OFF]
Enable Net Totalizers [Net tot. = OFF]
Enable Quick Start Menu [Quick start = ON]
Default Display Page [Disp. fn. = X]
Default Lock [Disp. lock = OFF]
Select English (GB) or Italian (IT) for display text.
To change the existing setting, press enter to access the function. Use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Set the desired contrast level to optimize viewing. The allowable range is 0 to 9.
To change the existing setting, press enter to access the function. Use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
Display/User Interface Inactivity Time Interval [Disp. times = s XXX]
Display Refresh Frequency [D. rate = Hz X]
This function sets the time interval for returning to normal operation following a period of inactivity when using the display/user interface. Allowable settings range from 20 to 255 seconds.
To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode.
Table of Contents
This function sets the display refresh rate. It is used to optimize viewing. It does
not aect the data refresh rate. The settings are 1, 2, 5 or 10 Hz.
To change the existing setting, press enter to access the function. Use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 30 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
2 ALARM(S)
0.000GalP+
0
2 ALARM(S)
0
0.000
Gal/m
GalT+
0.000GalP+
0.204GalT-
0.000GalP-
2 ALARM(S)
0
0.000
Gal/m
GalT+
0.000GalP+
0.204GalT-
0.000GalP-
0
Enable Partial Totalizers [Part. tot. = OFF]
This function enables or disables the display of partial totalizer data. When enabled, the following pages are visible on the display.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
9
5
2
Gal/m
Gal
T+
Gal
P+
2 ALARM(S)
6
P-
Gal
2 ALARM(S)
0.000
0.000
0.000GalT+
0.000GalP+
0.204GalT-
0.000
Enable Negative Totalizers [Neg. tot. = OFF]
This function enables or disables the display of negative totalizer data. When enabled, the following pages are visible on the display.
10
0.204GalT-
2 ALARM(S)
3
Gal/m
Gal
T-
Gal
P-
2 ALARM(S)
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and exit
the function. Press escape to exit the program mode. T
0.204
0.000
5
11
2 ALARM(S)
0.000GalP-
6
P-
Gal
2 ALARM(S)
able of Contents
0.000GalT+
0.000GalP+
0.204GalT-
0.000
ONICON Incorporated 727.447.6140 Page 31 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
2 ALARM(S)
0
+0.000
-0.204
Gal/m
Gal
Gal
TN PN
Enable Net Totalizers [Net tot. = OFF]
Enable Quick Start Menu [Quick start = ON]
This function enables or disables the display of net totalizer data.
Note: This function is only available when negative totalizers are enabled [Neg.
tot.= ON] in DISPLAY menu (menu 9). When enabled, the following page is visible
on the display.
4
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
This function enables or disables the display of the quick start menu page.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
ONICON Incorporated 727.447.6140 Page 32 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
2 ALARM(S)
0
+0.00
Gal/m
0
0
2 ALARM(S)
0
+0.000
-0.204
Gal/m
Gal
Gal
TN PN
2 ALARM(S)
0
0.000
Gal/m
GalT+
0.000GalP+
0.204GalT-
0.000GalP-
3 ALARM(S)
EXCITATION ERROR PIPE EMPTY
2 ALARM(S)
0.000GalP+
2 ALARM(S)
0.127
E1 V
0.168E2 V >10000.0E1 R kΩ >10000.0E2 R kΩ
0
Gal/m
ft/s +0.000
100
75
50
25
0
Default Display Page [Disp. fn = X]
1*
5
9
This function sets the default display page. This page will be displayed rst when
the transmitter exits the programming mode or when power is cycled. There are 13 possible display pages. They are numbered 0 -12. The display pages, with their corresponding numbers, are shown below. Not all pages may be enabled on your meter.
2
Gal/m
Gal
T+
Gal
P+
2 ALARM(S)
6
P-
Gal
2 ALARM(S)
10
0.000
0.000
0.000GalT+
0.000GalP+
0.204GalT-
0.000
3
Gal/m
Gal
T-
Gal
P-
2 ALARM(S)
7*
11
4
0.204
0.000
8*
0.000GalT+
2 ALARM(S)
12*
0*
*Factory default settings pages
Default Lock [Disp. lock = OFF]
0.204GalT-
2 ALARM(S)
To change the existing setting, press enter to access the function. Use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents
This function locks the default display page. When enabled, only the default page is visible on the display.
To change the existing setting, press enter to access the function and use the up
arrow key to turn the function on or off. Press enter to accept the change and
exit the function. Press escape to exit the program mode. Table of Contents
2 ALARM(S)
0.000GalP-
ONICON Incorporated 727.447.6140 Page 33 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.9 MENU 11 - FUNCTIONS
Reset Total Forward (+) Totalizer [T+ reset]
Reset Partial Forward (+) Totalizer [P+ reset]
Reset Total Forward (+) Totalizer [T+ reset]
Reset Partial Forward (+) Totalizer [P+ reset]
Reset Total Reverse (-) Totalizer [T- reset]
Reset Partial Reverse (-) Totalizer [P- reset]
Load Factory Default Sensor Body Settings [Load Sens.f.def]
Load Factory Default Transmitter Settings [Load Conv.f.def]
To reset the following totals or reload factory defaults, press enter to access
the function and use the escape (ESC) key to confirm. Press enter to accept the
change and exit the function. Press escape to exit the program mode. Table of
Contents
This function resets the total forward (+) totalizer.
This function resets the partial forward (+) totalizer.
Reset Total Reverse (-) Totalizer [T - reset]
Reset Partial Reverse (-) Totalizer [P - reset]
Load Factory Default Sensor Body Settings [Load Sens.f.def]
Load Factory Default Transmitter Settings [Load Conv.f.def]
This function resets the total reverse (-) totalizer.
This function resets the partial reverse (-) totalizer.
This function loads the factory default operating parameters for the sensor body.
Note: Reloading these values will void the factory calibration.
This function loads the factory default operating parameters for the sensor body.
ONICON Incorporated 727.447.6140 Page 34 onicon.com
FT-3100 TRANSMITTER PROGRAMMING GUIDE
3.10 MENU 12 - DIAGNOSTIC
Self Diagnostic Test [Self test]
Test Display [Display test]
Self Diagnostic Test [Self test]
Test Display [Display test]
Simulate Flow Rate [Flow sim. = OFF]
Firmware Information [Firmware info]
Serial Number [S/N = XXXXXX]
Working Time [WT = XXXX:XX:XX:XX]
This function interrupts normal operation and performs a complete test cycle on the measurement input circuits and on the coil excitation circuitry.
At the end of test the transmitter will revert to normal operation. This function is automatically performed whenever power is cycled. This function also restarts the transmitter.
Press enter to activate this function., Press escape when: “CONFIRM EXEC.?” appears on the display to start the test. Press any other key to exit without performing the test.
This function performs a pixel test on the display.
Press enter to perform the test.
Simulate Flow Rate [Flow sim. = OFF]
Firmware Information [Firmware info]
Serial Number [S/N = XXXXXX]
This function activates an internal signal that simulates a ow rate. This allows the outputs to be tested. The simulated ow rate is set as a percentage of the full scale ow rate.
When enabled, the following ashing icon is displayed on the upper left side of
the display.
To enable simulation, press enter to access the function and use the up arrow key to turn the function on. Press enter to accept the change and exit the function
The ow rate simulation percentage can be set by pressing the enter key from any ow display page. Use the right arrow to move the cursor and the up arrow key to change the value. Press enter to accept the ow rate percentage and
return to the display
Press the enter key followed by the escape key to end the simulation.
Displays a read only record of the current rmware and the date it was loaded.
Display a read only record of the electronics serial number
Working Time
Display of the accumulated operating time for the meter.
[WT = XXXX:XX:XX:XX]
ONICON Incorporated 727.447.6140 Page 35 onicon.com
Flow and Energy Measurement
FT-3100 TRANSMITTER PROGRAMMING GUIDE
For questions regarding this manual, please contact the ONICON Technical Support
Department at 727-447-6140.
DOC-0000680
ONICON
ONICON Incorporated 11451 Belcher Road South, Largo, FL 33773 USA
727-447-6140 • www.onicon.com • sales@onicon.com
April 2019
Loading...