OneTemp Calsys 1200 User Manual

Dry Block Calibrators
Wide Temperature Range
CALsys 1200 offer a wide temperature range from 250 °C to 1200 °C
Lightweight, portable
The CALsys 1200 block is ideal for Industrial/ Laboratory field use. It only weighs about 8 kg, and it is small enough to carry around.
Speed
The CALsys 1200 extremely quick to reach various temperatures, i.e. it cools down to 250°C in 150 minutes and heats up room temp to +1200 °C in 60 minutes.
Thermal & Cable Solutions
Calsys 1200
Portable, Lightweight, highly accurate temperature Calibrator for Industrial/ Laboratory field use
Accuracy and performance
The CALsys 1200 is an easily portable unit that also provides excellent calibration accuracy with stability ±0.3°C at 1200 °C.
Accredited calibration
Each CALsys 1200 is delivered with an accredited calibration certificate.
Computer Interface
The communication port(RS-232/ USB) enables communication with selected CALsys 1200 calibrators for automation calibration a n d d o c u m en ta t io n t hu s i t m a d e documentation easy.
CALsys 1200 offers easy to use portable temperature calibrator with high temperature range from 250 to 1200°C. It is a highly stable standard furnace for calibrating thermocouples / RTD. This calibrator can be used on site for high temperature calibration and also find application in glass, electric power, automotive and material process industry. The comparison volume is a metallic block of special material, which has a diameter of 37mm and 215mm long. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment. Our newly designed CALsys 1200 model offers better esthetic design and performance wise upgraded to
next level.
With the Tempsens make Compact Temperature Calibrator, you have chosen an extremely effective instrument which we hope will live up to all your expectations. This is a fast, timesaving, and reliable true industrial temperature calibrator designed for on-site use.
During the past several years, we have acquired extensive knowledge of industrial temperature calibration. This expertise is reflected in our products which are all designed for daily use in an industrial environment.
Temperature range
250 °C to 1200 °C
Accuracy
±2 °C
Stability
±0.01°C at 250°C
±0.02°C at 700°C
±0.03°C at 1200°C
Radial uniformity
±0.20°C at 250°C
±0.24°C at 700°C
±0.36°C at 1200°C
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
0.04°C Hysteresis
0.02 °C
Immersion depth
160 mm
Insert OD dimensions
37 mm
Method of Control
Self tunned PID controller
Heating time
60 Min
Cooling time
150 Min ( 1200 °C to 250 °C)
Resolution
0.1 °C up to 999°C
Display
LCD, °C or °F user-selectable
Size (H x W x D)
405(H) x 205(W) x 285(D) mm
Weight
13Kg
Power requirements
230 VAC, 1.5 KW(50 Hz)
Computer interface
RS - 232
Calibration
Accredited calibration certificate provided
Environmental operating conditions
0 °C to 40 °C, 0 % to 90 % RH (non-condensing)
Specifications valid in environmental conditions
13 °C … 33 °C
SPECIFICATIONS
Graphical Representation
0.45
0.40
0.35
0.30
0.25
0.20
0.15
Temperature in °C
0.10
0.05
CALsys 1200
0
250 700 1200
Temperature in °C
STABILITY
UNIFORMITY
Loading...